K Number
K243114
Date Cleared
2025-05-02

(214 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SQA-iOw Sperm Quality Analyzer is an automated analyzer intended for in-vitro diagnostic use to determine the following parameters in semen:

Measured parameters:

  • Sperm Concentration/ Total Sperm Concentration, millions/mL
  • Motile Sperm Concentration (MSC), millions/mL
  • Progressively Motile Sperm Concentration (PMSC), millions/mL (combines Rapidly and Slowly Progressive Motile Sperm Concentration, millions/mL)
  • Normal Forms (% Normal Morphology), %

Derived parameters:

  • Total Motility / Total Motile (PR + NP), %
  • Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
  • Non-Progressive (NP), %
  • Immotile (IM), %

The SQA-iOw is intended for CLIA Waived settings. The SQA-iOw does not provide a comprehensive evaluation of a male's fertility status and is intended for in vitro use only.

Device Description

The SQA-iOw Sperm Quality Analyzer is a PC-based analytical medical device that tests human semen samples. The device works with a computer application that manages the device, and information related to the patient, the sample, the test results and the facility.

After collection and preparation, 0.6 mL of semen sample is aspirated into a disposable SQA capillary sample delivery system and inserted into the SQA-iOw measurement chamber. The testing process takes approximately 75 seconds. The system performs an automatic self-test and auto-calibration upon start up, and checks device stability before each sample is run.

The SQA-iOw Sperm Quality Analyzer utilizes proprietary software code to both perform analysis of semen parameters and present those results on the user interface. This software is installed on a PC as a cloud-based application ("app") and is designed to perform all functions and features of the SQA-iO device, controlled by the user through a proprietary graphical user interface (GUI).

The SQA-iOw Sperm Quality Analyzer software analyzes semen parameters using signal processing technology. Sample testing is performed by capturing electrical signals as sperm moves through a light source in the SQA-iO optical block. These light disturbances are converted into electrical signals which are then analyzed by the SQA-iOw software. The SQA-iOw software applies proprietary algorithms to interpret and express these electrical signals and report them as various semen parameters.

The SQA-iOw Sperm Quality Analyzer package provides the SQA-iOw device and USB cable. SQA disposable capillaries, cleaning kits and related testing supplies and test kits are supplied individually.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the SQA-iOw Sperm Quality Analyzer meets them, based on the provided FDA 510(k) clearance letter:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA clearance letter does not explicitly list predefined quantitative acceptance criteria in a dedicated table format. Instead, it describes two precision studies and a method comparison study, concluding that the results "met the acceptance criteria." For the method comparison, it refers to "Passing-Bablok regression" with "Slopes, y-intercepts, and correlation coefficients, along with the 95% confidence intervals, were reported." The implicit acceptance criteria are typically that these statistical measures fall within a pre-specified range demonstrating equivalence to the predicate device.

Given the information provided, we can infer the acceptance criteria for the parameters measured and the reported performance.

Parameter CategoryTest TypeAcceptance Criteria (Implicit from conclusion)Reported Device Performance (Summary)
Precision (Control Material)Repeatability (Within-run), Between-day, Between-operator, Between-site, Total ImprecisionStDev and %CV met the acceptance criteria (specific values not provided in extract).All reported SDs and %CVs for Controls Level 1, Level 2, and Negative Control were low, indicating high precision. For example, Total %CV for Control Level 1 was 1.84%, and for Level 2 was 4.01%. Total SD and %CV for Negative Control were 0.00%.
Precision (Native Samples)Repeatability (Within-run), Between-operator, Total ImprecisionStDev and %CV met the acceptance criteria for all reported parameters (specific values not provided in extract).All reported SDs and %CVs for Sperm Concentration, MSC, PMSC, Morphology, Motility, Progressive Motility, Non-Progressive Motility, and Immotile were reported, with the conclusion that they "met the acceptance criteria." For instance, Total %CV for Sperm Concentration ranged from 1.5% to 14.1%, for MSC 0.0% to 41.6%, for PMSC 4.0% to 173.2% (with some very high %CVs for low-level samples), for Morphology 6.5% to 244.9% (with some very high %CVs for low-level samples), for Motility 4.2% to 11.0%, for Progressive Motility 6.1% to 261.7% (with some very high %CVs for low-level samples), for Non-Progressive Motility 6.4% to 76.7% (with some high %CVs for low-level samples), and for Immotile 1.8% to 10.4%. The conclusion states all met acceptance criteria, suggesting that higher %CV for low-level samples was considered acceptable within the context of clinical relevance for those low values.
Method ComparisonPassing-Bablok Regression: Intercept, Slope, Correlation CoefficientSlopes, y-intercepts, and correlation coefficients, along with the 95% confidence intervals, demonstrated clinical equivalence to the predicate device (specific ranges not provided in extract).CONCENTRATION: Intercept 0.05 (-0.4799 to 0.2610), Slope 0.98 (0.9718 to 0.9836), Correlation 1.0 (0.9974 to 0.9982).MOTILITY: Intercept 2.1 (1.2174 to 3.0000), Slope 0.9 (0.9189 to 0.9565), Correlation 0.96 (0.9493 to 0.9659).PROGRESSIVE MOTILITY: Intercept -0.7 (-1.4516 to 0.0000), Slope 1.0 (0.9286 to 0.9677), Correlation 1.0 (0.9683 to 0.9787).NON-PROGRESSIVE MOTILITY: Intercept -0.3 (-1.0000 to 0.0000), Slope 1.3 (1.2500 to 1.4000), Correlation 0.7 (0.6944 to 0.7850).IMMOTILE: Intercept 4.0 (3.0417 to 5.0000), Slope 0.9 (0.9200 to 0.9583), Correlation 0.9 (0.9130 to 0.9411).MORPHOLOGY: Intercept -1.0 (-1.0000 to -0.0455), Slope 1.0 (0.9091 to 1.0000), Correlation 1.0 (0.9563 to 0.9706).MSC: Intercept 0.3 (0.05708 to 0.5580), Slope 0.9 (0.9344 to 0.9571), Correlation 1.0 (0.9889 to 0.9925).PMSC: Intercept -0.3 (-0.5450 to -0.0968), Slope 0.9 (0.9149 to 0.9364), Correlation 1.0 (0.9894 to 0.9929).

2. Sample Size and Data Provenance

  • Sample Size for Test Set:
    • CLIA Waived User Precision Study (Control Material): 270 measurements in total (3 sites x 9 users (3 per site) over 3 days per site x 3 levels x 10 replicates of each level).
    • CLIA Waived User Precision Study (Native Samples): 216 measurements total (9 native semen samples x 2 replicates per sample x 3 users/site x 4 time points).
    • Method Comparison Study: 380 donor semen samples.
  • Data Provenance (Country of Origin and Retrospective/Prospective):
    • The Method Comparison Study was conducted across "Three U.S. sites."
    • The Precision studies were also multi-site, with the control material study having "3 sites". The native sample precision study was "across two sites."
    • The data appears to be prospectively collected for the purpose of these studies, as detailed study designs are provided, including number of sites, users, days, replicates, and samples. The samples used in the method comparison were "donor semen samples."

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts:
    • For the Method Comparison Study, there were "One or more TRAINED OPERATORS per site" (3 sites) who generated reference SQA-V results.
  • Qualifications of Experts:
    • The experts (TRAINED OPERATORS) were described as "fully trained and considered appropriate for generating reference SQA-V results." Their specific professional qualifications (e.g., medical technologists, clinical lab scientists) or years of experience are not explicitly stated.

4. Adjudication Method for the Test Set

  • The document implies that the ground truth for the method comparison study was established by the "TRAINED OPERATORS" using the predicate device (SQA-V). There is no mention of an adjudication process (e.g., 2+1, 3+1 consensus) among multiple experts to establish a "true" ground truth beyond the output of the predicate device operated by trained users. The samples were assayed "in singleton and in a blinded fashion" using both methods, suggesting a direct comparison rather than multi-reader adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No explicit MRMC comparative effectiveness study was described in terms of human readers improving with AI vs. without AI assistance. The study compares the performance of a new device (SQA-iOw operated by waived users) against a predicate device (SQA-V operated by trained users). It's a method comparison for an automated device, not an AI-assisted human reader study.

6. Standalone (Algorithm Only) Performance

  • The SQA-iOw is described as an "automated analyzer" that "utilizes proprietary software code to both perform analysis of semen parameters" and "applies proprietary algorithms to interpret and express these electrical signals and report them as various semen parameters." The performance measurements detailed (precision studies and method comparison) represent the standalone performance of the device/algorithm in processing samples and generating results for the specified semen parameters. There is no human-in-the-loop component in the measurement process itself.

7. Type of Ground Truth Used

  • The ground truth for the Method Comparison Study was established using the results from the predicate device (SQA-V) operated by trained users. This serves as a "reference standard" or "comparative method" rather than an absolute ground truth such as pathology or outcomes data.
  • For the Precision Studies, the ground truth is statistical variability around the mean measurements of control materials and native samples.

8. Sample Size for the Training Set

  • The document does not provide information on the sample size used for the training set for the SQA-iOw's algorithms. The studies described are validation (test set) studies, not algorithm development or training data descriptions.

9. How Ground Truth for Training Set was Established

  • The document does not provide information on how the ground truth for the training set was established, as it focuses on the validation studies. It only mentions that the device "applies proprietary algorithms" but not how these algorithms were developed or trained.

FDA 510(k) Clearance Letter - SQA-iOw Sperm Quality Analyzer

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 2, 2025

Medical Electronic Systems Ltd.
Taly Vider Cohen
Regulatory Affairs and IP Director
Alon Hatavor St. 20
Caesarea, 3088900
Israel

Re: K243114
Trade/Device Name: SQA-iOw Sperm Quality Analyzer
Regulation Number: 21 CFR 864.5220
Regulation Name: Automated differential cell counter
Regulatory Class: Class II
Product Code: POV
Dated: January 30, 2025
Received: January 30, 2025

Dear Taly Vider Cohen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243114 - Taly Vider Cohen Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243114 - Taly Vider Cohen Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu -S

Min Wu, Ph.D.
Branch Chief
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K243114
Device Name: SQA-iOw Sperm Quality Analyzer

Indications for Use (Describe)

The SQA-iOw Sperm Quality Analyzer is an automated analyzer intended for in-vitro diagnostic use to determine the following parameters in semen:

Measured parameters:

  • Sperm Concentration/ Total Sperm Concentration, millions/mL
  • Motile Sperm Concentration (MSC), millions/mL
  • Progressively Motile Sperm Concentration (PMSC), millions/mL (combines Rapidly and Slowly Progressive Motile Sperm Concentration, millions/mL)
  • Normal Forms (% Normal Morphology), %

Derived parameters:

  • Total Motility / Total Motile (PR + NP), %
  • Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
  • Non-Progressive (NP), %
  • Immotile (IM), %

The SQA-iOw is intended for CLIA Waived settings. The SQA-iOw does not provide a comprehensive evaluation of a male's fertility status and is intended for in vitro use only.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 1 of 8

Date: 02.MAY.2025

SQA-iOw 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

807.92 (a)(1):

Name: Medical Electronic Systems, LTD
Address: ALON HATAVOR ST. 20 ZONE 6
CAESAREA INDUSTRIAL PARK
CAESAREA, 38900, ISRAEL
Phone: 972 54 209-1712
Contact: Ms. Taly Vider Cohen

807.92 (a)(2): Device name, trade name and common name, and Classification

Trade Name: SQA-iOw Sperm Quality Analyzer
Common Name: SQA-iOw Sperm Quality Analyzer
Classification: Class II, POV
21 CFR 864.5220

807.92 (a)(3): Identification of the legally marketed predicate devices

SQA-iOw Sperm Quality Analyzer is substantially equivalent to a predicate device, the SQA-V sperm analyzer, (MES Ltd., Israel), cleared under K021746, September 20, 2002. Both of these testing devices utilize human sperm to measure a variety of male fertility factors.

807.92 (a)(4): Device Description

The SQA-iOw Sperm Quality Analyzer is a PC-based analytical medical device that tests human semen samples. The device works with a computer application that manages the device, and information related to the patient, the sample, the test results and the facility.

After collection and preparation, 0.6 mL of semen sample is aspirated into a disposable SQA capillary sample delivery system and inserted into the SQA-iOw measurement chamber. The testing process takes approximately 75 seconds. The system performs an automatic self-test and auto-calibration upon start up, and checks device stability before each sample is run.

The SQA-iOw Sperm Quality Analyzer utilizes proprietary software code to both perform analysis of semen parameters and present those results on the user interface. This software is installed on a PC as a cloud-based application ("app") and is designed to perform all functions and features of the SQA-iO device, controlled by the user through a proprietary graphical user interface (GUI).

The SQA-iOw Sperm Quality Analyzer software analyzes semen parameters using signal processing technology. Sample testing is performed by capturing electrical signals as sperm moves through a light source in the SQA-iO optical block. These light disturbances are converted into electrical signals which are then analyzed by the SQA-iOw software. The SQA-iOw software applies proprietary algorithms to interpret and express these electrical signals and report them as various semen parameters.

The SQA-iOw Sperm Quality Analyzer package provides the SQA-iOw device and USB cable. SQA disposable capillaries, cleaning kits and related testing supplies and test kits are supplied individually.

Page 6

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 2 of 8

807.92 (a)(5): Intended Use

The SQA-iOw Sperm Quality Analyzer is an automated analyzer intended for in-vitro diagnostic use to determine the following parameters in semen:

Measured parameters:

  • Sperm Concentration/ Total Sperm Concentration, millions/mL
  • Motile Sperm Concentration (MSC), millions/mL
  • Progressively Motile Sperm Concentration (PMSC), millions/mL (combines Rapidly and Slowly Progressive Motile Sperm Concentration, millions/mL)
  • Normal Forms (% Normal Morphology), %

Derived parameters:

  • Total Motility / Total Motile (PR + NP), %
  • Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
  • Non-Progressive (NP), %
  • Immotile (IM), %

The SQA-iOw is intended for CLIA Waived settings. The SQA-iOw does not provide a comprehensive evaluation of a male's fertility status and is intended for in vitro use only.

807.92 (a)(6): Technological Similarities and Differences to the Predicate

SQA-iOw Sperm Quality Analyzer is substantially equivalent to the predicate device; SQA-V Sperm Quality Analyzer, (MES Ltd., Israel), cleared under K021746. The SQA-iOw is substantially equivalent in terms of general intended use, sample type, male fertility factor measurements, and in vitro use. The various device features are compared in the table below.

SQA-iOw vs. SQA-V Predicate

ElementNew product SQA-iOw Sperm Quality AnalyzerPredicate SQA-V: 510(k) K021746
Intended useSee "intended use" section above.The SQA-V is a point-of-care, in vitro use, electro-optical medical device with on-screen visualization for semen analysis performed by healthcare professionals (trained lab technicians- device is CLIA moderate). The SQA-V does not provide a comprehensive evaluation of a male's fertility status.
Intended UserHealthcare professionals (CLIA waived, moderate, and high complexity settings)Healthcare professionals (trained lab technicians, CLIA moderate or high complexity)

Page 7

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 3 of 8

ElementNew product SQA-iOw Sperm Quality AnalyzerPredicate SQA-V: 510(k) K021746
Semen parametersThe SQA-iOw reports the following directly measured quantitative semen parameters:1. Sperm Concentration/ Total Sperm Concentration, millions/mL2. Motile Sperm Concentration (MSC), millions/mL3. Progressively Motile Sperm Concentration (PMSC), millions/mL (combines Rapidly and Slowly Progressive Motile Sperm Concentration, millions/mL)4. Normal Forms (% Normal Morphology), %The SQA-iOw also reports the following derived semen parameters:5. Total Motility / Total Motile (PR + NP), %6. Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)7. Non-Progressive (NP), %8. Immotile (IM), %The SQA-V reports the same Semen Parameters
WHO complianceWHO 6WHO 4/ WHO 5
Sample typeHuman semenSame
Male fertility factorYesSame
TechnologyDesk-top unit consists of a light source and optical sensors, connected to a PC that runs the software containing algorithms for the assessment of semen parameters. This software is installed on the user's PC as a cloud-based application. See additional explanations in the discussion section below.Desk-top unit consists of a light source, optical sensors, built-in video microscopy and an internal computer containing algorithms for the assessment of semen parameters.
Testing capillarySQA-V testing capillarySame
External controlsUse of QwikCheck Beads (cleared under K041600) for performing quality controlSame

(b)(1): Brief Description of Nonclinical Data

  • Linearity/dynamic ranges (Assay measuring range)- See K220828.
  • Analytical specificity (interference)- See K220828.
  • Sample stability- see K220828.
  • Analytical sensitivity (detection limits: limits of blank and detection/quantitation)- See K220828.

807.92 (b)(2): Brief Description of Clinical Data

Clinical Precision

MES performed two waived-user precision studies as follows:

  1. Precision using control material
  2. Precision using native samples.

Page 8

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 4 of 8

1. CLIA Waived user Precision study using control material

Objective and Study Design

To estimate precision/repeatability of the SQA-iOw device based on the CLSI EP05-A3 standard, a study was designed as follows: 3 sites x 9 users (3 per site) over 3 days per site x 3 levels x 10 replicates of each level (5 runs in duplicate) = 270 measurements in total. Three levels of bead controls (same lots across sites) and three lots of capillaries (different lots across sites) were used. Data analyses included ANOVA analysis with the following components: within-run, between-operator, between-day, and between-site repeatability using standard deviations (SDs) and Coefficients of Variation (CV, %).

Table 1: Controls Level 1

SiteNMeanRepeatability (Within-run)Between-dayBetween-run/ OperatorBetween site / instrumentTotal
SD%CVSD%CVSD
All sites combined9047.80.120.24%0.400.83%0.35

Table 2: Controls Level 2

SiteNMeanRepeatability (Within-run)Between-dayBetween-run/ OperatorBetween site / instrumentTotal
SD%CVSD%CVSD
All sites combined9024.90.060.23%0.572.28%0.45

Table 3: Negative Control

SiteNMeanRepeatability (Within-run)Between-dayBetween-run/ OperatorBetween site / instrumentTotal
SD%CVSD%CVSD
All sites combined900.00.000.00%0.000.00%0.00

Conclusion

The StDev and %CV met the acceptance criteria for the three levels.

2. CLIA Waived user Precision study using native samples

Objective and Study Design

To determine precision of the SQA-iOw device based on CLSI EP05-A3 guidance, as follows, across two sites: 9 total native semen samples x 2 reps per sample x 3 users/site x 4 time points = 216 measurements total, 24 results per sample. Precision analysis included within-run, between-operator standard deviations (SD) and Coefficients of Variation (CV, %), and total imprecision (reproducibility) using the estimated components.

Page 9

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 5 of 8

Results

The following precision results were obtained by the waived users testing native semen samples using SQA-iOw Waiver devices:

Table 4: SQA-iOw Sperm Concentration Precision

Note: SDs are used for low-level samples (<10)

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
12455.421.1252.0%0.925
224116.882.9132.5%1.370
32488.581.6011.8%1.773
424228.401.8410.8%3.697
52497.630.8110.8%1.215
62415.811.3168.3%1.807
72414.700.7405.0%1.775

Table 5: SQA-iOw Motile Sperm Concentration (MSC) Precision

Note: SDs are used for low-level samples (<10)

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
12427.060.9763.6%0.230
22457.632.1373.7%1.000
32441.781.3673.3%0.428
424120.413.4102.8%2.324
52432.433.1699.8%1.906
6241.600.0000.0%0.000
7241.600.0000.0%0.000
8249.420.6326.7%1.219
9244.320.62714.5%1.684

Table 6: SQA-iOw Progressive Motile Sperm Concentration (PMSC) Precision

Note: SDs are used for low-level samples (<10)

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
12422.500.9214.1%0.432
22449.032.5515.2%1.129
32433.521.5354.6%0.447
424108.703.6103.3%2.497
52424.693.10312.6%1.915
6240.030.041163.3%0.014
7240.090.245267.2%0.088
8246.800.5357.9%0.652
9242.970.53818.1%1.083

Page 10

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 6 of 8

Table 7: SQA-iOw Morphology Precision

Note: SDs are used for low-level samples (<10)

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
1247.670.7079.2%0.354
2247.960.7369.2%0.382
3247.080.5007.1%0.204
4249.670.4084.2%0.479
5244.040.61215.2%0.354
6240.130.204163.3%0.204
7240.330.764229.1%0.289
8248.380.7368.8%0.500
9243.500.91326.1%0.722

Table 8: SQA-iOw Motility Precision

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
12448.832.2904.7%1.145
22449.372.7575.6%1.378
32447.243.0316.4%1.516
42452.734.4048.4%2.202
52433.242.6417.9%1.320
62429.661.1093.7%0.555
72431.080.9553.1%0.478
82459.593.0395.1%1.519
92430.913.0289.8%1.514

Table 9: SQA-iOw Progressive Motility Precision

Note: SDs are used for low-level samples (<10)

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
12440.592.2195.5%1.110
22442.032.8806.9%1.440
32437.933.0738.1%1.536
42447.604.4969.4%2.248
52425.312.59410.2%1.297
6240.471.111234.1%0.556
7241.751.02958.7%0.514
82442.972.8826.7%1.441
92421.242.85213.4%1.426

Page 11

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 7 of 8

Table 10: SQA-iOw Non-progressive Motility Precision

Note: SDs are used for low-level samples (<10)

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
1248.242.50430.4%1.252
2247.342.82238.4%1.411
3249.302.93031.5%1.465
4245.133.52068.6%1.760
5247.932.71334.2%1.357
62429.191.7586.0%0.879
72429.331.6725.7%0.836
82416.622.87517.3%1.438
9249.672.85029.5%1.425

Table 11: SQA-iOw Immotile Precision

Sample #NMeanRepeatabilityBetween-operatorTotal
SD%CVSD
12451.172.2904.5%1.145
22450.632.7575.4%1.378
32452.763.0315.7%1.516
42447.274.4049.3%2.202
52466.762.6414.0%1.320
62470.341.1091.6%0.555
72468.920.9551.4%0.478
82440.413.0397.5%1.519
92469.093.0284.4%1.514

Conclusion

The StDev and %CV met the acceptance criteria for all reported parameters for the samples.

Method Comparison Study

Objective:

To demonstrate clinical equivalence of the SQA-iOw device operated by WAIVED USERS vs. the SQA-V performed by TRAINED USERS when testing matched native human semen samples.

Study Design:

  • Sites: Three U.S. sites
  • Waived testing: 9 WAIVED USERS (3 per site).
  • Comparative testing: One or more TRAINED OPERATORS per site who are fully trained and considered appropriate for generating reference SQA-V results.
  • Sample size: A total of 380 donor semen samples distributed approximately equally across 3 sites and 9 operators and assayed in singleton and in a blinded fashion using both methods.

Page 12

Medical Electronic Systems
SQA-iOw Sperm Quality Analyzer
Page 8 of 8

Parameters for comparison in the clinical studies:

Directly measured parameters:

  • Sperm Concentration/ Total Sperm Concentration, millions/ml
  • Motile Sperm Concentration (MSC), millions/ml
  • Progressively Motile Sperm Concentration (PMSC), millions/ml (combines Rapidly and Slowly Progressive Motile Sperm Concentration, millions/mL)
  • Normal Forms (% Normal Morphology), %

Calculated parameters:

  • Total Motility / Total Motile (PR + NP), %
  • Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
  • Non-Progressive (NP), %
  • Immotile (IM), %

Data Analysis: Statistical Methods and Acceptance Criteria

Stats: Passing-Bablok regression with "lab results" (SQA-V results obtained by the trained user) on the x-axis. Slopes, y-intercepts, and correlation coefficients, along with the 95% confidence intervals, were reported.

Results:

Table 12: SQA-iOw Intended User vs. SQA-V Expert User

ParameterNRangeIntercept (95% CI)Slope (95% CI)Correlation (95% CI)
CONCENTRATION, M/mL3806.0 – 272.5-0.05 (-0.4799 to 0.2610)0.98 (0.9718 to 0.9836)1.0 (0.9974 to 0.9982)
MOTILITY, %3800.0 – 100.02.1 (1.2174 to 3.0000)0.9 (0.9189 to 0.9565)0.96 (0.9493 to 0.9659)
PROGRESSIVE MOTILITY, %3800.0 – 88.0-0.7 (-1.4516 to 0.0000)1.0 (0.9286 to 0.9677)1.0 (0.9683 to 0.9787)
NON-PROGRESSIVE MOTILITY, %3800.0 – 30.0-0.3 (-1.0000 to 0.0000)1.3 (1.2500 to 1.4000)0.7 (0.6944 to 0.7850)
IMMOTILE, %3800.0 – 99.04.0 (3.0417 to 5.0000)0.9 (0.9200 to 0.9583)0.9 (0.9130 to 0.9411)
MORPHOLOGY, %3800.0 – 22.0-1.0 (-1.0000 to -0.0455)1.0 (0.9091 to 1.0000)1.0 (0.9563 to 0.9706)
MSC, %3800.2 – 151.70.3 (0.05708 to 0.5580)0.9 (0.9344 to 0.9571)1.0 (0.9889 to 0.9925)
PMSC, %3800.0 – 139.5-0.3 (-0.5450 to -0.0968)0.9 (0.9149 to 0.9364)1.0 (0.9894 to 0.9929)

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

The SQA-iOw testing confirms that the device can be used according to its intended use. The information and data provided in this 510(k) submission identifies no new safety or effectiveness issues for this device type. Therefore, the SQA-iO is safe and performs effectively based on its intended use.

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”