K Number
K222744
Date Cleared
2022-10-11

(29 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sol-M Luer Lock Syringe is intended for aspiration of medications and fluids into the body.

Device Description

The Sol-M Luer Lock Syringe is used to aspirate and inject fluid/medication into the body. The syringe is a sterile, single-use, 3-part syringe with luer lock conical connection. The Sol-MTM Luer Lock Syringe consists of the following components: Barrel, Plunger, Gasket. In addition to being used manually, the Sol-MTM Luer Lock Syringe can also work with a powerdriven syringe pump, with the luer lock tip connected to the IV administration line to deliver specific volume of fluid/medication. The syringe is individually blister packaged, and EtO sterilized with SAL of 10-6. The Sol-M™ Luer Lock Syringe is provided in 1, 3, 5, 10, 20, 30, and 60ml volume.

AI/ML Overview

The provided text describes the Sol-M Luer Lock Syringe (K222744) and its acceptance criteria, particularly focusing on its suitability for use with power-driven syringe pumps.

Here's an analysis of the provided information, framed by your requested criteria:

The study primarily focuses on demonstrating the suitability of the Sol-M Luer Lock Syringe for use with power-driven syringe pumps, referring to ISO 7886-2 as the key standard.

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that the device was tested according to ISO 7886-2:2020. The acceptance criteria are implicitly those defined by this standard. The reported performance is that the device met these criteria.

Acceptance Criteria Category (based on ISO 7886-2)Reported Device Performance
Critical Dimensions (for fit in pump)Met acceptance criteria
Short-term Flow Rate Error (with pump)Met acceptance criteria
Pump Forces (required to move plunger)Measured and made available
Syringe Compliance (fluid displaced)Met acceptance criteria
Manual Use (ISO 7886-1)Previously demonstrated
Material/Design/IndicationsUnchanged from predicate
Shelf Life (3 years)Verified in testing
BiocompatibilityLeveraged from predicate
SterilizationUnchanged from predicate

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the specific number of syringes used for each test (e.g., for critical dimensions, flow rate, compliance tests). It only refers to "testing per ISO 7886-2."
  • Data Provenance: Not explicitly stated, but based on the context of a 510(k) submission, this would be prospective testing conducted by the manufacturer, Sol-Millennium Medical Inc. The country of origin of the data is generally where the manufacturer performs their testing, which for Sol-Millennium Medical Inc. is based in Suwanee, Georgia.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This is a bench testing (non-clinical) submission involving a physical medical device (syringe). "Ground truth" in this context refers to standardized measurements and performance specifications outlined in ISO 7886-2, not expert consensus on diagnostic images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, especially those involving human interpretation of data where consensus is needed to establish a "ground truth" or resolve disagreements. For bench testing against a standard like ISO 7886-2, objective measurements are taken, compared against predefined limits, and either pass or fail.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret data (e.g., radiologists reading images) and the AI aims to improve their performance. This submission is for a physical medical device (syringe) and focuses on engineering performance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This is for a physical medical device, not an algorithm or AI. The testing described is "standalone" in the sense that the syringe's performance is measured directly, not in conjunction with human interpretation of its output.

7. Type of Ground Truth Used

The "ground truth" for the test set is established by international standards and specifications, specifically ISO 7886-2:2020 (for use with power-driven syringe pumps) and ISO 7886-1:2017 (for manual use). These standards define acceptable ranges for dimensional requirements, flow rate error, pump forces, and syringe compliance.

8. Sample Size for the Training Set

Not applicable. There is no AI/algorithm component requiring a training set mentioned in this submission. This is a traditional medical device (syringe) bench test.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue.

October 11, 2022

Sol-Millennium Medical Inc. Ying Zhao Sr. Manager, Regulatory Affairs 315 Shawnee North Dr. Suite 100 Suwanee, Georgia 30024

Re: K222744

Trade/Device Name: Sol-M Luer Lock Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: September 9, 2022 Received: September 12, 2022

Dear Ying Zhao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K222744

Device Name SOL-M Luer Lock Syringe

Indications for Use (Describe)

The Sol-M Luer Lock Syringe is intended for aspiration of medications and fluids into the body.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Sol-Millennium Medical Inc.
Sol-M™ Luer Lock Syringe

510(k) SUMMARY

K222744

Date PreparedOctober 7, 2022
SubmitterSol-Millennium Medical Inc.315 Shawnee North Dr. Suite 100Suwanee, GA 30024Phone: 815-578-2200
Primary Contact PersonYing ZhaoSol-Millennium Medical Inc.Sr. Manager, Regulatory AffairsPhone: 919-324-2268Email: yzhao@sol-m.com
Secondary ContactPersonPJ PasiaSol-Millennium Medical Inc.Director, Regulatory AffairsPhone: 847-363-1264Email: ppasia@sol-m.com
Device InformationTrade/Device Name: Sol-MTM Luer Lock SyringeRegulatory Number: 21 CFR 880.5860Regulatory Name: Piston SyringeRegulatory Class: IIProduct Code: FMFReview Panel: General Hospital
Predicate DeviceK101359: Sol-MTM Luer Lock SyringeManufacturer: Sol-Millennium Medical Inc.

{4}------------------------------------------------

INDICATIONS FOR USE

The Sol-MIM Luer Lock Syringe is intended for aspiration and injection of medications and fluids into the body.

REASON FOR SUBMISSION

The purpose of this submission is to evaluate the suitability of the Sol-M™ Luer Lock Syringe (formerly known as InviroStripe Luer Lock Syringes) cleared under K101359 by Sol-Millennium to be used with power-driven pumps and to add statements on the labeling indicating compatibility with pumps. There is no change to Indications for Use, material, dimensions, design, and packaging compared with the predicate device.

DEVICE DESCRIPTION

The Sol-M Luer Lock Syringe is used to aspirate and inject fluid/medication into the body. The syringe is a sterile, single-use, 3-part syringe with luer lock conical connection. The Sol-MTM Luer Lock Syringe consists of the following components:

  • . Barrel: The lubricated hollow cylinder has gradient markings on it. The tip of the barrel has a luer lock fitting for the user to attach a hypodermic needle or administration line.
  • Plunger: The plunger is used to pull back to aspirate fluids or depress to inject fluids. ●
  • Gasket: The gasket maintains the fluid in the barrel between the syringe nozzle and the ● plunger.

In addition to being used manually, the Sol-MTM Luer Lock Syringe can also work with a powerdriven syringe pump, with the luer lock tip connected to the IV administration line to deliver specific volume of fluid/medication.

The syringe is individually blister packaged, and EtO sterilized with SAL of 10-6. The Sol-M™ Luer Lock Syringe is provided in 1, 3, 5, 10, 20, 30, and 60ml volume.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The design and technological features of the subject device are the same as existing Sol-M™ Luer Lock Syringe. The existing Sol-MTM Luer Lock Syringe is identical to InviroStripe Standard Luer Lock Syringe cleared under K101359 by Inviro Medical, Inc. The brand name changed to Sol-M100 from InviroStripe when Sol-Millennium acquired Inviro Medical, Inc in 2011.

The difference between the subject device and the predicate device is that the subject device is suitable for use with power-driven syringe pumps. These two devices are identical on Indication for Use, material, dimensions, design, and technology. The similarities and differences are illustrated in the table below.

{5}------------------------------------------------

Element ofComparisonSubject DeviceSol-MTM Luer Lock Syringe(K222744)Predicate DeviceSol-MTM Luer Lock Syringe(K101359)Comment
Product CodeFMFFMFSame
Syringe TypePiston, syringePiston, syringeSame
Indication for UseThe Sol-MTM Luer LockSyringe is intended foraspiration and injection ofmedications and fluids intothe body.The Sol-MTM Luer LockSyringe is used to injectmedicine or vaccines into, orwithdraw fluids from, thebody.Different/SeeComment #1
Principle of OperationThree-piece piston syringe.Plunger is used to fill syringeas well as discharge the fluidfor both manual and pump-driven operation.Three-piece piston syringe.Plunger is used to fill syringeas well as discharge the fluid.Different/SeeComment #2
Specific Drug UseGeneral useGeneral useSame
Tip TypeLuer lock, ISO 80369-7Luer lock, ISO 80369-7Same
Nominal Capacity1, 3, 5, 10, 20, 30, and 60ml1, 3, 5, 10, 20, 30, and 60mlSame
Barrel Marking SpecsGraduated scaleGraduated scaleSame
Gradations legibilityBold MarkingsBold MarkingsSame
DeviceComponentBarrelPolypropylenePlungerPolypropyleneGasketIR RubberPolypropylenePolypropyleneIR RubberSame
LubricantCompositionSilicone oilSilicone oilSame
LubricantAmount/cm²≤ 0.25 mg/cm²≤ 0.25 mg/cm²Same
Barrel transparencyClearClearSame
BiocompatibilityISO 10993Per 21 CFR 801ISO 10993Per 21 CFR 801Same
LabelingInclude "Suitable for Usewith Power-Driven SyringePumps" statementDifferent/SeeComment #3
Sterilization MethodEtOEtOSame
Sterility Level (SAL)10-610-6Same
Shelf Life3 years5 yearsDifferent/SeeComment #4
Manual UseISO 7886-1ISO 7886-1Same
Critical DimensionsISO 7886-2Not applicableDifferent/SeeComment #5
Short-term Flow RateErrorISO 7886-2Not applicableDifferent/SeeComment #6
Pump ForcesISO 7886-2Not applicableDifferent/SeeComment #7
Syringe ComplianceISO 7886-2Not applicableDifferent/SeeComment #8

Comparison of Technology and Features

{6}------------------------------------------------

Comment #1:

The Indication for Use statement is reworded for clarification only, no change to the indication of the device.

Comment #2:

The subject device can be used for both manual injection with infusion pumps, while the predicate device can only be used for manual injection. The difference was evaluated by well-established methods in the FDA-recognized standard ISO 7886-2:2020 Sterile Hypodermic Syringes for Single Use - Part 2: Syringes for Use with Power-Driven Syringe Pumps. The Sol-MTM Luer Lock Syringe was subjected to design controls and tested to this specific standard to support substantial equivalence. Test methods and acceptance criteria utilized by Sol-Millennium have no changes or deviations from those included in ISO 7886-2. Testing results demonstrated that the subject device is as safe and effective as the predicate device.

Comment #3:

Labeling is updated to add pump compatible statement. As addressed in Comment 2, verification data demonstrated that no safety and effectiveness issues are raised.

Comment #4:

Due to commercialization demand for the subject device, 3-year shelf-life testing per ISO 7886-2 was conducted. The difference does not raise safety and effectiveness questions for its intended purpose.

Comment #5:

Testing per ISO 7886-2 was conducted to demonstrate the satisfaction of dimensional requirements which are critical to the fit of the syringe in a syringe pump and flow accuracy. Using the subject device with pumps does not raise any safety or effectiveness issue.

Comment #6:

Testing per ISO 7886-2 was conducted to demonstrate the maximum permissible flow rate error by subject device met the acceptance criteria. The syringe, when used with pump, can deliver medication or fluid at consistent flow rate. There is no safety and effectiveness questions raised.

Comment #7:

The required pump force to move plunger along barrel was measured and made available to pump manufacturers per ISO 7886-2. This information can be shared between pump and syringe manufacturers to ensure correct software programing by pump manufacturers.

Comment #8:

Testing per ISO 7886-2 was conducted to demonstrate the maximum allowable fluid displaced by the subject device met the acceptance criteria. The syringe, when used with pump, can delivery accurate volume of medication or fluid as be used as manually. There is no safety and effectiveness questions raised.

{7}------------------------------------------------

BIOCOMPATIBILITY

Biocompatibility testing was leveraged from the predicate device submission since there were no change to materials and manufacturing process and no biological concern was introduced.

STERILIZATION

There was no change in the sterilization process or materials involved with the sterilization, and the sterility assurance level remains unchanged.

SHELF LIFE

Shelf-life verification studies were conducted at T=3 under real time conditions to verify that the claimed shelf life of 3 years is supported. There were no failures identified and the test confirmed that the claimed expiration date is supported.

SUMMARY OF NON-CLINICAL TESTS (BENCH TESTING)

The Sol-M "1" Luer Lock Syringe had been previously tested in compliance with ISO 7886-1 for demonstrating manual use covered under K101359. To demonstrate that the device is also suitable for use with syringe pumps, Sol-Millennium has performed non-clinical testing according to ISO 7886-2. The performance and design testing conducted serve as the basis for establishing substantial equivalence; the testing included the following.

  • . ISO 7886-2:2020 Sterile hypodermic syringes for single use - Part 2: Syringes for use with power-driven syringe pumps
  • ISO 7886-1:2017 Sterile hypodermic syringes for single use Part 1: Syringes for manual . use

The compatibility with syringe pumps was considered the technical characteristic which could potentially introduce new risks to the subject device. Compliance to ISO 7886-2 Sterile hypodermic syringe for single use - Part 2: Syringes for use with power-driven syringe pumps, serves as the mitigation to the identified safety concerns.

CLINICAL TESTING

Clinical studies were not deemed necessary to support the substantial equivalence decision.

CONCLUSION

Sol-Millennium considers the Sol-MTM Luer Lock Syringe substantially equivalent to the predicate device. The subject device's Indications for Use, dimensional, materials, and technology are unchanged compared to the predicate device cleared under K101359. The potential risks introduced by adding pump compatibility were successfully mitigated by design control activities. The results of the testing demonstrated that the Sol-M™ Luer Lock Syringe

{8}------------------------------------------------

performs as intended and performs as well as the legally marketed predicate devices for the same intended use. Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).