K Number
K163131
Manufacturer
Date Cleared
2017-05-11

(184 days)

Product Code
Regulation Number
872.4200
Panel
DE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The A-dec NLZ electric motor system is control unit that drives a direct current (DC) electric micromotor that is activated by means of a foot control. It is intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, prophylaxis and endodontic treatment, with use of a straight, right-angle ISO E-tvpe handpiece attachment of equal, gear-reducing, or gear-increasing speed.

Device Description

The A-dec NLZ electric motor system is intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment, with use of a straight, right-angle or contra-angle ISO E-type handpiece attachment of equal, gear-reducing, or gear-increasing speed.

Control Unit, Motor and Motor Tubing are constituents of the A-dec NLZ electric motor system under this 510(k) application. The A-dec NLZ electric motor system is intended for use only with the A-dec 300 (K082985) or A-dec 500 (K032756) delivery system. The A-dec 300 (K082985) or A-dec 500 (K032756) delivery system offers operation modes for use with the A-dec NLZ electric motor system.

The Control Unit of the A-dec NLZ electric motor system is contained within the control head of the A-dec 300 (K082985) or A-dec 500 (K032756). The subject device acquires +24 VAC electrical power from the delivery system to the Control Unit. User can control the motor system through Touch Pad and Foot Controller of the delivery system. The control unit receives instructions such as motor rotation/stop, speed setting value, torque setting value, and LED turning ON/OFF from Touch Pad and Foot Controller. The Control Unit drives the Motor using instructed rotation speed and torque via Motor Tubing. The delivery system provides standard mode (range of rotation speed is 1,000 - 40,000 rpm) and endo mode (range of rotation speed is 100 - 5,000 rpm).

The Motor Tubing is connected to the Control Unit via a lead wire and air/water supply via tubing inside the control head of the delivery system. The other side of the Motor Tubing is attached to the Motor.

AI/ML Overview

The provided document is a 510(k) summary for the A-dec NLZ Electric Motor System. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain the level of detail typically required to answer your specific questions regarding acceptance criteria and a detailed study proving performance, especially in the context of AI/ML devices or studies involving human readers and ground truth.

This document describes a traditional dental device (electric motor system), not an AI/ML device. Therefore, many of your questions, particularly those about AI/ML specific details like training sets, ground truth establishment for training data, MRMC studies, or standalone algorithm performance, are not applicable to the information provided.

Based on the available information, here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The document states that the device was compliant with ISO 14457 and met the requirements of other standards. However, it does not provide a table with specific acceptance criteria (e.g., numerical thresholds for different performance metrics) and corresponding reported performance values for those criteria. It only provides a comparison table of features and ranges with the predicate device.

FeatureAcceptance Criteria (Implied by standard compliance)Reported Device Performance (A-dec NLZ Electric Motor System)
Bench TestingConformity to ISO 14457All samples compliant with ISO 14457
Electrical SafetyConformity to IEC 60601-1Test results indicate no concern
EMCConformity to IEC 60601-1-2Test results indicate no concern
BiocompatibilityConformity to ISO 10993-5, -10No biocompatibility concern raised
SterilizationConformity to FDA guidancePerformed in accordance with FDA guidance
Software ValidationConformity to FDA guidanceConducted in accordance with FDA guidance
Backflow (Motor Tubing)No backflow into Motor/Motor tubingNo backflow detected
Extractables (Water Circuit)No extractables/leachables detectedExtractables/leachables not detected
Ranges (compared to predicate, not acceptance criteria)
Standard Mode Rotation Speed1,000 - 40,000 rpm (Predicate: 1,000 - 40,000 rpm)1,000 - 40,000 rpm
Endo Mode Rotation Speed100 - 5,000 rpm (Predicate: 100 - 5,000 rpm)100 - 5,000 rpm
Standard Mode Max Torque- (Predicate: 4.10 Ncm)4.00 Ncm (Measured value)
Endo Mode Torque Range0.30 - 3.00 Ncm (Predicate: 0.30 - 3.00 Ncm)0.30 - 3.00 Ncm

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document mentions "All A-dec NLZ electric motor system samples" for bench testing but does not specify a numerical sample size. It refers to "samples" being compliant, but no specific count is provided. There is no information about data provenance as this is a device performance test, not a data-driven clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is not an AI/ML product that would require expert-established ground truth for a test set in the same way. The "ground truth" for this device's performance would be objective measurements against engineering and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept applies to human expert review for establishing ground truth, which is not relevant here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device. No MRMC study was mentioned or implied.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm, but a physical electric motor system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the performance tests mentioned (bench testing, electrical safety, EMC, biocompatibility, backflow, extractables), the "ground truth" is established by the specified international and national standards (e.g., ISO 14457, IEC 60601-1, IEC 60601-1-2, ISO 10993 series) and FDA guidance documents. Compliance is determined by objective measurements meeting the criteria set forth in these standards.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that uses a "training set."

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.