K Number
K232713

Validate with FDA (Live)

Date Cleared
2023-11-28

(84 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 16cm to 36cm (about 6½"-14½"), 22cm to 32cm (about 8¾"-12½"), 22cm to 42cm (about 8¾"-16½") or 22cm to 45cm (about 8¾"-17¾").

The cuff with arm circumference range of 1636cm is intended for children older than 6 years old or adults. The cuffs with arm circumference range of 2232cm or 2242cm or 2245cm, which are intended for adult population.

It is intended indoor use only.

Device Description

Blood Pressure Monitor TMB-2266 is designed to measure systolic pressure, diastolic pressure and pulse rate of population at or over 3 years old by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".

The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, five types of cuffs have been clinically validated to be matched with the device, suitable for population aged at or over three and with arm circumference from 16cm~45cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is medical witch power supply, PCB, thermistor, pressure pump, motor, release valve and pressure sensor.

The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ± 25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results.

An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

AI/ML Overview

The provided document describes the K232713 Blood Pressure Monitor, which is a digital monitor for measuring blood pressure and pulse rate. The information below is extracted from an FDA 510(k) summary concerning the device's substantial equivalence to predicate devices.

1. Acceptance Criteria and Reported Device Performance

The device's performance acceptance criteria are primarily derived from the ISO 81060-2 standard for non-invasive sphygmomanometers. The reported performance indicates that the device met these criteria.

Acceptance Criteria (Standard)Reported Device Performance
ISO 81060-2:2018+A1:2020: Mean error and standard deviation of differences in systolic and diastolic pressure not beyond specified limits.All data's mean error and standard deviation of differences in systolic and diastolic pressure is not beyond the limits set as per ISO 81060-2:2020. No adverse effect and/or complications were found in the study. This was confirmed in four clinical studies for different cuff models.
IEC 60601-1: Basic safety and essential performanceCompliance demonstrated through testing.
IEC 60601-1-2: Electromagnetic compatibilityCompliance demonstrated through testing.
IEC 80601-2-30: Basic safety and essential performance of automated non-invasive sphygmomanometersCompliance demonstrated through testing.
IEC 60601-1-11: Requirements for medical electrical equipment and systems used in home healthcareCompliance demonstrated through testing.
Biocompatibility (ISO 10993-1/5/10/23): Cytotoxicity, Sensitization, IrritationAll patient contact parts meet the requirements of ISO 10993-1/5/10/23 (cytotoxicity, sensitization, and irritation). The device is considered surface contacting for a duration of exceeding 24 hours but not exceeding 30 days.
Bluetooth Wireless (47 CFR Part 15, Subpart C 15.247): Compliance for wireless moduleThe wireless module was validated according to 47 CFR Part 15, Subpart C 15.247, demonstrating compliance.

2. Sample Size and Data Provenance for Test Set (Clinical Studies)

  • Sample Sizes:
    • Study 1: 87 qualified subjects
    • Study 2: 86 qualified subjects
    • Study 3: 86 qualified subjects
    • Study 4: 102 qualified subjects (65 adults, 37 children aged 3-12)
  • Data Provenance: The document does not explicitly state the country of origin. The studies are described as "clinical studies," implying they were prospective.

3. Number of Experts and Qualifications for Ground Truth Establishment

The document does not provide details on the number of experts or their specific qualifications for establishing the ground truth measurements during the clinical studies. However, for blood pressure monitor validation studies conforming to ISO 81060-2, trained observers are typically used to obtain reference blood pressure measurements using a validated auscultatory method.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1). For ISO 81060-2 studies, it is common to have multiple observers simultaneously or sequentially measure blood pressure using a reference method, and their measurements (or an average) are used as the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No information is provided about a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human interpretation.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes, the clinical studies describe the standalone performance of the blood pressure monitor, which operates autonomously to measure blood pressure and pulse rate (algorithm only). The studies evaluated the device's accuracy against a recognized reference standard (implied by ISO 81060-2).

7. Type of Ground Truth Used

The ground truth for the clinical studies was established through comparative measurements against a reference method, which for ISO 81060-2 studies typically involves simultaneous or sequential auscultatory measurements performed by trained observers.

8. Sample Size for the Training Set

The document does not provide information on the sample size used for the training set. This is typical for medical device submissions that focus on the clinical validation of the final product and do not disclose details about internal development or machine learning training data.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set was established, as details about the device's internal development and training are not part of this 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 28, 2023

Guangdong Transtek Medical Electronics Co., Ltd. Jerry Fan, RA Manager Zone A, No.105, Dongli Road, Torch Development District Zhongshan, Guangdong 528437 China

Re: K232713

Trade/Device Name: Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 31, 2023 Received: September 5, 2023

Dear Jerry Fan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232713

Device Name Blood Pressure Monitor

This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 16cm to 36cm (about 6½"-14½"), 22cm to 32cm (about 8¾"-12½"), 22cm to 42cm (about 8¾"-16½") or 22cm to 45cm (about 8¾"-17¾").

The cuff with arm circumference range of 1636cm is intended for children older than 6 years old or adults. The cuffs with arm circumference range of 2232cm or 22~45cm, which are intended for adult population.

It is intended indoor use only.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Expiration Date: 06/30/2023 See PRA Statement below.

Form Approved: OMB No. 0910-0120

{4}------------------------------------------------

510(k) Summary

Prepared in accordance with the requirement of 21 CFR Part 807.92

Prepared Date: 08/31/2023

1. Submitter
Name:Guangdong Transtek Medical Electronics Co., Ltd.
Address:Zone A, No. 105, Dongli Road, Torch Development District, 528437 Zhongshan,
Guangdong, China
Contact:Jerry Fan
Title:RA Manager
E-mail:gt-rateam@transtekcorp.com
TEL:+86 15728668528

2. Subject Device

Trade/Device NameBlood Pressure Monitor
ModelTMB-2266
Common NameBlood Pressure Monitor
ClassificationClass II
Product CodeDXN
Submission TypeTraditional 510(k)

3. Predicate Device

/Predicate DeviceReference Device
Manufacturer:Guangdong Transtek MedicalElectronics Co., Ltd.Guangdong Transtek MedicalElectronics Co., Ltd.
Device Name /Model:Blood Pressure Monitor / LS802-GSWelch Allyn 901123 Digital BloodPressure Device / ProBP TM 2000
510(k) Number:K202891K181832

4. Device Description

Blood Pressure Monitor TMB-2266 is designed to measure systolic pressure, diastolic pressure and pulse rate of population at or over 3 years old by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals. Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".

The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, five types of cuffs have been clinically validated to be matched with the device, suitable for population aged at or over three and with arm circumference from 16cm~45cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is medical witch power supply, PCB,

{5}------------------------------------------------

thermistor, pressure pump, motor, release valve and pressure sensor.

The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ± 25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results.

An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.

5. Indications for use

This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 16cm to 36cm (about 6½ "-14½"), 22cm to 32cm (about 8%"-12½"), 22cm to 42cm (about 8¾"-16½") or 22cm to 45cm (about 8¾"-17¾").

The cuff with arm circumference range of 16~36cm is intended for children older than 6 years old or adults.

The cuffs with arm circumference range of 2232cm or 2242cm or 22~45cm, which are intended for adult population.

It is intended indoor use only.

{6}------------------------------------------------

  1. Comparison to Predicate Device
FeaturesSubject DevicePredicate Device(K202891)Reference Device(K181832)Note
ModelTMB-2266LS802-GSProBP™ 2000/
ApplicantGuangdong Transtek MedicalElectronics Co., Ltd.Guangdong Transtek MedicalElectronics Co., Ltd.Guangdong Transtek MedicalElectronics Co., Ltd.Same
Device NameBlood Pressure MonitorBlood Pressure MonitorWelch Allyn 901123 Digital BloodPressure DeviceSame
Product CodeDXNDXNDXNSame
ClassificationClass IIClass IIClass IISame
Regulation #21 CFR 870.113021 CFR 870.113021 CFR 870.1130Same
Type of UseOTCOTCOTCSame
Intended Use /Indication for UseThis Blood Pressure Monitor is adigital monitor intended for use inmeasuring blood pressure and pulserate with arm circumference rangingfrom 16cm to 36cm (about 6 1/3"-145/8"),22cm to 32cm (about 83/4"-12½"),22cm to 42cm (about 83/4"-16½") or22cm to 45cm (about 83/4"-173/4").The cuff with arm circumferencerange of 16~36cm is intended forchildren older than 6 years old orThe Transtek Blood Pressure Monitoris digital monitors intended for use inmeasuring blood pressure andheartbeat rate with armcircumference ranging from 22cm to45cm (about 83/4"-17½"). The devicecan be used to detect irregularheartbeat. It is intended for adultindoor use only.The Welch Allyn ProBP 2000 Digitalblood pressure device is intended foruse in measuring blood pressure andheart rate in patients at least 3 yearsof age or older with armcircumferences between 15 cm to 55cm (approximately 5.9 to 21.7inches).The Welch Allyn ProBP 2000automatically measures systolic anddiastolic pressure and pulse rate. TheSimilar, referto Note1
The cuffs with arm circumference range of 2232cm or 2242cm or 22~45cm, which are intended for adult population.It is intended indoor use only.device is intended to be used by clinicians and medically qualified personnel.
Patient PopulationPopulation at or over 3 years oldAdultAt least 3 years of age or olderSimilar, refer to Note2
PrincipleOscillometric methodOscillometric methodOscillometric methodSame
Anatomical SiteUpper ArmUpper ArmUpper ArmSame
Where used (hospital, home, ambulance, etc.)HomeHomeMedical InstitutionsSame
Power Supply4*1.5V AAA batteries;Or by DC 5V adapter4*1.5V AA BatteryOr by DC 6V adapter4*1.5V AA BatteryOr by DC 6V adapterSimilar, refer to Note3
Human FactorsBlood pressureBlood pressureBlood pressureSame
Measurement ItemsMeasuring systolic and diastolic blood pressure and pulse rate of intended population, including irregular pulse rhythm detection.Measuring systolic and diastolic blood pressure and pulse rate of adult individual, including irregular pulse rhythm detection.Measuring systolic and diastolic blood pressure and pulse rate of intended population, including irregular pulse rhythm detection.Same
Cuff DeflationAutomatic deflationAutomatic deflationAutomatic deflationSame
Blood Pressure0mmHg ~ 299mmHg,0mmHg ~ 299mmHg,0mmHg ~ 300mmHg,Same
Measurement5°C - 40°C within ±3mmHg (0.4kPa)5°C - 40°C within ±3mmHg (0.4kPa)5°C - 40°C within ±3mmHg (0.4kPa)Same
Measurement
DisplayLCDLCDLCDSame
Memory2×1996099Different, refer to Note4
Operation EnvironmentTemperature: 5°C~40°C;Relative Humidity: 15%90% RH;Atmospheric: 700hPa1060hPaTemperature: 5°C~40°C;Relative Humidity: 15%90% RH;Atmospheric: 70kPa106kPaTemperature: 5°C to 40°C; Relative Humidity: ≤85% RHAtmospheric Pressure: 86kPa to 106kPaSame
Storage andtransportationEnvironmentTemperature: -20℃60℃;Relative humidity≤93%RH, non-condensing;Atmospheric: 500hPa1060hPaTemperature: -20℃60℃;Relative humidity≤93%RH, non-condensingAtmospheric: 500hPa1060hPaTemperature: -20°C to 60°CRelative Humidity: 10% RH - 93% RHAtmospheric Pressure: 50kPa - 106kPaSame
PerformanceCompliance with IEC 80601-2-30Compliance with IEC 80601-2-30Compliance with IEC 80601-2-30Same
ClinicalCompliance with ISO 81060-2Compliance with ISO 81060-2Compliance with ISO 81060-2Same
BiocompatibilityAll patient contact parts meet therequirements of ISO 10993-1/5/10/23All patient contact parts meet therequirements of ISO 10993-1/5/10All patient contact parts meet therequirements of ISO 10993-1/5/10Same, referto Note5
Electrical SafetyCompliance with IEC 60601-1 andIEC 60601-1-11Compliance with IEC 60601-1 andIEC 60601-1-11Compliance with IEC 60601-1Same
EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2Compliance with IEC 60601-1-2Same
WirelessBluetoothLTENo wireless functionDifferent,
refer to
Note6

adults.

Guangdong Transtek Medical Electronics Co., Ltd.

Vol_008_510(k) Summary

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Justification for difference:

Note 1:

The subject device shares a similar intended use with the predicate device and reference device. The substantial difference between the subject device and the predicate devices regarding the intended use lies in the arm circumference range. The subject device has the same intended population as reference device, though different from the predicate device. Used on its intended population, the subject device with matched cuffs has also been validated according to IEC 80601-2-30 and ISO 81060-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Note 2:

The subject device has the same intended population as reference device, though different from the predicate device. Used on its intended population, the subject device with matched cuffs has also been validated according to IEC 80601-2-30 and ISO 81060-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Note 3:

Same as the predicate device and reference device, the subject device can be powered either by battery or adapter, though there is tiny difference in the specifications of applicable battery and adapter. The subject device with its matched battery and adapter has been validated according to IEC 60601-1. IEC 60601-1-11. IEC 80601-2-30 and IEC 60601-1-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Note 4:

The subject device enjoys a larger memory room for storing measured data than both the predicate device and reference device. This difference, however, won't affect the normal measuring function of the devices. Besides, the subject device has also been validated according to IEC 60601-1, IEC 60601-1-2 and IEC 80601-2-30. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Note 5:

The cuffs of both the subject and predicate device are in compliance with international standards of compatibility. The difference displayed above is caused by the update of standards. The new cuffs are tested and proved to meet requirements of the latest ISO 10993-5/10/23, in terms of cytotoxicity, sensitization and irritation respectively. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Note 6:

The wireless module employed is different between the subject device and the predicate devices. Guangdong Transtek Medical Electronics Co., Ltd. 008- 7 / 9 Vol_008_510(k) Summary

{11}------------------------------------------------

However, it serves the same purpose, that is to transfer measurement results. As for the wireless technology of the subject device, it has also been validated according to 47 CFR Part 15, Subpart C 15.247. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

Conclusion:

Based on the comparison and analysis in this submission, it can be concluded that: the subject device is substantially equivalent to the predicate device and reference device regarding safety and effectiveness.

7. Performance Data

The following performance data were provided in support of the substantial equivalence determination:

Biocompatibility testing:

The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The biocompatibility testing includes the following tests:

  • Cytotoxicity
  • Sensitization
  • Irritation ●

The subject device is considered as surface contacting for a duration of exceed 24 hours but not exceed 30 days.

Non-clinical data

The device has been tested according to following standards:

  • IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirement for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
  • IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.
  • IEC 60601-1-11: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • FDA Guidance for Non-Automated Sphygmomanometer

Clinical data

device was was was and tested according to ISO 81060-2:2018+A1:2020 Non-invasive The sphygmomanometers - Part 2: Clinical validation of intermittent automated measurement type. Four clinical studies were performed on the device matching with different models of cuff. The study

Guangdong Transtek Medical Electronics Co., Ltd. Vol_008_510(k) Summary

{12}------------------------------------------------

population consisted of 87, 86, 86 and 102 qualified subjects respectively. For the first three studies, all subjects were adults. While for the last one, 65 subjects aged over 18 and 37 at the age of 3-12 were included. All data's mean error and standard deviation of differences in systolic, diastolic pressure is not beyond the limits set as per ISO 81060-2:2020. No adverse effect and/or complications are found in this study.

8. Conclusion

Based on the information presented in this 510(k) premarket notification submission, a conclusion can be drawn that the proposed subject device is considered substantially equivalent to the predicate device. The differences between the subject device and the predicate devices were successfully tested with relevant standards and FDA guidance, and do not affect equivalent safety and effectiveness or raise new issues concerning safety and effectiveness.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).