Search Results
Found 2 results
510(k) Data Aggregation
(348 days)
The PROPHYflex 4 Handpiece and Accessories is a dental handpiece intended for perform supragingival and subgingival prophylaxis treatments. The prophylaxis treatments are obtained by the projection of water, air and appropriate dental prophylaxis powders, onto the treatment site. PROPHYflex 4 Handpiece and Accessories in supra-gingival configuration is intended for use in the cleaning and polishing of teeth by removal soft deposits and areas of discoloration as well as preparing teeth for fillings or veneers and cleaning fixed bands and brackets on orthodontic appliances. PROPHYflex 4 Handpiece and Accessories in subgingival configuration is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
The PROPHYflex 4 is an air polishing device/powder jet handpiece that projects a mixture of water and air with sodium bicarbonate, calcium carbonate or fine glycine power onto tooth surfaces. The device includes the handpiece main body with adjustable spray feature, two cannulas, a wrench for installing the cannula, two interchangeable gripping sleeves, powder container with cover, a cleaning burr and nozzle pin, and a set of spare o-rings. An optional adapter and perio tip can be installed for subgingival treatments.
The powder is filled into the powder container by the user. When activated through foot control, air is pushed into the powder container, swirling the powder and pushing it through the device and out of the cannula. The spray is only mixed with the air and powder at the very tip of the cannula, providing an even and gentle cleaning effect. The amount of powder can be regulated via three different spray settings by the adjusting ring, helping the users to perform individual treatments such as subgingival biofilm removal, removal of discoloration and bacterial plaque, orthodontics, cleaning prior to fissure sealing, prosthetics, conservative and aesthetic dentistry by providing the possibility to adjust the cleaning effects to the clinical situation of the patient at hand.
The components of the PROPHYflex 4, including the powder container, is to be completely reprocessed after each use.
The three powder levels are as follows:
- Highest Level: Suitable for supragingival treatment and offers full cleaning power .
- Medium Level: Suitable for sub- and supragingival treatment and enables . gentle cleaning at a reduced powder quantity
- . Lowest Level: Helps rinsing the powder off the tooth and blowing the powder from the system after the treatment
During supraqinqival treatments (e.g. stain removal), the device doesn't touch the patient but will be placed 3 – 5 mm from the area that is to be cleaned. During subgingival treatments (biofilm management) with the perio tip, the tip is inserted into the pocket up to 5mm and angled towards the tooth.
The provided text is a 510(k) summary for a dental handpiece, the PROPHYflex 4. It focuses on demonstrating substantial equivalence to predicate devices through technological characteristic comparison and non-clinical performance testing.
Crucially, this document explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product."
Therefore, a study demonstrating the device meets acceptance criteria via clinical performance (which would involve human readers, ground truth established by experts, MRMC studies, etc.) was explicitly not performed and is not required for this type of device according to the provided FDA submission.
However, I can extract the acceptance criteria and reported performance based on the non-clinical (bench) testing, and other relevant information from the document, framing it within the context of a "study" as requested, even if it's a bench study rather than a clinical one.
Here's a breakdown based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for the PROPHYflex 4 are implicitly defined by its ability to perform as well as, or better than, its predicate devices across various technical specifications and to comply with relevant international standards. The study proving these criteria are met consists of non-clinical bench testing.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are essentially demonstrated equivalence or improvement over the predicate device (AIR-FLOW Handy 3.0 Plus (K151912)) and compliance with specified ISO standards.
Feature / Acceptance Criterion (Implicit) | Acceptance Criterion (Specific Value from Predicate or Standard) | Reported Device Performance (PROPHYflex 4) | Met? |
---|---|---|---|
Indications for Use (Subgingival) | Non-surgical removal of subgingival plaque in pockets up to 4 mm (Predicate H3.0 Plus), Up to 5mm (Reference Predicate H3.0 PERIO) | Non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. | Yes |
Max Mode Operation | Present (Predicate H3.0 Plus) | Present | Yes |
Internal Spray | Present (Predicate H3.0 Plus) | Present | Yes |
Coupling Dimensions | Complies with ISO 7785-1, ISO 14457, and ISO 9168 (Predicate H3.0 Plus) | Complies with ISO 7785-1, ISO 14457, and ISO 9168. (Specifically mentioning KaVo MULTIflex-Coupling, Sirona Coupling compatibility) | Yes |
Biocompatibility | According to ISO 10993-1 (Predicate H3.0 Plus) | Biocompatible according to ISO 10993-1 (Biological evaluation of medical devices), ISO 10993-5 (In vitro cytotoxicity), ISO 10993-10 (Irritation/sensitization), ISO 10993-18 (Chemical characterization). | Yes |
Noise Level | 74.9 dB(A) (Predicate H3.0 Plus) | 56.6 - 69.0 dB(A) (Quieter than predicate) | Yes |
Operating Temperature (Grip Sleeve) | 19.93°C - 23.33°C (Predicate H3.0 Plus) | 18.0°C - 25.1°C (Wider range, but within acceptable dental operating temperatures) | Yes |
Maximum Air Pressure | 2.7 – 3.5 bar (39 – 51 psi) (Predicate H3.0 Plus) | 3.2 - 5 bar (46 - 73 psi) (Higher pressure capability) | Yes |
Air and water supply | 10.3 – 13.5 Nl/min (Predicate H3.0 Plus) | 6.4 – 14.3 Nl/min (Comparable range) | Yes |
Maximum Water Pressure | 1.0 – 2.2 bar (15 – 32 psi) (Predicate H3.0 Plus) | 1.0 - 2.5 bar (15 - 36 psi) (Higher pressure capability) | Yes |
Spray Water Flow | 106 ml/min (Predicate H3.0 Plus) | 28 – 54 ml/min (Lower flow, noted as a difference but not raising new concerns) | Yes |
Compliance with Powder Jet Handpiece Standard | DIN 19402 (Predicate H3.0 Plus) | DIN EN ISO 20608 (Dentistry - Powder jet handpieces and powders), DIN 19402 (Powder Jet Handpieces) | Yes |
Additional Operational Modes | N/A (Predicate H3.0 Plus only has Max Mode) | Middle Mode (reduced powder), Spray-function Mode (water only) | N/A (Improvement over predicate) |
Device Dimensions (Length) | 200 mm (Predicate H3.0 Plus) | Up to 190 mm (10 mm shorter) | Yes |
Device Dimensions (Largest Diameter) | 43 mm (Predicate H3.0 Plus) | Up to 40 mm (3 mm thinner) | Yes |
Reprocessing Method | N/A (Standard dental practice) | Components are to be completely reprocessed after each use (standard for dental handpieces) | Yes |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify exact "sample sizes" in terms of number of devices tested for each bench test. These are typically covered under the manufacturer's internal quality system documentation and test protocols, which are summarized rather than detailed in a 510(k). The tests are performed on representative units of the device.
- Data Provenance: The testing was non-clinical (bench testing).
- Country of Origin: The manufacturer is Kaltenbach & Voigt GmbH, located in Biberach, Germany. Therefore, the testing would likely have been conducted in Germany or by approved testing facilities.
- Retrospective or Prospective: This refers to the collection of patient data, not bench testing. Bench testing is inherently prospective in its execution, demonstrating performance against specifications.
3. Number of Experts Used to Establish Ground Truth and Qualifications of Those Experts
- This question is not applicable as the study was purely non-clinical bench testing. There were no "experts" establishing a clinical "ground truth" for a test set, as no clinical test set involving human subjects was used. The "ground truth" for bench testing is derived from engineering specifications, established medical device safety and performance standards (e.g., ISO standards), and comparative performance against legally marketed predicate devices.
4. Adjudication Method for the Test Set
- This question is not applicable as there was no test set requiring human expert adjudication (e.g., image interpretation). Bench testing results are typically objectively measured against defined criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was explicitly NOT done. The document states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product." Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- This question is not applicable in the context of a dental handpiece. The device is a physical instrument, not an AI algorithm. Its "performance" is its mechanical, fluidic, and material characteristics demonstrated through bench testing.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance validation is based on:
- International Standards: Compliance with ISO standards (e.g., ISO 7785-1, ISO 14457, ISO 9168, ISO 10993 series, ISO 20608).
- Predicate Device Performance: Demonstrating that the PROPHYflex 4's performance characteristics (dimensions, noise, pressure, flow, etc.) are comparable to or improved over the legally marketed predicate device (AIR-FLOW Handy 3.0 Plus (K151912)) and other reference devices.
- Engineering Specifications: Internal design specifications and performance targets set by the manufacturer for the device.
8. The Sample Size for the Training Set
- This question is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a "training set" of data.
9. How the Ground Truth for the Training Set Was Established
- This question is not applicable as this is a physical medical device, not an AI/ML algorithm; therefore, there is no "training set" or "ground truth" establishment for it in this context.
Ask a specific question about this device
(307 days)
The Pelton & Crane P50 series Dental Operative Units and Accessories are intended to supply power to and serve as a base for other dental devices and accessories by providing air, water, vacuum and low voltage electrical power to dental instruments and dental handpieces. The Pelton & Crane P50 series Dental Operative Units and Accessories are designed for use by a trained professional in the field of dentistry.
The P50 series Dental Operative Unit and Accessories serves as a base that includes components to deliver air, water, electrical power, and vacuum to dental handpieces, instruments, and accessories. The controls are contained in a Doctor's Unit, an Assistant's Unit, and a Cuspidor. Additional parts include mount arms, patient dental chair, foot control, and a junction box that houses a power supply and air/water regulators. The unit may also include a dental operating light and a monitor. Various handpieces, instruments, and accessories can be added to the P50 series Dental Operative Unit which Pelton & Crane does not manufacture but does provide a means to connect them to the P50 series Dental Operative Units. These include, but not limited to, pneumatic handpieces, electric motors with handpieces, scalers, intra-oral cameras, curing lights, air/water syringes, SE and HVE vacuum instruments. Water quality is maintained automatically by the preprogramed cleaning functions for the water lines, cuspidor spouts, and instrument hoses including motors and syringes. Instructions are provided for waterline treatment and suction cleaning agents. Water supply is available by either the municipal water supply or a self-contained water bottle.
The retrieved document is a 510(k) Pre-Market Notification for the Pelton & Crane P50 series Dental Operative Unit and Accessories. This notification focuses on demonstrating substantial equivalence to a predicate device rather than providing a study proving performance against specific acceptance criteria for a novel device. Therefore, much of the requested information regarding acceptance criteria, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies is not applicable or available in this type of submission.
However, I can extract the relevant information that is present:
1. A table of acceptance criteria and the reported device performance
The document does not present specific acceptance criteria in the form of a table with numerical thresholds for performance metrics. Instead, it provides a comparison table (Table 5.3) outlining technological characteristics between the proposed device (P50 series Dental Operative Unit and Accessories) and the predicate device (ESTETICA Dental Treatment Unit and Accessories, K161488). The document states that "The performance testing results provided in this submission supports that the proposed device performs as well as the predicate devices for its intended use" and that "Hence, the device is deemed to be substantially equivalent to the ESTETICA Dental Treatment Unit and Accessories (K161488)."
Here is a summary of the comparative characteristics, serving as the basis for "performance" as described in this 510(k) submission:
Feature | Proposed Device (P50 series Dental Operative Unit and Accessories) | Predicate Device (ESTETICA Dental Treatment Unit and Accessories) |
---|---|---|
Intended Use / Indications | Intended to supply power to and serve as a base for other dental devices and accessories by providing air, water, vacuum, and low voltage electrical power to dental instruments and dental handpieces. Designed for use by a trained professional in the field of dentistry. | Intended to supply power to and serve as a base for other dental devices and accessories by providing air, water, vacuum, and low voltage electrical power to dental instruments and dental handpieces. Designed for use by a trained professional in the field of general dentistry. |
Regulation Number | 21 CFR 872.6640 | 21 CFR 872.6640 |
Regulation Title | Dental operative unit and accessories | Dental operative unit and accessories |
Regulation Class | I | I |
Product Code | EIA | EIA |
Power and Utility Supply | 115V/230V AC electrical supply, compressed air and water | 100V/240V AC electrical supply, compressed air and water |
Protection Class | Class 1 Equipment | Class 1 Equipment |
Applied Parts | Type B | Type B |
Control of Air and Water | Uses pneumatically controlled valves to control the flow of air and water. On/off and intensity controlled by foot pedal. | Uses pneumatically controlled valves to control the flow of air and water. On/off and intensity controlled by foot pedal. |
Dentist Element Positions | 6 handpiece positions + 1 additional | 6 handpiece positions + 1 additional |
Max. Motors | Up to 4 motors (electrical motors with and without light) | Up to 4 motors (electrical motors with and without light) |
Electric Motors | INTRA LUX KL 703 LED, INTRA LUX S600 LED | INTRA LUX KL 703 LED, INTRA LUX S600 LED |
User Interface | EasyTouch Display, color. Hands-free operation via foot control. | ESTETICA E70/E80 Vision: EasyTouch Display, color. ESTETICA E50 Life: Display, with several keys (foil) and Status LED. Hands-free operation via foot controller. |
Syringe | 3 Functional syringe without heater and light or Quick Clean air/water syringe | Multifunctional syringe with internal heater for water and light or 3 Functional syringe without heater and light |
Turbines | Turbines with light, MULTIflex couplings | Turbines with light, MULTIflex couplings |
Control of Coolant | Water pressure controlled by the unit, Air pressure controlled by the unit | Water pressure controlled by the unit, Air pressure controlled by the unit |
Scaler | PiezoLED with light, SONICflex, CAVITRON G139 | PiezoLED with light, SONICflex |
Intraoral Camera | DIAGNOcam 2170 | DIAGNOcam 2170 |
Curing Light | Satelec Mini LED | Satelec Mini LED |
X-ray Viewer | N/A | Panorama X-ray image viewer or small X-ray viewer |
Endodontic Functions | Implantology functions, Endodontic treatment functions, Adapted pump for saline solution | Implantology functions, Endodontic treatment functions, Adapted pump for saline solution |
Interfaces | USB Interface | USB Interface |
Assistant Element Positions | Up to 3 suction positions, 1 handpiece position (syringe) | 2/3 suction positions, 2 handpiece positions (syringe and curing light) |
Assistant Element UI | Touch pad (foil) with several keys and status LEDs | Touch pad (foil) with several keys and status LEDs |
Assistant Element Syringe | 3 Functional syringe without heater and light or Quick Clean air/water syringe | Multifunctional syringe with internal heater for water and light or 3 Functional syringe without heater and light |
Assistant Element Curing Light | Satelec Mini LED | Satelec Mini LED |
Suction | Suction devices, Saliva ejector, HVE (High volume evacuator), Surgical suction devices | Suction devices, Saliva ejector, HVE (High volume evacuator), Surgical suction devices |
Assistant Element Interfaces | No USB interface | USB interface |
Chair Patient Load | SP30 Chair - 450 lbs, SP18 Chair - 350 lbs, SP17 Chair - 350 lbs | ESTETICA E70/E80 Vision: 396 lbs, ESTETICA E50 Life: 407 lbs (Standard Chair), ESTETICA E50 Life: 297 lbs (Compact Chair) |
Headrest | Manually operated | Manually operated or motorized headrest |
Upholstery | Naugasoft or Ultraleather upholstery | Skai upholstery |
Armrest | Armrest, retractable, rotate | Armrest, retractable |
Chair User Interface | Foot-switch and touchpad | Foot-switch |
Chair Movement | Synchronized Backrest and seat movements, User programmable positions, Backrest hydraulic driven, Seat and seat inclination hydraulic driven | Synchronized Backrest and seat movements, User programmable positions, Backrest motor driven, ESTETICA E80 Vision: Horizontal movement, seating bench lifting, Seat and seat inclination motor driven |
Cuspidor bowl | Fixed bowl optional | Manually adjustable or Motor driven cuspidor bowl |
Amalgam separator | N/A | Amalgam separator |
Water Supply | Connection to municipal water or Self-contained water bottle | Connection to municipal water with safety separation via "air gap". Self-contained Water bottle option for E50. |
Water unit Heating | No heater | Water heater |
Waterline Treatment Solution | Waterline Treatment solution | Waterline Treatment solution |
Water / Suction Hose Treatment | Automated instrument cleaning, Manual suction hose cleaning w/suction line cleaner | Automated instrument cleaning, Automated suction hose cleaning w/suction line cleaner |
Water treatment | Waterline Treatment solution | Waterline Treatment solution |
Purge | Purge function | Purge function |
Pre-programmed waterline cleaning functions | Automatic for water lines, cuspidor, spouts, and instrument hoses including motors and syringes | Automatic for water lines, cuspidor, spouts, and instrument hoses including motors and syringes |
Daily Cleaning | Automated instrument flush with water before start up and after down times (Recommend 2 minutes per instrument). Automatic after patient cleaning of water and drainage system via fresh water and cleaning solution if used (Recommended 20 seconds). | Automated instrument flush with water before start up and after down times (Recommend 2 minutes per instrument). Automatic after patient cleaning of water and drainage system via fresh water and cleaning solution if used (Recommended 20 seconds). |
Weekly Cleaning | Automated flush with cleaning solution | Automated flush with cleaning solution |
Water Cleaner | KaVo OXYGENAL 6 | KaVo OXYGENAL 6 |
Water Line Treatment | Recommended 45 minutes cleaning time to complete. | Recommended 45 minutes cleaning time to complete. |
Recommended disinfectors | In accordance to EN ISO 15883-1 using alkaline cleaning agents | In accordance to ISO 15883-1 using alkaline cleaning agents |
Hygiene Center | External holder for hoses and syringes during automated waterline cleaning | External holder for hoses and syringes during automated waterline cleaning |
Operating Light | Helios dental light 3000 - LED, Helios dental light 1800 - LED | LEDview - LED |
Monitor | N/A | 19″ or 22″ monitor on the light support frame |
Foot Control | Standard DCI foot switch with chair foot control, Electronic foot control | Wireless Foot control available, Electronic foot control |
PC Interface | Yes | Yes |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The submission relies on "Non-Clinical Test Data" to demonstrate substantial equivalence, which primarily involves performance bench testing against international standards and comparative performance testing of integrated accessories versus cleared stand-alone devices. There is no mention of a traditional "test set" in the context of clinical data for performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as there was no clinical test set requiring ground truth establishment by experts for a specific performance metric.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable as there was no clinical test set requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a "Dental Operative Unit and Accessories" and is a physical medical device, not an AI-powered diagnostic or assistive tool for human readers. Its primary function is to provide power, air, water, and vacuum for other dental instruments.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical dental unit, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical performance testing, the "ground truth" was adherence to established international standards (e.g., AAMI ES60601-1, IEC 60601-1-2, ISO 10993-1, etc.) and comparative performance to cleared stand-alone predicate devices for integrated accessories. Specific outcomes data or expert consensus on clinical findings are not mentioned as "ground truth" in this context.
8. The sample size for the training set
This is not applicable as this is not an AI/ML device requiring a training set.
9. How the ground truth for the training set was established
This is not applicable as this is not an AI/ML device requiring a training set.
Summary of the Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the acceptance criteria is described as "Non-Clinical Test Data" and primarily involves:
- Performance bench testing: Conducted according to international standards for dental operative units to determine conformance in regards to basic safety and essential performance. Relevant standards cited include:
- AAMI ES60601-1:2005 +AC1; A2
- IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007)
- IEC 60601-1-2 Edition 3: 2007-03 (EMC)
- IEC 80601-2-60 Edition 1.0 2012-02 (Dental equipment specific)
- ISO 9168 Third edition 2009-07-15 (Hose connectors)
- ISO 14457 First edition 2012-09-15 (Handpieces and motors)
- ISO 7494-2 Second edition 2015-04-01 (Air, water, suction)
- ISO 7494-1 Second edition 2011-08-15 (General requirements)
- Biocompatibility testing: Completed for applicable components, referencing ISO 10993-1 and ISO 7405. Reference devices (Spirit - K143696 and ELECTROmatic - K163317) were used due to identical patient contacting parts.
- Software documentation: For moderate level of concern per FDA Guidance Document for Software Contained in Medical Devices, referencing IEC 62304.
- Comparative performance testing: The functions of the integrated accessories were compared to "cleared stand-a-lone device[s]."
- Risk Management: Compliance with ISO 14971:2007, with the Risk Management File reviewed by Intertek for 60601-1 certification.
The document explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product."
Ask a specific question about this device
Page 1 of 1