K Number
K073367
Device Name
HEMOSIL FIBRINOGEN-C
Date Cleared
2007-12-27

(27 days)

Product Code
Regulation Number
864.7340
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
HemosIL Fibrinogen-C is intended for the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems. For in vitro diagnostic use.
Device Description
The Fibrinogen-C kit uses an excess of thrombin to convert fibrinogen to fibrin in diluted plasma. At high thrombin and low fibrinogen concentration, the rate of reaction is a function of the fibrinogen concentration.
More Information

K931721 HemosIL Fibrinogen-C

Not Found

No
The summary describes a standard in vitro diagnostic assay based on a chemical reaction (Clauss method) and does not mention any AI or ML components.

No
The device is described as being for "quantitative determination of fibrinogen" and "for in vitro diagnostic use," indicating it is used for diagnosis, not treatment.

Yes

The "Intended Use / Indications for Use" section explicitly states, "For in vitro diagnostic use."

No

The device description clearly states it is a "kit" and describes a chemical reaction involving "reagents" and "plasma," indicating it is a physical in vitro diagnostic (IVD) device, not software only.

Yes, this device is an IVD (In Vitro Diagnostic).

The "Intended Use / Indications for Use" section explicitly states: "For in vitro diagnostic use."

N/A

Intended Use / Indications for Use

HemosIL Fibrinogen-C is intended for the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems. For in vitro diagnostic use.

Product codes (comma separated list FDA assigned to the subject device)

KQJ

Device Description

The Fibrinogen-C kit uses an excess of thrombin to convert fibrinogen to fibrin in diluted plasma. At high thrombin and low fibrinogen concentration, the rate of reaction is a function of the fibrinogen concentration.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Summary of Performance Data:

  • Interference:
    Current Parameters: Heparin Up to 1 U/mL, Hemoglobin Up to 375 mg/dL, Triglycerides Up to 890 mg/dL, Bilirubin Up to 21 mg/dL, Fibrinogen Degradation Products Up to 100 µg/mL
    Modified Parameters: Heparin Up to 1 U/mL, Hemoglobin Up to 375 mg/dL, Triglycerides Up to 750 mg/dL, Bilirubin Up to 21 mg/dL, Fibrinogen Degradation Products: New testing showed interference from FDP. Therefore, the insert will indicate that Fibrinogen-C assay results on the ACL TOP may be affected by degradation products (fibrin or fibrinogen) in the plasma assayed. This statement is consistent with the language already in use for the other ACL instrument platforms. An associated literature reference regarding FDP interference will be added.

  • Precision:
    Current Parameters: Control Level - Normal Mean 365 mg/dL, Within Run CV 7.9%; Low Fibrinogen Mean 93 mg/dL, Within Run CV 7.7%
    Modified Parameters: Control Level - Normal Mean 303 mg/dL, Within Run CV 4.5%; Low Fibrinogen Mean 107 mg/dL, Within Run CV 5.1%

  • Linearity:
    Current Parameters: 80 - 700 mg/dL
    Modified Parameters: 35 - 1000 mg/dL

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K931721 HemosIL Fibrinogen-C

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 864.7340 Fibrinogen determination system.

(a)
Identification. A fibrinogen determination system is a device that consists of the instruments, reagents, standards, and controls used to determine the fibrinogen levels in disseminated intravascular coagulation (nonlocalized clotting within the blood vessels) and primary fibrinolysis (the dissolution of fibrin in a blood clot).(b)
Classification. Class II (special controls). A control or fibrinogen standard intended for use with a fibrinogen determination system is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

0

K073367

Submitted by:

Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421

Contact Person:

Carol Marble, Regulatory Affairs Director Phone No .: 781-861-4467 Fax No .: 781-861-4207

Prepared:

November 29, 2007

Device Name:

HemosIL Fibrinogen-C

Regulatory Information:

KQJSystem, Fibrinogen Determination
864.7340Fibrinogen Determination SystemClass II

Predicate Device:

K931721 HemosIL Fibrinogen-C

Device Intended Use:

HemosIL Fibrinogen-C is intended for the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.

Device Description:

The Fibrinogen-C kit uses an excess of thrombin to convert fibrinogen to fibrin in diluted plasma. At high thrombin and low fibrinogen concentration, the rate of reaction is a function of the fibrinogen concentration.

Reason for Submission:

The HemosIL Fibrinogen-C test parameters on the ACL TOP are being modified as follows:

  • . Separate low-end and high-end reflex test parameters are being added to expand the linearity claims on the ACL TOP from the current labeled 80-700 mg/dL to 35-1000 mg/dL.
  • . The calibration levels are now prepared by aspirating the calibrator directly from the calibrator vial (direct dilution instead of serial dilution).
  • The robustness of the assay is increased by optimizing the incubation time and math model. .

Statement of Technological Characteristics of the Device Compared to Predicate Device:

The performance of the optimized test parameters for HemosIL Fibrinogen-C on the ACL TOP is substantially equivalent to the performance of the current legally marketed test parameters.

  • . There are no changes in the reagent's intended use, formulation or operating principle and no changes to the instrument system software, intended use, operating principle or labeled performance claims with this submission.
    DEC 2 7 2007

1

510(k) Summary (Cont.)

ート - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

Summary of Performance Data:

Limitations
Current ParametersModified Parameters
No Interference
HeparinUp to 1 U/mLHeparinUp to 1 U/mL
HemoglobinUp to 375 mg/dLHemoglobinUp to 375 mg/dL
TriglyceridesUp to 890 mg/dLTriglyceridesUp to 750 mg/dL
BilirubinUp to 21 mg/dLBilirubinUp to 21 mg/dL
Fibrinogen Degradation Products
Up to 100 µg/mLFibrinogen Degradation Products:
New testing showed interference from FDP.
Therefore, the insert will indicate that
Fibrinogen-C assay results on the ACL TOP
may be affected by degradation products
(fibrin or fibrinogen) in the plasma assayed.
This statement is consistent with the
language already in use for the other ACL
instrument platforms. An associated
literature reference regarding FDP
interference will be added.
Precision
Current ParametersModified Parameters
Control LevelMeanWithin Run CVControl LevelMeanWithin Run CV
Normal365 mg/dL7.9%Normal303 mg/dL4.5%
Low Fibrinogen93 mg/dL7.7%Low Fibrinogen107 mg/dL5.1%
Linearity
Current ParametersModified Parameters
80 - 700 mg/dL35 - 1000 mg/dL

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines forming its body and wings. The bird is positioned above a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA". The text is arranged around the circumference of the circle, with the bird symbol placed in the center.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 2 7 2007

Instrumentation Laboratory Co. C/O Carol Marble 113 Hartwell Avenue Lexington, Massachusetts 02421

Re: K073367

Trade/Device Name: Hemosil Fibrinogen-C Regulation Number: 21 CFR 864.7340 Regulation Name: Fibrinogen Determination System Regulatory Class: Class II Product Code: KQJ Dated: November 29, 2007 Received: November 30, 2007

Dear Ms. Marble:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

3

Page 2 - Instrumentation Laboratory Co.

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Tabat Broderf

Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

4

Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: HemosIL Fibrinogen-C

Indications for Use:

HemosIL Fibrinogen-C is intended for the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.

For in vitro diagnostic use.

ﮯ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use ____________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Jacqueline Bautista
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Ko73367

Special 510(k): HemosIL Fibrinogen-C