K Number
K110584
Date Cleared
2011-08-22

(174 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ASAHI Neurovascular Guide Wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

Device Description

The ASAHI CHIKAI Neurovascular Guide Wire is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 200 cm and 300 cm lengths. The extension wire is connected to the end of the guide wire outside the body for 200 cm wire. The guide wire is constructed from a stainless steel core wire with platinum-nickel and stainless steel coils. The coil assembly consists of an inner coil and an outer coil, and the coil assembly is soldered to the core wire. The coil assembly construction is similar to the 510k cleared ASAHI SION PTCA guide wire with K100578. The distal end of the guide wire has a radiopaque tip to achieve visibility, and is available in a straight configuration and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire. The proximal section of the guide wire for 300 cm wire is coated with PTFE.

AI/ML Overview

The provided text describes a 510(k) summary for the ASAHI CHIKAI Neurovascular Guide Wire, outlining its technical characteristics, intended use, and performance data to demonstrate substantial equivalence to predicate devices. However, it does not contain information describing acceptance criteria or a study that establishes performance against those criteria in the way typically expected for an AI/ML device (e.g., sensitivity, specificity, or FROC analysis).

The document is for a medical device (a guide wire), not an AI/ML diagnostic or prognostic tool. Therefore, the questions regarding acceptance criteria in terms of accuracy metrics, sample sizes for test/training sets, expert consensus for ground truth, adjudication methods, or MRMC studies are not applicable in this context. The "performance data" referred to is related to the physical and biological properties of the guide wire itself.

Here's a breakdown of the available information based on the provided text, and where the requested AI/ML specific information is not present:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria or performance metrics in a table format that would typically be used for an AI/ML device. Instead, it lists various types of in vitro bench tests and biocompatibility evaluations that were conducted. The "acceptance criteria" here are implied to be successful completion of these tests, demonstrating the device performs as intended and is substantially equivalent to predicate devices.

Test CategorySpecific TestReported Device Performance
In vitro bench testingTensile StrengthMeets predetermined performance criteria; performs as intended.
Turns to Failure (Torque Strength)Meets predetermined performance criteria; performs as intended.
Torqueability (Torque Response)Meets predetermined performance criteria; performs as intended.
Tip FlexibilityMeets predetermined performance criteria; performs as intended.
Coating AdhesionMeets predetermined performance criteria; performs as intended.
Catheter compatibility (Slipping Ability)Meets predetermined performance criteria; performs as intended.
Particulate testingMeets predetermined performance criteria; performs as intended.
Shelf life testingMeets predetermined performance criteria; performs as intended.
Biocompatibility/evaluationSystemic Toxicity StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
In Vitro Hemolysis StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
Intracutaneous StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
Cytotoxicity StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
Sensitization StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
Pyrogen StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
Plasma Recalcification TimeEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
Coagulation StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
In Vivo Thromboresistance StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
C3a Complement Activation StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.
SC5b-9 Complement Activation StudyEstablished by successful use of same materials and manufacturing process in 510(k) approved Asahi Guide Wire products.

The document states: "Enclosed within this submission is performance data that demonstrates that the ASAHI CHIKAI Neurovascular Guide Wire meets all predetermined performance criteria." and "This 510(k) notice includes mechanical and functional bench testing that demonstrates that the ASAHI CHIKAI Neurovascular Guide Wire performs as intended."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The tests mentioned are laboratory bench tests and biocompatibility studies, not clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The "ground truth" for a physical medical device typically refers to engineering specifications and material science standards, not expert interpretations of medical images or diagnoses.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a concept used in diagnostic studies (often for AI/ML) where expert disagreement needs to be resolved. It does not apply to the physical performance testing of a guide wire.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical medical instrument, not an AI/ML diagnostic or assistance tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical instrument, not an AI/ML algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For this type of device, the "ground truth" is based on established engineering standards for mechanical properties and validated scientific methods for biocompatibility (e.g., ISO standards for tensile strength, cytotoxicity tests, etc.). No "expert consensus," "pathology," or "outcomes data" in the clinical sense is explicitly mentioned as ground truth for these performance tests.

8. The sample size for the training set

Not applicable. This device does not involve machine learning or a "training set."

9. How the ground truth for the training set was established

Not applicable. This device does not involve machine learning or a "training set."


Summary of Device and Study Type:

The provided document describes a 510(k) Premarket Notification for a physical medical device, a neurovascular guide wire. The "study" referenced is a series of in vitro bench tests and biocompatibility evaluations designed to demonstrate the device's physical performance, material safety, and substantial equivalence to existing predicate devices. It is not an AI/ML device, and therefore, many of the requested criteria related to AI/ML performance assessment are not present or applicable.

{0}------------------------------------------------

AUG 2 2 2011

KI10584

510(K) SUMMARY

510(K) SUMMARY

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

APPLICANTAsahi Intecc Co., Ltd.1703 Wakita-cho, Moriyama-kuNagoya, Aichi 463-0024Japan
OFFICIALCORRESPONDENTYoshi TeraiPresident, CEOAsahi Intecc USA, Inc.2500 Red Hill Avenue, Suite 210Santa Ana, CA 92705Tel: (949) 756-8252FAX (949) 756-8165e-mail: yoshi.terai@asahi-intecc.com
TRADE NAME:ASAHI CHIKAI Neurovascular Guide Wire
COMMON NAME:Guide Wire
CLASSIFICATIONNAME:Wire, Guide, Catheter
DEVICECLASSIFICATION:Class 2 per 21 CFR §870.1330
PRODUCT CODEDQX - Catheter Guide Wire
PREDICATE DEVICE:1. Micro Therapeutics, Inc. / EV3 - SilverSpeed Hydrophilic Guidewire -510(k) K9932572. Asahi - JoWire Neo's PTCA Guide Wire - 510(k) K0227623. BSC/SciMed Life Systems, Inc. - Transend EX Platinum Guidewire -510(k)K971254Additional referenced 510(k) cleared device:4. Asahi - ASAHI SION PTCA Guide Wire - 510(k) K100578

Description of the Device Subject to Premarket Notification:

The ASAHI CHIKAI Neurovascular Guide Wire is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 200 cm and 300 cm lengths. The extension wire is connected to the end of the guide wire outside the body for 200 cm wire. The guide wire is constructed from a stainless steel core wire with platinum-nickel and stainless steel coils. The coil assembly consists of an inner coil and an outer coil, and the coil assembly is soldered to the core wire. The coil assembly construction is similar to the 510k cleared ASAHI SION PTCA guide wire with K100578. The distal end of the guide wire has a radiopaque tip to achieve visibility, and is available in a straight configuration and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire. The proximal section of the guide wire for 300 cm wire is coated with PTFE.

{1}------------------------------------------------

K110584

510(K) SUMMARY

INDICATION FOR USE:

ASAHI Neurovascular Guide Wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

TECHNICAL CHARACTERISTICS:

Comparisons of the ASAHI CHIKAI Neurovascular Guide Wire and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices. The ASAHI CHIKAI Neurovascular Guide Wire is similar in design - device dimensional specifications, and intended use, manufacturing process, operating principle, shelf life and sterilization process are the same and materials that have been used in other predicate devices in that its core wire, tip coils and solders remain the same.

PERFORMANCE DATA:

Enclosed within this submission is performance data that demonstrates that the ASAHI CHIKAI Neurovascular Guide Wire meets all predetermined performance criteria.

All components that come in direct contact with the patient have a long history of use in medical devices and are proven to be biocompatible for use in the vasculature. Furthermore, this submission contains reference to predicate ASAHI devices that use the same materials as used in the subject device.

In vitro bench testing, including tensile strength, torque strength, torqueability, tip flexibility, coating adherence, catheter compatibility, particulate testing and shelf life testing as listed below were conducted on the ASAHI CHIKAI Neurovascular Guide Wire. This 510(k) notice includes mechanical and functional bench testing that demonstrates that the ASAHI CHIKAI Neurovascular Guide Wire performs as intended.

The biocompatibility has been established by the successful use of the same materials and manufacturing process in currently 510(k) approved Asahi Guide Wire products.

Performance test/evaluation summary: Device performance: Tensile Strength Turns to Failure (Torque Strength) Torqueability (Torque Response) Tip Flexibility Coating Adhesion Slipping Ability of Guide Wire with Microcatheter Additional bench testing Particulate testing

Biocompatibility/evaluation: Systemic Toxicity Study In Vitro Hemolysis Study

{2}------------------------------------------------

510(K) SUMMARY

Intracutaneous Study Cytotoxicity Study Sensitization Study Pyrogen Study Plasma Recalcification Time Coagulation Study In Vivo Thromboresistance Study C3a Complement Activation Study SC5b-9 Complement Activation Study

SUMMARY/CONCLUSION:

The ASAHI CHIKAI Neurovascular Guide Wire characteristics are substantially equivalent to the specified predicate device and other currently marketed devices for the same indication for use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and a wavy line representing the body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Asahi Intecc USA, Inc c/o Yoshi Terai, President CEO 2500 Red Hill Avenue, Suite 210 Santa Ana, CA 92705

AUG 2 2 2011

Re: K110584

Trade/Device Name: Asahi Chikai Neurovascular Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Neurovascular catheter guide wire Regulatory Class: Class II Product Code: MOF Dated: August 4, 2011 Received: August 5, 2011

Dear Mr. Terai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the

{4}------------------------------------------------

quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.lda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

Erich Fromm

Malvina B. Eydelman, M. Director

Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): Kl(10584

Device Name: ASAHI CHIKA! Neurovascular Guide Wire

Indications for Use:

ASAHI Neurovascular Guide Wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

Prescription Use_ X

(Parl 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

JOE HUTTER

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page

KI 10584 510(k) Number

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.