Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K202144
    Date Cleared
    2022-09-23

    (784 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    A.B. Dental Device Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are dental prosthetics placed onto a dental implant or screw retained abutment, to provide support for dental prosthetic restorations. The Base Abutments and Adhesive Sleeves include: Titanium bases/sleeves with a pre-machined implant connection or screw retained abutment connection on which the CAD/CAM designed upper structure may be fitted to complete a two-piece dental abutment. The Base Abutments and Adhesive Sleeves include an abutment/sleeve screw for fixation to the underlying implant/screw retained abutment. These dental prosthetics may be used for single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with AB Dental Implants with internal connection and screw retained connections for standard and narrow platforms.

    All digitally designed abutments and/or copings for use with AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are intended to be sent to AB Dental validated milling center for manufacture.

    Device Description

    A.B. Dental Devices® CADCAM Products consists of; Base Abutments and Adhesive Sleeves which are used with Standard Platform (Internal Hex connection) implants, Narrow Platform implants and Screw retained (P64 and P14/P16 MultiUnit) abutments.

    Base Abutments (standard or narrow platform) and Adhesive Sleeves (screw retained abutment platform) are prefabricated abutments/sleeves made to be used as final abutments/ sleeves that will be used as adhesive connection (cement retained) of the prosthetic restoration on implant platform or on Multiunit platforms.

    The design and manufacturing of the patient-specific abutments take consideration the shape of the final prosthesis based on the patient's intra-oral scanning using the manufacturing. All digitally designed abutments and/or copings for use with AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are intended to be sent to AB Dental validated milling center for manufacture.

    This submission adds the following Base Abutments and Adhesive sleeves:

    Ti Base Anti-Rotational Abutments (Standard or Narrow platforms) - Base Abutment
    Ti Base Rotational Abutments (Standard or Narrow platforms) - Base Abutment
    P64 Ti Conical Adhesive Sleeve - Adhesive Sleeve
    Ti Straight Adhesive Sleeves - Adhesive Sleeve

    Base Abutments;
    The titanium bases are used as part of a two-piece abutment, where the base is premanufactured from titanium alloy (Ti-6Al-4V ELI) and the top half is a CAD/CAM zirconia superstructure, milled at a validated milling center. These pieces are cemented together to form the final abutment.

    Raw material Zirconium blanks
    Dental Direkt Zirconia Blanks by Dental Direkt GmbH, cleared under K183569

    Dental Cement
    PANAVIA SA Cement Universal by Kuraray Noritake Dental Inc., cleared under K183537

    Adhesive Sleeves
    The adhesive sleeves are all Rotational (non-engage) for bridges restoration.

    The adhesive sleeves are made of Titanium alloy (Ti-6Al-4V ELI) and will be installed on straight or angular screw-retained abutments.

    The Multiunit (screw retained abutments already cleared devices) are P14/P16 platform: The below listed devices are previously cleared items that the sleeves are intended to be used with:
    P14-3,17-1, P14-3,17-3, P14-3,30-1, P14-3,30-3, P14-3.75,17-1, P14-3.75,17-3, P14-3.75,30-1, P14-3.75,30-3, P16-3.75,1, P16-3.75,2, P16-3.75,3, P16-3.75,4, P16-3.75,5, P64-3,1, P64-3,17-0.5, P64-3,17-2, P64-3,2, P64-3,3, P64-3,30-0.5, P64-3,30-2, P64-3.75,1, P64-3.75,2, P64-3.75,3, P64-3.75,4, P64-3.75,5, P64-3.75,17-0.5, P64-3.75,17-2, P64-3.75,17-3, P64-3.75,30-0.5, P64-3.75,30-2, P64-3.75.30-3

    AI/ML Overview

    The provided document is a 510(k) Summary for a dental device, not an AI/ML device. Therefore, it does not contain information related to AI/ML acceptance criteria or studies proving device performance from an AI/ML perspective.

    However, I can extract the acceptance criteria and performance data for the dental device as presented in the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for the device in the typical sense of numerical thresholds for metrics like accuracy, sensitivity, or specificity. Instead, the submission focuses on demonstrating "substantial equivalence" to a predicate device through comparison of technological characteristics and non-clinical performance data.

    The "acceptance criteria" are implied by the comparison to the predicate device and the successful completion of various non-clinical tests. The "reported device performance" refers to the results of these non-clinical tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Characteristics Comparison
    Material identical to predicate (Ti-6Al-4V ELI)Met: Subject device uses Ti-6Al-4V ELI.
    Platform compatibility (Internal implant connection)Met: Subject device uses Standard Platform, Narrow platform, which is comparable to Predicate's Internal Hex, Conical Connection, and Multiunit/2-CONnect platforms as applicable to the specific abutment type.
    Angulation (0° Straight structure)Met: Minimum and Maximum abutment angle (°): Straight structure with 0° angle.
    Design (Anti-rotation/Rotational for Base Abutments, SleeveMet: Base Abutments: Anti-rotation or Rotational (Non engaging). Adhesive Sleeves: Sleeve design.
    Zirconia compatibility (FDA Cleared Zirconia, manufactured by validated milling center)Met: Uses FDA Cleared Zirconia (Dental Direkt Zirconia Blanks by Dental Direkt GmbH, K183569), to be manufactured by validated milling center.
    Sterility (End user sterilized)Met: End user sterilized.
    Non-clinical Performance
    Mechanical worst-case analysisConducted: Engineering analysis used to determine if additional components constitute a new mechanical worst-case. (Implicitly, no new worst-case requiring further testing beyond what was done).
    Cytotoxicity per ISO 10993-5Met: Testing per ISO 10993-5 (Nelson Lab. Study Number 1502319-S01) was completed. (Result is implicitly passing).
    Fatigue testing at 40° per ISO 14801Met: Fatigue testing at 40° per ISO 14801 on the two worst-case constructs of the subject submission was completed. (Result is implicitly passing).
    Sterilization validation per standardsMet: Sterilization validation according to ANSI/AAMI/ISO 17665 and ISO 11138 was completed. (Result is implicitly passing).
    MRI compatibility (MR Conditional)Met: MRI testing conducted per ASTM F2052-21, ASTM F2213-17, ASTM F2182-19, and ASTM F2119-2013 demonstrates that the claimed products are MR Conditional, allowing safe scanning under stated conditions.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as this is a 510(k) submission for a non-AI/ML dental device. The "test set" refers to physical samples for engineering and biological testing, not a dataset of patient information. The provenance of the manufacturing is Israel (A.B. Dental Devices Ltd., 19 Hayalomim St., Ashdod 7761117 Israel) and the testing labs are mentioned (e.g. Nelson Lab).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. "Ground truth" in the context of AI/ML refers to accurately labeled data against which an algorithm's performance is measured. For this dental device, "ground truth" would be established by validated engineering specifications, material properties, and regulatory standards. The experts involved would be engineers, material scientists, and regulatory affairs specialists. No specific number or qualifications are mentioned for establishing these in the document itself, beyond stating that an "Engineering analysis" was performed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods are typically used in clinical studies involving interpretation by multiple human readers. For the non-clinical tests described, the results are derived from physical measurements and laboratory procedures, not interpretations subject to adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a dental prosthetic and does not involve AI or human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a dental prosthetic and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests, the "ground truth" is based on:

    • Established engineering specifications and material properties: For design comparisons (e.g., material, dimensions, angulation).
    • Validated laboratory testing standards: For cytotoxicity (ISO 10993-5), fatigue (ISO 14801), sterilization (ANSI/AAMI/ISO 17665, ISO 11138), and MRI compatibility (various ASTM standards).
    • Predicate device characteristics: The predicate device K181580 serves as a benchmark for demonstrating substantial equivalence.

    8. The sample size for the training set

    This is not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    This is not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182455
    Date Cleared
    2019-07-15

    (311 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    A.B. Dental Device Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    A.B. DENTAL DEVICES® Dental Implants System consists of narrow (XXX-3 model number) and standard (XXX-3.75 model number) platform internal hex implants that are used with screw retained, cement retained, or overdenture abutments. This submission adds the following abutments:

    • Expansion of the P4 angled abutments to include: .
      • Narrow platform abutments (standard and long lengths, 15 and 25 degrees) o
      • Narrow platform, anatomic abutments (15 degrees) o
      • o Additional standard platform abutments (15 and 25 degrees)
      • Additional standard platform, anatomic abutments (15 and 25 degrees) O
    • . Expansion of the P5 ball attachments to include narrow platform, 20 degree abutments
    • Expansion of the P14 multiunit, angled abutments to include:
      • Additional sleeve materials (Delrin and CoCr) O
      • Narrow platform abutments (17 and 30 degrees) O
      • P25 AB LOC for angular adaptor O
    • Addition of P64 multiunit, straight and angled product line (narrow and standard platform, 17 ● and 30 degrees, with Delrin and Ti alloy sleeves), including its healing cap.
    • . Modification to the P4a-S screw to improve its performance characteristics
    AI/ML Overview

    The provided document is a 510(k) summary for the A.B. DENTAL DEVICES® Dental Implants System. It describes the device, its indications for use, and a comparison to predicate devices, focusing on the expansion of existing abutment lines and the addition of new ones. The document details the performance testing conducted to demonstrate substantial equivalence, primarily bench testing for mechanical properties and rationale for other characteristics.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance values in the manner typically seen for algorithmic performance. Instead, it describes various tests and states that the differences in the subject device do not raise new issues or are addressed through testing to establish substantial equivalence to predicate devices. The "acceptance criteria" here are implicitly meeting the performance of the predicate devices or demonstrating that the modifications do not create a new worst-case scenario.

    Aspect of PerformanceAcceptance Criteria (Implicit)Reported Device Performance
    Mechanical Strength- Subject device does not create a new mechanical worst-case compared to predicate (e.g., shorter moment arm, more material).- Engineering analysis confirmed that new components do not constitute a new mechanical worst-case.
    - Meets or exceeds performance of predicate devices when tested per ISO 14801.- Dynamic testing per ISO 14801 was performed on worst-case constructs, demonstrating substantial equivalence (e.g., narrow platform constructs demonstrated substantial equivalence to Straumann Crossfit SLA 3.3 mm (K111357), P64-3.75 shown equivalent to standard platform P14 abutment).
    - Modified screw (P4a-S) meets static torque to failure requirements and dynamic testing per ISO 14801.- Static torque to failure testing conducted for modified P4a-S screw. Static and dynamic testing per ISO 14801 performed in the same construct. The results imply successful performance to support equivalence, though specific values are not provided.
    BiocompatibilityMaterials (Ti-6Al-4V ELI, Delrin, CoCr) are biocompatible and identical to those used in predicate devices.- Application of ISO 10993-1, with a scientific rationale that biocompatibility testing was not needed as materials and manufacturing methods are identical to the sponsor's own predicate devices.
    Cleaning/SterilityModifications do not create a new worst-case for cleaning or sterility. Sterilization process is validated per ISO 17665.- Scientific rationale used to determine that additional components do not constitute a new cleaning or sterility worst-case. Sterilization validated per ISO 17665. The same color anodization (gold) has been cleared on other abutments in K162482, suggesting no new sterility issues related to this aesthetic change.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document states that "worst-case construct(s)" were used for mechanical testing per ISO 14801. However, it does not specify the sample size (number of units) for these tests. It also does not provide information on the data provenance (e.g., country of origin, retrospective or prospective) as the tests are bench-based engineering tests rather than clinical studies with patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to this type of device and study. The "ground truth" here is based on engineering principles, material properties, and established test standards (like ISO 14801), not on expert consensus for clinical interpretations or diagnoses.

    4. Adjudication Method for the Test Set

    This information is not applicable as the tests are objective engineering measurements and evaluations against established standards, not subjective assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device is a dental implant system (hardware), not an AI or software as a medical device (SaMD) that involves human readers or clinical image interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This is not applicable as the device is a dental implant system, not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Engineering principles and rationale: Used to determine if components constitute a "new worst-case" mechanically or for cleaning/sterility.
    • Standardized bench testing: Specifically mentioned ISO 14801 for mechanical fatigue testing and static torque to failure.
    • Biocompatibility standards: Application of ISO 10993-1.
    • Sterilization standards: Validation per ISO 17665.
    • Comparison to predicate devices: The established performance and characteristics of the legally marketed predicate devices serve as a benchmark for demonstrating substantial equivalence.

    8. The Sample Size for the Training Set

    This is not applicable. There is no "training set" in the context of this device submission, which involves physical implants and abutments rather than machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set mentioned or implied for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181381
    Date Cleared
    2018-11-23

    (183 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    A.B Dental Device Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    A.B. DENTAL DEVICES® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants systems. Narrow platform implants are to be used with straight abutments only. The following components are being added:

    • . P0-3 narrow platform healing caps, 4 and 7 mm length
    • PON-3.75 standard platform, narrow healing caps, 4 and 7 mm length .
    • P3-5 standard platform, straight abutment, 5 mm diameter, 5 mm length ●
    • P3S-3 narrow platform, anatomic, straight abutment, 1, 2, 3 mm shoulder lengths
    • . P3SW-3.75 standard platform, anatomic, straight abutment, wide diameter, 1, 2, 3 mm shoulder lengths
    • P3W-3 narrow platform, straight abutment, wide diameter, 9 mm length
    • P5-3 narrow platform, ball attachment, 4, 5, 6 mm height
    • P5-3.75-20 standard platform, 20º ball attachment, 4 and 6 mm height ●
    • P9HG-3.75 standard platform, composed abutment, gold alloy
    • P9R and P9HR -3.75 standard platform, composed abutment, CoCr alloy
    • . P25-3 narrow platform, A.B. LOC abutments, 0, 1, 2, 3, 4 5 mm height
    • . P25-3.75 standard platform. A.B. LOC abutments. 0. 1. 2 mm height
    • . P64-3.75 standard platform, straight adaptor abutment, 1, 2, 3, 4, 5 mm heights
    • P64 plastic and Ti alloy sleeve
    • P64 healing cap ●
    • PK standard platform, anatomic anti-rotation abutment, 1, 2, 3, 4 mm height
    • PK plastic sleeve
    • . PK healing cap
    AI/ML Overview

    This document is a 510(k) premarket notification for the A.B. Dental Devices® Dental Implants System. It describes the device, its indications for use, and a comparison to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets those criteria in a typical medical device AI context.

    The "Performance Data" section lists several types of engineering analysis and biocompatibility/sterilization/fatigue testing, but these are for the physical dental implant system itself, not for an AI/algorithm-based diagnostic or assistive device. There is no mention of a "device performance" in terms of clinical metrics like sensitivity, specificity, accuracy, or an effect size for human readers.

    Therefore, for the specific questions asked about acceptance criteria and a study proving device performance (especially questions related to AI/algorithm performance), the provided text does not contain the requested information.

    Here's why and what's missing:

    1. A table of acceptance criteria and the reported device performance: Not present. The document focuses on comparing the new components (healing caps, abutments) to previously cleared predicate devices based on material, diameter, platform, length, design, and sterility. There are no performance metrics or acceptance criteria for a diagnostic/assistive function.

    2. Sample size used for the test set and the data provenance: Not applicable as there is no "test set" for an AI algorithm. The performance data mentioned relates to physical testing (e.g., fatigue testing), but not clinical trials or data for an algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no ground truth established for an AI algorithm.

    4. Adjudication method for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is not for an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI algorithm.

    7. The type of ground truth used: Not applicable.

    8. The sample size for the training set: Not applicable.

    9. How the ground truth for the training set was established: Not applicable.

    In summary, this document is a 510(k) submission for conventional dental implant components, not an AI-powered medical device. As such, it does not contain the information requested about acceptance criteria and studies proving the performance of an AI/algorithm-based device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1