KWS · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Orthopedic · 21 CFR 888.3660 · Class 2
Overview
| Product Code | KWS |
|---|---|
| Device Name | Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented |
| Regulation | 21 CFR 888.3660 |
| Device Class | Class 2 |
| Review Panel | Orthopedic |
| Implant | Yes |
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Classification Rationale
Class II. The special controls for this device are:
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Cleared Devices (190)
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K254084 | AETOS Shoulder System - CONCELOC Glenoids | Smith & Nephew, Inc. | Apr 6, 2026 | SESE |
| K252418 | Catalyst F1x Shoulder System | Catalyst Orthoscience, Inc. | Nov 6, 2025 | SESE |
| K252416 | AETOS Shoulder System Meta Humeral Prosthesis Size 0 | Smith & Nephew, Inc. | Oct 29, 2025 | SESE |
| K241292 | MyShoulder Planner (5.3SSWPL) | Medacta International S.A. | Jan 30, 2025 | SESE |
| K241878 | Tornier Humeral Reconstruction System (Tornier HRS); Tornier Perform Humeral System - Stem (Tornier PHS-Stem) | Stryker Corporation (Tornier, Inc.) | Dec 6, 2024 | SESE |
| K241609 | Tornier Humeral Reconstruction System (Tornier HRS) | Stryker Corporation (Tornier, Inc.) | Sep 10, 2024 | SESE |
| K230513 | Arthrex Univers Apex OptiFit Humeral Stem | Arthrex, Inc. | Sep 26, 2023 | SESE |
| K222592 | AltiVate® Anatomic Shoulder AG e+ with Markers | Encore Medical L.P. | Jun 23, 2023 | SESE |
| K230572 | AETOS Shoulder System | Smith & Nephew, Inc. | Jun 7, 2023 | SESE |
| K220847 | AETOS Shoulder System | Smith & Nephew, Inc. | Jan 3, 2023 | SESE |
| K222317 | Catalyst EA Convertible Stemmed Shoulder | Catalyst Orthoscience, Inc. | Nov 10, 2022 | SESE |
| K213387 | AltiVate® Anatomic Shoulder AG e+ with Markers | Encore Medical L.P. | Jul 7, 2022 | SESE |
| K213785 | SMR TT Hybrid Glenoid, SMR 3-Pegs Cemented Glenoid | Limacorporate | May 13, 2022 | SESE |
| K213827 | TITAN Total Shoulder System (TSS), TITAN Reverse Shoulder System (RSS), TITAN Humeral Resurfacing Arthroplasty System (HRA), Modular Radial Head System (MRH), Katalyst Bipolar Radial Head System (Katalyst) | Smith & Nephew, Inc. | Mar 11, 2022 | SESE |
| K212356 | Exactech® Equinoxe® Laser Cage Glenoid | Exactech, Inc. | Mar 4, 2022 | SESE |
| K212435 | Comprehensive Humeral Fracture Positioning Sleeves | Biomet, Inc. | Dec 10, 2021 | SESE |
| K210861 | ReUnion Reversible Fracture System (RFX), ReUnion Reverse Shoulder Arthroplasty System (RSA), ReUnion Total Shoulder Arthroplasty System (TSA) | Stryker GmbH | Sep 3, 2021 | SESE |
| K210050 | Augmented VaultLock Glenoid | Arthrex, Inc. | Jun 7, 2021 | SESE |
| K203315 | BLUEPRINT Patient Specific Instrumentation | Tornier S.A.S. | Apr 15, 2021 | SESE |
| K203230 | Global Shoulder and Delta CTA systems | Depuy Lnc. | Apr 2, 2021 | SESE |
| K203026 | AltiVate® Anatomic Pegged Glenoid with Markers | Encore Medical L.P. | Dec 23, 2020 | SESE |
| K202289 | ReUnion Reversible Fracture System (RFX), ReUnion Reverse Shoulder Arthroplasty System (RSA), ReUnion Total Shoulder Arthroplasty System (TSA) | Stryker GmbH | Dec 16, 2020 | SESE |
| K193180 | Alliance Augmented Glenoid | Zimmer, Inc. | Feb 24, 2020 | SESE |
| K191247 | SmartSPACE Shoulder System | Techmah Medical, LLC | Nov 15, 2019 | SESE |
| K191814 | Alliance Glenoid | Zimmer, Inc. | Oct 23, 2019 | SESE |
| K190286 | Materialise Glenoid Positioning System | Materialise NV | Jul 8, 2019 | SESE |
| K191318 | AEQUALIS FLEX REVIVE Shoulder System | Tornier, Inc. | Jun 14, 2019 | SESE |
| K190521 | Aequalis Ascend Flex Shoulder System | Tornier S.A.S. | Jun 12, 2019 | SESE |
| K190290 | AltiVate Reverse Humeral Stem, 48mm | Encore Medical L.P. | Apr 26, 2019 | SESE |
| K182799 | Arthrex Univers II Shoulder Prosthesis System: Titanium Humeral Heads | Arthrex, Inc. | Mar 8, 2019 | SESE |
| K183039 | ReUnion Reversible Fracture System (RFX), ReUnion Reverse Shoulder Arthroplasty System (RSA), ReUnion Total Shoulder Arthroplasty System (TSA) | Stryker GmbH | Feb 15, 2019 | SESE |
| K183042 | SMR Lateralized Connectors with screws | Lima Corporate S.P.A. | Jan 25, 2019 | SESE |
| K181555 | Arthrex Fracture Adapter Hemi Shoulder Prosthesis | Arthrex, Inc. | Dec 21, 2018 | SESE |
| K181742 | ReUnion Reversible Fracture System (RFX), ReUnion Reverse Shoulder Arthroplasty System (RSA), ReUnion Total Shoulder Arthroplasty System (TSA) | Stryker GmbH | Sep 27, 2018 | SESE |
| K181420 | Aequalis Flex Revive Shoulder System | Tornier, Inc. | Sep 20, 2018 | SESE |
| K171841 | Arthrex Shoulder System | Arthrex, Inc. | Mar 29, 2018 | SESE |
| K172054 | Materialise Glenoid Positioning System | Materialise NV | Nov 3, 2017 | SESE |
| K172351 | AltiVate Reverse Humeral Stem, AltiVate Reverse Small Spacer, Altivate Reverse, Small Hemi-Adapter, AltiVate Reverse, Small Socket Insert | Encore Medical L.P. | Oct 30, 2017 | SESE |
| K172210 | ReUnion Reversible Fracture System (RFX), ReUnion Reverse Shoulder Arthroplasty System (RSA), ReUnion Total Shoulder Arthroplasty System (TSA) | Stryker Trauma AG | Oct 2, 2017 | SESE |
| K170910 | Medacta Anatomic Shoulder Prosthesis | Medacta International S.A. | Aug 16, 2017 | SESE |
| K170893 | Materialise Glenoid Positioning System | Materialise NV | Jun 19, 2017 | SESE |
| K162800 | BLUEPRINT Patient Specific Instrumentation | Tornier S.A.S. | Feb 22, 2017 | SESE |
| K162697 | Arthrex Glenoid Intelligent Reusable Instrument System, Arthrex OrthoVis Preoperative Plan, SmartBase for Arthrex Glenoid IRIS, ArthrexVIP Web Portal | Custom Orthopaedic Solutions, Inc. | Feb 10, 2017 | SESE |
| K161789 | BLUEPRINT Patient Specific Instrumentation | Tornier S.A.S. | Dec 27, 2016 | SESE |
| K161863 | ReUnion Reversible Fracture System (RFX), ReUnion Reverse Shoulder Arthroplasty (RSA), ReUnion Total Shoulder Arthroplasty (TSA) | Stryker GmbH | Nov 23, 2016 | SESE |
| K162024 | AltiVate Anatomic Shoulder System | Encore Medical L.P. | Nov 21, 2016 | SESE |
| K161108 | Arthrex VaultLock Glenoid | Arthrex, Inc. | Aug 29, 2016 | SESE |
| K160975 | Aequalis PerFORM+ Shoulder System | Tornier, Inc. | Jun 10, 2016 | SESE |
| K160555 | BLUEPRINT Patient Specific Instrumentation | Tornier S.A.S. | Jun 10, 2016 | SESE |
| K153602 | Materialise Glenoid Positioning System | Materialise NV | Apr 26, 2016 | SESE |
| K153722 | SMR 3-Pegs Glenoids | Lima Corporate S.P.A. | Apr 6, 2016 | SESE |
| K160085 | Anatomical Shoulder System, Anatomical Shoulder Domelock System, Anatomical Shoulder Fracture System | Zimmer GmbH | Mar 8, 2016 | SESE |
| K153215 | SmartBase for Arthrex Glenoid IRIS | Custom Orthopaedic Solutions, Inc. | Feb 19, 2016 | SESE |
| K152047 | INTEGRA TITAN Modular Total Shoulder System Fin-Lock Glenoid | Ascension Orthopedics | Dec 23, 2015 | SESE |
| K152267 | Remedy® Shoulder Spacer | Osteoremedies, LLC | Dec 15, 2015 | SESE |
| K153115 | Arthrex Univers Apex, Size 5 Stem | Arthrex, Inc. | Nov 23, 2015 | SESE |
| K151500 | Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) | Custom Orthopaedic Solutions | Oct 5, 2015 | SESE |
| K151293 | Aequalis Ascend Flex Shoulder System, Aequalis Reversed Shoulder Prosthesis | Tornier S.A.S. | Sep 24, 2015 | SESE |
| K143256 | SMR Modular Glenoid | Lima Corporate S.P.A. | May 28, 2015 | SESE |
| K150583 | Aequalis PerFORM+ Shoulder System | Tornier, Inc. | Apr 23, 2015 | SESE |
| K143374 | BLUEPRINT Patient Specific Instrumentation | Tornier S.A.S. | Apr 8, 2015 | SESE |
| K142403 | Anatomical Shoulder System/Anatomical Shoulder Domelock System/Anatomical Shoulder Fracture System | Zimmer GmbH | Nov 24, 2014 | SESE |
| K142072 | GLENOID INTELLIGENT REUSABLE INSTRUMENT SYSTEM (GLENOID IRIS) | Custom Orthopaedic Solutions, Inc. | Nov 24, 2014 | SESE |
| K142413 | TITAN Modular Total Shoulder System | Ascension Orthopedics | Nov 14, 2014 | SESE |
| K133349 | SMR TT METAL BACK GLENOID | Lima Corporate S.P.A. | Jun 3, 2014 | SESE |
| K133834 | DEPUY GLOBAL UNITE SHOULDER SYSTEM | Depuy(Ireland) | Apr 3, 2014 | SESE |
| K140063 | EQUINOXE EXTRA SHORT HUMERAL HEAD | Exactech, Inc. | Feb 11, 2014 | SESE |
| K133462 | DUOCENTRIC REVERSED SHOULDER PROSTHESIS | Aston Medical | Feb 7, 2014 | SESE |
| K130048 | REVERSE SHOULDER PROSTHESIS MONOBLOCK | Encore Medical L.P. | Oct 29, 2013 | SESE |
| K131633 | ARTHREX UNIVERS APEX | Arthrex, Inc. | Sep 27, 2013 | SESE |
| K131559 | MATCH POINT SYSTEM(TM), SURGICASE CONNECT, MATCH POINT SYSTEM(TM) GUIDES | Materialise NV | Sep 27, 2013 | SESE |
| K131277 | SEVIIN FRACTURE SHOULDER | Ingen Orthopedics, LLC | Aug 28, 2013 | SESE |
| K131129 | CAS PSI SHOULDER | Zimmer Cas | Aug 20, 2013 | SESE |
| K130050 | TITAN REVERSE SHOULDER SYSTEM | Ascension Orthopedic | Jun 18, 2013 | SESE |
| K130642 | SMR 3-PEGS GLENOIDS | Lima Corporate S.P.A. | Jun 12, 2013 | SESE |
| K131112 | LIMACORPORATE SMR CTA HUMERAL HEADS | Lima Corporate S.P.A. | Jun 7, 2013 | SESE |
| K130129 | UNIVERS REVERS SHOULDER PROSTHESIS SYSTEM | Arthrex, Inc. | May 31, 2013 | SESE |
| K123122 | GLENOID INTELLIGENT REUSABLE INSTRUMENT SYSTEM (GLENOID IRIS) | Cleveland Clinic | Apr 5, 2013 | SESE |
| K123982 | TURON GLENOID PEGGED, E+, TURON KEELED GLENOID, E+ | Encore Medical L.P. | Jan 23, 2013 | SESE |
| K122698 | AEQUALIS ASCEND FLEX SHOULDER SYSTEM | Tornier S.A.S. | Jan 8, 2013 | SESE |
| K121220 | EQUINOXE GLENOID, UHMWPE 16 PS=OSTERIOR AUGMENT PEGGED, SMALL, LEFT/RIGHT | Exactech, Inc. | Nov 20, 2012 | SESE |
| K121493 | ASCEND SHOULDER SYSTEM | Tornier, Inc. | Jun 20, 2012 | SESE |
| K120794 | ASCEND SHOULDER SYSTEM | Tornier, Inc. | Apr 13, 2012 | SESE |
| K112438 | TITAN MODULAR TOTAL SHOULDER SYSTEM | Ascension Orthopedic | Apr 11, 2012 | SESE |
| K113413 | ASCEND SHOULDER SYSTEM | Tornier, Inc. | Mar 9, 2012 | SESE |
| K120044 | ARTHREX UNIVERSII XL GLENOID-PEGGED, ARTHREX UNIVERS II XL GLENOID-KEELED | Arthrex, Inc. | Mar 8, 2012 | SESE |
| K113012 | PROMOS MODULAR SHOULDER SYSTEM | Smith & Nephew, Inc. | Jan 5, 2012 | SESE |
| K113309 | EXACTECH EQUINOXE CAGE GLENOIDS | Exactech, Inc. | Dec 8, 2011 | SESE |
| K111902 | AEQUALIS SHOULDER SYSTEM | Tornier | Nov 3, 2011 | SESE |
| K111212 | SMR SHOULDER SYSTEM- REVISION STEMS | Lima Corporate S.P.A. | Jul 28, 2011 | SESE |
| K110847 | SMR SHOULDER SYSTEM - CTA HUMERAL HEADS | Lima Corporate S.P.A. | Jul 15, 2011 | SESE |
| K111596 | TOTAL SHOULDER SYSTEM | Shoulder Innovations, LLC | Jul 5, 2011 | SESE |
| K111379 | EQUINOXE UHMWPE 12* POSTERIOR AUGMENT PEGGED GLENOID, SMALL / MEDIUM / LARGE / EXTRA-LARGE, LEFT / RIGHT | Exactech, Inc. | Jun 8, 2011 | SESE |
| K110905 | TOTAL SHOULDER SYSTEM | Shoulder Innovations, LLC | May 24, 2011 | SESE |
| K103835 | REUNION TSA SYSTEM-HUMERAL PRESS-FIT & CEMENTED STEMS MODEL 5569-YY-20XX (YY=P,C; XX=07-17,06L-12L) REUNION TSA SYSTEM N | Howmedica Osteonics Corp. | May 5, 2011 | SESE |
| K110865 | AEQUALIS ASCEND MODULAR ANATOMIC SHOULDER SYSTEM | Tornier, Inc. | Apr 27, 2011 | SESE |
| K102924 | AEQUALIS ASCEND MODULAR ANATOMIC SHOULDER SYSTEM | Tornier, Inc. | Jan 31, 2011 | SESE |
| K102670 | TOTAL SHOULDER SYSTEM | Shoulder Innovations, LLC | Jan 24, 2011 | SESE |
| K103466 | ARTHREX UNIVERS II SHOULDER PROSTHESIS | Arthrex, Inc. | Dec 22, 2010 | SESE |
| K103419 | EXACTECH EQUINOXE R UHMWPE POSTERIOR AUGMENT PEGGED GLENOIDS TM | Exactech, Inc. | Dec 13, 2010 | SESE |
| K093599 | ARROW ANATOMICAL SHOULDER SYSTEM | Fournitures Hospitalieres Industrie | Nov 24, 2010 | SESE |
| K103104 | SPECIAL ORTHOPAEDIC SOLUTIONS GLENOID SHOULDER COMPONENT | Howmedica Osteonics Corp. | Nov 22, 2010 | SESE |
| K103007 | AFFINITI TOTAL AND HEMI-SHOULDER SYSTEM | Tornier, Inc. | Nov 9, 2010 | SESE |
| K093430 | EXACTECH EQUINOXE XL KEELED GLENOID, PEGGED GLENOID, EXACTECH EQUINOXE CAGE, GLENOID CONTD. | Exactech, Inc. | Sep 2, 2010 | SESE |
| K100448 | ASCENSION MODULAR TOTAL SHOULDER SYSTEM | Ascension Orthopedics, Inc. | Aug 30, 2010 | SESE |
| K091196 | GRS GLENOID RESURFACING SYSTEM | Arthrosurface, Inc. | Oct 27, 2009 | SESE |
| K092122 | GLOBAL SHOULDER STEPTECH ANCHOR PEG GLENOID | DePuy Orthopaedics, Inc. | Jul 27, 2009 | SESE |
| K091751 | DEPUY DELTA XTEND REVERSE SHOULDER XLK CUP | DePuy Orthopaedics, Inc. | Jul 14, 2009 | SESE |
| K083435 | ARTHREX UNIVERS II SHOULDER PEGGED GLENOID | Arthrex, Inc. | Jun 26, 2009 | SESE |
| K082120 | AEQUALIS REVERSED FRACTURE SHOULDER PROTHESIS | Tornier | Oct 24, 2008 | SESE |
| K081707 | AEQUALIS SHOULDER FRACTURE AND SHOULDER SYSTEMS | Tornier, Inc. | Jul 17, 2008 | SESE |
| K081620 | MODIFICATION TO DEPUY DELTA XTEND REVERSE SHOULDER SYSTEM | DePuy Orthopaedics, Inc. | Jun 30, 2008 | SESE |
| K080402 | ENCORE SHOULDER PRIMARY HUMERAL STEM, MODEL 500-00, ENCORE SHOULDER HUMERAL HEAD, MODEL 520-38 | Encore Medical L.P. | Mar 28, 2008 | SESE |
| K071032 | ARTHREX UNIVERS II SHOULDER PROSTHESIS | Arthrex, Inc. | Aug 20, 2007 | SESE |
| K071090 | TRABECULAR METAL GLENOID- B/F COMPLETE SHOULDER SOLUTION | Zimmer Trabecular Metal Technology | Jun 29, 2007 | SESE |
| K060694 | MODULAR HYBRID GLENOID | Biomet Manufacturing Corp | Dec 11, 2006 | SESE |
| K060988 | TOTAL AND HEMI SHOULDER SYSTEM | Dvo Extremity Solutions, LLC | Dec 6, 2006 | SESE |
| K062029 | ANATOMICAL SHOULDER FRACTURE SYSTEM | Zimmer GmbH | Oct 31, 2006 | SESE |
| K062113 | SOLAR REUNION FRACTURE STEM | Howmedica Osteonics Corp. | Oct 3, 2006 | SESE |
| K060874 | DEPUY GLOBAL AP SHOULDER SYSTEM | DePuy Orthopaedics, Inc. | Jun 28, 2006 | SESE |
| K060209 | AEQUALIS SHOULDER FRACTURE SYSTEM & AEQUALIS SHOULDER SYSTEM | Tornier | Mar 2, 2006 | SESE |
| K052472 | DEPUY GLOBAL SHOULDER CROSSLINK GLENOID | DePuy Orthopaedics, Inc. | Oct 6, 2005 | SESE |
| K051623 | ANATOMICAL SHOULDER KEELED GLENOID | Zimmer GmbH | Jul 19, 2005 | SESE |
| K043346 | GENESIS TOTAL SHOULDER REPLACEMENT, MODEL 100-01-00 | Genesis Medical LLC | Jan 31, 2005 | SESE |
| K043077 | MODIFICATION TO AEQUALIS SHOULDER FRACTURE SYSTEM; AEQUALIS SHOULDER SYSTEM | Tornier | Dec 17, 2004 | SESE |
| K043061 | THE TRABECULAR METAL GLENOID - BIGLIANI/FLATOW THE COMPLETE SHOULDER SOLUTION, MODEL 4316-XX-XX | Zimmer Trabecular | Nov 23, 2004 | SESE |
| K042021 | EXACTECH EQUINOXE SHOULDER SYSTEM | Exactech, Inc. | Oct 26, 2004 | SESE |
| K041339 | AEQUALIS SHOULDER SYSTEM | Tornier | Jun 17, 2004 | SESE |
| K040610 | VERSA-DIAL HUMERAL HEAD | Biomet Manufacturing, Inc. | Jun 4, 2004 | SESE |
| K033506 | COMPREHENSIVE HUMERAL FRACTURE POSITIONING SLEEVE | Biomet, Inc. | Feb 2, 2004 | SESE |
| K032679 | AEQUALIS SHOULDER FRACTURE SYSTEM | Tornier | Dec 15, 2003 | SESE |
| K031449 | MODIFICATION TO THE TRABECULAR METAL GLENOID - BIGLIANI/FLATOW THE COMPLETE SHOULDER SOLUTION | Implex Corp. | Sep 23, 2003 | SESE |
| K030259 | AMATOMICAL SHOULDER WITH REMOVABLE HEAD | Centerplus Orthopedics, Inc. | Apr 24, 2003 | SESE |
| K030344 | MODULAR NEER 3 HUMERAL STEMS AND MODULAR NEER 3 HUMERAL HEADS | Smith & Nephew, Inc. | Mar 4, 2003 | SESE |
| K020345 | ARTHREX UNIVERS FRACTURE PROSTHESIS | Arthrex, Inc. | Sep 6, 2002 | SESE |
| K012788 | VAPORIZED HYDROGEN PEROXIDE (VHP) STERILIZATION OF UNMWPE COMPONENTS IN THE COFIELD TOTAL SHOULDER SYSTEM, THE NEER II | Smith & Nephew, Inc. | Mar 15, 2002 | SESE |
| K012378 | 3D HUMERAL HEADS | Smith & Nephew, Inc. | Dec 13, 2001 | SESE |
| K003324 | PLASMA SPRAYED SHOULDER | Encore Orthopedics, Inc. | Dec 6, 2001 | SESE |
| K012212 | MODIFICATION TO AEQUALIS SHOULDER SYSTEM | Tornier | Aug 15, 2001 | SESE |
| K010124 | ARTHREX UNIVERS 30 SHOULDER PROSTHESIS | Arthrex, Inc. | Apr 6, 2001 | SESE |
| K003801 | ANATOMICAL PRESS-FIT HUMERAL STEM | Sulzer Orthopedics, Inc. | Feb 28, 2001 | SESE |
| K003566 | COFIELD ECCENTRIC AND LATERAL OFFSET HUMERAL HEADS | Smith & Nephew, Inc. | Feb 14, 2001 | SESE |
| K003832 | SULZER ORTHOPEDICS ANATOMICAL SHOULDER CEMENTED HUMERAL STEM | Sulzer Orthopedics, Inc. | Jan 4, 2001 | SESE |
| K003728 | MODIFICATION TO AEQUALIS SHOULDER FRACTURE SYSTEM | Tornier | Dec 20, 2000 | SESE |
| K002683 | WMT ULTRA-FIT MODULAR SHOULDER SYSTEM | Wrightmedicaltechnologyinc | Nov 22, 2000 | SESE |
| K001706 | WMT MODULAR SHOULDER SYSTEM | Wrightmedicaltechnologyinc | Aug 16, 2000 | SESE |
| K994393 | AEQUALIS UNIVERSAL SHOULDER GLENOID | Tornier | Jun 22, 2000 | SESE |
| K001046 | BIOPRO GLENOID COMPONENT W/ KEEL & W/ POSTS | Biopro, Inc. | Jun 8, 2000 | SESE |
| K994350 | NEER III TOTAL SHOULDER SYSTEM | Smith & Nephew, Inc. | May 23, 2000 | SESE |
| K994392 | AEQUALIS SHOULDER FRACTURE SYSTEM | Tornier | Mar 10, 2000 | SESE |
| K000575 | GLOBAL ADVANTAGE EXTENDED HUMERAL HEAD | DePuy Orthopaedics, Inc. | Mar 10, 2000 | SESE |
| K992899 | BIO-MODULAR SHOULDER OFFSET HUMERAL HEADS | Biomet, Inc. | Oct 21, 1999 | SESE |
| K992119 | BIO-MODULAR SHOULDER SYSTEM | Biomet, Inc. | Sep 13, 1999 | SESE |
| K980244 | AEQUALIS SHOULDER PROSTHESIS | Tornier | Aug 23, 1999 | SESE |
| K992065 | GLOBAL ADVANTAGE SHOULDER, GLOBAL ADVANTAGE HUMERAL STEM, GLOBAL ADVANTAGE ECCENTRIC HEAD | DePuy Orthopaedics, Inc. | Jul 12, 1999 | SESE |
| K990598 | OSTEONICS SOLAR SHOULDER HUMERAL BEARING HEAD | Howmedica Osteonics Corp. | Mar 15, 1999 | SESE |
| K990136 | ANATOMICA GLENOID COMPONENT | Sulzer Orthopedics, Inc. | Mar 1, 1999 | SESE |
| K990137 | ANATOMICA HUMERAL STEM/HEAD COMPONENTS | Sulzer Orthopedics, Inc. | Feb 18, 1999 | SESE |
| K981487 | DEPUY GLOBAL SHOULDER GLENOID | DePuy Orthopaedics, Inc. | Jul 13, 1998 | SESE |
| K974044 | GLOBAL TOTAL SHOULDER ECCENTRIC HUMERAL HEAD | DePuy Orthopaedics, Inc. | Jan 22, 1998 | SESE |
| K971715 | SELECT SHOULDER OFFSET HUMERAL HEAD COMPONENTS | Sulzer Orthopedics, Inc. | Aug 7, 1997 | SESE |
| K962224 | SELECT SHOULDER CONCENTRIC HUMERAL HEAD COMPONENTS | Intermedics Orthopedics | Aug 28, 1996 | SESE |
| K962082 | OSTEONICS ALL POLYETHYLENE GLENOID SHOULDER COMPONENT | Osteonics Corp. | Aug 13, 1996 | SESE |
| K960906 | FOUNDATION GLENOID COMPONENT | Encore Orthopedics, Inc. | Apr 15, 1996 | SESE |
| K955944 | OSTEONICS N2/VAC PACKAGING PROCESS | Osteonics Corp. | Apr 9, 1996 | SESE |
| K955767 | COFIED MODULAR SHOULDER SYSTEM | Smith & Nephew Richards, Inc. | Apr 8, 1996 | SN |
| K955731 | OSTEONICS SHOULDER HUMERAL COMPONENTS | Osteonics Corp. | Mar 5, 1996 | SN |
| K952928 | AEQUALIS SHOULDER SYSTEM FOR TOTAL & HEMISHOULDER APPLICATIONS | Tornier | Sep 27, 1995 | SESE |
| K950521 | OSTEONICS ALL POLYETHYLENE GLENOID SHOULDER COMPONENT | Osteonics Corp. | Jun 16, 1995 | SESE |
| K950651 | TOTAL SHOULDER PROSTHESIS | Encore Orthopedics, Inc. | May 30, 1995 | SN |
| K935802 | SELECT(R) SHOULDER ALL POLY GLENOID | Intermedics Orthopedics | May 22, 1995 | SESE |
| K943300 | DEPUY POROUS COATED GLOBAL SHOULDER | Depuy, Inc. | Oct 31, 1994 | SN |
| K920362 | 3M MODULAR SHOULDER SYSTEM | 3M Health Care, Ltd. | Jul 22, 1993 | SESE |
| K904880 | BUECHEL-PAPPAS HUMERAL STEM COMPONENT | Endotec, Inc. | Jan 28, 1991 | SESE |
| K904879 | BEUCHEL-PAPPAS GLENOID COMPONENT | Endotec, Inc. | Jan 28, 1991 | SESE |
| K903886 | OSTEONICS SR GLENOID COMPONENTS | Osteonics Reconstructive Products Div. | Nov 20, 1990 | SN |
| K902855 | NEW JERSEY TOTAL SHOULDER GLENOID COMPONENT | Depuy, Inc. | Sep 20, 1990 | SESE |
| K901434 | PAPILLOTOME | Telemed Systems, Inc. | Jun 8, 1990 | SESE |
| K895226 | NEER II TOTAL SHOULDER SYSTEM & PROSTHESIS | 3M Company | Nov 3, 1989 | SESE |
| K884943 | INTERMEDICS SELECT SHOULDER METAL-BACKED GLENOID | Intermedics Orthopedics | Apr 27, 1989 | SN |
| K883816 | SHOULDER MICROSTRUCTURED HUMERAL PROSTHESIS | Osteonics Corp. | Nov 10, 1988 | SESE |
| K881750 | NORMAL HUMERAL PROSTHESIS | Osteonics Corp. | Aug 30, 1988 | SESE |
| K881836 | INTERMEDICS SELECT SHOULDER ALL PLASTIC GLEN. COMP | Intermedics Orthopedics | Aug 29, 1988 | SESE |
| K870262 | FENLIN II SHOULDER PROSTHESIS | Zimmer, Inc. | Apr 3, 1987 | SESE |
| K862535 | OSTEONICS SHOULDER PROSTHESIS | Osteonics Corp. | Sep 2, 1986 | SESE |
| K855183 | COFIELD SHOULDER SYSTEM | Richards Medical Co., Inc. | Mar 26, 1986 | SESE |
| K852365 | LINK SHOULDER PROSTHESIS ENDO-MODEL SYSTEM | Link America, Inc. | Dec 16, 1985 | SESE |
| K843624 | COFIELD SHOULDER SYSTEM | Richards Medical Co., Inc. | Jan 10, 1985 | SESE |
| K833410 | MONOSPHERICAL GLENOID PROSTH. KEEL | Howmedica Corp. | Nov 7, 1983 | SESE |
| K801970 | UCLA SHOULDER | Howmedica Corp. | Dec 31, 1980 | SESE |