Search Results
Found 3866 results
510(k) Data Aggregation
(247 days)
Shanghai United Imaging Healthcare Co., Ltd.
Ask a specific question about this device
(30 days)
VivaChek Biotech (Hangzhou) Co., Ltd
Ask a specific question about this device
(265 days)
Bosch Healthcare Solutions GmbH
Ask a specific question about this device
(61 days)
Fisher & Paykel Healthcare Ltd
Ask a specific question about this device
(81 days)
Shandong Guyue Healthcare Appliance Co., Ltd.
Ask a specific question about this device
(22 days)
Siemens Healthcare GmbH
Ask a specific question about this device
(142 days)
Carestream Health Inc.
Ask a specific question about this device
(85 days)
TOUCHBEAUTY BEAUTY & HEALTH (SHENZHEN) CO., LTD
The Red Light is intended to treat full face wrinkles.
The Blue Light is intended to treat mild to moderate inflammatory acne.
The Yellow Light is intended to treat wrinkles.
The GLAM LED light therapy mask is a home use wearable light emitting diode phototherapy device whose purpose is to produce an even, cool, narrow band of light for the treatment of full face wrinkles and mild to moderate acne of the face. The outer shell of the mask is manufactured from Polyethylene terephthalate (PET). The inner shell is a clear Polycarbonate (PC). The Light emitting diodes are mounted behind the clear Polycarbonate. The LEDs generate the light. The ear hooks are made of Acrylonitrile butadiene styrene (ABS) and the silicone goggle protect the eyes from LED lights. Unfold the ear hooks and place the mask on your face, the mask will automatically activate the light therapy mode. The LEDs produce blue, red and yellow light in the visible spectrum (Blue:415nm +/- 10nm, Yellow: 590nm +/-10nm, Red: 625nm +/-10nm.). The device works by emitting the specified wavelengths to treat full face of wrinkles or to treat mild to moderate inflammatory acne. Press the touch switch on the right ear hook twice to select the light therapy mode you want to use. Each mode operates in a 15-minute cycle. After 15 minutes, the device automatically deactivates the light mode and enters the standby mode.
This FDA 510(k) clearance letter for the GLAM LED Facial Mask (TB-2386F) does not include any information regarding clinical testing, acceptance criteria for device performance related to efficacy (e.g., wrinkle reduction or acne treatment), or studies comparing the device's performance against such criteria.
The letter focuses on the substantial equivalence based on:
- Technological Comparison: Comparing light wavelengths, power density, and treatment time to predicates.
- Safety Standards Compliance: Verification through non-clinical tests (electrical safety, EMC, biocompatibility, photobiological safety, software verification).
Therefore, I cannot provide details on the specific acceptance criteria for efficacy or a study proving the device meets those criteria from the provided document. The 510(k) summary explicitly states: "No animal or clinical study is included in this submission."
However, if we were to hypothesize what acceptance criteria and a study might look like for a device with these indications, and then illustrate what would be missing from this document:
Hypothetical Acceptance Criteria and Performance Study (Not Found in Provided Document):
Since the provided document explicitly states no clinical study was included, the following tables and sections represent what would typically be expected for a device claiming therapeutic efficacy, but none of this information is present in the provided 510(k) clearance letter.
Hypothetical Acceptance Criteria and Reported Device Performance (If Clinical Data Existed)
Acceptance Criteria (Hypothetical) | Reported Device Performance (Hypothetical) |
---|---|
Red Light (Wrinkle Treatment): | |
Primary Endpoint: A statistically significant reduction (e.g., >20%) in the appearance of fine lines and wrinkles (e.g., Fitzpatrick Wrinkle Scale, or qualitative dermatological grading) across the full face after 8-12 weeks of treatment, compared to baseline or a control group. | Not reported in 510(k) letter. If available, this section would state the measured percentage reduction in wrinkles, p-values, confidence intervals, and the specific grading scale used. Example: "Mean reduction of 32% in Fitzpatrick Wrinkle Scale scores for fine lines after 10 weeks (p 70% of subjects reporting improvement) in overall skin appearance, texture, or satisfaction as reported by subjects via a validated questionnaire. |
Ask a specific question about this device
(51 days)
Hangzhou Clongene Biotech Co.,Ltd.
The CLUNGENE Multi-Drug Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:
Drug | Calibrator | Cut-off (ng/mL) |
---|---|---|
Morphine (MOP/OPI300) | Morphine | 300 |
Morphine (MOP/OPI2000) | Morphine | 2,000 |
Methamphetamine (mAMP/MET1000) | D-Methamphetamine | 1,000 |
Methamphetamine (mAMP/MET500) | D-Methamphetamine | 500 |
Cocaine (COC300) | Benzoylecgonine | 300 |
Cocaine (COC150) | Benzoylecgonine | 150 |
Marijuana (THC) | 11-nor-9-THC-9-COOH | 50 |
Methylenedioxymethamphetamine (MDMA) | D,L-Methylenedioxymethamphetamine | 500 |
Buprenorphine (BUP) | Buprenorphine | 10 |
Propoxyphene (PPX) | D-Propoxyphene | 300 |
Amphetamine (AMP1000) | D-Amphetamine | 1,000 |
Amphetamine (AMP500) | D-Amphetamine | 500 |
Phencyclidine (PCP) | Phencyclidine | 25 |
Methadone metabolite (EDDP) | 2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
Oxycodone (OXY) | Oxycodone | 100 |
Oxazepam (BZO) | Oxazepam | 300 |
Nortriptyline (TCA) | Nortriptyline | 1,000 |
Secobarbital (BAR) | Secobarbital | 300 |
Methadone (MTD) | Methadone | 300 |
6-Monoacetylmorphine (6-MAM) | 6-Monoacetylmorphine | 10 |
Fentanyl (FYL) | Fentanyl | 1 |
The single or multi-test cups can consist of any combination of the above listed drug analytes, but only one cut off concentration under same drug condition will be included per device.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The CLUNGENE Multi-Drug Home Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:
Drug (Identifier) | Cut-off (ng/mL) |
---|---|
Morphine (MOP/OPI2000) | 300 or 2000 |
Methamphetamine (mAMP/MET) | 500 or 1,000 |
Cocaine (COC) | 150 or 300 |
Marijuana (THC) | 50 |
Methylenedioxymethamphetamine (MDMA) | 500 |
Buprenorphine (BUP) | 10 |
Propoxyphene (PPX) | 300 |
Amphetamine (AMP) | 500 or 1,000 |
Phencyclidine (PCP) | 25 |
Methadone metabolite (EDDP) | 300 |
Oxycodone (OXY) | 100 |
Oxazepam (BZO) | 300 |
Nortriptyline (TCA) | 1,000 |
Secobarbital (BAR) | 300 |
Methadone (MTD) | 300 |
6-Monoacetylmorphine (6-MAM) | 10 |
Fentanyl (FYL) | 1 |
The single or multi-test cup offers any combination from above 1 to 17 drugs, but only one cut off concentration under same drug condition will be included per device. |
The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC-MS/MS is the preferred confirmatory method.
It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.
CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
This document provides details on the performance characteristics of the CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup. Since this is an in vitro diagnostic device (specifically, a drug screening test), the acceptance criteria and study design are typically focused on analytical performance (accuracy, precision, analytical specificity) rather than a multi-reader multi-case (MRMC) comparative effectiveness study, which is more common for imaging AI. Similarly, "human readers improving with AI vs without AI" is not applicable here as the device is the test, not an aid to human interpretation of another modality.
Here's the breakdown based on the provided text:
Acceptance Criteria and Reported Device Performance
The core acceptance criterion for qualitative drug screening tests like this is accurate detection around a specific cutoff concentration. The reported performance demonstrates the device's ability to correctly classify samples as positive or negative relative to these cutoffs.
Table of Acceptance Criteria and Reported Device Performance (Analytical Precision/Reproducibility)
The "Acceptance Criteria" column represents the desired performance for a qualitative assay around its cutoff. For positive results, this means detecting drug concentrations above the cutoff, and for negative results, it means not detecting concentrations below the cutoff. The provided precision data shows the number of positive (+) and negative (-) results out of 50 tests for various concentrations relative to the cutoff. An ideal performance would show 100% positive for concentrations above cutoff and 100% negative for concentrations below, with roughly 50/50 split at the cutoff itself (due to inherent variability).
Drug (Cut-off ng/mL) | Acceptance Criteria (Implicit for qualitative assay) | Reported Device Performance (Accuracy as evidenced by reproducibility at various concentrations) Number of negative/positive results out of 50 tests. Values are aggregated across 3 lots where available. |
---|---|---|
MOP300 | All samples >cutoff should test positive; all samples |
Ask a specific question about this device
(209 days)
Heager Gmbh
The Heager Medical Laser Family device is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures. The device's 980nm laser is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology. The device is further indicated for laser assisted lipolysis. The device's 1470nm laser is intended for delivery of laser light to soft tissue in non-contact mode during general surgery procedures, indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.
The Heager Medical Laser Family generates a 980nm wavelength laser to act on a target tissue to achieve hemostasis, ablation, and coagulation of the target tissue. The Heager Medical Laser Family generates a 1470 nm wavelength laser that acts on the water molecules of the target tissue to achieve the function of treatment of reflux of the saphenous veins associated with varicose veins and varicosities.
The Heager Medical Laser Family has following characteristics:
- Dual wavelengths
- Temperature monitoring system
This document is a 510(k) clearance letter for a medical laser device, not a study report. Therefore, it does not contain the specific details about acceptance criteria and study results you requested in the format typically found in a clinical or performance study.
Based on the provided FDA 510(k) clearance letter for the Heager Medical Laser Family, here's what can be extracted and what information is not available within this document:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative, measurable way for device performance in terms of clinical outcomes or diagnostic accuracy. Instead, it demonstrates substantial equivalence to predicate devices based on technical specifications and adherence to recognized electrical and safety standards.
The "reported device performance" is primarily presented as technical specifications and compliance with safety standards rather than clinical performance metrics.
Feature/Standard | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Wavelength | Within specified range | 980nm±20nm, 1470nm±20nm |
Laser Output Power | Within specified range | 980nm±20nm cw 20W, 1470nm±20nm cw 15W |
Aiming Beam | Within specified range | Diode laser of 650nm, power Max. 10mW, adjustable brightness |
Pulse Width Range | Appropriately wide for intended use, similar to reference devices | 0.1s - 10s (stepping 0.1s) |
Electrical Safety | Compliance with IEC 60601-1 ed. 3.2, IEC 60825-1 ed. 2.0, IEC 60601-2-22 ed. 4.0 | Complies with IEC 60601-1:2005+A1:2012+A2:2020, IEC 60825-1:2014, IEC 60601-2-22:2019 (Note: IEC 60825-1 is listed as ed 2.0 in section 7.0 and 2014 in section 12, need clarification if 2014 is equivalent to or supersedes 2.0) |
Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 ed. 4.1 | Complies with IEC 60601-1-2:2014+A1:2021 |
User Interface (IEC 60601-1-6) | Compliance with IEC 60601-1-6 ed. 3.2 | Complies with IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION |
Biocompatibility of contact materials (implied for fiber) | Compliance with ISO 10993 series | Fibers must meet ISO 10993 series standards |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not available in the provided document. As this is a 510(k) for a laser surgical instrument, the primary "test set" would be the device itself tested against engineering and safety standards, not a clinical data set in the way an AI/diagnostic device would have. The document explicitly states: "No clinical study implemented for the Diode laser therapy device."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not available. Given that "No clinical study implemented," there would be no ground truth established by experts in a clinical context for the purpose of a study demonstrating effectiveness. The "ground truth" here is the adherence to engineering standards and technical specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not available. Adjudication methods are typically associated with clinical studies involving human observers or raters, which were not conducted for this 510(k) submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and not available. This is a laser surgical instrument, not an AI software/diagnostic device that would involve human readers or AI assistance in interpretation.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable and not available. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" in this 510(k) context is primarily compliance with recognized electrical, safety, and performance standards (e.g., IEC standards) and technical specifications of the device, demonstrated through non-clinical testing.
8. The sample size for the training set
This information is not applicable and not available. There is no "training set" in the context of an AI/machine learning model for this type of laser device 510(k) submission.
9. How the ground truth for the training set was established
This information is not applicable and not available for the same reason as point 8.
Ask a specific question about this device
Page 1 of 387