Search Filters

Search Results

Found 36 results

510(k) Data Aggregation

    K Number
    K251331
    Date Cleared
    2025-10-27

    (180 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate by using the arm cuff. The device can be used in medical facilities or at home, and only for indoor use. It is supplied for OTC use.

    Device Description

    The blood pressure monitor is a fully automatic, non-invasive upper arm measurement device using oscillometric methodology to measure systolic pressure, diastolic pressure and pulse rate. The device features an inflatable cuff that wraps around the arm, with a built-in pressure sensor and transducer that analyze arterial pulsations to determine blood pressure values. Measurement results are clearly displayed on the LCD screen.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.

    Device Description

    The Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa. The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low. The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients. The product is provided non-sterile, and not to be sterilized by the user prior to use.

    AI/ML Overview

    Acceptance Criteria and Study for Blood Pressure Monitor (RAK series)

    The provided document describes the acceptance criteria and the study that proves the Blood Pressure Monitor (RAK series) meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Blood Pressure Monitor are based on various international standards, primarily IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers, and ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. These standards define the accuracy requirements for blood pressure measurement devices.

    Acceptance Criteria (Standard Reference)Specific RequirementReported Device Performance
    Blood Pressure AccuracyMean error of differences (device vs. reference)Not over the limits of ISO 81060-2:2018
    (ISO 81060-2:2018)Standard deviation of differences (device vs. reference)Not over the limits of ISO 81060-2:2018
    Absolute difference for individual subject (within 5 mmHg)Implicitly met if mean error & SD are within limits
    Blood Pressure Measurement Range25-255 mmHg25-255 mmHg (verified by IEC 60601-1 and IEC 80601-2-30)
    Heart Rate Measurement Range40-199 bpm40-199 bpm
    Heart Rate Accuracy± 5%± 5%
    Electrical SafetyCompliance with IEC 60601-1: 2005+AMD1 (2012) +AMD2 (2020)Device complies
    EMCCompliance with IEC 60601-1-2: 2014/AMD1:2020Device complies
    BiocompatibilityCompliance with ISO 10993-1, ISO 10993-5, ISO 10993-10Device complies (Cytotoxicity, Skin Sensitization, Irritation)
    UsabilityCompliance with IEC 60601-1-6:2010+A1:2013+A2:2020Device complies
    Home Healthcare EnvironmentCompliance with IEC 60601-1-11:2015 +A1:2020Device complies

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 100 adult subjects were used for the clinical validation.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study. It also does not specify if the study was retrospective or prospective, but clinical validation studies for device clearance are typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document refers to a "clinical test" for validation in accordance with ISO 81060-2:2018. This standard typically requires concurrent or sequential measurements by a trained observer using a standard reference sphygmomanometer (e.g., mercury sphygmomanometer). However, the specific number of experts (e.g., clinicians, technicians) involved in establishing the ground truth measurements for the test set and their qualifications are not explicitly stated in the provided text.

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method used for the clinical test set. ISO 81060-2:2018 outlines specific protocols for obtaining reference blood pressure measurements, often involving multiple observers and an agreed-upon method for handling discrepancies, but the specific details for this study are not provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size of how much human readers improve with AI vs without AI assistance. The device in question is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The clinical validation, in accordance with ISO 81060-2:2018, evaluates the accuracy of the algorithm and device against a reference standard without human-in-the-loop assistance. The results state that "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018," indicating direct measurement of the device's accuracy.

    7. Type of Ground Truth Used

    The ground truth used for the clinical validation was established by concurrent or sequential measurements using a reference manometer (as per ISO 81060-2:2018 methodology). This is a gold standard for blood pressure validation studies, effectively functioning as an "expert consensus" on the true blood pressure readings, established by highly trained personnel following a strict protocol.

    8. Sample Size for the Training Set

    The document does not provide information regarding a "training set" or its sample size. This is typical for medical devices that are validated directly against established performance standards using a test set, rather than being developed using machine learning models that require distinct training and testing datasets. The device's underlying algorithm is likely based on established oscillometric principles and refined through engineering, not trained on a large dataset in the way an AI model would be.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, information on how its ground truth was established is not applicable or provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241351
    Date Cleared
    2024-08-21

    (100 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Blood Pressure Monitor is intended for use in measuring blood pressure and pulse rate in patients with arm circumferences from 16 to 36 cm (6.3 to 14.1 inch), 22 to 42cm (8.6 to 16.5 inch), 22 to 45cm (8.6 to 17.7 inch) or 40 to 52cm (15.7 to 20.5 inch).

    Cuff model AC1636-01, arm circumference range is 16-36cm (6.3 to 14.1 inch), which is intended for children older than 3 years old or adults without conditions of diabetes, pregnancy, or preeclampsia.

    Cuff model AC2245-021, arm circumference range is 2245cm (8.6 to 17.7 inch), which is intended for adult population or those who with conditions of diabetes, pregnancy, or pre-eclampsia. Cuff model AC2242-41 and cuff model AC4052-04, arm circumference range are 22-42cm (8.6 to 16.5 inch) and 4052cm (15.7 to 20.5 inch) respectively, which are intended for adults without conditions of diabetes, pregnancy, or pre-eclampsia.

    It is intended indoor use only.

    Device Description

    The Blood Pressure Monitor is designed to measure the systolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".

    The main components of the Blood Pressure Monitor is the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 160mm and 520 mm, includes the inflatable bladder and polyester shell. The device consists of the microprocessor, the pressure sensor, the operation keys, the electromagnetic deflation control valve and the LCD.

    The devices embed a Cellular Wireless network connections module that allows it to connect to receiving end. Once measurement is over, the LCD of device displays results, and the device will start to send out data such as systolic, diastolic, pulse rate, date and time by Wireless method and protocol.

    AI/ML Overview

    Here is a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ANSI/AAMI/ISO 81060-2:2019 and ISO 81060-2:2018/Amd.1:2020)Reported Device Performance (TMB-2092-G with new cuffs)
    Criterion 1: Mean difference between reference BP and device readings for SBP-0.02 ± 2.12 mmHg (AC2242-41 study)
    -0.56 ± 1.58 mmHg (AC4052-04 study)
    Criterion 1: Mean difference between reference BP and device readings for DBP0.17 ± 1.97 mmHg (AC2242-41 study)
    -0.24 ± 1.63 mmHg (AC4052-04 study)
    Criterion 2: Mean difference between reference BP and device readings for SBP-0.02 ± 1.49 mmHg (AC2242-41 study)
    -0.56 ± 0.98 mmHg (AC4052-04 study)
    Criterion 2: Mean difference between reference BP and device readings for DBP0.17 ± 1.51 mmHg (AC2242-41 study)
    -0.24 ± 1.12 mmHg (AC4052-04 study)
    Overall conclusion for both new cuffs (AC2242-41 and AC4052-04)Fulfilled both validation criteria 1 and 2 of ANSI/AAMI/ISO 81060-2:2019 and ISO 81060-2:2018/Amd.1:2020.

    Note: The exact numerical acceptance criteria for ANSI/AAMI/ISO 81060-2:2019 are not explicitly stated in the provided text, but the text confirms that the device fulfilled these criteria. Typically, Criterion 1 relates to the mean difference and standard deviation of differences, and Criterion 2 relates to the standard deviation of differences for individual subjects.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 172 general adult subjects in total across two clinical studies.
      • First clinical study (AC2242-41): 86 subjects, with 258 datasets collected.
      • Second clinical study (AC4052-04): 86 subjects, with 258 datasets collected.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be prospective clinical validation studies for the accuracy of the blood pressure monitor with the new cuff sizes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. However, blood pressure monitor clinical validation (like ISO 81060-2) typically involves multiple trained observers (e.g., two or three) to take reference blood pressure measurements using a validated sphygmomanometer (often mercury or validated auscultatory method) to establish the ground truth. These observers are usually trained clinical professionals.

    4. Adjudication method for the test set

    The document does not explicitly state the adjudication method. In ISO 81060-2 studies, ground truth is typically established by multiple observers (as mentioned above) taking simultaneous or nearly simultaneous measurements, and a consensus or average of these readings serves as the reference and deviations between observers are often within predefined limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not mentioned. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the clinical studies described evaluate the performance of the device itself in measuring blood pressure and pulse rate, without direct human intervention in the measurement process (beyond applying the cuff and initiating the measurement). The results (-0.02 ± 2.12 mmHg, etc.) represent the standalone performance of the algorithm.

    7. The type of ground truth used

    The ground truth was established by reference blood pressure measurements. In the context of ANSI/AAMI/ISO 81060-2, this typically refers to auscultatory measurements performed by human observers using a standardized reference sphygmomanometer.

    8. The sample size for the training set

    The document does not provide information regarding a separate training set. The clinical studies mentioned are validation studies (test sets) to prove the accuracy of the device against established standards. For an oscillometric blood pressure monitor, the algorithm itself is developed based on physiological models and extensive internal testing, not typically a separate "training set" in the machine learning sense that would be reported in this manner for a 510(k) submission.

    9. How the ground truth for the training set was established

    As no training set is explicitly mentioned in the context of this 510(k) submission for an oscillometric BPM, how its ground truth was established is not provided. The "ground truth" discussed for these types of devices relates to the clinical accuracy validation against reference measurements, as detailed in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240832
    Date Cleared
    2024-06-07

    (73 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾ -12½´) or 22cm to 42cm (about 834"-161/2"). It is intended for adult indoor use only.

    Device Description

    Blood Pressure Monitor BBZ32-AA01 is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method". The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, two types of cuffs have been clinically validated to be matched with the device, suitable for adults with arm circumference from 22cm~42cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is PCB, thermistor, pressure pump, motor, release valve and pressure sensor. The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results. An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

    AI/ML Overview

    The Guangdong Transtek Medical Electronics Co., Ltd. Blood Pressure Monitor (BBZ32-AA01) underwent a clinical study to demonstrate its performance against established acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated against the criteria outlined in ISO 81060-2:2018/Amd.1:2020 for non-invasive sphygmomanometers. The document does not explicitly present these as a table of "acceptance criteria," but rather as two validation criteria (Criterion 1 and Criterion 2) that the device must fulfill. The reported device performance is indicated as having fulfilled both these criteria.

    CriterionAcceptance Criteria (Implicit from ISO 81060-2:2018/Amd.1:2020)Reported Device Performance (Mean Difference +/- Standard Deviation)
    Impulse Test (e.g., Blood Pressure Accuracy)(Criterion 1) Mean difference and standard deviation of differences between reference blood pressure and device readingsSystolic BP: 0.17 ± 2.57 mmHg; Diastolic BP: 0.79 ± 2.09 mmHg
    Impulse Test (e.g., Blood Pressure Accuracy)(Criterion 2) Mean difference and standard deviation of differences between reference blood pressure and device readingsSystolic BP: 0.17 ± 1.99 mmHg; Diastolic BP: 0.79 ± 1.68 mmHg

    Note: The exact numerical thresholds for acceptance for Criterion 1 and 2 from ISO 81060-2:2018/Amd.1:2020 are not explicitly stated in this document but are implied to have been met by the reported performance.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 86 qualified subjects.
    • Data Provenance: The document does not explicitly state the country of origin but mentions the manufacturer is "Guangdong Transtek Medical Electronics Co., Ltd." in China. It is a prospective clinical study, as indicated by "The clinical study was performed on the device..."

    3. Number of Experts and Qualifications

    The document does not provide information on the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with 10 years of experience). It only refers to "reference BPs" for comparison.

    4. Adjudication Method

    The document does not specify an adjudication method like 2+1 or 3+1 for establishing ground truth. The "reference BPs" suggest a gold standard measurement was used, but the process for ensuring its accuracy is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was done to assess human readers' improvement with or without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

    6. Standalone Performance

    Yes, a standalone performance study was done. The clinical study described evaluates the algorithmic (device) performance in measuring blood pressure and pulse rate without human-in-the-loop assistance for interpretation or adjudication of results.

    7. Type of Ground Truth Used

    The ground truth used was "reference BPs." This implies that the device readings were compared against highly accurate and established blood pressure measurements, likely obtained by a validated manual method (e.g., auscultatory method by trained professionals) or another highly accurate reference device.

    8. Sample Size for the Training Set

    The document does not mention a training set sample size. This suggests the blood pressure monitor is likely based on an established oscillometric algorithm rather than a machine learning model that requires a dedicated training set.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned (implying a rule-based or conventional algorithm rather than a machine learning model), the method for establishing ground truth for a training set is not applicable or detailed in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240254
    Date Cleared
    2024-05-31

    (121 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾"-12½") or 22cm to 42cm (about 8¾"-16½"). It is intended for adult indoor use only.

    Device Description

    Blood Pressure Monitor TMB-2296-BT is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method". The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HI-121H material. The two types of cuffs have been clinically validated to be matched with the device, suitable for adults with arm circumference from 22cm~42cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is medical switch power supply, PCB, thermistor, pressure pump, motor, release valve and pressure sensor. The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results. An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

    AI/ML Overview

    The provided text describes the 510(k) submission for a Blood Pressure Monitor (Model TMB-2296-BT) and its substantial equivalence to a predicate device. The information focuses on regulatory compliance, non-clinical, and clinical performance data, primarily related to the accuracy of blood pressure and pulse rate measurements.

    1. Table of Acceptance Criteria and Reported Device Performance

    The main acceptance criteria for blood pressure monitors is typically defined by standards like ISO 81060-2. The document specifically states that the device was tested according to ISO 81060-2:2018/Amd.1:2020 and "fulfilled both validation criteria 1 and 2".

    The ISO 81060-2 standard defines the following criteria:

    • Criterion 1: The mean difference between the device readings and reference blood pressure (BP) measurements should be within ±5 mmHg, with a standard deviation of 8 mmHg or less.
    • Criterion 2: For each subject, the difference between the device reading and the reference BP should be within 5 mmHg for at least 65% of the subjects, within 10 mmHg for at least 85% of the subjects, and within 15 mmHg for at least 95% of the subjects.

    Here's a table summarizing the reported device performance against these generally accepted criteria (as defined by ISO 81060-2):

    Acceptance Criteria (ISO 81060-2)Reported Device Performance (Study 1)Reported Device Performance (Study 2)
    Criterion 1 (Mean Difference & SD)
    Mean difference ≤ ±5 mmHgSBP: 0.21 mmHg, DBP: 0.66 mmHgSBP: -1.62 mmHg, DBP: 0.12 mmHg
    Standard deviation ≤ 8 mmHgSBP: 2.59 mmHg, DBP: 2.12 mmHgSBP: 2.80 mmHg, DBP: 3.01 mmHg
    Criterion 2 (Within specified range)
    Mean difference ≤ ±5 mmHgSBP: 2.07 mmHg, DBP: 1.76 mmHg (presumably mean difference for criterion 2 check)SBP: 2.35 mmHg, DBP: 2.60 mmHg (presumably mean difference for criterion 2 check)
    Specific percentages (e.g., 65% within 5mmHg)Explicit percentages not provided, but stated "fulfilled both validation criteria."Explicit percentages not provided, but stated "fulfilled both validation criteria."

    Note: The document states "fulfilled both validation criteria 1 and 2 of the ISO 81060-2:2018/Amd.1:2020," indicating that the device met the required statistical thresholds for accuracy. The provided mean differences and standard deviations are the direct results from the studies, demonstrating compliance with Criterion 1. While specific percentages for Criterion 2 are not listed, the statement of fulfillment confirms compliance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Study 1: 88 qualified subjects. From these, 262 datasets were collected.
      • Study 2: 87 qualified subjects. From these, 259 datasets were collected.
    • Data Provenance: Not explicitly stated regarding the country of origin. The studies are described as "clinical studies," implying a prospective design where the data was collected specifically for the validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide details on the number or qualifications of experts for establishing ground truth. For blood pressure monitor validation according to ISO 81060-2, ground truth is established by multiple independent reference measurements, typically performed by trained observers using a mercury sphygmomanometer or a validated auscultatory device. The standard requires specific training and quality control for these observers, but their specific "qualifications" (e.g., as radiologists) are not applicable or mentioned. The primary focus is on the accuracy of the reference measurements, not expert interpretation.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an "adjudication method" in the context of expert review, as ISO 81060-2 validation primarily relies on direct comparison of the device's readings against a set of rigorously collected reference measurements. Multiple readings are taken per subject, and statistical analyses (mean difference, standard deviation) are applied directly to these comparative measurements, rather than requiring expert adjudication of "results" in the typical sense of image interpretation. The standard outlines specific procedures for simultaneous measurements by multiple observers to minimize bias in reference readings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI/CAD systems that assist human readers in interpreting medical images (e.g., detecting lesions). For a blood pressure monitor, the study evaluates the device's direct measurement accuracy against a gold standard, not its comparative effectiveness with or without human assistance in interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the clinical validation described in the document is a standalone performance study. The device, a blood pressure monitor, measures blood pressure and pulse rate directly and automatically. The reported performance metrics (mean difference, standard deviation) are of the device itself, without human intervention in the measurement process (beyond proper cuff placement and initiation of the measurement).

    7. The Type of Ground Truth Used

    The ground truth used was reference blood pressure measurements obtained through a validated clinical method (implied to be auscultatory or an equivalent gold standard method, as per ISO 81060-2) against which the automated device's readings were compared. This is a form of "outcomes data" in the sense of comparing the device's output to a clinically accepted reference measurement for the physiological parameter it is designed to measure.

    8. The Sample Size for the Training Set

    The document does not provide information about a "training set" or its sample size. For a blood pressure monitor that uses an oscillometric method, the device's algorithm for determining BP values from oscillation waveforms is pre-programmed/calibrated. It's not typically a machine learning model that undergoes a 'training phase' on a distinct dataset in the way an AI image analysis algorithm would. The clinical studies described are for validation (test set performance) of the final device, not for training internal algorithms.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, there is no mention of a traditional "training set" in the context of an AI/ML algorithm that would require a separate ground truth establishment for training. The ground truth for the device's internal calibration/oscillometric algorithm (if applicable) would have been established during its development and calibration phases, likely using established physiological models and comparisons to known accurate pressure measurements, but these details are not part of this 510(k) submission. The clinical studies establish the accuracy of the final, already "trained" or calibrated, device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240102
    Date Cleared
    2024-04-29

    (104 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾˜-12½´´), 22cm to 42cm (about 8% -16½"), 22cm to 45cm (about 8¾ -17¾ ") or 40cm to 52cm (about 15¾ -20½ "). It is intended for adult indoor use only.

    Device Description

    Blood Pressure Monitor TMB-2287-B is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals.

    Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".

    The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, five types of cuffs have been clinically validated to be matched with the device, suitable for population aged at or over three and with arm circumference from 22cm~52cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is medical switch power supply, PCB, thermistor, pressure pump, motor, release valve and pressure sensor.

    The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ± 15% of the average, the irregular pulse symbol will be displayed along with measurement results.

    An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

    With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS, and the hardware function is belong to Device-MDDS, they are not subject to FDA laws and regulations applicable to devices.

    AI/ML Overview

    The provided text details the performance data for the Guangdong Transtek Medical Electronics Co., Ltd. Blood Pressure Monitor (TMB-2287-B), demonstrating its substantial equivalence to a predicate device. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Criterion)Reported Device Performance (TMB-2287-B)
    Blood Pressure Measurement Accuracy (ISO 81060-2:2018/Amd.1:2020, Criterion 1)Study 1 (87 subjects): Mean differences between reference and device readings:- Systolic BP: -0.77 ± 2.09 mmHg- Diastolic BP: -0.08 ± 1.69 mmHgStudy 2 (91 subjects): Mean differences between reference and device readings:- Systolic BP: 0.2 ± 2.96 mmHg- Diastolic BP: 0.76 ± 3.42 mmHg
    Blood Pressure Measurement Accuracy (ISO 81060-2:2018/Amd.1:2020, Criterion 2)Study 1 (87 subjects): Mean differences between reference and device readings:- Systolic BP: -0.77 ± 1.58 mmHg- Diastolic BP: -0.08 ± 1.17 mmHgStudy 2 (91 subjects): Mean differences between reference and device readings:- Systolic BP: 0.2 ± 2.46 mmHg- Diastolic BP: 0.76 ± 2.83 mmHg
    Biocompatibility (ISO 10993-1/5/10/23)All patient contact parts meet the requirements. Testing included: Cytotoxicity, Sensitization, Irritation.
    Electrical Safety (IEC 60601-1, IEC 60601-1-11)Compliance with IEC 60601-1 (General requirements for basic safety and essential performance) and IEC 60601-1-11 (Requirements for medical electrical equipment and systems used in the home healthcare environment).
    Electromagnetic Compatibility (EMC) (IEC 60601-1-2)Compliance with IEC 60601-1-2.
    Automated Non-Invasive Sphygmomanometers (IEC 80601-2-30)Compliance with this standard.
    Wireless (47 CFR Part 15, Subpart C & FDA Guidance "Radio Frequency Wireless Technology")Compliance with 47 CFR Part 15, Subpart C, and the FDA Guidance for Radio Frequency Wireless Technology in Medical Devices. (Note: The wireless software and hardware functions are classified as Non-device-MDDS and Device-MDDS, respectively, due to their limited role in data transfer/storage without controlling device functions, and are not subject to FDA device regulations).
    Blood Pressure Measurement Range0 mmHg ~ 299 mmHg, within ±3 mmHg (0.4 kPa) for 5°C - 40°C.
    Pulse Rate Measurement Range40-199 beats/minute, ±5%.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Study 1: 87 qualified subjects (258 datasets collected).
      • Study 2: 91 qualified subjects (272 datasets collected).
    • Data Provenance: The document does not explicitly state the country of origin or whether the studies were retrospective or prospective. However, clinical validation reports typically involve prospective data collection for accuracy studies. The submitter is based in Zhongshan, Guangdong, China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document states that the clinical data was collected according to ISO 81060-2:2018/Amd.1:2020 which involves "reference BPs." While this standard requires comparisons to a reference measurement (typically auscultatory measurements by trained observers), the exact number and specific qualifications of the experts (e.g., medical professionals making the reference measurements) are not explicitly stated in the provided text.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method for establishing the ground truth beyond stating that "reference BPs" were used in accordance with ISO 81060-2. This standard typically involves multiple observers taking reference measurements, and criteria are usually in place for discrepancies, but the specific method (e.g., 2+1, 3+1) is not detailed here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. The studies described are clinical validation studies comparing the device's readings against a reference method, not studies comparing human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, the device itself is an automated non-invasive sphygmomanometer. The clinical validation, comparing the device's measurements to reference blood pressure readings, represents a standalone performance evaluation of the algorithm and hardware without human interpretation or intervention in the measurement process itself, beyond initial setup.

    7. The Type of Ground Truth Used

    The ground truth for the blood pressure measurement accuracy studies was established using reference blood pressure measurements, as required by the ISO 81060-2 standard. This typically involves auscultatory measurements performed by trained medical professionals.

    8. The Sample Size for the Training Set

    The document does not provide information regarding a training set or its sample size. This is a medical device clearance, not a submission for an AI/ML algorithm that typically requires separate training data. The device's underlying oscillometric algorithm would have been developed and refined, but the specific "training set" for that development is not part of this regulatory submission details.

    9. How the Ground Truth for the Training Set was Established

    As no "training set" is disclosed in the context of this regulatory document, there is no information on how its ground truth might have been established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233146
    Date Cleared
    2024-03-19

    (174 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the of adult by using the arm cuff (22-42cm), it can be used in medical facilities or at home. It is supplied for OTC use.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a Blood Pressure Monitor. It does NOT contain information about an AI/ML medical device, nor does it describe an acceptance criteria study with the requested details.

    Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth, or training set information.

    The provided text focuses on the regulatory clearance process for a standard medical device (blood pressure monitor), detailing regulations, marketing provisions, and administrative contact information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233130
    Date Cleared
    2023-12-20

    (84 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is a digital monitor in measuring blood pressure and pulse rate with a wrist circumference ranging from 13.5cm to 21.5cm (about 51/3"-81/2"). It is intended for adult indoor use only.

    Device Description

    Blood Pressure Monitor TMB-2285-BT is designed to measure systolic pressure, diastolic pressure and pulse rate of adult population by a non-invasive technique, with an inflatable cuff wrapped around the wrist. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method". The main components of the Blood Pressure Monitor include main unit and wrist cuff. For the outer housing of the main unit, it's made of ABS material. The cuff model is WC1321-04, suitable for adults with wrist circumference of 13.5cm to 21.5cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is PCB, thermistor, pressure pump, motor, release valve and pressure sensor. The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results. An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Transtek Blood Pressure Monitor (K233130)

    Based on the provided FDA 510(k) summary for the Transtek Blood Pressure Monitor (K233130), the device's acceptance criteria and the study proving it meets these criteria are outlined below.

    Please note: The document primarily focuses on demonstrating substantial equivalence to a predicate device through adherence to recognized standards and clinical validation. Specific "acceptance criteria" in the sense of predefined thresholds for performance metrics derived from an AI/ML model are not explicitly detailed as this is a traditional medical device (blood pressure monitor) clearance, not an AI/ML software as a medical device (SaMD). The acceptance criteria are therefore inferred from compliance with the relevant industry standards, particularly ISO 81060-2 for clinical accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Criterion)Reported Device Performance (Result and Metric)
    Clinical Accuracy (ISO 81060-2:2018/Amd.1:2020)
    Criterion 1: Mean difference between device reading and reference for SBP/DBP.SBP: -0.27 ± 3.18 mmHg (Mean ± Standard Deviation)
    DBP: -0.80 ± 2.51 mmHg (Mean ± Standard Deviation)
    Criterion 2: Mean difference between device reading and reference for SBP/DBP.SBP: -0.27 ± 2.60 mmHg (Mean ± Standard Deviation)
    DBP: -0.80 ± 2.06 mmHg (Mean ± Standard Deviation)
    Biocompatibility (ISO 10993-1/5/10/23)All tests (Cytotoxicity, Sensitization, Irritation) passed.
    Electrical Safety (IEC 60601-1, IEC 60601-1-11)Compliance confirmed.
    Electromagnetic Compatibility (IEC 60601-1-2)Compliance confirmed.
    Performance - Non-Invasive Sphygmomanometers (IEC 80601-2-30)Compliance confirmed.
    Wireless (47 CFR Part 15, Subpart C & FDA Guidance)Compliance confirmed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 258 datasets collected from 86 subjects.
    • Data Provenance: Not explicitly stated, but given the company's location (Guangdong, China), it is highly probable the study was conducted retrospectively or prospectively within China. The document does not specify if the data was retrospective or prospective, but clinical validation studies are typically prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth. For blood pressure monitors, the "ground truth" (reference BPs) in clinical validation is typically established by trained technicians or clinicians using validated reference methods (e.g., auscultatory method with mercury manometer) according to the ISO 81060-2 standard.

    4. Adjudication Method for the Test Set

    The document does not mention an explicit adjudication method for the test set. Given the nature of blood pressure measurement validation according to ISO 81060-2, the comparison is typically against a reference measurement performed by trained personnel using a standardized method, not a consensus of multiple independent human readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not performed. This type of study is more common for diagnostic imaging AI algorithms where the AI assists human readers. For a standalone blood pressure monitor, the primary evaluation is its direct accuracy against a reference standard.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone performance study was done. The clinical data presented (mean differences and standard deviations) directly assesses the device's accuracy in measuring blood pressure and pulse rate without human-in-the-loop assistance for the measurement process itself, beyond the initial setup for obtaining the reference readings. The device is intended for over-the-counter use, implying standalone operation.

    7. Type of Ground Truth Used

    The ground truth used was reference blood pressure measurements obtained from subjects, likely using a validated auscultatory method as per the ISO 81060-2 standard (often involving two observers for accuracy). The document refers to "reference BPs."

    8. Sample Size for the Training Set

    The document does not provide information regarding a specific "training set" or its size. This device is a traditional non-invasive blood pressure monitor, not an AI/ML-based device that typically undergoes a separate model training phase. The "study" described is a clinical validation study for accuracy, which serves as the test set for the device's performance.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set in the context of an AI/ML algorithm for this traditional blood pressure monitor, this question is not applicable. The device's underlying "algorithm" (oscillometric method) is a well-established engineering principle, not a data-trained AI model requiring a separate ground-truth-established training set. The clinical validation proves the accuracy of this established method as implemented in the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232713
    Date Cleared
    2023-11-28

    (84 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 16cm to 36cm (about 6½"-14½"), 22cm to 32cm (about 8¾"-12½"), 22cm to 42cm (about 8¾"-16½") or 22cm to 45cm (about 8¾"-17¾").

    The cuff with arm circumference range of 1636cm is intended for children older than 6 years old or adults. The cuffs with arm circumference range of 2232cm or 2242cm or 2245cm, which are intended for adult population.

    It is intended indoor use only.

    Device Description

    Blood Pressure Monitor TMB-2266 is designed to measure systolic pressure, diastolic pressure and pulse rate of population at or over 3 years old by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".

    The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, five types of cuffs have been clinically validated to be matched with the device, suitable for population aged at or over three and with arm circumference from 16cm~45cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is medical witch power supply, PCB, thermistor, pressure pump, motor, release valve and pressure sensor.

    The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ± 25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results.

    An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

    AI/ML Overview

    The provided document describes the K232713 Blood Pressure Monitor, which is a digital monitor for measuring blood pressure and pulse rate. The information below is extracted from an FDA 510(k) summary concerning the device's substantial equivalence to predicate devices.

    1. Acceptance Criteria and Reported Device Performance

    The device's performance acceptance criteria are primarily derived from the ISO 81060-2 standard for non-invasive sphygmomanometers. The reported performance indicates that the device met these criteria.

    Acceptance Criteria (Standard)Reported Device Performance
    ISO 81060-2:2018+A1:2020: Mean error and standard deviation of differences in systolic and diastolic pressure not beyond specified limits.All data's mean error and standard deviation of differences in systolic and diastolic pressure is not beyond the limits set as per ISO 81060-2:2020. No adverse effect and/or complications were found in the study. This was confirmed in four clinical studies for different cuff models.
    IEC 60601-1: Basic safety and essential performanceCompliance demonstrated through testing.
    IEC 60601-1-2: Electromagnetic compatibilityCompliance demonstrated through testing.
    IEC 80601-2-30: Basic safety and essential performance of automated non-invasive sphygmomanometersCompliance demonstrated through testing.
    IEC 60601-1-11: Requirements for medical electrical equipment and systems used in home healthcareCompliance demonstrated through testing.
    Biocompatibility (ISO 10993-1/5/10/23): Cytotoxicity, Sensitization, IrritationAll patient contact parts meet the requirements of ISO 10993-1/5/10/23 (cytotoxicity, sensitization, and irritation). The device is considered surface contacting for a duration of exceeding 24 hours but not exceeding 30 days.
    Bluetooth Wireless (47 CFR Part 15, Subpart C 15.247): Compliance for wireless moduleThe wireless module was validated according to 47 CFR Part 15, Subpart C 15.247, demonstrating compliance.

    2. Sample Size and Data Provenance for Test Set (Clinical Studies)

    • Sample Sizes:
      • Study 1: 87 qualified subjects
      • Study 2: 86 qualified subjects
      • Study 3: 86 qualified subjects
      • Study 4: 102 qualified subjects (65 adults, 37 children aged 3-12)
    • Data Provenance: The document does not explicitly state the country of origin. The studies are described as "clinical studies," implying they were prospective.

    3. Number of Experts and Qualifications for Ground Truth Establishment

    The document does not provide details on the number of experts or their specific qualifications for establishing the ground truth measurements during the clinical studies. However, for blood pressure monitor validation studies conforming to ISO 81060-2, trained observers are typically used to obtain reference blood pressure measurements using a validated auscultatory method.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1). For ISO 81060-2 studies, it is common to have multiple observers simultaneously or sequentially measure blood pressure using a reference method, and their measurements (or an average) are used as the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No information is provided about a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human interpretation.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Yes, the clinical studies describe the standalone performance of the blood pressure monitor, which operates autonomously to measure blood pressure and pulse rate (algorithm only). The studies evaluated the device's accuracy against a recognized reference standard (implied by ISO 81060-2).

    7. Type of Ground Truth Used

    The ground truth for the clinical studies was established through comparative measurements against a reference method, which for ISO 81060-2 studies typically involves simultaneous or sequential auscultatory measurements performed by trained observers.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. This is typical for medical device submissions that focus on the clinical validation of the final product and do not disclose details about internal development or machine learning training data.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established, as details about the device's internal development and training are not part of this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232621
    Date Cleared
    2023-11-22

    (85 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Blood Pressure Monitor is intended for use in measuring blood pressure and pulse rate in patients with arm circumferences from 16 to 36 cm (6.3 to 14.1 inch) or 22 to 45cm (8.6 to 17.7 inch).

    Cuff model AC1636-01, arm circumference range is 16-36cm (6.3 to 14.1 inch), which is intended for children older than 3 years old or adults without conditions of diabetes, pregnancy, or pre-eclampsia.

    Cuff model AC2245-021, arm circumference range is 22-45cm (8.6 to 17.7 inch), which is intended for adult population or those who with conditions of diabetes, pregnancy, or pre-eclampsia.

    It is intended indoor use only.

    Device Description

    The Blood Pressure Monitor is designed to measure the systolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".

    The main components of the Blood Pressure Monitor is the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 160mm and 450 mm, includes the inflatable bladder and polyester shell. The device consists of the microprocessor, the pressure sensor, the operation keys, the electromagnetic deflation control valve and the LCD.

    The devices embed a Cellular Wireless network connections module that allows it to receiving end. Once measurement is over, the LCD of device displays results, and the device will start to send out data such as systolic, diastolic, pulse rate, date and time by Wireless method and protocol.

    AI/ML Overview

    The provided text describes the acceptance criteria and study proving the device meets those criteria for a Blood Pressure Monitor.

    Here's the breakdown of the information requested:

    Acceptance Criteria and Reported Device Performance

    The clinical performance acceptance criteria for this blood pressure monitor are based on the ISO 81060-2:2018/Amd.1:2020 standard, specifically Criterion 1 and Criterion 2. These criteria define the acceptable mean difference and standard deviation between the device readings and reference blood pressure measurements.

    Table of Acceptance Criteria and Reported Device Performance

    CriterionMeasurementAcceptance Criteria (Mean Difference ± Standard Deviation)Reported Performance (Mean Difference ± Standard Deviation)Study(s) this applies to
    Criterion 1SBP≤ ±5 mmHg
    DBP≤ ±5 mmHg
    Criterion 2SBP & DBPStandard deviation of differences ≤ 8 mmHg
    The percentage of differences within 5 mmHg: at least 65%
    The percentage of differences within 10 mmHg: at least 85%
    The percentage of differences within 15 mmHg: at least 95%

    Detailed Reported Performance from Clinical Studies:

    Clinical StudyCuff ModelSubjectsSBP Mean Difference ± SD (Criterion 1)DBP Mean Difference ± SD (Criterion 1)SBP Mean Difference ± SD (Criterion 2)DBP Mean Difference ± SD (Criterion 2)Fulfilled Criteria (ISO 81060-2:2018/Amd.1:2020)
    First Clinical Study (AC1636-01)AC1636-0136 children (aged 9.8±5.1) and 52 adults (aged 52.6±28.2)0.25 ± 3.02 mmHg0.96 ± 2.54 mmHg0.25 ± 2.55 mmHg0.96 ± 2.32 mmHgBoth Criterion 1 and 2 and
    Second Clinical Study (AC2245-021)AC2245-02186 subjects (aged 49.1±18.5)-1.1 ± 3.51 mmHg-0.2 ± 3.70 mmHg-1.1 ± 2.22 mmHg-0.2 ± 2.82 mmHgBoth Criterion 1 and 2
    Third Clinical Study (diabetes)AC2245-02190 diabetic patients-1.11 ± 3.50 mmHg-1.24 ± 3.21 mmHg-1.11 ± 3.19 mmHg-1.24 ± 3.01 mmHgBoth Criterion 1 and 2
    Fourth Clinical Study (pregnant)AC2245-02191 pregnant women (12-40 weeks pregnant: 38.5% normotensive, 30.8% hypertensive, 30.8% preeclampsia)-2.74 ± 5.61 mmHg0.13 ± 4.48 mmHg-2.74 ± 5.11 mmHg0.13 ± 4.02 mmHgCriterion 2 (Only Criterion 2 for pregnant as per standard)
    The text indicates that the device fulfilled both validation criteria 1 and 2 of the ISO 81060-2:2018/Amd.1:2020 for the first three studies and fulfilled validation criteria 2 for the pregnant patient study. This implies the reported mean differences and standard deviations were within the limits specified by the standard.

    Study Details:

    1. Sample sizes used for the test set and the data provenance:

      • Total Clinical Study Subjects: 138 general adult subjects (aged 3-12 years for children and general adults), 90 diabetic patients, and 91 pregnant women.
      • First Clinical Study (AC1636-01): 261 datasets collected from 88 subjects (36 children and 52 adults)
      • Second Clinical Study (AC2245-021): 258 datasets collected from 86 subjects (general adults)
      • Third Clinical Study (diabetes): 268 valid BP dataset pairs collected from 90 diabetic patients
      • Fourth Clinical Study (pregnant): 270 valid datasets collected from 91 pregnant subjects
      • Data Provenance: Not explicitly stated, but typically clinical studies for regulatory submissions are prospective. The location (country of origin) is not mentioned in the provided text, but the manufacturer is based in China.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document implies that the ground truth (reference blood pressure measurements) was established by standard clinical methods using a reference device or trained observers as required by ISO 81060-2:2018/Amd.1:2020. However, the exact number and qualifications of experts (e.g., radiologists, cardiologists) that would typically establish such ground truth in the context of blood pressure measurement (e.g., supervising the auscultatory method or reading a reference device) are not specified in the provided text. The standard usually requires trained observers for the reference measurements.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • The document does not explicitly state an adjudication method. For blood pressure measurement validation studies, this typically involves simultaneously taking measurements with the device under test and a calibrated reference device (or multiple trained observers using the auscultatory method). The ISO 81060-2 standard outlines the methodology for obtaining reference measurements, often involving multiple mercury sphygmomanometers and trained observers. No further details on adjudication are provided.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a study to validate a blood pressure monitor, not an AI-assisted diagnostic device. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable and was not performed.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes. The tests described are for the standalone performance of the blood pressure monitor, which uses an oscillometric algorithm to measure blood pressure without active human interpretation during the measurement process itself, beyond initiation and cuff placement. The performance data presented are for the device (algorithm) measuring blood pressure against a reference.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth was established by reference blood pressure measurements obtained from clinical studies, presumably following the methodology detailed in ISO 81060-2:2018/Amd.1:2020. This standard typically relies on simultaneously recording blood pressure using a validated reference method (e.g., auscultatory method by trained observers using mercury sphygmomanometers, or other validated reference devices) and comparing it to the device under test.
    7. The sample size for the training set:

      • The document describes clinical validation studies (test sets) for the device. It does not provide information on the sample size used for the training set for the device's underlying algorithm development. Blood pressure monitors often use long-established algorithms, but if machine learning was used, the training set details are not provided.
    8. How the ground truth for the training set was established:

      • Since details of a specific training set are not provided, information on how its ground truth was established is also not available in the provided text.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4