K Number
K240832

Validate with FDA (Live)

Date Cleared
2024-06-07

(73 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾ -12½´) or 22cm to 42cm (about 834"-161/2"). It is intended for adult indoor use only.

Device Description

Blood Pressure Monitor BBZ32-AA01 is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method". The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, two types of cuffs have been clinically validated to be matched with the device, suitable for adults with arm circumference from 22cm~42cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is PCB, thermistor, pressure pump, motor, release valve and pressure sensor. The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results. An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

AI/ML Overview

The Guangdong Transtek Medical Electronics Co., Ltd. Blood Pressure Monitor (BBZ32-AA01) underwent a clinical study to demonstrate its performance against established acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The device's performance was evaluated against the criteria outlined in ISO 81060-2:2018/Amd.1:2020 for non-invasive sphygmomanometers. The document does not explicitly present these as a table of "acceptance criteria," but rather as two validation criteria (Criterion 1 and Criterion 2) that the device must fulfill. The reported device performance is indicated as having fulfilled both these criteria.

CriterionAcceptance Criteria (Implicit from ISO 81060-2:2018/Amd.1:2020)Reported Device Performance (Mean Difference +/- Standard Deviation)
Impulse Test (e.g., Blood Pressure Accuracy)(Criterion 1) Mean difference and standard deviation of differences between reference blood pressure and device readingsSystolic BP: 0.17 ± 2.57 mmHg; Diastolic BP: 0.79 ± 2.09 mmHg
Impulse Test (e.g., Blood Pressure Accuracy)(Criterion 2) Mean difference and standard deviation of differences between reference blood pressure and device readingsSystolic BP: 0.17 ± 1.99 mmHg; Diastolic BP: 0.79 ± 1.68 mmHg

Note: The exact numerical thresholds for acceptance for Criterion 1 and 2 from ISO 81060-2:2018/Amd.1:2020 are not explicitly stated in this document but are implied to have been met by the reported performance.

2. Sample Size and Data Provenance

  • Test Set Sample Size: 86 qualified subjects.
  • Data Provenance: The document does not explicitly state the country of origin but mentions the manufacturer is "Guangdong Transtek Medical Electronics Co., Ltd." in China. It is a prospective clinical study, as indicated by "The clinical study was performed on the device..."

3. Number of Experts and Qualifications

The document does not provide information on the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with 10 years of experience). It only refers to "reference BPs" for comparison.

4. Adjudication Method

The document does not specify an adjudication method like 2+1 or 3+1 for establishing ground truth. The "reference BPs" suggest a gold standard measurement was used, but the process for ensuring its accuracy is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study was done to assess human readers' improvement with or without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

6. Standalone Performance

Yes, a standalone performance study was done. The clinical study described evaluates the algorithmic (device) performance in measuring blood pressure and pulse rate without human-in-the-loop assistance for interpretation or adjudication of results.

7. Type of Ground Truth Used

The ground truth used was "reference BPs." This implies that the device readings were compared against highly accurate and established blood pressure measurements, likely obtained by a validated manual method (e.g., auscultatory method by trained professionals) or another highly accurate reference device.

8. Sample Size for the Training Set

The document does not mention a training set sample size. This suggests the blood pressure monitor is likely based on an established oscillometric algorithm rather than a machine learning model that requires a dedicated training set.

9. How the Ground Truth for the Training Set Was Established

Since no training set is mentioned (implying a rule-based or conventional algorithm rather than a machine learning model), the method for establishing ground truth for a training set is not applicable or detailed in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 7, 2024

Guangdong Transtek Medical Electronics Co., Ltd. Jerry Fan RA Manager Zone A. No. 105. Dongli Road. Torch Development District Zhongshan, Guangdong 528437 China

Re: K240832

Trade/Device Name: Blood Pressure Monitor (BBZ32-AA01) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 26, 2024 Received: March 26, 2024

Dear Jerry Fan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240832

Device Name

Blood Pressure Monitor (BBZ32-AA01)

Indications for Use (Describe)

This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8%"-12½") or 22cm to 42cm (about 83/4"-161/2").

It is intended for adult indoor use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

{4}------------------------------------------------

510(k) Summary

Prepared in accordance with the requirement of 21 CFR Part 807.92

Prepared Date: 03/26/2024

  1. Submitter Guangdong Transtek Medical Electronics Co., Ltd. Name: Zone A, No. 105, Dongli Road, Torch Development District, 528437 Zhongshan, Address: Guangdong, China Contact: Jerry Fan Title: RA Manager E-mail: gt-rateam@transtekcorp.com TEL: +86 15728668528

2. Subject Device

Trade/Device NameBlood Pressure Monitor
ModelBBZ32-AA01
Common NameBlood Pressure Monitor
ClassificationClass II
Product CodeDXN
Submission TypeTraditional 510(k)

3. Predicate Device

Manufacturer:Omron Healthcare, Inc.
Device Name / Model:Noninvasive blood pressure measurement system /HEM-7311
510(k) Number:K133379
ClassificationClass II
Product CodeDXN

4. Device Description

Blood Pressure Monitor BBZ32-AA01 is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals.

Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".

The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, two types of cuffs have been clinically validated to be matched with the device, suitable for adults with arm circumference from 22cm~42cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is PCB, thermistor, pressure pump, motor, release valve

{5}------------------------------------------------

and pressure sensor.

The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results.

An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is intended to transmit clinical data from Point B only, and not involved in active patient monitoring. Therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS, and the hardware function is belong to Device-MDDS, they are not subject to FDA laws and regulations applicable to devices.

5. Indications for use

This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾ -12½´) or 22cm to 42cm (about 834"-161/2").

It is intended for adult indoor use only.

ItemSubject DevicePredicate DeviceNote
510(k)ApplyingK133379/
ModelBBZ32-AA01HEM-7311/
ApplicantGuangdong Transtek MedicalElectronics Co., Ltd.OMRON HEALTHCARE, INC.Same
Device NameBlood Pressure MonitorBlood Pressure MonitorSame
Product CodeDXNDXNSame
ClassificationClass IIClass IISame
Regulation #21 CFR 870.113021 CFR 870.1130Same
Type of UseOTCOTCSame
Intended Use /Indication forUseThis Blood Pressure Monitor is adigital monitor intended for use inmeasuring blood pressure and pulserate with arm circumference rangingfrom 22cm to 32cm (about 8¾"-12½")The device is a digital monitorintended for use in measuring bloodpressure and pulse rate in adult patientpopulation.The device detects the appearance ofSimilar, referto Note 1.
or 22cm to 42cm (about 8¾"-16½")It is intended for adult indoor use only.irregular heartbeats duringmeasurement and gives a warningsignal with readings.
Armcircumference22cm to 32cm (about 8¾"-12½") or22cm to 42cm (about 8¾"-16½")Small 17-22cm, Standard 22-32cm,Large 22-42cmSimilar, referto Note 2.
PatientPopulationAdultAdultSame
PrincipleOscillometric methodOscillometric methodSame
Anatomical SiteUpper ArmUpper ArmSame
Where used(hospital, home,ambulance, etc.)HomeHomeSame
Power Supply4*1.5V AAA batteries4*AA batteries or AC adapterSimilar, referto Note 3.
Human FactorsBlood pressureBlood pressureSame
MeasurementItemsMeasuring systolic and diastolic bloodpressure and pulse rate of intendedpopulation, including irregular pulserhythm detection.Measuring systolic and diastolic bloodpressure and pulse rate of intendedpopulation, including irregular pulserhythm detection.Same
Cuff DeflationAutomatic deflationAutomatic pressure release valveSame
Blood PressureMeasurement0mmHg ~ 299mmHg,5°C - 40°C within ±3mmHg (0.4kPa)0mmHg ~ 299mmHg, within±3mmHg or 2% of readingSame
Pulse RateMeasurement40-199 beat/minute, ±5%40-180 beat/minute, within ±5% ofreadingSimilar, referto Note 2.
DisplayLCDLCDSame
Memory2×992×100Similar, referto Note 3.
OperationEnvironmentTemperature: 5°C~40°C;Relative Humidity: 15%90% RH;Atmospheric: 700hPa1060hPaTemperature: 10°C~40°C;Relative Humidity: 15%~90% RH;Similar, referto Note 3.
Storage andtransportationEnvironmentTemperature: -20°C60°C;Relative humidity≤93%RH,non-condensing;Atmospheric: 500hPa1060hPaTemperature: -20°C~60°C;Relative humidity 10 to 95%RH,non-condensing;Similar, referto Note 3.
PerformanceCompliance with IEC 80601-2-30Compliance with IEC 80601-2-30Same
ClinicalCompliance with ISO 81060-2Compliance with ANSI/AAMI/ISO81060-2Same
BiocompatibilityAll patient contact parts meet theCompliance with ISO 10993-1Same
requirements of ISO 10993-1/5/10/23
Electrical SafetyCompliance with AAMI ANSI ES60601-1 and IEC 60601-1-11Compliance with AAMI ANSI ES60601-1 and IEC 60601-1-11Same
EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2Same
WirelessBluetoothNot publicly availableSimilar, referto Note 3.

6. Comparison to Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

Justification for difference:

Note 1:

Although there are some differences between the "Intended Use / Indication for Use" of the subject device and predicate device, it is only expressed differently, and the general meaning is the same, and they both use in measuring blood pressure and pulse rate. Thus, the difference here does not raise any issues concerning safety and effectiveness.

Note 2:

Although there are some differences between the "Arm circumference" and "Pulse Rate Measurement" of the subject device and the predicate device, both the subject device and predicate device can be applied to 22-32cm and 22-42cm arm circumference, the pulse rate range of the subject device is similar to that of the predicate device, and the subject device has also been validated according to IEC 80601-2-30 and ISO 81060-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Note 3:

Although there are some differences between the "Power Supply", "Memory", "Operation Environment", "Storage and transportation Environment" and "Wireless" of the subject device and the predicate device, both the subject device and predicate device comply with the requirement of standard AAMI ANSI ES 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 80601-2-30, and the subject device also complies with the guidance "Content of Premarket Submissions for Device Software Functions". As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Conclusion:

Based on the comparison and analysis in this submission, it can be concluded that: the subject device is substantially equivalent to the predicate device and reference device regarding safety and effectiveness.

7. Performance Data

The following performance data were provided in support of the substantial equivalence determination:

{8}------------------------------------------------

Biocompatibility testing:

The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing includes the following tests:

  • Cytotoxicity
  • Sensitization
  • Irritation

The subject device is considered as surface contacting for a duration not exceed 24hours.

Non-clinical data

The device has been tested according to following standards:

  • IEC 60601-1: Medical electrical equipment Part 1: General requirements for basic safety ● and essential performance
  • IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirement for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
  • IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.
  • IEC 60601-1-11: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • FDA Guidance for Non-Automated Sphygmomanometer

Wireless testing:

  • 47 CFR Part 15, Subpart B, Subpart C
  • Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013)

Clinical data

device was tested according to ISO 81060-2:2018/Amd.1:2020 Non-invasive The sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type AMENDMENT 1. The clinical study was performed on the device matching with cuffs ranging from 22-32cm and 22-42cm. The study population consisted of 86 qualified subjects. 258 datasets were collected from 86 subjects with aged 52±13.5. The mean differences between reference BPs and device readings were 0.17±2.57/0.79±2.09mmHg for systolic BP (SBP)/diastolic BP (DBP) of criterion 1, and 0.17 ± 1.99/0.79 ± 1.68mmHg for SBP/DBP of criterion 2. BBZ32-AA01 fulfilled both validation criteria 1 and 2 of the ISO 81060-2:2018/Amd.1:2020.

8. Conclusion

Based on the information presented in this 510(k) premarket notification submission, a conclusion can be drawn that the proposed subject device is considered substantially equivalent to the predicate device. The differences between the subject device and the predicate devices were

{9}------------------------------------------------

successfully tested with relevant standards and FDA guidance, and do not affect equivalent safety and effectiveness or raise new issues concerning safety and effectiveness.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).