K Number
K213974
Device Name
GEM FLOW COUPLER Monitor (GEM1020M-2)
Date Cleared
2022-04-20

(121 days)

Product Code
Regulation Number
870.2100
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively.
Device Description
The GEM FLOW COUPLER System is designed for the detection of blood flow in vessels. The GEM FLOW COUPLER System consists of components as described below: 1. FLOW COUPLER Device (Product Code MVR) - single-use implantable rings used to secure anastomosis of small vessels. This includes a 20MHz ultrasonic Doppler transducer (probe) which attaches to one of the FLOW COUPLER rings, and an external lead. 2. FLOW COUPLER Monitor (Product Code DPW) - portable monitor that provides the audible output of the FLOW COUPLER Device's pulsed Doppler ultrasound signal. A varying audible signal is produced when the FLOW COUPLER Device (probe) detects flow.
More Information

Not Found

No
The description focuses on a Doppler ultrasound system for detecting blood flow and does not mention any AI or ML capabilities. The performance studies are standard electrical and functional tests, not AI/ML model validation.

Yes
The device is described as assisting in the end-to-end anastomosis of veins and arteries (a surgical procedure), and its function is to detect blood flow and confirm vessel patency intra-operatively and post-operatively. These actions contribute directly to the treatment and monitoring of a patient's medical condition.

Yes

The FLOW COUPLER System, which includes the device, is intended to "detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site." This purpose directly relates to diagnosing the presence and quality of blood flow.

No

The device description clearly outlines both hardware components (implantable rings, Doppler probe, external lead, portable monitor) and software components (processing the ultrasound signal for audible output). It is a system with both hardware and software.

Based on the provided information, the FLOW COUPLER Device is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens derived from the human body (like blood, urine, tissue) outside of the body to provide information for diagnosis, monitoring, or screening.
  • FLOW COUPLER Function: The FLOW COUPLER Device is an implantable device used within the body during and after surgical procedures. It uses ultrasound to detect blood flow in vivo (within the living organism) at the anastomotic site.

The device's function is to monitor blood flow directly within the patient's vessels, not to analyze samples taken from the patient. This clearly places it outside the scope of In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively.

Product codes (comma separated list FDA assigned to the subject device)

DPW, MVR

Device Description

The GEM FLOW COUPLER System is designed for the detection of blood flow in vessels. The GEM FLOW COUPLER System consists of components as described below:

  1. FLOW COUPLER Device (Product Code MVR) - single-use implantable rings used to secure anastomosis of small vessels. This includes a 20MHz ultrasonic Doppler transducer (probe) which attaches to one of the FLOW COUPLER rings, and an external lead.

  2. FLOW COUPLER Monitor (Product Code DPW) - portable monitor that provides the audible output of the FLOW COUPLER Device's pulsed Doppler ultrasound signal. A varying audible signal is produced when the FLOW COUPLER Device (probe) detects flow.

The predicate device (K191252) submission included an additional remote monitoring capability for the remote retrieval of FLOW COUPLER Monitor data via the GEMflow App (which is part of the GEM Cloud System). This optional remote monitoring capability was not extensively utilized and has been excluded from the subject device. The remote monitoring capability is not available with the subject device.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

vessels, veins and arteries

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Nurses and surgeons
Hospital Operating Room and Post-Anesthesia Care Unit

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Synovis conducted performance testing to evaluate safety and effectiveness of the FLOW COUPLER Monitor to support substantial equivalence to the predicate device, as well as to evaluate that product requirements are met. The following testing of the FLOW COUPLER Monitor has been performed:

  • FLOW COUPLER Monitor cleaning per IEC 60601-1:2005+AMD1:2012 section 11.6.6 and FLOW COUPLER Monitor Cleaning Conditioning, with passing inspection for damage, post-test leakage, and dielectric test. The monitor remained functioning after cleaning.
  • FLOW COUPLER Monitor Design Inspection and Functional Demonstration, confirming the power switch meets design and functional specifications.
  • FLOW COUPLER Monitor Audio Quality Testing, with results showing audio level greater than 70dB, Signal to Noise ratio of the output electrical signal to the speaker greater than 20 at all tested frequencies, and Total Harmonic Distortion (THD) within specification at varying frequencies.
  • FLOW COUPLER Monitor EMC Testing, compliant with IEC 60601-1-2 Edition 4.1: 2020 Class A for Emissions, Immunity for Professional Healthcare Facility Environment, and 47 CFR, Part 15:2022. §15.107 and §15.109. Class A.

All applicable verification and validation testing was completed with passing results per the predefined acceptance criteria for each test case. These results support that the modified device performs as intended and meets all functional requirements and performance specifications.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K191252

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.2100 Cardiovascular blood flowmeter.

(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 20, 2022

Synovis Micro Companies Alliance, Inc. Julie Carlston Senior Manager, Regulatory Affairs (a Subsidiary of Baxter International Inc.) 439 Industrial Lane Birmingham, Alabama 35211

Re: K213974

Trade/Device Name: GEM FLOW COUPLER Monitor (GEM1020M-2) Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW, MVR Dated: March 11, 2022 Received: March 14, 2022

Dear Julie Carlston:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Deborah Fellhauer, RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K213974

Device Name GEM FLOW COUPLER Monitor (GEM1020M-2)

Indications for Use (Describe)

The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

-Submitter Synovis Micro Companies Alliance, Inc. (A Subsidiary of Baxter International, Inc.) 439 Industrial Lane Birmingham, AL 35211-4464 Establishment Registration Number: 1062741

Julie Carlston Senior Regulatory Affairs Manager at Synovis Micro Companies Alliance, Inc. (a subsidiary of Baxter International Inc.) Telephone: 651.796.7511 Fax: 651.642.9018 Email: Julie Carlston@baxter.com

Date Prepared: April 14, 2022

II. Subject Device

Device Trade NameGEM FLOW COUPLER Monitor (GEM1020M-2)
Common NameCardiovascular Blood Flowmeter
Classification NameFlowmeter, Blood, Cardiovascular
Regulation Number: 21 CFR 870.2100
Classification: Class II

Implantable Clip
Regulation Number: 21 CFR 878.4300
Classification: Class II |
| Product Code | MVR, DPW |

III. Predicate Device

Device Trade NameGEM FLOW COUPLER System; K191252
Common NameCardiovascular Blood Flowmeter
Classification NameFlowmeter, Blood, Cardiovascular
Regulation Number: 21 CFR 870.2100
Classification: Class II
Implantable Clip
Regulation Number: 21 CFR 878.4300
Classification: Class II
Product CodeMVR, DPW

4

IV. Device Description

The GEM FLOW COUPLER System is designed for the detection of blood flow in vessels. The GEM FLOW COUPLER System consists of components as described below:

  1. FLOW COUPLER Device (Product Code MVR) - single-use implantable rings used to secure anastomosis of small vessels. This includes a 20MHz ultrasonic Doppler transducer (probe) which attaches to one of the FLOW COUPLER rings, and an external lead.

FLOW COUPLER Monitor (Product Code DPW) - portable monitor that provides the 2. audible output of the FLOW COUPLER Device's pulsed Doppler ultrasound signal. A varying audible signal is produced when the FLOW COUPLER Device (probe) detects flow.

The predicate device (K191252) submission included an additional remote monitoring capability for the remote retrieval of FLOW COUPLER Monitor data via the GEMflow App (which is part of the GEM Cloud System). This optional remote monitoring capability was not extensively utilized and has been excluded from the subject device. The remote monitoring capability is not available with the subject devic.

V. Intended Use / Indications for Use

The Intended Use and Indications for Use remain the same as the predicate device (K191252).

Intended Use

The FLOW COUPLER Device and System is intended to be used in the anastomosis of veins and arteries normally encountered in microvascular and vascular reconstructive procedures and in the detection of blood flow and confirmation of vessel patency following end-to-end anastomosis of vessels.

Indications for Use

The FLOW COUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FLOW COUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FLOW COUPLER Device is used in conjunction with the FLOW COUPLER Monitor, the FLOW COUPLER System is intended to detect blood flow and confirm vessel patency intraoperatively and post-operatively at the anastomotic site. Post-operatively, blood flow can be detected on an as needed basis for up to 7 days. The FLOW COUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days postoperatively.

VI. Comparison of Technological Characteristics with the Predicate Device

The overall functionality and features necessary to support the device's intended use remain unchanged from the previous device clearance (K191252). Minor modifications to the product hardware, firmware and labeling have been made in this submission:

. Power management update (monitor firmware update): revised logic for handling a battery-protection shutdown scenario.

5

Power button update (monitor hardware update): recessed the button to be . flush with the device bezel and increased the contact force.

Audio system update (monitor hardware update): updated speaker assembly ● and capacitors on Printed Circuit Board Assembly to reduce background noise and enhance overall audio clarity.

List of approved cleaning solutions update (labeling update): updated . Instructions for Use with an expanded list of approved cleaning solutions.

. Disabling of the optional remote monitoring capability: updated monitor firmware to disable the connection to the cloud backend server (GEM Cloud System) to disable the optional remote monitoring capability.

The technological characteristics and specifications of the FLOW COUPLER Monitor have been evaluated against its predicate device (previous version of the subject device) to determine equivalence.. A summary of the comparison between the predicate and subject devices is provided in Table 1 below.

AttributePredicate DeviceSubject Device
Product NameGEM FLOW COUPLER SystemSame
510(k)-holderSynovis Micro Companies Alliance, Inc.
(A Subsidiary of Baxter International, Inc.)Same
510(k) NumberK191252K213974
Product CodeMVR, DPWSame
Regulation21 CFR 878.4300
21 CFR 870.2100Same
ClassificationClass IISame
Intended UseThe FLOW COUPLER Device and System is intended to be used
in the anastomosis of veins and arteries normally encountered
in microvascular and vascular reconstructive procedures and
in the detection of blood flow and confirmation of vessel
patency following end-to-end anastomosis of vessels.Same
Indications for UseThe FLOW COUPLER Device is a single use, implantable device
that is intended to be used in the end-to-end anastomosis of
veins and arteries normally encountered in microsurgical and
vascular reconstructive procedures. The FLOW COUPLER
Device includes a pair of permanently implanted rings which
secure the anastomosis and a removable Doppler probe that is
press-fit onto one of the rings. When the FLOW COUPLER
Device is used in conjunction with the FLOW COUPLER
Monitor, the FLOW COUPLER System is intended to detect
blood flow and confirm vessel patency intra-operatively and
post-operatively at the anastomotic site. Post-operatively,
blood flow can be detected on an as needed basis for up to 7
days. The FLOW COUPLER Doppler probe is not intended to be
a permanent implant and should be removed 3 to 14 days
postoperatively.Same
Product Use
EnvironmentHospital Operating Room and Post-Anesthesia Care UnitSame
End Users of ProductNurses and surgeonsSame
Dimensions6.17" D x 8.18" W x 3.20" HSame
Weight1.8 lb. (0.8 kg)Same
Acoustical OutputAcoustical Power will be less than 94 mW/cm2Same
DepthNot selectableSame
Transmission
Frequency20 MHzSame
Transmission
CharacteristicPulsed wave transmission, continuous receptionSame
Pulse Repetition
Frequency156.25 KHzSame
Energy Used and/or
Delivered12 Volts direct current supplied by A/C to D/C converter
Rechargeable internal lithium ion battery packSame
Power InputInput Voltage: 90-264 VAC
Frequency: 47-63 HzSame
Power OutputTotal Output Power: 30 watts
Voltage: 12VDC
Current: 2.5A
Output Regulation ± 5%Same
Power management update
Monitor firmwareAnalog Front End, Doppler phase shift detection, plus FFT
filtering, noise reduction, and background noise reductionRevised logic for handling a
proper battery-protection
shutdown scenario to prevent
complete battery drainage and
inoperable monitor
Power button update
Power buttonLocation: rear of unit
Elevation: raised (3.50 mm)Location: rear of unit
Elevation: flush with device
bezel
Activation force: 400gf
Audio system update
Printed Circuit Board
Assembly (PCBA)PCBA per K191252PCBA components have been
changed to improve audio
quality
Speaker Assembly57.5mm wide, 3W Max. Power speaker
Concave speaker grill60mm wide, 4W Max. Power
speaker
Convex speaker grill
List of approved cleaning solutions update
Cleaning instructionsWipe the Monitor with a dry or water-moistened soft cloth.
Ensure any residual organic material is removed.
Do not pour or spray liquid directly on the Monitor. Allow to
air dry before use.Wipe the Monitor with a dry or
water-moistened soft cloth,
Isopropyl Alcohol, Ammonium
Hydroxide based surface
cleanser, Ammonium Chloride
based surface cleanser or 2%
bleach solution. Ensure any
residual organic material is
removed. Do not pour or spray
liquid directly on the Monitor.
Allow to air dry before use.
Disabling of the optional remote monitoring capability
Remote Access
Function (Optional)Android and iOS App to access audio files remotelyDisabled optional remote
monitoring capability

Table 1 - FLOW COUPLER Monitor Predicate and Subject Devices - Comparison Table

6

7

VII.Performance Data

Synovis conducted performance testing to evaluate safety and effectiveness of the FLOW COUPLER Monitor to support substantial equivalence to the predicate device, as well as to evaluate that product requirements are met. The following testing of the FLOW COUPLER Monitor has been performed.

Performance TestingAcceptance CriteriaResult
FLOW COUPLER Monitor cleaning per
IEC 60601-1:2005+AMD1:2012 section
11.6.6
FLOW COUPLER Monitor Cleaning
ConditioningPassing inspection for the damage and post-test leakage and dielectric test. Functioning monitor after the cleaning.Pass
FLOW COUPLER Monitor Design
Inspection and Functional
DemonstrationThe power switch meets design and functional specificationsPass
FLOW COUPLER Monitor Audio Quality
TestingAudio level greater than 70dB Signal to Noise ratio of the output electrical signal to the speaker shall be greater than 20 at all tested frequencies Total Harmonic Distortion (THD) within specification at varying frequenciesPass
FLOW COUPLER Monitor EMC TestingIEC 60601-1-2 Edition 4.1: 2020 Class A for Emissions, Immunity for Professional Healthcare Facility Environment And 47 CFR, Part 15:2022. §15.107 and §15.109. Class APass

Applicable verification and validation testing was completed with passing results per the predefined acceptance criteria for each test case. These results support that the modified device performs as intended, and meets all functional requirements and performance specifications.

VIII.Conclusions

The FLOW COUPLER System indications for use and principles of operation remain unchanged (are the same as the predicate device). Performance testing has been completed to evaluate safety and effectiveness of the subject device as it compares to its predicate. The conclusion drawn from the riskbenefit assessment and nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device. No new issues of safety and effectiveness have been identified.