K Number
K212340
Manufacturer
Date Cleared
2021-12-17

(142 days)

Product Code
Regulation Number
870.1250
Panel
NE
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EMBOGUARD Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravasular catheter into a selected blood vessel in the neurovascular system. The balloon provides temporary vascular occlusion during angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval Devices.

Device Description

EMBOGUARD Balloon Guide Catheter is a dual lumen, braid-reinforced, variable stiffness catheter with an eccentric inflation lumen, a radiopaque marker on the distal end and a bifurcated luer hub on the proximal end. A compliant balloon is mounted on the distal end. The distal end of the device shaft has a hydrophilic coating. Balloon Guide Catheter dimensions are indicated on the product label.

AI/ML Overview

This document describes the regulatory submission for the EMBOGUARD™ Balloon Guide Catheter (K212340). As such, the "acceptance criteria" and "device performance" refer to the results of bench and animal testing demonstrating substantial equivalence to a predicate device, rather than performance metrics for an AI algorithm. Therefore, many of the requested fields are not applicable in this context.

Here's a breakdown of the provided information, addressing your points where relevant:

1. Table of Acceptance Criteria and Reported Device Performance

The submission details numerous performance tests to demonstrate substantial equivalence to a predicate device. The general acceptance criterion for all of these tests is "Pass," meaning the device met the pre-determined specifications or demonstrated comparable performance to the predicate. The reported device performance for all listed tests is "Pass," indicating that all samples met the pre-determined acceptance criteria.

Study NameAcceptance CriteriaReported Device Performance (all samples)
Visual Inspection and Dimensional VerificationMet visual & dimensional specificationsPass
Visual Surface InspectionMet visual surface requirementsPass
Torque DurabilityCapable of 360 degrees hub rotationPass
Torque to FailureMet pre-determined acceptance criteriaPass (characterized only)
Torque TransmissionTorque transmission characterizedPass (characterized successfully after min 720° hub rotation)
Tensile StrengthMet tensile strength requirementsPass
Flexibility and Kink ResistanceAcceptable flexibility & kink resistancePass
Kink to FailureDetermined bend radius at which kink occursPass (characterized at different sections down to 2.5mm mandrel)
Catheter Lubricity and DurabilityMet coating lubricity & durability specificationsPass
Coating Length and LocationMet coating length & location specificationsPass
ParticulatesParticulates comparable to cleared controlsPass (characterized only)
Coating IntegrityCoating characterized pre- & post-simulated usePass (characterized only)
RadiopacityMet radiopaque characteristics requirementsPass
Simulated UseDemonstrated device performance in simulated anatomyPass
Simulated Clot Retrieval TestingDemonstrated device performance (incl. clot retrieval)Pass
User EvaluationDemonstrated device performance in simulated anatomyPass
Catheter Deliverability and Withdrawal ForceMet required tracking force specificationPass
Catheter Luer Hub Dimensional and Performance TestingMet requirements for small bore connectorsPass
Inner Lumen Integrity - PressureMet pressure requirementsPass
Inner Lumen Integrity - AspirationMet aspiration air leakage requirementsPass
Hub Shaft Pressure IntegrityMet minimum inner diameter specificationPass
Introducer Sheath Compatibility (Insertion and Withdrawal)Met required insertion & withdrawal forcePass
Introducer Sheath Compatibility (Re-insertion and Re-withdrawal)Device integrity maintained post re-insertion & re-withdrawalPass
Tip StiffnessSimilar to predicate devicePass
Balloon LocationMet balloon location specificationPass
Balloon Inflation / DeflationMet inflation & deflation time specificationsPass
Balloon FatigueNo degradation after 20 inflation cyclesPass
Balloon OD and ComplianceMet max recommended inflation volume vs. balloon diameter specificationsPass
Balloon ConcentricityMet balloon diameter specifications on each sidePass
Balloon BurstCapable of withstanding 2x and 2.5x recommended inflation volumePass
Conditioning, Distribution, and Shelf Life Aging - DeviceMet all specifications at baseline & after 1-year agingPass
Conditioning, Distribution, and Shelf Life Aging - PackagingMaintained packaging strength & integrityPass
Visual Inspection Packaging and LabelingMet packaging & labeling visual & adherence requirementsPass
System Visual Inspection (Dilator)Met visual & dimensional specificationsPass
Effective Length, Outer Diameter, Inner Diameter, Taper Diameter (Dilator)Met visual & dimensional specificationsPass
Tensile Strength (Dilator)Withstood peak tensile strengthPass
Kink Resistance (Dilator)No kinks, defects, or damagePass
Luer Dimensional (Dilator Hub)Met dimensional specificationsPass
Fluid Leakage by Pressure Decay (Dilator Hub)No leakage under pressurePass
Sub-atmospheric Pressure Air Leakage (Dilator Hub)No air leakagePass
Stress Cracking (Dilator Hub)No cracking under axial force & torquePass
Resistance to Separation from Axial Load (Dilator Hub)No separation from reference connectorPass
Resistance to Separation from Unscrewing (Dilator Hub)No overriding of threadsPass
Resistance to Overriding (Dilator Hub)No overriding of threadsPass
Torque Durability (Dilator Hub)No separation/embolization after 1 rotationPass
Visual Inspection (Luer Activated Valve)Met visual specificationsPass
Leak Test (Luer Activated Valve)No leakage during inflation/deflationPass
Activation (Luer Activated Valve)Allowed flow, no leakage when activated, prevented flow when syringe removedPass
LAV Male & Female Luer Dimensional (Luer Activated Valve)Met dimensional specificationsPass
Fluid Leakage by Pressure Decay, Sub-atmospheric Pressure Air Leakage (LAV)No leakage under pressurePass
Stress Cracking (LAV)No cracking under axial force & torquePass
Resistance to Separation from Axial Load (LAV)No separation from reference connectorPass
Resistance to Separation from Unscrewing (LAV)No separation from reference connectorPass
Resistance to Overriding (LAV)No overriding of threadsPass
Visual Inspection (Rotating Hemostasis Valve)Met visual specificationsPass
ID (Male Luer) (Rotating Hemostasis Valve)Met ID specificationPass
ID (Compression Seal) (Rotating Hemostasis Valve)Met ID specificationPass
ID (Side Port) (Rotating Hemostasis Valve)Met ID specificationPass
Seal Integrity (Pressure) (Rotating Hemostasis Valve)No leakage or collapse under pressurePass
Seal Integrity (Aspiration) (Rotating Hemostasis Valve)No leakage or collapse under aspirationPass
Gauging (RHV Hub Luer)Met conical part of lock fittingPass
Liquid Leakage (RHV Hub Luer)No leakage sufficient to form a falling dropPass
Air Leakage (RHV Hub Luer)No continued formation of air bubblesPass
Separation Force / Unscrewing Torque (RHV Hub Luer)Fitting remained attached to reference fittingPass
Ease of Assembly (RHV Hub Luer)Met requirements for small bore connectorsPass
Resistance to Overriding (RHV Hub Luer)No overriding of threads or lugsPass
Stress Cracking (RHV Hub Luer)Withstood stress crackingPass

2. Sample Size Used for the Test Set and Data Provenance

  • Bench Testing: The specific sample sizes for each bench test are not explicitly stated in the document. However, the results consistently state "All samples met the pre-determined acceptance criteria," implying that a sufficient number of samples were tested to gain confidence in the results according to the referenced standards. The data provenance is from internal testing conducted by Neuravi Ltd.
  • Animal Studies: The document mentions "Acute animal study has been performed." The specific number of animals is not provided. The model used was a swine model. This is prospective data from animal testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable. The device is a medical catheter, and its performance is evaluated through physical and biological testing, not by expert interpretation of images or other data that would require "ground truth" to be established by experts in the context of an AI study.

4. Adjudication Method for the Test Set

This information is not applicable. The device is a medical catheter, and its performance is evaluated through physical and biological testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was conducted. This device is a physical medical device, not an AI algorithm requiring human reader performance evaluation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This information is not applicable. The device is a physical medical device, not an AI algorithm.

7. Type of Ground Truth Used

The "ground truth" for this device's performance is established through adherence to recognized international standards, internal specifications, and comparison to a predicate device. For example:

  • Biocompatibility: Based on established biological response criteria defined in ISO 10993 series standards (e.g., non-cytotoxic, non-sensitizing, non-pyrogenic).
  • Bench Testing: Based on pre-determined product performance specifications derived from relevant ISO standards (e.g., ISO 10555, ISO 11070) and FDA guidance documents.
  • Animal Studies: Based on evaluation of usability, effectiveness, safety (histological evaluation), and comparison to the predicate device in a swine model.

There is no "expert consensus," "pathology," or "outcomes data" in the sense of a medical diagnostic ground truth for an AI algorithm.

8. Sample Size for the Training Set

This information is not applicable. The device is a physical medical device, not an AI algorithm.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable. The device is a physical medical device, not an AI algorithm.


Summary for this specific device:

The K212340 submission for the EMBOGUARD™ Balloon Guide Catheter focuses on demonstrating substantial equivalence to an existing predicate device (8F FlowGate Balloon Guide Catheter, K153729) through extensive non-clinical testing.

  • Acceptance Criteria: For all tests, the acceptance criteria were pre-determined specifications, often derived from international standards (e.g., ISO 10993, ISO 10555, ISO 11070, ISO 80369) and FDA guidance documents. For comparison studies, the criterion was similarity or equivalence to the predicate device.
  • Device Performance: The device "Passed" all listed biocompatibility, sterilization/shelf-life, and bench tests, indicating it met the pre-determined acceptance criteria for each.
  • Study Types: The studies included:
    • Biocompatibility Testing: Covering cytotoxicity, sensitization, intracutaneous irritation, systemic toxicity, hemocompatibility (hemolysis, complement activation, thrombogenicity), and genotoxicity.
    • Sterilization and Shelf Life Studies: Validation of EtO sterilization and 1-year shelf life.
    • Bench Testing: A comprehensive suite of physical and mechanical tests covering visual inspection, dimensions, torque, tensile strength, flexibility, kink resistance, lubricity, coating, particulates, radiopacity, simulated use (including clot retrieval), user evaluation, compatibility with ancillary devices, lumen integrity, and balloon performance. All these were direct physical tests of the device.
    • Animal Studies: An acute animal study in a swine model to assess usability, effectiveness, safety, and histological response compared to the predicate device.
  • No clinical studies were performed.
  • The document does not detail any AI component, and thus questions related to AI-specific metrics (sample sizes for training/test sets, expert ground truth, MRMC studies) are not relevant to this submission. The "ground truth" here refers to the validated physical and biological properties of the device as measured against established scientific and regulatory standards.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).