Search Results
Found 22 results
510(k) Data Aggregation
(7 days)
The Option™ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions:
· Pulmonary thromboembolism when anticoagulants are contraindicated
• Failure of anticoagulant therapy in thromboembolic disease
· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced
· Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter can only be performed by the jugular approach.
The Option™ELITE Vena Cava Filter 100cm System is indicated for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions:
· Pulmonary thromboembolism when anticoagulants are contraindicated
· Failure of anticoagulant therapy in thromboembolic disease
· Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced
· Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entional Procedure for Filter Retrieval" in patients who no longer require a filter in patients who no longer require a filter. Retrieval of the filter can only be performed by the jugular approach.
The Option™ELITE Vena Cava Filter is designed for the prevention of recurrent pulmonary embolism via percutaneous delivery in the inferior vena cava (IVC).
The Option™ELITE Vena Cava Filter 100cm System is designed for IVC filter insertion, delivery, deployment and placement via the popliteal and antecubital approach.
The self-centering Option™ELITE Filter is laser cut from nickel - titanium alloy (Nitinol) tubing. The Option™ELITE Filter consists of shape memory Nitinol struts emanating from a central location and is designed for optimal clot capture. Retention anchors (retention hooks) are located at the caudal portion of the filter. These anchors are intended for filter fixation to the vessel wall. The Option™ELITE Filter is intended to be used in caval diameters up to 30mm. A retrieval hook is centrally located at the cranial extremity.
The constrained Option™ELITE Filter is flexible and expands to the internal diameter of the IVC upon deployment. The Option™ELITE Filter imparts an outward radial force on the luminal surface of the vena cava to ensure proper positioning and stability. The Option™ ELITE Filter is designed to prevent pulmonary embolism while maintaining caval patency through central filtration.
The introduction kit is comprised of a filter housed in a filter cartridge. Catheter Sheath Introducer (5F ID). Angiographic Vessel Dilator with an open end, and a Pusher with deployment marker.
The Angiographic Vessel Dilator has side holes and 2 radiopaque markers, separated by 32mm (between the marker bands), that provide linear measurement of the inferior vena cava and assists in angiographic visualization when radiopaque contrast is delivered. The pusher advances the filter through the Catheter Sheath Introducer up to the deployment marker, and is then used to fix the filter in place during uncovering. The location of the distal end of the Catheter Sheath Introducer can be controlled by rotating the entire device to position the Catheter Sheath Introducer in the center of the vena cava. The Filter Cartridge houses the Option™ELITE Filter. The body of the Cartridge has text and colored arrows printed on it that identify assembly orientation, femoral is printed in green and jugular is printed in blue. The arrow of the desired access site will point into the Catheter Sheath Introducer hub. The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when used in conjunction with the delivery of radiopaque contrast media to the vena cava.
This document describes a 510(k) premarket notification for the Option™ELITE Vena Cava Filter System and Option™ELITE Vena Cava Filter 100cm System. The purpose of the submission is to incorporate results from the PRESERVE study summary data into the Instructions for Use (IFU).
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not include a specific table of acceptance criteria or reported device performance metrics in the traditional sense of a performance study. This submission is for incorporating existing clinical study results (PRESERVE study) into the device's labeling, rather than presenting new performance data for a modified device. The document explicitly states:
"No new operating principles have been introduced with the subject device. The subject device operates using the identical fundamental scientific technology as the predicate device; therefore, no performance testing was necessary nor was any performed. The clinical testing was previously provided to FDA under the Investigational Device Exemption and the PRESERVE study."
Because no new performance testing was conducted for this specific 510(k) submission, there are no new device performance metrics to report against an acceptance criterion in this document. The existing performance would be based on the prior clearance of the predicate devices and the PRESERVE study.
2. Sample Size Used for the Test Set and Data Provenance
The document refers to the "PRESERVE study." However, it does not provide details on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for this study within the provided text. To obtain this information, one would need to consult the original PRESERVE study documentation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The provided document does not contain information on the number of experts used to establish ground truth or their qualifications for the PRESERVE study's test set.
4. Adjudication Method for the Test Set
The provided document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set of the PRESERVE study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC comparative effectiveness study is typically used for AI-assisted diagnostic devices to assess the impact of AI on human reader performance. Since the device is a Vena Cava Filter System, it is a physical medical device, not a diagnostic imaging AI algorithm. Therefore, no MRMC comparative effectiveness study was done for this device.
6. Standalone (Algorithm Only) Performance
The device is a physical Vena Cava Filter System, not an algorithm or software-only device. Therefore, the concept of "standalone performance" for an algorithm does not apply to this device.
7. Type of Ground Truth Used
Given that this is a physical medical device (vena cava filter) and the PRESERVE study would be a clinical trial, the "ground truth" would likely be derived from patient outcomes, imaging data, and clinical assessments related to the filter's effectiveness in preventing pulmonary embolism and its safety profile. However, the document does not explicitly state the type of ground truth used for the PRESERVE study.
8. Sample Size for the Training Set
The concept of a "training set" typically applies to machine learning algorithms. Since the device is a physical medical device and not an AI/ML product, no training set was used in the context of algorithm development. Clinical trials, like PRESERVE, gather data for validation and effectiveness, not for training an algorithm.
9. How Ground Truth for the Training Set Was Established
As noted above, no training set was used in the context of an algorithm for this physical device. Therefore, how ground truth for a training set was established is not applicable.
In summary, the provided FDA 510(k) clearance letter and summary primarily focus on demonstrating substantial equivalence to a predicate device and incorporating pre-existing clinical study (PRESERVE study) data into the device's labeling. It does not present new performance data or detail the methodology of the PRESERVE study in the way one might expect for a new device's initial clearance or an AI/ML product. To get specific details about the PRESERVE study (sample size, ground truth, expert involvement, etc.), one would need to refer to its full study report or publications.
Ask a specific question about this device
(98 days)
The TLAB® Transvenous Liver Biopsy System is intended to be used for percutaneous transjugular and transfemoral venous liver access during diagnostic and interventional procedures.
The TLAB® Transvenous Liver Biopsy System is a single use, disposable, sterile device. The TLAB® Transvenous Liver Biopsy System consists of the following components: 18-Gauge (18Ga) Flexcore Biopsy Needle, 7 French (7Fr) Introducer Sheath with a curved metal stiffener, 5 French (5Fr) Straight Catheter, 5 French (5Fr) Curved Catheter, Tissue Removal Swabs, Bending Tool (used for transfemoral access).
Here's a breakdown of the acceptance criteria and study information for the TLAB® Transvenous Liver Biopsy System, based on the provided text:
This document describes a medical device (TLAB® Transvenous Liver Biopsy System) not an AI/ML device, therefore, sections related to AI/ML specific criteria (such as MRMC studies, standalone algorithm performance, training set details) are not applicable.
Acceptance Criteria and Reported Device Performance
Device Trade Name: TLAB® Transvenous Liver Biopsy System (TF-18C)
Regulation Number: 21 CFR 870.1340 - Catheter Introducer
Regulatory Class: Class II
Product Code: DYB
The device demonstrated substantial equivalence to predicate devices (TLAB® Transjugular Liver Access Set K022634 and Traveler Portal Vein Access Set K213638) based on a comparison of technological characteristics, indications for use, materials, packaging, principle of operation, design features, sterilization process, and in-vitro performance testing.
Acceptance Criteria Category | Specific Criteria / Test Performed | Reported Device Performance / Outcome |
---|---|---|
Performance Testing | Dimensional | Met acceptance criteria |
Visual | Met acceptance criteria | |
Radiopacity | Met acceptance criteria | |
Corrosion Resistance | Met acceptance criteria | |
Simulated Use | Met acceptance criteria | |
Tensile Strength | Met acceptance criteria | |
Torque Strength | Met acceptance criteria | |
Liquid Leakage | Met acceptance criteria | |
Luer Functional Testing (ISO 80369-7) | Met acceptance criteria | |
Catheter Functional Testing (ISO 10555-1) | Met acceptance criteria | |
Resistance to Breakage (ISO 9626) | Met acceptance criteria | |
Shipping Qualification Testing | Met acceptance criteria | |
Design Validation | Met acceptance criteria | |
Summative Usability | Met acceptance criteria | |
Biocompatibility (ISO 10993-1:2020) | Cytotoxicity (ISO 10993-5) | Previously performed, applicable to new device; met criteria |
Sensitization (ISO 10993-23) | Previously performed, applicable to new device; met criteria | |
Irritation or Intracutaneous Reactivity (ISO 10993-10) | Previously performed, applicable to new device; met criteria | |
Material Mediated Pyrogen (ISO 10993-11) | Previously performed, applicable to new device; met criteria | |
Acute Systemic Toxicity (ISO 10993-11) | Previously performed, applicable to new device; met criteria | |
Hemocompatibility (ISO 10993-4): ASTM Hemolysis Assay, Direct and Extract Methods (ISO) | Previously performed, applicable to new device; met criteria | |
Hemocompatibility (ISO 10993-4): Complement Activation Assay, SC5b-9 Method with Comparison Article (ISO) | Performed anew; met criteria | |
Hemocompatibility (ISO 10993-4): Partial Thromboplastin Time (PTT) Assay with Comparison Article (ISO) | Performed anew; met criteria | |
Hemocompatibility (ISO 10993-4): Heparinized Platelet and Leukocyte Count Assay with Comparison Article (ISO) | Performed anew; met criteria | |
Sterilization | Minimum SAL 10-6, EtO | Met acceptance criteria |
Shelf-Life | 3 years | Met acceptance criteria (based on predicate) |
Study Details
-
Sample size used for the test set and the data provenance:
- The document primarily describes bench-top testing (in-vitro performance) and biocompatibility testing. Specific sample sizes for each non-clinical test are not explicitly stated in the provided text, but it mentions that "A series of testing was conducted".
- Data Provenance: All studies were performed following approved protocols under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58. This indicates the testing was conducted in a controlled laboratory environment. No specific country of origin or retrospective/prospective nature for these non-clinical tests is given beyond GLP compliance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable as the described studies are non-clinical, bench-top performance and biocompatibility tests. They do not involve human interpretation or subjective clinical "ground truth" that would require expert consensus. The acceptance criteria themselves serve as the 'ground truth' for these engineering and materials tests.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This question is not applicable as the described studies are non-clinical, bench-top performance and biocompatibility tests. Adjudication methods are typically used in clinical studies or studies involving subjective human interpretation (e.g., image reading) where disagreement among experts needs resolution.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. Therefore, no MRMC comparative effectiveness study was performed in the context of AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. There is no algorithm to test in a standalone manner.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For Performance Testing: The "ground truth" is defined by the acceptance criteria established in accordance with protocols based on guidance and industry standards. These include established engineering and material science standards (e.g., ISO for luer, catheter, and breakage testing) that define acceptable performance.
- For Biocompatibility Testing: The "ground truth" is defined by the internationally recognized standards for biocompatibility (ISO 10993 series) and FDA Guidance, which specify test methods and acceptable limits for biological responses.
-
The sample size for the training set:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. There is no machine learning model that requires a training set.
-
How the ground truth for the training set was established:
- This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device.
Ask a specific question about this device
(107 days)
The Cleaner™ Plus 18F Thrombectomy System is indicated for mechanical de-clotting, aspiration, and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral venous vasculature.
The Cleaner Plus™ 18F Thrombectomy System is a single use device used to provide thrombectomy in the peripheral venous vasculature. The device provides additional features, such as aspiration and over-the-wire device placement.
The disposable system consists of: (1) the Aspiration Catheter & Dilator, (2) the Handpiece that includes an aspiration control and an integrated Maceration Wire, and a Peel-Away Introducer and (3) the external vacuum reservoir with pump known as the Aspiration Canister.
The system is inserted percutaneously into the vessel using an introduction catheter, the system Aspiration Catheter and Dilator may be placed over-the-wire to the site of thrombus. The device macerates intra-lumen and wall adherent thrombus with a maceration wire. The macerated thrombus is removed from the vessel using an aspiration system. The aspiration of the clot can be performed simultaneous or independently.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
The provided text describes information about the Cleaner™ Plus 18F Thrombectomy System, a medical device, and the basis for its FDA clearance through a 510(k) premarket notification (K233909). The information primarily focuses on establishing substantial equivalence to a predicate device (K211798), rather than detailing a study that proves the device meets specific acceptance criteria in the context of an AI/ML medical device.
Therefore, many of the requested items related to AI/ML device study design (e.g., sample size for test sets, expert ground truth establishment, MRMC studies, standalone algorithm performance, training set details) are not applicable or cannot be extracted from this document, as it concerns a mechanical thrombectomy system, not an AI/ML-driven device.
However, I can provide the available information regarding the device's performance assessment and the basis for its clearance.
Acceptance Criteria and Device Performance (Non-AI/ML Medical Device)
The document does not present a table of numerical "acceptance criteria" in the same way one might find for an AI/ML model's performance metrics (e.g., AUC, sensitivity, specificity thresholds). Instead, the acceptance criteria for this medical device are qualitative, based on established industry standards, guidance, and the successful outcome of various performance and safety tests demonstrating substantial equivalence to a predicate device.
The "reported device performance" refers to the successful completion and meeting of predefined parameters for each test, indicating the device performs as intended and is as safe and effective as the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
As this is a mechanical device, not an AI/ML system, the "acceptance criteria" are the successful completion of various tests and the demonstration of substantial equivalence. The "reported device performance" refers to the outcomes of these tests.
Test Category / Characteristic | Acceptance Criteria (Qualitative) | Reported Device Performance (Outcome) |
---|---|---|
Substantial Equivalence | Demonstrates comparable indications for use, mechanism of action, materials, technological characteristics, principle of operation, design features, and sterilization process to the predicate device, introducing no additional clinical risk. | Established based on in vitro performance testing, similarities in indications for use, materials, technological characteristics, principle of operation, design features, and sterilization process with the predicate device. Confirmed by in-vitro and in-vivo tests. Conclusion: Cleaner Plus 18F Thrombectomy System is substantially equivalent to the predicate device and introduces no additional clinical risk. |
In-Vitro Performance Testing | Meet acceptance criteria based on requirements outlined in guidance and industry standards for each specific test. | Cleaner Plus 18F Thrombectomy System Performance: Visual, Dimensional, Simulated Use, Leak Testing. |
Handpiece Performance: Aspiration and Maceration Performance, Tensile and Torque, Kink Testing, Catheter Torque. | ||
Canister Performance: Simulated Use Maceration and Aspiration Performance. | ||
Maceration Wire Performance: Bond Strength, Resistance to Corrosion, Shape Retention, Flexing and Fracture, Kink Radius, Fatigue, Endurance, Torque, Tensile, Dimensional, Visual. | ||
Handpiece and Catheter Performance: Peel-Away Introducer Removal Force, Maceration Wire RPM, Wire to Coupler Tensile, Peel-Away Introducer Functionality, Tip Collapse. | ||
Other Tests: Particulates, Canister Performance (Visual, Functional, Tensile - joint strength), Luer Compliance, Pump Performance, IEC Compliance, Software Testing, Shipping Qualification. All tests met acceptance criteria. | ||
Biocompatibility | Compliant with ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration ( |
Ask a specific question about this device
(140 days)
The Cleaner™ Pro Thrombectomy System is indicated for the removal of fresh, soft thrombi and emboli from the vessels of the peripheral venous vasculature, and for the infusion of physician-specified fluids, including thrombolytics. The Cleaner™ Pro Thrombectomy System is not intended for use in the pulmonary vasculature for treating of pulmonary embolism.
The Cleaner Pro Thrombectomy System is a single use device in the removal of fresh, soft emboli and thrombi and for the infusion of physician-selected fluids through the side-port of the aspiration catheter.
The disposable system consists of: (1) the Aspiration Catheter with Dilator, (2) the Handpiece with Flushing Adapter, and (3) the external vacuum reservoir with pump known as the Aspiration Canister.
The Aspiration Catheter with Dilator may be placed over-the-wire to navigate the device to the target site. Once in the target site, to complete the system, the Aspiration Canister is connected to the handpiece. The dilator is removed, and the device is activated by the user to aspirate soft emboli and thrombi. The clot is aspirated from the distal portion of the device through the handpiece and then collected in the aspiration canister reservoir.
Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.
The provided text is a 510(k) Summary for the Cleaner™ Pro Thrombectomy System. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than proving the device meets specific acceptance criteria based on clinical outcomes or standalone AI algorithm performance.
Therefore, many of the requested details about acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and training sets are not present in this type of regulatory submission. This document describes the bench-top and animal performance testing to support the claim of substantial equivalence for a medical device, which is different from studies demonstrating efficacy of an AI algorithm or performance directly tied to a specific set of clinical acceptance criteria.
However, I can extract the information that is available:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a table of specific acceptance criteria with quantitative performance metrics for a clinical outcome study. Instead, it lists performance testing conducted to demonstrate substantial equivalence to the predicate devices. The acceptance criteria for these tests are generally that the device performs comparably to the predicate or meets established engineering/design specifications for safety and functionality.
Test Category | Specific Tests Performed (examples) | Acceptance Criteria (Implied) | Reported Performance (Implied) |
---|---|---|---|
In-Vitro (Bench-top) Performance Testing | Catheter - Dimensional, Visual, Radiopacity, Hemostasis Valve Leak, System Leak, Burst, Kink Radius, Tensile Break, Torsional Break, Aspiration Tip Collapse. Dilator - Dimensional, Leak, Burst, Functional, Visual, Tensile Break, Radiopacity. Aspiration Canister - Visual, Functional, Tubing Tensile Break, Weld Strength. Handpiece - Functional, Simulate Use. System - Functional, Vacuum Decay, Noise Level, Integrity, Visual, Simulated Use. Shipping Qualification, IEC-60601 Compliance, Software Validation. | The device must meet predefined engineering specifications and perform comparably to predicate devices for these physical and functional characteristics, ensuring safety and intended function. | "The below were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence." (General statement that tests were successful) |
Biocompatibility Testing | Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Irritation or Intracutaneous Reactivity (ISO 10993-10), Material Mediated Pyrogen (ISO 10993-11), Acute Systemic Toxicity (ISO 10993-11), Hemocompatibility (ISO10993-4) - In-vitro Blood Assay Complement Activation, SC5b-9 Heparinized Platelet and Leucocyte Counts, Partial Thromboplastin Time (PTT), ASTM Hemolysis Assay (Direct and Extract Methods). | Compliance with ISO 10993 standards for the specified biological endpoints, ensuring the device is biocompatible for its intended use. | "Biocompatibility is established for the Cleaner Pro Thrombectomy System according to ISO 10993-1:2020 as an external communicating, circulating blood with limited duration |
Ask a specific question about this device
(57 days)
The 10F Sheath and Dilator Set is indicated for introduction of therapeutic devices into the vasculature, excluding coronary and neuro vasculature.
The 10F Sheath and Dilator Set is designed for single use and intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature. The 10F Sheath and Dilator Set consists of a 10F reinforced Introducer Sheath and a matching 10F radiopaque Dilator.
The 10F Introducer sheath is a coil reinforced sheath with a Platinum Iridium radiopaque marker band. The sheath has a hemostasis valve and a side port with stopcock. The 10F dilator is made of radiopaque, non-reinforced material with an atraumatic tip to minimize blood loss. The device does not include any coating.
The provided document describes a 510(k) premarket notification for the "10F Sheath and Dilator Set" and establishes its substantial equivalence to a predicate device based on in vitro performance testing and technological characteristics. It does not include information about an AI-powered device or a study proving that an AI device meets acceptance criteria. Thus, I cannot populate the requested table or answer the questions related to AI device performance.
However, I can extract the acceptance criteria and the study performed for the 10F Sheath and Dilator Set (a non-AI medical device).
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Radiopacity | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Visual | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Dimensional | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Simulative Use | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Leakage (Sheath and Dilator) | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Kink Test (System) | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Tensile (Sheath and Dilator) | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Luer testing | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Shipping Qualification | (Implicit performance standard) | Met acceptance criteria demonstrating substantial equivalence. |
Biocompatibility | **According to ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration |
Ask a specific question about this device
(265 days)
The L-Cath Single and Dual Lumen Catheters and the L-Cath Midline Catheters are indicated for short or long term peripheral access to the central venous system for the administrations, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
The L-Cath™ Single and Dual Lumen and L-Cath™ Midline Catheters are sterile, single use devices, sold as standalone devices which includes the following accessories:
• Trim Tool
• Tape measure
• Stylet (include in the final catheter assembly as applicable)
• Polyurethane catheter
These catheters are also sold as a component in a Basic Kit (catheter and introducer packaged together) which includes the following:
- BD Introsyte Autoguard Introducer or Splittable Introducer ●
- Stylet (include in the final catheter assembly as applicable)
- Trim Tool
- Tourniquet
- Tape measure
The provided text describes a 510(k) premarket notification for Argon Medical Devices' L-Cath™ Single and Dual Lumen Catheters and L-Cath™ Midline Catheters. This document focuses on demonstrating substantial equivalence to an existing predicate device (BD L-Cath™ Midline Catheter, BD Cath™ Single Lumen, BD Cath™ Dual Lumen Peripherally Inserted Central Catheters PICC, K091670), rather than proving the device meets clinical acceptance criteria based on studies involving AI or human readers.
Therefore, many of the requested criteria in your prompt (e.g., acceptance criteria for device performance in terms of AI metrics, sample sizes for test/training sets in an AI study, number of experts, adjudication methods, MRMC studies, standalone AI performance, ground truth establishment) are not applicable to this type of regulatory submission. This document details the engineering and biocompatibility aspects of the device to show it functions safely and effectively similarly to devices already on the market.
However, I can extract the relevant information from the provided text regarding the performance testing conducted to support the substantial equivalence claim.
Here's a breakdown of the requested information based on the provided text, highlighting what is applicable and what is not:
Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria
This submission is a 510(k) for substantial equivalence, not a clinical study proving performance against AI-related metrics. The "acceptance criteria" here refer to engineering, biocompatibility, and sterilization performance standards, demonstrating the modified device is as safe and effective as its predicate.
1. Table of Acceptance Criteria and Reported Device Performance
The document states: "The subject device met all the predetermined acceptance criteria derived from the above listed reference standards and internal test protocols and demonstrated substantially equivalent performance compared to the predicate device." While specific numerical acceptance criteria (e.g., "pressure burst must exceed X psi") are not explicitly detailed in a table, the document lists the types of performance tests conducted and confirms successful completion.
Acceptance Criteria Category (Implied) | Performance Test Conducted (Reported) | Device Performance (Reported) |
---|---|---|
Mechanical/Physical Performance | Catheter – Joint Leak Test | Met acceptance criteria (Substantially equivalent to predicate) |
Catheter – Dimensional | Met acceptance criteria (Substantially equivalent to predicate) | |
Catheter – Pressure Burst | Met acceptance criteria (Substantially equivalent to predicate) | |
Catheter – Durometer | Met acceptance criteria (Substantially equivalent to predicate) | |
Catheter – Radiopacity | Met acceptance criteria (Substantially equivalent to predicate) | |
Stylet – Retaining force | Met acceptance criteria (Substantially equivalent to predicate) | |
Luer Functional Mechanical Bonds – Pull force | Met acceptance criteria (Substantially equivalent to predicate) | |
Particulates | Met acceptance criteria (Substantially equivalent to predicate) | |
Material Durability | Met acceptance criteria (Substantially equivalent to predicate) | |
Supplemental Testing for Design Change (Luer Lock material change) | ||
Luer Integrity (Functional) Report 2023-042-RPT | Met acceptance criteria | |
ISO 80369-7 Compliance | Met acceptance criteria | |
Luer Integrity (Functional); Functional Leak Test Report 2023-134-RPT | Met acceptance criteria | |
Biocompatibility | Pyrogenicity (ISO 10993-11) | Met acceptance criteria (Substantially equivalent to predicate) |
Cytotoxicity (ISO 10993-5) | Met acceptance criteria (Substantially equivalent to predicate) | |
Hemocompatibility (ISO 10993-4) | Met acceptance criteria (Substantially equivalent to predicate) | |
Irritation/Sensitization (ISO 10993-10) | Met acceptance criteria (Substantially equivalent to predicate) | |
Systemic Toxicity (ISO 10993-11) | Met acceptance criteria (Substantially equivalent to predicate) | |
Ethylene oxide Residuals (ISO 10993-7) | Met acceptance criteria (Substantially equivalent to predicate) | |
Local effects after implantation (ISO 10993-7) | Met acceptance criteria (Substantially equivalent to predicate) | |
Supplemental Testing for Material Change (Luer lock material) | ||
Systemic Toxicity (10993-17) Report 2022-026-RPT | Met acceptance criteria | |
Genotoxicity (10993-3) | Met acceptance criteria | |
Sterilization | 100% EtO | Met acceptance criteria (SAL 10-6) |
ISO 11135:2014 & AAMI TIR28:2016 compliance | Met acceptance criteria |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated in terms of number of units tested for each criterion, but implied to be sufficient for demonstrating compliance with ISO standards and internal protocols.
- Data Provenance: The tests were conducted internally by Argon Medical Devices, as indicated by the company providing the summary. The regulatory context implies these are laboratory tests, not clinical data from patients. They are likely prospective bench and lab tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This is a medical device 510(k) submission based on engineering and biocompatibility testing, not an AI performance study requiring human expert ground truth for image or diagnostic interpretation. The "ground truth" here is established by validated test methods and reference standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. See point 3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI-assisted diagnostic device. Clinical testing (human studies) was explicitly stated as "not required for the determination of substantial equivalence."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. See point 5.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Engineering and Biocompatibility Test Standards: The "ground truth" for the device's performance is established by recognized international standards (e.g., ISO 10993 series for biocompatibility, ISO 80369-7 for luer connectors, ISO 11135 for sterilization) and internal validated test protocols designed to assess the device's physical properties and safety characteristics.
8. The sample size for the training set
- Not Applicable. This document describes a medical device, not an AI algorithm. There is no "training set."
9. How the ground truth for the training set was established
- Not Applicable. See point 8.
Summary of the Study Proving Acceptance:
The study presented in this 510(k) summary is a series of bench testing, laboratory analyses (biocompatibility), and sterilization validations designed to demonstrate that the modified L-Cath™ catheters are substantially equivalent to the previously cleared predicate devices. The primary change was the material used in the Luer Lock assembly.
The "study" involved:
- Performance Testing: Assessing mechanical characteristics like leak integrity, dimensions, pressure burst, durometer, radiopacity, retaining force, pull force, particulates, and material durability. These tests were stated to be the "SAME as K091670" (the predicate's testing) with additional supplemental testing specific to the Luer Lock material change (Luer Integrity, ISO 80369-7 Compliance).
- Biological Comparison (Biocompatibility Testing): Evaluation of potential biological risks including pyrogenicity, cytotoxicity, hemocompatibility, irritation/sensitization, systemic toxicity, ethylene oxide residuals, and local effects after implantation. Supplemental testing (Systemic Toxicity, Genotoxicity) was performed due to the Luer Lock material change.
- Sterilization Validation: Confirmation that the device can be consistently sterilized to meet a sterility assurance level (SAL) of 10-6, in accordance with ISO 11135:2014 and AAMI TIR28:2016.
The conclusion is that the modified device "met all the predetermined acceptance criteria derived from the above listed reference standards and internal test protocols and demonstrated substantially equivalent performance compared to the predicate device." This means the tests confirmed the new device performs within acceptable limits, similar to the predicate, and does not raise new questions of safety or effectiveness.
Ask a specific question about this device
(24 days)
Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are intended for the percutaneous removal of retrievable inferior vena cava (IVC) filters that are no longer medically required, via jugular approach.
The Single-Loop Snare Retrieval Kit and Triple-Loop Snare Retrieval Kit are single use devices. The disposable system consists of: [A] 9F (ID) Inner Sheath, [B] 11F (ID) Outer Sheath, [C] 8F (OD) Dilator, [D] Hemostasis Valve with Sideport, [E] High Pressure Stopcock, [F] 7F [2.4mm] (OD) x 76cm Snare Catheter with Tuohy-Borst Y-Port Adapter, 20mm x 93cm Single-Loop Snare (fully expanded) or 30mm x 93cm Triple-Loop Snare (fully expanded) with Torque Handle. The snares have radiopaque loops and are preloaded in the snare catheter. The snare catheter, inner sheath, and outer sheath have a radiopaque marker band at the distal tip for enhanced fluoroscopic visualization.
The provided text describes a 510(k) premarket notification for a medical device, specifically retrieval kits for inferior vena cava (IVC) filters. The submission focuses on a modification to an existing device.
Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a 510(k) submission for a modification to an existing device, the "acceptance criteria" are not explicitly stated as numerical targets in a table. Instead, the acceptance criteria are implicit in meeting the performance standards of the predicate device and ensuring the modification doesn't negatively impact safety or effectiveness. The reported device performance is that all tests met their respective acceptance criteria.
Acceptance Criteria (Implicit from Predicate & Safety) | Reported Device Performance |
---|---|
Material Properties (Outer Sheath Modification): | |
Outer Sheath - Visual | Met acceptance criteria |
Outer Sheath - Dimensional | Met acceptance criteria |
Outer Sheath - Tensile Strength | Met acceptance criteria |
Outer Sheath - Liquid Leakage | Met acceptance criteria |
Outer Sheath - Simulated Use | Met acceptance criteria |
Outer Sheath - Delamination | Met acceptance criteria |
Biocompatibility (for modified material): | |
Cytotoxicity (ISO 10993-5) | Met acceptance criteria |
Sensitization (ISO 10993-10) | Met acceptance criteria |
Irritation or Intracutaneous Reactivity (ISO 10993-23) | Met acceptance criteria |
Material Mediated Pyrogen (ISO 10993-11) | Met acceptance criteria |
Acute Systemic Toxicity (ISO 10993-11) | Met acceptance criteria |
Hemocompatibility (ISO 10993-4): | Met acceptance criteria |
- ASTM Hemolysis, Direct and Extract | Met acceptance criteria |
- Complement Activation, SC5b-9 | Met acceptance criteria |
- Partial Thromboplastin (PTT) | Met acceptance criteria |
- Platelet and Leukocyte Count | Met acceptance criteria |
- In vitro Blood Loop | Met acceptance criteria |
Leveraged Non-Clinical Testing (from predicate K191758): | |
Visual/Dimensional (Inner Sheath, Dilator, Delivery Catheter, Single-Loop Snare & Triple-Loop Snare) | Met acceptance criteria |
Leak Test (Inner Sheath, Dilator and Delivery Catheter) | Met acceptance criteria |
Burst Test | Met acceptance criteria |
Pull Test/Tensile Strength (Inner Sheath, Dilator, Delivery Catheter, Single-Loop Snare & Triple-Loop Snare) | Met acceptance criteria |
Corrosion Resistance (Single-Loop Snare & Triple-Loop Snare) | Met acceptance criteria |
Torque Response (Snare Assembly) | Met acceptance criteria |
Radiopacity | Met acceptance criteria |
Luer Testing | Met acceptance criteria |
Contrast Medium Injection | Met acceptance criteria |
High Pressure Stopcock Testing | Met acceptance criteria |
Flexural Modulus and Tip Flexibility Testing | Met acceptance criteria |
Radial Force Testing | Met acceptance criteria |
Design Validation Testing and Summative Usability Testing | Met acceptance criteria |
The document explicitly states: "Test results demonstrate that all acceptance criteria were met; therefore, the device meets the established product specifications."
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes used for each individual non-clinical test. It only lists the types of tests performed. The data provenance is not explicitly mentioned in terms of country of origin, but it is implied to be from the manufacturer's internal testing. The tests are non-clinical (bench testing, biocompatibility) rather than human subject studies, so the terms "retrospective or prospective" are not applicable in this context.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this submission. The tests performed are non-clinical (mechanical, materials, biocompatibility) and do not involve expert interpretation or ground truth establishment in the way clinical studies with AI algorithms would. The "ground truth" for these tests is based on objective measurements against established engineering and biocompatibility standards (e.g., ISO and ASTM standards).
4. Adjudication Method for the Test Set
This is not applicable as the tests are non-clinical and do not involve human interpretation or adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This submission is for a physical medical device (retrieval kit) with a material modification, not an AI-powered diagnostic or interpretive device. The document explicitly states: "Clinical testing was not required for the determination of substantial equivalence."
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
A standalone performance study was not done. This is not an AI device.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical tests is based on:
- Engineering specifications and standards: For mechanical and dimensional tests (e.g., tensile strength, leak test, burst test, radiopacity, Luer testing, flexural modulus, radial force).
- Biocompatibility standards: For biocompatibility tests (e.g., ISO 10993 series for cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity, hemocompatibility).
- Simulated use conditions: For tests like simulated use and design validation.
8. The Sample Size for the Training Set
This is not applicable as this is not an AI device and therefore has no "training set."
9. How the Ground Truth for the Training Set was Established
This is not applicable as this is not an AI device and therefore has no "training set" or associated ground truth establishment process.
Ask a specific question about this device
(76 days)
The Kodiak™ Dual Port Coaxial Introducer Kit is indicated to introduce therapeutic or diagnostic devices into the vasculature.
The Kodiak™ Dual Port Coaxial Introducer Kit is intended to introduce therapeutic or diagnostic devices into the vasculature. The Kit Model 381416000 contains a coaxial 16F and 14F braided reinforced sheaths with a fluoropolymer coating, an embedded radiopaque marker band, and an over molded hub assembly containing two hemostasis valves and side port with stopcock. It also contains a removable Y-connector with two 7F (ID) ports with hemostasis valves, a 14F dilator, a 16ga blunt flushing needle, and a high-pressure Luer adapter. After percutaneous access is established and a working guidewire is placed, the introducer system is inserted over the guidewire and advanced under imaging. Therapeutic or diagnostic devices are inserted through the system. The Kodiak™ Dual Port Coaxial Introducer Kit is a single use device.
The provided text is a 510(k) Summary for the Kodiak™ Dual Port Coaxial Introducer Kit, which is a medical device for introducing therapeutic or diagnostic devices into the vasculature. The document details the device's characteristics and performance testing to demonstrate substantial equivalence to a predicate device.
It is important to note that this document does not contain information about an AI/ML powered device. Therefore, many of the requested criteria regarding AI/ML studies (like multi-reader multi-case studies, standalone algorithm performance, training set details, ground truth establishment methods for AI) are not applicable.
Below is the information that can be extracted relevant to the performance of the Kodiak™ Dual Port Coaxial Introducer Kit as a non-AI medical device:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Performance Testing | System Leak | Not explicitly stated in the provided text, but implied to be compliance with applicable standards/protocols designed to ensure product safety and clinical function. | "shown to meet the acceptance criteria" |
Tensile Strength | Not explicitly stated | "shown to meet the acceptance criteria" | |
Tip Strength | Not explicitly stated | "shown to meet the acceptance criteria" | |
Radiopacity | Not explicitly stated, but would likely involve the device being sufficiently visible under fluoroscopy. | "shown to meet the acceptance criteria" | |
Kink Testing | Not explicitly stated, but would likely involve maintaining lumen patency and structural integrity when bent to a certain degree. | "shown to meet the acceptance criteria" | |
Dimensional | Not explicitly stated, but would refer to conformity to specified measurements (e.g., length, diameter) as detailed in the device description. | "shown to meet the acceptance criteria" | |
Functional Fit | Not explicitly stated, but would relate to the proper interaction and assembly of components (e.g., sheath with guidewire, dilator, Y-connector). | "shown to meet the acceptance criteria" | |
Simulated Use | Not explicitly stated, but would involve simulating real-world clinical scenarios to ensure the device performs as intended without failure. | "shown to meet the acceptance criteria" | |
Luer Testing | Not explicitly stated, but refers to the integrity and leak-proof nature of Luer connections according to ISO standards. | "shown to meet the acceptance criteria" | |
Power Injection Testing | Not explicitly stated, but would involve testing the device's ability to withstand pressures from power injection without rupture or leakage. | "shown to meet the acceptance criteria" | |
Tip Buckling | Not explicitly stated, but would relate to the resistan ce of the introducer tip to buckle or deform during insertion. | "shown to meet the acceptance criteria" | |
Particulate | Not explicitly stated, but would involve testing for the absence of unacceptable levels of particulates that could be introduced into the patient. | "shown to meet the acceptance criteria" | |
Packaging Integrity | Not explicitly stated, but would involve ensuring the sterile barrier and physical protection of the device until point of use. | "shown to meet the acceptance criteria" | |
Biocompatibility Testing | Cytotoxicity (ISO 10993-5) | Not explicitly stated, but would be based on ISO 10993-5 guidelines for cytotoxicity testing. | "shown to meet the acceptance criteria" |
Sensitization (ISO10993-10) | Not explicitly stated, but based on ISO 10993-10 guidelines. | "shown to meet the acceptance criteria" | |
Irritation/Intracutaneous Reactivity (ISO10993-23) | Not explicitly stated, but based on ISO 10993-23 guidelines (formerly ISO 10993-10). | "shown to meet the acceptance criteria" | |
Acute Systemic Toxicity (ISO10993-11) | Not explicitly stated, but based on ISO 10993-11 guidelines. | "shown to meet the acceptance criteria" | |
Pyrogenicity (ISO 10993-11) | Not explicitly stated, but based on ISO 10993-11 guidelines. | "shown to meet the acceptance criteria" | |
Hemolysis Direct/Indirect (ISO 10993-4) | Not explicitly stated, but based on ISO 10993-4 guidelines. | "shown to meet the acceptance criteria" | |
Thromboresistance (ISO 10993-4) | Not explicitly stated, but based on ISO 10993-4 guidelines, ensuring the device does not promote thrombus formation. | "shown to meet the acceptance criteria" | |
Partial Thromboplastin Time (ISO 10993-4) | Not explicitly stated, but based on ISO 10993-4 guidelines, assessing interactions with the coagulation system. | "shown to meet the acceptance criteria" | |
Complement Activation (ISO 10993-4) | Not explicitly stated, but based on ISO 10993-4 guidelines, assessing interactions with the complement system. | "shown to meet the acceptance criteria" | |
Platelet and Leukocyte Count (ISO 10993-4) | Not explicitly stated, but based on ISO 10993-4 guidelines, assessing effects on blood cellular components. | "shown to meet the acceptance criteria" |
The document states, "A series of testing was conducted in accordance with protocols based on requirements outlined in guidance and industry standards and the testing below was shown to meet the acceptance criteria that was determined to demonstrate substantial equivalence."
Study Information (for non-AI device)
-
Sample size used for the test set and the data provenance:
The document does not specify the exact sample sizes (number of units tested) for each individual performance or biocompatibility test. It only states that "a series of testing was conducted." The data provenance is internal testing performed by the device manufacturer, Argon Medical Devices, Inc. The data would be prospective for the purpose of this 510(k) submission, as these are tests specifically conducted for this new device. -
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is a non-AI medical device. The "ground truth" for its performance is established by objective physical and chemical testing against engineering specifications and international standards, not by expert interpretation of data. -
Adjudication method for the test set:
Not applicable for this type of device testing. Results are typically quantitative measurements compared against predefined acceptance limits. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a non-AI medical device. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a non-AI medical device. -
The type of ground truth used:
For performance testing, the "ground truth" refers to the objective physical and chemical properties and functional capabilities of the device as measured by standardized test methods against established specifications and international standards (e.g., ISO 10993 for biocompatibility, Luer connector standards, internal engineering specifications). -
The sample size for the training set:
Not applicable. This is a non-AI medical device, so there is no "training set." -
How the ground truth for the training set was established:
Not applicable. This is a non-AI medical device, so there is no "training set" or corresponding ground truth establishment process in the context of AI/ML.
Ask a specific question about this device
(181 days)
The Cleaner Plus™ Thrombectomy System is indicated for mechanical de-clotting, aspiration, and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral venous vasculature.
The Cleaner Plus™ Thrombectomy System is a single use device used to provide thrombectomy in the peripheral venous vasculature. The device provides additional features, such as aspiration and over-the wire device placement. The disposable system consists of: (1) the Aspiration Catheter & Dilator, (2) the Handpiece that includes system controls, and an integrated Maceration Wire, and a Peel-Away Introducer and (3) the Aspiration Canister. The Aspiration Catheter with Dilator may be placed over-the-wire to navigate the device to the therapeutic site. The dilator and guidewire are removed, and the Maceration Wire, using the Peel-Away introducer is advanced through the hemostasis valve of the Aspiration Catheter to the therapeutic site and connected to the handpiece. To complete the system, the provided Aspiration Canister is connected to the handpiece to provide aspiration. The Handpiece provides controls to turn on/off maceration and/or the application of suction. Mechanical thrombectomy is achieved by rotating a flexible stainless-steel maceration wire powered by a motor inside the handpiece. The aspiration source is provided to aspirate macerated clot from the distal portion of the device through the handpiece and captures the macerated clot in the Aspiration Canister reservoir. The Aspiration Canister includes a switch to initiate the pump, and LEDs that indicate the level of the vacuum.
The provided text describes a 510(k) submission for a device modification to the Cleaner Plus™ Thrombectomy System. This is a medical device, not an AI/ML algorithm. Therefore, many of the requested criteria, such as "Multi-reader multi-case (MRMC) comparative effectiveness study," "standalone (i.e., algorithm-only) performance," and "ground truth for the test and training set," are not applicable.
The submission focuses on demonstrating the substantial equivalence of a modified device to a predicate device through non-clinical performance testing.
Here's the relevant information extracted from the document:
1. Table of Acceptance Criteria and Reported Device Performance
The document states that "Test results demonstrate that all acceptance criteria were met." However, the specific acceptance criteria values are not detailed in the provided text. The types of tests performed indicate the areas for which acceptance criteria would have been established.
Acceptance Criteria Category (implied from tests) | Reported Device Performance |
---|---|
Corrosion Resistance | Met acceptance criteria |
Tensile Break | Met acceptance criteria |
Simulated Use (Handpiece performance with Helical component) | Met acceptance criteria |
Simulated Use (Aspiration Performance) | Met acceptance criteria |
Torque Strength (Helical Component to Wire) | Met acceptance criteria |
Wire Fatigue | Met acceptance criteria |
Particulates | Met acceptance criteria |
Biocompatibility (Cytotoxicity - MEM Elution) | Met acceptance criteria |
Biocompatibility (Sensitization – Guinea Pig Maximization) | Met acceptance criteria |
Biocompatibility (Irritation - Intracutaneous Reactivity) | Met acceptance criteria |
Biocompatibility (Systemic Toxicity - Material Mediated Pyrogen) | Met acceptance criteria |
Biocompatibility (Systemic Toxicity - Acute Systemic Toxicity) | Met acceptance criteria |
Biocompatibility (Hemocompatibility – Hemolysis Indirect) | Met acceptance criteria |
2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes for the individual tests. The tests were non-clinical (laboratory/bench testing), so data provenance from countries or whether it was retrospective/prospective is not applicable in the typical sense.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as this is a medical device performance study, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are typically for clinical studies or studies involving expert interpretation, not for bench testing of device components.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device, not an AI/ML system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical device, not an AI/ML system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The documentation did not specify a "ground truth" in the context of clinical or diagnostic outcomes. For performance testing of a physical device, the "ground truth" would be established physical measurements and material properties relevant to the device's function and safety (e.g., tensile strength, torque, aspiration volume, biocompatibility assays).
8. The sample size for the training set
Not applicable. This is a medical device, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is a medical device, not an AI/ML algorithm that requires a training set with established ground truth.
Summary of Study Type:
The study described is a non-clinical verification testing of a medical device modification to support substantial equivalence to a predicate device. The tests included various engineering performance tests (e.g., corrosion, tensile, torque, fatigue, simulated use) and biocompatibility assessments. Clinical and animal testing were not required for this determination of substantial equivalence.
Ask a specific question about this device
(274 days)
The Semi-Automatic Biopsy Instrument and associated Co-Axial Introducer Needle are used to obtain samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, pancreas, and other masses.
The Semi-Automatic Biopsy Instrument is used to obtain soft tissue biopsies. It is composed of a spring-loaded biopsy needle fitted into a plastic handle permitting single handed specimen collection. The position of the needle may be visualized by x-ray, CT, or ultrasound. The needle has numerically ordered centimeter markings to facilitate precise depth placement. The adjustable instrument allows for a specimen notch size of 10mm or 20mm, providing clinical flexibility. The device has an adjustable co-axial adapter built into the device for either the 10mm or 20mm notch settings. The stylet travels up to 27mm from when the device is in a fully charged position (prepared for a collection with the 20 mm notch size) to when the stylet is fully extended. The needle has an echogenic tip. The semi-automatic biopsy instrument is available in several needle gauge sizes and lengths. The plunger is color coded according to the various gauge sizes, e.g., green=14-gauge, purple=16-gauge, pink=18-gauge, and yellow=20-gauge.
The provided document is a 510(k) summary for a medical device (Semi-Automatic Biopsy Instrument). It lacks the specific details required to answer your request about acceptance criteria and a study proving that an AI/ML-based device meets those criteria.
The document describes a traditional medical device (biopsy instrument) and its performance testing, which are primarily engineering and material tests, not clinical studies involving AI performance, human readers, or ground truth establishment relevant to AI.
Therefore, I cannot provide the requested information based on the provided text. The document does not contain:
- A table of acceptance criteria and reported device performance related to AI/ML.
- Sample sizes for test sets in an AI context.
- Data provenance for AI/ML training/test sets.
- Number of experts and their qualifications for ground truth in an AI context.
- Adjudication method for AI test sets.
- MRMC comparative effectiveness study details.
- Standalone AI performance (algorithm only).
- Type of ground truth (expert consensus, pathology, outcomes data) for AI/ML.
- Sample size for training set for AI/ML.
- How ground truth for the training set was established for AI/ML.
The "Performance Testing" section in the document refers to physical and material tests for the biopsy instrument itself (e.g., Dimensional, Peak Tensile Strength, Simulated Use, Radiopacity, Echogenicity, Corrosion Resistance). This is entirely different from the kind of performance testing relevant to AI/ML medical devices.
Ask a specific question about this device
Page 1 of 3