Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K163167
    Date Cleared
    2017-08-09

    (268 days)

    Product Code
    Regulation Number
    888.3560
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
    B. Failed osteotomy or unicompartmental replacements.
    C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
    D. The non-porous (uncoated and coated with CoCr beads without Titanium) components may only be used with cement.
    E. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
    F. Stemmed baseplates of the CKS Plus Knee System are intended for cemented use only

    Device Description

    The CKS Plus extension to the Consensus Knee System (CKS) is a primary fixed bearing total knee system offering flexibility to restore knee function using either cruciate retaining (CR) or posterior stabilizing (PS) components with the option of tibial cancellous screw, tibial intramedullary (IM) stem fixation, and tibial augments

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "CKS Plus Knee System." It details the device, its indications for use, and claims substantial equivalence to predicate devices. Crucially, as a 510(k) summary for a medical device (a knee prosthesis), the focus is on demonstrating substantial equivalence through comparison to existing devices and bench testing, not on clinical performance studies involving AI or complex statistical analyses of diagnostic accuracy. Therefore, many of the requested elements for an AI/diagnostic device study (like sample sizes for test/training sets, expert qualifications, MRMC studies, standalone performance, and detailed ground truth establishment) are not applicable to this type of document.

    The document primarily discusses non-clinical performance data (bench testing) to verify the safety and effectiveness of the CKS Plus Knee System's components, focusing on mechanical properties and stability rather than diagnostic accuracy or human interpretation.

    Here's an attempt to address the request based only on the provided text, highlighting what is (and isn't) present:

    A. Table of Acceptance Criteria and Reported Device Performance

    The document does not present acceptance criteria in a formal table with reported performance side-by-side, as would be typical for a diagnostic device's clinical study. Instead, it lists the types of bench tests performed and implies that the device "meets" the requirements by stating it passed. The criteria are implicit in the ASTM standards and other specified tests.

    Acceptance Criteria (Implicit from Test Standards/Purpose)Reported Device Performance (Implied from "Testing Performed")
    Adequate tray fatigue strength (per ASTM F1800-12)"the tray would not fail under fatigue when one compartment collapses"
    Tibiofemoral joint stability (per ASTM F1223-14)Tested for PS-C insert per ASTM F1223-14
    Adequate tibial insert locking mechanism strengthTested per FDA's Class II Special Controls Guidance to ensure adequate connection strength and ease of insertion
    No failure in PS Post Fatigue Testing"PS Post Fatigue Testing, Report" (implies successful completion)
    No failure in Baseplate Fatigue Testing"Baseplate Fatigue Testing, Report" (implies successful completion)
    No failure in Tibial Insert Dislocation Testing"Tibial Insert Dislocation Testing, Report" (implies successful completion)
    No failure in VitalitE PS Post Fatigue Testing"VitalitE PS Post Fatigue Testing, Report" (implies successful completion)
    No failure in Tray Fatigue (FEA Worst Case)"Tray Fatigue: FEA Worst Case Scenario, Report" (implies successful completion)
    No failure in Stem Fatigue (FEA Worst Case)"Stem Fatigue: FEA Worst Case Scenario, Report" (implies successful completion)
    Pyrogenicity absence"Pyrogenicity testing has been performed" (implies successful completion)

    B. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in terms of number of devices or components tested. The document mentions "bench testing" was carried out, but does not provide specific sample quantities for each test.
    • Data Provenance: The tests are "bench testing," meaning they are performed in a laboratory setting on components or full devices, not on human subjects. Thus, there is no country of origin for human data, nor is it retrospective or prospective in that context.

    C. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Experts: Not applicable. This is not a study requiring expert consensus for ground truth on diagnostic performance. The "ground truth" for mechanical testing is adherence to validated ASTM standards and engineering principles.

    D. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This is not an image-based or diagnostic study requiring adjudication of expert interpretations.

    E. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size.

    • MRMC Study: No, an MRMC study was not done. This device is a knee prosthesis, not an AI or diagnostic tool that relies on human interpretation of cases. Therefore, there is no "effect size of how much human readers improve with AI vs without AI assistance" to report.

    F. If a Standalone (algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. This is a physical medical device (knee implant), not an algorithm or AI system.

    G. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for the performance claims for this device is based on engineering standards, ASTM specifications, and mechanical testing results. For example, ASTM F1800-12 defines criteria for tray fatigue, and success is determined by meeting those predefined engineering benchmarks.

    H. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This document describes the testing of a physical medical device, not the training of an AI algorithm.

    I. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set Establishment: Not applicable, as there is no training set mentioned or implied for an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160515
    Device Name
    PS2 Knee System
    Date Cleared
    2016-12-19

    (299 days)

    Product Code
    Regulation Number
    888.3560
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PS2 KNEE SYSTEM is designed as a system and is not intended for substitution of components from other systems.

    The indications for use are:

    • A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
    • B. Failed osteotomy or unicompartmental replacements.
    • C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
    • D. The PS2 KNEE SYSTEM is intended for cemented use.
    Device Description

    The PS2 Knee System is a primary fixed bearing total knee system offering flexibility to restore knee function using either cruciate retaining (CR) or posterior stabilizing (PS) components with the option of tibial cancellous screw and tibial stem fixation. The uncoated femoral components (CR and PS) and uncoated tibial baseplates (pegged, pegless, and holed) are made from cast CoCr alloy (ASTM F75), and are intended for cemented use only. The tibial inserts (CR and PS) and all-poly patellar components (oval and round) are made from UHMWPE (ASTM F648) or VitalitE (ASTM F2695). The PS2 tibial baseplate employs a modular keel, which is compatible with the previously cleared Consensus Revision Knee System (CRKS) stem and taper plug made from Titanium alloy (ASTM F1472). The PS2 holed baseplate is compatible with the previously cleared Consensus Knee System (CKS) 6.5mm cancellous bone screw made from Titanium alloy (ASTM F136). The PS2 baseplate is supplied with four preassembled cement dams made from UHMWPE and supplied with a distal plug made from UHMWPE intended for assembly in the operating room when stems are not desired.

    AI/ML Overview

    The provided document describes the PS2 Knee System, a medical device, and its substantial equivalence to legally marketed predicate devices, rather than detailing the acceptance criteria and the study proving it meets those criteria in the context of an AI/ML device.

    Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, data provenance, expert qualifications, ground truth, or MRMC comparative effectiveness studies because this document is for a traditional medical device (knee prosthesis), not an AI/ML device.

    The document discusses:

    • Indications for Use: What the device is intended for.
    • Substantial Equivalence: How the device is similar to previously cleared devices.
    • Non-Clinical Performance Data: Bench testing performed on the device's components to ensure safety and effectiveness, following various ASTM standards and FDA guidance for traditional medical devices. This includes topics like fatigue testing, joint stability, contact characteristics, locking mechanisms, and wear testing.

    This information is typical for a 510(k) submission for a non-AI/ML device where the primary pathway to market is demonstrating substantial equivalence to a predicate device based on material properties, design, and mechanical performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133919
    Date Cleared
    2014-07-14

    (203 days)

    Product Code
    Regulation Number
    888.3560
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used with the components of the Consensus Knee System:

    The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems.

    • A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
    • B. Failed osteotomy or unicompartmental replacements.
    • C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
    • D. The non-porous (uncoated and coated with CoCr beads without Titanium) components may only be used with cement.
    • E. The porous coated (CoCr beads with Titanium) components may be used with or without cement.

    When used as a component of the Consensus Revision Knee System:

    The Consensus® Revision Knee System is designed as a system and is only intended to be used with compatible components of the Consensus® Knee System. The revision knee femoral and tibial components are intended for cemented use only.

    The indications for use are:

    • A. Primary intervention of rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle.
    • B. Post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion. dvsfunction, or prior patellectomy)
    • C. Failed osteotomy or unicompartmental replacements.
    • D. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
    • E. The salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery
    • F. Moderate valgus, varus, or flexion deformities
    Device Description

    The VitalitE Tibial Insert is a vitamin E version of the UHMWPE tibial insert that has been a component of both the Consensus Knee System (CKS) and the Consensus Revision Knee System (RKS). The new insert still serves as the gliding surface between the metallic femoral component and the metallic tibial base plate of these knee systems. The new VitalitE insert will also be available in a cruciate retaining (CR) or ultra-congruent/ PCL substituting designs in sizes 0 through 6 and in thicknesses from 10 to 22 mm. The proprietary locking mechanism design on the underside of the insert will remain unchanged. The only difference will be the vitamin E additive in the UHMWPE which is a-tocopherol. This is the exact same polymer that is used in the Consensus VitalitE Acetabular insert but with a slightly lower radiation dose. Consensus will also offer our current CKS all-poly patellae in the new vitamin E UHMWPE material. This includes both the round and oval configurations with either a 7.5 mm or 10 mm thickness.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the Consensus Orthopedics VitalitE Tibial Insert & Patellar Components. This is a medical device, and the evaluation for such devices often focuses on substantial equivalence to existing legally marketed predicate devices, rather than a direct clinical performance study that proves specific acceptance criteria for a new AI/software device.

    Therefore, the requested information elements (especially those related to AI/software performance metrics like sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance, and specific ground truth establishment methods) are not applicable in the context of this 510(k) submission for a physical medical implant.

    However, I can extract the relevant information regarding the acceptance criteria and how the device meets them based on the provided text, reinterpreting some of your categories to fit the context of a physical implant's regulatory review.

    Here's the breakdown:

    Acceptance Criteria and Device Performance for Consensus Orthopedics VitalitE Tibial Insert & Patellar Components

    The core "acceptance criteria" for this device, within the scope of a 510(k) submission, revolve around demonstrating substantial equivalence to existing predicate devices. This means showing that the new device is as safe and effective as a legally marketed device that does not require premarket approval (PMA).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (based on Substantial Equivalence)Reported Device Performance/Evidence Provided
    Design Equivalence: New device has the same design/form as predicate devices.The VitalitE tibial insert "is the exact same design" as cleared for use with the Consensus Knee System (K932837, K953443 & K110950) and Consensus Revision Knee System (K100542). It is compatible with all metallic femoral and tibial components from those systems. The proprietary locking mechanism design on the underside remains unchanged.
    Material Equivalence: New device uses materials that are either identical or demonstrably equivalent to predicate devices, especially regarding the key differentiating feature (Vitamin E).The UHMWPE (GUR 1020) with α-tocopherol (Vitamin E) is the "same polymer" as cleared in K130652 for the Consensus VitalitE Acetabular Insert (with a slightly lower radiation dose). The use of Vitamin E infused UHMWPE tibial inserts for knee systems has been cleared for use in DJO/Encore systems via K091956 and K103223. The UHMWPE used in the current Consensus tibial inserts is GUR 1050 (implying the new GUR 1020 with Vitamin E is being compared for equivalence).
    Performance Equivalence: The new device performs equivalently to predicate devices in critical functional aspects.Non-Clinical Performance Data (Bench Testing): Included wear testing of the worst-case tibial insert. Included anterior disassembly testing of the locking mechanism (push-out testing) on a range of various sizes and types of inserts. The report states: "All of the observed results indicate that the VitalitE Tibial Inserts and Patellas are substantially equivalent to devices currently marketed." Biocompatibility Testing: Performed. Material Characterization: Performed.
    Processing and Shelf-Life Equivalence: Manufacturing, packaging, and sterilization processes maintain equivalence.The VitalitE inserts are "cleaned, packaged, and sterilized using the same validated processes as the current inserts." They will also have a "5 year shelf-life" (implicitly equivalent to or validated as acceptable for existing products).
    Intended Use Equivalence: Indications for Use are consistent with predicate devices.The "Indications for Use" statement (on page 5) lists specific primary intervention scenarios (rheumatoid arthritis, osteoarthritis, etc.), failed osteotomies, replacement of unsatisfactory components, and various deformities, which are standard for these types of knee systems and align with predicate devices.
    Safety and Effectiveness: Overall safety and effectiveness is comparable to legally marketed predicates.The FDA's final letter of substantial equivalence determination indicates that "the device is substantially equivalent... to legally marketed predicate devices." And the company states: "Therefore, the devices are as safe, as effective, and perform at least as safely and effectively as legally marketed predicates."

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not specified in terms of specific number of units for each bench test. For physical implant testing, "worst-case" scenarios are typically selected rather than a large statistical sample of every size/type. The document mentions "a range of various sizes and types of inserts" for the disassembly testing.
    • Data Provenance: The data appears to be from internal bench testing and analysis conducted by Consensus Orthopedics, Inc. The document doesn't specify country of origin for the data, but the company is based in El Dorado Hills, CA, USA, implying the testing likely occurred in the US or at a certified lab. The data is retrospective in the sense that it was generated for this specific submission to demonstrate equivalence to already marketed devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k) for a physical implant. "Ground truth" as a concept with expert consensus is typically used for AI/software devices where human interpretative judgment is being compared to an algorithm's output. For physical implants, "ground truth" is established through engineering and material science standards, validated testing methods, and regulatory requirements (e.g., ISO standards, ASTM standards, FDA guidance).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for reconciling discrepancies in human expert interpretations, especially in clinical image analysis for AI. For engineering bench testing, results are typically quantitative and compared against predefined pass/fail criteria or against control/predicate device performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. MRMC studies are specific to evaluating inter-reader variability and the impact of AI assistance on human performance, primarily in imaging diagnostics. This submission is for a physical knee implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical knee implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For physical implants demonstrating substantial equivalence, the "ground truth" is typically a combination of:
      • Validated engineering standards: e.g., ASTM standards for mechanical properties and wear.
      • Biocompatibility standards: demonstrating the material is safe for implantation.
      • Performance of legally marketed predicate devices: The "known good" performance of existing devices serves as the benchmark.
      • Material characterization: Chemical composition, physical properties, and manufacturing integrity are verified against specifications.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" in the context of a 510(k) for a physical medical device. This concept is relevant for machine learning algorithms.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130652
    Date Cleared
    2013-12-17

    (280 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VitalitE Acetabular Insert for the Consensus CS2™ Acetabular Cup System is indicated for use with the CONSENSUS® Hip System, TaperSet Hip System, CS2 Hip System, or UNISYN™ Hip System for the following indications:
    A) Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
    B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
    C) Proximal femoral fractures.
    D) Avascular necrosis of the femoral head.
    E) Non-union of proximal femoral neck fractures.
    F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities
    Acetabular components are indicated for cemented and cementless use.
    Consensus femoral stems are indicated for cemented and cementless use.
    UniSyn, TaperSet, and CS2 femoral stems are indicated for cementless use only.
    HA coated implants are indicated for cementless use only.

    Device Description

    The Consensus CS2™ Acetabular Cup System consists of a titanium alloy shell (ASTM F620 or F136) and a polyethylene liner (ASTM F648). The system is compatible with the femoral components of all cleared hip systems manufactured by Consensus Orthopedics, Inc.: Consensus Hip System (CHS), TaperSet Hip System (THS), CS2 Hip System (CS2HS), and UniSyn Hip System (UniSyn). The CS2 Acetabular Cup System can be used with the CoCr, zirconia, or Biolox delta femoral heads currently marketed by Consensus and with 22mm, 32mm, and 36mm inner diameters. The liners are available in either neutral or hooded versions and with either standard or lateral offsets. The shell is designed for uncemented press-fit or cemented use to the prepared acetabulum, and is designed to mate with the insert via secure insert/shell locking mechanism. The shell comes with or without holes for additional screw fixation. The new acetabular liners will be made from UHMPWE that contains vitamin E (a-tocopherol).

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the VitalitE Acetabular Insert for the CS2™ Acetabular Cup System. The submission primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices rather than proving performance against specific acceptance criteria through a standalone clinical study.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics for the VitalitE Acetabular Insert to be evaluated against in a clinical study. Instead, the submission relies on the concept of "substantial equivalence" to predicate devices. The performance data is primarily "Non-Clinical Performance Data."

    Acceptance Criteria CategorySpecific Acceptance Criteria (Not Explicitly Stated as Numerical Criteria)Reported Device Performance (Summary from Submission)
    Material Equivalence"The vitamin E poly used in the new Consensus CS2 insert is identical to the poly cleared for the StelKast EXp acetabular liner in K094035 and K122773 (with the exception of the radiation dosing level)."The new VitalitE Acetabular Insert uses a vitamin E cross-linked UHMWPE material that is "identical" to that used in cleared predicate devices (K094035, K122773), with a noted exception in radiation dosing.
    Design Equivalence"The new VitalitE Acetabular Insert for the CS2 Acetabular Cup System is identical in design to the cross-linked UHMWPE acetabular insert previously cleared for Consensus in the following 510(k)s: K021466, K070061, and K100933."The design of the new insert is "identical" to previously cleared cross-linked UHMWPE acetabular inserts (K021466, K070061, K100933).
    Bench Testing PerformanceImplied criteria for wear, push-out, and torque-out performance at least as safe and effective as predicates. (Specific numerical criteria not given)"Non-clinical testing and analysis were provided, including bench testing, material characterization, and biocompatibility testing. Bench testing of the new stem design included wear testing of the worst case acetabular insert, in addition to disassembly testing which included push-out, and torque-out tests." "All of the observed results indicate that the VitalitE Acetabular Insert is substantially equivalent to devices currently marketed."
    BiocompatibilityImplied criteria for biocompatibility at least as safe and effective as predicates. (Specific criteria not given)Biocompatibility testing was performed and included in the overall assessment of substantial equivalence.
    Overall Safety & Effectiveness"as safe, as effective, and performs at least as safely and effectively as legally marketed predicates."Based on material, characterization data, geometry, and mechanical testing, the device is deemed "substantially equivalent" and thus "as safe, as effective, and performs at least as safely and effectively as legally marketed predicates."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not describe a clinical study with a "test set" of patient data. The "testing" mentioned is entirely non-clinical (bench testing, material characterization, and biocompatibility). Therefore, there is no sample size for a test set or data provenance (country of origin, retrospective/prospective) related to patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. Since no clinical study involving patient data was conducted to establish ground truth, no experts were used for this purpose.

    4. Adjudication Method for the Test Set:

    Not applicable. No clinical test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is not a diagnostic device requiring interpretation by human readers. It is a medical implant, and the submission focuses on its physical and material properties.

    6. Standalone (Algorithm Only) Performance Study:

    Not applicable. This is not a software algorithm device. It is a physical medical device (an acetabular insert).

    7. Type of Ground Truth Used:

    Ground truth in the typical sense of clinical outcomes or expert consensus for patient data is not used in this submission. The "ground truth" here is established through:

    • Material properties: Chemical composition and physical characteristics of the UHMWPE.
    • Design specifications: Geometric and structural adherence to established designs.
    • Bench test results: Mechanical performance (wear, push-out, torque-out) measured against engineering standards or comparison to predicate device performance under similar test conditions.

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. No training set was used.

    Summary of the Study:

    The "study" described in K130652 is a non-clinical substantial equivalence submission. It relies on:

    • Comparison of material properties: The vitamin E polyethylene is stated to be "identical" to material in already cleared devices.
    • Comparison of design: The new insert's design is "identical" to previously cleared inserts.
    • Bench testing: Wear testing of the "worst-case acetabular insert" and disassembly testing (push-out, torque-out) were performed. The results are not detailed but are stated to "indicate that the VitalitE Acetabular Insert is substantially equivalent to devices currently marketed."
    • Biocompatibility testing: Mentioned as part of the non-clinical data.

    The "study" concludes that based on these non-clinical evaluations, the device is "as safe, as effective, and performs at least as safely and effectively as legally marketed predicates." This approach is typical for 510(k) clearances when a new device is fundamentally similar to existing, legally marketed devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122512
    Device Name
    CS2 HIP SYSTEM
    Date Cleared
    2012-11-05

    (80 days)

    Product Code
    Regulation Number
    888.3358
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CS2TM Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.

    The indications for use are:

    • A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
    • B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
    • C. Proximal femoral fractures.
    • D. Avascular necrosis of the femoral head.
    • E. Non-union of proximal femoral neck fractures.
    • F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural. abnormalities

    This CS2TM hip stem is indicated for cementless use.

    Indications for Use of the CONSENSUS® BIPOLAR or UNIPOLAR:

    • A. Primary replacement of the femoral head and neck with very little if any acetabular degradation noted.
    • B. Rheumatoid, osteo, and post traumatic arthritis.
    • C. Proximal femoral fractures.
    • D. Avascular necrosis of the femoral head.
    • E. Non-unions of proximal femoral neck fractures.
    • F. Revision of failed total hip arthroplasty.
    • G. Treatment of malunion or nonunion acetabular fractures.

    The CONSENSUS® BIPOLAR or UNIPOLAR are intended for cementless use.

    Device Description

    The Consensus CS2TM Hip System (CS2HS) is a monolithic; titanium alloy fit & fill hip stem design with a proximal, plasma sprayed, porous CPTi coating. The stem has a cylindrical geometry with a bullet tip and is available in sizes designated as 10 mm to 22 mm. The stems feature a neck shaft angle of 128° and a 12/14. Morse taper trumilon. The stems are available with, or without a collar. The CS2 Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System. The stem is. compatible with previously cleared CoCr heads, zirconia heads, Biolox delta heads, unipolar heads, bipolar heads, UHMWPE inserts and acetabular cups.

    AI/ML Overview

    This 510(k) summary describes a submission for a new hip stem design, the CS2™ Hip System, and primarily focuses on demonstrating its substantial equivalence to previously cleared devices through non-clinical performance data. As such, it does not detail acceptance criteria and a study proving those criteria are met in the same way a clinical trial or a performance study for an AI/ML device would.

    Here's an analysis based on the provided text, addressing the points where information is available and noting where it is not applicable or not provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    For medical devices like hip stems, "acceptance criteria" are typically related to mechanical properties, material integrity, and biological compatibility, which are demonstrated through bench testing and characterization, not performance metrics like accuracy or AUC. The "reported device performance" refers to the results of these non-clinical tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Strength/Durability (e.g., fatigue resistance)Bench testing of the new stem design included: - Distal fatigue testing - Proximal fatigue testing of the worst-case stem. These tests were conducted "consistent with the 'Guidance for Industry and FDA Staff Non-clinical Information for Femoral Stem Prostheses'." (The specific quantitative results or pass/fail thresholds are not detailed in this summary.)
    Porous Coating Characteristics (e.g., porosity)CPT plasma sprayed coating characterization "meets the definition of porosity per 21 CFR 888.3358." (Specific quantitative porosity values are not detailed.)
    Material CharacterizationBased on "material, characterization data, geometry and mechanical testing" (Specific characterization data is not detailed.)
    Substantial Equivalence to Predicate Devices"All of the observed results indicate that the CS2 Hip System is substantially equivalent to devices currently marketed. Therefore, the device is as effective, and performs at least as safely and effectively as legally marketed predicates."

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of this 510(k) summary, as it describes non-clinical bench testing. The "test set" would be the physical samples of the hip stem components subjected to mechanical and material characterization tests. The number of samples for each specific test (e.g., how many stems for fatigue testing) is not specified.
    • Data Provenance: The data comes from "Non-clinical Performance Data" and "bench testing." This implies internal laboratory testing conducted by Consensus Orthopedics, Inc. (the sponsor). There is no mention of country of origin of the data as it's not patient data, nor is it retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This is not applicable for this type of submission. Ground truth, in the sense of expert consensus on diagnostic images or clinical outcomes, is not relevant to non-clinical bench testing of a hip implant's mechanical properties. The "ground truth" here is established by engineering standards and regulatory requirements for medical device performance.

    4. Adjudication Method for the Test Set

    • Not applicable as there is no "adjudication" of expert opinions for a test set. Non-clinical tests follow established protocols, and results are measured objectively against predefined engineering specifications or regulatory guidance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC study was not done. This type of study is relevant for diagnostic devices that involve interpretation by human readers (e.g., radiologists, pathologists) and often for AI/ML-assisted diagnostic tools. This 510(k) pertains to a hip implant's physical and mechanical properties.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone performance study as described is not applicable. This device is a physical implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • For the non-clinical testing, the "ground truth" is based on:
      • Engineering Standards and Specifications: The mechanical tests (fatigue testing) were conducted "consistent with the 'Guidance for Industry and FDA Staff Non-clinical Information for Femoral Stem Prostheses'." This guidance provides the framework and methodologies for evaluating the safety and effectiveness of such devices.
      • Regulatory Definitions: The coating characterization "meets the definition of porosity per 21 CFR 888.3358."
      • Essentially, the "ground truth" is adherence to established industry-specific test methods and regulatory requirements for proving the safety and effectiveness of a hip prosthesis.

    8. The Sample Size for the Training Set

    • Not applicable. There is no training set for a physical hip implant. Training sets are typically associated with machine learning algorithms.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable for the reason stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120595
    Date Cleared
    2012-10-12

    (227 days)

    Product Code
    Regulation Number
    888.3360
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. Revision of failed femoral head replacement, hip arthroplasty or other hip procedures.
    A. Proximal femoral fractures.
    B. Avascular necrosis of the femoral head.
    C. Non-union of proximal femoral neck fractures.
    D. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, pseudarthrosis conversion, and structural abnormalities.
    UNISYN stems used with roughened and plasma coated bodies are intended for cemented or uncemented use. UNISYN stems used with plasma/HA or HA coated bodies are intended for uncemented use only.

    Device Description

    The UniSyn Hip System was originally cleared as a modular hip system for Hayes Medical under the name "Triton", modifications to the system were cleared in a subsequent submission by Hayes under the UniSyn name. Hayes Medical changed its name to Consensus Orthopedics in 2008. This system consists of three primary components: the neck, the body, and the stem. This submission is to addition of a set of modified ("Plus") stems.
    These minor modifications in body thickness are designed to allow for better contact between the implant and the inner surface of the medullary canal during revision surgery. The sizes of stems offered with the modified design are within the ranges previously cleared for the UniSyn system. The Plus stems are offered in neutral or +1 mm styles, short or long, in 10 to 20 mm diameters, and lengths from 110 to 210 mm. There is no change to the manufacturing, packaging, or sterilization processes. This change will not affect the indications for use nor any of the labeling.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the UNISYN Plus stem, which is part of the UNISYN Hip System. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices, rather than establishing new safety and effectiveness through a comprehensive study with acceptance criteria and performance metrics in the way a novel device might.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    For this type of 510(k) submission, the "acceptance criteria" and "reported device performance" are primarily focused on demonstrating that the modified device (UNISYN Plus stem) maintains the same fundamental characteristics and performance as its predicate devices. The document does not present a table with specific quantitative acceptance criteria (e.g., accuracy, sensitivity, specificity) and corresponding performance results. Instead, it relies on the concept of substantial equivalence.

    The key "criteria" for this submission are that the new "Plus" stems for the UniSyn hip systems employ the:

    • Same materials
    • Basic design features
    • Manufacturing processes
    • Packaging
    • Indications as the predicate stems for the UHS.

    The "reported device performance" is implicitly that these criteria are met, leading to the conclusion of substantial equivalence.

    Specific Performance Data Mentioned:

    • Distal fatigue performance: "Distal fatigue performance of the UniSyn Plus grit blasted stem was evaluated per ISO 7206-4 on a worst case example."
      • Acceptance Criteria (Implied): The fatigue performance must be equivalent to, or better than, the predicate device and meet the standards set by ISO 7206-4.
      • Reported Performance: Not explicitly stated with a numerical result, but the conclusion is that "No other nonclinical bench testing was deemed necessary over that used in support of the predicates based on identical design and materials," implying satisfactory performance.

    2. Sample Size for Test Set and Data Provenance:

    • Sample Size for Test Set: The document explicitly states "a worst case example" was evaluated for distal fatigue performance. This means the sample size for this specific test was very small, likely N=1 or a limited number of test articles representing the worst-case scenario. It is a bench test, not a clinical test involving human subjects.
    • Data Provenance: The data comes from non-clinical bench testing conducted by the manufacturer (Consensus Orthopedics, Inc.). It is not from human subjects, so concepts like "country of origin" or "retrospective/prospective" do not apply.

    3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications:

    Not applicable. The "ground truth" for a device like a hip stem is typically established through engineering specifications, material science standards, and established biomechanical performance criteria (e.g., ISO standards). There are no "experts" in the sense of clinicians establishing a diagnostic ground truth for a test set.

    4. Adjudication Method for the Test Set:

    Not applicable. This is a non-clinical bench test, not a clinical study involving human assessment or adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. This type of study is typically done for diagnostic imaging devices or AI-assisted diagnostic tools. The UNISYN Plus stem is a physical orthopedic implant, making an MRMC study irrelevant. There is no human reader improvement with AI assistance discussed as there is no AI component.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable. There is no algorithm or AI component in the UNISYN Plus stem. Performance assessment is based on physical material and design testing.

    7. Type of Ground Truth Used:

    The ground truth used for evaluating the UNISYN Plus stem's performance relies on engineering specifications, material properties, and adherence to recognized international standards such as ISO 7206-4 for fatigue testing. It is not expert consensus, pathology, or outcomes data in the clinical sense, but rather a demonstration of compliance with established biomechanical and material science criteria.

    8. Sample Size for the Training Set:

    Not applicable. This device does not involve a "training set" in the context of machine learning or algorithms. Its development and testing are based on traditional engineering design and physical performance verification.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set. The "ground truth" for the overall design and performance of the hip system is established through a combination of biomechanical design principles, material science knowledge, and adherence to relevant ISO standards, accumulated over years of orthopedic implant development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121935
    Date Cleared
    2012-09-10

    (70 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TaperSet™ Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.
    The indications for use are:

    • A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
    • B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
    • C. Proximal femoral fractures.
    • D. Avascular necrosis of the femoral head.
    • E. Non-union of proximal femoral neck fractures.
    • F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
      The TaperSet™ hip stem is indicated for cementless use.
    Device Description

    The TaperSet Hip System (THS) is a monolithic, titanium alloy tapered hip stem design with a proximal, plasma sprayed, porous CP Ti coating. The stem has a dual wedge geometry and is available in both standard and 7mm lateral offsets in sizes currently designated as 7.5mm to 24mm. The stems feature a neck shaft angle of 135° and a 12/14 Morse taper trunnion. The TaperSet Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System. The stem is compatible with previously cleared CoCr heads, zirconia or Biolox delta ceramic heads, unipolar heads, bipolar heads. UHMWPE inserts and acetabular cups. The modification addressed here is the addition of two stems in sizes 5mm and 6mm.
    The new stems are manufactured from forged titanium alloy (Ti-6AI-4V ELI, ASTM F620) and the proximal portion of the femoral stem component is plasma sprayed with commercially pure titanium (C. P. Ti. ASTM F1580). The new femoral stems are available in both standard and 7mm lateralized options and have the identical neck and taper design, and porous coating as the predicate TaperSet stems cleared under K102399 in design and indications.

    AI/ML Overview

    The provided FDA 510(k) summary (K121935) describes the TaperSet Hip System Small Stems, which are orthopedic implants, not an AI/ML powered device. Therefore, many of the requested categories related to AI/ML device performance and study design are not applicable to this submission.

    However, I can extract the relevant information regarding the acceptance criteria and the non-clinical study that demonstrates the device's performance based on mechanical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    The stems were to achieve 5 million cycles of loading at a maximum load of 517 lbs. (2300 N) without failure in accordance with ASTM F2068-2009 and ISO 7206-4:2010(E). This applies to the worst-case configuration (the 5mm lateral femoral stem).The submission states: "Based upon engineering analysis and testing of the design modification the new 5 & 6 mm femoral stems for the TaperSet Hip System by Consensus are substantially equivalent to devices currently marketed. Therefore, the device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicates." While not explicitly stating "passed," the determination of substantial equivalence, which is based on this testing, implies that the acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document refers to testing of a "worst-case example," specifically "the 5mm lateral femoral stem." This implies that at least one of these stems was tested. The document also states "The stems were to achieve...", suggesting that multiple stems may have been tested to confirm consistency, but the exact number isn't specified beyond "the 5mm lateral femoral stem was determined to be the worst-case example per analysis report (R11-007)." The study is a non-clinical, laboratory-based mechanical test.
    • Data Provenance: This is not applicable in the context of clinical data for a mechanical test. The testing was performed on newly manufactured devices to evaluate their mechanical properties.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This is not applicable as the study involves mechanical testing of an orthopedic implant, not the evaluation of a device by human experts or the establishment of a ground truth for diagnostic purposes. The "ground truth" here is the physical performance of the device under specific loading conditions, measured by engineering standards.

    4. Adjudication Method for the Test Set

    This is not applicable for a non-clinical mechanical test. The test results are determined by instrumentation and adherence to engineering standards, not through human adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. This is not an AI/ML device or a device that requires human interpretation of outputs.

    6. If a Standalone Performance (i.e. algorithm only without human-in-the-loop performance) was Done

    No. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The "ground truth" for this study is defined by established biomechanical and engineering standards for hip prostheses:

    • ASTM F2068-2009 (Standard Specification for Femoral Prostheses – Mechanical Performance)
    • ISO 7206-4:2010(E) (Implants for surgery — Partial and total hip joint prostheses — Part 4: Determination of endurance properties of stemmed femoral components)

    The acceptance criteria (5 million cycles at 517 lbs/2300 N without failure) are directly derived from these standards.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121263
    Date Cleared
    2012-05-22

    (26 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TaperSet™ Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.
    The indications for use are:

    • A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
    • B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
    • C. Proximal femoral fractures.
    • D. Avascular necrosis of the femoral head.
    • E. Non-union of proximal femoral neck fractures.
    • F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
      The TaperSet™ hip stem is indicated for cementless use.
    Device Description

    The TaperSet Hip System (THS) is a monolithic, titanium alloy tapered hip stem design with a proximal, plasma sprayed, porous CPTi coating. The stem has a dual wedge geometry and is available in both standard and 7mm lateral offsets in sizes designated as 7.5mm to 24mm. The stems feature a neck shaft angle of 135° and a 12/14 Morse taper trunnion. The TaperSet Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System. The stem is compatible with previously cleared CoCr heads, zirconia or Biolox delta ceramic heads, unipolar heads, bipolar heads, UHMWPE inserts and acetabular cups. The modification addressed here is the addition of a line of stems with a reduced distal profile (RDP) in sizes 10.5mm to 24mm.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the TaperSet™ Hip System RDP Stems, structured according to your request.

    It's important to note that the provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a performance study with detailed acceptance criteria and standalone algorithm performance as might be typical for AI/ML device submissions. Therefore, some of your requested information (e.g., sample sizes for test/training sets, expert qualifications, MRMC studies, standalone performance with specific metrics) is not applicable or not provided in this type of regulatory submission for a traditional medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Material EquivalenceThe new device must utilize materials equivalent to the predicate device."The new Reduced Distal Profile (RDP) stems for the TaperSet Hip System have the identical neck and taper design, porous coating, and sizing as the predicate TaperSet stems cleared under K 102399 in design and indications." (Implied: Material is titanium alloy with CPTi coating, same as predicate).
    Design EquivalenceThe new device's design features, where modified, must not introduce new safety or efficacy considerations that differ significantly from the predicate, or must be demonstrably safe and effective through engineering analysis."The new Reduced Distal Profile (RDP) stems for the TaperSet Hip System have the identical neck and taper design, porous coating, and sizing as the predicate TaperSet stems cleared under K 102399 in design and indications." "A reduced distal profile femoral hip stem design was previously cleared by FDA under K101086 for the Biomet Taperloc Complete Hip System."
    Indications for UseThe indications for use of the new device must be identical or substantially similar to the predicate device.The indications for use listed are identical to those of the predicate device (K102399).
    Functional EquivalencePerformance characteristics affected by the modification (e.g., mechanical properties, fit) must be demonstrated as equivalent or superior through appropriate testing or analysis. (Implied: No adverse impact on mechanical integrity or fit due to RDP)."No additional non-clinical testing was performed because there was no proximal design change from the predicate TaperSet and distally the new smallest size is still larger than the smallest predicate TaperSet stem which was previously tested." "Based upon engineering analysis of the design modification the new reduced distal profile stems... are substantially equivalent."
    Safety and EffectivenessThe new device must be as safe and effective as the legally marketed predicate device(s). This is the overarching criterion for 510(k) clearance, demonstrated by equivalence in materials, design, indications, and performance (or justification for lack of new testing)."Therefore, the device is as safe, as effective, and performs at least as safely and effectively as legally marketed predicates."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable. This submission relies on an "engineering analysis of the design modification" and comparison to predicate devices, rather than a specific "test set" of patient data as might be used for software or diagnostic devices.
    • Data Provenance: Not applicable. The "study" here is a regulatory submission demonstrating substantial equivalence based on prior clearances and engineering analysis.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No "ground truth" was established by experts for a test set in the context of this 510(k) submission. Regulatory experts at the FDA ("Division of Surgical, Orthopedic, and Restorative Devices") reviewed the submission for substantial equivalence.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. There was no test set requiring ground truth adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a traditional orthopedic implant, not an AI/ML device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a physical orthopedic implant. There is no algorithm or standalone performance to evaluate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable in the conventional sense of clinical ground truth. The "ground truth" for this submission is the established safety and effectiveness of the predicate devices and the regulatory framework that deems substantial equivalence as sufficient for clearance. The "truth" is established by demonstrating that the modified device does not raise new questions of safety or effectiveness compared to legally marketed devices.

    8. The sample size for the training set

    • Not applicable. There is no training set as it's a physical device, not a software or AI/ML product.

    9. How the ground truth for the training set was established

    • Not applicable for the reasons stated above.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The "study" in this context is the 510(k) premarket notification process itself, which relies on demonstrating substantial equivalence to existing, legally marketed predicate devices.

    The proof described is:

    • Comparison to Predicate: The new TaperSet™ Hip System RDP Stems were compared to the predicate TaperSet stems (K102399) and another previously cleared RDP stem design (Biomet Taperloc Complete Hip System, K101086).
    • Identical Features: The new RDP stems maintain "identical neck and taper design, porous coating, and sizing" with the predicate TaperSet stems.
    • Engineering Analysis: Consensus Orthopedics performed an "engineering analysis of the design modification."
    • Justification for No Additional Testing: No new non-clinical testing (like mechanical bench testing) was deemed necessary because:
      • There was "no proximal design change from the predicate TaperSet."
      • Distally, the "new smallest size is still larger than the smallest predicate TaperSet stem which was previously tested." This implies that the previous testing on the predicate covered potentially more challenging (smaller) scenarios.
    • Conclusion of Substantial Equivalence: Based on this engineering analysis and comparison, the manufacturer concluded and the FDA agreed that the device is "substantially equivalent" to legally marketed predicates, and therefore "as safe, as effective, and performs at least as safely and effectively as legally marketed predicates."

    This type of submission avoids extensive new clinical trials or performance testing by leveraging the safety and efficacy track record of existing, similar devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110950
    Date Cleared
    2011-06-27

    (84 days)

    Product Code
    Regulation Number
    888.3565
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems.

    • A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
    • B. Failed osteotomy or unicompartmental replacements.
    • C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
    • D. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
    Device Description

    The Consensus Total Knee System (CKS) is a primary fixed-bearing total knee system that has been on the market since the mid-1990's.

    The CKS has been designed to replicate the natural anatomy of the knee in order to restore knee function. It has been developed to preserve and utilize healthy ligamentous structures. For cases where the soft tissues are not functional, the PCL substituting tibial inserts or the posterior stabilized system are available for increased stability.

    The CKS incorporates femoral, tibial, and patellar components and all associated instrumentation needed for implantation. The CKS can be used for total knee replacement with posterior cruciate ligament (PCL) retaining or substituting.

    The femoral components are provided in left and right side versions and are designed to replicate natural kinematic motion between the femur, tibia and patella. The Consensus femoral component is designed to provide uniform contact zones in the coronal plane throughout the range of motion when the knee is properly aligned. The femoral component is also designed with a large distal radius to optimize contact areas and reduce contact stress. The trochlear groove in the femur is designed to allow the load from the patella to be evenly distributed on the femur with adequate lateral constraint.

    The CKS metallic components are available in non-porous and porous coated variants for cemented use and in a porous coated (CoCr beads with Titanium) version for uncemented use.

    AI/ML Overview

    This document describes the non-clinical performance data for the Consensus® Knee System Line Extensions. The data focuses on material properties and mechanical performance rather than AI or human reader studies.

    Here's the breakdown of the requested information based on the provided text, with an emphasis on addressing the relevant sections and noting where information is not applicable:

    1. A table of acceptance criteria and the reported device performance

    SpecificationAcceptance CriteriaVerification Results
    Porous Coating: CoCr beads with Ti coating
    Microstructure of the modified surfaceN/A (Detailed parameters are reported, no specific acceptance criteria)Bead to Bead Neck Diameter 0.33 mmPore Size 0.432 mmVolume % Porosity 37%Coating Thickness 0.889 mm
    Corrosion of the modified surfaceEqual to or improved corrosion resistance when compared with CoCr beads using ASTM F746 & G61Critical Potential Breakdown PotentialCoCr Beads 1290 mV 1200 mVTi Coated CoCr Beads 1315 mV 1200 mV
    Static tensile strengthThe static tensile strength will exceed 20 MPa.Static Tensile Strength of 58.32 MPa
    Static shear strengthThe static shear strength will exceed 20 MPa.Static Shear Strength of 58.32 MPa
    Shear fatigue strengthThe shear fatigue strength will exceed 10 million cycles.10 million cycles achieved with a strength of 13.78 MPa
    Rotating beam fatigue strengthThe rotating beam fatigue strength will exceed 10 million cycles.10 million cycles achieved with a strength of 206.7 MPa
    Excessive abrasionN/A (Results are reported, no specific acceptance criteria defined)200N load: Avg. mass loss 0.006 g; Avg. thickness loss 6%1500N load: Avg. Mass loss 0.179 g; Avg. thickness loss 23%
    RLP Femoral Components
    Articulating surface of the RLP componentsSimilar contact area and surface stress distributions.RLP had the same contact area as the original CKS. The PS RLP matches the articulating surface of the RLP.
    Size 0 Tibial Base Plate and Insert
    Size 0 tibial baseplate/insert assembly push-in/push-out forceSimilar push-in/push-out loads when compared with existing baseplate/insert combinations.Minimum push-out load was 428 lbs with failure mode being deformation of anterior snap recess in poly insert; Similar to other insert/baseplate combinations. Components can be easily inserted by hand.
    Thicker Tibial Insert
    Insert thickness per FDA Guidance Jan. 16, 2003Insert thickness must be >6mm.Insert thickness was greater than 6mm.
    Porous Coated Metal Backed Patella
    Articulating surface of the porous coated metal backed patellaIdentical articulating surface.Articulating surface of the porous coated metal backed patella is identical to existing patella.
    Mating geometry between the UHMWPE and metal back of the porous coated metal backed patellaIdentical mating geometry.The mating geometry of the porous coated metal backed patella is identical to the existing metal backed patella.

    The remaining information requested is largely not applicable (N/A) to this specific type of device and study. The provided document describes a 510(k) submission for line extensions of a knee prosthetic system, focusing on non-clinical performance data related to material properties, mechanical strength, and design conformity. It does not involve any AI, diagnostic imaging, or human reader performance studies.

    Here's an explanation for each point:

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify a "sample size" in the context of clinical trials or data sets for an AI model. For mechanical testing, the number of samples tested for each specification is not explicitly stated, although it's implied that multiple samples were tested to achieve "Avg. mass loss," etc.
    • Data Provenance: N/A. This is non-clinical mechanical and material testing, not data collected from human subjects or from a specific geographical origin. It's likely laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. This study does not involve expert evaluation for establishing ground truth as it's not a diagnostic or AI-based device. The "ground truth" here refers to the physical properties and performance metrics measured in a lab setting according to established standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Adjudication methods are relevant for studies involving human interpretation or clinical outcomes, which is not the case here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is not an AI-assisted diagnostic or interpretation device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This device does not involve any algorithm or AI requiring standalone performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device's performance is based on objective engineering and material science standards and measurements. This includes:
      • ASTM standards (e.g., F746 & G61 for corrosion).
      • Pre-defined mechanical thresholds (e.g., static tensile strength > 20 MPa, fatigue strength > 10 million cycles).
      • Physical design specifications (e.g., "Identical articulating surface," "Insert thickness must be >6mm").

    8. The sample size for the training set

    • N/A. There is no AI or machine learning model, so no training set is applicable.

    9. How the ground truth for the training set was established

    • N/A. As there is no training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110542
    Date Cleared
    2011-04-01

    (36 days)

    Product Code
    Regulation Number
    888.3358
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Consensus hip systems are designed for total or partial hip arthroplasty and are only intended to be used with compatible Consensus components per the appropriate system specific indications.

    The general indications for use are:

    • A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
    • B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
    • C. Proximal femoral fractures.
    • D. Avascular necrosis of the femoral head.
    • E. Non-union of proximal femoral neck fractures.
    • F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.

    Consensus hip system implants are intended for uncemented or cemented use per the system specific indications.

    Device Description

    The Consensus hip systems are semi-constrained, hip prosthesis designed for either The Consensuring bio surgery. They include the Consensus® Hip System (CHS), the Unisyn™ Hip System, and the TaperSet™ Hip System (THS). All three hip systems utilize the exact same 12/14 Morse taper trunnion. These hip stems are compatible with previously cleared CoCr heads, zirconia heads, unipolar heads, bipolar heads, UHMWPE inserts and acetabular cups.

    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about acceptance criteria or a study that proves a device meets those criteria. The document is a 510(k) summary for Consensus Orthopedics, Inc.'s hip systems, discussing their device description, indications for use, and substantial equivalence to predicate devices. It explicitly states:

    "Clinical Performance Data: No clinical studies were performed."

    The "Non-Clinical Performance Data" section mentions "All required testing per 'Guidance Document for the Preparation of Premarket Notifications of Ceramic Ball Hip Systems' were performed," and component testing according to ISO standards, but it does not provide specific acceptance criteria or report device performance against such criteria. It only lists the types of tests done.

    Therefore, I cannot provide the requested table or information about sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies, as these details are not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1