Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K242541
    Device Name
    GEM ZIPCLIP
    Date Cleared
    2025-01-15

    (142 days)

    Product Code
    Regulation Number
    878.4300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ZIPCLIP is indicated for use in surgical procedures to occlude blood vessels.

    Device Description

    The GEM ZIPCLIP (ZIPCLIP) is a sterile, single-use automatic microclip applier. Each device contains 15 titanium microclips. When applied on a vessel, microclips remain with the patient as permanent implants. ZIPCLIP is for prescription use only and the use environment is the operating room. The ZIPCLIP device is a pre-loaded, disposable, single-patient use mechanical assembly comprised of plastic and metal components. The applier device has (2) scissor-like handles that, when driven medially, close the distal jaws forming a closed clip the device in the open configuration. When released, the handles return to their resting state and the closed clip disengages from the applier. Simultaneously, a new clip is automatically loaded into the distal jaws for consecutive firing. When ZIPCLIP is empty a lockout clip (anodized gold in color) deploys between the jaws, preventing them from closing. The ZIPCLIP is a mechanical assembly comprised of plastic and metal components. The clips are composed exclusively of titanium (Grade 1) and are supplied sterile in a preloaded channel that is incorporated into the number of clips per applier is fifteen (15). The clips are stacked and contained internal to the device. Clips cannot be reloaded once the stack is deployed and the applier is disposed of once emptied. Note, the ZIPCLIP preloaded microclips that can be used with the ZIPCLP device.

    AI/ML Overview

    Here is an analysis of the provided text regarding the acceptance criteria and study for the device, organized according to your request.

    Please note: The provided document is a 510(k) summary for a medical device (GEM ZIPCLIP, an implantable clip) and primarily focuses on demonstrating substantial equivalence to a predicate device. It describes performance testing but does not detail acceptance criteria in the format of specific thresholds for metrics like sensitivity, specificity, accuracy, or human reader improvement, which are typical for AI/ML-based medical devices or diagnostic tools. The document describes a traditional medical device (a surgical clip) and its mechanical performance rather than the performance of a software algorithm. Therefore, many of your requested points related to AI/ML device studies (e.g., ground truth establishment for training sets, MRMC studies, standalone algorithm performance, AI assistance effect size) are not applicable to this document's content.


    Acceptance Criteria and Device Performance for GEM ZIPCLIP

    As per the provided 510(k) Summary for the GEM ZIPCLIP, the device is a sterile, single-use automatic microclip applier containing titanium microclips for occluding blood vessels. The performance testing described is focused on the mechanical and functional aspects of the device, rather than the diagnostic capabilities typical of AI/ML software.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a formal table of specific, quantitative acceptance criteria (e.g., success rates, tensile strength thresholds) with corresponding numerical reported performance values. Instead, it states the purpose of the testing and implies successful meeting of implicit performance criteria necessary for safe and effective occlusion.

    Acceptance Criteria Category (Implied)Description from DocumentReported Device Performance (Implied)
    Functional PerformanceReliable deployment of clips for blood vessel ligation."Bench performance testing to verify that ZIPCLIP reliably deploys the clips and ligates vessels was performed."
    Occlusion EffectivenessSafe and effective occlusion of blood vessels."ZIPCLIP deployment of microclips on simulated vessels and microclip ligation of simulated vessels was performed using benchtop models to verify the ZIPCLIPs ability to safely and effectively occlude blood vessels."
    MR-Conditional LabelingDevice is appropriate for MR-Conditional labeling."Additional testing was performed to verify the ZIPCLIP microclip is appropriate for MR-Conditional labeling..."
    Safety and EffectivenessSafe and effective for the intended user population."...and validate that ZIPCLIP is safe and effective for the intended user population."
    Substantial EquivalenceNo new questions of safety or effectiveness compared to the predicate device."The design/ technological differences were found to not affectiveness and are supported by design verification and validation activities performed. The non-clinical testing performed supports the subject device for its intended use."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact number of test units (e.g., appliers, clips) or the number of vessels or simulations used in the bench testing. It broadly states that "bench performance testing" was performed.
    • Data Provenance: The testing was "bench testing" conducted by Baxter (Synovis Micro Companies Alliance is part of Baxter, as indicated by the contact email). This implies a controlled laboratory environment.
      • Country of Origin: Not explicitly stated, but typically assumed to be the country of the manufacturer or its testing facilities (likely USA, given the FDA submission).
      • Retrospective or Prospective: Not applicable as this relates to clinical data. The bench testing would be considered prospective in its execution (i.e., tests were planned and executed to gather performance data).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This concept is not applicable to this type of device and study. The "ground truth" for a surgical clip is its physical performance (e.g., does it close properly, does it occlude a vessel, does it stay closed). This is assessed via engineering and mechanical testing, not by expert interpretation of images or clinical outcomes in the diagnostic sense.

    4. Adjudication Method for the Test Set

    Not Applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where multiple human readers or experts are involved in interpreting complex data (e.g., medical images) to establish a consensus "ground truth." For the mechanical testing of a surgical clip, direct measurements and observable outcomes determine performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC study was not done. This type of study is specifically designed for evaluating diagnostic devices, especially those involving AI, to measure the impact of AI assistance on human reader performance. As the GEM ZIPCLIP is a mechanical surgical device and not a diagnostic tool, an MRMC study is not applicable. Therefore, there is no effect size reported for human readers improving with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not Applicable. The GEM ZIPCLIP is a physical medical device (surgical clip applier), not a software algorithm. Therefore, standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance testing is:

    • Physical Verification/Measurement: Successful deployment, proper closure of clips, and effective occlusion on "simulated vessels" using "benchtop models."
    • Material Properties Testing: Verification of titanium material (Grade 1) and MR-Conditional properties.
    • Functional Observation: Observing the mechanical actions of the applier (e.g., automatic loading, lockout mechanism).

    Essentially, the "ground truth" is established by direct engineering and biomechanical testing demonstrating the functional capabilities and material characteristics of the device.

    8. The Sample Size for the Training Set

    Not Applicable. This device is not an AI/ML algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the GEM ZIPCLIP device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233394
    Date Cleared
    2024-01-28

    (117 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FLOW NOW is intended for monitoring blood flow in vessels.

    FLOW NOW is indicated for monitoring blood flow in peripheral vessels during and following reconstructive microvascular procedures, re-implantation, and free flap transfers. Postoperatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW NOW Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days postoperatively.

    Device Description

    The GEM FLOW NOW (FLOW NOW) device is comprised of an implantable non-active silicone vessel sleeve (Figure 2), a 20 MHz ultrasonic Doppler transducer (probe) and probe wire, suture sleeve, probe connector, and an external lead that connects to a monitor. FLOW NOW has a soft, pliable silicone "vessel sleeve" that retains the Doppler probe via press-fit into a silicone bore. The device comes with vessel sleeve and probe already assembled.

    The FLOW NOW Device is intended for monitoring blood flow in vessels. It is indicated for monitoring blood flow in peripheral vessels intraoperatively, and following reconstructive micro-vascular procedures, re-implantation, and free-flap transfers. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW NOW Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively. The FLOW NOW device is intended for use with both currently cleared FLOW COUPLER Monitors (GEM1020M and GEM1020M-2).

    The probe is removed from the patient 3 to 14 days after implantation, once monitoring is complete. The probe is separated from the vessel sleeve and removed from the patient's body. non-invasively, by the surgeon pulling on the probe wire outside the patient's body. The silicone vessel sleeve remains permanently implanted. The FLOW NOW device can be used to monitor blood flow through veins and arteries ranging from 2.0 to 4.0 mm in outside diameter.

    AI/ML Overview

    The provided text is a 510(k) Summary for the GEM FLOW NOW (GEM2770-FN) device. It details the device's technical characteristics and claims of substantial equivalence to a predicate device but does not contain information about the acceptance criteria or a study proving device performance against those criteria in the context of human reader improvement with AI assistance (MRMC study) or standalone algorithm performance.

    The document primarily focuses on non-clinical bench testing to demonstrate that the GEM FLOW NOW met defined specifications and comparable performance to its predicate device. It explicitly states that "Clinical performance testing was not required for the GEM FLOW NOW device."

    Therefore, I cannot provide the requested information regarding acceptance criteria, reported device performance in those specific contexts, sample sizes for test sets, data provenance, expert details, adjudication methods, MRMC study effect sizes, or standalone algorithm performance.

    Here's what I can extract based on the provided text, focusing on the available performance data:

    1. A table of acceptance criteria and the reported device performance

    The document mentions that "Pre-determined performance specifications were tested, and verification and validation activities were conducted to demonstrate that the GEM FLOW NOW met the defined criteria." However, the specific quantitative acceptance criteria and the detailed reported device performance are not provided in this document. It only states a general conclusion: "The GEM FLOW NOW met acceptance criteria and demonstrated comparable performance to the predicate device for the equivalent indications for use."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided as the performance data described is non-clinical bench testing, not a clinical study with a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. The document discusses non-clinical bench testing, not studies requiring expert review of patient data for ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical performance testing was not required for the GEM FLOW NOW device." This means no study involving human readers or AI assistance was conducted or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone algorithm performance study was not done. The device is a physical diagnostic tool (ultrasonic pulsed doppler imaging system) and not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical bench testing, the ground truth would be based on engineering specifications and measurements from the physical device. Specific details are not provided.

    8. The sample size for the training set

    This information is not applicable/provided. The device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. The device is not an AI/ML algorithm that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221843
    Date Cleared
    2022-09-14

    (82 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GEM™ BIOVER disposable microvascular clamps are instruments which are used for all microsurgical procedures. They are used to occlude vessels during anastomosis, which is necessary as a result of vessel damage or thrombosis.

    Device Description

    The GEM™ Biover Microvascular Clamps (Clamps) are arterial and venous clamps in both single and double clamp configurations, available in varying sizes. The clamps are sterile, disposable clamps for single use. The clamps are used in microsurgery, plastic surgery, and hand surgery for the temporary occlusion of blood vessels during an anastomosis.

    The clamps are available for arteries and veins. The clamping force is higher for the arterial clamps than for the veinous clamps. The clamps are identified by a product reference code and by the color of the clamp. The clamps for veins are green, and the clamps for arteries are yellow.

    The clamps are packaged individually in two Tyvek pouches (double pouched). The devices in the Tyvek pouches are gamma sterilized and are placed in cartons of 10. The clamps are used once and are disposed of after use.

    AI/ML Overview

    This document is a 510(k) summary for the GEM™ Biover Microvascular Clamps, asserting their substantial equivalence to a previously cleared predicate device. This type of submission generally focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than proving performance against specific quantitative acceptance criteria in a robust statistical study.

    Based on the provided text, the device is not an AI/ML powered device, nor is it a diagnostic device that requires high-level performance metrics, such as sensitivity, specificity, or AUC. The submission is for a physical medical device (microvascular clamps). Therefore, the information requested in the prompt regarding AI/ML performance metrics, sample sizes for test/training sets, expert adjudication, MRMC studies, and ground truth establishment is not applicable to this document.

    The document states:

    • "The GEM Biover microvascular clamps remain unchanged from the predicate device."
    • "There are no engineering or performance changes to the device or its packaging."

    The "performance data" section mentions:

    • "The label design change and the outer carton sales unit design change were assessed and validated in a Packaging Summative Human Factors/Usability Study."
    • "Historical complaint data were reviewed and indicate no use-related concerns, thus, there are no use-related risks or complaints that trigger a need for further Human Factors validation."

    This indicates that the "performance" considered for this submission relates to packaging and user interface (labeling/carton design), and the study conducted was a human factors/usability study, not a clinical performance study measuring accuracy or efficacy of a medical algorithm.

    Therefore, I cannot populate the table or answer the specific questions related to acceptance criteria, AI/ML performance, and ground truth for a diagnostic or AI-powered device, as these details are not present and are not relevant to this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213974
    Date Cleared
    2022-04-20

    (121 days)

    Product Code
    Regulation Number
    870.2100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively.

    Device Description

    The GEM FLOW COUPLER System is designed for the detection of blood flow in vessels. The GEM FLOW COUPLER System consists of components as described below:

    1. FLOW COUPLER Device (Product Code MVR) - single-use implantable rings used to secure anastomosis of small vessels. This includes a 20MHz ultrasonic Doppler transducer (probe) which attaches to one of the FLOW COUPLER rings, and an external lead.
    2. FLOW COUPLER Monitor (Product Code DPW) - portable monitor that provides the audible output of the FLOW COUPLER Device's pulsed Doppler ultrasound signal. A varying audible signal is produced when the FLOW COUPLER Device (probe) detects flow.
    AI/ML Overview

    The medical device in question is the GEM FLOW COUPLER Monitor (GEM1020M-2).

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Performance TestingAcceptance CriteriaReported Device Performance
    FLOW COUPLER Monitor cleaning per IEC 60601-1:2005+AMD1:2012 section 11.6.6 and Cleaning ConditioningPassing inspection for the damage and post-test leakage and dielectric test. Functioning monitor after the cleaning.Pass
    FLOW COUPLER Monitor Design Inspection and Functional DemonstrationThe power switch meets design and functional specifications.Pass
    FLOW COUPLER Monitor Audio Quality TestingAudio level greater than 70dB. Signal to Noise ratio of the output electrical signal to the speaker shall be greater than 20 at all tested frequencies. Total Harmonic Distortion (THD) within specification at varying frequencies.Pass
    FLOW COUPLER Monitor EMC TestingIEC 60601-1-2 Edition 4.1: 2020 Class A for Emissions, Immunity for Professional Healthcare Facility Environment. And 47 CFR, Part 15:2022. §15.107 and §15.109. Class A.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for each specific test set (e.g., how many monitors were tested for cleaning, how many for audio quality). However, the tests described are primarily non-clinical performance tests of the device's hardware, firmware, and cleaning protocols. Therefore, the "data provenance" is typically within a controlled laboratory or manufacturing environment, rather than clinical patient data. There is no information regarding country of origin of data or whether it was retrospective or prospective, as these are not relevant to these types of engineering/performance tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    For the types of tests conducted (e.g., cleaning, design, audio quality, EMC), the "ground truth" is established by industry standards (e.g., IEC 60601-1:2005, IEC 60601-1-2 Edition 4.1: 2020, 47 CFR, Part 15:2022) and the manufacturer's internal design and functional specifications. These are objective measures rather than subjective expert consensus. Therefore, the concept of "experts establishing ground truth" in the manner of medical image interpretation is not directly applicable here. The "experts" would be the engineers and technicians performing the tests and verifying compliance with the established standards and specifications. Their qualifications are implicitly in their ability to perform such tests and interpret the standards.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense for non-clinical performance and engineering tests. The tests are designed with objective acceptance criteria, and the results (e.g., measured values, pass/fail based on a threshold) are typically directly compared against these criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document describes a 510(k) Pre-market Notification for modifications to an existing device, the GEM FLOW COUPLER Monitor. The changes are primarily hardware and firmware updates (power management, power button, audio system, remote monitoring disabling) and labeling updates (cleaning solutions). The device itself (a cardiovascular blood flowmeter) provides audible output to a human user to detect blood flow and confirm vessel patency. It does not involve AI or image interpretation, and therefore, an MRMC study or an assessment of human reader improvement with AI assistance is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This device is a monitor that provides audible output to a human user (surgeon, nurse) for the detection of blood flow. It is intrinsically a "human-in-the-loop" device in its intended use. It does not operate as a standalone algorithm for diagnosis or decision-making.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance tests is based on:

    • International and Federal Standards: e.g., IEC 60601-1, IEC 60601-1-2, 47 CFR Part 15.
    • Manufacturer's Design and Functional Specifications: These are internal objective criteria derived from engineering requirements.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm or a diagnostic tool that requires a "training set" in the machine learning sense. The performance tests are for hardware and software functionality, not for training a model.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093310
    Date Cleared
    2010-02-01

    (102 days)

    Product Code
    Regulation Number
    878.4300
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K040163
    Date Cleared
    2004-04-16

    (81 days)

    Product Code
    Regulation Number
    878.4300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Microvascular Anastomotic COUPLER is to be used in the anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures.

    Device Description

    The device is a sterile (gamma irradiation), single-use, implantable pair of rings molded out of high density polyethylene with six stainless steel pins on each of the ring. A probe-holder feature is molded on each of the paired rings to be a point of attachment for a sensor/probe at the site of anastomosis.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Microvascular Anastomotic Coupler and does not contain information regarding acceptance criteria or a study proving the device meets said criteria.

    The document outlines the device's indications for use, technological comparison to a predicate device (GEM Microvascular Anastomotic Coupler; K861985), and asserts substantial equivalence in terms of biocompatibility, manufacturing, performance, sterilization, shelf life, packaging, safety, and efficacy to the predicate device.

    Therefore, I cannot provide the requested information from the given text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1