GLENOID INTELLIGENT REUSABLE INSTRUMENT SYSTEM (GLENOID IRIS)
Applicant
Cleveland Clinic
Product Code
KWS · Orthopedic
Decision Date
Apr 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Indications for Use
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant. The Glenoid IRIS is indicated for use in planning and placing the central glenoid guide pin for the DePuy Anchor Peg Glenoid (APG) component of the DePuy AP Shoulder System, the DePuy Global StepTech Glenoid component, or the DePuy Delta Xtend Reverse Shoulder metaglene component as an alternative to the standard instruments provided for placing the guide pin with these implant systems. The Glenoid IRIS is not indicated for use in hemi-shoulder arthroplasty. The labeling and indications for use for each of these DePuy shoulder systems remain the same as described in DePuy's labeling.
Device Story
System uses patient CT data and 3D modeling for preoperative planning of glenoid component orientation. Surgeon-approved plan transferred to patient-specific instrument (Glenoid SmartBone - Pin Trajectory) and adjustable reusable instrument (Glenoid IRI). During surgery, Glenoid IRI set with patient-specific settings to guide placement of standard 2.5 mm Steinmann pin. Used in clinical setting by surgeons to facilitate accurate guide pin placement for total shoulder systems. Replaces standard manual instruments; improves surgical precision by aligning guide pin with preoperatively planned trajectory. Benefits patient through potentially improved implant positioning.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including verification and validation, biocompatibility (ISO 10993-10; USP Class VI), packaging integrity, and performance studies using cadaver and sawbones models.
Technological Characteristics
Patient-specific manual instrument system. Materials meet biocompatibility standards (ISO 10993-10; USP Class VI). System includes patient-specific guides and adjustable reusable instruments. Operates via mechanical guidance of a 2.5 mm Steinmann pin. Software used for preoperative planning and 3D modeling.
Indications for Use
Indicated for patients undergoing total shoulder arthroplasty requiring central glenoid guide pin placement for DePuy AP Shoulder System (APG component), DePuy Global StepTech Glenoid, or DePuy Delta Xtend Reverse Shoulder metaglene. Not indicated for hemi-shoulder arthroplasty.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
DePuy Global Shoulder AP Shoulder System (K060874)
K162697 — Arthrex Glenoid Intelligent Reusable Instrument System, Arthrex OrthoVis Preoperative Plan, SmartBase for Arthrex Glenoid IRIS, ArthrexVIP Web Portal · Custom Orthopaedic Solutions, Inc. · Feb 10, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
## APR 0 5 2013
510(k) SPONSOR / MANUFACTURER:
Custom Orthopaedic Solutions, Inc. A subsidiary of Cleveland Clinic 10000 Cedar Avenue Cleveland, Ohio 44106
CONTACT PERSON:
Stephen J. Peoples, VMD, MS Peoples & Associates Consulting, LLC 5010 Lodge Pole Lane Fort Wayne, IN 46814 Tele: (260) 645 - 0327 Email: SPeoplesVMD@gmail.com
TRADE NAME:
Glenoid Intelligent Reusable Instrument System (Glenoid IRIS)
COMMON NAMES:
Total shoulder replacement instruments
| Product | Product Code | Regulation and<br>Classification Name | Device Class |
|---------------------------------------------------------------------|--------------|--------------------------------------------------------------------------------|--------------|
| Glenoid Reusable<br>Intelligent Instrument<br>System (Glenoid IRIS) | KWS | 21 CFR 888.3660<br>Shoulder, semi-<br>constrained metal /<br>polymer, cemented | II |
#### PREDICATE DEVICES:
DePuy Global Shoulder AP Shoulder System (K060874 DePuy Global Shoulder Glenoid (K981487) DePuy Global StepTech Anchor Peg Glenoid (K092122) DePuy Delta Xtend Reverse Shoulder System (062250)
#### DEVICE DESCRIPTION:
The Glenoid Intelligent Reusable Instrument System is composed of two (2) manual instruments intended for use to facilitate preoperative planning and intraoperative placement of the glenoid component in total shoulder replacement. CT data and 3D modeling is used to provide preoperative planning of total shoulder glenoid component or reverse shoulder metaglene component orientation according to each patient's glenoid anatomy. The preoperatively planned and surgeon approved component orientation is subsequently transferred to the patient's glenoid during surgery by the use of a patient specific instrument and an adjustable instrument with patient specific settings. A patient specific Glenoid SmartBone - Pin Trajectory instrument is used to set the adjustable reusable instrument, the Glenoid Intelligent Reusable Instrument (Glenoid IRI), with the settings necessary to reflect the guide pin trajectory embedded in the Glenoid SmartBone. The Glenoid IRI is then used to guide the placement of the standard 2.5 mm pin (Steinmann pin) by the surgeon that is used in preparation of the glenoid for implantation of the glenoid component. All other steps of the surgical procedure are accomplished according to each system's standard surgical technique.
{1}------------------------------------------------
#### INTENDED USE AND INDICATIONS:
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
The Glenoid IRIS is indicated for use in planning and placing the central glenoid guide pin for the DePuy Anchor Peg Glenoid (APG) component of the DePuy AP Shoulder System, the DePuy Global StepTech Glenoid component, or the DePuy Delta Xtend Reverse Shoulder metaglene component as an alternative to the standard instruments provided for placing the guide pin with these implant systems. The Glenoid IRIS is not indicated for use in hemi-shoulder arthroplasty.
The labeling and indications for use for each of these DePuy shoulder systems remain the same as described in DePuy's labeling.
#### BASIS OF SUBSTANTIAL EQUIVALENCE:
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is intended to be used as an alternative method and instrument system for placing the 2.5 mm glenoid guide pin that is used for preparing the glenoid for implantation with the DePuy Global AP Shoulder Glenoid component, the DePuy Global StepTech Anchor Peg Glenoid component, or the DePuy Delta Xtend Reverse Shoulder System metaglene component. The Glenoid IRIS replaces the standard instruments and techniques for the placement of the guide pin in these systems as described in each system's surgical technique. The Glenoid IRIS has the same intended use and indications for use and similar function, surgical technique, design." materials, and performance as the standard instruments used for placing the elenoid guide pin in these systems. All other surgical steps are completed according to the steps described in the surgical techniques for each system using the standard instruments supplied for each system. The indications for use of the DePuy shoulder systems, with which the Glenoid IRIS is intended to be used, are not changed by the Glenoid IRIS and remain the same as described in the labeling for these shoulder systems.
#### Non-Clinical Testing
The following testing was performed to demonstrate substantial equivalency of the Glenoid IRIS to the predicate devices.
- . Performance testing - verification and validation testing
- . Performance testing - biocompatibility and toxicity (ISO 10993-10; USP Class VI)
- . Performance testing - cadaver study
- . Performance testing - sawbones study
- . Performance testing - software validation
- . Performance testing - packaging integrity testing
#### Clinical Testing
Clinical testing was not necessary to determine substantial equivalence between the Glenoid IRIS and the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Custom Orthopaedic Solutions, Inc. A subsidiary of Cleveland Clinic c/o Stephen J. Peoples, VMD, MS Peoples & Associates Consulting, LLC 5010 Lodge Pole Lane Fort Wayne, IN, 46814
Letter dated: April 5, 2013
Re: K123122
Trade/Device Name: Glenoid Intelligent Reusable Instrument System (Glenoid IRIS) Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS Dated: March 11, 2013 Received: March 14, 2013
Dear Dr. Peoples,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the iddictions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual resustration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly and in warranties. We remind you; however, that device labeling must be truthful and not misleding.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. Stephen J. Peoples ·
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K123122
Device Name: Glenoid Intelligent Reusable Instrument System
### Intended Use and Indications for Use:
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
The Glenoid IRIS is indicated for use in planning and placing the central glenoid guide pin for the DePuy Anchor Peg Glenoid (APG) component of the DePuy AP Shoulder System, the DePuy Global StepTech Glenoid component, or the DePuy Delta Xtend Reverse Shoulder metaglene component as an alternative to the standard instruments provided for placing the guide pin with these implant systems. The Glenoid IRIS is not indicated for use in hemi-shoulder arthroplasty.
The labeling and indications for use for each of these DePuy shoulder systems remain the same as described in DePuy's labeling.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
page 1 of 1
## Elizabeth፣២៦Erank -S
Division of Orthopedic Devices
11
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.