K Number
K120164

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-06-08

(141 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.

Device Description

The Angioslide PROTEUSTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9). One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure. The shaft (4) comprises the outer shaft (6) and the inner shaft (7). The distal end of the balloon (A) is connected to the inner shaft and the proximal end of the balloon (B) is connected to the outer shaft. The inner shaft is connected to the proximal hub (10) which is connected to the pulling knob (1) and the outer shaft is connected to the handle grip (3). The pulling knob lock (11) locks the handle grip and the pulling knob together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by pressing on pulling knob lock ( 1 l ) and pulling the pulling knob away from the handle (1). The embolic capture feature involves a single-use suction mechanism that works through inward folding of the balloon, which creates negative pressure within the capture cavity for debris capture and removal. The reduced pressure in the capture cavity causes some of the particles that are released during the procedure to flow into the cavity for containment and removal. The balloon size and diameter are printed on the strain relief (12). Refer also to the package label for information about catheter length, balloon nominal and rated burst pressure, balloon size, balloon compliance, guidewire compatibility and sheath compatibility.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the PROTEUS™ PTA Catheter with Embolic Capture Feature, based on the provided text:

Acceptance Criteria and Device Performance

The device underwent extensive non-clinical bench testing, sterilization, packaging, and shelf-life testing to demonstrate conformance. The table below summarizes the acceptance criteria and reported device performance for these non-clinical tests.

Note: The provided document is a 510(k) summary for a submission that claims substantial equivalence to a predicate device (K111750), implying that the new device has identical technological characteristics for both PTA balloon function and embolic capture. Therefore, the "study" described here is a suite of design verification and validation tests rather than a comparative clinical trial.

Test DescriptionAcceptance CriteriaReported Device Performance
Visual Inspection - External SurfaceExternal surface of catheter effective length, including the distal end, is free from extraneous matter and surface defects (90% Confidence, 90% Reliability)PASS
Visual Inspection - Distal TipDistal tip is smooth, rounded, tapered, or similarly finished. (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Distal Bond O.D.In Tolerance (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Soft Tip LengthIn Tolerance (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Guidewire Inner LumenIn Tolerance (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Wrapped Balloon O.D.The balloon must pass aperture < 2.00mm at 37deg C (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Catheter Overall LengthIn tolerance (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Catheter Overall Effective LengthIn tolerance (90% Confidence, 90% Reliability)PASS
Dimensional Inspection - Knob HeightIn tolerance (90% Confidence, 90% Reliability)PASS
Handle Assembly BurstBurst Pressure ≥ 12 atm (90% Confidence, 90% Reliability)PASS
Handle Assembly Fatigue≥ 10 Inflation/Deflation Cycles to RBP (95% Confidence, 90% Reliability)PASS
Handle Assembly LeakageNo LeakagePASS
Balloon Inflation and Deflation Time90% Confidence, 90% Reliability; No leakage during inflation; Inflation Time: ≤ 14.0 sec, Deflation Time: ≤ 30.6 secPASS
Tensile Strength - Distal Balloon to Inner Shaft (Peel)90% Confidence, 90% Reliability; Force at Break ≥ 10NPASS
Tensile Strength - Distal Balloon to Inner Shaft (Shear)90% Confidence, 95% Reliability; Force at Break ≥ 10NPASS
Tensile Strength - Outer Shaft to T Connector90% Confidence, 95% Reliability; Force at Break ≥ 10NPASS
Tensile Strength - Inflation Tube to T Connector90% Confidence, 90% Reliability; Force at Break ≥ 15NPASS
Tensile Strength – Inflation Tube to Inflation Luer90% Confidence, 90% Reliability; Force at Break ≥ 15NPASS
Tensile Strength - Cylinder to T Connector90% Confidence, 90% Reliability; Force at Break ≥ 15NPASS
Tensile Strength - Inner Shaft to Pulling Rod90% Confidence, 90% Reliability; Force at Break ≥ 10NPASS
Tensile Strength - Pulling Rod to Proximal Luer90% Confidence, 95% Reliability; Force at Break ≥ 15NPASS
Tensile Strength - Pulling Rod to Knob Base90% Confidence, 90% Reliability; Force at Break ≥ 15NPASS
Tensile Strength - Distal Cap to Shells90% Confidence, 90% Reliability; Force at Break ≥ 15NPASS
Tensile Strength - Proximal Cap to Shells90% Confidence, 90% Reliability; Force at Break ≥ 15NPASS
Tensile Strength - Knob Base to Knob CoverForce at Break ≥ 15N (90% Confidence, 90% Reliability)PASS
Corrosion ResistanceNo signs of corrosion after treatmentPASS
Stroke LengthMinimum Stroke Length: ≥ 70% of the associated Balloon Working Length; Maximum Stroke Length: Balloon can be deflated after reaching stroke limit.PASS
Simulated Use in Tortuous Anatomy Model - guidewire compatibility90% Confidence / 90% Reliability; Catheter can be mounted over a .035” guidewirePASS
Simulated Use in Tortuous Anatomy Model - Advance/Retract/Deploy/Fold Balloon/Withdraw after Procedure90% Confidence, 90% Reliability; Completely folded balloon passes through identified Introducer Sheath at the end of procedure.PASS
Simulated Use in Tortuous Anatomy Model - Kink Resistance90% Confidence, 90% Reliability; No permanent deformations (kinks) are present once removed from the tortuous anatomy model.PASS
Flow Characteristics90% Confidence, 90% Reliability; Distal flow observed in uninflated and deflated state, occlusion of distal flow in inflated state (straight and bend configurations)PASS
Labeling Validation90% Confidence, 90% Reliability; Instructions for performing type 1 and type 2 recovery methods can be performed successfully per instructions.PASS
Packaging Validation - Pouch Label Visual InspectionLabel is securely attached to the pouch surface. Across entire label, details/print is legible and free of smudging or creasing (90% Confidence, 90% Reliability)PASS
Packaging Validation - Pouch Seal Visual InspectionPouch Seals at locations A, B, and C are visibly free of wrinkles, channels, visible bubbles, foreign material, transparent areas, and non-uniform seal width. (90% Confidence, 90% Reliability)PASS
Packaging Validation - Bubble Leak TestingNo leaks are observed from seals or surface of the pouch. (90% Confidence, 90% Reliability)PASS
Packaging Validation - Seal Strength TestingContract Manufacturer and Supplier Seal: Peel Force ≥ 5.0N (90% Confidence, 90% Reliability)PASS
Pyrogenicity (LAL Test)Must meet spec of 0.5 EU/ml for general blood contacting medical devicesMet (successful completion)
EtO ResidualsWithin the ISO 10993-7 standard specificationMet
Sterility Assurance Level (SAL)1x10^-6Met
Shelf Life (Device Function and Performance)All functional and performance specifications met after sterilization and accelerated aging (3-year shelf life)Met

Study Details for PROTEUS™ PTA Catheter

Based on the provided K120164 510(k) summary, the "study" refers to a series of non-clinical bench tests for design verification and validation, as well as tests for biocompatibility, sterilization, packaging, pyrogenicity, and shelf life. This is a 510(k) submission, which typically relies on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting new clinical trials for effectiveness.

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: The sample sizes for Design Verification and Validation testing "were based on required confidence/reliability levels as a result of risk analysis performed for the PROTEUS™ PTA Balloon Catheter, or per recommendations within the FDA Guidance 'Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems.'" Specifically, the number of samples utilized varied depending on whether the data was variable or attribute.
      • For the "four corners" approach, at least four sizes (largest/smallest diameters and lengths) plus a middle catheter size were tested for robustness.
      • Many tests specify a confidence/reliability level (e.g., 90% Conf., 90% Rel. or 95% Conf., 90% Rel.), which dictates the minimum sample size for attribute or variable data.
    • Data Provenance: The tests were "in vitro bench testing" conducted by Angioslide, Ltd., an Israeli company. The data is prospective, generated specifically for this submission. The origin of the data itself is from laboratory bench tests.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This document describes bench testing of a medical device's physical and functional characteristics, not an AI or diagnostic device that requires expert ground truth. Therefore, no experts (e.g., radiologists) were used to establish ground truth in the traditional sense of medical image interpretation or diagnosis. The "ground truth" for these tests are the pre-defined engineering specifications and performance standards outlined in the acceptance criteria.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of clinical or image-based findings. The results of the bench tests (e.g., whether a dimension is "in tolerance" or if "no leakage" is observed) are direct physical measurements and observations against pre-defined criteria.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (PTA Catheter) and not an AI or diagnostic imaging device. Therefore, no MRMC study, human readers, or AI assistance was involved.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm or AI system.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the bench tests, the "ground truth" is defined by engineering specifications, regulatory standards (e.g., ISO, FDA guidance documents), and internal design requirements. For example, "In Tolerance" for a dimension, "No Leakage" for a seal, "Force at Break ≥ 10N" for tensile strength. For pyrogenicity, it's a specific endotoxin limit (0.5 EU/ml).
  7. The sample size for the training set:

    • Not applicable. This is not an AI or machine learning device that requires a training set. The "training" for this device would be its manufacturing process and design iterations, not data sets for an algorithm.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set mentioned or implied for this type of device.

{0}------------------------------------------------

K120164

JUN - 8 2012

510(k) Summary

PROTEUS™ PTA Catheter with Embolic Capture Feature

Introduction

This document contains the 510(k) summary for the modified PROTEUS™ PTA Catheter with Embolic Capture Feature. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

Applicant Name and Address:

Name:Angioslide, Ltd.
Address:7 Halamish St.Business and Industry Park - NorthCaesarea 30889Israel
Tel: 011-972-4-6187200Fax: 011-972-4-6187222
Official Contact:Clay AnselmoCEOReglera LLC
Address:11925 W. I-70 Frontage Rd. NorthSuite 900Wheat Ridge, CO 80033
Phone:303-223-4303
Fax:303-832-6700
E-mail:anselmoc@reglera.com

Summary Preparation Date: 1/18/2012

Device Name and Classification:

Trade Name:PROTEUS™ PTA Balloon Catheter withEmbolic Capture Feature
Common Name:Percutaneous Transluminal AngioplastyBalloon Catheter
Classification Name:Catheter, Percutaneous
Classification:Class II, 21 CFR 870.1250
Product CodeDQY/LIT

{1}------------------------------------------------

Predicate Devices:

The modified PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is claimed to be substantially equivalent to the following legally marketed predicate device:

  • Angioslide PROTEUSTM PTA Balloon Catheter with Embolic Capture Feature . (K111750)
    Performance Standards: There are no mandatory performance standards for this device.

Device Description (see Figure 1)

Image /page/1/Figure/6 description: This image shows a diagram of a medical device with labeled parts. The device includes a handle (3), a needle (4, 5), and a connector (2). Detail A shows a close-up of the needle tip (9), while Detail B shows the needle base (6). Other labeled components include 1, 7, 8, 10, 11, and 12.

Figure 1

The Angioslide PROTEUSTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9).

One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.

{2}------------------------------------------------

The shaft (4) comprises the outer shaft (6) and the inner shaft (7). The distal end of the balloon (A) is connected to the inner shaft and the proximal end of the balloon (B) is connected to the outer shaft. The inner shaft is connected to the proximal hub (10) which is connected to the pulling knob (1) and the outer shaft is connected to the handle grip (3). The pulling knob lock (11) locks the handle grip and the pulling knob together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by pressing on pulling knob lock ( 1 l ) and pulling the pulling knob away from the handle (1). The embolic capture feature involves a single-use suction mechanism that works through inward folding of the balloon, which creates negative pressure within the capture cavity for debris capture and removal. The reduced pressure in the capture cavity causes some of the particles that are released during the procedure to flow into the cavity for containment and removal.

The balloon size and diameter are printed on the strain relief (12). Refer also to the package label for information about catheter length, balloon nominal and rated burst pressure, balloon size, balloon compliance, guidewire compatibility and sheath compatibility.

Indications for Use:

The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, illo-femoral, poplited, tibial, peroneal, and profunda arteries.

The PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.

Comparison of Technological Characteristics

The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is an over the wire co-axial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip. The catheter is compatible with a 0.035" guidewire.

The balloon catheter technological characteristics of the PROTEUS™ PTA Balloon Catheter are identical to those of the PROTEUS™ PTA Balloon Catheter (K111750). In both devices, lesion dilation is achieved by means of an inflatable balloon.

The PROTEUS™ PTA Balloon Catheter overall length, catheter sheath sizing, balloon diameter, balloon length, balloon nominal pressure, balloon rated burst pressure and end hole diameter are the same as those of the PROTEUS™ PTA Balloon Catheter (K111750).

The embolic capture technological characteristics of the PROTEUS™ PTA Balloon Catheter are identical to those of the PROTEUS™ PTA Balloon Catheter (K111750). In both devices, the containment and removal of embolic material is achieved by proximal vessel occlusion, by means of an inflatable balloon, and subsequent capture and removal of embolic material.

Summary of Non-Clinical Testing

{3}------------------------------------------------

In vitro bench testing of the Angioslide PROTEUS™ PTA Balloon Catheter was conducted in accordance with Angioslide's Risk Analysis and all applicable FDA Guidance documents and ISO standards, including:

ISO 10555-1 - Sterile, Single Use Intravascular Catheters- Part 1: General Requirements

ISO 10555-4 - Sterile, Single Use Intravascular Catheters- Part 4: Balloon Dilatation Catheters

FDA Guidance - Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems , January 13, 2005

FDA Guidance - Coronary and Carotid Embolic Protection Devices - Premarket Notification [510(k)] Submissions, February 15, 2008

All bench testing, unless otherwise specified, was conducted using finished devices which were sterilized by the final validated sterilization process.

Design Verification and Validation:

Design verification and validation testing of the PROTEUSTM PTA Balloon Catheter was performed using a "four corners" approach (i.e. 2 x 2 factorial of the largest and smallest balloon diameters and lengths. In addition to these four sizes, a middle catheter size was tested to provide supplemental data to support the robustness of the overall design of the Angioslide PROTEUSTM PTA Balloon Catheter. Sample sizes used for Design Verification and Validation testing were based on required confidence/reliability levels as a result of risk analysis performed for the PROTEUS™ PTA Balloon Catheter, or per recommendations within the FDA Guidance "Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems." Specifically, the number of samples utilized for each test depended on whether the data to be collected was variable data or attribute data in nature.

{4}------------------------------------------------

Test DescriptionAccept/Reject CriteriaResults
Visual Inspection -External SurfaceExternal surface of catheter effectivelength, including the distal end, is freefrom extraneous matter and surfacedefects90% Confidence, 90% ReliabilityPASS
Visual Inspection -Distal TipDistal tip is smooth, rounded, tapered, orsimilarly finished.90% Confidence, 90% ReliabilityPASS
DimensionalInspection - DistalBond O.D.In Tolerance90% Confidence, 90% ReliabilityPASS
DimensionalInspection - SoftTip LengthIn Tolerance90% Confidence, 90% ReliabilityPASS
DimensionalInspection -Guidewire InnerLumenIn Tolerance90% Confidence, 90% ReliabilityPASS
DimensionalInspection -Wrapped BalloonO.D.The balloon must pass aperture <2.00mm at 37deg C90% Confidence, 90% ReliabilityPASS
DimensionalInspection -Catheter OverallLengthIn tolerance90% Confidence, 90% ReliabilityPASS
DimensionalInspection -Catheter OverallEffective LengthIn tolerance90% Confidence, 90% ReliabilityPASS
DimensionalInspection - KnobHeightIn tolerance90% Confidence, 90% ReliabilityPASS
Handle AssemblyBurstBurst Pressure ≥ 12 atm90% Confidence, 90% ReliabilityPASS
Handle AssemblyFatigue≥ 10 Inflation/Deflation Cycles to RBP95% Confidence, 90% ReliabilityPASS
Test DescriptionAccept/Reject CriteriaResults
Handle AssemblyLeakageNo LeakagePASS
Balloon Inflationand Deflation Time90% Confidence, 90% ReliabilityNo leakage during inflationInflation Time: ≤ 14.0 sec,Deflation Time: ≤ 30.6 secPASS
Tensile Strength -Distal Balloon toInner Shaft (Peel)90% Confidence, 90% ReliabilityForce at Break ≥ 10NPASS
Tensile Strength -Distal Balloon toInner Shaft (Shear)90% Confidence, 95% ReliabilityForce at Break ≥ 10NPASS
Tensile Strength -Outer Shaft to TConnector:90% Confidence, 95% ReliabilityForce at Break ≥ 10NPASS
Tensile Strength -Inflation Tube to TConnector:90% Confidence, 90% ReliabilityForce at Break ≥ 15NPASS
Tensile Strength –Inflation Tube toInflation Luer90% Confidence, 90% ReliabilityForce at Break ≥ 15NPASS
Tensile Strength -Cylinder to TConnector90% Confidence, 90% ReliabilityForce at Break ≥ 15NPASS
Tensile Strength -Inner Shaft toPulling Rod90% Confidence, 90% ReliabilityForce at Break ≥ 10NPASS
Tensile Strength -Pulling Rod toProximal Luer90% Confidence, 95% ReliabilityForce at Break ≥ 15NPASS
Tensile Strength -Pulling Rod toKnob Base90% Confidence, 90% ReliabilityForce at Break ≥ 15NPASS
Tensile Strength -Distal Cap to Shells90% Confidence, 90% ReliabilityForce at Break ≥ 15NPASS
Tensile Strength -Proximal Cap toShells90% Confidence, 90% ReliabilityForce at Break ≥ 15NPASS
Test DescriptionAccept/Reject CriteriaResults
Tensile Strength-Knob Base to KnobCoverForce at Break ≥ 15N90% Confidence, 90% ReliabilityPASS
CorrosionResistanceNo signs of corrosion after treatmentPASS
Stroke LengthMinimum Stroke Length:≥ 70% of the associated BalloonWorking LengthMaximum Stroke Length:Balloon can be deflated after reachingstroke limit.PASS
Simulated Use inTortuous AnatomyModel - guidewirecompatibility90% Confidence / 90% ReliabilityCatheter can be mounted over a .035”guidewirePASS
Simulated Use inTortuous AnatomyModel -Advance/Retract/Deploy/FoldBalloon/Withdrawafter Procedure90% Confidence, 90% ReliabilityCompletely folded balloon passesthrough identified Introducer Sheath atthe end of procedure.PASS
Simulated Use inTortuous AnatomyModel - KinkResistance90% Confidence, 90% ReliabilityNo permanent deformations (kinks) arepresent once removed from the tortuousanatomy model.PASS
FlowCharacteristics90% Confidence, 90% ReliabilityDistal flow observed in uninflated anddeflated state, occlusion of distal flow ininflated state (straight and bendconfigurations)PASS
LabelingValidation90% Confidence, 90% ReliabilityInstructions for performing type 1 andtype 2 recovery methods can beperformed successfully per instructions.PASS
Test DescriptionAccept/Reject CriteriaResults
PackagingValidation - PouchLabel VisualInspectionLabel is securely attached to the pouchsurface.Across entire label, details/print islegible and free of smudging or creasingPASS
90% Confidence, 90% Reliability
PackagingValidation - PouchSeal VisualInspectionPouch Seals at locations A, B, and C arevisibly free of wrinkles, channels,visible bubbles, foreign material,transparent areas, and non-uniform sealwidth.90% Confidence, 90% ReliabilityPASS
PackagingValidation -Bubble LeakTestingNo leaks are observed from seals orsurface of the pouch.90% Confidence, 90% ReliabilityPASS
PackagingValidation - SealStrength TestingContract Manufacturer and SupplierSeal: Peel Force $\geq$ 5.0N90% Confidence, 90% ReliabilityPASS

PROTEUS™ PTA Balloon Catheter Design Verification and Validation Summary Table

{5}------------------------------------------------

.

{6}------------------------------------------------

{7}------------------------------------------------

:

{8}------------------------------------------------

Biocompatibility Testing:

Based on Risk Analysis no additional biocompatibility testing was required for this modification.

Sterilization:

The PROTEUS™ PTA Balloon Catheter is packaged and sterilized using substantially equivalent materials, methods, and sterilization parameters used for most commercially available PTA balloon catheter products. The PROTEUSTM PTA Balloon Catheter is sterilized by ethylene oxide (EtO) sterilization providing a Sterility Assurance Level (SAL) of Ix10°. EtO sterilization validation was completed in accordance with ISO 11135 "Medical devices -Validation and routine control of ethylene oxide sterilization". Testing of EtO residuals was performed in accordance with ISO 10993-7 "Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals." Results of the sterilization validation were found to meet all acceptance criteria and confirmed a SAL of 1x10", and EtO/ECh residuals were found to be within the ISO 10993-7 standard specification.

Packaging:

Packaging validation testing was conducted on the PROTEUS™ PTA Balloon Catheter packaging which included visual inspection (per ASTM F1886), leak testing (per ASTM F2096/ASTM F1929), and pouch seal strength testing (per ASTM F88). Prior to the start of packaging integrity testing, all samples were pre-conditioned by sterilization and subjected to transportation simulation per ASTM D4332-01 and ASTM D4169 respectively. All results of the packaging validation testing were found to meet acceptance criteria.

Pyrogenicity:

The PROTEUS™ PTA Balloon catheter is non-pyrogenic based successful completion of pyrogenicity testing:

Description of the methodused to make thedetermination:The Limulus amebocyte lysate (LAL) test is used fordetection and quantification of bacterial endotoxin on eachlot prior to release.
Identification of the testingendpoint reached andrationale for selecting thatendpoint:Must meet spec of 0.5 EU/ml for general blood contactingmedical devices
Description of theextraction technique usedto obtain the test fluid fromthe test device:Flow through method, volume of LAL water used:40ml/device
Identification of thereference method used:FDA Guideline on Validation of the Limulus AmebocyteLysate (LAL) Test as an End-Product Endotoxin Test,Dec 1987,USP <85> Bacterial Endotoxins Test

{9}------------------------------------------------

Shelf Life:

The PROTEUS™ PTA Balloon Catheter is currently validated for a three (3) year shelf life. After sterilization and accelerated aging, device function and performance verification/validation testing was conducted on the PROTEUS™ PTA Balloon Catheter and its packaging system that included the following:

  • Device
    • Visual Inspection o
    • Dimensional Inspection
    • Balloon Burst Pressure Testing 0
    • Balloon Fatigue Testing O
    • Balloon Compliance Testing 0
    • Balloon Inflation/Deflation Testing o
    • Catheter Leakage Testing O
    • Bond Tensile Strength Testing o
    • Simulated Use Testing 0
    • Stroke Length Testing o
    • Flow Characteristics Testing o
    • o Corrosion Resistance Testing

Packaging .

  • O Visual Inspection (per ASTM F1886)
  • o Leak Testing (per ASTM F2096/ASTM F1929)
  • o Pouch Seal Strength Testing (per ASTM F88)

The results of the testing described above were found to meet all acceptance criteria, and demonstrate that the PROTEUS™ PTA Balloon Catheter and its packaging system meets all functional and performance specifications at the product's three year shelf-life. Accelerated aging results are being confirmed through real time aging studies.

Substantial Equivalence Conclusion:

Non-clinical verification and validation of the PROTEUS™ PTA Balloon Catheter was performed through extensive bench testing, sterilization, packaging and shelf life testing. Results of the testing demonstrated that the PROTEUS™ PTA Balloon Catheter design met all specifications and is adequate for its intended use. Additionally, the test results demonstrated substantial equivalence of the PROTEUS™ PTA Balloon Catheter to its predicate device.

In conclusion, the PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the following legally marketed predicate device:

  • PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature, Angioslide -K11750

{10}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/10/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and features an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the eagle in a circular fashion. The seal is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN - 8 2012

Angioslide c/o Clay Anselmo CEO Reglera, LLC 11925 West I-70 Frontage Road North Suite 900 Wheat Ridge, CO 80033

Re: K120164

Trade/Device Name: PROTEUS PTA Balloon Catheter with Embolic Capture Feature Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 11, 2012 Received: May 14, 2012

Dear Mr. Anselmo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{11}------------------------------------------------

Page 2 - Mr. Clay Anselmo

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

M.S. Killebrew

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{12}------------------------------------------------

Indications for Use

510(k) Number (if known): 上(ac) 164

Device Name: PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature

Indications for Use:

The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.

The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Miz Keller

Page 1 of 1

(Division Sign-Off) Division of Cardiovascular Devices

K120164 510(k) Number_

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).