Search Filters

Search Results

Found 37 results

510(k) Data Aggregation

    K Number
    K211910
    Manufacturer
    Date Cleared
    2021-07-19

    (28 days)

    Product Code
    Regulation Number
    870.5175
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Recommended for balloon atrioseptostomy, an accepted technique in most pediatric cardiology centers for the palliation of several congenital cardiac defects. Balloon atrioseptostomy is performed in conjunction with diagnostic cardiac catheterization and has been carried out after the diagnosis of several congenital cardiac defects: transposition of the great arteries, total anomalous pulmonary venous drainout pulmonary obstruction, tricuspid atresia, mitral stenosis, mitral atresia, and pulmonary atresia with intact ventricular septum.

    Device Description

    Not Found

    AI/ML Overview

    There is no information in the provided text regarding acceptance criteria or a study that proves a device meets specific acceptance criteria. The document is an FDA 510(k) clearance letter for the Z-6 Atrioseptostomy Catheter, confirming its substantial equivalence to a predicate device. It defines the indications for use but does not contain details about a performance study, acceptance criteria, sample sizes, expert involvement, or ground truth establishment.

    Therefore, I cannot provide the requested information in a table or fulfill the other points as the necessary data is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190974
    Manufacturer
    Date Cleared
    2019-05-14

    (29 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Recommended for introduction of balloons, catheters and other diagnostic and interventional devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.

    Device Description

    The NuMED D'VILL Introducer is recommended for introduction of balloons, catheters and other diagnostic and interventional devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The introducer consists of a dilator and sheath with hemostasis valve and side port on the proximal end of the sheath assembly. There is a single image band embedded in the distal end of the sheath tubing for imaging purposes. The sheath is Pebax braided with stainless steel and a PTFE liner and will accommodate a 0.035" guidewire. The dilator is LDPE. The D'VILL is available in 10, 12 and 14F sizes in both 30 and 85cm lengths, as well as, 12 and 14F in a 65 cm length.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "D'Vill Introducer." This submission is a "Special 510(k)" to obtain marketing clearance for an additional size (65cm length) of an already cleared device, the D'VILL Introducer – K171206.

    The crucial information to understand here is that a Special 510(k) is typically used for modifications to a manufacturer's own legally marketed device where the modification does not alter the fundamental scientific technology of the device and does not raise new questions of safety and effectiveness. This often means that the performance testing is focused on demonstrating that the new modification performs equivalently to the existing device within the established parameters, rather than proving the overall safety and effectiveness of a completely new device from scratch.

    Therefore, the "acceptance criteria" and "study" described in this document are not a clinical study with human patients, but rather bench testing to demonstrate that the new size performs similarly to the predicate device.

    Here's the breakdown based on the provided document:


    Acceptance Criteria and Device Performance for the D'Vill Introducer (New 65cm Length)

    The device in question is an additional size (65cm length) of an already cleared product. The performance testing focused on demonstrating substantial equivalence to the predicate device rather than establishing entirely new safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Test NameReported Device Performance
    Surface InspectionAll tests met their acceptance criteria and specifications.
    Size Designation – Sheath and DilatorAll tests met their acceptance criteria and specifications.
    Freedom from Leakage – Sheath and Hemostasis ValveAll tests met their acceptance criteria and specifications.
    Peak Tensile Force – 3 locationsAll tests met their acceptance criteria and specifications.
    Strength of Union Between Hub and DilatorAll tests met their acceptance criteria and specifications.
    Kink / FlexibilityAll tests met their acceptance criteria and specifications.
    Luer Testing (stress cracking, resistance to separation and overriding)All tests met their acceptance criteria and specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the precise number of units tested for each performance test. It only states, "A complete list of tests performed are provided below. All tests met their acceptance criteria and specifications."
    • Data Provenance: This is bench testing of a medical device, not a human study. The data is prospective, generated from testing newly manufactured devices. The country of origin for the testing is not explicitly stated, but given the FDA submission, it would typically be conducted in a controlled environment, often by the manufacturer or a contracted lab.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • This information is not applicable as the ground truth for this type of device modification (addition of a new size) is established through engineering and materials testing standards, not clinical expert consensus. The "ground truth" is adherence to established physical and mechanical performance specifications.

    4. Adjudication Method for the Test Set

    • Not applicable. Bench testing results are typically measured against predetermined specifications, not adjudicated by experts in the context of diagnostic interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. An MRMC study is relevant for diagnostic devices that involve human interpretation of images or other data. This device is a catheter introducer, a physical medical device, and its performance is evaluated through engineering and materials testing, not diagnostic accuracy by human readers.

    6. Standalone (Algorithm Only) Performance

    • Not applicable. This is a physical medical device, not an AI algorithm.

    7. Type of Ground Truth Used

    • The ground truth used for this type of device is based on engineering specifications, material properties, and established industry standards for the performance of introducer catheters. This includes measurements of physical dimensions, force tolerance, leak resistance, and biocompatibility (by equivalence to predicate materials).

    8. Sample Size for the Training Set

    • Not applicable. This device is not an AI algorithm or a diagnostic tool that requires a "training set" in the machine learning sense. The "training" for the device's development involves engineering design, material selection, and manufacturing processes honed over time, implicitly incorporating knowledge from previous designs and testing.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. See point 8. The "ground truth" (or design principles) for manufacturing such a device is established through biomedical engineering principles, regulatory standards (like ISO standards for medical devices), and a history of manufacturing similar devices.

    Summary of the Study:

    The "study" was a series of in-vitro (bench) performance tests designed to demonstrate that the new 65cm length of the D'Vill Introducer performs equivalently to the existing, predicate D'Vill Introducer (K171206). The tests covered physical dimensions, mechanical strength, and functional integrity (e.g., leakage, kink resistance). All tested parameters met the prespecified acceptance criteria, leading to the conclusion of substantial equivalence for the new size. Biocompatibility was addressed by stating that the materials and manufacturing processes are identical to the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171206
    Manufacturer
    Date Cleared
    2017-10-20

    (178 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Recommended for introduction of balloons, catheters and other diagnostic and interventional devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.

    Device Description

    The NuMED D'VILL Introducer is recommended for introduction of balloons, catheters and other diagnostic and interventional devices. The introducer consists of a dilator, and sheath with hemostasis valve and side port on the proximal end of the sheath assembly. There is a single image band embedded in the distal end of the sheath tubing for imaging purposes. The sheath is Pebax braided with stainless steel and a PTFE liner and will accommodate a 0.035" guidewire. The dilator is LDPE. The D'VILL is available in 10, 12 and 14F sizes and 30 and 85cm lengths.

    AI/ML Overview

    The document provided is a 510(k) summary for the D'Vill Introducer, a medical device. It describes the device, its intended use, and substantial equivalence to predicate devices, but it does not contain information about a study proving the device meets specific acceptance criteria based on performance data.

    The "Performance Testing" section on page 4 lists several tests performed:

    • Surface Inspection
    • Size Designation – Sheath and Dilator
    • Freedom from Leakage – Sheath and Hemostasis Valve
    • Peak Tensile Force - 3 locations
    • Strength of Union Between Hub and Dilator
    • Dimensional Requirements – Sheath, Stopcock, and Dilator
    • Luer Hub Testing
    • Kink/Flexibility Testing – Sheath and Dilator
    • Sterilization Testing
    • Biocompatibility Evaluation – Short duration contact with circulating blood (< 24 hours)
    • Shelf Life Testing

    The document states, "All tests met their acceptance criteria and specifications." However, it does not provide the specific acceptance criteria for each test or the reported results.

    Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance, nor can I answer questions related to sample size, ground truth, experts, or comparative effectiveness studies, as this information is not present in the provided text.

    The information requested, such as specific performance metrics (e.g., leakage rate, tensile strength values, specific dimensions) and their acceptance thresholds, as well as the actual test results, would typically be found in a more detailed test report or design verification document, which is not included in this 510(k) summary. These summaries focus on demonstrating substantial equivalence rather than providing detailed raw performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161451
    Manufacturer
    Date Cleared
    2016-09-22

    (119 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Not Found

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving a device meets those criteria. The document is an FDA 510(k) clearance letter for the NuMED, Inc. BIB Stent Placement Catheter. It indicates that the device has been found substantially equivalent to a predicate device, allowing it to be marketed, but does not include performance data or details of a study.

    The sections you've highlighted regarding "Indications for Use" and what appears to be garbled text or non-English characters do not provide the requested information. Therefore, I cannot complete the table or answer the specific questions about acceptance criteria or a study based on the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160598
    Manufacturer
    Date Cleared
    2016-06-20

    (110 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Recommended for temporary occlusion of the aorta.

    Device Description

    The REBOA Balloon Occlusion Catheter is a coaxial catheter recommended for temporary occlusion of the aorta. The outer body is made of polymeric tubing, and the inner tubing is comprised of a multi-layer extrusion of polyamide (Vestamid PA12) that surrounds a braid of 304 LV Stainless Steel. The catheter features a proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon. This balloon is of the non-compliant variety and is designed to insert through the smallest possible introduction sleeve. The through lumen terminates at the tip of the catheter, and will accept the passage of the appropriate guidewire. All catheter sizes will have radiopaque platinum marker band(s), centered or under the balloon shoulders, for placement using fluoroscopy. There are also markings on the catheter shaft to aid in placement. The catheter is packaged in a polyethylene loop and is double packed in two heat sealed Tyvek pouches.

    AI/ML Overview

    This document describes testing and acceptance criteria for the REBOA Catheter, a device intended for temporary occlusion of the aorta.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test PerformedAcceptance CriteriaREBOA ResultsPredicate Device - PTS-X Results
    Visual InspectionThe catheters shall be free from contamination, discoloration, and any form of damage that could impact the proper functioning of the device.All catheters were visually inspected without any anomalies.All catheters were visually inspected without any anomalies.
    Balloon Preparation TestEach catheter shall be prepped per the procedure without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.
    Diameter and Profile TestThe balloon diameter at rated burst pressure / volume shall be within +/- 10% of the labeled balloon diameter and the samples should fit through the selected introducer with no problems.All catheters met the acceptance criteria.All catheters met the acceptance criteria.
    Balloon DistensibilityThe results must demonstrate that the balloon diameter are within +/- 10% of the labeled diameter at the RBP / rated volume and will not be significantly increased at increasingly higher pressures.All data obtained demonstrates that the balloon diameter is within +/- 10% of the labeled diameter at the rated volume. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.All data obtained demonstrates that the balloon diameter is within +/- 10% of the labeled diameter at the RBP. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.
    Balloon Minimum Burst Strength / VolumeThe results must show statistically that with at least 95% confidence, 99.9% of the balloons will not burst at or below the maximum recommended rated burst pressure / volume.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Repeated Balloon Inflation (Balloon Fatigue) TestNo breaks allowed.No Breaks.No breaks.
    Balloon Inflation/Deflation TestInflation achieved in less than 12 seconds and deflation achieved in less than 59 seconds.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Balloon Deflatability TestThere should be no interference with balloon deflation.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Tip Pull and Torque TestMust withstand at least 8 turns without breaking.No breaks.No breaks.
    Bond Strength TestAll bonds must withstand at least 8.9 Newtons / 2 lbs. of pull strength.All bonds met the established acceptance criteria.All bonds met the established acceptance criteria.
    Catheter Body Maximum Pressure TestAll samples must withstand 700 psi.All samples met the established acceptance criteria.All samples met the established acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes used for each in-vitro test. It mentions "All catheters" or "All samples" met the criteria, suggesting that a representative sample was tested for each evaluation.

    The data provenance is from in-vitro testing, meaning the tests were conducted in a laboratory setting, not on human subjects. There is no information regarding country of origin, retrospective or prospective data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable as the "ground truth" for the in-vitro tests conducted on the REBOA Catheter is based on objective, measurable physical and mechanical properties, not expert interpretation of diagnostic data.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests are objective physical measurements, not requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of human readers improve with AI vs without AI assistance:

    This information is not applicable as the REBOA Catheter is a medical device, not an AI diagnostic tool that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable for the same reason as point 5.

    7. The Type of Ground Truth Used:

    The ground truth for the in-vitro tests is based on objective, measurable physical and mechanical properties (e.g., diameter, pressure resistance, burst strength, inflation/deflation times, pull strength).

    8. The Sample Size for the Training Set:

    This information is not applicable as there is no mention of a "training set" in the context of this device. The testing described is for performance verification of the final device, not for training a model or algorithm.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable for the same reason as point 8.

    Summary of the Study:

    The study detailed in the 510(k) Summary is an in-vitro performance verification study of the REBOA Catheter. It compares the performance of the REBOA Catheter against pre-defined acceptance criteria and, in some cases, against a predicate device (PTS-X PTV Catheter and Fogarty Occlusion Catheter). The study aims to demonstrate that the REBOA Catheter meets the necessary physical and mechanical specifications for its intended use, establishing substantial equivalence to legally marketed predicate devices. The "study" here refers to a series of laboratory tests, not a clinical trial or an AI algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160889
    Manufacturer
    Date Cleared
    2016-05-20

    (50 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for CP Stent / Covered CP Stent placement in vessels over 8mm in diameter.

    Device Description

    The NuMED BIB® Stent Placement Catheter Model 420 is indicated for CP Stent / Covered CP Stent placement in vessels over 8mm in diameter. The catheter is triaxial in construction with two lumens being used to inflate the balloons while one lumen is being used for tracking over a guidewire. The purpose of the double balloon catheter is to apply an incremental inflation for the purpose of dilating a stent. The inner balloon provides initial expansion of the stent and also acts as a tool to hold the stent on the catheter prior to the outer balloon being inflated. The outer balloon is then inflated providing the remainder of the expansion. There are radiopaque platinum marker bands under the balloon shoulders for placement using fluoroscopy. The balloon material is clear. The catheter balloon diameters are stamped onto the inflation extensions and are labeled with balloon diameter x balloon length and the catheter lot number.

    AI/ML Overview

    The provided text describes a medical device, the NuMED BIB® Stent Placement Catheter, and its substantial equivalence to a predicate device. It includes performance data from in vitro testing and a clinical trial. However, the document does not explicitly state acceptance criteria in a quantitative format, nor does it present the study results specifically framed as meeting such criteria. Instead, it leverages prior testing and a clinical trial to demonstrate substantial equivalence.

    Based on the provided text, here's an attempt to extract and infer the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    As explicit acceptance criteria with numerical targets are not stated, this table will present the key performance indicators from the leveraged studies, which implicitly served as the basis for performance evaluation for substantial equivalence.

    Acceptance Criterion (Inferred from study objectives)Reported Device Performance
    Bench Testing (In Vitro)
    Visual inspectionLeveraged from K050857 and P150028; assumed to have passed.
    Balloon Preparation TestLeveraged from K050857 and P150028; assumed to have passed.
    Diameter and Profile TestLeveraged from K050857 and P150028; assumed to have passed.
    Balloon DistensibilityLeveraged from K050857 and P150028; assumed to have passed.
    Balloon Minimum Burst Strength / VolumeLeveraged from K050857 and P150028; assumed to have passed.
    Repeated Balloon Inflation (Balloon Fatigue) TestLeveraged from K050857 and P150028; assumed to have passed.
    Balloon Inflation/Deflation TestLeveraged from K050857 and P150028; assumed to have passed.
    Balloon Deflatability TestLeveraged from K050857 and P150028; assumed to have passed.
    Tip Pull and Torque TestLeveraged from K050857 and P150028; assumed to have passed.
    Bond Strength TestLeveraged from K050857 and P150028; assumed to have passed.
    Catheter Body Maximum Pressure TestLeveraged from K050857 and P150028; assumed to have passed.
    Biocompatibility Testing
    Cytotoxicity (L929)Leveraged from K050857 and P150028; assumed to have passed.
    Sensitization (ISO Guinea Pig Maximization Test)Leveraged from K050857 and P150028; assumed to have passed.
    Irritation (ISO Rabbit Intracutaneous Reactivity)Leveraged from K050857 and P150028; assumed to have passed.
    Systemic Toxicity (ISO Mouse Systemic Injection)Leveraged from K050857 and P150028; assumed to have passed.
    Material-Mediated Pyrogenicity (USP Rabbit Pyrogenicity)Leveraged from K050857 and P150028; assumed to have passed.
    Hemocompatibility (Hemolysis)Leveraged from K050857 and P150028; assumed to have passed.
    Clinical Performance (Primary Safety Endpoints)
    Rate of serious adverse events within 30 days8.9%
    Rate of post-procedure paradoxical hypertension7.5%
    Clinical Performance (Primary Effectiveness Endpoints)
    Reduction in arm-leg systolic blood pressure gradient from baseline to 12 months30 ± 22 mmHg
    Length of hospital stay1.1 ± 0.3 days
    CP Stent successfully mounted and implanted with acceptable rate of procedure-related adverse events and clinically acceptable systolic blood pressure gradientsDemonstrated

    2. Sample size used for the test set and the data provenance

    • Test Set Description: The clinical study is referred to as the COAST trial.
    • Sample Size: 112 patients were enrolled.
    • Data Provenance: The COAST trial was a "prospective, multi-center, single arm study." While the specific country of origin is not explicitly stated, "multi-center" implies data from several locations, typically within a regulatory region (e.g., USA for FDA approval).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The provided text only refers to clinical trial data (COAST study) for safety and effectiveness. It does not mention the use of experts to establish a "ground truth" for the test set in the context of diagnostic performance or image interpretation often associated with AI/ML devices. Therefore, this question is not applicable to the submitted document. The clinical outcomes served as the definitive measure of device performance.

    4. Adjudication method for the test set

    The document does not describe an adjudication method for the test set, as no "ground truth" establishment by multiple experts is discussed for diagnostic purposes. Clinical trial outcomes are typically assessed by study investigators and monitored, but not through an adjudication process in the sense of resolving discrepancies between expert opinions on a data point.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such MRMC comparative effectiveness study is mentioned. The device is a physical medical device (stent placement catheter), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical medical device; it does not have a standalone algorithm or AI component.

    7. The type of ground truth used

    For the clinical study (COAST trial), the "ground truth" for evaluating the device's performance was:

    • Clinical Outcomes/Events:
      • Rates of serious adverse events within 30 days.
      • Rates of post-procedure paradoxical hypertension.
      • Reduction in arm-leg systolic blood pressure gradient from baseline to 12 months.
      • Length of hospital stay.
    • Successful implantation: The text states, "The results demonstrated that the CP Stent could be successfully mounted and implanted using the BiB Stent Placement Catheter..."

    For the bench and biocompatibility testing, the "ground truth" was established based on pre-defined engineering specifications, material standards, and biological safety guidelines, which the device components were expected to meet.

    8. The sample size for the training set

    The document does not describe a "training set" in the context of AI/ML. The device is a physical medical device. The "leveraged" data from K050857 and P150028 could be considered as prior performance data used for development and comparison, but not a "training set" in the AI sense.

    9. How the ground truth for the training set was established

    As there is no "training set" for an AI/ML algorithm described, this question is not applicable. The prior performance data from K050857 and P150028 were established through engineering tests, biocompatibility assessments, and clinical evaluation as per regulatory requirements for medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131869
    Manufacturer
    Date Cleared
    2013-08-08

    (45 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K122012
    Device Name
    Z-MED Z-MED II
    Manufacturer
    Date Cleared
    2012-10-04

    (86 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Balloon Aortic Valvuloplasty

    Device Description

    The NuMED Z-MED™ catheter is a coaxial catheter recommended for Balloon Aortic Valvuloplasty. The catheter's inner and outer shafts are constructed of polyamide tubing. The catheter features a molded proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. This balloon is of the non-compliant variety. Both the shaft size and the guidewire size vary according to balloon diameter. The distal lumen terminates at the tip of the catheter and will accept the passage of the appropriate guidewire. The lumen has radiopaque platinum marker band(s), centered or under the balloon shoulders, for placement using fluoroscopy. The catheter is white in color and the balloon material is clear. The catheter balloon diameter and name is stamped onto the Y sleeve and the balloon extension is labeled with the catheter specifications and lot number. The catheter is packaged in a polyethylene sheath and is double packed in two heat sealed Tyvek pouches. The NuMED Z-MED II™ catheter is a coaxial catheter recommended for Balloon Aortic Valvuloplasty. The catheter's inner and outer shafts are constructed of polyamide tubing. The catheter features a molded proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. This balloon is of the non-compliant variety. Both the shaft size and the guidewire size vary according to balloon diameter. The distal lumen terminates at the tip of the catheter and will accept the passage of the appropriate guidewire. The lumen has radiopaque platinum marker band(s), centered or under the balloon shoulders, for placement using fluoroscopy. The catheter is white in color and the balloon material is clear. The catheter balloon diameter and name is stamped onto the Y sleeve and the balloon extension is labeled with the catheter specifications and lot number. The catheter is packaged in a polyethylene sheath and is double packed in two heat sealed Tyvek pouches.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance Study

    The K122012 submission for the Z-MED and Z-MED II Catheters primarily relies on in-vitro (bench) testing and a review of existing clinical data for safety and effectiveness. No specific algorithms or human-in-the-loop performance studies are described as this is a medical device (catheter), not an AI/software device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text details in-vitro test results for the Z-MED and Z-MED II catheters, as well as two predicate devices (NuCLEUS-X and NuCLEUS). The table below focuses on the Z-MED and Z-MED II results as the subject devices.

    Test PerformedAcceptance CriteriaZ-MED ResultsZ-MED II Results
    Visual InspectionThe catheters shall be free from contamination, discoloration, and any form of damage that could impact the proper functioning of the device.All catheters were visually inspected without any anomalies.All catheters were visually inspected without any anomalies.
    Balloon Preparation TestEach catheter shall be prepped per the procedure without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.
    Diameter and Profile TestThe balloon diameter at rated burst pressure shall be within +/- 10% of the labeled balloon diameter and the samples should fit through the selected introducer with no problems.All catheters met the acceptance criteria.All catheters met the acceptance criteria.
    Balloon DistensibilityThe results must demonstrate that the balloon diameter is within +/- 10% of the labeled diameter at the RBP and will not be significantly increased at increasingly higher pressures.All data obtained demonstrates that the balloon diameter is within +/- 10% of the labeled diameter at the RBP. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.All data obtained demonstrates that the balloon diameter is within +/- 10% of the labeled diameter at the RBP. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.
    Balloon Minimum Burst StrengthThe results must show statistically that with at least 95% confidence, 99.9% of the balloons will not burst at or below the maximum recommended rated burst pressure.(Specific pressures listed for various balloon sizes, all indicating performance at or above the acceptance criteria)(Specific pressures listed for various balloon sizes, all indicating performance at or above the acceptance criteria)
    Repeated Balloon Inflation (Balloon Fatigue) TestNo breaks allowed.No Breaks.No Breaks.
    Balloon Inflation/Deflation TestInflation achieved in less than 12 seconds and deflation achieved in less than 20 seconds.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Balloon Deflatability TestThere should be no interference with balloon deflation.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Tip Pull and Torque TestMust withstand at least 10 turns without breaking.No breaks.No breaks.
    Bond Strength TestAll bonds must withstand at least 3 lbs. of pull strength.All bonds met the established acceptance criteria.All bonds met the established acceptance criteria.
    Catheter Body Maximum Pressure TestAll samples must withstand 30 ATM (400psi).>30 ATM>30 ATM

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for In-Vitro Testing: The document does not explicitly state the numerical sample size for each in-vitro test (e.g., how many catheters were visually inspected or burst tested). It refers to "All catheters" or "All samples" meeting criteria, indicating that a sufficient number were tested.
    • Data Provenance (Clinical):
      • Country of Origin: The clinical evaluation report was prepared to meet the requirements of the Medical Device Directive MOD 93/42/EEC (European Union standard). The clinical use of the Z-MED and Z-MED II catheters for BAV and BMV is mentioned in "those countries where they are used," and for BPV in "the United States."
      • Retrospective/Prospective: The clinical information gathered appears to be a retrospective review of existing clinical literature, device history, previous sales, and reported complaints. It references "Numerous clinical studies reported in literature" and "200 patients" without incidents from BAV prior to TAVI.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and not provided in the context of this submission. The tests described are in-vitro bench tests for a physical medical device (catheter), which do not typically involve experts establishing ground truth in the way it's done for diagnostic AI/software. The ground truth for in-vitro tests is based on objective physical measurements and engineering specifications.

    For the clinical context, the "ground truth" refers to observations from actual patient use, as reported in clinical literature and through device monitoring. No specific number or qualification of experts reviewing patient outcomes is provided beyond the implicit understanding that medical professionals conducted the procedures and reported outcomes.

    4. Adjudication Method for the Test Set

    This is not applicable for the in-vitro bench testing described. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or for establishing ground truth in cases where there's variability in interpretation (e.g., image analysis by multiple radiologists). The in-vitro tests are objective physical measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study is not applicable and not mentioned. This type of study is relevant for diagnostic devices, especially those involving human interpretation of data, often comparing human performance with and without AI assistance. The Z-MED and Z-MED II are interventional catheters, not diagnostic devices or AI software. There's no AI component mentioned to compare human readers with or without.

    6. Standalone (Algorithm Only) Performance Study

    This is not applicable and not mentioned. The devices are physical catheters, not algorithms or software. Therefore, there is no standalone (algorithm only) performance to evaluate.

    7. Type of Ground Truth Used

    • In-Vitro Testing: The ground truth for the in-vitro tests is established by the predefined engineering specifications and acceptance criteria outlined in the table (e.g., balloon diameter within +/- 10% of labeled, no breaks, inflation/deflation times). These are objective, measurable physical parameters.
    • Clinical Evaluation: The "ground truth" for clinical safety and performance is based on real-world patient outcomes and reported incidents/adverse events from existing clinical use and literature reviews.

    8. Sample Size for the Training Set

    This is not applicable. The Z-MED and Z-MED II are physical medical devices undergoing traditional bench testing and clinical review, not AI models that require training sets.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121859
    Device Name
    TYSHAK NUCLEUS
    Manufacturer
    Date Cleared
    2012-08-31

    (66 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis.
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention.
    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving device performance. The document is an FDA 510(k) clearance letter for the NuMED Inc. Tyshak NuCLEUS™ Catheter, which primarily addresses regulatory classification and substantial equivalence to predicate devices, along with indications for use. It does not include details on performance studies, acceptance criteria, sample sizes, expert qualifications, or ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN080015
    Manufacturer
    Date Cleared
    2012-06-11

    (1265 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NuMED NuCLEUS and NuCLEUS-X BAV Catheters are indicated for Balloon Aortic Valvuloplasty.

    Device Description

    The NuCLEUS BAV and the NuCLEUS-X BAV are both coaxial catheters, with the NuCLEUS-X currently cleared in the United States under K081680 for Balloon Pulmonic Valvuloplasty (BPV). The NuCLEUS BAV is currently only marketed outside of the United States. The outer body of both devices is made of {Redacted as b4} tubing, while the inner shaft (or tubing) material is the only difference between the two devices. In the NuCLEUS-X Catheter, the inner tubing is comprised of {Redacted as b4}while in the NuCLEUS Catheter, the inner tubing is comprised of {Redacted as b4} The sponsor states that the {Redacted as b4}makes it less likely to stretch when force is used. From an engineering perspective, in terms of composition, testing, and intended use, FDA believes there is no significant difference in these technological characteristics between the inner tubing of the two catheters based upon the non-clinical performance testing results. Furthermore, these two devices would be expected to perform in a similar manner in the clinical setting during balloon aortic valvuloplasty procedures. The catheters also feature a molded proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. The balloon is designed with a waist formed into the middle of the balloon to allow accurate balloon placement. Upon reaching a specified pressure, the waist will expand to the rated balloon diameter and dilate the valve to the rated diameter. The distal lumen terminates at the tip of the catheters and will accept the passage of the 0.035" guidewire. This lumen has 3 radiopaque platinum marker bands. One under each of the balloon shoulders and one located at the "waist" or center of the balloon for placement using fluoroscopy. The catheters are packaged in a {Redacted as b4} sheath and double packed in two {Redacted as b4} sealed Tyvek pouches.

    AI/ML Overview

    This document describes the regulatory evaluation of the NuMED NuCLEUS and NuCLEUS-X BAV Catheters. The assessment includes non-clinical bench studies and the leveraging of extensive clinical experience with similar devices and the BAV procedure in general.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and performance data for the NuMED NuCLEUS-X and NuCLEUS BAV Catheters are primarily derived from non-clinical bench testing:

    Test PerformedAcceptance CriteriaNuCLEUS-X BAV ResultsNuCLEUS BAV Results
    Visual InspectionCatheters shall be free from contamination, discoloration, and any form of damage that could impact the proper functioning of the device.All catheters were visually inspected without any anomalies.All catheters were visually inspected without any anomalies.
    Balloon Preparation TestEach catheter shall be prepped per the procedure without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.
    Diameter and Profile TestThe balloon diameter at rated burst pressure shall be within +/- 10% of the labeled balloon diameter and the samples should fit through the selected introducer with no problems.All catheters met the acceptance criteria.All catheters met the acceptance criteria.
    Balloon DistensibilityThe results must demonstrate that the balloon diameter is within +/- 10% of the labeled diameter at the RBP and will not be significantly increased at increasingly higher pressures.All data obtained demonstrated that the balloon diameter is w/in +/-10% of the labeled diameter at the RBP. All data obtained demonstrated that the diameter of the balloons will not be significantly increased at increasingly higher pressures.All data obtained demonstrated that the balloon diameter is within +/-10% of the labeled diameter at the RBP. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.
    Repeated Balloon Inflation (Balloon Fatigue Test)No breaks allowed.30 Samples - no breaks30 Samples - no breaks
    Balloon Minimum Burst StrengthThe results must show statistically that with at least 95% confidence, 99.9% of the balloons will not burst at or below the maximum recommended rated burst pressure.20 samples of smallest diameter-shortest length, smallest diameter-longest length, largest diameter-shortest length, largest diameter-longest length, and all diameters in between (various sizes listed).20 samples of smallest diameter-shortest length, smallest diameter-longest length, largest diameter-shortest length, largest diameter-longest length, and all diameters in between (various sizes listed).
    Balloon Inflation/Deflation TestInflation achieved in less than 12 seconds and deflation achieved in less than 20 seconds.Twenty samples of all Diameters. All catheters met the established acceptance criteria.Twenty Samples of all Diameters. All catheters met the established acceptance criteria.
    Balloon Inflstability TestThere should be no interference with balloon deflation.Twenty Samples of all Diameters. All catheters met the established acceptance criteria.Twenty Samples of all Diameters. All catheters met the established acceptance criteria.
    Tip Pull and Torque TestMust withstand at least 10 turns without breaking.Twenty random samples of each shaft size. No breaks reported.Twenty random samples of each shaft size. No breaks reported.
    Bond Strength TestAll bonds must withstand at least 3 lbs. of pull strength.Twenty samples of each shaft size. All bonds met the established acceptance criteria.Twenty samples of each shaft size. All bonds met the established acceptance criteria.
    Catheter Body Maximum Pressure TestAll samples must withstand 30 ATM (400psi).Twenty samples of each guidewire size. >30ATMTwenty samples of each guidewire size. > 400 psi

    Note: The document references specific sizes for the Balloon Minimum Burst Strength for each device, which are too extensive to list individually in this summary table, but they are detailed in the provided text.

    2. Sample Size Used for the Test Set and Data Provenance

    For the formal performance testing (bench studies), samples were taken across various models and sizes of both the NuCLEUS-X and NuCLEUS catheters.

    • Visual Inspection, Balloon Preparation, Diameter & Profile, Balloon Distensibility, Balloon Inflation/Deflation, Balloon Inflstability, Bond Strength, Catheter Body Max Pressure: "All catheters" or "Twenty samples of all Diameters" or "Twenty samples of each guidewire size" were tested and met the criteria.
    • Repeated Balloon Inflation (Balloon Fatigue Test): 30 samples for both NuCLEUS-X and NuCLEUS.
    • Balloon Minimum Burst Strength: 20 samples for different configurations (smallest diameter-shortest length, smallest diameter-longest length, largest diameter-shortest length, largest diameter-longest length, and all diameters in between) for both NuCLEUS-X and NuCLEUS.
    • Tip Pull and Torque Test: Twenty random samples of each shaft size for both NuCLEUS-X and NuCLEUS.

    Data Provenance: The non-clinical (bench) data was generated by the manufacturer, NuMED, Inc. The clinical data is based on existing, published literature and clinical experience from various sources globally. The document does not specify a country of origin for the clinical data in most cases, but some references mentioned (e.g., European Heart Journal, Revista Espanola de Cardiogia) indicate international provenance. The clinical data is retrospective, drawing from previously documented clinical studies and experiences.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For the clinical evaluation, the "ground truth" and assessment of the devices and BAV procedure were established by:

    • Sponsor-provided literature: NuMED provided "two clinical reports containing extensive published literature" summarizing the clinical use of their devices.
    • FDA Reviewers: The FDA specifically states, "The specifically-cited clinical experience using the NuMED NuCLEUS family of catheters for BAV is a scientifically sound approach that was agreed upon by the scientific and clinical reviewers of this de novo." These reviewers are presumably experts in cardiology and medical device evaluation within the FDA, though their specific qualifications (e.g., years of experience as cardiologists) are not detailed in this document.

    For the bench testing, the ground truth is based on engineering specifications and direct physical measurements, not expert consensus on qualitative data.

    4. Adjudication Method for the Test Set

    For the bench testing, no "adjudication method" in the sense of expert review consensus is mentioned. The tests have clear pass/fail acceptance criteria demonstrated by direct measurements.

    For the clinical evaluation, the "adjudication" was effectively performed by the FDA's "scientific and clinical reviewers" who assessed the provided literature, including the NuMED NuCLEUS family of catheters, and determined that the body of evidence supported the safety and effectiveness. This is implicit in the FDA's "FDA Comment" sections on each cited paper. No formal N+1 reader adjudication scheme is described regarding the clinical literature review.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This document describes the evaluation of a physical medical device (balloon catheter) for regulatory approval, not an AI-powered diagnostic or interpretive tool. Therefore, no MRMC comparative effectiveness study involving human readers with or without AI assistance was conducted or is relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. As mentioned above, this is a physical medical device, not an algorithm. The "standalone performance" is represented by the extensive bench testing which evaluates the device's physical properties and functionality.

    7. The Type of Ground Truth Used

    • Bench Testing: The ground truth for bench studies is based on engineering specifications and measurable physical properties. For example, balloon diameter within +/- 10% of labeled, inflation/deflation times, burst pressure, bond strength, and visual inspection for anomalies. These are objective measures against predefined criteria.
    • Clinical Evaluation: The ground truth for the clinical aspect is primarily based on published clinical literature and extensive real-world clinical experience (outcomes data) with similar devices and the BAV procedure, as well as specific mentions of the NuMED NuCLEUS catheters in successful procedures. This includes demonstrating:
      • Well-accepted use of BAV as a palliative treatment for severe aortic stenosis.
      • BAV as a preparatory step for TAVR.
      • Successful outcomes where NuMED NuCLEUS catheters were used.
      • The similarity in design between NuCLEUS and NuCLEUS-X making existing NuCLEUS clinical data applicable to NuCLEUS-X.

    8. The Sample Size for the Training Set

    Not Applicable in the context of machine learning. For the clinical evaluation, the "training set" equivalent would be the vast body of worldwide clinical experience with BAV, which is described as "extensive," "many years," and involving "numerous BAV clinical studies." This represents a very large, but unquantified, patient population. For the specific NuMED NuCLEUS family, the cited clinical studies involved cohorts of 20, 212, and 50 patients, and various unspecified "patients" in other papers.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable for machine learning. For the clinical evaluation, the "ground truth" (i.e., the established understanding of the BAV procedure and the performance of these catheters) was established through:

    • Accumulated Clinical Data: Decades of clinical practice and research on Balloon Aortic Valvuloplasty (BAV) using various devices.
    • Published Studies: Peer-reviewed medical literature documenting the efficacy, safety, and outcomes of BAV procedures, including specific mentions of NuMED NuCLEUS catheters.
    • FDA Assessment: The FDA's "scientific and clinical reviewers" critically evaluated this body of literature and experience to form their conclusions regarding the safety and effectiveness of the device and procedure.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4