Search Results
Found 3 results
510(k) Data Aggregation
(60 days)
Bonebridge Osteosynthesis Plating System (SALGINA 2.5mm Distal Radius System, CASCELLA 3.5mm Superior
The SALGINA 2.5mm Distal Radius System is indicated for
- · Intra-articular fractures of the Distal Radius.
- · Extra-articular fractures of the Distal Radius
The CASCELLA 3.5mm Superior Clavicle System is indicated for:
- fixation of fractures of the clavicle shaft
- fixation of fractures of the lateral clavicle
- malunions of the clavicle
- · non-unions of the clavicle
The LEPORELLO 3.5mm Olecranon System is indicated for
- · fractures
- osteotomies
- nonunions
of the olecranon and proximal ulna.
The Bonebridge Osteosynthesis Plating System is intended for treating fractures of various bones. It consists of plates, locking and non-locking screws for fixation and corresponding instruments. The plating system is further subdivided into variants/types based on the anatomical location of the fracture.
Plates and screws are made of stainless steel (ISO 5832-1 or ASTM F138 or ASTM F139) or pure titanium (ASTM F67 or ISO 5832-2). All materials used are biocompatible, corrosion-resistant and nontoxic in a biological environment. Surgical instruments are made of stainless steel (ASTM F899. ISO 7153-1. ISO 5832-1. and ASTM F138/139). medical grade PEEK, medical grade EPDM terpolymer, or medical grade silicone. All plates are sterilized with gamma irradiation and delivered sterile. Screws and instruments are delivered non-sterile. Devices supplied in a non-sterile condition must be cleaned and steam sterilized prior to surgical use. Non-clinical testing has demonstrated the devices are MR Conditional.
The provided text describes a 510(k) premarket notification for the Bonebridge Osteosynthesis Plating System. The focus of the submission is on demonstrating substantial equivalence to predicate devices through non-clinical testing. This document does NOT describe a study that proves a device meets acceptance criteria for an AI/software-as-a-medical-device (SaMD) product, nor does it provide information relevant to a multi-reader, multi-case (MRMC) study or the establishment of ground truth by expert consensus for an AI model.
The information provided pertains to the mechanical, material, and labeling aspects of a physical bone fixation system, not a diagnostic AI device. Therefore, I cannot extract the requested information regarding AI device performance, sample sizes for test/training sets, expert qualifications, ground truth establishment, or MRMC study details from this document.
The document only states that:
- Clinical investigations were not required due to a clinical evaluation including a literature review and verification/validation activities.
- Usability evaluation was successfully completed in accordance with IEC 62366-1, with no significant usability issues identified.
The non-clinical tests summarized are related to the physical properties and safety of the implantable device:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't present acceptance criteria in a table format with specific quantitative metrics for the "acceptance criteria" and "reported device performance" in the context of an AI device. Instead, it lists various non-clinical tests performed on the physical implant and states whether they "Pass" or "Fail" against their respective standards. These are not performance metrics for an AI algorithm.
Non-Clinical Test | Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|---|
Sterilization Validation | Minimal dose of 25kGy validated using VDmax25 method (ISO 11137-2) for SAL of 10-6. Cleaning and sterilization procedures validated (ISO 17664, ISO 17665-1) at 132°C for 4 minutes with 20 min drying time. | Pass |
Packaging Validation | Validation of sterile packaging in accordance with ISO 11607 1/2 and ASTM F1980. Transport simulation (ISTA 2A) followed by Dye-Penetration (ASTM F1929), Visual inspection (ASTM F1886/1886M), Seal strength (ASTM F88/F88M), and Microbial barrier testing (DIN 58953-6, Chapter 2.14). | Pass |
Biocompatibility | Biological Evaluation and toxicological risk assessment for biological safety according to ISO 10993-series and FDA guidance. | Pass |
Mechanical Testing | Static and dynamic comparative testing demonstrating substantial equivalence of each subject plate variant to corresponding predicate devices with respect to maximum force (static) and maximum force for a given number of cycles (dynamic). Predefined acceptance criteria for mechanical performance successfully met. Design verification successfully completed for compatibility, anatomical shape, and appearance. | Pass |
MRI Safety | Assessment of displacement force and torque effects in the main static magnetic field at 3Tesla (ASTM F2052-21, ASTM F2213-17). Assessment of heating effects due to RF-field during MR scans at 1.5Tesla and 3Tesla (ASTM F2182-19). Assessment of image artifacts at 3Tesla (ASTM F3110-24). The tested implant and associated product family can be claimed as MR conditional. | Pass |
2. Sample size used for the test set and data provenance:
Not applicable. This is for a physical device, and no AI test set is mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. Ground truth for an AI algorithm is not established in this document.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study was done or mentioned, as this is not an AI device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable.
8. The sample size for the training set:
Not applicable.
9. How the ground truth for the training set was established:
Not applicable.
In summary, the provided FDA 510(k) clearance letter and associated summary pertain to the substantial equivalence of a physical bone fixation system with existing predicate devices based on non-clinical engineering and material properties testing. It does not contain any information about acceptance criteria or studies related to AI or software-as-a-medical-device.
Ask a specific question about this device
(79 days)
Bonebridge Osteosynthesis Plating System
The Bonebridge POYA 3.5mm Posteromedial Tibia System is intended for adult patients and indicated for internal fixation of posteromedial proximal tibia fractures, including buttressing of fractures of the proximal and metaphyseal areas of the tibia. Additionally, the longer plates are indicated for fractures extending to the tibial shaft.
The POYA 3.5mm Medial Proximal Tibia System is intended for adult patients and indicated to buttress metaphyseal fractures of the medial tibia plateau, split- type fractures of the medial tibia plateau, medial split fractures with associated depressions and split or depression fractures of the medial tibia plateau. Also, for use in the fixation of osteopenic bone and fixation of nonunions and malunions of the medial proximal tibia.
The SALTI 2.5/3.5mm Ankle System is intended for fixation of the ankle in adult patients and particularly in osteopenic bone. Specifically,
- the Medial and Anterolateral Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia.
- . The Posterior Distal Tibia Plate is intended to buttress partial articular fractures and bone fragments of the distal tibia.
- . The Lateral Distal Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.
The Bonebridge Osteosynthesis Plating System is intended for treating fractures of various bones. It consists of plates, locking and non-locking screws for fixation and corresponding instruments. The plating system is further subdivided into variants/types based on the anatomical location of the fracture.
Plates and screws are made of stainless steel (ISO 5832-1 or ASTM F138 or ASTM F139) or pure titanium (ASTM F67 or ISO 5832-2). All materials used are biocompatible, corrosion-resistant and nontoxic in a biological environment. Surgical instruments are made of stainless steel (ASTM F899, ISO 7153-1, ISO 5832-1, and ASTM F138/139), medical grade PEEK, medical grade EPDM terpolymer, or medical grade silicone, All plates are sterilized with gamma irradiation and delivered sterile. Screws and instruments are delivered non-sterile. Devices supplied in a non-sterile condition must be cleaned and steam sterilized prior to surgical use. Non-clinical testing has demonstrated the devices are MR Conditional.
The provided document is a 510(k) premarket notification letter from the FDA for a medical device called the "Bonebridge Osteosynthesis Plating System." This type of document is centered on demonstrating substantial equivalence to existing legally marketed devices, rather than establishing acceptance criteria and proving performance through novel clinical studies in the way one might expect for a new, innovative diagnostic or AI device.
Therefore, the information typically requested in your prompt regarding acceptance criteria and performance against those criteria (such as sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) is not present in this document.
The document primarily focuses on non-clinical tests to show that the new device is as safe and effective as its predicates. Here's a breakdown of the available information in relation to your request:
1. Table of acceptance criteria and the reported device performance:
The document states that "All predefined acceptance criteria were successfully met" for mechanical testing, but it does not explicitly list these acceptance criteria or the specific numerical performance values for each. It only reports a "Pass" for each category of non-clinical testing.
Test Category | Acceptance Criteria (Explicitly stated in document?) | Reported Device Performance |
---|---|---|
Sterilization validation | Minimal dose of 25kGy (for gamma irradiation) | Pass |
SAL of 10^-6 (for gamma irradiation) | Pass | |
132°C (270F) for 4 minutes and 20 min drying time (for steam sterilization) | Pass | |
Packaging validation | N/A (referenced standards ISO 11607 1/2, ASTM F1980, ISTA 2A, ASTM F1929, ASTM F1886/1886M, ASTM F88/F88M, DIN 58953-6 Chapter 2.14) | Pass |
Biocompatibility | N/A (referenced standards ISO 10993-series and FDA guidance) | Pass |
Mechanical testing | Substantially equivalent to predicate devices for maximum force (static) and maximum force for a given number of cycles (dynamic). | Pass |
MRI safety | "MR conditional" considering local SAR | Pass |
Assessment of displacement force and torque effects (ASTM F2052-21, ASTM F2213-17) | Pass | |
Assessment of heating effects (ASTM F2182-19) | Pass | |
Assessment of image artifacts (ASTM F2119-07 (2013)) | Pass | |
Usability tests | No significant usability issues (based on primary objectives prior Application/ Usability Risk Assessment update) | Successful |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
This information is not provided in the document. The document describes mechanical testing and refers to "design verification" which would involve testing prototypes, but does not detail the sample sizes for these tests, nor does it refer to patient data for performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
This information is not applicable/not provided. The "ground truth" concept is highly relevant for diagnostic devices or AI, where human experts (e.g., radiologists, pathologists) determine the correct diagnosis or outcome. For a bone plating system, performance is primarily assessed through biomechanical testing and material properties, not through expert-labeled patient data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not applicable/not provided for the reasons stated above. Adjudication relates to resolving disagreements among experts when establishing ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable/not provided. MRMC studies are typically for evaluating the impact of AI or diagnostic tools on human reader performance, which doesn't apply to a bone plating system.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This is not applicable/not provided. This device is a physical bone plating system, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
As discussed, the concept of "ground truth" in the diagnostic/AI sense is not applicable here. The "truth" for this device's performance is derived from standardized engineering and material testing (e.g., mechanical load tests, biocompatibility assessments, sterilization efficacy).
8. The sample size for the training set:
This information is not applicable/not provided. There is no "training set" for this type of physical medical device. Training sets are relevant for machine learning algorithms.
9. How the ground truth for the training set was established:
This information is not applicable/not provided.
In summary: The document is a 510(k) submission for a bone plating system and focuses on demonstrating substantial equivalence through non-clinical testing. It does not contain information related to acceptance criteria or studies in the context of diagnostic or AI device performance as described in your prompt. The "acceptance criteria" where mentioned implicitly refer to passing engineering specifications and regulatory standards.
Ask a specific question about this device
(266 days)
Bonebridge Osteosynthesis Plating System
The SALGINA® 2.5mm Volar Distal Radius System is indicated for
- · Intra-articular fractures of the Distal Radius
- · Extra-articular fractures of the Distal Radius
The CASCELLA® 3.5mm Superior Clavicle System is indicated for:
- fixation of fractures of the clavicle shaft
- · fixation of fractures of the lateral clavicle
- malunions of the clavicle
- · non-unions of the clavicle
The TAMINA® and TAMINA®TF Proximal Humerus System is indicated for:
· Dislocated two-, three-, and four fractures of the proximal humerus including fractures involving osteopenic bone.
- · Pseudarthroses (non-unions) of the proximal humerus
- · Osteotomies of the proximal humerus
The POYA® 3.5mm Lateral Proximal Tibia System is indicated for the internal fixation of fractures of the proximal tibia in adults and skeletally mature adolescents including:
- simple fractures
- · comminuted fractures
- · lateral wedge, medial wedge as well as bicondylar combination of lateral and medial wedge fractures
- · depression fractures
- · periprosthetic fractures
- · nonunions, malunions, tibial osteotomies and osteopenic bone
The POYA® 3.5mm Posteromedial Tibia System is indicated for internal fixation of posteromedial proximal tibia fractures including buttressing of fractures of the proximal areas of the tibia.
The DALVAZZA® 2.5mm Distal Ulna System is indicated for
- · fixation of fractures
- nonunions
of the distal ulna.
The LEPORELLO® 3.5mm Olecranon System is indicated for
- fractures
- · osteotomies
- non-unions
of the olecranon and proximal ulna.
The Bonebridge Osteosynthesis Plating System is intended for treating fractures of various bones. It consists of plates, locking and non-locking screws for fixation and corresponding instruments. The plating system is further subdivided into variants/types based on the anatomical location of the fracture.
Plates and screws are made of stainless steel (ISO 5832-1 or ASTM F138 or ASTM F139) or pure titanium (ASTM F67 or ISO 5832-2). All materials used are biocompatible, corrosion-resistant and nontoxic in a biological environment. Surgical instruments are made of stainless steel (ISO 5832-1 or ASTM F138 or ASTM F899), medical grade PEEK, medical grade EPDM terpolymer, or medical grade silicone.
All plates are sterilized with gamma irradiation and delivered sterile. Screws and instruments are delivered non-sterile. Devices supplied in a non-sterile condition must be cleaned and steam sterilized prior to surqical use. Non-clinical testing has demonstrated the devices are MR Conditional.
The provided text describes a 510(k) premarket notification for the Bonebridge Osteosynthesis Plating System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving independent effectiveness through a clinical study with detailed acceptance criteria for diagnostic performance. Therefore, the information typically requested regarding acceptance criteria for diagnostic accuracy, sample sizes for test/training sets, expert ground truth establishment, or multi-reader multi-case studies is not directly applicable or available in this document.
However, the document does list various performance tests conducted to establish substantial equivalence. I will present the information related to device performance and the testing conducted, framed within the principles of acceptance criteria where possible.
Here's a breakdown of the requested information based on the provided text, acknowledging the limitations inherent in a 510(k) summary:
1. Table of acceptance criteria and reported device performance:
The document doesn't explicitly state "acceptance criteria" with numerical thresholds for performance like sensitivity or specificity, as it's not a diagnostic device. Instead, it describes various engineering and material tests. The acceptance criterion for these tests is generally that the device performs equivalently to or better than the predicate device, or meets relevant international standards.
Test Category | Acceptance Criteria (Implied from text) | Reported Device Performance |
---|---|---|
Sterilization Validation | Gamma irradiation: Achieve a Sterility Assurance Level (SAL) of 10^-6 as per ISO 11137-2 (VDmax25 method). |
Steam sterilization: Successfully validated according to ISO 17664 and ISO 17665-1 at 132°C (270F) for 4 minutes with 20 min drying time. | Gamma irradiation: The minimal dose of 25kGy is validated using VDmax25 method as described in ISO 11137-2 and confirmed a Sterility Assurance Level SAL of 10-6.
Steam sterilization: Cleaning and sterilization procedures have been successfully validated in accordance with ISO 17664 and ISO 17665-1 at 132°C (270F) for 4 minutes and 20 min drying time. |
| Packaging Validation | Successful validation according to ISO 11607 1/2 and ASTM F1980. Includes transport simulation (ISTA 2A) followed by specific packaging verification tests (Dye-Penetration, Visual Inspection, Seal Strength, Microbial Barrier Testing). | Validation of the sterile packaging has been successfully performed in accordance with ISO 11607 1/2 and ASTM F1980. Furthermore, a transport simulation was conducted according ISTA 2A followed by these packaging verification tests: Dye-Penetration, ASTM F1929; Visual inspection, ASTM F1886/1886M; Seal strength, ASTM F88/F88M; Microbial barrier testing, DIN 58953-6, Chapter 2.14. |
| Mechanical Testing | Plates: Predefined acceptance criteria for static and dynamic testing successfully met, including statistical analysis and comparative testing to predicate devices.
Screws: Successfully tested in accordance with ASTM F543 and includes comparative testing to predicate devices.
Design Verification: Compatibility of implants and instruments, and assessment of anatomical shape and appearance successfully completed. | Plates: Static and dynamic testing has been performed and included statistical analysis and comparative testing to the predicate devices. The predefined acceptance criteria were successfully met.
Screws: Tested successfully in accordance with ASTM F543: Standard Specification and Test Methods for Metallic Medical Bone Screws and includes comparative testing to predicate devices.
Design verification was successfully completed and included compatibility of implants and instruments as well as assessment of anatomical shape and appearance. |
| MRI Safety | MR conditional status based on tests for displacement force and torque, heating effects due to RF-field, and image artifacts. (Tests according to ASTM F2052-15, ASTM F2213-17, ASTM F2182-11a, ASTM F2119-07). | The Bonebridge Osteosythesis Plating System is MR conditional considering local SAR based on the following tests: Assessment of displacement force and torque effects in the main static magnetic field at 3Tesla. Additionally, the expected magnetic force in a stronger magnetic field gradient of 30T/m was extrapolated. (According to ASTM F2052-15 and ASTM F2213-17); Assessment of heating effects due to the RF-field during MR scans at 1.5Tesla and 3Tesla according to ASTM F2182-11a; Assessment of image artifacts at 3Tesla according to ASTM F2119-07 (2013).
| Usability | No significant usability issues are identified, meeting predefined acceptance criteria from the study. Study participants must be able to use the products safely and effectively. | Summative usability evaluation studies in accordance with IEC 62366-1 support that there are no significant usability issues due to the study acceptance criteria of the primary objectives prior Application/ Usability Risk Assessment update. Therefore, the summative usability evaluation studies of the Bonebridge Osteosythesis Plating System are considered as successful. The study participants were able to use the products safely and effectively. |
| Biocompatibility | Assessment performed in accordance with ISO 10993-1. | A biological assessment has been performed in accordance with ISO 10993-1. |
| Clinical Evaluation| Not required if existing interventions for relevant indications match surgical technique, device design, and material, and if verification and validation activities are conclusive. | Based on the results of the literature review and the results of verification and validation activities it has been concluded that clinical investigations are not required, since surgical technique, device design and material match established interventions for the relevant indications. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This information is not provided in the given text. The document refers to "testing" performed (e.g., static and dynamic testing for plates, testing for screws, packaging validation, MRI safety tests), but it does not specify the sample sizes of devices or materials used for these engineering tests. There's no "test set" in the context of diagnostic data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable/provided as this is an orthopedic implant (osteosynthesis plating system), not an AI/diagnostic device that would require expert-established ground truth for a test set. Usability studies were conducted with "study participants," but their qualifications or number for establishing ground truth are not detailed.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable/provided. As noted above, this device is not an AI/diagnostic device.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable/provided. This is an orthopedic implant, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable/provided. This is an orthopedic implant, not an algorithm/AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
This information is not applicable/provided in the context of diagnostic accuracy. For the engineering tests, the "ground truth" would be established by the specifications of the materials, the design requirements, and the standards (e.g., ISO, ASTM) against which the device was tested. For usability, the ground truth is the successful and safe operation by participants.
8. The sample size for the training set
This information is not applicable/provided. This device is an orthopedic implant and does not involve AI or a "training set" in the machine learning sense.
9. How the ground truth for the training set was established
This information is not applicable/provided. As above, it is not an AI device.
Ask a specific question about this device
Page 1 of 1