K Number
K240871
Date Cleared
2024-11-18

(234 days)

Product Code
Regulation Number
870.1330
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synxess Neurovascular Guidewire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guidewire is intended for use only in the neuro vasculature.

Device Description

The Synxess Neurovascular Guidewire is available in diameters of 0.014 inch (0.35 mm) and 0.010 inch (0.25 mm) and has 200 cm length. The device is comprised of a core wire, coils, and coatings, and is supplied with a shaping mandrel and a torque device. The guidewire distal tip is straight and shapeable. The device is provided sterile and is for single use.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the Synxess Neurovascular Guidewire, and details the non-clinical testing performed to establish substantial equivalence to predicate devices. It does not describe an AI/ML-driven device or a study involving human readers and AI assistance. Therefore, many of the requested points in the prompt (e.g., AI performance metrics, MRMC study, expert ground truth establishment for AI, training set details) are not applicable to this document.

However, I can provide information based on the provided text regarding the device's acceptance criteria and the study that proves it meets them, focusing on the non-clinical performance testing conducted for a guidewire.

Here's the breakdown based on the provided document:

Acceptance Criteria and Device Performance for Synxess Neurovascular Guidewire

The acceptance criteria for this non-AI device are based on passing various performance bench tests and biocompatibility tests in accordance with relevant international standards and FDA guidance. The study proving the device meets these criteria is the battery of non-clinical tests performed.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for each test are implicitly "Pass" or "Met acceptance criteria" as defined by the referenced standards and guidance documents. The reported device performance for all listed tests was successful.

TestStandards, Guidance, Test MethodReported Device Performance
Visual InspectionFDA Guidewire Guidance and ISO 11070Pass
Dimensional VerificationFDA Guidewire Guidance and ISO 11070Pass
Simulated UseFDA Guidewire GuidancePass
Tensile Strength and Tip PullFDA Guidewire Guidance and ISO 11070Pass
Torque StrengthFDA Guidewire GuidancePass
TorqueabilityFDA Guidewire GuidancePass
Coating IntegrityFDA Guidewire GuidancePass
Particulate EvaluationFDA Guidewire GuidancePass
LubricityFDA Guidewire GuidancePass
Corrosion ResistanceFDA Guidewire Guidance and ISO 11070Pass
Kink ResistanceFDA Guidewire GuidancePass
Tip FlexibilityFDA Guidewire GuidancePass
RadiopacityFDA Guidewire Guidance and ISO 11070Pass
Fracture TestISO 11070Pass
Flexing TestISO 11070Pass
Shaping Mandrel Visual and Dimensional InspectionThe shaping mandrel was examined for signs of impurities or defects. The mandrel length and outer diameter were measured.Met acceptance criteria
Shaping Mandrel Corrosion ResistanceISO 11070Pass
Shaping Mandrel Tensile StrengthThe tensile strength between the shaping mandrel rod and handle was measured.Met acceptance criteria
Torque Device Visual Inspection and Simulated UseThe torque device was examined for signs of impurities or defects. Simulated use evaluated the torque device compatibility and performance.Met acceptance criteria
Biocompatibility Tests
Cytotoxicity Study Using MTT MethodISO 10993-5Pass
Sensitization Study Guinea Pig Maximization TestISO 10993-10Pass
Intracutaneous Reactivity Test In RabbitsISO 10993-23Pass
Acute Systemic Toxicity Study In MiceISO 10993-11Pass
Pyrogen Test In RabbitsISO 10993-11 / USPPass
Hemolysis Assay-Direct Contact And Extract MethodsISO 10993-4 / ASTM F756-17Pass
Complement Activation SC5b-9 AssayISO 10993-4Pass
Thromboresistance Evaluation In DogsISO 10993-4Pass
Sterilization Validation
EO Sterilization Process ValidationISO 11135:2014Validated
Sterilant ResidualsISO 10993-7:2008Met acceptance criteria
Shelf-Life Validation
Shelf-Life (3 years)Testing of accelerated aged devicesValidated

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample sizes (e.g., number of guidewires) used for each individual performance bench test or biocompatibility test. It only states that the device "passed all performance bench testing." For biocompatibility, some tests specify animal models (e.g., guinea pig, rabbits, mice, dogs), but the specific number of animals is not provided.
  • Data Provenance: The document generally indicates that the tests were performed in accordance with "internal requirements, referenced guidance and international standards." The country of origin for the data is implied to be from the manufacturer, Enlight Medical Technologies (Shenzhen) Co., Ltd., located in China. The data would be considered prospective as it was generated specifically for this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • This question is not applicable as this is a physical medical device (guidewire), not an AI/ML-driven diagnostic device requiring expert clinical interpretation for ground truth. The "ground truth" for these tests is defined by the physical and chemical properties measured against established engineering and biological standards.

4. Adjudication Method for the Test Set

  • This question is not applicable for the reasons stated above. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where human interpretation of medical images or patient data requires consensus, which is not relevant here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No. An MRMC study was not performed because this is a physical medical device, not an AI/ML-driven system that assists human readers in diagnostic tasks. The document explicitly states: "No clinical testing was deemed necessary to support the substantial equivalence of the subject device."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an algorithm-only device.

7. The Type of Ground Truth Used

  • The "ground truth" for this device's performance is established by engineering standards (e.g., measurements of tensile strength, torqueability, flexibility, dimensions) and biological/chemical standards for biocompatibility (e.g., cytotoxicity assays, pyrogen tests, hemolysis assays). These are objective, quantifiable measures against predefined acceptance criteria from international standards (e.g., ISO 11070, ISO 10993 series) and FDA guidance.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. This is a physical device, not an AI/ML model.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.