The Aequalis Ascend Flex Shoulder System is intended for use as: a replacement of shoulder joints in primary anatomic or in primary reverse a replacement of other shoulder joints devices in case of revisions if sufficient bone stock remains. The Aequalis Ascend Flex Shoulder System also allows for conversions from anatomic to reverse shoulder prosthesis in case of revision.
Device Story
Aequalis Ascend Flex Shoulder System is a semi-constrained shoulder prosthesis; used for total shoulder replacement or hemiarthroplasty. System includes titanium humeral stems (coated/uncoated), cobalt-chrome (CoCr) humeral heads, and UHMWPE glenoid components. Reversed configuration utilizes a titanium tray, UHMWPE insert, and CoCr/titanium glenoid components. Device implanted by orthopedic surgeons in clinical settings to relieve pain and restore mobility. Subject submission adds 135° and 140° reversed angle combinations and an 'Inlay' surgical technique to existing 145° options. Implants function by replacing damaged joint surfaces; reversed configuration shifts center of rotation to compensate for rotator cuff deficiency. Benefits include pain relief and improved joint function for patients with severe cuff pathology or failed prior arthroplasty.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via non-clinical bench testing, specifically fatigue testing for the new angle combinations in both anatomic and reversed configurations.
Indicated for patients with shoulder joint disability due to rheumatoid arthritis, non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), functional deformity, humeral head fractures, traumatic arthritis, or revision of other devices. Anatomic configuration requires intact/reconstructable rotator cuff. Reverse configuration requires functional deltoid muscle and massive, non-repairable rotator cuff tear.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Aequalis™ Ascend™ Flex Shoulder System (K122698, K151293)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
June 12, 2019
Tornier SAS Lamia Askri Regulatory Affairs Specialist 161 rue Lavoisier 38330 Montbonnot Saint Martin France
Re: K190521
Trade/Device Name: Aequalis™ Ascend™ Flex Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder ioint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS, KWT, HSD, PHX Dated: May 6, 2019 Received: May 13, 2019
Dear Lamia Askri:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Raquel Peat, PhD, MPH, USPHS For: Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190521
#### Device Name
Aequalis Ascend Flex Shoulder System
#### Indications for Use (Describe)
#### IN ANATOMIC:
The stem and head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement.
The Aequalis Ascend Flex Shoulder System is to be used only in patients with an intact or reconstructable rotator ouff. where it is intended to provide increased mobility and to relieve pain. The Aequalis Ascend Flex Shoulder System is indicated for use as a replacement of shoulder joints disabled by:
- Rheumatoid arthritis with pain
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- Revision of other devices if sufficient bone stock remains
## IN REVERSE:
The Aequalis Ascend Flex Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- Rheumatoid arthritis
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- Revision of the devices if sufficient bone stock remains.
The reversed adapter is indicated for use as components of the Aequalis Ascend Flex Shoulder System total shoulder replacement and for transformation of the Aequalis Ascend Flex Shoulder System into reverse shoulder prosthesis without the removal of the humeral stem during revision surgery for patients with a functional deltoid muscle. The components are permitted to be used in the transformation from anatomic to reverse if the humeral stem is well fixed, the patient has a functional deltoid muscle; the arthropathy is associated with a massive and non-repairable rotator cuff-tear.
#### Notes:
- all components are single use.
- the coated humeral stem is intended for cemented or cementless use,
- the non-coated humeral stem is for cemented use only.
- the all-poly glenoid components are intended for cemented use only.
- the glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
- Titanium humeral heads are intended for patients with suspected sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for
**FORM FDA 3881 (7/17)**
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### patients who lack a suspected material sensitivity to cobalt alloy.
#### Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Aequalis Ascend Flex Shoulder System
Section 4 - page 3 / page 3
{4}------------------------------------------------
# Summary of Safety and Effectiveness information Special 510(k) Premarket - Aequalis Ascend Flex Shoulder System
Regulatory authority: Safe Medical Devices Act of 1990, 21 CFR 807.92 Date Prepared: February 28,2019
## Administrative Information
| Name: | Tornier SAS |
|----------|--------------------------------------------------------------|
| Address: | 161 rue Lavoisier<br>38330 Montbonnot Saint Martin<br>France |
| Contact Person: | Lamia ASKRI |
|-----------------|-------------------------------|
| Title: | Regulatory Affairs Specialist |
| Phone: | +33476613540 |
| Fax: | +33476613533 |
| Email: | lamia.askri@wright.com |
### Device Information
| Name of Device: | Aequalis <sup>TM</sup> Ascend <sup>TM</sup> Flex Shoulder System |
|-------------------|------------------------------------------------------------------------------------------|
| Common Name (s): | Shoulder Prosthesis |
| Regulatory Class: | Class II |
| Regulation: | 21 CFR § 888.3660, Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis. |
| Product Codes: | KWS, KWT, HSD, PHX |
## Predicate Device Information
| Predicate: | Aequalis™ Ascend™ Flex Shoulder System |
|----------------|----------------------------------------|
| 510(k) Number: | K122698, K151293 |
#### Device Description:
The Aequalis Ascend Flex Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear.
The Aequalis Reversed Shoulder System is intended to relieve pain and significant disability following massive and non repairable cuff-tear associated to arthropathy and following massive cuff-tear arthroplasty. It is a semi-constrained system composed of a humeral and a glenoid parts.
The Aequalis Ascend Flex Shoulder System consists of:
- * In an Anatomic configuration: a titanium humeral stem offered in Titanium Plasma Spray (Ti PS) coated and un-coated stem versions, a compatible humeral head (CoCr) with a compatible UHMWPE Aequalis glenoid; or UHMWPE Affiniti Anatomic glenoid. The Aequalis Ascend Flex Shoulder System stem and head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with a glenoid, as a total shoulder joint replacement.
- * In a Reversed configuration: a titanium humeral stem offered in Titanium Plasma Spray (Ti PS) coated and un-coated stem versions, a reversed adapter with compatible Aequalis
{5}------------------------------------------------
Reversed glenoid implants. The reversed adapter is comprised of two components: a titanium tray and a UHMWPE reversed insert. The Aequalis Reversed glenoid implants is comprised of four components: Baseplate: made from Titanium; Glenoid sphere: made from of CoCr; Screw (baseplate/to glenoid sphere): made from CoCr and Fixation screws: made from Titanium.
This submission corresponds to the addition of 135° & 140° reversed final angle combinations and the Inlay technique, in addition to the existing 145° angle without any change in the implants design.
## Intended Use
The Aequalis Ascend Flex Shoulder System is intended for use as:
- a replacement of shoulder joints in primary anatomic or in primary reverse
- a replacement of other shoulder joints devices in case of revisions if sufficient bone stock remains.
The Aequalis Ascend Flex Shoulder System also allows for conversions from anatomic to reverse shoulder prosthesis in case of revision.
## Indications For Use
IN ANATOMIC:
The stem and head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement.
The Aequalis Ascend Flex Shoulder System is to be used only in patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and to relieve pain.
The Aequalis Ascend Flex Shoulder System is indicated for use as a replacement of shoulder joints disabled by:
- Rheumatoid arthritis with pain -
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) -
- Correction of functional deformity .
- . Fractures of the humeral head
- . Traumatic arthritis
- . Revision of other devices if sufficient bone stock remains
#### IN REVERSE:
The Aequalis Ascend Flex Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- Rheumatoid arthritis .
- . Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- -Correction of functional deformity
- -Fractures of the humeral head
- Traumatic arthritis -
- Revision of the devices if sufficient bone stock remains -
The reversed adapter is indicated for use as components of the Aequalis Ascend Flex Shoulder System total shoulder replacement and for transformation of the Aequalis Ascend Flex Shoulder System into a reverse shoulder prosthesis without the removal of the humeral stem during revision surgery for patients with a functional deltoid muscle. The components are permitted to be used in the transformation from anatomic to reverse if the humeral stem is well fixed, the patient has a
Copyright Wright Medical CORP-FORM-2698 Rev A
Section 11 - page 2 / page 5
{6}------------------------------------------------
functional deltoid muscle; the arthropathy is associated with a massive and non-repairable rotator cuff-tear.
Notes:
- All components are single use. -
- -The coated humeral stem is intended for cemented or cementless use.
- = The non-coated humeral stem is intended for cemented use only.
- -All poly glenoid components are intended for cemented use only.
- The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
- . Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack a suspected material sensitivity to cobalt alloy.
## Comparison of Technological Characteristics with the Predicate Device
The proposed changes to the predicate device include the addition of 135° & 140° reversed final angle and an alternate surgical option (Inlay), in addition to the existing 145° angle. These changes consist of added stem/ insert combinations without any change in the implants design. The subject device has the same design of implants, the same materials, the same manufacturing principle, the same method of fixation, the same packaging and the same sterilization process as the predicate device.
The new combinations do not raise new issues of safety or effectiveness and are supported by performance testing.
### Non-clinical Performance Testing
Non-clinical bench testing and process validations were performed to demonstrate substantial equivalence of the subject device to the predicate device.
- Fatigue test in anatomic and reversed configuration .
## Clinical Testing
Clinical studies were not required to demonstrate substantial equivalence between the subject device and the predicate device.
#### Conclusions
The AequalisTM Ascend™ Flex Shoulder System does not raise new questions of safety or effectiveness. with performance testing. The results of performance testing for the new angle combinations of the Aequalis Ascend Flex Shoulder System support substantial equivalence to the current Aequalis Ascend Flex Shoulder System (K122698, K151293).
Copyright Wright Medical CORP-FORM-2698 Rev A
Section 11 - page 3 / page 5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.