Search Results
Found 21 results
510(k) Data Aggregation
K Number
K242612Device Name
Option ELITE Vena Cava Filter System (352506070E 352506100E)
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2024-09-10
(7 days)
Product Code
DTK
Regulation Number
870.3375Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Option™ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions:
· Pulmonary thromboembolism when anticoagulants are contraindicated
• Failure of anticoagulant therapy in thromboembolic disease
· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced
· Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter can only be performed by the jugular approach.
The Option™ELITE Vena Cava Filter 100cm System is indicated for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions:
· Pulmonary thromboembolism when anticoagulants are contraindicated
· Failure of anticoagulant therapy in thromboembolic disease
· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced
· Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entional Procedure for Filter Retrieval" in patients who no longer require a filter in patients who no longer require a filter. Retrieval of the filter can only be performed by the jugular approach.
Device Description
The Option™ELITE Vena Cava Filter is designed for the prevention of recurrent pulmonary embolism via percutaneous delivery in the inferior vena cava (IVC).
The Option™ELITE Vena Cava Filter 100cm System is designed for IVC filter insertion, delivery, deployment and placement via the popliteal and antecubital approach.
The self-centering Option™ELITE Filter is laser cut from nickel - titanium alloy (Nitinol) tubing. The Option™ELITE Filter consists of shape memory Nitinol struts emanating from a central location and is designed for optimal clot capture. Retention anchors (retention hooks) are located at the caudal portion of the filter. These anchors are intended for filter fixation to the vessel wall. The Option™ELITE Filter is intended to be used in caval diameters up to 30mm. A retrieval hook is centrally located at the cranial extremity.
The constrained Option™ELITE Filter is flexible and expands to the internal diameter of the IVC upon deployment. The Option™ELITE Filter imparts an outward radial force on the luminal surface of the vena cava to ensure proper positioning and stability. The Option™ ELITE Filter is designed to prevent pulmonary embolism while maintaining caval patency through central filtration.
The introduction kit is comprised of a filter housed in a filter cartridge. Catheter Sheath Introducer (5F ID). Angiographic Vessel Dilator with an open end, and a Pusher with deployment marker.
The Angiographic Vessel Dilator has side holes and 2 radiopaque markers, separated by 32mm (between the marker bands), that provide linear measurement of the inferior vena cava and assists in angiographic visualization when radiopaque contrast is delivered. The pusher advances the filter through the Catheter Sheath Introducer up to the deployment marker, and is then used to fix the filter in place during uncovering. The location of the distal end of the Catheter Sheath Introducer can be controlled by rotating the entire device to position the Catheter Sheath Introducer in the center of the vena cava. The Filter Cartridge houses the Option™ELITE Filter. The body of the Cartridge has text and colored arrows printed on it that identify assembly orientation, femoral is printed in green and jugular is printed in blue. The arrow of the desired access site will point into the Catheter Sheath Introducer hub. The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when used in conjunction with the delivery of radiopaque contrast media to the vena cava.
Ask a Question
K Number
K241145Device Name
TLAB® Transvenous Liver Biopsy System (TF-18C)
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2024-08-01
(98 days)
Product Code
DYB
Regulation Number
870.1340Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The TLAB® Transvenous Liver Biopsy System is intended to be used for percutaneous transjugular and transfemoral venous liver access during diagnostic and interventional procedures.
Device Description
The TLAB® Transvenous Liver Biopsy System is a single use, disposable, sterile device. The TLAB® Transvenous Liver Biopsy System consists of the following components: 18-Gauge (18Ga) Flexcore Biopsy Needle, 7 French (7Fr) Introducer Sheath with a curved metal stiffener, 5 French (5Fr) Straight Catheter, 5 French (5Fr) Curved Catheter, Tissue Removal Swabs, Bending Tool (used for transfemoral access).
Ask a Question
K Number
K233909Device Name
Cleaner Plus 18F Thrombectomy System - Cleaner Plus 18F Aspiration Catheter, Cleaner Plus 18F Handpiece with Maceration Wire, Cleaner Plus 18F Aspiration Canister
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2024-03-28
(107 days)
Product Code
QEW, KRA
Regulation Number
870.5150Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Cleaner™ Plus 18F Thrombectomy System is indicated for mechanical de-clotting, aspiration, and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral venous vasculature.
Device Description
The Cleaner Plus™ 18F Thrombectomy System is a single use device used to provide thrombectomy in the peripheral venous vasculature. The device provides additional features, such as aspiration and over-the-wire device placement.
The disposable system consists of: (1) the Aspiration Catheter & Dilator, (2) the Handpiece that includes an aspiration control and an integrated Maceration Wire, and a Peel-Away Introducer and (3) the external vacuum reservoir with pump known as the Aspiration Canister.
The system is inserted percutaneously into the vessel using an introduction catheter, the system Aspiration Catheter and Dilator may be placed over-the-wire to the site of thrombus. The device macerates intra-lumen and wall adherent thrombus with a maceration wire. The macerated thrombus is removed from the vessel using an aspiration system. The aspiration of the clot can be performed simultaneous or independently.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
Ask a Question
K Number
K232679Device Name
Cleaner Pro Thrombectomy System; Cleaner Pro Aspiration Catheter with Handpiece; Cleaner Pro Aspiration Canister
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2024-01-19
(140 days)
Product Code
QEW, KRA
Regulation Number
870.5150Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Cleaner™ Pro Thrombectomy System is indicated for the removal of fresh, soft thrombi and emboli from the vessels of the peripheral venous vasculature, and for the infusion of physician-specified fluids, including thrombolytics. The Cleaner™ Pro Thrombectomy System is not intended for use in the pulmonary vasculature for treating of pulmonary embolism.
Device Description
The Cleaner Pro Thrombectomy System is a single use device in the removal of fresh, soft emboli and thrombi and for the infusion of physician-selected fluids through the side-port of the aspiration catheter.
The disposable system consists of: (1) the Aspiration Catheter with Dilator, (2) the Handpiece with Flushing Adapter, and (3) the external vacuum reservoir with pump known as the Aspiration Canister.
The Aspiration Catheter with Dilator may be placed over-the-wire to navigate the device to the target site. Once in the target site, to complete the system, the Aspiration Canister is connected to the handpiece. The dilator is removed, and the device is activated by the user to aspirate soft emboli and thrombi. The clot is aspirated from the distal portion of the device through the handpiece and then collected in the aspiration canister reservoir.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
Ask a Question
K Number
K233432Device Name
10F Sheath and Dilator Set
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2023-12-08
(57 days)
Product Code
DYB
Regulation Number
870.1340Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The 10F Sheath and Dilator Set is indicated for introduction of therapeutic devices into the vasculature, excluding coronary and neuro vasculature.
Device Description
The 10F Sheath and Dilator Set is designed for single use and intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature. The 10F Sheath and Dilator Set consists of a 10F reinforced Introducer Sheath and a matching 10F radiopaque Dilator.
The 10F Introducer sheath is a coil reinforced sheath with a Platinum Iridium radiopaque marker band. The sheath has a hemostasis valve and a side port with stopcock. The 10F dilator is made of radiopaque, non-reinforced material with an atraumatic tip to minimize blood loss. The device does not include any coating.
Ask a Question
K Number
K230669Device Name
L-Cath Single and Dual Lumen Catheters, L-Cath Midline Catheters
Manufacturer
Argon Medical Devices
Date Cleared
2023-11-30
(265 days)
Product Code
LJS, FOZ
Regulation Number
880.5970Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The L-Cath Single and Dual Lumen Catheters and the L-Cath Midline Catheters are indicated for short or long term peripheral access to the central venous system for the administrations, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
Device Description
The L-Cath™ Single and Dual Lumen and L-Cath™ Midline Catheters are sterile, single use devices, sold as standalone devices which includes the following accessories:
• Trim Tool
• Tape measure
• Stylet (include in the final catheter assembly as applicable)
• Polyurethane catheter
These catheters are also sold as a component in a Basic Kit (catheter and introducer packaged together) which includes the following:
- BD Introsyte Autoguard Introducer or Splittable Introducer ●
- Stylet (include in the final catheter assembly as applicable)
- Trim Tool
- Tourniquet
- Tape measure
Ask a Question
K Number
K232443Device Name
Single-Loop Snare Retrieval Kit, Triple-Loop Snare Retrieval Kit
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2023-09-07
(24 days)
Product Code
MMX
Regulation Number
870.5150Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are intended for the percutaneous removal of retrievable inferior vena cava (IVC) filters that are no longer medically required, via jugular approach.
Device Description
The Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are single use devices. The disposable system consists of: [A] 9F (ID) Inner Sheath, [B] 11F (ID) Outer Sheath, [C] 8F (OD) Dilator, [D] Hemostasis Valve with Sideport, [E] High Pressure Stopcock, [F] 7F [2.4mm] (OD) x 76cm Snare Catheter with Tuohy-Borst Y-Port Adapter, 20mm x 93cm Single-Loop Snare (fully expanded) or 30mm x 93cm Triple-Loop Snare (fully expanded) with Torque Handle. The snares have radiopaque loops and are preloaded in the snare catheter. The snare catheter, inner sheath, and outer sheath have a radiopaque marker band at the distal tip for enhanced fluoroscopic visualization.
Ask a Question
K Number
K230631Device Name
Kodiak Dual Port Coaxial Introducer Kit
Manufacturer
Argon Medical Devices
Date Cleared
2023-05-22
(76 days)
Product Code
DYB
Regulation Number
870.1340Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Kodiak™ Dual Port Coaxial Introducer Kit is indicated to introduce therapeutic or diagnostic devices into the vasculature.
Device Description
The Kodiak™ Dual Port Coaxial Introducer Kit is intended to introduce therapeutic or diagnostic devices into the vasculature. The Kit Model 381416000 contains a coaxial 16F and 14F braided reinforced sheaths with a fluoropolymer coating, an embedded radiopaque marker band, and an over molded hub assembly containing two hemostasis valves and side port with stopcock. It also contains a removable Y-connector with two 7F (ID) ports with hemostasis valves, a 14F dilator, a 16ga blunt flushing needle, and a high-pressure Luer adapter. After percutaneous access is established and a working guidewire is placed, the introducer system is inserted over the guidewire and advanced under imaging. Therapeutic or diagnostic devices are inserted through the system. The Kodiak™ Dual Port Coaxial Introducer Kit is a single use device.
Ask a Question
K Number
K223176Device Name
Cleaner Plus Thrombectomy System, Cleaner Plus Aspiration Catheter, Cleaner Plus Handpiece with 65cm Maceration Wire, Cleaner Plus Handpiece with 135cm Maceration Wire, Cleaner Plus Aspiration Canister
Manufacturer
Argon Medical Devices, Inc.
Date Cleared
2023-04-10
(181 days)
Product Code
KRA, QEW
Regulation Number
870.1210Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Cleaner Plus™ Thrombectomy System is indicated for mechanical de-clotting, aspiration, and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral venous vasculature.
Device Description
The Cleaner Plus™ Thrombectomy System is a single use device used to provide thrombectomy in the peripheral venous vasculature. The device provides additional features, such as aspiration and over-the wire device placement. The disposable system consists of: (1) the Aspiration Catheter & Dilator, (2) the Handpiece that includes system controls, and an integrated Maceration Wire, and a Peel-Away Introducer and (3) the Aspiration Canister. The Aspiration Catheter with Dilator may be placed over-the-wire to navigate the device to the therapeutic site. The dilator and guidewire are removed, and the Maceration Wire, using the Peel-Away introducer is advanced through the hemostasis valve of the Aspiration Catheter to the therapeutic site and connected to the handpiece. To complete the system, the provided Aspiration Canister is connected to the handpiece to provide aspiration. The Handpiece provides controls to turn on/off maceration and/or the application of suction. Mechanical thrombectomy is achieved by rotating a flexible stainless-steel maceration wire powered by a motor inside the handpiece. The aspiration source is provided to aspirate macerated clot from the distal portion of the device through the handpiece and captures the macerated clot in the Aspiration Canister reservoir. The Aspiration Canister includes a switch to initiate the pump, and LEDs that indicate the level of the vacuum.
Ask a Question
K Number
K213232Device Name
NovaCore Semi-Automatic Biopsy Instrument
Manufacturer
Argon Medical Devices
Date Cleared
2022-07-01
(274 days)
Product Code
KNW
Regulation Number
876.1075Why did this record match?
Applicant Name (Manufacturer) :
Argon Medical Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Semi-Automatic Biopsy Instrument and associated Co-Axial Introducer Needle are used to obtain samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, pancreas, and other masses.
Device Description
The Semi-Automatic Biopsy Instrument is used to obtain soft tissue biopsies. It is composed of a spring-loaded biopsy needle fitted into a plastic handle permitting single handed specimen collection. The position of the needle may be visualized by x-ray, CT, or ultrasound. The needle has numerically ordered centimeter markings to facilitate precise depth placement. The adjustable instrument allows for a specimen notch size of 10mm or 20mm, providing clinical flexibility. The device has an adjustable co-axial adapter built into the device for either the 10mm or 20mm notch settings. The stylet travels up to 27mm from when the device is in a fully charged position (prepared for a collection with the 20 mm notch size) to when the stylet is fully extended. The needle has an echogenic tip. The semi-automatic biopsy instrument is available in several needle gauge sizes and lengths. The plunger is color coded according to the various gauge sizes, e.g., green=14-gauge, purple=16-gauge, pink=18-gauge, and yellow=20-gauge.
Ask a Question
Page 1 of 3