(449 days)
The EXp Acetabular Shell Liner is intended for use in reconstruction of the articulating surface of the acetabular portion of the hip that is severely disabled and/or very painful resulting from:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed total hip Arthroplasty.
Cemented and Uncemented Applications
The EXp Acetabular Shell Liner is made of polyethylene to which Vitamin E has been added. It is available in both hooded and non-hooded options. The liner is part of a complete total hip system and will be used in conjunction with an acetabular shell, femoral head and femoral stem in total hip arthroplasty. The femoral heads which are to be mated with the EXp Liner are made of Biolox forte, Biolox Delta, or CoCr alloy.
The provided text describes a 510(k) summary for the EXp Acetabular Shell Liner. This documentation focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing rather than conducting a controlled clinical study with acceptance criteria often seen in AI/ML device submissions. Therefore, many of the requested elements for AI/ML device evaluation (like sample sizes for test sets, expert ground truth, MRMC studies, training set details) are not applicable or present in this specific type of submission.
However, I can extract and structure the information related to the performance claims and the studies that support them.
Acceptance Criteria and Study for EXp Acetabular Shell Liner
The submission details several claims regarding the oxidative stability and mechanical performance of the EXp material and the EXp liners, which can be interpreted as the performance goals or "acceptance criteria" for the device to be deemed substantially equivalent. The "reported device performance" are the results of the non-clinical tests conducted to support these claims.
1. Table of Acceptance Criteria and Reported Device Performance
| Claim/Acceptance Criteria | Reported Device Performance |
|---|---|
| EXp material is more resistant to oxidation and retains mechanical performance better than conventional UHMWPE after oxidative aging. Criteria: Ultimate load (per ASTM F2183) does not decrease during oxidative aging (per ASTM F2003); Resistance to oxidation (per ASTM F2102) after aging per ASTM F2003. | After 4 weeks of aging per ASTM F2003: - Maximum ASTM F2102 oxidation index for EXp: increased from 0.1 ± 0.02 to 0.2. - Maximum ASTM F2102 oxidation index for conventional GUR1050: increased from 0.1 ± 0.04 to 3.8 ± 0.2. - Ultimate load for EXp (ASTM F2183): remained constant (63.3 ± 8.9 N non-aged, 73.1 ± 5.2 N 4-week-aged). - Ultimate load for conventional GUR1050: decreased from 71.5 ± 3.0 N (non-aged) to an embrittled condition (no sample could be machined after 4 weeks; 45.7 ± 5.9 N after 2 weeks). |
| Morphology of EXp UHMWPE is consistent with conventional UHMWPE. Criteria: No consolidation defects or voids when imaged at high magnification under scanning electron microscopy. | Freeze fracture analysis and scanning electron microscopy showed both EXp UHMWPE and 25kGy GUR 1050 material demonstrated no consolidation defects or voids at high magnification. |
| Vitamin E blended into EXp UHMWPE does not elute from the material. Criteria: No Vitamin E detected in hexane or isopropanol (IPA) extracts using GC-MS and LC-MS. | GC-MS and LC-MS analysis confirmed no Vitamin E was extracted from EXp material when refluxed in hexane at 74°C for 24 hours (detection limits: 50-250 ppb) or soaked in IPA at room temperature for 18 hours (detection limits: 1000 ppm). |
| EXp liners remain resistant to oxidation after 5 million cycles of wear testing and artificial aging. Criteria: Resistance to oxidation (per ASTM F2102) after wear testing (ISO 14242-1) followed by accelerated aging (ASTM F2003). | After 5 million cycles of wear testing (ISO 14242-1) and subsequent 4 weeks of oxidative aging (ASTM F2003): - Maximum oxidation index (ASTM F2102) for EXp liners: increased from 0.1 ± 0.02 (non-aged EXp material) to 0.2 ± 0.02 (4-week-aged EXp material). |
| EXp liners experience no failures from dynamic impingement testing. Criteria: Resistance to rim fracture under fatigue loading conditions to 1.0 MC (million cycles) per ASTM F2582, with no failure of the locking mechanism. | Dynamic impingement testing per ASTM F2582: - Three (3) artificially aged EXp liners were loaded to engage the rim with the femoral neck at 4.6 Nm (70% of static peak dislocation moments). - All three samples reached run-out (1.0 MC) with no fractures observed and no failure of the locking mechanism. |
Regarding the other requested information, which is more relevant to AI/ML device studies:
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Not applicable. This submission is for a physical medical device (acetabular liner) and relies on non-clinical, in vitro, and mechanical bench testing, not data analysis from patient studies. The "samples" refer to physical specimens of the material or device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. Ground truth in this context is established by standardized physical, chemical, and mechanical testing protocols (e.g., ASTM, ISO standards), not human expert consensus on images or clinical data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are typically for subjective assessments or discrepancy resolution in clinical or image-based studies. This submission relies on objective physical measurements.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a physical device, not an AI/ML diagnostic or assistive tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for the performance claims is derived from standardized physical and chemical measurements according to established industry standards (e.g., ASTM F2183, ASTM F2003, ASTM F2102, ISO 14242-1, ASTM F2582).
8. The sample size for the training set
- Not applicable. This device does not involve a "training set" in the context of AI/ML or statistical modeling. The material properties are inherent to its manufacturing process and composition.
9. How the ground truth for the training set was established
- Not applicable. As above, no training set is described for this physical device.
{0}------------------------------------------------
510(k) Summary
| Manufacturer: | StelKast, Inc.200 Hidden Valley RoadMcMurray, PA 15317 | MAR 2 4 2011 |
|---|---|---|
| Device Trade Name: | EXp Acetabular Shell Liner | |
| Contact: | Mr. Donald A. StevensVice Chairman(888) 273-1583 | |
| Prepared by: | Musculoskeletal Clinical Regulatory Advisers, LLC1331 H Street, NW, 12th FloorWashington, DC 20005Phone: (202)552-5800 | |
| Date Prepared: | March 14, 2011 | |
| Common Name: | Acetabular Shell Liner | |
| Classification: | 21 CFR 888.3358, Hip joint metal/polymer/metal semiconstrainedporous-coated uncemented prosthesis | |
| Class: | II | |
| Product Codes: | OQG, OQH, OQI, LPH, LWJ, JDI, MAY, LZO |
Indications For Use:
The EXp Acetabular Shell Liner is intended for use in reconstruction of the articulating surface of the acetabular portion of the hip that is severely disabled and/or very painful resulting from:
-
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
-
- Rheumatoid arthritis.
-
- Correction of functional deformity.
-
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
-
- Revision of previously failed total hip Arthroplasty.
Cemented and Uncemented Applications
{1}------------------------------------------------
Device Description:
The EXp Acetabular Shell Liner is made of polyethylene to which Vitamin E has been added. It is available in both hooded and non-hooded options. The liner is part of a complete total hip system and will be used in conjunction with an acetabular shell, femoral head and femoral stem in total hip arthroplasty. The femoral heads which are to be mated with the EXp Liner are made of Biolox forte, Biolox Delta, or CoCr alloy.
Predicate Devices:
Liners
Comparative information presented in the 510(k) supports the substantial equivalence of the EXp Acetabular Liner with respect to its indications for use, design, materials, and function.
This 510(k) demonstrates the substantial equivalence of the EXp Acetabular Liner to the following predicate devices: Acetabular Liners in the Stelkast Provident Hip System (K935484) and the Stelkast ProForm Hip System (K950827); Biomet RingLoc Acetabular Component with ArCom Polyethylene (K032396 and K970501). Previously cleared polyethylene acetabular liners to which Vitamin E has been added include the Biomet E-Poly (Vitamin E) Acetabular Liners (K050327).
All of these acetabular liners have the same intended use, have the same general design and available sizes, and are made of UHMWPE.
Non-clinical testing was performed on the EXp Acetabular Liner to determine tensile strength, impact strength, compressive strength, small punch strength, thermal properties, free radical concentration. oxidation resistance, swell ratio, hip simulator wear under normal and abrasive conditions, wear particle characterization, rim impingement, liner push-out, lever-out, and torque-out resistance, GCMS analysis of hexane extract postwear testing, fusion defect characterization, fatigue crack propagation, trans-vinylene index, and biocompatibility (i.e., mutagenicity, irritation, sensitization, and cytotoxicity testing).
The results of the performed tests demonstrate that the EXp Acetabular Liner is substantially equivalent to legally marketed predicate devices.
Femoral Heads
The StelKast Biolox Delta Ceramic Femoral heads are substantially equivalent to previously cleared Biolox Delta Ceramic Femoral Heads: Smith & Nephew (K083762); Biomet (K073102); Zimmer (K071535); and Howmedica Osteonics (K051588). The StelKast Biolox forte Ceramic Femoral Heads were approved for use with StelKast femoral stems in P040051. The CoCr Femoral Heads were cleared for use with StelKast femoral stems in K934162.
Testing in support of the ceramic femoral heads includes static burst strength, fatigue, post-fatigue burst strength, pull-off, and rotational stability.
{2}------------------------------------------------
Claims for the EXp Liner
Oxidative Stability and Mechanical Performance of the EXp material
-
- EXp is more resistant to oxidation than conventional UHMWPE and its ultimate load, as measured per ASTM F2183, does not decrease during oxidative aging per ASTM F2003, The EXp UHMWPE is a compression molded GUR 1020 blended with Vitamin E. crosslinked with gamma irradiation and terminally sterilized using ethylene oxide. For comparison, a well recognized industry standard material, conventional GUR 1050 material, sterilized using 25kGy in an inert environment was tested in parallel. The EXp UHMWPE demonstrated resistance to oxidation, as measured using ASTM F2102, after aging per ASTM F2003. Specifically, after four weeks of aging per ASTM F2003, the maximum ASTM F2102 oxidation index increased from 0.1 ± 0.02 to 0.2 for the EXp material and from 0.1 ± 0.04 to 3.8 ± 0.2 for the conventional GUR1050 material.
Consistent with the oxidation index data, the resulting mechanical performance of the materials, as determined using the ASTM F2183 small punch test. showed that the EXp material retained its mechanical performance while that of the 25kGv GUR 1050 material decreased. Specifically, the ultimate load for the EXp material remained relatively constant. 63.3 ± 8.9 N and 73.1 ± 5.2 N. for the non-aged and 4-week-aged material, respectively. The ultimate load for the conventional material decreased from 71.5 ± 3.0 N for the non-aged material to an embrittled condition in which no small punch sample could be machined. For reference, the ultimate load after two weeks of aging had reduced to 45.7 ± 5.9 N.
- EXp is more resistant to oxidation than conventional UHMWPE and its ultimate load, as measured per ASTM F2183, does not decrease during oxidative aging per ASTM F2003, The EXp UHMWPE is a compression molded GUR 1020 blended with Vitamin E. crosslinked with gamma irradiation and terminally sterilized using ethylene oxide. For comparison, a well recognized industry standard material, conventional GUR 1050 material, sterilized using 25kGy in an inert environment was tested in parallel. The EXp UHMWPE demonstrated resistance to oxidation, as measured using ASTM F2102, after aging per ASTM F2003. Specifically, after four weeks of aging per ASTM F2003, the maximum ASTM F2102 oxidation index increased from 0.1 ± 0.02 to 0.2 for the EXp material and from 0.1 ± 0.04 to 3.8 ± 0.2 for the conventional GUR1050 material.
All EXp samples were machined from GUR1020- E. which is blended by percentage weight with Vitamin E. compression molded, crosslinked with gamma irradiation, and subsequently ethylene oxide sterilized. All conventional samples were machined from GUR1050, ram extruded UHMWPE and subsequently gamma sterilized using 25-40 kGv in an inert environment. Bench testing is not necessarily indicative of clinical performance.
-
- The morphology of the EXp UHMWPE is consistent with conventional UHMWPE. Material samples of both the EXp UHMWPE and 25kGy GUR 1050 material were subjected to freeze fracture analysis. Both materials demonstrated no consolidation defects or voids when imaged at high magnification under scanning electron microscopy.
All EXp samples were machined from GUR1020- E, which is blended by percentage weight with Vitamin E, compression molded, crosslinked with gamma irradiation, and subsequently ethylene oxide sterilized. All conventional samples were machined from GUR1050, ram extruded UHMWPE and subsequently gamma sterilized using 25-40 kGy in an inert environment. Bench testing is not necessarily indicative of clinical performance.
- The morphology of the EXp UHMWPE is consistent with conventional UHMWPE. Material samples of both the EXp UHMWPE and 25kGy GUR 1050 material were subjected to freeze fracture analysis. Both materials demonstrated no consolidation defects or voids when imaged at high magnification under scanning electron microscopy.
-
- The vitamin E blended into the EXp UHMWPE does not elute from the EXp material during hexane extraction or isopropanol (IPA) extraction. Gas chromatography-mass spectrometry (GC-MS) and liquid chromatography -mass spectrometry (LC-MS) analysis of hexane solvent used for extraction of the EXp material confirmed that no Vitamin E was extracted from the material when refluxed at 74°C for 24 hours. GC-MS and LC-MS analysis of IPA solvent used for extraction of the EXp material confirmed
{3}------------------------------------------------
that no Vitamin E was extracted from the material when soaked at room temperature for 18 hours. The GC-MS and LC-MS technique have detection limits of 50-250 ppb and 1000 ppm, respectively.
All EXp samples were machined from GUR1020- E. which is blended by percentage weight with Vitamin E, compression molded, crosslinked with gamma irradiation, and subsequently ethylene oxide sterilized. Bench testing is not necessarily indicative of clinical performance.
Oxidative Stability and Mechanical Performance of the EXp Liners
-
- EXpremains resistant to oxidation after 5 million cycles of wear testing and artificial aging. Wear testing followed by accelerated aging provides a method to assess the possibility that repeated loading experienced during wear testing may change the distribution or content of the Vitamin E in the EXp material; leaving the material susceptible to oxidation. Testing was conducted per ISO 14242-1 using an AMTI hip simulator with 20g/L bovine serum lubricant to a total cycle count of 5.0 million cycles. All samples were tested in their final sterilized form. Following wear testing, the EXp Liners were aged in an oxidative environment per ASTM F2003 for 4 weeks. The EXp Liners demonstrated a resistance to oxidation, as measured using ASTM F2102. Specifically, the maximum oxidation index increased from 0.1 ± 0.02 for the non-aged EXp material to 0.2 ± 0.02 for the four-week-aged EXp material.
EXp samples were machined from GUR1020- E, which is blended by percentage weight with Vitamin E, compression molded, crosslinked with gamma irradiation, and subsequently ethylene oxide sterilized. Bench testing is not necessarily indicative of clinical performance.
- EXpremains resistant to oxidation after 5 million cycles of wear testing and artificial aging. Wear testing followed by accelerated aging provides a method to assess the possibility that repeated loading experienced during wear testing may change the distribution or content of the Vitamin E in the EXp material; leaving the material susceptible to oxidation. Testing was conducted per ISO 14242-1 using an AMTI hip simulator with 20g/L bovine serum lubricant to a total cycle count of 5.0 million cycles. All samples were tested in their final sterilized form. Following wear testing, the EXp Liners were aged in an oxidative environment per ASTM F2003 for 4 weeks. The EXp Liners demonstrated a resistance to oxidation, as measured using ASTM F2102. Specifically, the maximum oxidation index increased from 0.1 ± 0.02 for the non-aged EXp material to 0.2 ± 0.02 for the four-week-aged EXp material.
-
- The EXp liners experienced no failures as a result of dynamic impingement testing. Dynamic impingement testing, per ASTM F2582, confirmed that artificially aged EXp liners demonstrated a resistance to rim fracture under fatigue loading conditions to 1.0MC. Specifically, three liners were loaded to engage the liner rim with the femoral neck at moments equal to 4.6 Nm (70% of the static peak dislocation moments). All three samples reached run-out equal to 1.0MC and no fractures of the liners were observed throughout testing nor did the locking mechanism fail.
All EXp samples were machined from GUR1020- E, which is blended by percentage weight with Vitamin E, compression molded, crosslinked with gamma irradiation, and subsequently ethylene oxide sterilized. Bench testing is not necessarily indicative of clinical performance.
- The EXp liners experienced no failures as a result of dynamic impingement testing. Dynamic impingement testing, per ASTM F2582, confirmed that artificially aged EXp liners demonstrated a resistance to rim fracture under fatigue loading conditions to 1.0MC. Specifically, three liners were loaded to engage the liner rim with the femoral neck at moments equal to 4.6 Nm (70% of the static peak dislocation moments). All three samples reached run-out equal to 1.0MC and no fractures of the liners were observed throughout testing nor did the locking mechanism fail.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
StelKast, Inc. % Mr. Donald A. Stevens Vice Chairman 200 Hidden Valley Road McMurray, Pennsylvania 15317
MAR 2 4 2011
Re: K094035 Trade/Device Name: EXp Acetabular Shell Liner Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: OQG, OQH, OQI, LPH, LWJ, JDI, MAY, LZO Dated: March 15, 2011 Received: March 17, 2011
Dear Mr. Stevens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the includions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drig, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, inte re of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not maching.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devine can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Donald A. Stevens
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and Issing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-100.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115801.html for the Center for Devices and Radiological Health's (CDRH's) Office of Cochilian 1962.1000 roll note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.h1m for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Aty B. Ruter
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: EXp Acetabular Liner
Indications for Use:
-
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
-
- Rheumatoid arthritis.
-
- Correction of functional deformity.
-
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
-
- Revision of previously failed total hip Arthroplasty.
Cemented and Uncemented Applications
Prescription Use YES AND/OR Over-The-Counter Use NO
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Mulkerson
(Division Sign-Offy Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K094035 Page j of 1
§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.
(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.