K Number
K200109
Device Name
OXY-1 System
Manufacturer
Date Cleared
2020-10-23

(280 days)

Product Code
Regulation Number
870.4350
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration.

Device Description

The Breethe® OXY-1 System provides extracorporeal circulation for full or partial cardiopulmonary bypass support for up to six hours. The OXY-1 System includes; a disposable pump and oxygenator, a pump driver, blood tubing, and a console for powering and controlling the pump and managing gas flow. These components are designed to operate together to reduce the overall equipment footprint at the bedside.

AI/ML Overview

The provided text is a 510(k) Summary for the Abiomed OXY-1 System, a cardiopulmonary bypass oxygenator. It outlines the device's characteristics, comparison to predicates, and performance data submitted to the FDA for substantial equivalence determination.

However, the document does not contain information related to a study involving acceptance criteria for an AI/ML medical device, human reader performance, ground truth establishment by experts, or MRMC studies. The performance data mentioned (Biocompatibility, Electrical Safety, Software V&V, and Performance Testing per FDA guidance) relates to the physical and functional aspects of the OXY-1 System hardware and software, typical for a Class II medical device of this type, not an AI/ML diagnostic or prognostic system.

Therefore, I cannot extract the information required by your request from the provided text, as the request is geared towards an AI/ML device study, which is not described herein.

To fulfill your request, I would need a document detailing the clinical validation or performance study of an AI/ML medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 23, 2020

Abiomed Inc. % Ken Ryder Senior Director Global Regulatory Affairs 1500 John Ave Suite 190 Halethorpe, Maryland 21227

Re: K200109

Trade/Device Name: OXY-1 System Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ, KFM, DWA, DWF Dated: September 21, 2020 Received: September 22, 2020

Dear Ken Ryder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200109

Device Name OXY-1 System

Indications for Use (Describe)

The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Product: Oxy-1 System Section: 5 – 510(k) Summary

510(k) Summary

I. SUBMITTER

Abiomed, Inc. 22 Cherry Hill Dr. Danvers, MA 01923 Phone: 978-646-1707

Contact Person: J. Kenneth Ryder Date Prepared: October 21, 2020

II. DEVICE

Name of Device: OXY-1 System Common or Usual Name: OXY-1 System Classification Name: Cardiopulmonary bypass oxygenator Regulatory Class: II Product Code: DTZ - Oxygenator, Cardiopulmonary Bypass KFM - Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type DWA - Control, Pump Speed, Cardiopulmonary Bypass DWF - Catheter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass

III. PREDICATE DEVICE(S)

Medtronic Affinity NT (Model 511) Oxygenator (K191029) Maquet, Rotaflow Pump (K991864) Maquet, Rotaflow Console (K991864) Maquet, HLM Tubing Set with Bioline coating (K080592)

IV. DEVICE DESCRIPTION

The Breethe® OXY-1 System provides extracorporeal circulation for full or partial cardiopulmonary bypass support for up to six hours. The OXY-1 System includes; a disposable pump and oxygenator, a pump driver, blood tubing, and a console for powering and controlling the pump and managing gas flow. These components are designed to operate together to reduce the overall equipment footprint at the bedside.

V. INDICATION FOR USE

The OXY-1 System is indicated for:

  • OXY-1 System The OXY-1 System is intended to be used for extracorporeal circulation. The OXY-1 System pumps, oxygenates and removes carbon dioxide from blood during cardiopulmonary bypass up to 6 hours in duration.
    The identified predicates are indicated for the following:

{4}------------------------------------------------

Product: Oxy-1 System
Section: 5 – 510(k) Summary Page 2 of 5
Medtronic - Affinity NT511 (K191029)The AFFINITY® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to beused in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxidefrom the blood and to cool or warm the blood during routine cardiopulmonary bypassprocedures up to 6 hours in duration.
Maquet - RotaflowPump – RF 32(K991864)The Rotaflow centrifugal pump system is intended for use in an extracorporeal perfusioncircuit to pump blood during short duration cardiopulmonary bypass procedures lasting6 hours or less.
Maquet - RotaflowConsole (K991864)The Rotaflow centrifugal pump system is intended for use in an extracorporeal perfusioncircuit to pump blood during short duration cardiopulmonary bypass procedures lasting6 hours or less.
Maquet - HLM TubingSet with Bioline coating(K080592)The HLM Tubing Sets with Bioline Coating are designed to be used in extracorporealcirculation during cardiopulmonary bypass procedures lasting six hours or less.

The subject device and the identified predicates have the same intended use for extracorporeal circulation during full or partial cardiopulmonary bypass. The Indications for Use for the OXY-1 System is substantially equivalent to the predicate devices. Differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicates.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Oxygenator TechnologicalCharacteristics /Performance SpecificationsSubject DeviceOxy-1 System (PLU Disposable)Predicate DeviceMedtronic - Affinity NT 511 (K191029)
Materials
MembranePolymethylpentenePolypropylene
HousingPolycarbonatePolycarbonate
PottingUrethaneUrethane
Design
Heat ExchangerNoYes
Membrane TypeHollow FiberHollow Fiber
Effective Membrane Surface Area2.4 m22.5 m2
Gas PathwayTwo Gas PathwaysSingle Gas Pathway
Oxygenator GeometryCylindricalCylindrical
Tubing Connectors3/8"3/8"
Priming Volume (Oxygenator Only)285 ml215 ml
SterileSAL 10^-6SAL 10^-6
Sterilization methodEthylene OxideEthylene Oxide
Performance Specifications
Blood Flow Rate0.5-5.0 Lpm0.5 - 7.0 Lpm
Gas TransferTested per "Guidance for Cardiopulmonary BypassOxygenators 510(k) Submissions" Dated November13, 2000Tested per "Guidance for CardiopulmonaryBypass Oxygenators 510(k) Submissions" DatedNovember 13, 2000
HemolysisAverage NIH of 0.0210 mg/100LAverage NIH of 0.0318 mg/100L
Pressure Drop (at 5 Lpm)Average of 38 mmHgAverage of 52 mmHg

{5}------------------------------------------------

Centrifugal PumpTechnologicalCharacteristics /Performance SpecificationsSubject DevicePredicate Device
Oxy-1 System (PLU Disposable)Maquet - Rotaflow Pump - RF 32 (K991864)
Materials
HousingPolycarbonatePolycarbonate
ImpellerPolycarbonatePolycarbonate
BearingPolyethylene, SapphirePolyethylene, Sapphire
Design
Pump OperationCentrifugal pumpCentrifugal pump
CouplingMagneticMagnetic
BearingSingle pivotSingle pivot
Priming Volume (Pump Only)34ml32ml
Tubing Connectors3/8"3/8"
SterileSAL 10^-6SAL 10^-6
Sterilization methodEthylene OxideEthylene Oxide
Performance Specifications
Blood Flow Rate0.5-5.0 Lpm0.5 - 9.0 Lpm
HemolysisAverage NIH of 0.0210 mg/100LAverage NIH of 0.0318 mg/100L
Controller & DriverTechnologicalCharacteristics /Performance SpecificationsSubject DeviceOxy-1 System (Console)Predicate DeviceMaquet - Rotaflow Console (K991864)
Design
CouplingMagneticMagnetic
Energy SourceBattery / MainsBattery / Mains
Other features
Sweep Gas SourceInternal / ExternalExternal
Sighing Gas SourceInternal / ExternalExternal
Software
Constant RPM ModeYesYes
Constant Flow ModeNoYes
-Display
--Blood FlowYesYes
--Pump RPM SettingsYesYes
--Internal Sweep Gas SourceSettingsYesN/A
--Alarm Limit SettingsYesYes
-Alarms
--Blood FlowYesYes
--BubbleYesYes
--Low BatteryYesYes
--System Failure AlarmsYesYes
Hardware
-Bubble SensorYesYes
-Flow SensorYesYes
Emergency Drive UnitHand CrankHand Crank
Performance Specifications

{6}------------------------------------------------

Controller & DriverTechnologicalCharacteristics /Performance SpecificationsSubject DevicePredicate Device
RPM Range0-4500 RPM0-5000 RPM
Blood Tubing TechnologicalCharacteristics /Performance SpecificationsSubject DevicePredicate Device
Oxy-1 System (PLU Disposable)Maquet - HLM Tubing Set with Bioline coating(K080592)
Materials
TubingPolyvinyl ChloridePolyvinyl Chloride
Surface TreatmentNoneBioline (albumin and heparin)
Design
Tubing Connectors3/8"3/8"

VII. PERFORMANCE DATA

The following performance data was provided in support of the substantial equivalence determination.

Biocompatibility Testing:

Biocompatibility testing of the OXY-1 PLU Disposable was conducted for circulating blood for less than 24 hours.

Electrical safety and electromagnetic compatibility (EMC):

The system complies with the IEC 60601-1 standard for electrical safety and the IEC 60601-1-2 standard for EMC.

Software Verification and Validation Testing:

The software is classified as "major" level of concern. Testing and documentation for a "major" level of concern are provided in the submission.

Performance Testing:

The following testing was conducted as recommended in "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff" dated November 13, 2000.

  • Gas Transfer Testing
  • Blood and Gas Path Integrity
  • Hemolysis
  • Hydraulic Performance
  • Volatile Organic Compounds
  • Particulates evaluation
  • ISO 80601-2-69
  • Software V&V
  • Reliability
  • Shelf Life

No animal studies or clinical studies were required or conducted.

{7}------------------------------------------------

Product: Oxy-1 System Section: 5 – 510(k) Summary

VIII. CONCLUSIONS

The OXY-1 System has the same intended use as the identified predicates, has been demonstrated to be substantially equivalent to the identified predicates through performance testing, and the differences do not raise different questions of safety and effectiveness.

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”