K Number
K220828
Date Cleared
2023-08-07

(503 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SQA-iO Sperm Quality Analyzer is an automated point-of-care in vitro use only medical device for semen analysis performed by healthcare professionals (trained lab technicians). The SQA-iO does not provide a comprehensive evaluation of a male's fertility status.

The SQA-iO provides direct and calculated quantitative measurements for the following parameters:

Directly measured parameters:

  • Sperm Concentration, M/mL ●
  • . Motile Sperm Concentration (MSC), M/mL
  • . Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapid and Slow PMSC, millions/mL)
  • . Normal Forms (Normal Morphology), %

Calculated parameters:

  • Total Motility (PR + NP), %
  • . Progressive Motility (PR), % (combines Rapidly and Slowly Progressive Motility, %)
  • Non-Progressive Motility (NP), % .
  • Immotile (IM), %
  • Functional Sperm Concentration (FSC), millions/mL ●
Device Description

The SQA-iO is a PC-based analytical medical device that tests human semen samples. The device works with a computer application that manages the device, and information related to the patient, the sample, test results and the facility.

After collection and preparation, 0.6 mL of semen sample is aspirated into a disposable SQA capillary sample delivery system and inserted into the SQA-iO measurement chamber. The testing process takes approximately 75 seconds. The system performs an automatic self-test and auto-calibration upon start up, and checks device stability before each sample is run.

The SQA-iO utilizes proprietary software code to both perform analysis of semen parameters and to present those results on the user interface. This software is installed on the user's PC as a cloud-based application ("app") and is designed to perform all functions and features of the SQA-iO device, controlled by the user through a proprietary graphical interface (GUI).

The SQA-iO software analyzes semen parameters using signal processing technology. Sample testing is performed by capturing electrical signals as sperm moves through a light source in the SQA-iO optical block. These light disturbances are converted into electrical signals which are then analyzed by the SQA-iO software. The SQA-iO software applies proprietary algorithms to interpret and express these electrical signals and report them as various semen parameters.

The SQA-iO package provides all the supplies necessary to perform semen analysis: SQA-iO device, USB cable, SQA disposable capillaries, and a cleaning kit.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" for the clinical study endpoints for the SQA-iO. However, it presents the results of a method comparison study against a predicate device (SQA-V) and indicates the desired range for the statistical measures (slopes, y-intercepts, and correlation coefficients) to demonstrate substantial equivalence.

ParameterAcceptance Criteria (Implied)Reported Device Performance (SQA-iO vs. SQA-V)
Concentration, M/mLSlopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -1.5 (CI: -2.0 to -0.7), Slope: 1.0 (CI: 1.0 to 1.0), Correlation: 1.0 (CI: 0.98 to 0.99)
Motility, %Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -3.0 (CI: -3.1 to -1.7), Slope: 1.0 (CI: 1.0 to 1.0), Correlation: 1.0 (CI: 0.95 to 0.97)
Progressive Motility, %Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -0.8 (CI: -1.0 to 0.0), Slope: 0.9 (CI: 0.9 to 1.0), Correlation: 1.0 (CI: 0.97 to 0.98)
Rapidly Progressive, %Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: 0.1 (CI: 0.0 to 0.3), Slope: 1.0 (CI: 0.9 to 1.0), Correlation: 0.9 (CI: 0.90 to 0.94)
Slowly Progressive, %Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -0.8 (CI: -1.0 to 0.0), Slope: 1.0 (CI: 0.9 to 1.0), Correlation: 0.9 (CI: 0.86 to 0.93)
Non-Progressive, %Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -1.9 (CI: -3.0 to -1.0), Slope: 1.2 (CI: 1.0 to 1.3), Correlation: 0.8 (CI: 0.71 to 0.83)
Immotile, %Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: 3.0 (CI: 1.0 to 5.0), Slope: 1.0 (CI: 1.0 to 1.0), Correlation: 1.0 (CI: 0.95 to 0.97)
MSC, M/mLSlopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -0.9 (CI: -1.7 to -0.6), Slope: 1.0 (CI: 1.0 to 1.0), Correlation: 1.0 (CI: 0.98 to 0.99)
PMSC, M/mLSlopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -0.4 (CI: -0.7 to -0.3), Slope: 1.0 (CI: 0.9 to 1.0), Correlation: 1.0 (CI: 0.99 to 1.00)
Rapid PMSC, M/mLSlopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: 0.0 (CI: -0.1 to 0.0), Slope: 1.0 (CI: 1.0 to 1.0), Correlation: 1.0 (CI: 0.96 to 0.98)
Slow PMSC, M/mLSlopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -0.1 (CI: -0.4 to -0.1), Slope: 1.0 (CI: 0.9 to 1.0), Correlation: 1.0 (CI: 0.98 to 0.99)
Morphology, % (n = 155)Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: 0.0 (CI: 0.0 to 0.1), Slope: 1.0 (CI: 0.9 to 1.0), Correlation: 1.0 (CI: 0.96 to 0.98)
FSC, M/mL (n = 155)Slopes between 0.9 and 1.2, Y-intercepts near zero, r ≥ 0.8Intercept: -0.1 (CI: -0.1 to 0.0), Slope: 0.9 (CI: 0.9 to 1.0), Correlation: 1.0 (CI: 0.97 to 0.99)
Conclusion from resultsAll criteria metThe data demonstrate slopes between 0.9 and 1.2, Y-intercepts near zero, and a correlation coefficient ("r") ≥ 0.8, indicating the device meets the criteria for substantial equivalence to the predicate device.

2. Sample size used for the test set and data provenance:

  • Test Set Sample Size: 165 matched native semen samples (for most parameters; 155 for Morphology and FSC).
  • Data Provenance: The study used "native human semen samples" from "consented donors" which suggests prospective collection. The study was conducted at "three sites," but the specific country of origin is not explicitly stated. The submitting company, Medical Electronic Systems LTD, is based in Israel, which could imply data from Israel or multinational sites.

3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

  • Number of experts: 12 SQA-iO operators across all sites who are "laboratory technicians or professionals who are familiar with semen analysis."
  • Adjudication Method for Ground Truth (Predicate Device Results): An "expert operator assayed the same sample in duplicate on the SQA-V predicate, and the means of the SQA-V results were used as the comparator results." This implies a form of internal consistency/averaging from a single expert reading (on the predicate device), rather than a multi-expert consensus for the ground truth itself.

4. Adjudication method for the test set:

The term "adjudication method" typically refers to resolving discrepancies between multiple readers or between an AI and a reader. In this context, the study compares the SQA-iO device's measurements against the a predicate device (SQA-V). The predicate device's results were established by an "expert operator" in duplicate, and the mean of these results was used as the comparator. This is not an adjudication method in the traditional sense of resolving expert disagreement for a ground truth label, but rather establishing a reference measurement from the predicate.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, an MRMC comparative effectiveness study was not done. The study performed was a method comparison study between the new device (SQA-iO) and a predicate device (SQA-V), not a study evaluating human reader performance with and without AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the study describes the standalone performance of the SQA-iO device. The SQA-iO is an automated device that provides "direct and calculated quantitative measurements" for various sperm parameters. The comparison was primarily between the SQA-iO's automated output and the SQA-V's output (also an automated device, with human operators for loading/starting). While human operators (lab technicians) interact with the SQA-iO, the device itself generates the quantitative results without human interpretation of the images or signals for the reported parameters.

7. The type of ground truth used:

The ground truth for the comparison study was the measurements obtained from the predicate device, the SQA-V sperm analyzer, which was operated by an "expert operator" in duplicate. This is a reference device comparison rather than a direct comparison to an absolute "ground truth" established by a gold standard like pathology or long-term outcomes. The SQA-V's results were treated as the established benchmark.

8. The sample size for the training set:

The document does not explicitly state the sample size for the training set used for the SQA-iO's proprietary algorithms. The clinical data section focuses on the performance of the device after its development. The non-clinical data mentions using "native human semen samples" for various bench studies, but these are for analytical performance validation (precision, linearity, etc.), not explicitly described as a training set for algorithm development.

9. How the ground truth for the training set was established:

Not explicitly mentioned in the document. The document describes analytical and clinical validation studies. Information about the specific methods for establishing ground truth during the original training and development of the SQA-iO's proprietary algorithms is not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 7, 2023

Medical Electronic Systems LTD Taly Cohen Regulatory Affairs and IP Director Alon Hatavor 20, Zone 6, Caesarea Industrial Park Caesarea. 3088900 Israel

Re: K220828

Trade/Device Name: SQA-iO Sperm Quality Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: POV Dated: March 10, 2022 Received: March 22, 2022

Dear Taly Cohen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Min Wu-S

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220828

Device Name SOA-iO Sperm Quality Analyzer

Indications for Use (Describe)

The SQA-iO Sperm Quality Analyzer is an automated point-of-care in vitro use only medical device for semen analysis performed by healthcare professionals (trained lab technicians). The SQA-iO does not provide a comprehensve evaluation of a male's fertility status.

The SQA-iO provides direct and calculated quantitative measurements for the following parameters:

Directly measured parameters:

  • Sperm Concentration, M/mL ●
  • . Motile Sperm Concentration (MSC), M/mL
  • . Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapid and Slow PMSC, millions/mL)
  • . Normal Forms (Normal Morphology), %

Calculated parameters:

  • Total Motility (PR + NP), %
  • . Progressive Motility (PR), % (combines Rapidly and Slowly Progressive Motility, %)
  • Non-Progressive Motility (NP), % .
  • Immotile (IM), %
  • Functional Sperm Concentration (FSC), millions/mL

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over The Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------

| X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K220828- 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

807.92 (a)(1):Name:Medical Electronic Systems, LTD
Address:ALON HATAVOR ST. 20 ZONE 6CAESAREA INDUSTRIAL PARKCAESAREA, 38900, ISRAEL
Phone:FAX:Contact:972 54 209-1712NAMs. Taly Vider Cohen

807.92 (a)(2): Device name- trade name and common name, and

Classification/Trade Name:

SQA-iO Sperm Quality Analyzer

Common Name: SQA-iO Sperm Quality Analyzer

Class II, POV Classification: 21 CFR 864.5220

807.92 (a)(3): Identification of the legally marketed predicate devices

SQA-iO is substantially equivalent to a predicate device, the SQA-V sperm analyzer, (MES Ltd., Israel), cleared under K021746, September 20, 2002. Both of these testing devices utilize fresh human sperm to measure a variety of male fertility factors.

807.92 (a)(4): Device Description

The SOA-iO is a PC-based analytical medical device that tests human semen samples. The device works with a computer application that manages the device, and information related to the patient, the sample, test results and the facility.

After collection and preparation, 0.6 mL of semen sample is aspirated into a disposable SQA capillary sample delivery system and inserted into the SOA-iO measurement chamber. The testing process takes approximately 75 seconds. The system performs an automatic self-test and auto-calibration upon start up, and checks device stability before each sample is run.

The SQA-iO utilizes proprietary software code to both perform analysis of semen parameters and to present those results on the user interface. This software is installed on the user's PC as a cloud-based application ("app") and is designed to perform all functions and features of the SQA-iO device, controlled by the user through a proprietary graphical interface (GUI).

The SQA-iO software analyzes semen parameters using signal processing technology. Sample testing is performed by capturing electrical signals as sperm moves through a light source in the SQA-iO optical block. These light disturbances are converted into electrical signals which are then analyzed by the SOAiO software. The SQA-iO software applies proprietary alqorithms to interpret and express these electrical signals and report them as various semen parameters.

The SQA-iO package provides all the supplies necessary to perform semen analysis: SQA-iO device, USB cable, SQA disposable capillaries, and a cleaning kit.

{4}------------------------------------------------

Medical Electronic Systems SQA-iO Sperm Quality Analyzer

807.92 (a)(5): Intended Use

The SQA-iO Sperm Quality Analyzer is an automated point-of-care in vitro use only medical device for semen analysis performed by healthcare professionals (trained lab technicians). The SQA-iO does not provide a comprehensive evaluation of a male's fertility status.

The SQA-iO provides direct and calculated quantitative measurements for the following parameters:

Directly measured parameters:

  • · Sperm Concentration, M/mL
  • Motile Sperm Concentration (MSC), M/mL .
  • Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapid and Slow . PMSC, millions/mL)
  • Normal Forms (Normal Morphology), %

Calculated parameters:

  • Total Motility (PR + NP), % .
  • . Progressive Motility (PR), % (combines Rapidly and Slowly Progressive Motility, %)
  • Non-Progressive Motility (NP), %
  • . Immotile (IM), %
  • Functional Sperm Concentration (FSC), millions/mL ●

807.92 (a)(6): Technological Similarities and Differences to the Predicate

SQA-iO is substantially equivalent to an FDA-cleared predicate device- SQA-V sperm quality analyzer, (MES Ltd., Israel), cleared under K021746, September 20, 2002. SQA-iO is substantially equivalent to this product in terms of general intended use, sample type, male fertility factor measurements, and in vitro use. Further, the SQA-iO is substantially equivalent to the SQA-V in terms of the assessed parameters.

SQA-iO vs. SOA-V Predicate

ElementNew productSQA-iO Sperm Quality AnalyzerPredicateSQA-V: 510(k) K021746
Intended useThe SQA-iO Sperm Quality Analyzer is an automatedpoint-of-care in vitro use medical device for semenanalysis performed by healthcare professionals(trained lab technicians). The SQA-iO does not providea comprehensive evaluation of a male's fertility status.The SQA V is a point-of-care, in vitrouse, electro-optical medical device withon-screen visualization for semenanalysis performed by healthcareprofessionals (trained lab technicians).
SemenparametersThe SQA-iO Sperm Quality Analyzer is an automatedpoint-of-care in vitro use only medical device forsemen analysis performed by healthcare professionals(trained lab technicians). The SQA-iO does notprovide a comprehensive evaluation of a male'sfertility status.The SQA-iO provides direct and calculated quantitativemeasurements for the following parameters:Directly measured parameters:• Sperm Concentration, M/mLSame

{5}------------------------------------------------

ElementNew productSQA-iO Sperm Quality AnalyzerPredicateSQA-V: 510(k) K021746
Motile Sperm Concentration (MSC), M/mL Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapid and Slow PMSC, millions/mL) Normal Forms (Normal Morphology), % Calculated parameters: Total Motility (PR + NP), % Progressive Motility (PR), % (combines Rapidly and Slowly Progressive Motility, %) Non-Progressive Motility (NP), % Immotile (IM), % Functional Sperm Concentration (FSC), millions/mL
Sample typeHuman semenSame
Male fertility factorYesSame
TechnologyDesk-top unit consists of a light source and optical sensors, connected to a PC that runs the software containing algorithms for the assessment of semen parameters. This software is installed on the user's PC as a cloud-based application.Desk-top unit consists of a light source, optical sensors, built-in video microscopy and an internal computer containing algorithms for the assessment of semen parameters.
External controlsUse of QwikCheck Beads (cleared under K041600) for performing quality controlSame

807.92(b)(1): Brief Description of Nonclinical Data

Medical Electronic Systems (MES) has conducted a series of analytical (bench) studies in support of the analytical claims of the SQA-iO. These studies were all performed in-house and included evaluations for: precision- Repeatability (Native Sample precision- Reproducibility (Controls precision), analytical sensitivity (limits of blank and detection/guantitation), linearity/dynamic ranges, analytical specificity (interference), sample stability, and cleaning robustness.

All the studies utilized native human semen samples (including seminal plasma for negative samples), with some samples being manually diluted with seminal plasma to achieve the necessary quantitation ranges. Semen samples were collected following WHO 6** ed. manual quidance for sample handling from consented donors and were assayed in a blinded fashion on the SQA-iO, and in parallel on the SQA-V comparator device, as applicable.

Precision- Repeatability (Native Sample Precision)

Eight samples, spanning low, middle and high levels of five parameters (concentration, motility, MSC, PMSC, and morphology, were assayed in duplicate on five devices over four time-points. Due to the limited stability of semen samples, each "day" in the statistical analysis represented different times of day with

{6}------------------------------------------------

Medical Electronic Systems SQA-iO Sperm Quality Analyzer

testing at the 0, 20, 40, and 60-minute time points resulting in: 2 replicates x 4 days = 40 replicate results per sample, for each of the five parameters.

Data analyses were performed to provide: within-run, between-day, betweenoperator/lot/instrument, and total precision and standard deviations (SDs) and percent coefficients of variation (%CVs) were calculated for each sample.

ConcentrationWithin-RunBetween-RunBetween-DayBetween-Operator/Lot/InstrumentTotal
SampleNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
1408.50.637.4%0.617.2%0.252.9%0.607.1%0.627.3%
24034.51.664.8%1.704.9%0.772.2%1.313.8%1.765.1%
34045.43.257.2%3.307.3%1.663.7%3.096.8%3.467.6%
44058.53.125.3%3.075.2%1.041.8%2.113.6%3.045.2%
54062.22.423.9%2.383.8%1.422.3%2.303.7%2.644.2%
640181.65.252.9%5.352.9%3.421.9%3.832.1%5.873.2%
740227.65.872.6%6.252.7%5.452.4%3.481.5%7.583.3%
840212.93.741.8%4.422.1%4.872.3%2.671.3%5.792.7%

SQA-iO Sperm Concentration Precision

SQA-iO Motility Precision

MotilityWithin-RunBetween-RunBetween-DayBetween-Operator/Lot/InstrumentTotal
SampleNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
1400.00.000.0%0.000.0%0.000.0%0.000.0%0.000.0%
24077.02.823.7%2.743.6%1.201.6%2.593.4%2.873.7%
34062.32.624.2%2.594.2%0.741.2%2.273.7%2.544.1%
44080.60.991.2%1.001.2%0.460.6%0.831.0%1.011.3%
54058.03.836.2%4.657.7%3.235.6%2.604.5%6.9912.1%
64043.91.814.1%1.994.5%1.182.7%1.373.1%2.044.6%
74030.72.297.5%2.528.3%2.227.2%0.943.1%3.039.9%
84049.91.523.0%1.773.5%1.523.0%1.282.6%2.054.1%

{7}------------------------------------------------

Medical Electronic Systems SQA-iO Sperm Quality Analyzer

MSCWithin-RunBetween-RunBetween-DayBetween-Operator/Lot/InstrumentTotal
SampleNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
1402.00.000.0%0.000.0%0.000.0%0.000.0%0.000.0%
24026.51.315.0%1.365.1%1.054.0%0.682.6%1.606.0%
34027.91.405.0%1.555.5%1.033.7%1.083.9%1.676.0%
44047.02.996.4%2.996.4%1.132.4%2.274.8%2.976.3%
54035.51.424.0%1.564.4%0.772.2%1.273.6%1.544.3%
64079.42.873.6%3.544.5%2.413.0%1.091.4%3.604.5%
74069.34.266.2%5.057.3%4.296.2%1.372.0%5.858.4%
840106.23.433.2%4.484.2%5.305.0%2.182.1%6.125.8%

SQA-iO Motile Sperm Concentration (MSC) Precision

SQA-iO Progressively Motile Sperm Concentration (PMSC) Precision

PMSCWithin-RunBetween-RunBetween-DayBetween-Operator/Lot/InstrumentTotal
SampleNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
1400.00.000.0%0.000.0%0.000.0%0.000.0%0.000.0%
24023.21.114.8%1.144.9%0.944.1%0.743.2%1.386.0%
34024.21.275.2%1.355.6%0.833.4%0.903.7%1.415.8%
44042.22.806.6%2.816.7%1.162.8%2.115.0%2.826.7%
54031.51.785.6%1.865.9%0.762.4%1.113.5%1.926.1%
64070.32.643.8%3.344.8%2.343.3%0.921.3%3.404.8%
74051.04.609.1%5.3410.6%5.2010.2%2.514.9%6.5412.8%
84093.43.583.8%4.394.7%5.325.7%2.212.4%6.146.6%

SQA-iO Normal Morphology Precision

Normal MorphologyWithin-RunBetween-RunBetween-DayBetween-Operator/Lot/InstrumentTotal
SampleNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
1400.00.000.0%0.000.0%0.000.0%0.000.0%0.000.0%
24015.40.875.7%0.875.7%0.332.2%0.785.1%0.926.0%
34011.21.009.0%1.008.9%0.252.2%0.898.0%0.988.8%
44016.50.784.7%0.835.0%0.372.2%0.593.6%0.855.1%
54010.20.585.7%0.616.0%0.414.0%0.454.4%0.666.5%
6407.20.354.8%0.395.4%0.192.6%0.263.6%0.415.6%
7403.60.4211.9%0.4613.0%0.3910.7%0.226.2%0.5515.1%
8408.50.485.6%0.536.3%0.516.0%0.354.2%0.688.0%

{8}------------------------------------------------

Precision- Reproducibility (Controls Precision)

This study was designed to establish the reproducibility of the SQA-iO utilizing three levels of QwikCheck beads (controls) assayed by three different operators in three locations over a 5-day period. Six replicates

of each level of QwikCheck Beads, representing varying concentration levels (High, Normal and Negative) were included, and the output parameter was sperm concentration.

For each level of control tested, the mean, SD and %CV were calculated, along with the %CV for repeatability and reproducibility (total precision).

SiteControl LevelConcentration Mean, M/mlRepeatabilityBetween-Day
SD%CVSD%CV
1Level High L145.70.260.6%0.811.8%
1Level Normal L224.00.040.2%0.552.3%
1Level NEG L30.00.000.0%0.000.0%
2Level High L147.30.050.1%0.831.8%
2Level Normal L224.80.030.1%0.130.5%
2Level NEG L30.00.000.0%0.000.0%
3Level High L146.70.030.1%0.230.5%
3Level Normal L225.60.010.0%0.662.6%
3Level NEG L30.00.000.0%0.000.0%

SQA-iO Repeatability and Between-Day variation

Between-LaboratoryBetween-DayReproducibility (Total Precision)
Control LevelSD%CVSD%CVSD%CV
Level High L10.831.8%0.621.3%0.731.6%
Level Normal L20.803.2%0.441.8%0.622.5%
Level Low L30.000.0%0.000.0%0.000.0%

Analytical sensitivity (limits of blank and detection/quantitation)

The objective of this study was to define the limit of blank (LoB), Limit of Detection (LoD) and limit of Quantitation (LoQ) of the SQA-iO system for sperm concentration.

Following the quidance of CLSI EP17-A2, the study included a BLANK (seminal plasma) and a LOW-LEVEL sample of sperm concentration. The samples were assayed 12 times on each of five SQA-iO devices (60 tests per level).

The data demonstrated the following:

Limit of Blank (LoB) = 0 M/mL Limit of Detection (LoD) = 1.73 M/mL Limit of Quantitation (LoQ) = 5.96 M/mL

Linearity (measurement range)

The dynamic range and linearity of the SQA-iO instrument (concentration, M/mL) were established by using three SOA-iO devices and one SQA-V reference system. Semen samples were prepared at nine semen concentration intervals ranging from low to high level (less than 2 to 400 M/mL). The test was performed in three SQA-iO devices per concentration level, and results were compared to the average of

{9}------------------------------------------------

the triplicate results of the SQA-V analyzer.

It was demonstrated that the SQA-iO linear regression coefficients "R" exceeded 0.9.

Interference

The interference study was based on the CLSI EP07, 3rd ed. quidelines, i.e., each contaminant was added as 1/20 part of the sample resulting in a 5%/95% ratio. The spiked and control samples were tested by the SQA-iO at two sperm concentration levels defined as "low `` (15-60 M/mL) and "high" (100–200 M/mL), and data were reported for concentration, motility, MSC, PMSC, and morphology. The data demonstrated no significant differences between spiked and controls at both levels and across all reported parameters.

Sample Stability

This study was performed to demonstrate the stability of semen motility parameters. It is a known fact that sperm motility decreases over time, and therefore MES proposes, in its SQA-iO labeling, to follow WHO 4th, 5th and 6th quidelines that semen should be tested within 1 hour of collection. In this study, the motility parameters of % motility, MSC and PMSC (the presumptive "worst-case" parameters for sample stability) were assessed at various time intervals, and results were compared to their initial (T0) results following the scheme below:

  • T0 within 1 hour of collection but after sample liquefaction
  • T1 2 hours post collection
  • T2 3 hours post collection
  • T3 4 hours post collection

The study results indicate that motility parameters are reasonably stable at 2 hours post collection, but individual samples can vary widely. Additionally, the study results support the WHO recommendations to perform semen analysis within 1 hour of collection.

Electrical Safety Tests

Electrical safety tests and EMC tests on the SQA-iO Sperm Quality Analyzer were performed according to the following standards:

  • IEC 61010-1 Safety requirements for electrical equipment for measurement, control, and . laboratory use Part 1: General requirements
  • IEC 61010-2-101 Safety requirements for electrical equipment for measurement, control and . laboratory use Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
  • . EMC: IEC 60601-1-2 General requirements for basic safety and essential performance related with electromagnetic compatibility of Medical Devices

It was demonstrated that the SQA-iO device has successfully passed all the tests specified by the above standards, according to all applicable standard requirements. There were no deviations or modifications made to the tested version of the SQA-iO device in order to pass any of the electrical safety and EMC tests.

Environmental Tests

The environmental test evaluating the transportation and environmental stress impact on the SQA-iO was performed according to ASTM 4332-14 and ASTM D4169-16. The tests included evaluating the SQA-iO in extreme climatic and transportation conditions, including desert conditions (high temperatures with low humidity), extreme cold (low temperatures), tropical conditions (high temperatures and high humidity), free fall (drop), bump, random vibrations, and low air pressure conditions expected in its application environment and during transportation.

{10}------------------------------------------------

It was demonstrated that the SQA-iO withstands the transportation and environmental conditions in compliance with the standard requirements as no deteriorational performance of the tested SQA-iO was observed following the environmental challenges.

807.92 (b)(2): Brief Description of Clinical Data

Method Comparison Study

Method comparison data were based on comparative testing of 165 matched native semen samples that were assayed according to WHO 6th criteria by the SQA-iO (on-test) and the SOA-V (reference) methods across three sites. At each site, one SQA-iO device and one SQA-V device were used.

The SQA-iO operators (n = 12 across all sites) were laboratory technicians or professionals who are familiar with semen analysis. In expert operator assayed the same sample in dublicate on the SQA-V predicate, and the means of the SQA-V results were used as the comparator results.

The data were analyzed by Passing-Bablok regression (SQA-V on x-axis), and the y-intercepts, slopes, and correlation coefficients, along with the 95% confidence intervals, are shown in the table, below.

ParameterInterceptCISlopeCICorrelationCI
CONCENTRATION, M/mL-1.5-2.0 to -0.71.01.0 to 1.01.00.98 to 0.99
MOTILITY, %-3.0-3.1 to -1.71.01.0 to 1.01.00.95 to 0.97
PROGRESSIVE MOTILITY, %-0.8-1.0 to 0.00.90.9 to 1.01.00.97 to 0.98
RAPIDLY PROGRESSIVE, %0.10.0 to 0.31.00.9 to 1.00.90.90 to 0.94
SLOWLY PROGRESSIVE, %-0.8-1.0 to 0.01.00.9 to 1.00.90.86 to 0.93
NON-PROGRESSIVE, %-1.9-3.0 to -1.01.21.0 to 1.30.80.71 to 0.83
IMMOTILE, %3.01.0 to 5.01.01.0 to 1.01.00.95 to 0.97
MSC, M/mL-0.9-1.7 to -0.61.01.0 to 1.01.00.98 to 0.99
PMSC, M/mL-0.4-0.7 to -0.31.00.9 to 1.01.00.99 to 1.00
RAPID PMSC, M/mL0.0-0.1 to 0.01.01.0 to 1.01.00.96 to 0.98
SLOW PMSC, M/mL-0.1-0.4 to -0.11.00.9 to 1.01.00.98 to 0.99
MORPHOLOGY, % (n = 155)0.00.0 to 0.11.00.9 to 1.01.00.96 to 0.98
FSC, M/mL (n = 155)-0.1-0.1 to 0.00.90.9 to 1.01.00.97 to 0.99

SQA-iO vs. SQA-V (n = 165)

The data demonstrate slopes between 0.9 and 1.2, Y-intercepts near zero, and a correlation coefficients. ("r") ≥0.8.

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

The SQA-iO testing confirms that the device can be used according to its intended use and in an equivalent manner to the predicate device. The information and data provided in this 510(k) submission identifies no new safety or effectiveness issues for this device type. The SQA-iO is safe and performs effectively based on its intended use.

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”