Tornier Humeral Reconstruction System (Tornier HRS)
Applicant
Stryker Corporation (Tornier, Inc.)
Product Code
KWS · Orthopedic
Decision Date
Sep 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Intended Use
IN ANATOMIC: The proximal body, stem, assembly screw, locking cap, optional spacer(s), and humeral head may be used together, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement. The Tornier HRS Shoulder System is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain. The Tornier HRS Shoulder System is indicated for use as a replacement of shoulder joints disabled by: Rheumatoid arthritis with pain Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) Correction of functional deformity Fractures of the humeral head Traumatic arthritis Revision of other devices if sufficient bone stock remains IN REVERSE: The Tornier HRS Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle with pain disabled by: Rheumatoid arthritis Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) Correction of functional deformity Fractures of the humeral head Traumatic arthritis Massive and non-repairable rotator cuff tear Revision of the devices if sufficient bone stock remains The reversed tray and polyethylene insert are indicated for use in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of the humeral assembly during revision surgery for patients with a functional deltoid muscle.
Device Story
Tornier HRS is a modular, convertible shoulder arthroplasty system for anatomic or reverse configurations. Components include proximal body, stem, assembly screw, locking cap, and optional spacers. System allows conversion from anatomic to reverse without removing humeral assembly during revision. Implanted by surgeons in clinical settings to replace glenohumeral articulation. Provides mobility and pain relief. Titanium alloy (Ti6A4V) used for body/stem; CoCr for screws/caps. Titanium humeral heads available for patients with cobalt sensitivity. Device is single-use, sterile, and gamma-irradiated. Subject of this submission is the addition of MR Conditional labeling; no changes to device design or materials.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical testing regarding Magnetic Resonance (MR) environment compatibility.
Technological Characteristics
Modular shoulder prosthesis. Materials: Ti6A4V (ASTM F-136) for proximal body, stem, spacers; CoCr (ISO 5832-12) for assembly screw, locking cap. Titanium plasma spray coating on body/stem. Cemented or cementless fixation. Sterilization: Gamma radiation (min 25 kGy, SAL 1x10-6).
Indications for Use
Indicated for patients requiring shoulder joint replacement due to rheumatoid arthritis, non-inflammatory degenerative joint disease (osteoarthritis/avascular necrosis), functional deformity, humeral head fractures, traumatic arthritis, or revision. Anatomic configuration requires intact/reconstructable rotator cuff. Reverse configuration requires functional deltoid muscle and is indicated for massive/non-repairable rotator cuff tears.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Aequalis Flex Revive Shoulder System Line Extension (K191318)
K130661 — ZIMMER TRABECULAR METAL REVERSE SHOULDER SYSTEM · Zimmer, Inc. · May 9, 2013
K241878 — Tornier Humeral Reconstruction System (Tornier HRS); Tornier Perform Humeral System - Stem (Tornier PHS-Stem) · Stryker Corporation (Tornier, Inc.) · Dec 6, 2024
K214001 — Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System · Biomet Manufacturing Corp · Mar 15, 2023
K181420 — Aequalis Flex Revive Shoulder System · Tornier, Inc. · Sep 20, 2018
K250713 — Equinoxe® Humeral Reconstruction Prosthesis; Equinoxe® Stemless Shoulder Implants · Exactech, Inc. · Dec 1, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
September 10, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Stryker Corporation (Tornier, Inc.) Jang Long Staff Specialist, Regulatory Affairs 10801 Nesbitt Avenue South Bloomington, Minnesota 55437
## Re: K241609
Trade/Device Name: Tornier Humeral Reconstruction System (Tornier HRS) Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS, PHX, HSD Dated: August 9, 2024 Received: August 9, 2024
Dear Jang Long:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely, Joseph P. Russell Digitally signed by Joseph P. Russell -5 -5 Date: 2024.09.10 11:11:10 -04'00' for: Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241609
Device Name
Tornier Humeral Reconstruction System (Tornier HRS)
### Indications for Use (Describe)
IN ANATOMIC:
The proximal body, stem, assembly screw, locking cap, optional spacer(s), and humeral head may be used together, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing sufface, or in conjunction with the glenoid, as a total replacement.
The Tornier HRS Shoulder System is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain. The Tornier HRS Shoulder System is indicated for use as a replacement of shoulder joints disabled by:
- Rheumatoid arthritis with pain
- · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- · Correction of functional deformity
- · Fractures of the humeral head
- Traumatic arthritis
- · Revision of other devices if sufficient bone stock remains
## IN REVERSE:
The Tornier HRS Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle with pain disabled by:
- Rheumatoid arthritis
- · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- · Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- · Massive and non-repairable rotator cuff tear
- · Revision of the devices if sufficient bone stock remains
The reversed tray and polyethylene insert are indicated for use in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of the humeral assembly during revision surgery for patients with a functional deltoid muscle.
## Notes:
- · All components are single use.
- · The coated humeral stem is intended for cemented or cementless use.
- · The all-poly glenoid components are intended for cemented use only.
- The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
· Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity.
The wear properties of Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack a suspected material sensitivity to cobalt
{4}------------------------------------------------
alov.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| 510(k) #: K241609 | 510(k) Summary | K241609 - Page 1 of 3<br>Prepared on: 2024-09-09 |
|------------------------------------|-------------------------------------------------------------------|--------------------------------------------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Stryker Corporation (Tornier, Inc.) | |
| Applicant Address | 10801 Nesbitt Avenue South Bloomington MN 55437 United States | |
| Applicant Contact Telephone | 612-258-5639 | |
| Applicant Contact | Ms. Long Jang | |
| Applicant Contact Email | jang.long@stryker.com | |
| Correspondent Name | Stryker Corporation (Tornier, Inc.) | |
| Correspondent Address | 10801 Nesbitt Avenue South Bloomington MN 55437 United States | |
| Correspondent Contact Telephone | 612-258-5639 | |
| Correspondent Contact | Ms. Jang Long | |
| Correspondent Contact Email | jang.long@stryker.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Tornier Humeral Reconstruction System (Tornier HRS) | |
| Common Name | Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented | |
| Classification Name | Shoulder joint metal/polymer semi-constrained cemented prosthesis | |
| Regulation Number | 888.3660 | |
| Product Code(s) | KWS, PHX, HSD | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K191318 | Aequalis Flex Revive Shoulder System Line Extension | KWS,PHX,HSD |
| K181420 | Aequalis Flex Revive Shoulder System | KWS,PHX,HSD |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
Tornier HRS (formerly branded Aequalis™ Flex Revive™ Shoulder System) is a fully convertible anatomic and reversed shoulder arthroplasty system that is designed to be used with existing Tornier implant system intended for total or partial replacement of the glenohumeral articulation. Tornier HRS includes a proximal body (metaphysis), a stem, assembly screw, and locking cap. The proximal body has a female taper that is compatible with Tornier Flex (formerly branded Ascend Flex) Humeral Heads and Tornier Flex (formerly branded Ascend Flex) reversed trays and poly inserts.
Tornier HRS is implanted by a surgeon and is designed to select the components to size the shoulder system for the patient. It allows the shoulder to be constructed in an anatomical or reversed configuration using cleared Tornier Flex (formerly branded
{6}------------------------------------------------
Ascend Flex) Humeral Heads and Tornier Flex (formerly branded Ascend Flex) reversed trays and inserts. In addition, Tornier HRS can be transformed from anatomic to reverse shoulder prosthesis without the humeral implant assembly during revision surgery.
The humeral length is measured to determine the overall humeral implant construct length is assembled from 120 mm (using the short proximal body and stem) to 300 mm (using the standard proximal body, spacers, and stem) in 10 mm increments with the spacers as needed and either of the two available lengths of the proximal body for patient specificity.
The proximal body, stem, and spacers are made from Ti6A4V per ASTM F-136. The proximal body and stem have a Titanium plasma spray coating. The assembly screw and locking cap are made of CoCr per ISO 5832-12. All implant parts are single use and packaged sterile, using gamma radiation at a minimum dose of 25 kGy to an SAL of 1x10-6.
# Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
### IN ANATOMIC:
The proximal body, stem, assembly screw, locking cap, optional spacer(s), and humeral head may be used together, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement.
The Tornier HRS Shoulder System is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain. The Tornier HRS Shoulder System is indicated for use as a replacement of shoulder joints disabled by:
- Rheumatoid arthritis with pain
- · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- · Revision of other devices if sufficient bone stock remains
### IN REVERSE:
The Tornier HRS Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle with pain disabled by:
- Rheumatoid arthritis
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- · Massive and non-repairable rotator cuff tear
- Revision of the devices if sufficient bone stock remains
The reversed tray and polyethylene insert are in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of the humeral assembly during revision surgery for patients with a functional deltoid muscle.
#### Notes:
- · All components are single use.
- The coated humeral stem is intended for cemented or cementless use.
- · The all-poly glenoid components are intended for cemented use only.
- The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
• Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack a suspected material sensitivity to cobalt alloy.
# Indications for Use Comparison
The indications for use of the subject and predicate devices are identical. There is no impact to the indications for use as a result of the revision to the labeling to add MR conditional safety information. Therefore, adding Magnetic Resonance (MR) Conditional Labeling does not raise different questions of safety or effectiveness.
# ogical Comparison
Based upon a comparison of the interials, summary of technological characteristics, and non-clinical evaluation, the
## 21 CFR 807.92(a)(5)
{7}------------------------------------------------
subject Tornier HRS is substantially equivalent to the current Tornier HRS (cleared as Aequalis Flex Revive in K191318 and K181420). There have been no changes to the technological characteristics of Tornier HRS as a result of the labeling to add MR conditional safety information. The subject devices have the same design and are manufactured from the same materials as the predicate devices.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The Tornier HRS system has been evaluated in a Magnetic Resonance Environment through non-clinical testing as outlined in the FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment – Guidance for Industry and FDA Staff", dated October 10, 2023. This testing was conducted to characterize the compatibility of Tornier HRS in the MR environment.
No clinical studies were performed.
The Tornier HRS system does not raise different questions of safety or effectiveness. Based upon a comparison of the intended use, materials, summary of technological characteristics, and non-clinical evaluation, the subject Tornier HRS is considered substantially equivalent to the current Tornier HRS (cleared as Aequalis Flex Revive in K191318 and K181420).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.