Search Results
Found 15 results
510(k) Data Aggregation
(54 days)
Abiomed Inc.
The Abiomed 14Fr Low Profile Introducer Kit is intended to facilitate access to the vascular system for the introduction and removal of the Impella CP Catheter and ancillary devices.
The Abiomed 14Fr Low Profile Introducer Kit (Gen 2) is a sterile, single-use, prescription device. The 14Fr Low Profile Introducer Kit consists of an introducer sheath and a tapered sheath dilator which is compatible with an 0.035" guidewire. The 14Fr Low Profile Introducer Kit is kitted with an 0.035" access guidewire, supplemental dilators and a luer adapter for convenience to help facilitate insertion. The 14Fr Low Profile Introducer Sheath consists of a sheath hub with three-way stopcock and flush port at its proximal end a sheath body at its distal end. The sheath hub features an introducer cap, hemostasis valve, side-port with three-way stopcock and flush port, a butterfly (suture pad), and connects to the dilator hub. The coil reinforced polymer sheath body has an insertion profile of 14Fr to allow the insertion and removal of the Impella CP Catheter and ancillary devices. The 14Fr Low Profile Dilator consists of a dilator body, a tapered tip at the distal end, and a hub at the proximal end, which connects with the sheath hub. Additionally, the 14Fr dilator has a hydrophilic coating to aid in insertion of the device into the vasculature.
This looks like a 510(k) summary for a medical device, specifically an introducer kit. The document describes the device, its intended use, and comparison to a predicate device. However, it does not contain the level of detail regarding acceptance criteria, study design, and performance metrics that would be expected for a typical AI/ML-based medical device described in the prompt's request.
Here's why and what information can be extracted:
This is NOT an AI/ML Medical Device.
The device described, the "Abiomed 14Fr Low Profile Introducer Kit," is a physical medical device (a catheter introducer) and not an AI/ML software device. This means the requested information about AI model performance, training sets, ground truth establishment methods, human reader studies, etc., is not applicable to this document. The "performance testing" section refers to traditional bench tests for physical device characteristics (e.g., visual inspection, mechanical testing, coating integrity, sterility, biocompatibility).
Therefore, I cannot populate most of the requested fields accurately as they pertain to AI/ML models.
However, I can extract the relevant information regarding the device's acceptance criteria and the type of study performed to show it met these criteria, as presented in the document for this physical device.
Information Extracted (Interpreted for a non-AI/ML device):
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria for a Physical Device: The document states that performance testing "met the acceptance criteria." While specific numerical criteria are not listed, the implied criteria would be successful completion of each bench test without failure.
- Reported Device Performance: "showed that the subject device, Abiomed 14Fr Low Profile Introducer Kit (Gen2), met the acceptance criteria and demonstrated substantial equivalence for its intended use."
Acceptance Criteria (Implied for physical device) Reported Device Performance Device passes specified bench tests Met acceptance criteria Demonstrates substantial equivalence Demonstrated substantial equivalence for intended use Biocompatibility: Safe for patient contact Passed ISO 10993-1 No new safety or performance issues identified None identified -
Sample size used for the test set and the data provenance:
- Sample Size: Not specified for individual tests.
- Data Provenance: Not applicable as these are bench tests on manufactured device samples, not patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth for a physical device is typically established through engineering specifications, material science, and regulatory standards, not expert consensus on patient data.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI-assisted device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This is not an AI algorithm. The device performance (standalone) was assessed through bench testing.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For a physical device, ground truth is based on engineering specifications, material properties, and established regulatory standards (e.g., ISO 10993 for biocompatibility, defined mechanical test parameters).
-
The sample size for the training set:
- Not applicable. This is not an AI/ML device.
-
How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(422 days)
Abiomed Inc.
The preCARDIA device is intended for use in selectively stopping or controlling blood flow of the inferior and superior vena cava in applications including perioperative procedures in patients requiring emergency control of hemorrhage, and for blood pressure monitoring.
The preCARDIA System is for use in selectively stopping or controlling blood flow of the inferior and superior vena cava, in applications including perioperative procedures in patients requiring emergency control of hemorrhage, and for blood pressure monitoring. The System includes a Balloon Occlusion Catheter (Catheter), Controller and Console. The Catheter has an atraumatic endovascular balloon that has been attached to the proximal portion of a commercially available Thermodilution (TD) Catheter. The single use, 110 cm disposable catheter comes in two balloon locations, 34 and 36 cm measured from the distal end of the catheter and is provided sterile. The catheter placement procedure is performed under fluoroscopy by clinicians trained in endovascular techniques. The balloon includes proximal and distal radiopaque markers to allow for visualization and confirmation of proper placement in the IVC or SVC. Following placement and securement of the catheter, the patient can be transferred to the operating room or other appropriate care location where selectively stopping or controlling blood flow in the SVC or IVC will continue alongside other standard of care treatments. The primary operating principle of the preCARDIA System is the selective stopping or controlling of the blood flow in the IVC or SVC through either sustained inflation or successive inflation and deflation cycles of the balloon as regulated by the Controller.
The provided text is a 510(k) summary for the preCARDIA Occlusion System. It details the device, its intended use, and a comparison with predicate devices. While it lists various performance tests conducted, it does not contain specific acceptance criteria or the reported device performance for an AI/ML-based device.
The document outlines bench testing and an animal study to evaluate the performance of the preCARDIA Catheter. However, it does not provide:
- A table of acceptance criteria and reported device performance.
- Sample sizes for a test set or data provenance for an AI/ML context.
- Information on expert ground truth establishment (number of experts, qualifications, adjudication method).
- Details on MRMC studies or a standalone algorithm performance.
- The type of ground truth used relevant to AI/ML (e.g., pathology, outcomes data).
- Sample size for a training set or how ground truth was established for AI/ML training data.
The text is for a physical medical device (vascular clamp/occlusion system), not an AI/ML algorithm. Therefore, the concepts requested in the prompt, such as "acceptance criteria for an AI/ML model," "training set," "test set," "ground truth," "expert consensus," and "MRMC study," are not applicable to the content provided.
Based on the provided document, I cannot fulfill the request as it pertains to AI/ML device validation. The document describes the regulatory submission for a physical medical device and its associated non-AI/ML performance testing.
Ask a specific question about this device
(139 days)
Abiomed Inc.
The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration.
The Abiomed OXY-1 System provides extracorporeal circulation for cardiopulmonary bypass support for up to six hours in duration. The OXY-1 System includes:
- A Disposable pump and oxygenator .
- A Pump Driver (blood pump) ●
- Blood tubing .
- A Console for controlling the pump and managing gas flow .
These components are designed to operate together, simplify operation, and reduce the overall equipment footprint at the bedside.
The provided text is a 510(k) summary for the Abiomed OXY-1 System, a cardiopulmonary bypass oxygenator. It is not a document describing an AI/ML-driven medical device, nor does it contain information about acceptance criteria or studies related to AI/ML device performance.
Therefore, I cannot extract the requested information regarding acceptance criteria and the study proving the device meets them, specifically in the context of an AI/ML device, as the provided document pertains to a traditional medical device (a cardiopulmonary bypass oxygenator).
The information requested in your prompt (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set details) are all relevant to the validation of AI/ML models. Since this document describes a physical medical device, these concepts are not applicable.
Ask a specific question about this device
(87 days)
Abiomed Inc.
The 14 Fr Low Profile Introducer Set is intended to facilitate access to the vascular system for the introduction and removal of the Impella CP Catheter and ancillary devices.
The Abiomed 14Fr Low Profile Introducer Set is a sterile, single-use, prescription device. The 14Fr Low Profile Introducer Set consists of an introducer sheath and a tapered sheath dilator which is compatible with an 0.035" quidewire. The 14Fr Low Profile Introducer Set is kitted with an 0.035" access quidewire, supplemental dilators and a luer adapter for convenience to help facilitate insertion.
The 14Fr Low Profile Introducer Sheath consists of a sheath hub with three-way stopcock and flush port at its proximal end a sheath body at its distal end. The sheath hub features an introducer cap, hemostasis valve, side-port with three-way stopcock and flush port, a butterfly (suture pad), and connects to the dilator hub. The coil reinforced polymer sheath body has an insertion profile of 14 Fr to allow the insertion and removal of the Impella CP Catheter and ancillary devices. The 14 Fr Low Profile Dilator consists of a dilator body, a tapered tip at the distal end, and a hub at the proximal end, which connects with the sheath hub. Additionally, the 14Fr dilator has a hydrophilic coating to aid in insertion of the device into the vasculature.
This document describes the FDA's 510(k) clearance for the Abiomed 14Fr Low Profile Introducer Set. The information provided heavily focuses on the comparison to predicate devices and performance testing. However, it does not explicitly state acceptance criteria in a quantitative format, nor does it detail a clinical study with human patients, as the clearance was based on bench testing and substantial equivalence to predicate devices.
Let's break down the available information based on your request:
1. A table of acceptance criteria and the reported device performance:
The document broadly states that the device "met the acceptance criteria," but it does not specify what those criteria are in a numerical or categorical table format for each test. Instead, it lists the types of tests performed.
Test Type | Reported Device Performance Statement |
---|---|
Visual Inspection and Dimensional Verification | Met acceptance criteria (implied by overall conclusion) |
Sheath System Verification | Met acceptance criteria (implied by overall conclusion) |
Simulated Use Testing | Met acceptance criteria (implied by overall conclusion) |
Packaging Validation | Met acceptance criteria (implied by overall conclusion) |
Biocompatibility Testing | Met acceptance criteria, safe for patient contact and duration (in accordance with ISO 10993-1) |
Sterilization Assessment | Met acceptance criteria (implied by overall conclusion) |
Leak Testing | Met acceptance criteria (implied by overall conclusion) |
Mechanical Testing | Met acceptance criteria (implied by overall conclusion) |
Coating Integrity | Met acceptance criteria (implied by overall conclusion) |
Particulate | Met acceptance criteria (implied by overall conclusion) |
Radiopacity | Radiopaque (sheath has internal metal coil) |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not specified for each individual bench test. The document mentions "Performance testing (bench) was completed," but doesn't detail the number of units tested for each specific test.
- Data Provenance: The document does not specify the country of origin of the data. The study was a retrospective evaluation based on bench testing, as it states "No clinical data were required to demonstrate substantial equivalence."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable as there was no test set involving human data or expert review to establish ground truth in the traditional sense for medical imaging or diagnostic devices. The acceptance was based on engineering and biocompatibility bench testing.
4. Adjudication method for the test set:
- Not applicable. There was no human-reviewed test set requiring adjudication in the context of this 510(k) submission.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, and its effect size:
- No, an MRMC comparative effectiveness study was not done. The document explicitly states, "No clinical data were required to demonstrate substantial equivalence." This type of study would involve human readers and clinical cases, which was not part of this clearance process.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a physical medical introducer set, not an AI algorithm. Therefore, "standalone" performance for an algorithm is not relevant. The performance evaluated was the device's physical and functional attributes through bench testing.
7. The type of ground truth used:
- The "ground truth" for this device's performance was established via engineering specifications, material science standards, and biocompatibility standards (e.g., ISO 10993-1). For example, "leak testing" would have a pre-defined maximum allowable leak rate as its ground truth, based on engineering requirements for a safe and effective introducer.
8. The sample size for the training set:
- Not applicable. This device is a physical medical device, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.
9. How the ground truth for the training set was established:
- Not applicable. As stated above, there is no "training set" for this type of device.
Ask a specific question about this device
(112 days)
Abiomed Inc.
INTENDED USE:
The Impella XR Sheath Set is intended to facilitate femoral access to the vascular system.
INDICATIONS FOR USE:
The Impella XR Sheath Set is intended for use for the percutaneous introduction of the Impella 2.5 Catheter and ancillary devices.
The Impella XR Sheath Set is a sterile, single-use, prescription device. The Impella XR Sheath Set consists of an introducer sheath (Impella XR Introducer Sheath) and a tapered sheath dilator (Impella XR Dilator) which is compatible with a 0.035" guidewire. The Impella XR Sheath Set is kitted with a 0.035" access quidewire and a supplemental dilator in the Impella XR Sheath Kit for convenience to help facilitate insertion.
The Impella XR Introducer Sheath consists of a sheath hub with a three-way stop cock and flush port at its proximal end and an expandable sheath body at its distal end. The sheath hub features an introducer cap, hemostasis valve, side-port with three-way stop cock and flush port, a butterfly (suture pad), and connects with the dilator hub. The sheath body has an insertion profile of 10 Fr and expands after removal of the dilator to allow the insertion of the Impella 2.5 catheter. The sheath body also features a hydrophobic coating to aid in the insertion of the introducer sheath.
The Impella XR Dilator consists of a dilator body, a tapered tip at the distal end, and a hub at the proximal end, which connects with the sheath hub.
The provided document describes the Impella XR Sheath Set, a medical device for introducing catheters, and its substantial equivalence to predicate devices. However, the document does NOT contain information about acceptance criteria and study details typical of an AI/ML medical device submission. Specifically, it does not include:
- A table of acceptance criteria with reported device performance for an AI/ML algorithm.
- Sample sizes for test sets, data provenance, or details about ground truth establishment by experts for AI/ML performance evaluation.
- Information on MRMC comparative effectiveness studies or standalone AI algorithm performance.
- Details regarding training set size or how ground truth was established for a training set (which are relevant to AI/ML devices).
The document is for a traditional catheter introducer device, and the performance testing mentioned (Visual Inspection, Dimensional Verification, Packaging, Sterilization, Shelf Life, Biocompatibility, and a GLP animal study) relates to the physical and biological characteristics of this type of hardware, not to an AI/ML algorithm's diagnostic or predictive performance.
Therefore, I cannot provide the requested information for an AI/ML device based on the given text.
If this were an AI/ML device submission, the requested information would typically be found in dedicated sections describing the clinical validation study, often including sensitivity, specificity, AUC, and other performance metrics against established acceptance criteria, along with detailed methodologies for data collection, expert review, and ground truth generation.
Ask a specific question about this device
(280 days)
Abiomed Inc.
The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration.
The Breethe® OXY-1 System provides extracorporeal circulation for full or partial cardiopulmonary bypass support for up to six hours. The OXY-1 System includes; a disposable pump and oxygenator, a pump driver, blood tubing, and a console for powering and controlling the pump and managing gas flow. These components are designed to operate together to reduce the overall equipment footprint at the bedside.
The provided text is a 510(k) Summary for the Abiomed OXY-1 System, a cardiopulmonary bypass oxygenator. It outlines the device's characteristics, comparison to predicates, and performance data submitted to the FDA for substantial equivalence determination.
However, the document does not contain information related to a study involving acceptance criteria for an AI/ML medical device, human reader performance, ground truth establishment by experts, or MRMC studies. The performance data mentioned (Biocompatibility, Electrical Safety, Software V&V, and Performance Testing per FDA guidance) relates to the physical and functional aspects of the OXY-1 System hardware and software, typical for a Class II medical device of this type, not an AI/ML diagnostic or prognostic system.
Therefore, I cannot extract the information required by your request from the provided text, as the request is geared towards an AI/ML device study, which is not described herein.
To fulfill your request, I would need a document detailing the clinical validation or performance study of an AI/ML medical device.
Ask a specific question about this device
(49 days)
Abiomed Inc.
The Abiomed 23 Fr Sheath is intended for introduction of pacing leads or catheters.
The Abiomed 23 Fr Sheath is packaged as a sterile kit, which includes an introducer sheath (consisting of sheath with sideport tubing, stopcock and hemostasis valve assembly), and dilators, and a guidewire. It is identical to the commercially available predicate device, the Oscor 23 Fr Adelante-S2 Introducers, K122084.
The provided text is related to an FDA 510(k) premarket notification for a medical device called the "Abiomed 23 Fr Sheath." It is a regulatory document affirming that the device is substantially equivalent to a legally marketed predicate device.
Crucially, this document does not describe the acceptance criteria or a study that proves the device meets specific performance criteria. Instead, it states that the device is identical to a previously cleared predicate device, and therefore, no additional qualification testing was required for substantial equivalence.
Therefore, I cannot provide the requested information from the given text as it explicitly states that no such studies or acceptance criteria comparisons were performed for this specific 510(k) submission.
Here's a breakdown based on the information provided, and why most of your questions cannot be answered from this document:
1. A table of acceptance criteria and the reported device performance
- Not Applicable. The document explicitly states: "Since the Abiomed 23 Fr Sheath is identical to its commercially available predicate device, no additional qualification testing was required for the substantial equivalence determination." Therefore, no new acceptance criteria or reported device performance data is presented.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. No new testing was performed for this 510(k) submission. The substantial equivalence is based on the device being identical to the predicate.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. No new ground truth establishment or expert review was part of this submission, as no new testing was conducted.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No test set was generated for this 510(k) submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This device is a catheter introducer sheath, not an AI or imaging device, so MRMC studies are not relevant. This document is a regulatory clearance for a physical medical device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not Applicable. This is not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable. No new ground truth was established for this submission.
8. The sample size for the training set
- Not Applicable. This is not an AI algorithm.
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI algorithm.
Summary from the provided text:
The Abiomed 23 Fr Sheath obtained 510(k) clearance (K201116) by demonstrating substantial equivalence to a predicate device, the Oscor Adelante-S Introducer Series (K122084).
The basis for this substantial equivalence is that the Abiomed 23 Fr Sheath is identical to its commercially available predicate device in terms of "design, packaging, manufacturing, sterilization, and labeling." Because of this asserted identity, no additional qualification testing was required for this specific 510(k) submission. Therefore, the document does not contain information about new acceptance criteria or studies performed to demonstrate performance.
Ask a specific question about this device
(49 days)
Abiomed, Inc.
The Impella CP Introducer is intended for introduction of pacing leads or catheters into the body.
The Impella CP Introducer is packaged as a sterile kit. which includes an introducer sheath (consisting of sheath with sideport tubing, stopcock and hemostasis valve assembly), and dilators, and a guidewire. It is identical to the commercially available predicate device, Oscor Introducer set, Model Adelante-S2 Introducer (cleared under K122084).
This document (K192769) outlines the 510(k) submission for the Impella CP Introducer. However, it explicitly states that no additional qualification testing was required for the substantial equivalence determination because the device is identical to its commercially available predicate device, the Oscor Introducer set, Model Adelante-S2 Introducer (cleared under K122084).
Therefore, the requested information regarding acceptance criteria and a study proving the device meets those criteria is not present in this document, as such a study was deemed unnecessary for this submission.
Here's a breakdown of why this information isn't available based on the provided text:
-
A table of acceptance criteria and the reported device performance: This is not available because no new performance testing was conducted. The device is considered to perform identically to its predicate.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as no new test set was used.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no new test set was used.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no new test set was used.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an introducer, not an AI or imaging device, so MRMC studies are irrelevant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable, as no new testing or ground truth establishment was required.
-
The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
-
How the ground truth for the training set was established: Not applicable, as this is not an AI/machine learning device.
In summary, the K192769 submission for the Impella CP Introducer relies entirely on its substantial equivalence to a predicate device (Oscor Introducer set, Model Adelante-S2 Introducer, K122084) due to their identical design, manufacturing, and performance characteristics. Therefore, the document does not contain details about new acceptance criteria or a study to prove their fulfillment.
Ask a specific question about this device
(342 days)
ABIOMED, INC.
The IMPELLA 2.5 PLUS Catheter is intended for partial circulatory support using an extracorporeal bypass control unit, for periods up to 6 hours. It is also intended to be used to provide partial circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass.
The IMPELLA 2.5 PLUS Catheter also provides pressure measurements which are useful in determining intravascular pressure.
The Impella 2.5 Plus is an update to the design of the IMPELLA 2.5, which was cleared by the FDA in 2008 (under K063723). The predicate device, the IMPELLA 2.5, and the Impella 2,5 Plus have identical designs and materials of construction. The only design modifications made to the IMPELLA 2.5 for the Impella 2.5 Plus are slight increases in the diameters of the inflow can ula impeller and pump housing (from 12 F to 14 F). As a result of this increase, the Impella 2.5 Plus has approximately 30% higher flow than the IMPELLA 2.5.
This 510(k) summary describes a medical device, the Impella 2.5 Plus, which is an updated version of a previously cleared device, the Impella 2.5. The submission focuses on demonstrating substantial equivalence to the predicate device due to minor design modifications (increased diameters of the inflow cannula impeller and pump housing).
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state formal acceptance criteria with numerical targets. Instead, the "acceptance criteria" are implied to be substantially equivalent performance to the predicate device, the IMPELLA 2.5, and compliance with general design and safety standards.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Functional Equivalence | |
- Operate as intended | - DVV testing demonstrated that the Impella 2.5 Plus operated as intended. |
- Substantial equivalence to predicate IMPELLA 2.5 | - DVV testing demonstrated that the Impella 2.5 Plus was substantially equivalent to the IMPELLA 2.5. |
- Identical functional scientific technology | - The Impella 2.5 Plus employs identical functional scientific technology as the IMPELLA 2.5. |
Power/Flow Performance | |
- Maintain adequate flow (or improve upon predicate) | - Mean Flow: Up to 3.3 L/Min (60 mmHg differential pressure) - This is an improvement over the predicate's 2.5 L/Min. |
- Appropriate pump speed | - Pump Speed: 46,000 RPM |
Safety and Compatibility | |
- No new risks from minor design differences | - Testing was completed to mitigate any possible new risks from the minor design differences. The results of bench testing demonstrated that the design changes did not raise any issues related to safety or effectiveness compared to the IMPELLA 2.5. |
- Compatibility with IMPELLA Controller | - Software Verification testing was completed to verify the Impella 2.5 Plus could be used with the IMPELLA Controller. Electrical Standards and Safety testing (IEEC 60601-1, Applied Part) was completed to verify compliance for use of the Impella 2.5 Plus with the IMPELLA Controller. |
- Compliance with electrical safety standards | - Electrical Standards and Safety testing (IEEC 60601-1, Applied Part) was completed to verify compliance. |
- Design modifications do not affect intended use or fundamental science | - Design modifications: do not affect the intended use of the device, AND do not alter the fundamental scientific technology of the device. |
Biocompatibility/Material Integrity | - The predicate device and Impella 2.5 Plus have identical designs and materials of construction (e.g., Polyurethane for catheter, cannula, and pigtail). Implicitly, this implies that previous biocompatibility and material integrity assurances for the predicate apply, and no new testing beyond comparison was deemed necessary for these aspects due to identical materials. |
2. Sample Size Used for the Test Set and Data Provenance
The submission indicates that the studies conducted were bench tests and verification & validation (V&V) testing. It does not specify a "test set" in the context of patient data or clinical samples.
- Sample Size: Not applicable in the sense of patient-based test sets. The sample size would refer to the number of devices tested in bench experiments. This detail is not provided in the summary.
- Data Provenance: Not applicable in the sense of country of origin or retrospective/prospective clinical data for this type of submission. The performance data is derived from in vitro (bench) testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. The ground truth for this device's performance is established through engineering and scientific measurements and comparisons against established specifications and the predicate device's performance. It does not involve expert image interpretation or clinical diagnosis.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical or image-based studies where expert consensus is required. This device relies on direct physical measurements and engineering verification.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance
Not applicable. This device is a mechanical circulatory support device, not an AI-powered diagnostic or assistive tool for human readers.
6. If a Standalone (i.e., Algorithm-only without human-in-the-loop) Performance Study Was Done
Not applicable. This device is a physical medical device, not an algorithm. Its "standalone" performance is its bench-test performance as described.
7. The Type of Ground Truth Used
The "ground truth" for this submission is established by:
- Engineering specifications and design requirements: The device's ability to meet its intended design parameters.
- Performance of the predicate device (IMPELLA 2.5): The Impella 2.5 Plus is compared against the known performance characteristics of its legally marketed predecessor.
- International standards: Compliance with relevant standards like IEC 60601-1 for electrical safety.
- Functional operation: Verification that the device "operated as intended."
8. The Sample Size for the Training Set
Not applicable. This is a medical device, not an AI model, so there is no training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(30 days)
ABIOMED, INC.
The IMPELLA Controller with Flow Control is an extracorporeal bypass control unit intended to be used to provide circulatory support for periods up to 6 hours. It is also intended to be used to provide circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass. The Impella Controller with Flow Control is intended to be used by trained healthcare professionals in healthcare facilities and medical transport (i.e. ambulance, helicopter, or fixed-wing aircraft) environments.
The IMPELLA Controller with Flow Control also displays pressure measurement readings, which are useful in determining intravascular pressure.
The IMPELLA Controller with Flow Control is identical to the IMPELLA Controller in its physical characteristics. The only difference is that an additional operating mode. Flow Control Mode, has been implemented via a software modification. Both the IMPELLA Controller and the IMPELLA Controller with Flow Control are identical microprocessor-based pump motor drivers incorporating an infusion system. They both are intended to be used by trained healthcare professionals in hospital and medical transport environments.
The provided text describes a 510(k) summary for the IMPELLA Controller with Flow Control. This submission primarily focuses on the device's substantial equivalence to a predicate device and its compliance with various electrical, safety, and software standards. It doesn't present an acceptance criteria table with reported device performance in the typical format of clinical or diagnostic studies. Instead, the "acceptance criteria" are implied by the standards and testing performed to demonstrate that the new Flow Control Mode is safe and effective and that the device as a whole is substantially equivalent to its predicate.
Here's an attempt to structure the information based on your request, with significant caveats due to the nature of the provided document:
Acceptance Criteria and Study Summary for IMPELLA Controller with Flow Control
The IMPELLA Controller with Flow Control is a medical device (a non-roller type cardiopulmonary blood pump) that received 510(k) clearance based on its substantial equivalence to a predicate device, the IMPELLA Controller (K093801). The primary change in the new device is a software modification to include a "Flow Control Mode." The studies performed were largely in vitro bench tests and compliance assessments against recognized standards to demonstrate that this new mode did not introduce new safety or effectiveness concerns and that the device's overall performance remained consistent with its predicate.
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a 510(k) summary based on substantial equivalence, specific numerical performance metrics for acceptance criteria and device performance, as might be found in a diagnostic accuracy study, are not explicitly provided. Instead, the acceptance criteria are met by demonstrating compliance with established standards and proper functioning of the new feature.
Acceptance Criteria Category | Specific Acceptance Criteria (Implied by document) | Reported Device Performance (as per document) |
---|---|---|
Electromagnetic Compatibility (EMC) | Conformance with IEC 60601-1-2, including IEC 61000-3-X and IEC 61000-4-X, and EN 55011. | Testing found "in conformance" with specified standards. |
Electrical Safety | Conformance with IEC 60601-1 - Part 1. | Testing found "in conformance" with specified standard. |
Packaging and Shipping | Conformance with ISTA 2A and EN 868. | Testing found "in conformance" with specified standards. |
Altitude and Vibration | Testing per RTCA/DO-160C during operation. | Testing performed "per" specified standard. |
Software Design & Testing | Compliance with "Guidance for Industry and FDA Staff - Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices" (2005). | Software design & testing "complied with" the FDA guidance document. |
System Durability (In Vitro) | New Flow Control Mode does not compromise system durability, similar to predicate. | System Durability Testing was completed. (Specific results not detailed, but implied to be acceptable as no new safety/effectiveness issues were raised). |
System Characterization (In Vitro) | Performance characteristics of the new Flow Control Mode meet specifications. | System Characterization Test was completed. (Implied acceptable performance). |
System Flow Characterization (In Vitro) | Flow Control Algorithm behaves as designed, and the two control modes (Flow Control and Speed Control) are equivalent for use with IMPELLA pump catheters. | System Flow Characterization Test was completed and showed "the Flow Control Algorithm behaves as designed, and that the 2 control modes are equivalent for use with the IMPELLA pump catheters." |
Overall Safety & Effectiveness | The IMPELLA Controller with Flow Control does not raise new issues with safety or effectiveness compared to the predicate device. | "The results of the bench testing demonstrated that the IMPELLA Controller with Flow Control did not raise new issues with safety or effectiveness." |
2. Sample Size for Test Set and Data Provenance
The "studies" described are primarily in vitro (bench) tests and compliance assessments for a medical device's control unit, not clinical trials involving human subjects or diagnostic image analysis. Therefore, concepts like "test set sample size" and "data provenance" (country of origin, retrospective/prospective) in the context of clinical data are not directly applicable here. The "test set" would refer to the physical device prototypes and test setups used in the various engineering and performance evaluations.
- Sample Size: Not specified in terms of units or duration beyond mentioning "System Durability Testing," "System Characterization Test," and "System Flow Characterization Test." These would involve testing an unspecified number of device units or components under defined conditions.
- Data Provenance: The tests are in vitro (bench testing) and compliance assessments performed by the manufacturer, ABIOMED, Inc., presumably at their facilities. There's no mention of country of origin for clinical data as no clinical data is presented for this specific 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This 510(k) summary does not involve determining "ground truth" through expert review in the sense of clinical diagnosis or image interpretation. The "ground truth" for the engineering and performance tests would be defined by the technical specifications, international standards, and the expected behavior of the device as designed. The expertise involved would be that of the engineers and quality assurance personnel conducting the tests and verifying compliance. No specific number or qualification of "experts" is provided in the document for establishing such ground truth.
4. Adjudication Method for the Test Set
Not applicable. This is not a study involving human readers or diagnostic interpretation that would require an adjudication method like 2+1 or 3+1. The results of the engineering tests would be documented, analyzed against predefined acceptance criteria (specifications and standards), and validated by engineers and quality control personnel.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This 510(k) pertains to a hardware/software control unit for a blood pump, not an imaging device or diagnostic algorithm that would typically involve human readers interpreting cases with and without AI assistance. The "AI" in this context is the software's "Flow Control Algorithm," which controls the mechanical pump.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)
Yes, a form of standalone performance study was done for the algorithm, albeit in an in vitro context. The document states:
- "Performance characteristics of the IMPELLA Controller with Flow Control were tested in vitro to verify that the new Flow Control Mode met its specifications."
- "Additional specialized performance testing was completed to verify that the 2 control modes are substantially equivalent. This testing showed that the Flow Control Algorithm behaves as designed, and that the 2 control modes are equivalent for use with the IMPELLA pump catheters."
This demonstrates that the "Flow Control Algorithm" operating within the device was tested independently (without human intervention during the test itself, although humans initiated and monitored the tests) to ensure it performed as designed and met its specifications in vitro.
7. Type of Ground Truth Used
The "ground truth" for the in vitro performance studies and compliance testing was based on:
- Engineering Specifications: The pre-defined design requirements and specifications for the device's functionality, especially the new Flow Control Mode.
- International Standards: Adherence to established and recognized standards for medical devices (e.g., IEC 60601-1-2 for EMC, IEC 60601-1 Part 1 for electrical safety, RTCA/DO-160C for altitude/vibration).
- Predicate Device Performance: The established, safe, and effective performance of the existing IMPELLA Controller (K093801), against which the new device's overall safety and effectiveness were compared to establish substantial equivalence.
8. Sample Size for the Training Set
Not applicable. This submission does not describe a machine learning algorithm that requires a "training set" of data in the typical sense. The software modification is explicitly stated as implementing a "Flow Control Mode" and its algorithm's behavior was verified through in vitro testing. There's no indication that machine learning was used where a training set distinct from a test set would be defined.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of the information provided.
Ask a specific question about this device
Page 1 of 2