K Number
K251168

Validate with FDA (Live)

Device Name
Image Suite
Date Cleared
2025-09-04

(142 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Image Suite is a stand-alone radiographic imaging software designed to perform patient registration, review, report, archive, and print patient images. The software interacts with Carestream CR or DR to receive radiographic images for review, and has the ability to receive, archive, and review DICOM images from compatible third-party modalities (for example, US, MR, CT). The software can also update Carestream CR or DR device firmware and monitor device calibration. The software serves as a web server to support web clients for image viewing. Image Suite is intended for use by radiologists and trained healthcare professionals. The software presents images to medical professionals in a convenient digital medium so they can make diagnostic and/or therapeutic decisions.

This excludes mammography applications and Tablet Viewer applications in the United States.

Device Description

Image Suite software is a Picture Archiving and Communication System (PACS) that allows patient registration, image acquisition, processing, reviewing, reporting, archiving and printing of radiographic images from compatible Carestream image acquisition devices, such as digital radiography (DR) detectors or computed radiography (CR) systems. In addition to being designed to function with images acquired from Carestream compatible devices, Image Suite can receive DICOM images from compatible third-party modalities (for example, Ultrasound, MRI, CT). Third party images may be viewed and archived by Image Suite, but no changes to the raw image from the third-party device are possible.

This Image Suite submission adds features that are commercially available on other Carestream products. Image Suite utilizes the same image processing software (Eclipse) as the Carestream CR and DR devices that the features being detailed in this submission were previously cleared on.

These features include:

  • Companion Imaging
    • Tube & PICC Line /High Detail Visualization/Skeletal Enhancement (commercially available option in x-ray systems under K120062)
    • Pneumothorax visualization (does not detect the pneumothorax) (commercially available option in x-ray systems K120062)
    • Bone Suppression (previously cleared in K133442)
  • Multiple Looks (subset of the looks provided in the commercially available x-ray systems)
  • Smart Grid (cleared under K163157).

This submission will also address cumulative changes made since the last 510(k) submission K140271. Over time minor changes have been made to the Image Suite product that have not impacted safety or effectiveness of the product.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Image Suite

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 4, 2025

Carestream Health Inc.
Jessica Deryke
Regulatory Affairs Manager
150 Verona Street
Rochester, New York 14608

Re: K251168
Trade/Device Name: Image Suite
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: August 5, 2025
Received: August 5, 2025

Dear Jessica Deryke:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251168 - Jessica Deryke
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

Page 3

K251168 - Jessica Deryke
Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K251168

Please provide the device trade name(s).

Image Suite

Please provide your Indications for Use below.

Image Suite is a stand-alone radiographic imaging software designed to perform patient registration, review, report, archive, and print patient images. The software interacts with Carestream CR or DR to receive radiographic images for review, and has the ability to receive, archive, and review DICOM images from compatible third-party modalities (for example, US, MR, CT). The software can also update Carestream CR or DR device firmware and monitor device calibration. The software serves as a web server to support web clients for image viewing. Image Suite is intended for use by radiologists and trained healthcare professionals. The software presents images to medical professionals in a convenient digital medium so they can make diagnostic and/or therapeutic decisions.

This excludes mammography applications and Tablet Viewer applications in the United States.

Please select the types of uses (select one or both, as applicable).

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Image Suite Page 8 of 60

Page 5

510(k) Summary

510(k) Owner Name: Carestream Health, Inc.
510(k) Owner Address: 150 Verona Street
Rochester, NY, 14608
510(k) Owner Phone: 585-627-6505
510(k) Owner Fax: 585-627-8802
Contact Person & Info: Jessica DeRyke
Regulatory Affairs Manager
Jessica.deryke@carestream.com
585-489-7627
Date Summary Prepared: September 3, 2025

Device Classification Information:

Predicate

510(k) Submitter: Carestream Health, Inc.
Trade Name: Tablet Viewer Software for Image Suite
Device: System, Image Processing, Radiological
Regulation Description: Picture Archiving and Communication System
Review Panel: Radiology
Product Code: LLZ
Regulation Number: 21 CFR 892.2050
Device Class: II

Subject Device

510(k) Submitter: Carestream Health, Inc.
Trade Name: Image Suite
Device: System, Image Processing, Radiological
Regulation Description: Medical Image Management and Processing System
Review Panel: Radiology
Product Code: LLZ
Regulation Number: 21 CFR 892.2050
Device Class: II

Page 1 of 6

Page 6

K251168
Page 2 of 6

Device Description:

Image Suite software is a Picture Archiving and Communication System (PACS) that allows patient registration, image acquisition, processing, reviewing, reporting, archiving and printing of radiographic images from compatible Carestream image acquisition devices, such as digital radiography (DR) detectors or computed radiography (CR) systems. In addition to being designed to function with images acquired from Carestream compatible devices, Image Suite can receive DICOM images from compatible third-party modalities (for example, Ultrasound, MRI, CT). Third party images may be viewed and archived by Image Suite, but no changes to the raw image from the third-party device are possible.

This Image Suite submission adds features that are commercially available on other Carestream products. Image Suite utilizes the same image processing software (Eclipse) as the Carestream CR and DR devices that the features being detailed in this submission were previously cleared on.

These features include:

  • Companion Imaging
    • Tube & PICC Line /High Detail Visualization/Skeletal Enhancement (commercially available option in x-ray systems under K120062)
    • Pneumothorax visualization (does not detect the pneumothorax) (commercially available option in x-ray systems K120062)
    • Bone Suppression (previously cleared in K133442)
  • Multiple Looks (subset of the looks provided in the commercially available x-ray systems)
  • Smart Grid (cleared under K163157).

This submission will also address cumulative changes made since the last 510(k) submission K140271. Over time minor changes have been made to the Image Suite product that have not impacted safety or effectiveness of the product.

Indication for Use:

Image Suite is a stand-alone radiographic imaging software designed to perform patient registration, review, report, archive, and print patient images. The software interacts with Carestream CR or DR to receive radiographic images for review, and has the ability to receive, archive, and review DICOM images from compatible third-party modalities (for example, US, MR, CT). The software can also update Carestream CR or DR device firmware and monitor device calibration. The software serves as a web server to support web clients for image viewing. Image Suite is intended for use by radiologists and trained healthcare professionals. The software presents images to medical professionals in a convenient digital medium so they can make diagnostic and/or therapeutic decisions.

This excludes mammography applications and Tablet Viewer applications in the United States

Page 7

Traditional 510(k) Image Suite

Comparison of Technological Characteristics:

A substantial equivalence table (Table 1) provides the similarities and differences between the modified and predicate devices.

Table 1: Substantial Equivalence Table

Predicate: Image Suite (with Tablet Viewer Software)Subject Device: Image SuiteNew Feature to Image Suite
510(k) NumberK140271K251186N/A
Indications for UseCARESTREAM Image Suite is an image management system whose intended use is to receive, process, review, display, print and archive images and data from all imaging modalities.Tablet Viewer Software for Image Suite is used for patient management by clinicians in order to access and display patient data, medical reports, and medical images for diagnosis from different modalities including CR, DR, CT, MR, and US.Tablet Viewer Software for Image Suite provides wireless and portable access to medical images for remote reading or referral purposes from web browsers including usage with validated mobile devices. This device is not intended to replace the full Mini- PACS and should beImage Suite is a stand-alone radiographic imaging software designed to perform patient registration, review, report, archive, and print patient images. The software interacts with Carestream CR or DR to receive radiographic images for review, and has the ability to receive, archive, and review DICOM images from compatible third-party modalities (for example, US, MR, CT). The software can also update Carestream CR or DR device firmware and monitor device calibration. The software serves as a web server to support web clients for image viewing. Image Suite is intended for use by radiologists and trained healthcare professionals. The software presents images to medical professionals in a convenient digital medium so they can make diagnostic and/or therapeutic decisions.N/A

Page 3 of 6

Page 8

Traditional 510(k) Image Suite
Page 4 of 6

Page 9

Traditional 510(k) Image Suite
Page 5 of 6

used only when there is no access to the full Mini-PACS Web Viewer.This excludes mammography applications in the United States.This excludes mammography applications and Tablet Viewer applications in the United States
Image Processing SoftwareEclipseEclipse IINo
DisplayMulti- Monitor Display or iPad 2Multi- Monitor DisplayNo
Operating SystemWindows or iOSSameNo
BrowserInternet Explorer or SafariSameNo
DICOM supportYesSameNo
ModalitiesCarestream Health: CR and DR modalitiesThird party: CR, DR, CT, MR, and US modalitiesSameNo
Measurement CapabilitiesYesSameNo
Display MeasuresYesSameNo
2D DisplayYesSameNo
Cine ModeYesSameNo
Zoom ImagesYesSameNo
Rotate/Flip ImagesYesSameNo
Window Level AdjustmentYesSameNo
Layout adjustmentYesSameNo
Lossy vs. LosslessYesSameNo
Tube & PICC Line/ High Detail Visualization (Companion View)NoNew Image Suite Feature Cleared in K120062Yes
Pneumothorax (Companion View)NoNew Image Suite Feature Cleared in K120062Yes
Bone Suppression (Companion View)NoNew Image Suite Feature Cleared in K133442Yes
X-FactorNoNew Image Suite FeatureYes
Multiple LooksOne look availablePreviously available on Image Suite there are now Four looks availableYes
Smart GridNoNew Image Suite Feature Cleared in K163157Yes

Page 5 of 6

Page 10

Traditional 510(k) Image Suite
Page 6 of 6

Discussion of Nonclinical Testing

Performance testing was conducted to verify the design output met the design input requirements and to validate the device conformed to the defined user needs and intended uses. Software testing was conducted under simulated use conditions. Predefined acceptance criteria were met and demonstrated that the device is as safe and as effective as the predicate devices.

Conclusion of Substantial Equivalence:

The comparison of the predicate and subject devices demonstrates that they are substantially equivalent in terms of their intended use, technological characteristics, and performance outcomes. Both devices fulfill the same primary functions and achieve comparable results, ensuring their safety and efficacy within their defined applications. Any differences in design or components are non-significant and do not affect the overall operation or intended purpose of the devices. Therefore, these distinctions do not raise new questions regarding their safety or effectiveness, supporting the assertion of substantial equivalence.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).