K Number
K182455
Device Name
A.B. DENTAL DEVICES Dental Implants System
Date Cleared
2019-07-15

(311 days)

Product Code
Regulation Number
872.3630
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Description
A.B. DENTAL DEVICES® Dental Implants System consists of narrow (XXX-3 model number) and standard (XXX-3.75 model number) platform internal hex implants that are used with screw retained, cement retained, or overdenture abutments. This submission adds the following abutments: - Expansion of the P4 angled abutments to include: . - Narrow platform abutments (standard and long lengths, 15 and 25 degrees) o - Narrow platform, anatomic abutments (15 degrees) o - o Additional standard platform abutments (15 and 25 degrees) - Additional standard platform, anatomic abutments (15 and 25 degrees) O - . Expansion of the P5 ball attachments to include narrow platform, 20 degree abutments - Expansion of the P14 multiunit, angled abutments to include: - Additional sleeve materials (Delrin and CoCr) O - Narrow platform abutments (17 and 30 degrees) O - P25 AB LOC for angular adaptor O - Addition of P64 multiunit, straight and angled product line (narrow and standard platform, 17 ● and 30 degrees, with Delrin and Ti alloy sleeves), including its healing cap. - . Modification to the P4a-S screw to improve its performance characteristics
More Information

No
The summary describes a system of dental implants and abutments, focusing on mechanical properties, materials, and sterilization. There is no mention of AI or ML in the intended use, device description, or performance studies.

Yes

The device restores the patient's chewing function, which is a therapeutic effect.

No

The device description indicates it is a system of dental implants and abutments used for surgical and restorative applications to provide support for prosthetic devices. It is a treatment device, not a diagnostic one.

No

The device description explicitly details physical components (implants, abutments, screws, sleeves, healing caps) and the performance studies focus on mechanical and material properties, sterilization, and biocompatibility, all indicative of a hardware device. There is no mention of software as the primary or sole component.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the device is for surgical and restorative applications, placed in the bone to support prosthetic devices and restore chewing function. This is a direct intervention on the patient's body.
  • Device Description: The description details physical implants and abutments, which are surgically implanted components.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue, etc.) to provide information for diagnosis, monitoring, or treatment. IVDs are typically used in laboratories or point-of-care settings for analysis of biological samples.

The device described is a medical device used for surgical implantation and prosthetic support, not for in vitro diagnostic testing.

N/A

Intended Use / Indications for Use

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Product codes

NHA

Device Description

A.B. DENTAL DEVICES® Dental Implants System consists of narrow (XXX-3 model number) and standard (XXX-3.75 model number) platform internal hex implants that are used with screw retained, cement retained, or overdenture abutments. This submission adds the following abutments:

  • Expansion of the P4 angled abutments to include: .
    • Narrow platform abutments (standard and long lengths, 15 and 25 degrees) o
    • Narrow platform, anatomic abutments (15 degrees) o
    • o Additional standard platform abutments (15 and 25 degrees)
    • Additional standard platform, anatomic abutments (15 and 25 degrees) O
  • . Expansion of the P5 ball attachments to include narrow platform, 20 degree abutments
  • Expansion of the P14 multiunit, angled abutments to include:
    • Additional sleeve materials (Delrin and CoCr) O
    • Narrow platform abutments (17 and 30 degrees) O
    • P25 AB LOC for angular adaptor O
  • Addition of P64 multiunit, straight and angled product line (narrow and standard platform, 17 ● and 30 degrees, with Delrin and Ti alloy sleeves), including its healing cap.
  • . Modification to the P4a-S screw to improve its performance characteristics

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

bone of the upper or lower jaw

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

  • Engineering analysis to determine if additional components constitute a new mechanical worst-case.
  • Testing per ISO 14801 in air on the worst-case construct.
  • Static torque to failure testing of modified P4a-S screw. In addition, static and dynamic testing per ISO 14801 in the same construct was conducted.
  • . Scientific rationale to determine if additional components constitute a new cleaning or sterility worst-case
  • . Application of ISO 10993-1 to justify that biocompatibility testing was not needed as materials and manufacturing methods are identical to the sponsor's own predicate
  • Sterilization validation per ISO 17665

Key results: The A.B. Dental Implants System has same intended use and similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the A.B. Dental Implants Systems do not raise different issues. Performance data demonstrate that the A.B. Dental Implants System is substantially equivalent.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K162482 A.B. DENTAL DEVICES® Dental Implants System

Reference Device(s)

K131215 A.B. DENTAL DEVICES® Dental Implants System, K112440, DENTAL DEVICES® Dental Implants System, K111357 Straumann Crossfit SLA 3.3 mm

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, and the full name of the agency, "U.S. Food & Drug Administration," is in a smaller font size below the acronym.

July 15, 2019

A.B. Dental Device Ltd. % John Smith Partner Hogan Lovells US LLP 555 13th Street Washington, District of Columbia 20004

Re: K182455

Trade/Device Name: A.B. DENTAL DEVICES® Dental Implants System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 18, 2019 Received: June 18, 2019

Dear John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

510(k) Number (if known) K182455

Device Name

A.B. DENTAL DEVICES ® Dental Implants System

Indications for Use (Describe)

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

3

K182455 510(k) SUMMARY A.B. DENTAL DEVICES® Dental Implants System

Sponsor:

A.B. Dental Devices Ltd. 19 Havalomim St., Ashdod 7761117 lsrael

Contact Person:

Gabi Krauss RA Manager Phone: 972-8-8531388 Facsimile: 972-8-8522562

Date Prepared: July 15, 2019

Trade Name:A.B. DENTAL DEVICES® Dental Implants System
Common or Usual Name:Abutment, implant, dental, endosseous
Classification:21 CFR §872.3630 Endosseous dental implant abutment
Product Code:NHA

Predicate Devices

K162482 A.B. DENTAL DEVICES® Dental Implants System (primary) K131215 A.B. DENTAL DEVICES® Dental Implants System (reference) K112440, DENTAL DEVICES® Dental Implants System (reference) K111357 Straumann Crossfit SLA 3.3 mm (reference)

Device Description

A.B. DENTAL DEVICES® Dental Implants System consists of narrow (XXX-3 model number) and standard (XXX-3.75 model number) platform internal hex implants that are used with screw retained, cement retained, or overdenture abutments. This submission adds the following abutments:

  • Expansion of the P4 angled abutments to include: .
    • Narrow platform abutments (standard and long lengths, 15 and 25 degrees) o
    • Narrow platform, anatomic abutments (15 degrees) o
    • o Additional standard platform abutments (15 and 25 degrees)
    • Additional standard platform, anatomic abutments (15 and 25 degrees) O
  • . Expansion of the P5 ball attachments to include narrow platform, 20 degree abutments
  • Expansion of the P14 multiunit, angled abutments to include:
    • Additional sleeve materials (Delrin and CoCr) O
    • Narrow platform abutments (17 and 30 degrees) O
    • P25 AB LOC for angular adaptor O
  • Addition of P64 multiunit, straight and angled product line (narrow and standard platform, 17 ● and 30 degrees, with Delrin and Ti alloy sleeves), including its healing cap.
  • . Modification to the P4a-S screw to improve its performance characteristics

Indications for Use

4

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Technological Characteristics Comparison

K162482 IndicationsK182455 Indications
A.B. DENTAL DEVICES® Dental Implants SystemA.B. DENTAL DEVICES® Dental Implants System
is indicated for use in surgical and restorativeis indicated for use in surgical and restorative
applications for placement in the bone of the upperapplications for placement in the bone of the upper
or lower jaw to provide support for prostheticor lower jaw to provide support for prosthetic
devices, such as artificial teeth, in order to restoredevices, such as artificial teeth, in order to restore
the patient's chewing function. A.B. DENTALthe patient's chewing function. A.B. DENTAL
DEVICES® Dental Implants System is indicatedDEVICES® Dental Implants System is indicated
also for immediate loading when good primaryalso for immediate loading when good primary
stability is achieved and with appropriate occlusalstability is achieved and with appropriate occlusal
loading.loading.
Two Stage Implants: I22, I5, I55, I10.
P4 and P14 angled abutments are to be used only
with standard platform implants 3.5 mm in diameter
or larger.

The indications for the subject device and the indications cleared in K162482 are provided above. The only difference is the highlighted text which details a subset of AB Dental implants that were the subject of K162482 and compatibility limitations for the angled P4 and P14 abutments. The reference device Indications for Use do not include any component-specific additional language which would apply to the subject devices and therefore are not included in this comparison. In comparison to K162482, the subject device only differs in respect to indications related to the specific components contained within each submission. As K162482 dealt with implants (122, 15, l10) specific instruction related to these devices are contained in the indications but are omitted from this submission which does not modify these devices. The limitation to P4 and P14 abutments is also being removed, as this submission contains the required information to support use of angled abutments with narrow platform implants.

The purpose of this traditional 510(k) is to expand the current product line to include the new abutments listed below. No substantive changes are being made to the indications. Comparison tables for each modification are provided below.

Subject P4-3,(15 and 25)Predicate P4L-3.75,25
510(k) NumberSubjectK162482
MaterialTi-6Al-4V ELI with gold colored
anodizationTi-6Al-4V ELI
Diameter, mm4.23, 4.215.43

Cement Retained, Angled Abutments (P4) 1.

5

PlatformNarrowStandard
Angulation, º15, 2525
Length, mm8.3, 8.713.08
DesignAnti-rotationAnti-rotation, Long
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

Both standard length and long length narrow platform P4 abutments are being added as part of this 510(k). As their geometry is identical aside for differences in length, the standard length abutments are less worst-case than the long. Material, manufacturing methods (aside from color anodization), and sterility are identical between the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

Subject P4-3.75,(15 and 25)Predicate P4L-3.75,25
510(k) NumberSubjectK162482
MaterialTi-6Al-4V ELITi-6Al-4V ELI
Diameter, mm4.59, 4.685.43
PlatformStandardStandard
Angulation, º15, 2525
Length, mm9.05, 8.9513.08
DesignAnti-rotationAnti-rotation, Long
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

The two abutments have the same design, and differ only in length. As the subject abutments have a shorter length, they have a shorter moment arm and do not create a new worst-case. Material, manufacturing methods, and sterility are identical between the subject and predicate device.

Subject P4-3.75,15stPredicate P4-3.75,25st
510(k) NumberSubjectK162482
MaterialTi-6Al-4V ELITi-6Al-4V ELI
Diameter, mm4.74.2
PlatformStandardStandard
Angulation, º1525
Length, mm9.159.15
DesignAnti-rotation, StraightAnti-rotation, Straight
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

The geometries are identical except for the lower angulation, which does not present a new worstcase. Material, manufacturing methods, and sterility are identical between the subject and predicate device.

510(k) NumberSubject P4L-3,(15 and 25)Predicate P4L-3.75,25
SubjectSubjectK162482

6

MaterialTi-6Al-4V ELI with gold colored anodizationTi-6Al-4V ELI
Diameter, mm4.535.43
PlatformNarrowStandard
Angulation, º15, 2525
Length, mm13.3513.08
DesignAnti-rotation, LongAnti-rotation, Long
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

The subject device is similar to the predicate; however, small geometric differences, including a narrower platform, is addressed through bench testing. Material, manufacturing methods (aside from color anodization), and sterility are identical between the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

Subject P4S-3,15-(1,2,3)Predicate P4S-3.75,15-(1,3)
510(k) NumberSubjectK162482
MaterialTi-6Al-4V ELI with gold colored
anodizationTi-6Al-4V ELI
Diameter, mm4.1, 4.4, 4.44.3, 4.9
PlatformNarrowStandard
Angulation, º1515
Length, mm8.05, 9.2, 10.258.25, 9.8
Shoulder Height,
mm1, 2, 31, 3
DesignAnti-rotation, AnatomicAnti-rotation, Anatomic
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

The subject device is similar to the predicate; however, small geometric differences, including a narrower platform, is addressed through bench testing. Material, manufacturing methods (aside from color anodization), and sterility are identical between the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

Subject P4S-3.75,15-2Predicate P4S-3.75,15-(1,3)
510(k) NumberSubjectK162482
MaterialTi-6Al-4V ELITi-6Al-4V ELI
Diameter, mm4.54.3, 4.9
PlatformStandardStandard
Angulation, º1515
Length, mm9.48.25, 9.8
Shoulder Height, mm21, 3
DesignAnti-rotation, AnatomicAnti-rotation, Anatomic
LoadingSingle unitSingle unit

7

SterilityNon-SterileNon-Sterile

The addition of an intermediate shoulder length does not create a new worst-case. Material, manufacturing methods, and sterility are identical between the subject and predicate device.

Subject P4S-3.75,25-Predicate P4S-3.75,15-Predicate P4L-3.75,25
(1,2,3)(1, 3)
510(k)
NumberSubjectK132125K162482
MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI
Diameter,
mm5, 4.92, 4.834.3, 4.95.43
PlatformStandardStandardStandard
Angulation,
°251525
Length,
mm7.95, 9.1, 10.38.25, 9.813.08
Shoulder
Height, mm1, 2, 31,3N/A
DesignAnti-rotation, AnatomicAnti-rotation, AnatomicAnti-rotation, Long
LoadingSingle unitSingle unitSingle unit
SterilityNon-SterileNon-SterileNon-Sterile

These abutments are similar in design to the predicate P4S-3.75,15-(1,3) abutments and differ only in that they have a 25° angulation while the predicate abutments have 15° angulation. None-theless, an engineering rationale shows that the new abutments do not create a new-worst as compared to the previously cleared P4L-3.75,25 abutment. Specifically, while the angulation between the P4S-3.75,25 and the P4L-3.75,25 abutments are the subject P4S abutment is 2.5 mm shorter, and therefore has a shorter moment arm as compared to the predicate P4L-3.75,25 abutment. Material, manufacturing methods and sterility are identical between the subject and predicate device.

Subject P4SW-3.75,3Predicate P4L-3.75,25
510(k) NumberSubjectK162482
MaterialTi-6Al-4V ELITi-6Al-4V ELI
Diameter, mm5.35.43
PlatformStandardStandard
Angulation, º1525
Length, mm11.413.08
Shoulder Height, mm3N/A
DesignAnti-rotation, Anatomic, WideAnti-rotation, Long
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

8

An engineering rationale shows that the angulation and length of the subject abutment is less than predicate, resulting in a shorter moment arm. Further, the wide diameter abutment has more material than a standard diameter abutment. Therefore, this abutment does not create a new, worstcase. Material, manufacturing methods, and sterility are identical between the subject and predicate device.

Subject P5-3,20-(1.5,3,4,5)Predicate P5-3.75,20-(1,2,3,5)
510(k) NumberSubjectK132125
MaterialTi-6Al-4V ELI with gold colored
anodizationTi-6Al-4V ELI
PlatformNarrowStandard
Length, mm1.5, 3, 4, 51, 2, 3, 5
Angle, °2020
DesignBall AttachmentBall Attachment
LoadingMulti-unitMulti-unit
SterilityNon-SterileNon-Sterile

2. Overdenture Ball Attachment Abutments (P5)

The subject device is similar to the predicate; however, small geometric differences, including a narrower platform, is addressed through bench testing. Material, manufacturing methods (aside from color anodization), and sterility are identical between the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

3. Multiunit Angular Adaptor (P14)

| | Subject P14-3,17-(1 and
3) and P14-3,30-(1 and 3) | Predicate P14-
3.75,17-(1 and 3) and
P14-3.75,30-(1 and 3) | Predicate Straumann
Crossfit SLA 3.3 mm |
|------------------------|------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------|
| 510(k) Number | Subject | K132125 | K111357 |
| Material | Ti-6Al-4V ELI with gold
colored anodization | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform | Narrow | Standard | Narrow |
| Shoulder
Height, mm | 1,3 | 1,3 | 2.5, 4, 5.5 |
| Angle, ° | 17, 30 | 17, 30 | 17, 30 |
| Design | Multi-unit | Multi-unit | Multi-unit |
| Loading | Multi-unit | Multi-unit | Multi-unit |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |

Dynamic testing of the worst-case narrow platform construct shows the abutment is substantially equivalent to the Straumann Crossfit SLA 3.3 mm (K111357) which has similar dimensions and materials. Material, manufacturing methods and sterility (aside from color anodization) are identical between the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

9

Subject P14bPredicate P14bPredicate P14-bG
510(k) NumberSubjectK112440K132125
MaterialDelrinDelrin6019 Gold Alloy + Delrin
Connection
Method/
PlatformTaper fit+ screw /
P14 platformTaper fit+ screw /
P16 platformTaper fit+ screw /
P14 platform
Angle, °Mounts to 17°, 30° or
straight (0°) P16Mounts straight (0°) P16Mounts to 17°, 30° or
straight (0°) P16
Length, mm101010
DesignPlastic Sleeve for Angled
AdaptorPlastic Sleeve for
Straight AdaptorComposed Sleeve for
Angled Adaptor
LoadingMulti-unitMulti-unitMulti-unit
SterilityNon-SterileNon-SterileNon-Sterile

This sleeve has been previously cleared for use with the P16 straight abutment (cleared in K112440) and this submission seeks to expand its use with the angled P14 abutment. These sleeves are comparable to the P14-bG composed sleeves cleared for use with the P14 in K132125, except that the entire sleeve is made from plastic that is replaced with metal. Material, manufacturing methods, and sterility are identical for the subject and predicate device.

| | Subject P25-P14(1,2) | Predicate P5-P14(1,2) | Predicate P25-
3.75 |
|-----------------------------------|-------------------------------|----------------------------------------|------------------------|
| 510(k) Number | Subject | K132125 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Connection
Method/
Platform | Screws into
P14 platform | Screws into
P14 platform | Standard |
| Angle, ° | Mounts to 17, 30 | Mounts to 17, 30 | 0 |
| Length, mm | 1, 2 | 1, 2 | 3 - 7 |
| Design | AB LOC for Angular
Adaptor | Ball Attachment for Angular
Adaptor | AB LOC
Abutment |
| Loading | Multi-unit | Multi-unit | Multi-unit |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |

These abutments are similar in design to the predicate P5-P14 abutments in that both allow the P14 base to be used the proximal portion of another implant. For the predicate P5-P14 abutment, connection is made to a P5 abutment, while the subject device allows connection to a P25 abutment. The proximal geometry of the P25-P14 is identical to the proximal geometry of the P25 abutments cleared in K132125 and is compatible with the same class I exempt silicone caps. Material, manufacturing methods and sterility are identical between the subject and predicate device.

Single Unit Straight and Angled Adaptor (P64) 4.

Subject P64-3,(1,2,3)Predicate P16-3.75
-------------------------------------------

10

510(k) NumberSubjectK112440
MaterialTi-6Al-4V ELI with gold colored anodizationTi-6Al-4V ELI
PlatformNarrowStandard
Angle, °N/AN/A
Height1, 2, 31, 2, 3, 4, 5
DesignIntegrated screw, screw retainedIntegrated screw, screw retained
LoadingSingle unitSingle unit
SterilityNon-SterileNon-Sterile

Both systems contain an integrated screw to connect to standard platform implants and are made from Ti-6Al-4V ELI. The subject and predicate P64 have heights that vary from 1 - 3 mm. The subject device is contained within the same height range as the predicate and does not create a new worst-case. Material (aside from color anodization), manufacturing methods, and sterility are identical for the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

| | Subject P64-3,17-(0.5
and 2) and P64-3,30-(0.5
and 2) | Predicate Straumann
Crossfit SLA 3.3 mm |
|------------------------|-------------------------------------------------------------|--------------------------------------------|
| 510(k) Number | Subject | K111357 |
| Material | Ti-6Al-4V ELI with gold
colored anodization | Ti-6Al-4V ELI |
| Platform | Narrow | Narrow |
| Shoulder
Height, mm | 0.5, 2 | 2.5, 4, 5.5 |
| Angle, ° | 17, 30 | 17,30 |
| Design | Single-unit | Multi-unit |
| Loading | Multi-unit | Multi-unit |
| Sterility | Non-Sterile | Non-Sterile |

Dynamic testing of the worst-case narrow platform construct shows that the narrow platform abutment is substantially equivalent to the Straumann Crossfit SLA 3.3 mm (K111357) which has similar dimensions and materials. Material, manufacturing methods (aside from color anodization) and sterility are identical between the subject and predicate device. The same color anodization has been cleared on other abutments in K162482.

| | Subject P64-3.75,17-(0.5, 2, and 3) and
P64-3.75,30-(0.5, 2, and 3) | Predicate P14-3.75,17-(1 and 3) and
P14-3.75,30-(1 and 3) |
|------------------------|------------------------------------------------------------------------|--------------------------------------------------------------|
| 510(k) Number | Subject | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform | Standard | Standard |
| Shoulder
Height, mm | 0.5, 2, 3 | 1,3 |
| Angle, ° | 17, 30 | 17,30 |

11

DesignSingle-unitMulti-unit
LoadingMulti-unitMulti-unit
SterilityNon-SterileNon-Sterile

These abutments are similar in design to the predicate P14 abutments except that they are monolithic single-units while the predicate P14 abutments are composed of two different parts. Both abutments are intended for multi-unit loading and have the same maximum shoulder height and angulations. Dynamic testing of the worst-case narrow platform construct shows the P64-3.75 to be equivalent to the standard platform P14 abutment. Material, manufacturing methods and sterlity are identical between the subject and predicate device.

Subject P64bPredicate P14b
510(k) NumberSubjectK112440
MaterialDelrinDelrin
Connection
Method/ PlatformTaper fit+ screw /
P64 platformTaper fit+ screw /
P16 platform
Angle, °Mounts to 17°, 30° or straight
(0°)Mounts to straight (0°) P16
Length, mm1010
DesignPlastic Sleeve for Angled
AdaptorPlastic Sleeve for Straight Adaptor
LoadingMulti-unitMulti-unit
SterilityNon-SterileNon-Sterile

The two sleeves primarily differ in their distal geometry so that they fit their respective compatible abutments. Material, manufacturing methods and sterility are identical for the subject and predicate device.

Subject P64-bTPredicate P14-bT
510(k) NumberSubjectK132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
Connection
Method/ PlatformTaper fit+ screw /
P64 platformTaper fit+ screw /
P14 platform
Angle, °Mounts to 17°, 30° or straight (0°)Mounts to 17°, 30° or straight (0°)
P16
Length, mm1212
DesignMetal Sleeve for Angled AdaptorMetal Sleeve for Angled Adaptor
LoadingMulti-unitMulti-unit
SterilityNon-SterileNon-Sterile

The two sleeves primarily differ in their distal geometry so that they fit their respective compatible abutments. Its addition is supported by the same information as the P64 abutment itself, as the two are evaluated together via dynamic testing. Material, manufacturing methods and sterility are identical for the subject and predicate device.

12

Subject P0-P64,5Predicate P0-P14,(5 and 7)
510(k) NumberSubjectK132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
ConnectionP64P14/P16
Method/Platform
Angle, °N/AN/A
Height55,7
DesignTemporary Healing CapTemporary Healing Cap
SterilityNon-SterileNon-Sterile

The caps differ with respect to their abutment mating geometry to match their respective abutments. The subject cap is only available in 5 mm height, while the predicate is available in additional shorter and longer heights. Material, manufacturing methods, and sterility are identical for the subject and predicate device.

Performance Data

  • Engineering analysis to determine if additional components constitute a new mechanical worst-case.
  • Testing per ISO 14801 in air on the worst-case construct.
  • Static torque to failure testing of modified P4a-S screw. In addition, static and dynamic testing per ISO 14801 in the same construct was conducted.
  • . Scientific rationale to determine if additional components constitute a new cleaning or sterility worst-case
  • . Application of ISO 10993-1 to justify that biocompatibility testing was not needed as materials and manufacturing methods are identical to the sponsor's own predicate
  • Sterilization validation per ISO 17665

Substantial Equivalence

The A.B. Dental Implants System has same intended use and similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the A.B. Dental Implants Systems do not raise different issues. Performance data demonstrate that the A.B. Dental Implants System is substantially equivalent.

Conclusions

The A.B. Dental Implants System is substantially equivalent to the predicate device.