Search Results
Found 12 results
510(k) Data Aggregation
(422 days)
Kaltenbach & Voigt GmbH
The DEXIS CariVu 3-in-1 by KaVo acquires patient images as a diagnostic aid for detection of smooth surface caries, occlusal caries, proximal caries, secondary caries, and tooth cracks. The DEXIS Cari Vu 3-in-1 by KaVo is designed for use by a trained professional in the field of dentistry.
The device is used as an intraoral camera to acquire images in the oral cavity of the patient. The camera combines three different known applications from available intraoral cameras in one device. These applications are standard camera functionality, transillumination and fluorescence caries detection. Therefore, the camera uses different light colors for each of the three applications to illuminate the tooth. In combination with an optical system, CMOS sensor, and an internal image processing unit, a digital image of the tooth is generated. The live images are transferred as a standard mjpg video stream via USB interface to a running computer and can interface with a dental imaging software such as for example CONEXIO for basic intraoral camera management functionality.
The device may be used as an aid for monitoring caries lesion size and shape by comparing images taken with the Dexis CariVu 3-in-1 by KaVo over time, by using different reference points, such as tooth contour and position the device to optimize capturing tooth features previously recorded in a still or video image.
For easy use the camera is equipped with an autofocus mechanism which will provide a sharp representation of the tooth which is placed in the center of the imaqe. The autofocus provides the ability to have sharp images from close distances of about 2mm to about 120mm distance.
The different operation modes are described as follows:
1.) Color mode:
For normal color images, a white LED for illumination is used. With this light the camera provides coloured images or videos. The camera can be used with or without a Tip Vision Full HD which is required for illumination for the transillumination mode.
- Fluorescence mode:
For the fluorescence mode the blue LED illumination is used. The blue light excites bacterial porphorines as an aid to detect caries.
To enhance the image quality and to minimize external light influence on the fluorescence effect the accessories Fluorescence Cover Vision Full HD can be used optionally with the camera. The Fluorescence Cover Vision Full HD is a rubber cap which can be slipped on the front camera head and prevent or minimize external light on the images surface of the tooth. After each usaqe the Fluorescence Cover Vision Full HD will be reprocessed separately.
The camera can be used with or without a Tip Vision Full HD which is required for illumination for the transillumination mode. This mode is used to record images and videos to aid in the diagnosis of caries (mainly in the occlusal area) and plaque by means of fluorescent effect.
- Transillumination mode (by the use of near infrared light):
The transillumination mode provides images to aid in the detection of caries. For this mode near infrared light will be used. The near infrared light penetrates the side of the tooth near gingiva and illuminates the enamel of the tooth to produce gray scale images whereby intact enamel appears bright, and defects such as caries or cracks result in diffusion of light path and create an area of dark appearance.
To use this mode, the Tip Vision Full HD must be attached to the device as the LEDs for the illumination are located inside the Tip Vision Full HD. The Tip Vision Full HD touches the side of the tooth and gingiva and illuminates it from two sides whereas the image is taken from the top.
The DEXIS CariVu 3-in-1 by KaVo is available in two similar models. The differences between the models are a wall mounted version to store the device with a cable length of 2.5 meters (model 1.011.4444) and a table holder version to store the device on a dental unit with a cable length of 1.55 meters (model 1.013.1700). All critical components within the DEXIS CariVu 3in-1 by KaVo are common. A disposable sheath is used with the device to mitigate the potential for cross contamination.
The provided text does not contain detailed information about specific acceptance criteria or an in-depth study proving the device meets those criteria, especially in the format requested. The document is a 510(k) summary for the DEXIS CariVu 3-in-1 by KaVo, primarily focusing on demonstrating substantial equivalence to predicate devices rather than presenting the results of a primary clinical study with predefined acceptance criteria.
However, I can extract the following information from the text:
1. Table of acceptance criteria and reported device performance:
The document does not provide a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, accuracy, or AUC. It focuses on comparison of capabilities and safety standards with predicate devices.
2. Sample size used for the test set and the data provenance:
- Clinical Testing: "Clinical testing was not conducted on this product." This indicates that there was no clinical test set for performance evaluation in the context of this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and their qualifications:
- Since no clinical testing was performed, there were no experts used to establish ground truth for a test set.
4. Adjudication method for the test set:
- Since no clinical testing was performed, there was no adjudication method for a test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No MRMC study was done, as "Clinical testing was not conducted on this product."
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The device description indicates it's an intraoral camera with various modes (color, fluorescence, transillumination) "as a diagnostic aid for detection." It is designed "for use by a trained professional in the field of dentistry." This implies a human-in-the-loop scenario. No standalone performance data for an algorithm alone is provided.
7. The type of ground truth used:
- Not applicable as no clinical testing was performed to establish performance metrics against a ground truth.
8. The sample size for the training set:
- The document does not mention a training set or its size, as it states clinical testing was not conducted. This implies that the device's development and evaluation for this submission did not involve a supervised learning approach requiring a distinct training set in the context of a clinical performance study.
9. How the ground truth for the training set was established:
- Not applicable as no training set is mentioned in the context of clinical performance evaluation.
Summary of what is present in the document relevant to performance and equivalence:
The document primarily focuses on non-clinical bench testing to demonstrate performance according to international standards for medical electrical equipment, sterilization, biocompatibility, risk management, and cybersecurity. These are safety and engineering performance criteria, not clinical diagnostic performance criteria.
- Non-Clinical Test Data: Performance bench testing focused on:
- IEC 60601-1-2 Forth Edition (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance)
- ANSI AAMI ES60601-1 and IEC 60601-1:2005 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)
- CAN/CSA-C22.2 No. 60601-1 (Medical Electrical Equipment - Part 1: General Reg. for Basic Safety & Essential Performance)
- IEC 62304: 2006 Edition 1.1 (Medical device software - Software life cycle processes)
- ISO 17665-1 First edition (Sterilization of health care products - Moist heat)
- AAMI / ANSI ST79 2017 (Comprehensive guide to steam sterility assurance in health care facilities)
- ISO 10993-1 Fourth Edition (Biological evaluation of medical devices)
- ISO 7405 Second edition (Dentistry - Evaluation of biocompatibility of medical devices used in dentistry)
- ISO 14971 2007 and 2012 (Medical devices - Application of risk management to medical devices)
- Guidance for Industry and FDA Staff 2014-10-02 Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- Guidance for Industry and FDA Staff 2017-06-09 Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling
The conclusion of the 510(k) process states: "Based upon a comparison of indications for use and technological characteristics, together with results from non-clinical performance testing, we believe that DEXIS CariVu 3-in-1 by KaVo is substantially equivalent to the predicate devices." This highlights that the equivalence was demonstrated without direct clinical performance data for the new device.
Ask a specific question about this device
(348 days)
Kaltenbach & Voigt GmbH
The PROPHYflex 4 Handpiece and Accessories is a dental handpiece intended for perform supragingival and subgingival prophylaxis treatments. The prophylaxis treatments are obtained by the projection of water, air and appropriate dental prophylaxis powders, onto the treatment site. PROPHYflex 4 Handpiece and Accessories in supra-gingival configuration is intended for use in the cleaning and polishing of teeth by removal soft deposits and areas of discoloration as well as preparing teeth for fillings or veneers and cleaning fixed bands and brackets on orthodontic appliances. PROPHYflex 4 Handpiece and Accessories in subgingival configuration is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
The PROPHYflex 4 is an air polishing device/powder jet handpiece that projects a mixture of water and air with sodium bicarbonate, calcium carbonate or fine glycine power onto tooth surfaces. The device includes the handpiece main body with adjustable spray feature, two cannulas, a wrench for installing the cannula, two interchangeable gripping sleeves, powder container with cover, a cleaning burr and nozzle pin, and a set of spare o-rings. An optional adapter and perio tip can be installed for subgingival treatments.
The powder is filled into the powder container by the user. When activated through foot control, air is pushed into the powder container, swirling the powder and pushing it through the device and out of the cannula. The spray is only mixed with the air and powder at the very tip of the cannula, providing an even and gentle cleaning effect. The amount of powder can be regulated via three different spray settings by the adjusting ring, helping the users to perform individual treatments such as subgingival biofilm removal, removal of discoloration and bacterial plaque, orthodontics, cleaning prior to fissure sealing, prosthetics, conservative and aesthetic dentistry by providing the possibility to adjust the cleaning effects to the clinical situation of the patient at hand.
The components of the PROPHYflex 4, including the powder container, is to be completely reprocessed after each use.
The three powder levels are as follows:
- Highest Level: Suitable for supragingival treatment and offers full cleaning power .
- Medium Level: Suitable for sub- and supragingival treatment and enables . gentle cleaning at a reduced powder quantity
- . Lowest Level: Helps rinsing the powder off the tooth and blowing the powder from the system after the treatment
During supraqinqival treatments (e.g. stain removal), the device doesn't touch the patient but will be placed 3 – 5 mm from the area that is to be cleaned. During subgingival treatments (biofilm management) with the perio tip, the tip is inserted into the pocket up to 5mm and angled towards the tooth.
The provided text is a 510(k) summary for a dental handpiece, the PROPHYflex 4. It focuses on demonstrating substantial equivalence to predicate devices through technological characteristic comparison and non-clinical performance testing.
Crucially, this document explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product."
Therefore, a study demonstrating the device meets acceptance criteria via clinical performance (which would involve human readers, ground truth established by experts, MRMC studies, etc.) was explicitly not performed and is not required for this type of device according to the provided FDA submission.
However, I can extract the acceptance criteria and reported performance based on the non-clinical (bench) testing, and other relevant information from the document, framing it within the context of a "study" as requested, even if it's a bench study rather than a clinical one.
Here's a breakdown based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for the PROPHYflex 4 are implicitly defined by its ability to perform as well as, or better than, its predicate devices across various technical specifications and to comply with relevant international standards. The study proving these criteria are met consists of non-clinical bench testing.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are essentially demonstrated equivalence or improvement over the predicate device (AIR-FLOW Handy 3.0 Plus (K151912)) and compliance with specified ISO standards.
Feature / Acceptance Criterion (Implicit) | Acceptance Criterion (Specific Value from Predicate or Standard) | Reported Device Performance (PROPHYflex 4) | Met? |
---|---|---|---|
Indications for Use (Subgingival) | Non-surgical removal of subgingival plaque in pockets up to 4 mm (Predicate H3.0 Plus), Up to 5mm (Reference Predicate H3.0 PERIO) | Non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. | Yes |
Max Mode Operation | Present (Predicate H3.0 Plus) | Present | Yes |
Internal Spray | Present (Predicate H3.0 Plus) | Present | Yes |
Coupling Dimensions | Complies with ISO 7785-1, ISO 14457, and ISO 9168 (Predicate H3.0 Plus) | Complies with ISO 7785-1, ISO 14457, and ISO 9168. (Specifically mentioning KaVo MULTIflex-Coupling, Sirona Coupling compatibility) | Yes |
Biocompatibility | According to ISO 10993-1 (Predicate H3.0 Plus) | Biocompatible according to ISO 10993-1 (Biological evaluation of medical devices), ISO 10993-5 (In vitro cytotoxicity), ISO 10993-10 (Irritation/sensitization), ISO 10993-18 (Chemical characterization). | Yes |
Noise Level | 74.9 dB(A) (Predicate H3.0 Plus) | 56.6 - 69.0 dB(A) (Quieter than predicate) | Yes |
Operating Temperature (Grip Sleeve) | 19.93°C - 23.33°C (Predicate H3.0 Plus) | 18.0°C - 25.1°C (Wider range, but within acceptable dental operating temperatures) | Yes |
Maximum Air Pressure | 2.7 – 3.5 bar (39 – 51 psi) (Predicate H3.0 Plus) | 3.2 - 5 bar (46 - 73 psi) (Higher pressure capability) | Yes |
Air and water supply | 10.3 – 13.5 Nl/min (Predicate H3.0 Plus) | 6.4 – 14.3 Nl/min (Comparable range) | Yes |
Maximum Water Pressure | 1.0 – 2.2 bar (15 – 32 psi) (Predicate H3.0 Plus) | 1.0 - 2.5 bar (15 - 36 psi) (Higher pressure capability) | Yes |
Spray Water Flow | 106 ml/min (Predicate H3.0 Plus) | 28 – 54 ml/min (Lower flow, noted as a difference but not raising new concerns) | Yes |
Compliance with Powder Jet Handpiece Standard | DIN 19402 (Predicate H3.0 Plus) | DIN EN ISO 20608 (Dentistry - Powder jet handpieces and powders), DIN 19402 (Powder Jet Handpieces) | Yes |
Additional Operational Modes | N/A (Predicate H3.0 Plus only has Max Mode) | Middle Mode (reduced powder), Spray-function Mode (water only) | N/A (Improvement over predicate) |
Device Dimensions (Length) | 200 mm (Predicate H3.0 Plus) | Up to 190 mm (10 mm shorter) | Yes |
Device Dimensions (Largest Diameter) | 43 mm (Predicate H3.0 Plus) | Up to 40 mm (3 mm thinner) | Yes |
Reprocessing Method | N/A (Standard dental practice) | Components are to be completely reprocessed after each use (standard for dental handpieces) | Yes |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify exact "sample sizes" in terms of number of devices tested for each bench test. These are typically covered under the manufacturer's internal quality system documentation and test protocols, which are summarized rather than detailed in a 510(k). The tests are performed on representative units of the device.
- Data Provenance: The testing was non-clinical (bench testing).
- Country of Origin: The manufacturer is Kaltenbach & Voigt GmbH, located in Biberach, Germany. Therefore, the testing would likely have been conducted in Germany or by approved testing facilities.
- Retrospective or Prospective: This refers to the collection of patient data, not bench testing. Bench testing is inherently prospective in its execution, demonstrating performance against specifications.
3. Number of Experts Used to Establish Ground Truth and Qualifications of Those Experts
- This question is not applicable as the study was purely non-clinical bench testing. There were no "experts" establishing a clinical "ground truth" for a test set, as no clinical test set involving human subjects was used. The "ground truth" for bench testing is derived from engineering specifications, established medical device safety and performance standards (e.g., ISO standards), and comparative performance against legally marketed predicate devices.
4. Adjudication Method for the Test Set
- This question is not applicable as there was no test set requiring human expert adjudication (e.g., image interpretation). Bench testing results are typically objectively measured against defined criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was explicitly NOT done. The document states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product." Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- This question is not applicable in the context of a dental handpiece. The device is a physical instrument, not an AI algorithm. Its "performance" is its mechanical, fluidic, and material characteristics demonstrated through bench testing.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance validation is based on:
- International Standards: Compliance with ISO standards (e.g., ISO 7785-1, ISO 14457, ISO 9168, ISO 10993 series, ISO 20608).
- Predicate Device Performance: Demonstrating that the PROPHYflex 4's performance characteristics (dimensions, noise, pressure, flow, etc.) are comparable to or improved over the legally marketed predicate device (AIR-FLOW Handy 3.0 Plus (K151912)) and other reference devices.
- Engineering Specifications: Internal design specifications and performance targets set by the manufacturer for the device.
8. The Sample Size for the Training Set
- This question is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a "training set" of data.
9. How the Ground Truth for the Training Set Was Established
- This question is not applicable as this is a physical medical device, not an AI/ML algorithm; therefore, there is no "training set" or "ground truth" establishment for it in this context.
Ask a specific question about this device
(119 days)
Kaltenbach & Voigt GmbH
The SMARTmatic handpieces are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
The SMARTmatic electrical-driven handpieces listed in the table below are dental handpieces for the use by a trained professional in the field of general dentistry. The devices are electricalpowered handpieces that are reusable and ergonomically shaped. The devices can be sterilized by the steam autoclave method. Through the tube and the electrical motor connected to a dental unit, the SMARTmatic handpieces equipped with a handpiece connection according to ISO 3964 (Dentistry - Coupling dimensions for handpiece connectors - ISO 3964:1982) receive the energy for the gear, the cooling water and air for cutting treatment. Dental burs and other attachments will be used with the SMARTmatic handpieces per ISO 1797-1:2011 (Dentistry - Shanks for rotary instruments - Part 1: Shanks made of metals).
The request asks for a detailed description of acceptance criteria and the study proving the device meets these criteria. However, the provided document is a 510(k) premarket notification for a dental handpiece (SMARTmatic), which focuses on demonstrating substantial equivalence to a predicate device rather than on specific acceptance criteria and a detailed study proving performance against those criteria.
The document primarily highlights the substantial equivalence of the SMARTmatic handpieces to the MASTERmatic LUX handpieces (K143465) based on intended use, technological characteristics, and performance data. It relies on conformance to various ISO standards and a guiding document for dental handpieces. The document does not contain information on:
- A specific table of acceptance criteria with reported device performance against numerical targets.
- Sample sizes for a test set or data provenance (as it's not a clinical study on diagnostic accuracy).
- Number of experts, their qualifications, or adjudication methods for ground truth.
- Multi-reader multi-case (MRMC) comparative effectiveness studies.
- Standalone algorithm performance (as it's a mechanical device, not an AI algorithm).
- Sample size for a training set or how ground truth for a training set was established.
Therefore, many of the requested points cannot be extracted from this document as they are not relevant to a 510(k) submission for a dental handpiece.
However, I can extract information related to the performance data and "acceptance" from the perspective of substantial equivalence and adherence to recognized standards.
Here's the information that can be extracted and a clear indication of what is not present:
1. A table of acceptance criteria and the reported device performance
The document doesn't provide a table of acceptance criteria in the sense of specific numerical thresholds for performance metrics. Instead, it relies on conformance to international standards. The "Summary of the Technological Characteristics" table ([page 7] and [page 8]) compares the new device (SMARTmatic) to the predicate device (MASTERmatic LUX) across various design and technical specifications. The "acceptance criteria" here are effectively that the new device's characteristics are either the same as the predicate or, if different, that these differences do not raise new questions of safety or effectiveness, as demonstrated by non-clinical testing to relevant standards.
Summary of Technological Characteristics and Performance (as presented in the document):
Descriptive Information | SMARTmatic | MASTERmatic LUX (Predicate) |
---|---|---|
Indications for Use | The SMARTmatic handpieces are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry. | The MASTERmatic LUX handpieces are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry. |
Functional Principle | Through the micro motor connected to the dental treatment unit, the handpieces receive energy, cooling water, air for treatment, and light (for illumination) (Note: SMARTmatic does not have fiber optics). | Through the micro motor connected to the dental treatment unit, the handpieces receive energy, cooling water, air for treatment, and light for illumination the operating area. |
Air / Water Ports | No Spray / External Spray (according to model) | Internal Spray |
Fiberoptics | Without light system | With built-in light system |
Dimensions | Head size-Height: Up to 13.6 mm | |
Head size-Diameter: Up to 9.8 mm | ||
Length: Up to 93.4 mm | Head size-Height: Up to 13.6 mm | |
Head size-Diameter: Up to 10.2 mm | ||
Length: Up to 95.9 mm | ||
Type of Chuck | Push Button, Twist-tension Chuck, Snap-on & Screw-in | Push Button, Twist-tension Chuck |
Coupling Dimensions | Dimensions comply with ISO 3964 | Dimensions comply with ISO 3964 |
Accessories | Jet Needle, Drill Bit Stop (Bur Stop), Hook, Dental Burs and Prophylaxis Heads | Jet Needle, Drill Bit Stop (Bur Stop), Hook and Dental Burs |
Rotary Instruments | For Burs, Cutters and other attachments (with straight handpiece shank or with contra-angle shank) according to ISO 1797-1. | For Burs, Cutters and other attachments (with straight handpiece shank or with contra-angle shank) according to ISO 1797-1. |
Biocompatibility | Biocompatible according to ISO 10993-1 for both direct and indirect patient-contacting portions. Materials listed in tables (not provided in this excerpt). | Biocompatible according to ISO 10993-1 for both direct and indirect patient-contacting portions. |
Chuck Design | Push Button, Twist-tension Chuck, Snap-on & Screw-in | Push Button, Twist-tension Chuck |
Light Intensity | N/A (without light system) | 25,000 Lux |
Bur Retention Force | Min. 22 N (45 N straight handpieces) | Min. 22 N (45 N straight handpieces) |
Bur Extraction Force | 15N - 52N | 15N - 52N |
Maximum Air Pressure | 29 psi | 29 psi |
Maximum Water Pressure | 29 psi | 29 psi |
Speed Range (RPM's) | Up to 40,000 rpm | Up to 200,000 rpm |
Conformance Standards | ISO 14457, ISO 1797-1, ISO 3964, ISO 10993-1, ISO 17665-1, AAMI ANSI ST79 | ISO 14457, ISO 1797-1, ISO 3964, ISO 10993-1, ISO 17665-1 |
Sterilization | Sterilizable according to ISO 17665-1 | Sterilizable according to ISO 17665-1 |
Lubricant | KaVo QUATTROcare (K071288) | KaVo QUATTROcare (K071288) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- None stated directly. The document refers to "Performance bench testing according to international standards." This typically involves testing a defined number of device units or components, but the specific sample sizes are not provided in this summary.
- Data Provenance: Not explicitly stated, but the manufacturer is Kaltenbach & Voigt GmbH, based in Biberach, Germany. The tests are "Non-Clinical Test Data," indicating lab/bench testing, not data from human subjects. This would be considered prospective bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This information is for clinical studies or AI algorithm validation, not for a mechanical dental handpiece undergoing bench testing for substantial equivalence.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This is for clinical studies/AI algorithm validation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a mechanical dental handpiece, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a mechanical device, not an algorithm. Bench testing was done for the device itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For biocompatibility: Conformance to ISO 10993-1. The "ground truth" is that the materials used meet the specified biocompatibility standards.
- For mechanical performance (e.g., bur retention force, speed range, coupling dimensions): Conformance to relevant ISO standards (e.g., ISO 14457, ISO 1797-1, ISO 3964). The "ground truth" is whether the physical measurements and operational parameters of the device meet the requirements of these standards.
- For sterilization: Conformance to ISO 17665-1 and AAMI ANSI ST79. The "ground truth" is that the sterilization process specified for the device achieves the required sterility assurance levels.
8. The sample size for the training set
- Not applicable. This refers to AI/machine learning models.
9. How the ground truth for the training set was established
- Not applicable. This refers to AI/machine learning models.
Summary of the Study (as described in the document):
The "study" described is a non-clinical performance bench testing program to demonstrate substantial equivalence to a legally marketed predicate device (MASTERmatic LUX K143465).
- Methodology: Performance bench testing according to international standards (listed below) and biocompatibility and sterilization studies.
- Standards utilized for performance verification:
- ISO 3964 (Dentistry - Coupling dimensions for handpiece connectors)
- ISO 1797-1 (Dentistry - Shanks for rotary instruments - Part 1: Shanks made of metals)
- ISO 7405 / ISO 10993-1 (Dentistry - Evaluation of biocompatibility of medical devices used in dentistry / Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management system)
- ISO 14457 (Dentistry - Handpieces and motors)
- ISO 17665-1 (Sterilization of health care products - Moist heat - Part 1)
- AAMI ANSI ST79 (Comprehensive Guide To Steam Sterilization And Sterility Assurance In Health Care Facilities)
- Guidance for Industry and FDA Staff - Dental Handpieces - Premarket Notification [510(k)] Submissions.
- Conclusion: The tests demonstrated that despite some differences in technological characteristics (e.g., external vs. internal spray, presence/absence of fiber optics, speed range, specific chuck types, and minor dimensions), these differences do not raise new concerns regarding substantial equivalence to the predicate device. The SMARTmatic handpieces are deemed substantially equivalent based on intended use, indications, technological characteristics, principle of operation, features, and performance data derived from these non-clinical tests.
- Clinical Data: The document explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product."
Ask a specific question about this device
(90 days)
Kaltenbach & Voigt GmgH
The ELECTROmatic is intended to convert pneumatic output from a dental treatment center to drive the COMFORT drive motor handpiece and the INTRA LUX KL 703 LED motor for operation of electrically-driven dental handpieces. This device is designed for use by a trained professional in the field of general dentistry
The ELECTROmatic dental control unit is a stand-alone system for operating electrically-driven KaVo handpieces. External power supply provides electric power to the unit. The 4-hole tubing connected to the unit supplies chip /cooling air, water and pressure signal. The speed of the electric handpiece is controlled by air pressure. The control unit is positioned close to a treatment unit at the location preferred by the dentist. The ELECTROmatic system consists of a base unit with a motor hose, an electrical motor, a transformer, and a power cord.
The provided document is a 510(k) premarket notification for a medical device called "ELECTROmatic." This document primarily focuses on establishing substantial equivalence to a predicate device, rather than presenting a study demonstrating the device meets specific acceptance criteria in the way a novel AI or diagnostic device might.
The ELECTROmatic is a dental control unit, and the "acceptance criteria" discussed are largely related to meeting established industry standards for safety and performance, as well as demonstrating that its technological characteristics do not raise new concerns compared to the predicate device.
Here's an analysis based on your requested information, acknowledging that many of the points you asked for are not applicable to this type of device submission:
1. Table of Acceptance Criteria and Reported Device Performance
For this type of device (a dental control unit), acceptance criteria are primarily defined by compliance with recognized international and national standards for medical electrical equipment, dental instruments, and biocompatibility. The document states that performance bench testing was conducted to demonstrate conformance to these standards.
Acceptance Criterion (Standard Compliance) | Reported Device Performance |
---|---|
ISO 9168:2009 (Dentistry – Hose connectors for air-driven dental handpieces) | Performance bench testing conducted to determine conformance. (Implicit: Device meets requirements) |
ISO 14457:2012 (Dentistry – Handpieces and motors) | Performance bench testing conducted to determine conformance. (Implicit: Device meets requirements) |
IEC 60601-1:2005 (Medical electrical equipment – General requirements for basic safety and essential performance) | Performance bench testing conducted to determine conformance (IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007)). (Implicit: Device meets safety and essential performance requirements) |
IEC 60601-1-2:2007 (Medical electrical equipment – Electromagnetic compatibility) | Performance bench testing conducted to determine conformance (IEC 60601-1-2:2007 / EN 60601-1-2:2007). (Implicit: Device meets EMC requirements) |
ANSI/AAMI ES60601-1:2005/A2:2010 (Medical electrical equipment – General requirements for basic safety and essential performance) | Performance bench testing conducted to determine conformance. (Implicit: Device meets safety and essential performance requirements) |
IEC 62304:2006 (Medical device software – Software life cycle processes) | Performance bench testing conducted to determine conformance (IEC 62304:2006 -First edition). (Implicit: Software development followed recognized life cycle processes, ensuring safety and performance) |
ISO 7405:2008 & ISO 10993-1:2009 (Biocompatibility of medical devices) | Biocompatibility studies completed for the applicable components. (Implicit: Materials are biocompatible) |
Sterilization Studies (for applicable components) | Sterilization studies completed for the applicable components. (Implicit: Applicable components can be effectively sterilized) |
Substantial Equivalence (General) | The performance tests demonstrate that the differences in technological characteristics do not raise different concerns regarding substantial equivalence to the predicate device. The device satisfies all criteria of substantial equivalence and does not raise new concerns regarding substantial equivalence: (1) Indications for Use, (2) Technological Characteristics and (3) Performance Data. The new device does not introduce a fundamentally new scientific technology and the nonclinical tests demonstrate that the device is substantially equivalent. |
2. Sample size used for the test set and the data provenance
This document describes a premarket notification for a medical device that converts pneumatic output to electrical energy to drive dental handpieces. It is not a study involving patient data or a test set in the context of clinical performance of a diagnostic or AI device.
- Test Set Sample Size: Not applicable. The "tests" described are bench tests against engineering standards, not tests on a dataset of patient samples.
- Data Provenance: Not applicable. The "data" involves engineering measurements and compliance with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. There is no "ground truth" in the clinical sense being established by experts for a test set in this 510(k) submission. The design and performance verification are conducted by engineers and specialists in accordance with specified standards.
4. Adjudication method for the test set
Not applicable. There is no multi-reader or adjudication process for interpreting results in this submission. Compliance is determined through objective measurements against engineering and safety standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a dental control unit, not an AI diagnostic tool or system that assists human readers. No MRMC study was conducted.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm-only device. It's a hardware device with embedded software for controlling motor functions. Its "performance" refers to its ability to meet engineering specifications and safety standards, not a diagnostic output.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable in the clinical sense. The "ground truth" for this device's performance evaluation is its adherence to the specified technical and safety standards (e.g., ISO, IEC, ANSI/AAMI) through bench testing. The device is expected to operate within defined electrical, mechanical, and safety parameters, which are objectively measurable.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set. The software mentioned controls motor functions based on deterministic algorithms, not learned patterns from data.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for a machine learning model, no ground truth needed to be established in this context. The software's functionality is designed based on known physics and engineering principles, not derived from labeled data.
In summary:
The ELECTROmatic is a Class I dental control unit that demonstrated substantial equivalence to a predicate device by showing compliance with a range of recognized international and national standards for medical electrical equipment, dental instruments, software lifecycle processes, and biocompatibility through non-clinical bench testing. The evaluation focused on engineering performance and safety compliance, rather than clinical performance based on patient data or the development of a diagnostic AI algorithm.
Ask a specific question about this device
(190 days)
KALTENBACH & VOIGT GMBH
The ESTETICA Dental Treatment Unit and Accessories are intended to supply power to and serve as a base for other dental devices and accessories by providing air, water, vacuum and low voltage electrical power to dental instruments and dental handpieces. The ESTETICA Dental Treatment Unit and Accessories are designed for use by a trained professional in the field of general dentistry.
The ESTETICA Dental Treatment Unit and Accessories serves as a base that includes components to deliver air, water, electrical power, and vacuum to dental handpieces, instruments, and accessories. The device includes a patient chair, foot control, dentist element, assistant element and a water unit. The unit may also include a dental operating light and a monitor. Various handpieces and accessories can be added to the ESTETICA Dental Treatment Unit and Accessories. These include, but not limited to, pneumatic handpieces, electric motors with handpieces, scalers, intra-oral cameras, curing lights, air/water syringes and SE and HVE vacuum instruments.
The provided text describes the 510(k) premarket notification for the ESTETICA Dental Treatment Unit and Accessories. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. Here's a breakdown of the requested information based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document does not present specific acceptance criteria in a table format with corresponding reported device performance metrics in the way one might see for a diagnostic AI device (e.g., sensitivity, specificity thresholds). Instead, the acceptance criteria are implicitly defined by conformance to various international and national standards for dental operative units and medical electrical equipment. The "reported device performance" is broadly stated as "Performance bench testing according to international standards for dental operative units has been conducted to determine conformance" and "the performance of the ESTETICA Dental Treatment Unit and Accessories has been verified utilizing the following standards."
Here's an interpretation based on the provided text, outlining the areas of conformance:
Acceptance Criteria (Implicit by Conformance to Standards) | Reported Device Performance |
---|---|
Biocompatibility (ISO 7405, ISO 10993-1) | All applicable components meet biocompatibility requirements. |
Sterilization (ISO 17665-1, AAMI ANSI ST79) | Applicable components meet sterilization requirements. |
Software Documentation (FDA Guidance Document for Software Contained in Medical Devices) | Software documentation for moderate level of concern completed. |
Comparative Performance of Integrated Accessories | Performance of integrated accessories is comparable to cleared standalone devices. |
Wireless Foot Pedal - Software Verification/Validation | Software verification/validation of functions conducted. |
Wireless Foot Pedal - EMC Testing (IEC 60601-1-2) | EMC testing performed to evaluate communication loss risk. |
Wireless Foot Pedal - Electrical Safety (IEC 60601-1-1) | Electrical safety tests conducted. |
General Requirements & Test Methods for Dental Units (ISO 7494-1) | Performance verified per standard. |
Water and Air Supply for Dental Units (ISO 7494-2) | Performance verified per standard. |
Hose Connectors for Air Driven Dental Handpieces (ISO 9168) | Performance verified per standard. |
Handpieces and Motors (ISO 14457) | Performance verified per standard. |
Basic Safety and Essential Performance (Medical Electrical Equipment) (IEC 60601-1, ANSI/AAMI ES60601-1) | Conformance demonstrated. |
Electromagnetic Compatibility of Medical Electrical Equipment (IEC 60601-1-2) | Conformance demonstrated. |
Medical Device Software Life Cycle Processes (IEC 62304) | Conformance demonstrated. |
Risk Management (ISO 14971) | Application of risk management to medical devices demonstrated. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not detail specific sample sizes for particular performance tests. It refers to "Performance bench testing" and "Comparative performance testing" but does not specify the number of units or test subjects used in these tests. The provenance of the data (country of origin, retrospective/prospective) is also not specified, as these are non-clinical bench tests, not clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This product is a dental treatment unit, not an AI or diagnostic device that requires expert-established ground truth for its performance evaluation in the typical sense of medical image analysis. The "ground truth" here is adherence to engineering and safety standards, verified through bench testing. Therefore, there were no experts used to establish ground truth in the context of clinical interpretation or diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As a device undergoing non-clinical bench testing for conformance to standards, there is no "adjudication method" in the clinical sense (e.g., for conflicting expert opinions). Performance is evaluated against documented technical specifications and standard requirements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a dental operative unit, not an AI-assisted diagnostic or interpretative tool. Therefore, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This device does not involve an "algorithm only" component for diagnostic or interpretative tasks. It is a physical medical device that provides power and functions for dental procedures. Performance evaluations were done on its physical and electrical characteristics according to relevant standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance evaluation of the ESTETICA Dental Treatment Unit and Accessories is the set of specifications and requirements outlined in the international and national standards listed (e.g., ISO 7494-1, IEC 60601-1). Performance is assessed based on whether the device operates within the parameters defined by these standards. No expert consensus, pathology, or outcomes data were used for establishing ground truth in this context.
8. The sample size for the training set
Not applicable. This device is not an AI/ML product and does not involve a "training set" for an algorithm.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
Ask a specific question about this device
(126 days)
Kaltenbach & Voigt Gmbh
The MASTERmatic LUX handpieces are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
The MASTERmatic LUX electrical-driven handpieces - MASTERmatic LUX M25 L, MASTERmatic LUX M20 L, MASTERmatic LUX M10 L, MASTERmatic LUX M07 L, MASTERmatic LUX M29 L, MASTERmatic LUX M05 L are dental handpieces for the use by a trained professional in the field of general dentistry. The devices are electrical-powered handpieces that are reusable and ergonomically shaped, and provided with a fiber optic light system. The devices can be sterilized by the steam autoclave method. Through the tube and the electrical motor connected to a dental unit, the MASTERmatic LUX handpieces equipped with a handpiece connection according to ISO 3964 (Dentistry - Coupling dimensions for handpiece connectors - ISO 3964:1982) receive the energy for the gear, the cooling water and air for cutting treatment through pouring holes and light for illumination the operation area. Dental burs and other attachments according to ISO 1797-1 (Dentistry - Shanks for rotary instruments - Part 1: Shanks made of metals - ISO 1797-1:2011) will be used with the MASTERmatic LUX handpieces.
Here's a breakdown of the acceptance criteria and study information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (from referenced standards) | Reported Device Performance (MASTERmatic LUX) |
---|---|
Intended Use/Indications for Use: Removal of carious material, reducing hard tooth structure, cavity & crown preparations, root canal preparations, removal of fillings, processing & finishing tooth preparations/restorations, polishing teeth. Designed for use by a trained dental professional. | Same as acceptance criteria (stated in "Statement of Intended Use" and "Summary of the Technological Characteristics"). |
Functional Principle: Receives energy, cooling water, air for treatment, and light for illumination via micro motor connected to dental unit, complying with ISO 3964. | Same as acceptance criteria (stated in "Summary of the Technological Characteristics" and "Mechanism of Action"). |
Air/Water ports: Internal Spray | Internal Spray (stated in "Summary of the Technological Characteristics") |
Fiberoptics: Must have a light system. | With built-in light system (stated in "Summary of the Technological Characteristics") |
Speed Range: Up to 200,000 rpm | Up to 200,000 rpm (stated in "Summary of the Technological Characteristics") |
Rotary Instruments: Compatible with burs, cutters, and other attachments (straight or contra-angle shank) according to ISO 1797-1. | Compatible with burs, cutters, and other attachments (straight or contra-angle shank) according to ISO 1797-1 (stated in "Summary of the Technological Characteristics"). |
Biocompatibility: Direct and indirect patient-contacting portions must be biocompatible according to ISO 10993-1. | Complies with ISO 10993-1 (stated in "Summary of the Technological Characteristics"). |
Light Intensity: Approx. 25,000 LUX | Approx. 25,000 LUX (stated in "Summary of the Technological Characteristics"). |
Bur retention force: Min. 22 N (45 N for straight handpieces) | Min. 22 N (45 N straight handpieces) (stated in "Summary of the Technological Characteristics"). |
Sterilization: Sterilizable according to ISO 17665-1 (Moist heat). | Sterilizable according to ISO 17665-1 (stated in "Summary of the Technological Characteristics"). |
Compliance to Standards: ISO 14457 (Dentistry - Handpieces and motors), ISO 1797-1 (Dentistry - Shanks for rotary instruments), ISO 3964 (Dentistry - Coupling dimensions for handpiece connectors). | Complies with ISO 14457, ISO 1797-1, ISO 3964 (stated in "Summary of the Technological Characteristics" and "Non-Clinical Test Data"). |
2. Sample size used for the test set and the data provenance
The document explicitly states: "Non-Clinical Test Data". This indicates that the testing was bench testing and not performed on human subjects. Therefore, there is no "test set" in the context of patient data, nor is there any provenance related to country of origin for such data. The testing was conducted according to international standards for dental gear-driven handpieces.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable. The device underwent "Non-Clinical Test Data" (bench testing) and "Clinical testing has not been conducted on this product." Therefore, there was no human expert adjudication for a clinical ground truth.
4. Adjudication method for the test set
This section is not applicable. As no clinical studies were performed that required expert adjudication, no adjudication method was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. No MRMC study was conducted, as no clinical studies were performed and the device is a dental handpiece, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This section is not applicable. The device is a dental handpiece and does not involve an algorithm or AI. Performance was assessed through non-clinical bench testing.
7. The type of ground truth used
The ground truth used for performance verification was based on compliance with established international standards for dental devices (bench testing outcomes). This includes:
- ISO 3964 (Dentistry - Coupling dimensions for handpiece connectors)
- ISO 1797-1 (Dentistry - Shanks for rotary instruments)
- ISO 10993-1 (Biological evaluation of medical devices - Biocompatibility)
- ISO 14457 (Dentistry - Handpieces and motors)
- ISO 17665-1 (Sterilization of health care products - Moist heat)
8. The sample size for the training set
This section is not applicable. There is no mention of a training set as this is a physical medical device, not a machine learning model.
9. How the ground truth for the training set was established
This section is not applicable. There is no training set mentioned or implied in the provided document.
Ask a specific question about this device
(189 days)
KALTENBACH & VOIGT GMBH
EXPERTsurg: This KaVo product is intended for surgery to expose and dissect oral tissue structures or endodontic treatments (e.g. periodontal gap, gingiva, bone, jaw, extractions and implantations).
INTRA LUX S600 LED: The medical device is intended to drive / operate a dental handpiece / contra-angle handpiece equipped with a handpiece connection according to ISO 3964. This medical device is an electrical low-voltage motor for dental purposes according to ISO 11498 type 2 and classified as a type B application part.
SURGmatic S201 L / C; SURGmatic S201 XL / XC; SURGmatic S11 L / C: The medical device is intended for the following uses: Surgery such as setting an implant, bone augmentation, sinus lift, tooth extraction procedures, implantology, osteotomia, root resection and oral, jaw and facial surgery.
The MASTERsurg / EXPERTsurg is a dental surgical system out of:
- EXPERTsurg (surgical control unit)
- INTRA LUX S600 LED (surgical motor)
- SURGmatic S201 L / C (handpieces)
- SURGmatic S201 XL / XC (handpieces)
- SURGmatic S11 L / C (handpieces)
The MASTERsurg / EXPERTsurg is a system out of a surgical control unit (EXPERTsurg), a surgical motor (INTRA LUX S600 LED) and different handpieces (SURGmatic).
The surgical control unit (EXPERTsurg) consists of a surgical controller, a motor cable, an instruments tray, a holder and a tube set. The surgical motor (INTRA LUX S600 LED) consists of the motor, a stopper for maintenance purposes and exchange O-rings for the motor coupling.
As a functional principle the software-based surgical controls the speed and torque of a dental micro motor. The unit is equipped with a pump for the use with external irrigation tubing allowing irrigation of the working area. The surgical control unit is operated through the buttons on the tabletop console or via foot control. The surgical control unit is intended to be used with the INTRA LUX S600 LED motor. Straight or contra-angle handpieces with a handpiece connection according to ISO 3964 can be equipped. The instrument tray allows the dentist to deposit the SURGmatic handpieces in a safe position. The holder is intended to be used for general storage. The tube set is needed to deliver the external irrigation from the bottle (not part of this 510(k)) to the SURGmatic handpieces. A power cord provides electric power to the unit. The EXPERTsurg will be delivered with software on the surgical control unit. The EXPERTsurg follows the international standards for electrical safety and electromagnetic compatibility which are applicable for the use with human beings.
Each of the different handpieces (SURGmatic) consists of the handpiece, a spray clip (only SURGmatic S201 L / C and SURGmatic S201 XL / XC), a jet needle for cleaning purposes and a tube for the external spray.
The SURGmatic handpieces are electrical driven dental handpieces for the use by a trained professional in the field of general dentistry. The devices are electrical-powered handpieces that are reusable and ergonomically shaped, and provided with a fiber optic light system (L version). The devices can be sterilized by the steam autoclave method. Through the INTRA LUX S600 LED motor connected to the surgical control unit, the SURGmatic handpieces equipped with a handpiece connection according to ISO 3964 receives the energy, the cooling water and air for treatment and the light for illumination the operation area (L version). Surgical burs and cutters (with straight handpiece shank or with contra-angle shank) according to ISO 1797 can be used with the SURGmatic handpieces. The SURGmatic handpieces use external spray allowing external irrigation to the working area. The mechanism of action for the SURGmatic handpieces is as follows. The dental straight and contra-angle handpieces are electrical-driven handpieces which will be supplied with energy, air and light through the dental micro motor of the surgical control unit. Dental burs (not part of this 510(k)) according to ISO 1797 could be inserted into the chuck system of the handpieces. Based on the speed adjusted in the surgical control unit the handpiece bur rotates up to 40,000 rpm. The SURGmatic handpieces interacts with the patient through a rotating bur with the patient teeth according to the intended use.
The provided document is a 510(k) premarket notification for a dental surgical system. It focuses on demonstrating substantial equivalence to predicate devices rather than proving specific performance metrics against acceptance criteria in the context of a new, quantifiable medical AI device evaluation. Therefore, much of the requested information (such as sample sizes for test/training sets, number/qualifications of experts, adjudication methods, MRMC studies, effect sizes, specific ground truth types, etc.) is not applicable or available in this type of submission.
The document describes non-clinical testing performed to show conformance to standards and software validation, but it does not present a study designed to establish device performance against predefined acceptance criteria in the way one would for a diagnostic or AI-driven system.
However, I can extract information related to the device's characteristics and the non-clinical tests performed.
Here's a summary of the available information based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of "acceptance criteria" and "reported device performance" in the sense of quantifiable metrics for diagnostic accuracy or clinical outcomes. Instead, it demonstrates compliance with international standards and describes functional principles and technical characteristics that are deemed substantially equivalent to predicate devices.
The "performance" is implicitly demonstrated through conformance to established standards and validation of software functionality, rather than through meeting specific numerical thresholds for a novel diagnostic claim.
Characteristic / Test | Acceptance Criteria (Implicit from Standards) | Reported Device Performance / Compliance |
---|---|---|
Electrical Safety | Conformance to international standards for electrical safety | Successfully tested and conforms to international standards. |
Electromagnetic Compatibility (EMC) | Conformance to international standards for EMC | Successfully tested and conforms to international standards. |
Biocompatibility | Demonstrably safe for intended use (based on established studies) | Biocompatibility studies completed; device is safe for its intended use. |
Software Validation | Confirmation of device performance as designed | Software successfully validated to confirm performance. |
Functional Principle | Similar to predicate devices (e.g., control speed/torque, irrigation) | Identical to predicate device (CHIROPRO L SYSTEM K092214) |
Compliance to Standards | ISO 11498 type 2 (motor), ISO 14457, ISO 1797, ISO 3964 (handpieces) | Complies with these international standards. |
Motor Torque | N/A (compared to predicate, not an AC) | 5.5 Ncm (Predicate: 6.8 Ncm) |
Max Speed (Handpieces) | N/A (compared to predicate, not an AC) | Up to 40,000 rpm (Predicate: Up to 50,000 rpm) |
Flow Rate (Pump) | N/A (compared to predicate, not an AC) | 30 - 110 ml/min (Predicate: 30 - 125 ml/min) |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable. The submission describes non-clinical engineering and bench testing (electrical safety, EMC, biocompatibility, software validation), not a clinical study with a "test set" of patients or data in the typical sense of evaluating diagnostic performance.
- Data Provenance: The tests described are engineering and laboratory tests, not clinical data from patients. Therefore, information like "country of origin of data" and "retrospective or prospective" is not relevant.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. There is no mention of "ground truth" or expert review in the context of establishing performance for a diagnostic or AI system. The "ground truth" for non-clinical tests would be the specifications and requirements of the relevant international standards.
4. Adjudication method for the test set
Not applicable, as there is no human review of a "test set" to establish ground truth.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-driven or diagnostic device that involves human readers interpreting cases.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a dental surgical system, not an algorithm or AI device. The software validation confirms the functional performance of the device's control unit, not an independent algorithm's diagnostic accuracy.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For non-clinical tests, "ground truth" would be the defined specifications and parameters of the standards (e.g., a certain voltage, current, or biocompatibility threshold).
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device that involves a training set.
9. How the ground truth for the training set was established
Not applicable. There is no training set.
Ask a specific question about this device
(207 days)
KALTENBACH & VOIGT GMBH
The MASTERtorque LUX 8900 L is intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
The MASTERtorque LUX 8900 L air-driven handpieces are dental handpieces for the use by a trained professional in the field of general dentistry. The devices are air-powered handpieces that are reusable and ergonomically shaped, and provided with a fiber optic light system. The devices can be sterilized by the steam autoclave method. Through the MULTIflex coupling connected to a dental unit, the MASTERtorque LUX 8900 L receives the air for high speed turbine, the cooling water and air for cutting treatment through pouring holes and light for illumination the operation area. The MASTERtorque LUX 8900 L is equipped with the Direct Stop Technology which reduces the stopping time of the high speed turbine and so of the bur. The back suction also will be reduced due to the Direct Stop Technology. Additionally the MASTERtorque LUX 8900 L will be supplied with an exchange filter for the water lines inside of the dental air-driven handpiece. The exchange filter filters the spray water inside the product. The filter is considered to be a spare part. If there is an insufficient amount of cooling water out of the MASTERtorque LUX 8900 L according to the instructions.for use, this spare has to be exchanged. Further one there is a jet needle supplied with the MASTERtorque LUX 8900 L. By using this part the operator is able to clean the spray holes in the head of the product if there is an insufficient amount of cooling water out of the MASTERtorque LUX 8900 L according to the instructions for use.
The provided text describes a 510(k) premarket notification for a dental handpiece, the MASTERtorque LUX 8900 L. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific acceptance criteria through a clinical study. Therefore, the information requested in the prompt, which is typical for a study demonstrating a device meets acceptance criteria, is largely not applicable to this document.
Here's a breakdown based on the information available:
Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" in the way a clinical study would define them (e.g., a specific sensitivity or specificity threshold). Instead, it compares the technological characteristics and performance of the MASTERtorque LUX 8900 L to its predicate devices to demonstrate substantial equivalence.
Table of Performance Characteristics (from original document):
Descriptive Information | MASTERtorque LUX 8900 L | SUPERTORQUE HIGH-SPEED HANDPIECES (K073478) | HIGH SPEED AIR TURBINE HANDPIECES, PAR-4HEX/4HX SERIES, TWIN Power (K061701) |
---|---|---|---|
Indications for Use | The MASTERtorque LUX 8900 L is intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry. | Identical | Identical (The PAR-4HEX/ 4 HX series handpiece is for use by authorized persons in the practice of the dentistry.) |
Principle of operation | Through the tube and the MULTIflex coupling connected to a dental unit, the air-powered handpiece receives the air for high speed turbine, the cooling water and air for cutting treatment through pouring holes and light for illumination the operation area. | Identical | Identical |
Air / water ports | Up to four (4) | Identical | Three (3) |
Fiberoptics | With built-in light system | With and without built-in light system | With and without built-in light system |
Dimensions | Headsize-Height: 13.0 mm | ||
Headsize-Diameter: 12,5 mm | Headsize-Height: 15,0 mm | ||
Headsize-Diameter: 13,0 mm | Headsize-Height: 13,2 mm | ||
Headsize-Diameter: 10,5 mm | |||
Type of chuck | Push Button | Identical | Identical |
Power (approx.) | 23 watts | 18 watts | 22 watts |
Coupling Dimensions | Length with coupling: | ||
Approx. 141 mm | Length with coupling: | ||
Approx. 130 mm | Length with coupling: | ||
Approx. 122,5 mm | |||
Chemical composition of the patient-contacting portions of the device | Stainless steel (See details in Section XV) | German silver and stainless steel | Information not available |
Chemical composition of the water / air lines | Stainless steel, german silver (nickel - chromium coated), PEEK and Fluoride Rubber Viton (See details in Section XV) | German silver and stainless steel | Information not available |
Light Intensity | Approx. 25,000 LUX | Identical | Identical |
Bur retention force | Up to 24 Ncm | Identical | Information not available |
Operating Pressure | 30 - 61 psi (41 psi recommended) | > 40 psi recommended | 19 - 42 psi (35 psi recommended) |
Idling Speed | 340,000 - 400,000 rpms | Approx. 350,000 rpms | 340,000 - 400,000 rpms |
Run-down behavior | Direct Stop | Not applicable | Quick Stop |
Compliance to Standards | Complies with ISO 7785-1. | ||
ISO 7405 and ISO 9168 | Identical | Identical | |
Lubricant | KaVo QUATTROcare | ||
(K071288) | Identical | Information not available |
The "acceptance criteria" here are implicitly that the device's characteristics and performance are comparable to the predicate devices and comply with relevant international standards. The "reported device performance" is the direct listing of these characteristics.
Study Information (Not Applicable to this 510(k) Preamarket Notification)
-
Sample size used for the test set and the data provenance: This section of the document states: "Non-Clinical Test Data: Performance tests according to the international standards for dental air-driven handpieces have been conducted to determine the conformance to the state of the art." It also states: "Clinical Test Data: Clinical testing has not been conducted on this product." Therefore, there is no "test set" in the context of clinical data. For non-clinical performance tests, no sample size or data provenance is detailed.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical test set was used. "Ground truth" for the non-clinical performance tests would be defined by the standards and engineering measurements.
-
Adjudication method for the test set: Not applicable, as no clinical test set was used.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a dental handpiece, not an AI-assisted diagnostic device.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a mechanical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the non-clinical performance data, the "ground truth" would be established by the specifications and measurement techniques dictated by the relevant ISO standards (ISO 7785-1, ISO 7405, and ISO 9168).
-
The sample size for the training set: Not applicable. This is not a machine learning or AI device.
-
How the ground truth for the training set was established: Not applicable.
In summary: The provided document is a 510(k) submission, which focuses on demonstrating substantial equivalence to existing devices through non-clinical performance testing and comparison of technological characteristics. It explicitly states that no clinical testing was conducted. Therefore, most of the requested information regarding clinical study design, sample sizes, expert involvement, and ground truth establishment from clinical data is not present and not required for this type of regulatory submission. The "study" mentioned for acceptance is implicitly the non-clinical performance testing against international standards for dental air-driven handpieces.
Ask a specific question about this device
(323 days)
KALTENBACH & VOIGT GMBH
The DIAGNOcam 2170 Is a diagnostic ald for the detection of open or includent carles lesions above the gingiva and for monitoring the progress of such lesions.
Indications:
- Detection of smooth surface caries
- Detection of occlusal carries
- Detection of proximal carles
- Detection of Initial caries
- Detection of secondary caries
- Detection of cracks
The DIAGNOcam 2170 is a handheld laser fluorescence caries detection device which uses the DIFOTI technology (Digital Imaging Fiberoptic Transillumination) as the functional principle. KaVo DIAGNOcam 2170 delivers images, which are reminiscent of X-rays but which are completely radiation free - by means of a light that is especially adapted to this examination method. The tooth structures allow the passage of light from the entry site to the camera. Areas that block light transmission (e.g. carious lesions) show up clearly as well delimited, dark areas. A digital camera captures the actual situation and makes it visible in real-time on the screen. The USB connector provides electric power to the unit. The DIAGNOcam 2170 will be delivered with a firmware (unit) at the handheld device which is responsible for steering of the camera functions and a computer based software (image) which is responsible to show / display the pictures, store / save the pictures and the possibility of a life stream. The internal laser diode generates an exact wavelength being detectable by the CCD sensor. The DIAGNOcam 2170 follows the international standards for electrical safety and electromagnetic compatibility which are applicable for the use with human beings. The DIAGNOcam 2170 works with a laser wavelength at 788 nm and an output power of 2mW for each laser diode (2 sources). The maximum output power is 21mW/cm2.
Here's an analysis of the provided text regarding the DIAGNOcam 2170, focusing on acceptance criteria and study details:
Overall Conclusion: The provided document does not contain any acceptance criteria or a study proving the device meets them. It is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices, rather than presenting de novo performance studies against pre-defined acceptance criteria.
The document explicitly states: "Clinical testing has not been conducted on this product."
Therefore, I cannot populate the table or provide answers to most of your specific questions as the relevant information is not present in the provided text.
Here's how I can address the prompts based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Not specified | No clinical performance data or acceptance criteria are presented in this 510(k) summary. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable, as no clinical test data is presented.
- Data Provenance: Not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable, as no clinical test data is presented.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable, as no clinical test data is presented.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. The document explicitly states: "Clinical testing has not been conducted on this product." The device is a diagnostic aid, not an AI system.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- This question is largely not applicable in the context of this device. The DIAGNOcam 2170 is a hardware device (laser fluorescence caries detection device) that produces images for a dentist to interpret in real-time or from stored images. It's not an "algorithm only" device in the sense of AI performing a diagnosis without human input.
- However, the document mentions "the DIAGNOcam 2170 software has been successfully validated to confirm the performance of the device." This software handles camera functions, image display, storage, and live streaming. This validation is likely focused on the software's functional correctness rather than its diagnostic performance in a standalone capacity for caries detection.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable, as no clinical test data or ground truth establishment for diagnostic performance is presented.
8. The sample size for the training set
- Not applicable, as no machine learning or AI training is discussed. The device operates on the physical principle of transillumination for caries detection, not through a trained AI model.
9. How the ground truth for the training set was established
- Not applicable, as no machine learning or AI training is discussed.
Ask a specific question about this device
(310 days)
KALTENBACH & VOIGT GMBH
The ELECTROtorque TLC 4893 is intended to convert pneumatic output from a dental treatment center to electrical energy to drive the INTRAmatic KL 702 motor for operation of electricallydriven dental handpieces. These devices are designed for use by a trained professional in the field of general dentistry.
The ELECTROtorque TLC 4893 dental control unit is a stand-alone system for operating electrically-driven KaVo handpieces. External power supply provides electric power to the unit. The 4-hole tubing connected to the unit supplies chip /cooling air, water and pressure signal. The speed of the electric handpiece is controlled by air pressure. The control unit is positioned close a treatment unit at the location preferred by the dentist. The ELECTROtorque TLC 4893 system consists of a base unit with a motor hose, an electrical motor, a transformer, and a power cord. The ELECTROtorque TLC 4893 dental control unit is a software-driven device. The software controls the following features: (1) Motor control, (2) Motor start / stop behavior, (3) Motor speed, (4) Motor performance, (5) Measuring power consumption, (6) Monitoring power consumption, and (7) SAFEdrive. The new SAFEdrive software feature monitors the power consumption of electrical hand pieces to reduce the probability or severity of overheating, thus minimizing the risk of burns to the patient. The user can navigate through the software menu via the control panel (see Figure 2.0).
The INTRAmatic KL 702 motor is an electrical low-voltage motor for dental purposes according to ISO 11498 type 2. The motor connects onto the KaVo specific tubing of the dental treatment unit ELECTROtorque. The speed of the INTRAmatic KL 702 is controlled by air pressure of the dental treatment center. The converted pneumatic output signal (electrical energy) from a dental treatment center drives the motor to operate an electrically-driven dental handpiece. Electricallydriven dental handpiece, which conform to ISO 3964, can be attached on the motor. The INTRAmatic KL 702 is intended only for dental treatment by a dental professional.
The provided text describes a medical device, the ELECTROtorque TLC 4893 with INTRAmatic KL 702, which are dental handpieces and accessories. It details the device's function, its similarities and differences to predicate devices, and the non-clinical tests performed. It explicitly states that "Clinical testing has not been conducted on these products."
Therefore, I cannot provide information on acceptance criteria or a study proving the device meets those criteria from the given text, as no clinical studies were performed. The document focuses on demonstrating substantial equivalence to legally marketed predicate devices through non-clinical testing and functional similarity.
However, I can extract the information provided about the non-clinical testing that was performed:
Non-Clinical Test Data (Acceptance Criteria and Study Details)
1. Table of Acceptance Criteria and Reported Device Performance
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
SAFEdrive Software Function | Capable of detecting electrical handpieces to reduce probability/severity of overheating; minimizing patient burn risk. | The ELECTROtorque TLC 4893 software has been successfully validated to confirm the performance of the device. The software validation included testing of the new SAFEdrive software function to demonstrate that with this added function, the ELECTROtorque TLC 4893 is capable of detective electrical handpieces to reduce the probability or severity of overheating, thus minimizing the risk of burns to the patient. |
Electromagnetic Compatibility | In accordance with applicable recognized consensus standards. | Testing of electromagnetic compatibility has been conducted in accordance with applicable recognized consensus standards. |
Electrical Safety | In accordance with applicable recognized consensus standards. | Testing of electrical safety has been conducted in accordance with applicable recognized consensus standards. |
Sterilization (INTRAmatic KL 702) | Safe and effective. | Sterilization studies were done to determine the safety and effectiveness of the INTRAmatic KL 702. |
Biocompatibility (INTRAmatic KL 702) | Safe and effective. | Biocompatibility studies were done to determine the safety and effectiveness of the INTRAmatic KL 702. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- The document does not specify the sample size for any of the non-clinical tests.
- Data provenance is not explicitly mentioned (e.g., country of origin, retrospective/prospective). The tests were likely conducted internally or by a contracted lab.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable as no clinical studies or human expert ground truth establishment for software validation are detailed in the provided text. The "ground truth" for the non-clinical tests would be defined by engineering specifications, regulatory standards, or predetermined performance parameters.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable; no adjudication method for a test set is mentioned for these non-clinical tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable; no clinical studies, and therefore no MRMC studies, were conducted. The device (ELECTROtorque TLC 4893) includes a software feature (SAFEdrive) but is not described as an "AI" system in the context of improving human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The SAFEdrive software function was validated. This would be a standalone evaluation of the algorithm's performance in monitoring power consumption and detecting overheating. The text states: "The software validation included testing of the new SAFEdrive software function to demonstrate that with this added function, the ELECTROtorque TLC 4893 is capable of detective electrical handpieces to reduce the probability or severity of overheating, thus minimizing the risk of burns to the patient." This implies an algorithm-only performance assessment against predefined overheating conditions.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the SAFEdrive software validation, the ground truth would likely be established through engineered scenarios simulating various power consumption and overheating conditions. This would involve specific temperature and energy parameters as mentioned ("Temperature and energy studies have been conducted to determine the parameters for the new SAFEdrive software function").
- For electromagnetic compatibility, electrical safety, sterilization, and biocompatibility, the ground truth is established by adherence to applicable recognized consensus standards.
8. The sample size for the training set
- Not applicable; this device is not described as involving a machine learning/AI model that requires a training set in the context of the provided text. The software validation is for a rule-based or algorithmic function.
9. How the ground truth for the training set was established
- Not applicable for the same reasons as point 8.
Ask a specific question about this device
Page 1 of 2