K Number
K242528

Validate with FDA (Live)

Date Cleared
2025-01-10

(137 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.

Device Description

The Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa. The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low. The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients. The product is provided non-sterile, and not to be sterilized by the user prior to use.

AI/ML Overview

Acceptance Criteria and Study for Blood Pressure Monitor (RAK series)

The provided document describes the acceptance criteria and the study that proves the Blood Pressure Monitor (RAK series) meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Blood Pressure Monitor are based on various international standards, primarily IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers, and ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. These standards define the accuracy requirements for blood pressure measurement devices.

Acceptance Criteria (Standard Reference)Specific RequirementReported Device Performance
Blood Pressure AccuracyMean error of differences (device vs. reference)Not over the limits of ISO 81060-2:2018
(ISO 81060-2:2018)Standard deviation of differences (device vs. reference)Not over the limits of ISO 81060-2:2018
Absolute difference for individual subject (within 5 mmHg)Implicitly met if mean error & SD are within limits
Blood Pressure Measurement Range25-255 mmHg25-255 mmHg (verified by IEC 60601-1 and IEC 80601-2-30)
Heart Rate Measurement Range40-199 bpm40-199 bpm
Heart Rate Accuracy± 5%± 5%
Electrical SafetyCompliance with IEC 60601-1: 2005+AMD1 (2012) +AMD2 (2020)Device complies
EMCCompliance with IEC 60601-1-2: 2014/AMD1:2020Device complies
BiocompatibilityCompliance with ISO 10993-1, ISO 10993-5, ISO 10993-10Device complies (Cytotoxicity, Skin Sensitization, Irritation)
UsabilityCompliance with IEC 60601-1-6:2010+A1:2013+A2:2020Device complies
Home Healthcare EnvironmentCompliance with IEC 60601-1-11:2015 +A1:2020Device complies

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 100 adult subjects were used for the clinical validation.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study. It also does not specify if the study was retrospective or prospective, but clinical validation studies for device clearance are typically prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document refers to a "clinical test" for validation in accordance with ISO 81060-2:2018. This standard typically requires concurrent or sequential measurements by a trained observer using a standard reference sphygmomanometer (e.g., mercury sphygmomanometer). However, the specific number of experts (e.g., clinicians, technicians) involved in establishing the ground truth measurements for the test set and their qualifications are not explicitly stated in the provided text.

4. Adjudication Method for the Test Set

The document does not explicitly state the adjudication method used for the clinical test set. ISO 81060-2:2018 outlines specific protocols for obtaining reference blood pressure measurements, often involving multiple observers and an agreed-upon method for handling discrepancies, but the specific details for this study are not provided.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size of how much human readers improve with AI vs without AI assistance. The device in question is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

6. Standalone Performance Study

Yes, a standalone performance study was done. The clinical validation, in accordance with ISO 81060-2:2018, evaluates the accuracy of the algorithm and device against a reference standard without human-in-the-loop assistance. The results state that "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018," indicating direct measurement of the device's accuracy.

7. Type of Ground Truth Used

The ground truth used for the clinical validation was established by concurrent or sequential measurements using a reference manometer (as per ISO 81060-2:2018 methodology). This is a gold standard for blood pressure validation studies, effectively functioning as an "expert consensus" on the true blood pressure readings, established by highly trained personnel following a strict protocol.

8. Sample Size for the Training Set

The document does not provide information regarding a "training set" or its sample size. This is typical for medical devices that are validated directly against established performance standards using a test set, rather than being developed using machine learning models that require distinct training and testing datasets. The device's underlying algorithm is likely based on established oscillometric principles and refined through engineering, not trained on a large dataset in the way an AI model would be.

9. How the Ground Truth for the Training Set Was Established

As no training set is described, information on how its ground truth was established is not applicable or provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

January 10, 2025

Shenzhen Ruiankang Technology Co., Ltd. Xinhua Zhao General Manager Floor1 and 2. No.8. Zhugushi Chunyang Industrial Park Wulian Community, Longgang Street, Longgang District Shenzhen, Guangdong 518117 China

Re: K242528

Trade/Device Name: Blood Pressure Monitor (RAK262, RAK263, RAK266, RAK268, RAK269, RAK282, RAK283, RAK286, RAK288, RAK289, RAK291, RAK292, RAK293, RAK295, RAK296, RAK297, RAK298) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 12, 2024 Received: December 12, 2024

Dear Xinhua Zhao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Robert T. Kazmierski -S

for

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242528

Device Name

Blood Pressure Monitor (RAK260, RAK262, RAK263, RAK266, RAK268, RAK269, RAK282, RAK283, RAK286, RAK288, RAK289, RAK291, RAK292, RAK293, RAK295, RAK296, RAK297, RAK298)

Indications for Use (Describe)

The Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

{4}------------------------------------------------

510(k) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

1. Submitter:Shenzhen Ruiankang Technology Co.,LtdFloor1 and 2, No.8, Zhugushi Chunyang Industrial Park, Wulian Community,Longgang Street, Longgang District, Shenzhen City, Guangdong, China.TEL: (086)1590941123
Contact Person:Zhao Xinhua
Prepare date:2024-10-14
2. Device name andclassificationDevice Name: Blood Pressure Monitor
Models: RAK260, RAK262, RAK263, RAK266, RAK268, RAK269, RAK282, RAK283,RAK286, RAK288, RAK289, RAK291, RAK292, RAK293, RAK295, RAK296, RAK297,RAK298
Classification Name: 21 CFR 870.1130 Cardiovascular Diagnostic Devices
Product code: DXN
Regulatory Class: Class II
3. Reason forSubmissionSubmission for the 510(k) clearance.
4. PredicateDevice(s)Shenzhen AOJ Medical Technology Co., Ltd. , AOJ-30B
5. Device DescriptionThe Blood Pressure Monitor is designed as a battery driven automatic non-invasiveblood pressure monitor. It can automatically complete the inflation, deflation andmeasurement, which can measure systolic and diastolic blood pressure as well asthe pulse rate of adult person at upper arm within its claimed range and accuracyvia the oscillometric technique. The result will be displayed in the international unitmmHg or Kpa.The device has the data storage function in order for data reviewing, including thesystolic pressure, diastolic pressure, pulse rate and measurement time. The devicealso has low voltage indication, which will be triggered when the battery is low.The proposed device is intended to be used in medical facilities or at home. Andthe effectiveness of this sphygmomanometer has not been established in pregnant(including pre-eclamptic) patients.The product is provided non-sterile, and not to be sterilized by the user prior to use.
6. Indications for UseThe Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive

oscillometric technique at medical facilities or at home.

{5}------------------------------------------------

7. Predicate Device Comparison

Please refer to following table to find differences between the subject device and predicate device.

ITEMProposed DevicePredicate DeviceComparisonResult
ManufactureShenzhen RuiankangTechnology Co.,Ltd.Shenzhen AOJ MedicalTechnology Co., Ltd.Same
Indications for UseThe Blood Pressure Monitor isintended to measure thesystolic pressure and diastolicpressure, as well as the pulserate of adult person vianon-invasive oscillometrictechnique at medical facilitiesor at home.The Electronic Blood PressureMonitor is intended to measurethe systolic and diastolic bloodpressure as well as the pulserate of adult person vianon-invasive oscillometrictechnique in which an inflatableCUFF is wrapped around theupper arm at medical facilitiesor at home.Different¹
ContraindicationsNot KnownNot KnownSame
ApplicationscenarioMedical Facilities and HomeUseMedical Facilities and HomeUseSame
Operational Specifications
PrincipleOscillometricOscillometricSame
Measurement ItemSYS, DYS, Pulse RateSYS, DYS, Pulse RateSame
Intended patientpopulationAdultAdultSame
Intendedapplication siteUpper armUpper armSame
Blood pressuremeasurementrange25~255 mmHg30 - 255 mmHgDifferent²
Accuracy$\pm$ 3 mmHg$\pm$ 3 mmHgSame
Heart ratemeasurementrange40-199 bpm40-199 bpmSame
Accuracy$\pm$ 5%$\pm$ 5%Same
Cuff circumference22~32 cm22~36 cmDifferent³
ComponentsLCD/Key/Cuff/MCU/Pump/BatteriesLCD/Key/Cuff/MCU/Pump/BatteriesSame
Auto shutdownYESYESSame
OperatingenvironmentTemperature: 5°C~ 40°CHumidity: 15%-80% RH,Atmospheric pressure:70 kPa - 106 kPaTemperature: 5°C~ 40°CHumidity: 15%-90% RH,Atmospheric pressure:70 kPa - 106 kPaSame
StorageenvironmentAmbient Temperature:20°C to 55°CAmbient Temperature:20°C to 55°CSame
Table 1 Comparison between the predicate and the subject device

{6}------------------------------------------------

Relative Humidity: 10-93% RH,Atmospheric pressure:70 kPa - 106 kPaRelative Humidity: 10-93% RH,Atmospheric pressure:70 kPa - 106 kPa
Power supplyUSB DV 5V or 6V DC (4 *1.5V AAA batteries)6V DC (4 *1.5 V AAA batteries)Different4
Memory2*992*99Same
Compliance Standards
Bio-compatibilityISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-1ISO 10993-5ISO 10993-10Same
Electrical SafetyIEC 60601-1IEC 60601-1-11IEC 60601-1IEC 60601-1-11
EMCIEC 60601-1-2IEC 60601-1-2
PerformanceIEC 80601-2-30IEC 80601-2-30

Justification for the differences:

    1. Different Indications for Use
      The subject device shares a similar intended use with the predicate device, and the substantial difference between the subject device and the predicate device regarding the intended use lies in the arm circumference range. However, the subject device with matched cuffs has been validated according to IEC 80601-2-30. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effective.
    1. Different Blood pressure measurement range
      The Blood pressure measurement range of the subject device is different form the predicate device, but the measurement range of the subject device has been verified by IEC 60601-1 and IEC 80601-2-30, so the different range will be acceptable for the subject device.
    1. Different Cuff circumference
      The range of pressure is wider than the predicate device while the cuff circumference is narrower than predicate device. They have the same accuracy. The differences are very slightly and they both contain the blood pressure range of most people, and the measurement range of proposed device is fully verified according to IEC 80601-2-30, so the different range and accuracy will be acceptable for the subject device.
    1. Different Power supply
      The difference between subject device and predicate device is the batteries, the predicate device use 4 *1.5V AAA batteries while the subject device use USB DV5V or 4pcs AAA batteries, but the power supply safety of subject device is justified by the IEC60601-1 test reports. Thus this difference will not affect the safety and effectiveness.

As seen in the comparison tables, the subject and predicate devices have same design principle, similar design features and performance specifications. The different technological characteristics between the subject and predicate devices will not raise different questions of safety or effectiveness as demonstrated in the non-clinical and clinical evidence.

8. Performance Testing

{7}------------------------------------------------

Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.

Non-Clinical Testing:

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Blood pressure monitor was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the testing included the following tests, results of which demonstrate the biocompatibility of the subject device:

  • Cytotoxicity
  • Skin Sensitization
  • Skin Irritation

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005+AMD1 (2012) +AMD2 (2020) Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014/AMD1:2020 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.

Bench Testing

Bench testing was conducted and the results show that the subject device complies with the IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers. And blood pressure Accuracy meets the requirements defined in IEC 80601-2-30.

Usability Testing

Usability testing was conducted and the results show that the subject device complies with the IEC 60601-1-6:2010+A1:2013+A2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.

Home-used medical equipment requirements and environmental test:

Environmental testing was conducted and the test results show that the subject device complies with the IEC 60601-1-11:2015 +A1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

Software Verification and Validation Testing

Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.

{8}------------------------------------------------

Clinical data:

Clinical test has been performed in accordance with ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. The clinical test included 100 adult subjects. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018. No adverse effect and/or complication is found in the study.

9. Conclusion

Verification and validation testing was conducted on the subject device Blood pressure monitor and all testing passed pre-specified criteria. The subject device and the predicate device have very similar intended use and the differences in technological features do not raise different questions of safety and effectiveness. This premarket notification submission demonstrates that the subject device is substantially equivalent to the predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).