Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K172494
    Manufacturer
    Date Cleared
    2017-12-15

    (119 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Angioslide, Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and removal of embolic material (e.g. debris, thrombus) during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The Angioslide PROTEUSTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9).

    One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.

    The shaft (4) comprises the outer shaft (6) and the inner shaft (7). The distal end of the balloon (A) is connected to the inner shaft and the proximal end of the balloon (B) is connected to the outer shaft. The inner shaft is connected to the proximal hub (10) which is connected to the pulling knob (1) and the outer shaft is connected to the handle grip (3). The pulling knob lock (11) locks the handle grip and the pulling knob together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by pressing on pulling knob lock (11) and pulling the pulling knob away from the handle (1). The inward-folding of the balloon forms a cavity and allows for collection and removal of embolic material.

    AI/ML Overview

    The provided document describes the Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature. It is a medical device designed for peripheral transluminal angioplasty and for capturing embolic material during angioplasty in various peripheral arteries. The document is a 510(k) summary, demonstrating substantial equivalence to a predicate device (K133043).

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comprehensive table of non-clinical bench tests. All tests passed, indicating the device met its acceptance criteria. Below is a summarized version of the table provided in the document:

    TestAcceptance Criteria (Example)Reported Device Performance
    Visual inspection - external surfaceFree from extraneous matter and surface defects (
    Ask a Question

    Ask a specific question about this device

    K Number
    K133043
    Manufacturer
    Date Cleared
    2013-12-06

    (71 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ANGIOSLIDE, LTD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and removal of embolic material (e.g. debris, thrombus) during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (S) located near the distal atraumatic soft tip (9).

    One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.

    The shaft (4) comprises the outer shaft (6) and the inner shaft (7). The distal end of the balloon (A) is connected to the inner shaft and the proximal end of the balloon (B) is connected to the outer shaft. The inner shaft is connected to the proximal hub (10) which is connected to the pulling knob (1) and the outer shaft is connected to the handle grip (3). The pulling knob lock ( 1 ) locks the handle grip and the pulling knob together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by pressing on pulling knob lock (11) and pulling the pulling knob away from the handle (1). The inward-folding of the balloon forms a cavity and allows for collection and removal of embolic material.

    The balloon size and diameter are printed on the strain relief (12). Refer also to the package label for information about catheter length, balloon nominal and rated burst pressure, balloon size, balloon compliance, guidewire compatibility and sheath compatibility.

    AI/ML Overview

    The PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature underwent non-clinical bench testing to demonstrate its performance and safety, particularly regarding its use within a stent environment.

    1. Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Post-Dilatation, Minimum Balloon Burst Pressure (RBP)RBP ≥ 14atm (3x100 only), 12atm for all other sizes with 95% Confidence, 99.9% ReliabilityPASS
    Post-Dilatation, Balloon Fatigue (Repeated Inflation/Deflations)Inflation/Deflation Cycles ≥ 10 at 14atm (for 3x100 only), 12atm for all other sizes. No leakage, rupture, and/or herniation. Up to max 40 cycles with 95% Confidence, 90% ReliabilityPASS
    Post Dilatation Balloon Inflation/Deflation TestingInflation time: ≤ 14.0 sec, Deflation time: ≤ 30.6 sec. No leakage upon inflation with 90% Confidence, 90% ReliabilityPASS
    Post Dilatation, Simulated Use in Tortuous Anatomy Model - Guide Wire CompatibilityCatheter can be mounted over a 0.014" guide wire (3x100 only), 0.035" guide wire for all other sizes with 90% Confidence, 90% ReliabilityPASS
    Post Dilatation Simulated Use in Tortuous Anatomy Model - Introducer Sheath CompatibilityCompletely folded balloon passes through identified Introducer Sheath (5F, 6F, 7F) at the end of procedure with 90% Confidence, 90% ReliabilityPASS
    Post Dilatation Simulated Use in Tortuous Anatomy Model – Kink ResistanceNo permanent deformations (kinks) are present once removed from the tortuous anatomy model with 90% Confidence, 90% ReliabilityPASS
    Post-Dilatation Capture EfficiencyN/A - Characterization onlyPASS (Overall CE consistent with previously tested device sizes without multiple overlapping stents, despite slight increase in particulate generation with stents.)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document states that "Sample sizes used for Design Verification and Validation testing were based on required confidence / reliability levels as a result of risk analysis performed for the PROTEUS™ PTA Balloon Catheter, or per recommendations within the FDA Guidance "Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems." Specifically, the number of samples utilized for each test depended on whether the data to be collected was variable data or attribute data in nature." Specific numerical sample sizes for each test are not explicitly provided in the text.
    • Data Provenance: The tests were non-clinical in vitro bench tests conducted by Angioslide, Ltd. No patient data (retrospective or prospective) from specific countries was used for these tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    Not applicable. The study involved non-clinical bench testing, not human expert evaluation of clinical data to establish ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. The study involved non-clinical bench testing, not human expert review with adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No. This was a non-clinical bench test. No human readers or AI assistance were involved in evaluating efficacy.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical medical device (catheter), not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    In a non-clinical bench study for a physical device, "ground truth" is established by the specifications and measurable outcomes of the device's physical performance against predefined engineering and regulatory standards (e.g., burst pressure, inflation/deflation times, absence of leakage, ability to pass through a tortuous model). The "ground truth" is essentially the expected physical behavior and integrity of the device under simulated conditions.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical medical device undergoing performance testing, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As this is not an AI model, there is no training set or ground truth for a training set in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120164
    Manufacturer
    Date Cleared
    2012-06-08

    (141 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ANGIOSLIDE, LTD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The Angioslide PROTEUSTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9). One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure. The shaft (4) comprises the outer shaft (6) and the inner shaft (7). The distal end of the balloon (A) is connected to the inner shaft and the proximal end of the balloon (B) is connected to the outer shaft. The inner shaft is connected to the proximal hub (10) which is connected to the pulling knob (1) and the outer shaft is connected to the handle grip (3). The pulling knob lock (11) locks the handle grip and the pulling knob together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by pressing on pulling knob lock ( 1 l ) and pulling the pulling knob away from the handle (1). The embolic capture feature involves a single-use suction mechanism that works through inward folding of the balloon, which creates negative pressure within the capture cavity for debris capture and removal. The reduced pressure in the capture cavity causes some of the particles that are released during the procedure to flow into the cavity for containment and removal. The balloon size and diameter are printed on the strain relief (12). Refer also to the package label for information about catheter length, balloon nominal and rated burst pressure, balloon size, balloon compliance, guidewire compatibility and sheath compatibility.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the PROTEUS™ PTA Catheter with Embolic Capture Feature, based on the provided text:

    Acceptance Criteria and Device Performance

    The device underwent extensive non-clinical bench testing, sterilization, packaging, and shelf-life testing to demonstrate conformance. The table below summarizes the acceptance criteria and reported device performance for these non-clinical tests.

    Note: The provided document is a 510(k) summary for a submission that claims substantial equivalence to a predicate device (K111750), implying that the new device has identical technological characteristics for both PTA balloon function and embolic capture. Therefore, the "study" described here is a suite of design verification and validation tests rather than a comparative clinical trial.

    Test DescriptionAcceptance CriteriaReported Device Performance
    Visual Inspection - External SurfaceExternal surface of catheter effective length, including the distal end, is free from extraneous matter and surface defects (90% Confidence, 90% Reliability)PASS
    Visual Inspection - Distal TipDistal tip is smooth, rounded, tapered, or similarly finished. (90% Confidence, 90% Reliability)PASS
    Dimensional Inspection - Distal Bond O.D.In Tolerance (90% Confidence, 90% Reliability)PASS
    Dimensional Inspection - Soft Tip LengthIn Tolerance (90% Confidence, 90% Reliability)PASS
    Dimensional Inspection - Guidewire Inner LumenIn Tolerance (90% Confidence, 90% Reliability)PASS
    Dimensional Inspection - Wrapped Balloon O.D.The balloon must pass aperture
    Ask a Question

    Ask a specific question about this device

    K Number
    K120805
    Manufacturer
    Date Cleared
    2012-04-13

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ANGIOSLIDE, LTD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angroplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9).

    One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.

    AI/ML Overview

    Acceptance Criteria and Device Performance for PROTEUS™ PTA Catheter with Embolic Capture Feature

    This submission describes the non-clinical testing performed to demonstrate the substantial equivalence of the modified PROTEUS™ PTA Catheter with Embolic Capture Feature. The testing followed Angioslide's Risk Analysis, applicable FDA Guidance documents, and ISO standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual inspection - external surfaceExternal surface of catheter effective length, including the distal end, is free from extraneous matter and surface defects (
    Ask a Question

    Ask a specific question about this device

    K Number
    K111750
    Manufacturer
    Date Cleared
    2011-07-21

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ANGIOSLIDE, LTD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The PROTEUS™ PTA Balloon Catheter with Embolic Capture is an over the wire coaxial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip.

    The PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature consists of a foldable balloon with radiopaque markers which aid in positioning the balloon in the artery during the angioplasty procedure, the co-axial shaft comprising the inner shaft and the outer shaft, and the handle comprising the handle enclosure, the T-connector, the inflation port, the pulling rod with the guidewire port, and the pulling rod locks.

    Following embolic capture, the user can opt to use the PROTEUS™ Particle Visualization Kit (PVK) to facilitate the visualization of the embolic material captured by the PROTEUS™ PTA Balloon Catheter.

    AI/ML Overview

    The provided text describes a Special 510(k) submission for the PROTEUS™ PTA Catheter with Embolic Capture Feature. This submission is for modifications to an already cleared device, specifically the addition of a Particle Visualization Kit (PVK) accessory.

    Here's an analysis of the provided information regarding acceptance criteria and the study:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state specific quantitative acceptance criteria or provide a table of reported device performance in terms of metrics like sensitivity, specificity, accuracy, etc. This is common for Special 510(k) submissions focusing on modifications, especially when the modification is an accessory and the core function (embolic capture) of the device remains identical to a predicate.

    The primary "acceptance criterion" implied throughout is substantial equivalence to the predicate device (Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature K090364) in terms of indications for use, technological characteristics, and performance characteristics.

    The document states:

    • "The balloon technological characteristics of the modified version of PROTEUS™ PTA Balloon Catheter are substantially equivalent to those of the unmodified version of the device."
    • "The modified PROTEUSTM PTA Balloon Catheter overall length, catheter sheath sizing, balloon diameter, balloon length, balloon nominal pressure, balloon rated burst pressure and end hole diameter are identical to the unmodified version of the device."
    • "The embolic capture technological characteristics of the modified PROTEUSTM PTA Balloon Catheter are identical to those of the unmodified version of the device."

    The only specific testing mentioned relates to the new accessory:

    • "Verification and validation testing of the PROTEUS™ Particle Visualization Kit included simulated use testing."

    Without specific metrics or quantitative data from this simulated use testing, a formal table cannot be constructed.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document mentions "simulated use testing" for the Particle Visualization Kit but does not specify the sample size (e.g., number of simulated procedures, number of particles visualized).
    • Data Provenance: The testing was "simulated," implying in-house lab testing, not patient data from a specific country, nor retrospective or prospective in a clinical sense.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • The document does not mention the use of experts to establish ground truth for any test set. The nature of the device (a catheter and a visualization kit) and the type of testing described (simulated use for an accessory) likely did not necessitate independent expert ground truth establishment in the way a diagnostic AI algorithm would.

    4. Adjudication Method for the Test Set:

    • Since there's no mention of experts establishing a ground truth for a test set, no adjudication method was needed or described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was mentioned or performed. This type of study is typically done for diagnostic imaging devices or AI algorithms where human interpretation is involved. This submission is for a medical device (catheter) and an accessory, not a diagnostic imaging AI.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study:

    • Not applicable. The device is a physical catheter and a visualization kit, not an algorithm. Therefore, no standalone algorithm performance study was mentioned.

    7. Type of Ground Truth Used:

    • For the simulated use testing of the Particle Visualization Kit, the "ground truth" would likely be engineer-defined criteria and observations of whether the kit successfully facilitated the visualization of deliberately introduced embolic material. This is based on the functional description of the PVK. No mention of expert consensus, pathology, or outcomes data for this specific testing.

    8. Sample Size for the Training Set:

    • Not applicable. This device is hardware (catheter and accessory), not a machine learning or AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set.

    In summary, the provided document focuses on establishing substantial equivalence for a non-AI physical medical device and an accessory. It describes the device's characteristics and notes that non-clinical simulated use testing was performed for the new accessory, but it does not delve into detailed quantitative acceptance criteria or elaborate on study designs typical of AI/diagnostic device submissions. The core argument is that the modified device is identical or substantially equivalent in its critical features to an already cleared predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090364
    Manufacturer
    Date Cleared
    2010-03-23

    (403 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ANGIOSLIDE, LTD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide eXtraTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral translyminal angioplasty and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The Angioslide eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is a single-use disposable over-the-wire co-axial dual lumen catheter with a foldable balloon near the distal tip. The balloon catheter consists of the balloon located near the distal atraumatic soft tip, the telescopic shaft and the handle.

    One lumen is used for inflation of the balloon and is accessed via the inflation port. The other lumen, starting at the quidewire port. allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopague markers for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.

    The shaft comprises the outer shaft and the inner shaft. The distal end of the balloon is connected to the inner shaft and the proximal end of the balloon is connected to the outer shaft. The inner shaft is connected to the pulling rod and the outer shaft is connected to the handle body. The pulling rod lock locks the handle body and the pulling rod together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by unlocking the pulling rod lock counter-clockwise and retracting the pulling rod. The inward-folding of the balloon forms a cavity and allows for collection of embolic material.

    AI/ML Overview

    Here's an analysis of the Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature based on the provided 510(k) summary, focusing on acceptance criteria and the supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly lay out a table of acceptance criteria with corresponding performance metrics from the studies for the device itself in a quantitative manner. Instead, the document describes the types of tests performed and states that the "results of the MC-LEADER and MC-LEADER Supplemental Studies met performance goals derived from historical literature describing PTA performance." For the animal study, it states: "The results of this study demonstrated that the safety and performance of the Angioslide eXtra™ PTA Balloon Catheter was substantially equivalent to that of the control article when used in an in vivo model."

    However, based on the identified endpoints from the clinical studies, we can infer the performance goals the device aimed to meet:

    Acceptance Criterion (Inferred from Study Endpoints)Reported Device Performance
    Acute device successMet performance goals
    Acute procedural success (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1