Search Results
Found 12 results
510(k) Data Aggregation
(266 days)
SpineCraft, LLC
ASTRA and AVANT Navigated Reusable Instruments are indicated for preparation and placement of SpineCraft ASTRA Spine system pedicle screws during thoracolumbar sacroiliac spinal surgery to assist surgeon in precisely locating anatomical structures in either open, minimally invasive procedures, or percutaneous, procedures.
ASTRA and AVANT Navigated Reusable Instruments are specifically designed for use with Medtronic StealthStation® System S8 (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Use of the ASTRA and AVANT Navigated Reusable Instruments is limited to use only with ASTRA Spine System implants.
ASTRA-OCT Navigated Reusable instruments are indicated for preparation and placement of SpineCraft ASTRA-OCT Spine screws during cervico-thoracic spinal surgery to assist surgeon in precisely locating anatomical structures in open procedures.
ASTRA-OCT Navigated Reusable Instruments are specifically designed for use with Medtronic StealthStation® System S8 (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Use of the ASTRA-OCT Navigated Reusable Instruments is limited to use only with ASTRA-OCT Spine System implants.
The ASTRA, AVANT and ASTRA-OCT Navigation instruments are non-sterile, reusable surgical instruments designed for compatibility with the Medtronic NavLock Trackers and to ultimately provide seamless interaction with the Medtronic StealthStation® System. The ASTRA and AVANT Navigation instruments are for use with ASTRA Spine System pedicle screws and the ASTRA-OCT Navigation instruments are for use with ASTRA-OCT Spine System pedicle screws. The instruments are manufactured from medical grade stainless steel. The ASTRA-OCT navigation instruments are available in same or similar diameters and lengths as the corresponding predicate Medtronic navigated instruments. This includes awls, probes, drill bits, taps and screwdrivers.
This looks like a 510(k) summary for a medical device rather than a study evaluating the performance of an AI/ML powered device. As such, it does not contain the specific information requested in the prompt regarding acceptance criteria and a study proving the device meets those criteria for an AI/ML system.
However, I can extract information related to the device's performance testing based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document provides a list of performance tests conducted according to ASTM F2554-18, implying that the device performance demonstrated it "perform[s] as designed, are suitable for their intended use and are substantially equivalent to the cited corresponding predicate devices under the same test conditions." However, specific numerical acceptance criteria (e.g., "accuracy must be greater than X") and corresponding numerical performance results are not provided in this summary.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Meet requirements of ASTM F2554-18 for positional accuracy of computer-assisted surgical systems. | "Performance testing conducted on the ASTRA, AVANT and ASTRA-OCT Navigation instruments Systems demonstrated that they perform as designed, are suitable for their intended use and are substantially equivalent to the cited corresponding predicate devices under the same test conditions." |
Achieve specific accuracy for single point measurement. | Tested per ASTM F2554-18. Specific results not provided. |
Achieve specific accuracy for instrument axis rotation measurement. | Tested per ASTM F2554-18. Specific results not provided. |
Achieve specific accuracy for instrument angular position perpendicular to the system camera measurement. | Tested per ASTM F2554-18. Specific results not provided. |
Achieve specific accuracy for instrument angular position parallel to the system camera measurement. | Tested per ASTM F2554-18. Specific results not provided. |
Achieve specific accuracy for distance between points measurement. | Tested per ASTM F2554-18. Specific results not provided. |
Dimensional similarity to predicate devices. | "A detailed dimensional analysis and one-to-one comparison has been conducted for subject and predicate device to support the substantial equivalence." and "The ASTRA, AVANT and ASTRA-OCT Navigation instruments Systems are dimensionally similar to the cited corresponding predicate devices." |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This document describes non-clinical performance testing of physical surgical instruments, not a study involving patient data or an AI/ML algorithm's test set. Therefore, information regarding "sample size for the test set" and "data provenance" (country of origin, retrospective/prospective) is not applicable in the context of this 510(k) summary. The testing was laboratory-based.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as the document describes non-clinical performance testing of physical instruments, not an AI/ML algorithm requiring expert ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for the reasons stated above.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not performed as this device is a set of navigated surgical instruments, not an AI/ML assistance system for human readers. No clinical studies were performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to an AI/ML algorithm. The device described, "ASTRA & AVANT Navigation Instruments System and ASTRA-OCT Navigation Instruments System," is a set of non-sterile, reusable surgical instruments designed for compatibility with a navigation system (Medtronic StealthStation® System S8). It is not a standalone AI/ML algorithm. Non-clinical performance testing was done on the instruments themselves.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical performance testing, the "ground truth" would be established by the precise measurement systems and methodologies outlined in ASTM F2554-18 for determining positional accuracy. This is a technical standard measurement, not expert consensus, pathology, or outcomes data typically associated with AI/ML clinical studies.
8. The sample size for the training set
This is not applicable as there is no AI/ML algorithm being trained by this device.
9. How the ground truth for the training set was established
This is not applicable for the reasons stated above.
Ask a specific question about this device
(59 days)
SpineCraft, LLC
The ASTRA Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar, and sacral spine.
The ASTRA Spine System is indicated for non-cervical (T)-S2/Ilium) pedicle fixation and non-pedicle fixation in skeletally mature patients as an adjunct to following indications: degenerative disc disease (DDD - defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; trauma (i.e., fracture and/or dislocation); spinal stenosis; deformities (scoliosis, lordosis and/or kyphosis); spinal tumor; and failed previous fusion (pseudo-arthrosis).
When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (DDD - defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the ASTRA Spine System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the ASTRA Spine System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The ASTRA fenestrated screw when used with other components of the ASTRA Spine System is indicated to provide the surgeon with an open or minimally invasive approach for posterior spinal surgery. The ASTRA fenestrated screw is intended to be used with saline or radiopaque dye.
The ASTRA Spine System consists of Ø 5.5mm, Ø 6.0mm and Ø 6.2mm longitudinal, lordosed, contoured and revision rods, pedicle screws (monoaxial, and uniplanar), cannulated pedicle screws (standard and reduction monoaxial, standard, reduction & extended tab polyaxial and standard & reduction uniplanar), fenestrated screws (standard, reduction & extended tab standard & reduction uniplanar), hooks (standard & reduction), lateral iliac connectors, rod-to-rod connectors and transverse (cross) connectors. Most of the components are available in a variety of sizes to more closely match the patient's anatomy.
The safety and effectiveness of the ASTRA fenestrated screw has not been established when used in conjunction with bone cement or for use in patients with poor bone quality (e.g.,osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye.
Materials: Titanium alloy per ASTM F136 and CoCr alloy per ASTM F1537
This document is a 510(k) summary for a medical device (ASTRA Spine System), not a study evaluating an AI/ML powered device. As such, it does not contain the information required to answer your questions regarding acceptance criteria and a study proving the device meets those criteria for an AI/ML product.
Specifically, the document states: "No clinical studies were performed" (Page 6, Section 8), and the non-clinical tests described are mechanical tests for orthopedic implants (ASTM F1717 and ASTM F1798), not performance evaluations of an AI/ML algorithm.
Therefore, I cannot extract the following information from the provided text:
- A table of acceptance criteria and the reported device performance: This document reports mechanical test results against predicate devices, not AI performance metrics.
- Sample sizes used for the test set and the data provenance: Not applicable to a mechanical device test.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used: Not applicable.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
The document describes a spinal implant system, which is a physical device, and its substantial equivalence is demonstrated through mechanical testing against predicate devices, not through a study involving AI/ML performance.
Ask a specific question about this device
(100 days)
SpineCraft, LLC
The ORIO-Ti cervical intervertebral body fusion cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had at least six (6) weeks of non-operative treatment with an intervertebral cage. ORIO-Ti cervical intervertebral body fusion implants are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at the C-3 to C-7 disc levels using autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. ORIO-Ti cervical intervertebral body fusion implant is to be used with supplemental fixation that has been cleared for use in the cervical spine.
The ORIO-Ti lumbar intervertebral body fusion cage is indicated for intervertebral body fusion procedures in sketally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six (6) months of non-operative treatment. ORIO-Ti lumbar intervertebral body fusion implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via a transforaminal, open posterior, anterior/anterolateral, or lateral approach. The ORIO-Ti lumbar intervertebral body fusion implants are to be used with supplemental fixation that have been cleared for use in the lumbosacral spine.
The subject devices are intervertebral body fusion devices for use in cervical lumbar spinal surgery. They may also be referred to as interbody fusion devices or interbody cages. The devices are generally boxshaped or trapezoidal shaped with windows or pockets in their design to allow for the placement of autograft or allogenic bone graft. The ORIO-Ti Cage System is a Titanium alloy interbody fusion system comprised of various device configurations based on surgical approach and patient anatomy, and may be implanted via one of the following approaches: as a single device via an anterior cervical (ACDF), bilaterally in pairs via a posterior (PLIF) approach; as a single device via a transforaminal (TLIF) approach; or as a single device via an anterior (ALIF) approach. The exterior of the devices has "teeth" on the superior and inferior surfaces to help prevent the devices from migrating once they are implanted.
The provided text is a 510(k) summary for the ORIO-Ti Intervertebral Body Fusion Cage System. It primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and non-clinical testing.
Crucially, the document explicitly states under "8. Clinical Test Summary": "No clinical studies were performed."
Therefore, it is not possible to provide the requested information about acceptance criteria and a study that proves the device meets those criteria, as no clinical studies were conducted for this device. The information provided heavily emphasizes non-clinical testing and comparison to predicate devices, which is common for 510(k) submissions where clinical data is often not required if substantial equivalence can be demonstrated through other means.
Without clinical study data, none of the points you asked for (acceptance criteria table, sample size for test/training sets, data provenance, expert ground truth, adjudication, MRMC study, standalone performance, type of ground truth, training set ground truth establishment) can be answered.
Ask a specific question about this device
(34 days)
SpineCraft, LLC
The ASTRA Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic. lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra: degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis).
The ASTRA Spine System is also a sacral/iliac screw fixation system of the indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the ASTRA Spine System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the ASTRA Spine System is intended to treat pediatric patients diagnosed with the following conditions:
spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The subject ASTRA Spine System is top loading, multiple components, posterior system which consists of rods, cannulated and non-cannulated monoaxial, uniplanar and polyaxial screws, hooks, iliac connectors, rod connectors, and cross connectors. Most of the components are available in a variety of sizes to match the patient's anatomy more closely.
The subject ASTRA Spine System includes the pediatric use in the indications and other labeling updates to provide more clarity to the device insert.
The provided text is a 510(k) premarket notification for the ASTRA Spine System, which is a medical device. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as typically done for new drug applications or certain high-risk medical devices.
Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, device performance studies, sample sizes, ground truth establishment, or multi-reader multi-case studies.
Here's why and what the document does describe:
- Acceptance Criteria & Reported Device Performance: This section is explicitly missing because the submission is not focused on proving the device meets specific performance criteria through new studies. Instead, it argues that the device's design, materials, and function are similar enough to existing, legally marketed predicate devices to be considered substantially equivalent.
- Sample size used for the test set and the data provenance: Not applicable. No new test set in this context.
- Number of experts used to establish the ground truth for the test set and qualifications of those experts: Not applicable. No new ground truth establishment study.
- Adjudication method for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No. The document explicitly states "No clinical studies were performed in support of this submission."
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical spinal implant system, not a software algorithm.
- The type of ground truth used: Not applicable.
- The sample size for the training set: Not applicable. No training set for an algorithm.
- How the ground truth for the training set was established: Not applicable.
What the document does state regarding testing:
- Non-clinical Test Summary: "Non-clinical tests were not performed in support of this submission." This means no new bench testing or material testing data was submitted for this particular 510(k). The assumption is that the materials and design are already established as safe and effective by the predicate devices they are referencing.
- Clinical Test Summary: "No clinical studies were performed in support of this submission." This further reinforces that no human studies were conducted for this specific submission to prove performance or safety.
In summary, for a 510(k) premarket notification like this one, the "proof" the device meets acceptance criteria is primarily based on its substantial equivalence to already cleared predicate devices, meaning it shares similar technological characteristics and indications for use, and does not raise different questions of safety and effectiveness.
Ask a specific question about this device
(45 days)
SpineCraft LLC
The ASTRA-OCT Spine System implants are intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7) and the thoracic spine (T1-T3):
- · Traumatic spinal fractures and/or Traumatic dislocations;
- Instability or deformity;
- · Failed previous fusions (e.g. pseudoarthrosis);
- · Tumors involving the cervical/thoracic spine: and
· Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
The ASTRA-OCT Spine System implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the ASTRA-OCT Spine System rods may be connected to other occipital cervical thoracic or thoracolumbar stabilization rod systems ranging in diameter from 3.5mm, including the ASTRA or APEX Spine Systems, using corresponding connectors.
The ASTRA-OCT Spine System consists of a series of polyaxial screws, occipital screws, occipital plates, hooks, rods, lateral connectors, rod-to-rod connectors, set screws, and cross connectors.
Materials:
Titanium alloy per ASTM F136 CoCr allov per ASTM F1537
This FDA 510(k) summary describes a new medical device, the ASTRA-OCT Spine System, and its substantial equivalence to a predicate device. The information provided is for a traditional medical device (spinal implant) and does not include acceptance criteria or a study proving device meets acceptance criteria in the context of an AI/ML powered device.
Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for an AI/ML device.
The document states:
- "No clinical studies were performed" for the ASTRA-OCT Spine System.
- The substantiation for equivalence is based on non-clinical mechanical testing and comparison to predicate systems.
The requested information is typically found in submissions for AI/ML powered devices, which are assessed differently from traditional hardware devices like the ASTRA-OCT Spine System.
Ask a specific question about this device
(149 days)
SPINECRAFT LLC
The APEX-DL Spine System with APEX Spine System Components is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX-DL Spine System with APEX Spine System Components is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The APEX-DL Spine System with APEX Spine System Components is also a sacraliliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX-DL Spine System with APEX Spine System Components is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
When used for posterior non-cervical pediatic patients, the APEX-DL Spine System implants with APEX Spine System Components are indicated as an adjunct to treat adolescent idiopathic scoliosis. The APEX DL Spine System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The APEX-DL Spine System includes Monoaxial, Uniplanar, and Polyaxial Double Lead Thread Screws in the cannulated and non-cannulated versions and in regular and reduction (extended tab) versions. APEX-DL Spine System is a low profile thoracolumbar implant for use with wide range of patient statures. The APEX-DL Spine System Polyaxial screws feature a friction head, which is designed to provide precise reduction mechanism as a result of easier rod capturing. The APEX-DL Spine System also includes Lordosed Percutaneous Rods. The APEX-DL Spine System is compatible with the APEX Spine System 5.5mm and 6.0mm rods, hooks, side-by-side connectors, iliac connectors, cross connectors, and washers.
The APEX-DL Spine System is a spinal implant for immobilization and stabilization of spinal segments. The document indicates that no clinical studies were performed. The acceptance criteria and testing are based on non-clinical (mechanical) tests.
1. Table of Acceptance Criteria and Reported Device Performance
Test Type | Acceptance Criteria | Reported Device Performance |
---|---|---|
ASTM F1717 (Vertebrectomy Model) | ||
Static Compression Bending | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Static Torsion | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Dynamic Compression Bending | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
ASTM F1798 (Interconnection Mechanisms) | ||
Static Axial Gripping Capacity | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Axial Torque Gripping Capacity | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Static Flexion-Extension | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Dynamic Flexion-Extension | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Note: The acceptance criteria are implicitly defined as demonstrating mechanical properties that are "equal or higher" than the identified predicate devices, based on the testing standards.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated in the provided document. For each of the ASTM tests listed, a specific number of samples (implants/constructs) would have been used, but this detail is not provided.
- Data Provenance: The data is from non-clinical (mechanical) testing conducted presumably by SpineCraft, LLC or a contracted testing facility. It is not patient or human data; therefore, country of origin or retrospective/prospective does not apply in the typical sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is a non-clinical, mechanical testing study, not a study requiring expert clinical assessment or ground truth establishment in a medical imaging or diagnostic context. The "ground truth" here is the objective measurement of mechanical properties according to established ASTM standards.
4. Adjudication Method for the Test Set
Not applicable. This is a non-clinical, mechanical testing study. Adjudication methods like 2+1 or 3+1 are used for expert consensus in clinical or imaging studies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is a clinical study involving multiple human readers interpreting medical cases. The provided document explicitly states, "No clinical studies were performed."
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is a physical spinal implant, not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is based on objective mechanical measurements according to recognized industry standards (ASTM F1717 and ASTM F1798). The performance is then compared to "predicate systems" as a benchmark for substantial equivalence.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device undergoing mechanical testing, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As described in point 8, there is no "training set" for this type of device evaluation.
Ask a specific question about this device
(90 days)
SPINECRAFT LLC
The ASTRA System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis).
The ASTRA Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The ASTRA System is also a sacral/iliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
The ASTRA Spine System is a top loading, multiple component, posterior spinal fixation system which consists of rods, cannulated and non-cannulated monoaxial, uniplanar and polyaxial screws, hooks, iliac connectors, rod connectors, and cross connectors. Most of the components are available in a variety of sizes to more closely match the patient's anatomy.
Materials:
Titanium alloy
CoCr alloy
Here's an analysis of the provided text regarding the ASTRA Spine System, focusing on acceptance criteria and study details.
Important Note: The provided document is a 510(k) summary for a medical device (Spine System), not an AI/ML device. Therefore, many of the requested fields related to AI/ML specific studies (like sample sizes for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance) are not applicable to this type of medical device submission. This document focuses on demonstrating substantial equivalence to predicate devices through non-clinical (mechanical) testing.
1. Table of Acceptance Criteria and Reported Device Performance
For this medical device, the "acceptance criteria" are implicitly defined by performance equivalence to predicate devices in standardized mechanical tests. The "reported device performance" is a statement of comparative equivalence.
Acceptance Criteria (Implicit) | Reported Device Performance (Summary) |
---|---|
Static Compression Bending: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Static Torsion: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Dynamic Compression Bending: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Axial Gripping: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Torsional Gripping: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Static Flexion-Extension: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Dynamic Flexion-Extension: Must perform at least equivalent to predicate systems. | "The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems." |
Overall Substantial Equivalence: Must be substantially equivalent in intended use, design, material, performance, and function. | "The ASTRA Spine System is substantially equivalent to the predicate devices in terms of indications for use, design, material, performance and function." |
2. Sample Size Used for the Test Set and the Data Provenance
This is a physical device, and the "test set" refers to the tested device components.
- Sample Size: Not explicitly stated in terms of number of components tested for each test, but standard engineering practices for medical device testing would involve a sufficient number (e.g., n=5 or n=10 per test) to ensure statistical significance, though the exact numbers are not provided in this summary.
- Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting. There is no patient data provenance (e.g., country of origin, retrospective/prospective) since no human data was used for these mechanical tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- Not Applicable. For mechanical testing of a physical device, ground truth is established by standardized testing protocols (e.g., ASTM standards) and measured physical properties, not by expert consensus on clinical data. Engineers and lab technicians perform and analyze the tests according to these standards.
4. Adjudication Method for the Test Set
- Not Applicable. There is no adjudication in the sense of reconciling human expert opinions for clinical images or data. The results of mechanical tests are objective measurements against defined criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- Not Applicable. This is a physical spinal implant, not an AI/ML diagnostic or assistive device. No MRMC studies were performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not Applicable. This is a physical spinal implant; there is no algorithm or standalone performance.
7. The Type of Ground Truth Used
- Mechanical Test Standards and Predicate Device Performance: The "ground truth" for the non-clinical tests is based on the established performance characteristics and safety profiles of legally marketed predicate devices, as defined by FDA regulations for substantial equivalence, and adherence to relevant ASTM standards (ASTM F1717 and ASTM F1798).
8. The Sample Size for the Training Set
- Not Applicable. This is a physical device; there is no AI/ML model or "training set" in the context of machine learning. The device design and materials are based on engineering principles and knowledge of predicate devices.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no training set for an AI/ML model, this question is not relevant. The "ground truth" for developing the physical device is based on established biomechanical and medical understanding of spinal fixation and the performance of existing, cleared devices.
Ask a specific question about this device
(210 days)
SPINECRAFT LLC
The VELOX Anterior Cervical Plate System is interior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fision of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with:
· degenerative disc disease (as defined by neck pain of discogenion of the disc confirmed by patient history and radiographic studies),
- · spondylolisthesis,
· trauma (i.e. fractures or dislocations),
• tumors,
· deformity,
· pseudarthrosis.
· failed previous fusion,
· spinal stenosis.
The VELOX Anterior Cervical Plate System is intended for anterior screw fixation of the plate to the cervical spine. The fixation construct consists of a cervical plate that is attached to the vertebral body of the cervical spine with self-tapping bone screws using an anterior approach. The VELOX anterior cervical plate, screws and instruments were designed taking into consideration the whole procedure. Plates are available in a variety of lengths, addressing multiple levels of fixation. The VELOX plate incorporates vision ports that allow visualization of post-operative endplate/graft incorporation. Alignment notches on the cephalad and caudal ends of the plate facilitate precise midline placement and allow for Temporary Pin fixation of the plate. To accommodate normal cervical spine lordosis and, at the same time, minimize the need for additional plate contouring, the VELOX Anterior Cervical Plate comes with a pre-machined lordotic curve. Bone screws are available for fixed angle implantation in a variety of lengths.
The VELOX Anterior Cervical Plate is a medical device and is not subject to acceptance criteria in the same way an AI/ML diagnostic or prognostic device would be. Instead, its approval is based on demonstrating substantial equivalence to existing legally marketed predicate devices through non-clinical performance testing.
Here's an analysis of the provided information in the context of your request:
Acceptance Criteria and Device Performance (Derived from Substantial Equivalence Claim)
The "acceptance criteria" for a medical device like this are primarily demonstrating that its design, materials, and mechanical performance are equivalent or superior to predicate devices. The performance is reported in comparison to these predicates.
Acceptance Criteria Category | Reported Device Performance (VELOX ACP System) |
---|---|
Mechanical Performance (ASTM F1717-12) | |
Static Compression Bending | Results were equal or higher than predicate systems. |
Static Torsion | Results were equal or higher than predicate systems. |
Dynamic Compression Bending | Results were equal or higher than predicate systems. |
Explanation: In the context of a 510(k) submission for a traditional medical device, "acceptance criteria" are implied by the chosen predicate device's established safety and effectiveness. The manufacturer must demonstrate that the new device meets or exceeds the performance of these predicates.
Study Details (Non-Clinical)
-
Sample sizes used for the test set and data provenance:
- Test Set Sample Size: Not explicitly stated in terms of the number of specific test articles. However, the study references "testing included Static Compression Bending Tests, Static Torsion Tests and Dynamic Compression Bending Tests" according to ASTM F1717-12. The ASTM standard itself would specify the number of samples required for each test.
- Data Provenance: The testing was conducted by the manufacturer, SpineCraft, LLC, likely in a laboratory setting. The country of origin for the testing is implied to be within the USA, given the submitter's location in Westmont, IL, USA. This is a prospective non-clinical study designed for regulatory submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This is a non-clinical mechanical performance study, not a clinical study involving human patients or expert interpretation of data. "Ground truth" in this context refers to the measured physical properties and performance characteristics determined by the mechanical tests, not expert consensus on medical conditions.
-
Adjudication method for the test set:
- Not Applicable. As a non-clinical mechanical test, there is no expert adjudication process. The results are quantitative measurements against established engineering standards.
-
If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a mechanical device, not an AI/ML algorithm. No human readers or AI assistance are involved in its primary function or testing for regulatory approval.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a mechanical device, not an algorithm.
-
The type of ground truth used:
- Mechanical Test Results against ASTM Standards. The "ground truth" for this device's performance is derived from the objective measurements obtained through standardized mechanical testing (ASTM F1717-12). These results are then compared to the known performance of predicate devices.
-
The sample size for the training set:
- Not Applicable. This is a mechanical device, not an AI/ML algorithm that requires a training set. The design of the device is based on engineering principles and potentially prior experience with similar devices, but not a data-driven training set in the AI sense.
-
How the ground truth for the training set was established:
- Not Applicable. No training set was used.
Ask a specific question about this device
(92 days)
SPINECRAFT LLC
The APEX Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The APEX Spine System is also a sacraliliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis. trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
The purpose of this submission is to add x-large and uniplanar pedicle screws, open iliac connectors and open-closed side-by-side connectors to the APEX Spine System. The APEX Spine System X-Large Screws are similar to the APEX screws and are available in a variety of diameters and lengths and can be used of the components of the previously cleared APEX Spine System. The APEX Uniplanar Pedicle Screw is similar in principle to the APEX Polyaxial Pedicle Screw; however the design restricts motion to the sagittal plane. The Open Iliac Connectors and Open-Closed Side-by-Side Connectors are utilized when pedicle screws are not vertically aligned.
The provided document describes a 510(k) submission for the APEX Spine System Line Extension, which is a spinal fixation system. This type of submission relies on demonstrating substantial equivalence to predicate devices rather than proving safety and effectiveness through clinical trials with specific acceptance criteria related to a device's performance metrics like sensitivity, specificity, accuracy, etc.
Therefore, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" as typically applied to AI/ML or diagnostic devices does not apply in this context. This submission focuses on mechanical performance and substantial equivalence.
Here's an analysis based on the information provided, framed to address the spirit of your request:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a submission for a spinal implant system and not an AI/ML or diagnostic device, the acceptance criteria are based on mechanical testing standards and comparison to predicate devices, rather than performance metrics like sensitivity or specificity.
Acceptance Criteria (Mechanical Testing Standards) | Reported Device Performance |
---|---|
ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model (Static Compression Bending Tests, Static Torsion Tests, Dynamic Compression Bending Tests) | Results were equal or higher than the predicate systems. |
ASTM F1798: Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies used in Spinal Arthrodesis Implants (Static Axial Gripping, Static Axial Torque, Dynamic Flexion-Extension) | Results were equal or higher than the predicate systems. |
Substantial Equivalence: In terms of intended use, design, materials used, mechanical safety, and performance compared to predicate devices. | The APEX Spine System Line Extension was deemed substantially equivalent to the predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): Not explicitly stated in terms of number of screws or constructs, but the testing was performed on representative samples of the new components (x-large and uniplanar pedicle screws, open iliac connectors, and open-closed side-by-side connectors) added to the APEX Spine System. The ASTM standards dictate specific specimen requirements for each test.
- Data Provenance: The mechanical testing was performed in a laboratory setting, not on patient data. Thus, there is no "country of origin" or "retrospective/prospective" distinction in this context.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- Ground Truth for Mechanical Testing: For mechanical performance criteria, the "ground truth" is established by the ASTM standards themselves. These standards are developed by committees of engineers, scientists, and medical professionals. Compliance with these established standards demonstrates mechanical integrity.
- Experts: While not explicitly listed as "experts" for ground truth in the conventional sense, the underlying expertise comes from the materials scientists and biomechanical engineers who perform and interpret the mechanical tests, as well as the committees that develop the ASTM standards.
4. Adjudication Method for the Test Set
- Mechanical Testing: Adjudication is not applicable in the sense of expert review for diagnosis. The "adjudication" is based on comparing the quantitative results of the mechanical tests against the established performance criteria (e.g., strength, durability) and against the performance of the predicate devices. If the new device meets or exceeds the predicate device's performance, it's considered acceptable.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size
- No MRMC study was done. This type of study is relevant for diagnostic devices or AI algorithms where human readers interpret medical images. The APEX Spine System is a physical surgical implant.
6. If a Standalone (algorithm only without human-in-the-loop performance) study was done
- Not applicable. This concept pertains to AI algorithms. The APEX Spine System is a medical device.
7. The Type of Ground Truth Used
- Ground Truth: For this device, the "ground truth" is primarily defined by established mechanical engineering principles, material science standards (ASTM), and the demonstrated performance of clinically accepted predicate devices. The goal is to show that the new components perform at least as well as existing, cleared devices.
8. The Sample Size for the Training Set
- Not applicable. This submission is not for an AI/ML device, so there is no training set in the AI sense.
9. How the Ground Truth for the Training Set was Established
- Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.
Ask a specific question about this device
(117 days)
SPINECRAFT LLC
The APEX Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic. Iumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The APEX Spine System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative disc disease (defined as discogenic back pain with deqeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous, posterior approach with AIM Spine MIS instrumentation, the APEX Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, and/or lordosis), turnor, pseudoarthrosis, and failed prvious fusion in skeletally mature patients . Levels of fixation are for the thoracic. lumbar and sacral spine.
The purpose of this submission is to add cannulated screws to the APEX Spine System and add indications for use of the APEX Spine System in a posterior percutaneous approach and product code NKB. The APEX Spine System Cannulated Screws have exactly the same geometry of the APEX screws with the addition of a cannulation. The screws are available in a variety of diameters and lengths and can be used with the components of the previously cleared APEX Spine System.
Materials: Titanium alloy per ASTM F136
Function: The APEX Spine System is indicated for 1)non-cervical spinal fixation devices intended for posterior, non-pedicle fixation, 2)for non-cervical pedicle screw fixation and 3)hook fixation systems of the non-cervical spine, 4)for degenerative disc disease.
The APEX Spine System is a pedicle screw spinal system. The submission (K110906) addresses the addition of cannulated screws to the system and new indications for use in a posterior percutaneous approach.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Performance: |
- Static Compression Bending
- Static Torsion
- Dynamic Compression Bending | Met: Results were "equal or higher than the predicate systems" for all tests (ASTM F1717-09). |
| Material Composition: - Titanium alloy per ASTM F136 | Met: The device utilizes Titanium alloy per ASTM F136. |
| Biocompatibility: (Implicitly covered by substantial equivalence to predicates) | Met: Substantially equivalent to predicate devices with established biocompatibility. |
| Sterility: - Non-sterile, steam sterilized at hospital | Met: Device is non-sterile and intended for steam sterilization at the hospital. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated for specific tests. The document refers to "testing" and "results," implying a sufficient number of samples were tested to generate comparative data.
- Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted by the manufacturer or a contracted testing facility, following ASTM standards. This is a retrospective analysis in the context of device design and comparison to predicates.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not applicable. This submission focuses on the mechanical and material equivalence of a spinal implant system to predicate devices, not on diagnostic accuracy requiring expert interpretation or ground truth establishment in a clinical sense.
4. Adjudication Method for the Test Set
- Not applicable. As noted above, this submission does not involve human readers or clinical interpretation that would require an adjudication method. The "adjudication" here is the comparison of mechanical test results against established predicate device performance and ASTM standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. This type of study is not applicable to a spinal implant system focused on mechanical performance and material composition.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Not applicable. This device is a physical spinal implant, not an AI algorithm.
7. The Type of Ground Truth Used
- Engineering/Performance Standards: The "ground truth" used for testing was primarily based on:
- ASTM F1717-09: "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."
- Predicate Device Performance: The performance of previously cleared spinal systems (e.g., APEX Spine System (K062513), Viper Spine System (K090648), Polaris Spinal System (K103393), Synthes Pangea System (K052123)).
- Material Specifications: ASTM F136 for Titanium alloy.
8. The Sample Size for the Training Set
- Not applicable. There is no "training set" in the context of this kinesiologic and material equivalence submission for a physical medical device. The device's design is based on established engineering principles and prior device knowledge, not a data-driven training process.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no training set, this question is not relevant. The "ground truth" in development of such a device is rooted in biomechanical principles, material science, and the established safety and effectiveness of predicate devices.
Summary of the Study Proving Acceptance Criteria:
The study summarized in K110906 is a non-clinical bench testing study designed to demonstrate the substantial equivalence of the modified APEX Spine System (with added cannulated screws and percutaneous approach indication) to its predicate devices.
- Study Name: Non-clinical Test Summary
- Study Type: Bench Testing, Mechanical Performance Comparison.
- Methodology:
- Tests were conducted according to ASTM F1717-09, "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."
- Specific tests included: Static Compression Bending, Static Torsion, and Dynamic Compression Bending.
- The "acceptance criteria" here were that the results of these tests for the new device configuration should be "equal or higher" than those of the predicate devices.
- Results: The new device configuration for the APEX Spine System with cannulated pedicle screws met these performance criteria, with results reported as "equal or higher than the predicate systems."
- Conclusion: Based on these non-clinical tests, the manufacturer concluded that the APEX Spine System with the added cannulated pedicle screws is substantially equivalent to the predicate devices in terms of intended use, design, materials, performance, and function. No clinical studies were performed for this submission, as substantial equivalence was demonstrated through non-clinical testing and comparison to legally marketed predicate devices.
Ask a specific question about this device
Page 1 of 2