(147 days)
The Portrait VSM vital signs monitor is intended to monitor a single patient's vital signs at the site of care or during intra-hospital transport.
The noninvasive oscillometric blood pressure parameter is intended for measurement of systolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients.
The optional GE TruSignal pulse oximetry and accessories are indicated for the continuous nonitoring of functional Oxygen Saturation (SpO2) and pulse rate, including monitoring conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients.
The optional Masimo SET® pulse oximetry and accessories are indicated for the continuous nonitoring of functional Oxygen Saturation (SpO2) and pulse rate, during both no motion conditions, and for patients who are well or poorly perfused (low perfusion) for adult, pediatric and neonatal patients.
The optional Nellcor™ pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused.
The optional Welch Allyn® SureTemp® Plus electronic thermometer is intended to measure one of oral, axillary, and rectal temperature of adult and pediatric patients.
The optional Exergen TemporalScanner thermometer is intermittent measurement of human body temperature of patients of all ages.
The optional HeTaiDa electronic infrared non-touch thermometer is intermittent measurement of human body temperature of patients of all ages.
A wireless network connection is provided to transmit clinical data into various hospital information systems. An optional remote alarm cable connection is intended to complement visual and audible alarms and not replace the need for the presence of a caregiver.
The portable device is designed for use in hospital-type facilities. The Portrait VSM vital signs monitor can also be used in satellite areas or alternate care settings.
The Portrait VSM vital signs monitor is intended for use under the direct supervision of a licensed health care practitioner.
The Portrait VSM vital signs monitor is not intended for use during MRI.
"Portable" refers to the ability of the Portrait VSM vital signs monitor to be easily moved by the caregiver, such as on a roll stand.
The proposed Portrait™ VSM is a vital signs monitor which is developed based on primary predicate vital signs monitor VC150(K133810) with integrated NIBP and SpO2 design from a secondary predicate monitor B105M (K213490) and provided additional non-contact infrared body temperature measurement option by supporting OEM thermometer (K171888) previously cleared by FDA.
In addition to the added non-contact infrared body thermometer, the proposed monitor Portrait™ VSM also offer several enhancements:
New hardware platform
Adopted equivalent NIBP design from B105M(K213490)
Adopted equivalent SpO2 design from B105M(K213490)
Compatible with Recorder B1X5-REC
Support Round Advisor in spot check mode
Support automatically screens brightness adjustment.
Improved Early Warning Score
Addtional alarm management enhancement.
Additional cybersecurity enhancement
The proposed monitor Portrait™ VSM adopts larger 10-inch LCD touch screen with improved Industrial Design (ID) to be more portable and more compact for clinicians than the primary predicate monitor VC150 (K133810) while maintaining the same primary function and operation.
As with the predicate Monitor VC150 (K133810), the proposed Portrait™ VSM is Vital Signs Monitor, utilizing an LCD display and can measure the most commonly used vital signs of patient: Non-invasive Blood Pressure (NIBP), Pulse Rate (PR), Temperature (Temp), and Pulse Oxygen Saturation (SpO2).
Same as the predicate monitor VC150 (K133810), the proposed monitor Portrait™ VSM also has three choices for SpO2 include GE TruSignal™; Nellcor™ or Masimo SET® and temperature measurement from Exergen® TemporalScanner™, and Welch Allyn® SureTemp®.
Both the predicate monitor VC150 (K133810) and the proposed monitor Portrait™ VSM can be configured to be used for Spotchecking or for continuous morning, the device can send measured patients' data to EMR (Electronic Medical Record) system by interfacing to Hospital Information Systems (HIS) over a wired or wireless network.
Moreover, same as the predicate monitor VC150 (K133810), the proposed monitor Portrait™ VSM can be powered by battery or AC, has a carrying handle and can be placed on a shelf or table or mounted in a variety of ways using a mounting plate located on the bottom of the monitor.
The provided document is a 510(k) Summary for the GE Medical Systems Information Technologies, Inc. Portrait VSM vital signs monitor. It describes the device, its intended use, and a comparison to predicate devices, along with summaries of non-clinical and clinical testing.
However, the document explicitly states that "the proposed monitor Portrait™ VSM did not require clinical studies to support substantial equivalence." This means there is no detailed clinical study described in this document that proves the device meets specific acceptance criteria based on human-in-the-loop performance or expert-adjudicated ground truth, as would be expected for a complex AI/ML-driven device.
The document primarily focuses on demonstrating substantial equivalence to existing predicate devices through bench testing, compliance with consensus standards, and verification of hardware/software functionality, electrical safety, usability, and environmental performance.
Therefore, I cannot fulfill all parts of your request as the provided text does not contain a study with the specific elements you've asked for related to clinical performance verification with AI/ML and human readers.
However, I can extract the acceptance criteria (in terms of standards compliance and functional equivalence) and the reported performance from the non-clinical testing described.
Here's what can be extracted and what cannot be provided given the document's content:
Acceptance Criteria and Device Performance (Based on Non-Clinical Testing and Equivalence Claim)
Since no clinical study was required, the "acceptance criteria" for this device's submission are primarily based on demonstrating:
- Functional Equivalence to legally marketed predicate devices.
- Compliance with relevant electrical safety, EMC, usability, and performance standards for vital signs monitors.
- Verification of hardware and software specifications through bench testing.
Therefore, the table below will reflect the claimed equivalence and standards compliance as the "performance" rather than specific accuracy metrics against a clinical ground truth from a reader study.
| Acceptance Criterion (Based on Equivalence/Standards) | Reported Device Performance (from Non-Clinical Testing) |
|---|---|
| Functional Equivalence to Predicate VC150 (K133810) and B105M (K213490) | The Portrait™ VSM is described as having "essentially same" features and parameters as the VC150. It incorporates NIBP and SpO2 design from the B105M, and supports an OEM thermometer previously cleared (K171888). Minor differences (e.g., larger screen, capacitive touch, wired network, specific Masimo/Nellcor OEM boards) are highlighted as either equivalent or improvements that do not affect safety/effectiveness. |
| Non-Invasive Blood Pressure (NIBP) Performance | Compliance with IEC 80601-2-30:2018 ("Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers"). The NIBP design is identical to predicate B105M (K213490) and uses the SuperSTAT algorithm (K022834). |
| Pulse Oximetry (SpO2) Performance | Compliance with ISO 80601-2-61:2017+C1:2018 ("Particular requirements for basic safety and essential performance of pulse oximeter equipment"). GE TruSignal SpO2 design is identical to predicate B105M (K213490). Different Masimo (MS-2011SB, K053269) and Nellcor (NELL1-SR OxiMax, K060576) OEM boards are used, which are noted to have been implemented in predicate B105M (K213490). |
| Temperature Measurement Performance | Compliance with ISO 80601-2-56:2017+AMD1:2018 ("Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement") and ASTM E1112-00 (2018) ("Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature"). Supports Welch Allyn SureTemp, Exergen TemporalScanner, and the added HeTaiDa Non-Contact Infrared Body Thermometer (K203332). |
| General Safety (Electrical, EMC) | Compliance with IEC 60601-1:2005+A1:2012+A2:2020 (Electrical safety) and IEC 60601-1-2:2014+A1:2020 (EMC). Also passed IEC TR 60601-4-2:2016. |
| Alarm System Performance | Compliance with IEC 60601-1-8:2006+A1:2012+A2:2020 (General requirements for alarm systems). Alarm volume ranges (e.g., Min >= 45dBA, Max <= 85dBA) meet this standard. |
| Usability/Human Factors | Compliance with IEC 60601-1-6:2010+A1:2013+A2:2020 and IEC 62366-1:2015+A1:2020. Summative Usability testing was concluded with 15 US Clinical and 15 US Technical users. |
| Software Validation & Cybersecurity | Follows FDA software guidance on "Content of Premarket Submissions for Device Software Functions" (categorized as "Enhanced" documentation level), "General Principles of Software Validation," "Off-The-Shelf Software Use in Medical Devices," and relevant cybersecurity guidances. Software standards IEC 62304:2006+A1:2015 and ISO 14971:2019 were applied. Security Risk Assessment, Threat model, and Penetration testing were conducted. |
| Environmental Performance (Mechanical, Thermal, Humidity) | Verified to meet specifications for operating and storage conditions (e.g., Temp: 5-40°C operating, -20-60°C storage; Humidity: 15-90% non-condensing operating, 10-90% storage). |
| Reprocessing Efficacy | Follows FDA guidance "Reprocessing Medical Devices in Health Care Settings Validation Methods and Labeling." Efficacy validation met acceptance criteria. |
| Wireless Coexistence | EMC testing per IEC 60601-1-2:2020, adherence to FDA Guidance-Radio Frequency Wireless Technology in Medical Devices, and IEEE ANSI C63.27-2021. |
Regarding the specific questions about clinical studies, AI/ML, and human reading components:
- 2. Sample sized used for the test set and the data provenance: Not applicable as no clinical study for device performance was conducted/required according to the document. The document describes non-clinical bench testing and usability testing (15 US clinical, 15 US technical users) which are not designed to establish "performance" against a clinical ground truth for vital signs measurements.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth established for performance validation from a clinical study.
- 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No clinical study with adjudication.
- 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No. This device is a vital signs monitor, not an AI-driven image analysis or diagnostic tool that assists human readers in interpreting complex medical data. The "AI" concept as you've framed it (human readers improving with AI assistance) does not apply to this device's function or the described testing.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device's core functions (NIBP, SpO2, Temp) are algorithm-driven, and their performance is proven through compliance with relevant IEC/ISO standards and comparison to predicate devices, as detailed in the non-clinical bench testing section. This validates the "standalone" performance of these measurement algorithms against established testing methodologies for vital signs monitors.
- 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the NIBP, SpO2, and Temperature measurements, the "ground truth" for the non-clinical performance validation is derived from established testing protocols within the cited international standards (e.g., IEC 80601-2-30 for NIBP, ISO 80601-2-61 for SpO2, ISO 80601-2-56 for temperature). These standards define specific test setups, reference equipment, and acceptable deviation ranges for accuracy. It's not "expert consensus" or "pathology" in the diagnostic AI sense, but rather engineering-validated performance against known inputs and reference measurements.
- 8. The sample size for the training set: Not applicable. This document is for a vital signs monitor, not a new AI/ML algorithm that requires a "training set" in the context of deep learning or similar AI development.
- 9. How the ground truth for the training set was established: Not applicable. (See #8)
In summary, the provided document details a substantial equivalence submission for a vital signs monitor, where the acceptance criteria and proof are based on equivalence to existing devices and compliance with established performance and safety standards through extensive non-clinical bench testing and verification, rather than a clinical study involving AI/ML performance or human reader improvements.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 25, 2024
GE Medical Systems Information Technologies, Inc. Jung William Regulatory Affairs Director 9900 Innovation Drive Wauwatosa, Wisconsin 53226
Re: K233810
Trade/Device Name: Portrait VSM Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DXN, DQA, DPZ, FLL, MHX Dated: November 30, 2023 Received: November 30, 2023
Dear Jung William:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology Diagnostics and Monitoring Devices Office of Cardiovascular Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K233810
Device Name
Portrait VSM
Indications for Use (Describe)
The Portrait VSM vital signs monitor is intended to monitor a single patient's vital signs at the site of care or during intra-hospital transport.
The noninvasive oscillometric blood pressure parameter is intended for measurement of systolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients.
The optional GE TruSignal pulse oximetry and accessories are indicated for the continuous nonitoring of functional Oxygen Saturation (SpO2) and pulse rate, including monitoring conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients.
The optional Masimo SET® pulse oximetry and accessories are indicated for the continuous nonitoring of functional Oxygen Saturation (SpO2) and pulse rate, during both no motion conditions, and for patients who are well or poorly perfused (low perfusion) for adult, pediatric and neonatal patients.
The optional Nellcor™ pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused.
The optional Welch Allyn® SureTemp® Plus electronic thermometer is intended to measure one of oral, axillary, and rectal temperature of adult and pediatric patients.
The optional Exergen TemporalScanner thermometer is intermittent measurement of human body temperature of patients of all ages.
The optional HeTaiDa electronic infrared non-touch thermometer is intermittent measurement of human body temperature of patients of all ages.
A wireless network connection is provided to transmit clinical data into various hospital information systems. An optional remote alarm cable connection is intended to complement visual and audible alarms and not replace the need for the presence of a caregiver.
The portable device is designed for use in hospital-type facilities. The Portrait VSM vital signs monitor can also be used in satellite areas or alternate care settings.
The Portrait VSM vital signs monitor is intended for use under the direct supervision of a licensed health care practitioner.
The Portrait VSM vital signs monitor is not intended for use during MRI.
"Portable" refers to the ability of the Portrait VSM vital signs monitor to be easily moved by the caregiver, such as on a roll stand.
Type of Use (Select one or both, as applicable)
J Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular frame. The letters are stylized with curved lines, and the frame has decorative flourishes resembling droplets or leaves around the circle.
510(k) Summary for K233810
In accordance with 21 CFR 807.92(c) the following summary of information is provided:
| Owner/Contact/Date (807.92(a)(1): | |
|---|---|
| Date: | November 30th, 2023 |
| Submitter: | GE Medical Systems Information Technologies, Inc. |
| 9900 Innovation Drive | |
| Wauwatosa, WI 53226 | |
| USA | |
| Primary Contact Person: | William JungRegulatory Affairs DirectorGE HealthCare |
| Phone: 571-396-1558 | |
| E-mail: William.Jung@gehealthcare.com | |
| Secondary Contact Person: | |
| Joel KentDirector, Regulatory Affairs StrategyGE HealthCarePhone: 617-851-0943E-mail: joel.kent@gehealthcare.com | |
| Sun YanliRegulatory Affairs ManagerGE HealthCare | |
| Phone: +86-18051597363E-mail: yanli.sun@gehealthcare.com | |
| Device names (807.92(a)(2):Trade Name: | Portrait™ VSM |
| Common/Usual Name: | Multiparameter patient monitor (Monitor, Physiological |
| Patient | |
| (Without Arrhythmia Detection or Alarms)) | |
K233810 510(k) Summary Page 1 of 17
{6}------------------------------------------------
| GE Medical SystemsInformationTechnologies, Inc.9900 Innovation DriveWauwatosa, WI 53226, USA | |
|---|---|
| Classification Names: | 21 C.F.R. §870.2300 Cardiac monitor (includingcardiotachometer and rate alarm).21 C.F.R. §870.1130 Noninvasive blood pressure measurementsystem.21 C.F.R. §870.2700 Oximeter21 C.F.R. §870.2710 Ear oximeter.21 C.F.R. §880.2910 Clinical electronic thermometer.21 C.F.R. §870.1100 alarm, blood-pressure |
| Product Code:Subsequent ProductCode:Predicate Device(s)(807.92(a)(3)): | MWIDXN, DQA, DPZ, FLLPrimary Predicate: K133810 Vital Signs Monitor VC150Additional Predicate Devices:K213490 B105M Patient MonitorK171888 Non-contact infrared body thermometer |
| Device Description(807.92(a)(4)): | The proposed Portrait™ VSM is a vital signs monitor which isdeveloped based on primary predicate vital signs monitorVC150(K133810) with integrated NIBP and SpO2 design from asecondary predicate monitor B105M (K213490) and providedadditional non-contact infrared body temperature measurementoption by supporting OEM thermometer (K171888) previouslycleared by FDA.In addition to the added non-contact infrared body thermometer,the proposed monitor Portrait™ VSM also offer severalenhancements:New hardware platform●Adopted equivalent NIBP design from B105M(K213490)●Adopted equivalent SpO2 design from B105M(K213490)●Compatible with Recorder B1X5-REC●Support Round Advisor in spot check mode●Support automatically screens brightness adjustment.●Improved Early Warning Score●Addtional alarm management enhancement.●Additional cybersecurity enhancement●K233810510(k) Summary Page |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a circle. The letters are stylized with curved lines, giving them a classic and elegant appearance. The circle surrounding the letters has decorative flourishes at the top and bottom, resembling stylized leaves or swirls. The logo is presented in black against a white background, creating a strong contrast.
The proposed monitor Portrait™ VSM adopts larger 10-inch
LCD touch screen with improved Industrial Design (ID) to be more portable and more compact for clinicians than the primary predicate monitor VC150 (K133810) while maintaining the same primary function and operation.
As with the predicate Monitor VC150 (K133810), the proposed Portrait™ VSM is Vital Signs Monitor, utilizing an LCD display and can measure the most commonly used vital signs of patient: Non-invasive Blood Pressure (NIBP), Pulse Rate (PR), Temperature (Temp), and Pulse Oxygen Saturation (SpO2).
Same as the predicate monitor VC150 (K133810), the proposed monitor Portrait™ VSM also has three choices for SpO2 include GE TruSignal™; Nellcor™ or Masimo SET® and temperature measurement from Exergen® TemporalScanner™, and Welch Allyn® SureTemp®.
Both the predicate monitor VC150 (K133810) and the proposed monitor Portrait™ VSM can be configured to be used for Spotchecking or for continuous morning, the device can send measured patients' data to EMR (Electronic Medical Record) system by interfacing to Hospital Information Systems (HIS) over a wired or wireless network.
Moreover, same as the predicate monitor VC150 (K133810), the proposed monitor Portrait™ VSM can be powered by battery or AC, has a carrying handle and can be placed on a shelf or table or mounted in a variety of ways using a mounting plate located on the bottom of the monitor.
Indications for Use (807.92(a)(5)):
The Portrait™ VSM vital signs monitor is intended to monitor a single patient's vital signs at the site of care or during intrahospital transport.
The noninvasive oscillometric blood pressure parameter is intended for measurement of systolic, diastolic, and mean arterial
K233810 510(k) Summary Page 3 of 17
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circle. There are decorative flourishes resembling water droplets or stylized leaves surrounding the circle, adding a touch of elegance to the design. The logo is presented in black and white, giving it a classic and timeless appearance.
blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients.
The optional GE TruSignal pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate, including monitoring during conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients.
The optional Masimo SET® pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate, during both no motion and motion conditions, and for patients who are well or poorly perfused (low perfusion) for adult, pediatric and neonatal patients.
The optional Nellcor™ pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused.
The optional Welch Allyn SureTemp Plus electronic thermometer is intended to measure one of oral, axillary, and rectal temperature of adult and pediatric patients.
The optional Exergen TemporalScanner thermometer is intended for the intermittent measurement of human body temperature of patients of all ages.
The optional HeTaiDa electronic infrared non-touch thermometer is intended for the intermittent measurement of human body temperature of patients of all ages.
A wireless network connection is provided to transmit clinical data into various hospital information systems. An optional remote alarm cable connection is intended to complement visual and audible alarms and not replace the need for the presence of a caregiver.
K233810 510(k) Summary Page 4 of 17
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized monogram, enclosed within a circular frame. The frame has decorative flourishes resembling water droplets or stylized leaves, adding a touch of elegance to the design. The logo is presented in black and white, with the monogram and frame in black against a white background.
The portable device is designed for use in hospitals and hospitaltype facilities. The Portrait™ VSM vital signs monitor can also be used in satellite areas or alternate care settings.
The Portrait™ VSM vital signs monitor is intended for use under the direct supervision of a licensed health care practitioner. The Portrait™ VSM vital signs monitor is not intended for use during MRI.
"Portable" refers to the ability of the Portrait™ VSM vital signs monitor to be easily moved by the caregiver, such as on a roll stand.
Contraindications for using the monitor
The Portrait™ VSM vital signs monitor is not intended for use during MRI.
Technology (807.92(a)(6)):
The proposed monitor Portrait™ VSM is a vital signs monitor where features and parameters are essentially same as in predicate monitor VC150 (K133810). This 510 (k) introduces non-contact infrared body thermometer by supporting additional optional OEM module (K171888) previously cleared by FDA and enhancements of several features.
The fundamental technology of the proposed monitor Portrait™ VSM is same as the predicate device.
The proposed monitor Portrait™ VSM are substantially equivalent to the predicate devices.
A summary of the main changes compared to the predicate are listed below in the comparison table.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a circular emblem with the letters "GE" intertwined in the center. The letters are stylized and have a classic, elegant design. The logo is black and white.
Product Comparison versus Predicate Main Features:
The proposed monitor Portrait™ VSM shares equivalent indications for use, intended use patient
populations, and functional features as the predicate devices. The proposed monitor Portrait™ VSM consists of improvements and optional features outlined below that are substantially equivalent to the predicate device.
| Specification | Primary Predicate Vital Sings MonitorVC150(K133810) | Proposed Portrait™ VSM Vital Signs Monitor | Differences |
|---|---|---|---|
| • Indicationsfor Use | The VC150 is intended to monitor a single patient's vital signs at the site of care or during intra-hospital transport. | The Portrait™ VSM is intended to monitor a single patient's vital signs at the site of care or during intra-hospital transport. | Equivalent. |
| The noninvasive oscillometric blood pressureparameter is intended for measurement of systolic, diastolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients. | The noninvasive oscillometric blood pressureparameter is intended for measurement of systolic, diastolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients. | Same features found in the predicate Monitor VC150 are retained, with the addition of infrared non-touch intermittent temperature measurement. | |
| The optional GE TruSignal pulse oximetry and accessories are indicated for continuous noninvasive monitoring of functional oxygen saturation (SpO2) and pulse rate, including monitoring during conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients. | The optional GE TruSignal pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate, including monitoring during conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients. | The addition of the infrared non-touch intermittent temperature measurement is supported through previously cleared HeTaida thermometer (K203332) | |
| The optional Masimo SET® pulse CO-oximetry and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate, during both no motion and motion conditions, and for patients who are well or poorly perfused (low perfusion) for adult, pediatric and neonatal patients. | The optional Masimo SET® pulse CO-oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate, during both no motion and motion conditions, and for patients who are well or poorly perfused (low perfusion) for adult, pediatric and neonatal patients. | The proposed device only displays the measurement result from The HeTaida thermometer. | |
| The optional Nellcor™ oximetry and accessories are indicated for the continuous, noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused. | The optional Nellcor™ pulse oximetry and accessories are indicated for the continuous noninvasive monitoring of functional Oxygen Saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused. | ||
| The optional Welch Allyn SureTemp Plus electronic thermometer is intended to measure oral, axillary, and rectal temperature of adult and pediatric patients.The optional Exergen TemporalScanner thermometer is intended for the intermittent measurement of human body temperature of patients of all ages. | The optional Welch Allyn SureTemp Plus electronic thermometer is intended to measure one of oral, axillary, and rectal temperature of adult and pediatric patients.The optional Exergen TemporalScanner thermometer is intended for the intermittent measurement of human body temperature of patients of all ages. | ||
| The optional HeTaiDa electronic infrared non-touch thermometer is intended for the intermittent measurement of human body temperature of patients of all ages. |
{11}------------------------------------------------
| A wireless network connection is provided totransmit clinical data into various hospitalinformation systems. An optional remote alarmcable connection is intended to complement visualand audible alarms and not replace the need forthe presence of a caregiver.The portable device is designed for use in hospitalsand hospital-type facilities. The VC150 can also beused in satellite areas or alternate care settings."Portable" refers to the ability of the VC150 to beeasily moved by the caregiver, such as on a rollstand. The VC150 is not intended to be used forcontinuous monitoring during patient transport. | A wireless network connection is provided totransmit clinical data into various hospitalinformation systems. An optional remote alarmcable connection is intended to complement visualand audible alarms and not replace the need forthe presence of a caregiver.The portable device is designed for use in hospitalsand hospital-type facilities. The Portrait™ VSM canalso be used in satellite areas or alternate caresettings.The Portrait™ VSM is intended for use under thedirect supervision of a licensed health carepractitioner."Portable" refers to the ability of the Portrait™VSM to be easily moved by the caregiver, such ason a roll stand. The Portrait™ VSM is not intendedfor use during MRI. | Add few clarification to clarify thedevice need be used under thedirect supervision of a licensedhealth care practitioner and can'tbe used during the MRI andremoved one sentence regardingcontinuous monitoring to keepconsistent with GE other monitors.Nothing change for the deviceitself. Both the predicate and theproposed device can use in thesame condition, the modificationof the description in indication foruse is only to keep consistentdescription with GE other familyproducts.Therefore, the changes do notsignificantly affect safety and/oreffectiveness. | |
|---|---|---|---|
| • PatientPopulation | Adult, pediatric and neonate | Adult, pediatric and neonate | Identical |
| • UseEnvironment | The Patient Monitor is intended for use in multipleareas within a professional healthcare facility | The Patient Monitor is intended for use in multipleareas within a professional healthcare facility | Identical |
| • Intrahospitaltransport withina professionalhealthcarefacility | Yes | Yes | Identical |
| • Intendedapplication site | NIBP- limbSpO2- finger, toe, earTemperature- Oral/Rectal/Axillary/Forehead | NIBP- limbSpO2- finger, toe, earTemperature- Oral/Rectal/Axillary/Forehead | Identical |
| GE Medical SystemsInformationTechnologies, Inc.9900 Innovation DriveWauwatosa, WI 53226, USA | |||
| • AvailableParameters | Pulse rateNon-invasive blood pressure (NIBP)Oxygen saturation (SpO2)TemperatureRespiration rate (only available with Masimotechnology) | Pulse rateNon-invasive blood pressure (NIBP)Oxygen saturation (SpO2)TemperatureRespiration rate (only support manually enteredvalues of Respiration Rate) | EquivalentThe predicate monitor couldsupport measure respiration rate bychoose Masimo technology.The proposed monitor doesn'tprovide the respiration ratemeasurement due to employdifferent Masimo technology.respiration rate measurement is notprovided in Masimo board MS-2011SB (K053269). However, theproposed monitor supports caregiver to enter the respiration ratevalue manually for patient.The change does not affect safetyor effectiveness. |
| • PeripheralInterfaces | Network interfaceMains inputNurse call Connector3 USB ConnectorSpO2 connectorNIBP hose connection | Network interfaceMains inputNurse call Connector3 USB ConnectorSpO2 connectorNIBP hose connectionHDMI ConnectorRecorder Connector | EquivalentThe proposed device adds HDMI(High-Definition MultimediaInterface) as connector to provide aslave video interface for asecondary display.The proposed device also supportsconnect to thermal recorder B1X5-REC(K213490) for printingthrough recorder connector. Therecorder B1x5-REC(K213490) isidentical and cleared in thepredicate monitor B105M(K213490).The function was verified on theproposed device, and itdemonstrated has equivalence tothe predicate device.The change does not affect safetyor effectiveness. |
| Display size | 8.4-inch LCD display | 10.1-inch LCD display | EquivalentThe proposed device supportslarger LCD display with better userexperience.The change does not affect safetyor effectiveness. |
| Display type | Resistive touch screen | Capacitive touch screen | EquivalentThe proposed device uses thecapacitive touch screen due totechnology improvement ascapacitive touch screen has lowpower, long service life and stableoperation compare with resistivetouch screen.The change does not affect safetyor effectiveness. |
| OperatingSystem | Linux | Linux | Identical |
| NetworkingInterface | WLAN | WLANLAN | EquivalentThe proposed device additionalsupport wired networkconnectivity.The proposed device adoptsequivalent LAN design as B105M(K213490)The change does not affect safetyor effectiveness. |
| NetworkingProtocol | HL7Cerner-iBusLDAPNTP | HL7Cerner-iBusLDAPNTP | Identical |
| NetworksSupportedDevices | CARESCAPE Gateway | EMR Gateway Pro V2 | EquivalentThe proposed device supportsEMR Gateway Pro V2 which hasbetter cost/product ratio.Both CARESCAPE Gatewayand EMR Gateway Pro V2 acts asa communication conduit betweenthe EMR system and the monitorand use the same HL7 interfaceengine.The change does not affect safetyor effectiveness. |
| AudibleIndicators | High, Medium, and Low alarm have audible tonesthat differentiate them from each other. | High, Medium, and Low alarm have audible tonesthat differentiate them from each other. | Identical |
| VisualIndicators | Alarm light. Highly visible Red/Yellow/Cyan lightLCD Display: Red/Yellow/Cyan alarm notification | Alarm light. Highly visible Red/Yellow/Cyan lightLCD Display: Red/Yellow/Cyan alarm notification. | Identical |
| Alarm Volume | Adjustable.- Minimum level >= 40dBA,- Maximum level<= 90dBA | Adjustable.- Minimum level >= 45dBA,- Maximum level<= 85dBA | EquivalentSlight difference for range due todifferent hardware. Both rangesmeet the requirements ofIEC60601-1-8 standard.The change does not affect safetyor effectiveness. |
| Alarm Volumeadjustment | High/medium/low three priorities share same alarmvolume setting. | Provide two options:Option 1: High/Medium/Low three priorities sharesame alarm volume setting.Option 2: Separate alarm volume setting forLow priority alarms | EquivalentThe proposed device provided theadditional option for Separatealarm volume setting for lowpriority alarms.Separate alarm volume setting forlow priority alarms alreadyimplemented in the predicatemonitor B105M (K213490) in theidentical way.The change does not affect safetyor effectiveness. |
| AllowedMinimumAlarm VolumeAdjustmentRange | The allowed minimum volume adjustment range is50%-100%Adjustment the minimum volume is passwordprotected | The allowed minimum volume adjustment range is1-10 (equivale to 10% -100%)Adjustment the minimum volume is passwordprotected | EquivalentThe proposed device changed theminimum volume level range to10% to 100% to align with GEother monitors.Minimum volume adjustmentrange to 10% to 100% has beenimplemented in the predicatemonitor B105M (K213490) withthe identical way.The change does not affect safetyor effectiveness. |
| Work mode | Spot check modeMonitoring mode | Spot check modeMonitoring mode | Identical |
| NIBPmeasurement | Support | Support | EquivalentThe proposed device incorporatedthe NIBP design from the predicatmonitor B105M (213490).It adopts identical NIBP algorithmSuperSTAT (K022834), identicalhardware design with thepredicate Monitor B105M(K213490) except the NIBP circuitboard has been re-arranged toadjust the layout of the samehardware to fit new industrydesign size.The change does not affect safetyor effectiveness. |
| Masimo SpO2measurement | Masimo SpO2 | Masimo MS-2011SB (K053269) | EquivalentSame as primary predicate VC150(K133810), the proposed devicealso supports Masimo SpO2 butadopts different Masimo OEMboard MS-2011SB (K053269)which has been implemented inthe predicate monitor B105M(K213490) in equivalent way.The change does not affect safetyor effectiveness. |
| Nellcor SpO2measurement | Nellcor SpO2 | Nellcor NELL1-SR OxiMax OEM(K060576) | EquivalentSame as primary predicateVC150(K133810), the proposeddevice also supports Nellcor SpO2but adopts different Nellcor OEMboard NELL1-SR (K060576) whichhas been implemented in thepredicate monitor B105M(K213490) in equivalent way.The change does not affect safetyor effectiveness. |
| GE SpO2measurement | GE TruSginal SpO2 | GE TruSginal SpO2 | EquivalentThe proposed device adoptsidentical GE TruSginal SpO2 designas the predicate monitor B105M(K213490) try to share the mostelctric boards design amog the GEmonitors.The change does not affect safetyor effectiveness. |
| TemperatureMeasurement | Welch Allyn SureTempExergen Temporal Scanner Thermometer | Welch Allyn SureTempExergen Temporal Scanner ThermometerHeTaiDa Non-Contact Infrared BodyThermometer | Equivalent |
| The proposed device additionalsupport HeTaiDa Non-ContactInfrared Body Thermometer(K203332) which has been 510(k)cleared. | |||
| The change does not affect safetyor effectiveness. | |||
| PrintersSupported | Integrated Thermal Recorder | B1X5-REC thermal recorder | Equivalent |
| The B1X5-REC recorder providesthe user the same thermalrecording functions as the thermalrecorders that were used in thepredicate VC150 Vital SignsMonitor (K133810). | |||
| The B1X5-REC (K213490)already cleared in K201941 andused in the predicate MonitorB105M (K213490). | |||
| The change does not affect safetyand effectiveness. | |||
| Host Operatingtemperature | 5° to 40°C (41° to 104°F) | 5° to 40°C (41° to 104°F) | Identical |
| Operatinghumidity range | 5 to 95% non-condensing | 15 to 90% non-condensing | EquivalentSlightly different in Operatinghumidity range due to hardwaredifference.The change does not affect safetyand effectiveness. |
| Operatingatmosphericpressure | 700 to 1060 hPa (525 to 795 mmHg) | 700 to 1060 hPa (525 to 795 mmHg) | Identical |
| Storagetemperature | -20 to 50 °C (-4 to 122 °F) (Host) | -20 to 60 °C (-4 to 140 °F) (Host) | Equivalent |
| Storagehumidity | 5 to 95% non-condensing (Host) | 10 to 90% non-condensing (Host) | Slightly different in Storageenvironment range due to hardwaredifference. |
| Storageatmosphericpressure | 500 to 1060 hPa (525 to 795 mmHg) | 700 to 1060 hPa (525 to 795 mmHg) | The change does not affect safetyand effectiveness. |
| Power supply | From AC mains power adapte, internal battery | From AC, internal battery | Identical |
| PowerRequirements | 100 - 240 V ±10%, 50 /60 Hz | 100 - 240 V ±10%, 50 /60 Hz | Identical |
| PowerConsumption | <=48 VA | <=150 VA | Identical |
| Protection | Class I | Class I | Identical |
| Battery | Lithium Ion | Lithium Ion | Equivalent |
| The proposed monitor used adifferent battery compared withpredicate monitor VC150.However, the battery used byproposed monitor is identical withpredicate device B105M(K213490)The change does not affect safetyand effectiveness. | |||
| Height | 247±5mm | 275±5mm | Equivalent |
| Width | 242±5mm(Without Welch Allyn temperature option) | 265±5mm(Without Welch Allyn temperature option) | Physical specifications aredifferent due to different industrydesign.The change does not affect safetyand effectiveness. |
| Depth | 136±5mm | 175±5mm | |
| Weight | 2.8 kg (6.2 lb) | 3.8 kg (8.4 lb) |
{12}------------------------------------------------
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circle. There are decorative swirls emanating from the circle at four points around the perimeter. The logo is black and white.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo, which is a circular emblem with the letters 'GE' intertwined in a stylized script. The letters are surrounded by a decorative swirl pattern within the circle. The logo is presented in black and white, with the letters and swirl pattern in black against a white background.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. The circle is surrounded by decorative, swirling elements, giving the logo a classic and recognizable appearance. The logo is black and white.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. There are three decorative swirls or flourishes evenly spaced around the outside of the circle. The logo is black and white.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle has decorative flourishes or swirls extending outward from its perimeter at several points. The logo is presented in black against a white background.
Determination of Substantial Equivalence (807.92(b)(1):
Summary of Non-Clinical Tests:
Bench testing related to software, hardware and performance including applicable consensus standards was conducted on the Portrait™ VSM Vital Signs Monitor to demonstrate that the design meets the specifications.
This section addresses the non-clinical testing for Portrait™ VSM Vital Signs Monitor relied on for a determination of substantial equivalence to the primary predicate monitor K133810 VC150 and the secondary predicate monitor K213490 B105M.
Per the FDA guidance titled "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions; Guidance for Industry and Food and Drug Administration Staff, Document issued on December 20, 2019", the following was verified:
- Hardware Bench Testing ●
- . Alarms Bench Testing
- IEC 80601-2-30: 2018
- ISO 80601-2-56: 2017+AMD1:2018 ●
- ISO 80601-2-61: 2017+ C1:2018 ●
K233810 510(k) Summary Page 13 of 17
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a circle. The letters are stylized with curved lines, giving them a classic and elegant appearance. The circle surrounding the letters is also decorated with curved, leaf-like shapes at the top and bottom, adding to the logo's distinctive design. The logo is presented in black and white, with the letters and decorative elements in black against a white background.
The Portrait™ VSM Vital Signs Monitor meets the EMC requirements described in IEC 60601-1-2:2014+A1:2020 "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests". Compliance according to the "Electromagnetic Compatibility (EMC) of Medical Devices Guidance for Industry and Food and Drug Administration Staff, issued on June 6, 2022." The Portrait™ VSM Vital Signs Monitor also passed the testing for IEC TR 60601-4-2:2016 "Medical electrical equipment -Part 4-2: Guidance and interpretation -Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems. The Portrait™ VSM Vital Signs Monitor has been evaluated for electromagnetic compatibility and potential risks from common emitters in the Portrait™ VSM Vital Signs Monitor environment.
The Portrait™ VSM Vital Signs Monitor meet the electrical safety requirements of IEC 60601-1:2005+A1:2012+A2:2020 " Medical electrical equipment - Part 1: General requirements for basic safety and essential performance". This testing was performed by a recognized independent and Certified Body Testing Laboratory (CBTL) under the IECEE CB Scheme.
Additional data is provided for compliance to:
- . IEC 60601-1-8:2006+A1:2012+A2:2020: Medical electrical equipment - part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- . IEC 80601-2-30:2018: Medical electrical equipment Part 2-30: Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 80601-2-49:2018: Medical electrical equipment Part ● 2-49: Medical electrical equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
- ISO 80601-2-56:2017+A1:2018: Medical electrical ● equipment - Part 2-56: Particular requirements for basic
K233810 510(k) Summary Page 14 of 17
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border that features decorative, teardrop-shaped elements evenly spaced around the perimeter. The logo is presented in black and white, with the letters and border in black against a white background.
safety and essential performance of clinical thermometers for body temperature measurement
- ISO 80601-2-61:2017+C1:2018: Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ASTM E1112-00 (2018) Standard Specification for ● Electronic Thermometer for Intermittent Determination of Patient Temperature
Environmental (Mechanical, and Thermal Safety) testing, based on the Portrait™ VSM Vital Signs Monitor proposed uses and locations, was confirmed to meet the specifications listed in the requirements. The Portrait™ VSM Vital Signs Monitor specifications verification evidence is included for the following:
- Operating temperature
- Operating humidity ●
- Operating pressure ●
- Storage and transport temperature ●
- Storage and transport humidity
- Storage and transport pressure ●
- Mechanical stress
- Fluid ingress
- Packaging Bench Testing
The Portrait™ VSM Vital Signs Monitor follow the guidance "Reprocessing Medical Devices in Health Care Settings Validation Methods and Labeling, Document issued on March 17, 2015". Reprocessing efficacy validation has been conducted in accordance with the documented reprocessing instructions using worst-case devices/components of the Portrait™ VSM Vital Signs Monitor. The reprocessing efficacy validation met the acceptance criteria for the reprocessing efficacy validation tests.
The Portrait™ VSM Vital Signs Monitor follows the Applying Human Factors and Usability Engineering to Medical Devices; Guidance for Industry and Food and Drug Administration Staff, Document issued on: February 3, 2016 and the following standards:
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border that features decorative, teardrop-shaped elements evenly spaced around the perimeter. The logo is presented in black and white, with the letters and border in black against a white background.
- IEC 60601-1-6:2010+A1:2013+A2:2020: Medical ● electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62366-1:2015+A1:2020: Medical devices Part 1: ● Application of usability engineering to medical devices
Summative Usability testing has been concluded with 15 US Clinical, 15 US Technical users. The usability testing of the Portrait™ VSM Vital Signs Monitor follows the FDA Guidance for Industry and Food and Drug Administration Staff "Applying Human Factors and Usability Engineering to Medical Devices, Document issued on: February 3, 2016".
Wireless performance data was provided related to:
- . EMC testing per IEC 60601-1-2:2020, including electromagnetic immunity in the exclusion band and proximity fields from RF wireless communication.
- FDA Guidance-Radio Frequency Wireless Technology in Medical Devices, Document issued on: August 14, 2013
- IEEE ANSI C63.27-2021 American National Standard for ● Evaluation of Wireless Coexistence.
The Portrait™ VSM Vital Signs Monitor follow the FDA software guidance documents as outlined in this submission.
- Content of Premarket Submissions for Device Software . Functions. Guidance for Industry and Food and Drug Administration Staff. Document issued on June 14, 2023
- General Principles of Software Validation, Guidance for Industry and FDA Staff, January 11, 2002
- . Off-The-Shelf Software Use in Medical Devices, Document issued on August 11, 2023
- . Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software, Document issued on January 14, 2005
- Cybersecurity in Medical Devices: Ouality System Considerations and Content of Premarket Submissions. Guidance for Industry and Food and Drug Administration Staff, Document issued on September 27,2023
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a circle. There are decorative flourishes resembling droplets or stylized leaves surrounding the letters within the circle.
- Design Considerations and Pre-market Submission ● Recommendations for Interoperable Medical Devices. Document issued on September 6, 2017
Software testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions." Bases on the risk level of the Portrait™ VSM Vital Signs Monitor, the document level for this submission is considered as "Enhanced" documentation level. Software standards IEC 62304:2006+A1:2015 Medical device software -Software life cycle processes and risk management standard ISO 14971:2019 Medical devices - Application of risk management to medical devices were also applied to the design.
Patient safety, security, and privacy risks have been addressed in the design and development of Portrait™ VSM Vital Signs Monitor including a Security Risk Assessment, Threat model and Penetration testing. This includes system integrity controls, access controls, audit controls, network controls which address the General Principles and Security Capabilities outlined in the "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions. Draft Guidance for Industry and Food and Drug Administration Staff. Document issued on April 8, 2022."
Summary of Clinical Tests: Clinical (807.92(b)(2)):
The subject of this premarket submission, the proposed monitor Portrait™ VSM did not require clinical studies to support substantial equivalence.
GE HealthCare considers the proposed monitor Portrait™ VSM Conclusion (807.92(b)(3)): to be substantially equivalent to the predicate devices.
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).