Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120805
    Manufacturer
    Date Cleared
    2012-04-13

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angroplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

    The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.

    Device Description

    The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9).

    One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.

    AI/ML Overview

    Acceptance Criteria and Device Performance for PROTEUS™ PTA Catheter with Embolic Capture Feature

    This submission describes the non-clinical testing performed to demonstrate the substantial equivalence of the modified PROTEUS™ PTA Catheter with Embolic Capture Feature. The testing followed Angioslide's Risk Analysis, applicable FDA Guidance documents, and ISO standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual inspection - external surfaceExternal surface of catheter effective length, including the distal end, is free from extraneous matter and surface defects (<0.2mm² TAPPI). 90% Confidence, 90% ReliabilityPASS
    Visual inspection - distal tipDistal tip is smooth, rounded, tapered, or similarly finished. 90% Confidence, 90% ReliabilityPASS
    Distal Bond Outer DiameterIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Distal Balloon Cone to Distal TipIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Distal Tip to Proximal Outer Tube Tip LengthIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Soft Tip LengthIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Wrapped Balloon Outer DiameterIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Guide Wire Lumen Inner DiameterIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Proximal Bond Outer DiameterIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Marker Band SpacingIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Balloon Working LengthIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Balloon Outer DiameterIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Catheter Overall Effective LengthIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Catheter Overall LengthIn Tolerance. 90% Confidence, 90% ReliabilityPASS
    Minimum Balloon Burst Pressure (RBP)RBP ≥ 14atm. 95% Confidence, 99.9% ReliabilityPASS
    Balloon Fatigue (Repeated Inflation/Deflations)Inflation/Deflation Cycles ≥ 10 at 14atm (for 3x100). No leakage, rupture, and/or herniation up to max 40 cycles. 95% Confidence, 90% ReliabilityPASS
    Tensile Strength- Distal Balloon to Inner Tube (Peel)≥ 5N. 90% Confidence, 95% ReliabilityPASS
    Tensile Strength- Distal Balloon to Inner Tube (Shear)≥ 5N. 90% Confidence, 95% ReliabilityPASS
    Tensile Strength- Proximal Balloon to Outer Tube≥ 10N. 90% Confidence, 95% ReliabilityPASS
    Tensile Strength- Cylinder to T-Connector≥ 15N. 90% Confidence, 90% ReliabilityPASS
    Tensile Strength- Inflation Tube to T-Connector≥ 15N. 90% Confidence, 90% ReliabilityPASS
    Tensile Strength- Outer Tube to T-Connector≥ 15N. 90% Confidence, 90% ReliabilityPASS
    Tensile Strength- Inner Shaft to Pulling Rod≥ 10N. 90% Confidence, 95% ReliabilityPASS
    Tensile Strength- Pulling Rod to Proximal Luer≥ 15N. 90% Confidence, 90% ReliabilityPASS
    Tensile Strength- Pulling Rod to Knob Base≥ 15N. 90% Confidence, 90% ReliabilityPASS
    Tensile Strength- Cylinder to O-ring Cap≥ 15N. 90% Confidence, 90% ReliabilityPASS
    Liquid LeakageNo leakage. 90% Confidence, 90% ReliabilityPASS
    Balloon Inflation/Deflation TestingInflation Time: ≤ 14.0s. Deflation Time: ≤ 30.6s. 90% Confidence, 90% ReliabilityPASS
    Balloon ComplianceCompliance: ≤ 13%. Nominal Pressure: 8atm. RBP: 14atm. 90% Confidence, 90% ReliabilityPASS
    Flow Characteristics- Straight ConfigurationDistal flow observed in uninflated and deflated state, occlusion of distal flow in inflated state. 90% Confidence, 90% ReliabilityPASS
    Flow Characteristics- Extreme Angle ConfigurationDistal flow observed in uninflated and deflated state, occlusion of distal flow in inflated state. 90% Confidence, 90% ReliabilityPASS
    Stroke LengthMinimum Stroke Length: ≥ 70% of the associated Balloon Working Length. Maximum Stroke Length: Balloon can be deflated after reaching stroke limit.PASS
    Simulated Use in Tortuous Anatomy Model - guide wire compatibility90% Confidence / 90% Reliability. Catheter can be mounted over a 0.014" guide wire. 90% Confidence, 90% ReliabilityPASS
    Simulated Use in Tortuous Anatomy Model - Introducer Sheath Compatibility90% Confidence, 90% Reliability. Completely folded balloon passes through identified Introducer Sheath at the end of procedure. 90% Confidence, 90% ReliabilityPASS
    Simulated Use in Tortuous Anatomy Model - Kink Resistance90% Confidence, 90% Reliability. No permanent deformations (kinks) are present once removed from the tortuous anatomy model. 90% Confidence, 90% ReliabilityPASS
    Simulated Use in Tortuous Anatomy Model - Max Advancement ForceN/A - Characterization onlyN/A
    Simulated Use in Tortuous Anatomy Model - Max Collapse ForceN/A - Characterization onlyN/A
    Simulated Use in Tortuous Anatomy Model - Max Anatomy Retraction ForceN/A - Characterization onlyN/A
    Simulated Use in Tortuous Anatomy Model - Max Removal Retraction ForceN/A - Characterization onlyN/A
    Capture Efficiency ComparisonN/A - Characterization onlyN/A

    2. Sample Sizes Used for the Test Set and Data Provenance

    The sample sizes for testing varied depending on whether the data collected was variable or attribute data, which was determined by the risk analysis (DFMEA) performed for the PROTEUS™ PTA catheter and required confidence/reliability levels as per FDA Guidance "Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems." The documentation does not specify the exact number of units tested for each individual test but states that the number met the required statistical confidence and reliability levels.

    The data provenance is from in vitro bench testing conducted by Angioslide, Ltd. and is thus not from a specific country of origin in a clinical context (e.g., patient data). It is prospective in the sense that the tests were designed and conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The tests described are non-clinical (bench testing) of the physical and mechanical properties of the device, not an evaluation requiring expert interpretation of clinical ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. This was bench testing, not an evaluation relying on expert consensus for clinical interpretations. The results are objective measurements against defined criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted as this is a device performance evaluation (bench testing) and not an AI-based diagnostic device.

    6. Standalone Performance Study

    This document details standalone performance testing of the device itself (hardware), not an algorithm or AI without human interaction. The results demonstrate the device's ability to meet its specifications.

    7. Type of Ground Truth Used

    The "ground truth" for these tests was established by engineering specifications, regulatory standards (ISO), and defined acceptance criteria for the physical and mechanical properties of the device, such as dimensions, tensile strength, burst pressure, and flow characteristics.

    8. Sample Size for the Training Set

    Not applicable. This is a medical device (catheter) and not an AI/Machine Learning algorithm, therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1