K Number
K251811
Date Cleared
2025-08-25

(74 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Midwest Motor Control Module includes a control unit, ports for three motors, an apex location port, and a peristaltic pump port. The motor ports are compatible with the Midwest Power Lux and Midwest Power Lux Implant.

The Midwest Power Lux electric micromotor is intended for use as a drive for rotating and oscillating straight and contra-angle hand pieces for endodontic and general dental purposes.

The Midwest Power Lux Implant motor is intended for use in implant placement and includes a closed system for delivery of sterile solution via peristaltic pump operation.

The apex locator supports the dentist in the determination of the working length during the endodontic treatment.

These motors are to be driven by the control unit. The control unit is an internal component of a dental delivery system.

The use of this product is intended exclusively for duly qualified dental practitioners.

Device Description

This device is a system that includes a control module (internal component of a dental delivery system) that can operate up to three dental motors, an apex locator, and a peristaltic pump. The motors do not operate concurrently. The motors that can be attached include dental motors and dental implant motors. The motors included with this module in this subject system are the Midwest Power Lux and Midwest Power Lux Implant. Also included is the cable, file, and tip required for the operation of the apex location technology. The programming supports manual use of the apex locator as well as combined use of the locator with a motor. The included peristaltic pump and control allows for sterile solution irrigation when required.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the "Motor and Apex Module (MaAM)" (K251811) does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria related to the apex locator's accuracy or the motors' performance in a clinical setting.

The document primarily focuses on regulatory approval, substantial equivalence to predicate devices, and general safety and performance testing. While it mentions the device supports "determination of the working length during the endodontic treatment" (for the apex locator) and drives "rotating and oscillating straight and contra-angle hand pieces for endodontic and general dental purposes" (for the motors), it does not provide the specific performance metrics (e.g., accuracy, precision, torque output, speed stability) for these functions, nor the studies that verified them.

The "Non-Clinical and/or Clinical Tests Summary & Conclusions" section explicitly states: "Not applicable (no clinical data necessary)." This indicates that the FDA clearance was based on non-clinical testing, primarily focused on safety, electrical compatibility, and adherence to established industry standards (ISO 14457 for dental motors, IEC 62471 for photobiological safety, and ISO 3964 for dental coupling dimensions).

Therefore, based solely on the provided document, I cannot generate the requested table and study details. The information about performance acceptance criteria, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment for clinical effectiveness is absent.

However, I can extract and present the information that is available in the document regarding testing:


Summary of Device Performance and Testing (Based on Available Information in K251811)

The provided 510(k) clearance letter for the Motor and Apex Module (MaAM) establishes substantial equivalence to predicate devices based on non-clinical testing. The document explicitly states "Not applicable (no clinical data necessary)", indicating that the clearance did not involve clinical studies assessing the device's functional performance in a patient setting, such as the accuracy of the apex locator or the clinical efficacy of the motors.

The testing highlighted in the document focuses on compliance with international standards and mechanical/electrical safety.

1. Table of Acceptance Criteria and Reported Device Performance

Given that clinical performance acceptance criteria and respective reported device performance data are not detailed in the provided document, the table below reflects the types of non-clinical testing mentioned and the general conclusion of meeting specifications.

Acceptance Criterion (Type of Testing)Reported Device Performance
Dental Motors (ISO 14457)Device met specifications in accordance with ISO 14457 for dental motors. (Specific metrics not provided)
Photobiological Safety (IEC 62471)Device met specifications in accordance with IEC 62471:2006 for photobiological safety of lamps and lamp systems.
Dental Coupling Dimensions (ISO 3964)Device met specifications in accordance with ISO 3964:2018 AMD1:2018 for dental coupling dimensions.
Electrical Safety / EMC(Implicitly part of regulatory requirements; specific standards/results not detailed but assumed to be met for clearance).
Mechanical Integrity(Implicitly part of regulatory requirements; specific tests/results not detailed but assumed to be met for clearance).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified in the document. Testing was non-clinical (bench/laboratory).
  • Data Provenance: Non-clinical (bench testing in a laboratory). Country of origin is not specified for the test data itself, but the applicant (Dentsply Sirona Inc.) is located in the United States. The testing was retrospective in the sense that it was conducted prior to the submission for clearance.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Not applicable. The clearance was based on non-clinical, standard-driven testing. There was no clinical ground truth established by experts for performance assessment in a patient setting.

4. Adjudication Method for the Test Set

  • Not applicable. No clinical test set or subjective assessment requiring adjudication (like image interpretation) was mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC study was done. The document explicitly states "Not applicable (no clinical data necessary)". Therefore, no effect size of human readers improving with AI vs. without AI assistance can be reported. The device is not an AI-assisted diagnostic tool in the sense of image interpretation.

6. Standalone Performance (Algorithm Only)

  • Not applicable. The device is a physical medical device (motors, control module, apex locator) and not an algorithm or AI software for standalone performance evaluation in a diagnostic context. Its "performance" refers to its physical and electrical operation within specified limits, not its diagnostic accuracy.

7. Type of Ground Truth Used

  • Not applicable, for clinical performance. For the non-clinical tests, the "ground truth" was established by the specifications defined in the referenced ISO and IEC standards (e.g., a motor must achieve a certain speed, a coupling must fit, light emission must be within safety limits). There was no "expert consensus, pathology, or outcomes data" ground truth as would be used for diagnostic accuracy studies.

8. Sample Size for the Training Set

  • Not applicable. This device is hardware with embedded control software; it is not a machine learning or AI model that requires a training set in the typical sense of algorithm training on data.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As above, this device does not utilize a "training set" in a machine learning context.

In conclusion, the FDA 510(k) clearance for the MaAM (K251811) was granted based on non-clinical testing demonstrating conformance to relevant safety and performance standards for dental motors, photobiological safety, and dental coupling dimensions. The document explicitly states that no clinical data was necessary for this clearance, meaning functional performance (e.g., apex locator accuracy, motor clinical efficacy) in a patient setting was not assessed as part of this submission.

FDA 510(k) Clearance Letter - Motor and Apex Module (MaAM)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov

August 25, 2025

Dentsply Sirona Inc.
Diane Rutherford
Director Regulatory Affairs
221 West Philadelphia St
Suite 60W
York, Pennsylvania 17401

Re: K251811
Trade/Device Name: Motor and Apex Module (MaAM)
Regulation Number: 21 CFR 872.4200
Regulation Name: Dental Handpiece And Accessories
Regulatory Class: Class I, reserved
Product Code: EBW, EKX, LQY
Dated: June 12, 2025
Received: June 12, 2025

Dear Diane Rutherford:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251811 - Diane Rutherford Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

Page 3

K251811 - Diane Rutherford Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K251811

Please provide the device trade name(s).

Motor and Apex Module (MaAM)

Please provide your Indications for Use below.

The Midwest Motor Control Module includes a control unit, ports for three motors, an apex location port, and a peristaltic pump port. The motor ports are compatible with the Midwest Power Lux and Midwest Power Lux Implant.

The Midwest Power Lux electric micromotor is intended for use as a drive for rotating and oscillating straight and contra-angle hand pieces for endodontic and general dental purposes.

The Midwest Power Lux Implant motor is intended for use in implant placement and includes a closed system for delivery of sterile solution via peristaltic pump operation.

The apex locator supports the dentist in the determination of the working length during the endodontic treatment.

These motors are to be driven by the control unit. The control unit is an internal component of a dental delivery system.

The use of this product is intended exclusively for duly qualified dental practitioners.

Please select the types of uses (select one or both, as applicable).

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

Motor and Apex Module Page 11 of 56

Page 5

K251811

510(k) Summary

510(k)#: K251811 Prepared on: 2025-06-12

Contact Details

21 CFR 807.92(a)(1)

Applicant NameDentsply Sirona Inc
Applicant Address221 West Philadelphia St Suite 60W York PA 17401 United States
Applicant Contact Telephone+1-704-458-1942
Applicant ContactMs. Diane Rutherford
Applicant Contact Emaildiane.rutherford@dentsplysirona.com

Device Name

21 CFR 807.92(a)(2)

Device Trade NameMotor and Apex Module
Common NameDental handpiece and accessories
Classification NameController, Foot, Handpiece And Cord
Regulation Number872.4200
Product Code(s)EBW, LQY

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K163131A-dec NLZ Electric Motor SystemEBW
K150909Motor BL ISO E (a physical part of Intego)EIA
K151045Motor BL Implant (a physical part of Teneo)EIA
K233865X-Smart Pro; X-Smart Pro+EKX

Device Description Summary

21 CFR 807.92(a)(4)

This device is a system that includes a control module (internal component of a dental delivery system) that can operate up to three dental motors, an apex locator, and a peristaltic pump. The motors do not operate concurrently. The motors that can be attached include dental motors and dental implant motors. The motors included with this module in this subject system are the Midwest Power Lux and Midwest Power Lux Implant. Also included is the cable, file, and tip required for the operation of the apex location technology. The programming supports manual use of the apex locator as well as combined use of the locator with a motor. The included peristaltic pump and control allows for sterile solution irrigation when required.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Midwest Motor Control Module includes a control unit, ports for three motors, an apex location port, and a peristaltic pump port. The motor ports are compatible with the Midwest Power Lux and Midwest Power Lux Implant.

The Midwest Power Lux electric micromotor is intended for use as a drive for rotating and oscillating straight and contra-angle hand pieces for endodontic and general dental purposes.

Page 6

The Midwest Power Lux Implant motor is intended for use in implant placement and includes a closed system for delivery of sterile solution via peristaltic pump operation.

The apex locator supports the dentist in the determination of the working length during the endodontic treatment.

These motors are to be driven by the control unit. The control unit is an internal component of a dental delivery system.

The use of this product is intended exclusively for duly qualified dental practitioners.

Indications for Use Comparison

21 CFR 807.92(a)(5)

The proposed and predicate devices are intended for the same clinical use. Minor variations exist in the way the indications are described, but those variations do not impact the clinical use, safety or effectiveness of the device.

Technological Comparison

21 CFR 807.92(a)(6)

The Midwest Motor Control Module includes a control module (internal component of a dental delivery system) that can operate up to three dental motors, an apex locator, and a peristaltic pump. The motors do not operate concurrently. The motors that can be attached include dental motors and dental implant motors and are only able to be operated in conjunction with the control module; the motors are integral components of the system. The motors included with this module are the Midwest Power Lux and Midwest Power Lux Implant. Also included is the cable, file, and tip required for the operation of the apex location technology. The programming supports manual use of the apex locator as well as combined use of the locator with a motor. The included peristaltic pump and control allows for sterile solution irrigation when required.

The Midwest Motor Control Module and motors have the same technological characteristics as the predicate and reference devices. Both provide a controller to operate electric micromotors for dental procedures performed by qualified dental practitioners. The control module is contained within the dental delivery system and acquires electrical power from the delivery system. Users control the motor system through the delivery system interfaces. Control receives instructions from the interfaces such as motor rotation/stop, speed setting value, torque setting value, and LED turning ON/OFF. The control module drives the motors via the motor tubing.

For Apex location, the proposed and predicate device use established apex location determination technology which evaluates the electrical impedance between the file placed in the dental root canal and the lip placed on the mucosa of the patient. Both devices allow for two methods: Apex location via a file clamp without the hand piece or via a compatible contra-angle handpiece. Both devices share components, including lip clip, file clamp, file clamp cable, adapter cable for apex measurement.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Testing in accordance to ISO 14457 for the dental motors.
Testing in accordance with IEC 62471:2006, Photobiological safety of lamps and lamp systems
Testing in accordance with ISO 3964:2018 AMD1:2018, Dental coupling dimensions

Not applicable (no clinical data necessary)

All testing demonstrated the proposed device to meet specifications and therefore, equivalent to the identified predicate device(s).

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.