K Number
K153729
Manufacturer
Date Cleared
2016-01-27

(30 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Flow Gate Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.

Device Description

Like the predicate device, the 8F FlowGate2™ Balloon Guide Catheter is a coaxial- lumen, braidreinforced, variable stiffness catheter designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. A radiopaque marker is included on the distal end for angiographic visualization. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. A bifurcated luer hub on the proximal end allows attachments for flushing, inflation and aspiration. Modified FlowGate™ Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the 8F FlowGate2™ Balloon Guide Catheter. It outlines the acceptance criteria and testing performed to demonstrate its substantial equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance:

The document broadly states that all performed tests "Met established acceptance criteria." However, it does not provide specific quantitative acceptance criteria or detailed numerical results for each test. Instead, it offers a list of tests and a blanket statement of compliance.

TestReported Device Performance
Modifications to catheter shaft
Catheter Simulated Use (Retriever/Microcatheter Compatibility)Met established acceptance criteria.
Catheter Kink Resistance, Variables (Test Kink Resistance)Met established acceptance criteria.
Catheter Tip Flexibility (Catheter/Vessel Interaction)Met established acceptance criteria.
Torque TransmissionMet established acceptance criteria.
High Pressure LeakMet established acceptance criteria.
Distal Shaft TrackabilityMet established acceptance criteria.
Lumen CompatibilityMet established acceptance criteria.
6F Dilator Accessory
Dimensional VerificationMet established acceptance criteria.
Dilator Shape RetentionMet established acceptance criteria.
Air Leak ResistanceMet established acceptance criteria.
Liquid Leak ResistanceMet established acceptance criteria.
Tensile StrengthMet established acceptance criteria.
High Pressure LeakMet established acceptance criteria.
Catheter/Vessel Interaction (Catheter Tip Flexibility)Met established acceptance criteria.
Kink ResistanceMet established acceptance criteria.
Tuohy Borst Valve with Sideport Accessory
Hub GaugingMet established acceptance criteria.
Liquid LeakageMet established acceptance criteria.
Air LeakageMet established acceptance criteria.
Separation ForceMet established acceptance criteria.
Unscrewing TorqueMet established acceptance criteria.
Ease of AssemblyMet established acceptance criteria.
Resistance to OverridingMet established acceptance criteria.
Stress CrackingMet established acceptance criteria.
Catheter Simulated Use (Tuohy Borst Valve with Sideport seal over devices and wires)Met established acceptance criteria.
BiocompatibilityMet biological safety requirements per EN ISO 10993-1: 2009 for externally communicating medical devices with circulating blood contact for less than 24 hours.
SterilizationAdopted into a validated sterilization process in accordance with ISO 11135:2007 and AAMI TIR 28:2009.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample sizes (number of devices or cases) used for the verification and validation (test set) activities. It also does not explicitly state the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be primarily bench testing performed on manufactured devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The study described is a technical verification and validation of a medical device's physical and functional properties, not a clinical study involving human assessment of data requiring expert ground truth establishment in the typical sense (e.g., for diagnostic accuracy).

4. Adjudication Method for the Test Set:

Not applicable. The tests performed are objective technical assessments (e.g., measuring leak resistance, tensile strength, kink resistance), not subjective assessments requiring expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

No, an MRMC comparative effectiveness study was not done. This submission is for a percutaneous catheter, not a diagnostic or AI-assisted interpretation device that would typically involve human readers. The study focuses on demonstrating the substantial equivalence of the new device to existing predicate devices through material, design, and performance testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device is not an algorithm or AI system. It is a physical medical device (catheter). The testing focuses on the physical and functional performance of the device itself.

7. The Type of Ground Truth Used:

The "ground truth" for the tests performed is based on established engineering principles, industry standards (e.g., ISO 11135, EN ISO 10993-1), and internal performance specifications determined through design and risk assessment processes. For instance, for "High Pressure Leak," the ground truth is a pass/fail criterion based on whether the device exhibits leakage under specific pressure. These are objective measures rather than subjective expert consensus, pathology, or outcomes data in the context of diagnostic performance.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI or machine learning device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures in profile facing right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 27,2016

Concentric Medical, Inc. Mr. Shazia Hakim Senior Regulatory Affairs Specialist 301 East Evelyn Avenue Mountain View, California 94041

Re: K153729

Trade/Device Name: 8F FlowGate Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: December 22, 2015 Received: December 28, 2015

Dear Mr. Hakim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

William J. Heetderks -S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153729

Device Name 8F FlowGate Balloon Guide Catheter

Indications for Use (Describe)

Flow Gate Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Device Trade Name:8F FlowGate2™ Balloon Guide Catheter
Common Name:Percutaneous Catheter
Classification Name:Percutaneous Catheter, 21CFR 870.1250 Class II
Product Code:DQY
Submitter:Concentric Medical, Inc.301 E. Evelyn AvenueMountain View, CA 94041Tel 510 - 413-2636Fax 510 - 413-2588Facility Registration #2954917
Contact:Shazia HakimSenior Regulatory Affairs Specialist

Date Prepared: December 22, 2015

Predicate Device:

Primary Predicate Device
K131492(clearance grantedOctober 3, 2013)Modified FlowGate™ BalloonGuide Catheter
Reference Predicate Device
K122581(clearance grantedNovember 21, 2012)Concentric Balloon GuideCatheter

{4}------------------------------------------------

Device Description

Like the predicate device, the 8F FlowGate2™ Balloon Guide Catheter is a coaxial- lumen, braidreinforced, variable stiffness catheter designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. A radiopaque marker is included on the distal end for angiographic visualization. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. A bifurcated luer hub on the proximal end allows attachments for flushing, inflation and aspiration. Modified FlowGate™ Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label.

Accessories

The 8F FlowGate2™ Balloon Guide Catheter is packaged with a Dilator, Rotating Hemostasis Valve, Tuohy Borst Valve with Sideport, Peel Away Sheaths, Luer-Activated Valve, and Extension tubing.

Indications for Use

The Indications for Use are the same as the predicate devices and are as follows:

FlowGate" Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.

Technological Characteristics and Product Feature Comparison

The subject device, 8F FlowGate2™ Balloon Guide Catheter is substantially equivalent to the predicate devices in terms of:

  • indications for use
  • fundamental scientific technology
  • fundamental design
  • materials and processes for packaging and sterilization of devices

A tabular comparison of the specific technological characteristics between the predicate devices and subject device is provided below.

{5}------------------------------------------------

Product Feature Comparison of Subject Device with Predicate Devices (K122581, K131492)
FeatureReference Predicate Device (K122581)Primary Predicate Device (K131492)Subject DeviceRationale for difference (if applicable)
Indications for UseThe Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.Same except for product name as “FlowGate Balloon Guide Catheter”Same as predicate device (K131492)Not applicable
FeatureReference PredicateDevice (K122581)PrimaryPredicate Device(K131492)Subject DeviceRationale fordifference (ifapplicable)
Device DescriptionThe Concentric BalloonGuide Catheters arecoaxial-lumen, braid-reinforced, variablestiffness catheters designedfor use in facilitating theinsertion and guidance ofan intravascular catheterinto a selected blood vesselin the peripheral and neurovascular systems. Aradiopaque marker isincluded on the distal endfor angiographicvisualization. A compliantballoon is mounted on thedistal end to providetemporary vascularocclusion duringangiographic procedures. Abifurcated luer hub on theproximal end allowsattachments for flushing,inflation and aspiration.Balloon Guide Catheterdimensions and maximumrecommended ballooninflation volume areindicated on product label.Same except forproduct name as"FlowGateBalloon GuideCatheter"Same aspredicate device(K131492)Not applicable
Target PopulationPatients undergoingpercutaneousinterventional proceduresSameSameNot applicable
Anatomical SitesPeripheral and neurovasculatureSameSameNot applicable
Regulatory Status
Regulation Number21CFR 870.1250SameSameNot applicable
Regulation NamePercutaneous CatheterSameSameNot applicable
FeatureReference PredicateDevice (K122581)PrimaryPredicate Device(K131492)Subject DeviceRationale fordifference (ifapplicable)
Regulatory ClassIISameSameNot applicable
Product CodeDQYSameSameNot applicable
Materials
Outer Jacket-Proximal ShaftPurple Pebax® 7233Barium sulfateSameSameNot applicable
Outer Jacket- DistalShaftBlue Pebax 6333Barium sulfateSameSameNot applicable
Inner Jacket-Proximal ShaftWhite Polyamide BlendBarium sulfateSameSameNot applicable
Inner Jacket -Proximal TransitionNatural Pebax 7233Barium sulfateSameSameNot applicable
Inner Jacket -Proximal Mid ShaftNatural Pebax 5533,Barium sulfateSameSameNot applicable
Inner Jacket -Distal Mid ShaftNatural Pebax 4033SameSameNot applicable
Inner Jacket -Distal ShaftWhite Pebax 2533,Barium sulfateSameSameNot applicable
Distal TipWhite Pebax 2533,Barium sulfateWhite Pebax3533White Pebax2533Barium sulfateSame aspredicate device(K131492)Not applicable
Balloon MaterialNuSil MED 4035 siliconeelastomerNuSil MED 4025siliconeelastomerSame aspredicate device(K131492)Not applicable
Balloon BondAdhesiveLoctite 4011 CyanoacrylateLoctite 3974 UV AdhesiveSameSameNot applicable
FeatureReference PredicateDevice (K122581)PrimaryPredicate Device(K131492)Subject DeviceRationale fordifference (ifapplicable)
Braid304V Stainless SteelSameSameNot applicable
Braid distal endsecurementLoctite UV AdhesiveLoctite 4014CyanoacrylateSame aspredicate device(K131492)Not applicable
LinerEtched PTFESameSameNot applicable
Marker Band90% Platinum/10% IridiumSameSameNot applicable
Luer Hub AdhesiveLoctite 3321 UV AdhesiveSameSameNot applicable
2-Arm Luer HubPolyurethaneSameSameNot applicable
Strain ReliefWhite Pebax 4033White PolyolefinWhite Pebax4033White PolyolefinSame aspredicate device(K131492)Not applicable
Labeled ShaftOuter Diameter8F (2.7mm)SameSameNot applicable
Labeled Shaft InnerDiameter6.4F (0.084 inch /2.1 mm)SameSameNot applicable
Maximum outerdiameter alongeffective length0.108 inchSameSameNot applicable
Effective Lengths100 cm and 90 cm95 cm and 85 cmSame aspredicate device(K131492)Not applicable
RadiopaqueMarker Locationfrom Distal Tip2 mm0.75 mmSame aspredicate device(K131492)Not applicable
RadiopaqueMarker Length0.020 inchSameSameNot applicable
FeatureReference PredicateDevice (K122581)PrimaryPredicate Device(K131492)Subject DeviceRationale fordifference (ifapplicable)
Accessory DevicesProvidedDilator (1)Rotating Hemostasis Valve(1)Tuohy Borst Valve withsideport (1)Peel Away Sheaths (2)Luer-Activated Valve (1)Same aspredicatedevice,with addition ofExtensionTubing (1)Same aspredicatedevice, K131492except formodified Dilatortip shape andTuohy BorstValve withSideport withsmaller IDThe 6F Dilatorand Tuohy BorstValve withSideport havebeen modifiedto meet userpreferences.Bench testinghasdemonstratedthat thecompatibility ofmodified 6FDilator andTuohy BorstValve withSideport withthe subjectdevice does notaffect the safetyandeffectiveness ofthe device.
Tuohy Borst Valvewith SideiportmaterialABS bodypolypropylene capPVC tubingpolycarbonate luersilicone sealLDPE washer/ringSameSame aspredicatedevice, K131492except forPolycarbonatebodypolycarbonateсарPVC tubingpolycarbonateluersilicone sealPTFEwasher/ringThe Tuohy BorstValve withSideport hasbeen modifiedto meet userpreferences.Bench testinghasdemonstratedthat thecompatibility ofmodified TuohyBorst Valve withSideport withthe subjectdevice does notaffect the safetyandeffectiveness ofthe device.
FeatureReference PredicateDevice (K122581)PrimaryPredicate Device(K131492)Subject DeviceRationale fordifference (ifapplicable)
Dilator ShaftmaterialBlue PebaxBarium sulfateBismuth subcarbonate (attip only) Stainless Steelbraid wireSameSameNot applicable
Dilator HubmaterialWhite PolycarbonateSameSameNot applicable
Dilator Strain ReliefSleeve materialGreen PebaxSameSameNot applicable
Dilator adhesivematerialsCyanoacrylateSameSameNot applicable
Dilator Shaft OuterDiameter (nominal)0.072 inch (distal), 0.078inch (proximal)SameSameNot applicable
Dilator InnerDiameter (nominal)0.041 inch (distal), 0.050"(proximal)SameSameNot applicable
Dilator EffectiveLength (nominal)123 cmSameSameNot applicable
Dilator Distal TipAngle60°SameSameNot applicable
FeatureReference PredicateDevice (K122581)PrimaryPredicate Device(K131492)Subject DeviceRationale fordifference (ifapplicable)
Dilator Distal TipLength1.40 inchSame0.67 inchesThe tip shape ofthe 6F Dilatorhas beenshortened tomeet userpreferences.Bench testinghasdemonstratedthat thecompatibility ofthe shorterdistal tip lengthof the 6F Dilatorwith the subjectdevice does notaffect the safetyandeffectiveness ofthe device.
Packaging
Materials andConfigurationPolyethylene tubes forcatheter and dilator, HDPEsterile prep card andpackaging card, Tyvek andLLDPE/Nylon Film Pouch,Chipboard cartonSameSameNot applicable
SterilizationMethod100% EtOSameSameNot applicable
How SuppliedSterile, Single UseSameSameNot applicable

Product Feature Comparison of Subject Device with Predicate Devices (K122581, K131492)

{6}------------------------------------------------

Concentric Medical, Inc.

Premarket Notification, Special 510(k) 8F FlowGate²™ Balloon Guide Catheter

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Concentric Medical, Inc.

Premarket Notification, Special 510(k) 8F FlowGate²™ Balloon Guide Catheter

{10}------------------------------------------------

Concentric Medical, Inc. Premarket Notification, Special 510(k) 8F FlowGate²™ Balloon Guide Catheter

{11}------------------------------------------------

Concentric Medical, Inc. Premarket Notification, Special 510(k) 8F FlowGate²™ Balloon Guide Catheter

{12}------------------------------------------------

Risk Assessment

Risk assessment of the modifications has been conducted in accordance with EN ISO 14971:2012. Concentric Medical, Inc. has determined that the modifications to the predicate devices raise no new questions of safety or effectiveness.

Results of verification and validation testing have demonstrated the subject device is substantially equivalent to the predicate devices. Furthermore, the modifications did not result in any new failure modes nor were there any changes to existing failure modes.

Testing Summary

The results of verification and validation testing conducted on the subject device demonstrate that it performs as designed, and is suitable for its intended use. Specifically, the following tests were performed on the subject device:

TestConclusions
Modifications to catheter shaft
Catheter Simulated Use (Retriever/Microcatheter Compatibility)Met established acceptance criteria.
Catheter Kink Resistance, Variables (TestKink Resistance)Met established acceptance criteria.
Catheter Tip Flexibility (Catheter/Vessel Interaction)Met established acceptance criteria.
Torque TransmissionMet established acceptance criteria.
High Pressure LeakMet established acceptance criteria.
6F Dilator AccessoryMet established acceptance criteria.
Dimensional VerificationMet established acceptance criteria.
Dilator Shape RetentionMet established acceptance criteria.
Air Leak ResistanceMet established acceptance criteria.
Liquid Leak ResistanceMet established acceptance criteria.
Tensile StrengthMet established acceptance criteria.
High Pressure LeakMet established acceptance criteria.
Catheter/Vessel Interaction (Catheter Tip Flexibility)Met established acceptance criteria.

{13}------------------------------------------------

TestConclusions
Kink ResistanceMet established acceptance criteria.
Distal Shaft TrackabilityMet established acceptance criteria.
Lumen CompatibilityMet established acceptance criteria.
Tuohy Borst Valve with Sideport Accessory
Hub GaugingMet established acceptance criteria.
Liquid leakageMet established acceptance criteria.
Air LeakageMet established acceptance criteria.
Separation ForceMet established acceptance criteria.
Unscrewing TorqueMet established acceptance criteria.
Ease of AssemblyMet established acceptance criteria.
Resistance to OverridingMet established acceptance criteria.
Stress CrackingMet established acceptance criteria.
Catheter Simulated Use (Tuohy Borst Valvewith Sideport seal over devices and wires)Met established acceptance criteria.

Biocompatibility

Biocompatibility testing was completed for the previously cleared devices under K122581 and K131492. These results apply to the subject device because it uses the same materials and processes as previously cleared predicate devices. Biocompatibility testing was conducted on the accessories packaged with the previously cleared devices under K131492, which included the modified Tuohy Borst Valve with Sideport to be packaged with the proposed 8F FlowGate² BGC in compliance with EN ISO 10993-1. The results demonstrate that the subject device with its accessories meet biological safety requirements per EN ISO 10993-1: 2009 for externally communicating medical devices with circulating blood contact for less than 24 hours.

{14}------------------------------------------------

Sterilization and Shelf life

The subject device, including its accessories, are sterilized by 100% EtO and have been adopted into a validated sterilization process in accordance with the principles of ISO 11135:2007, ("Sterilization of health-care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices") and AAMI TIR 28:2009 ("Product Adoption and Process Equivalency for Ethylene Oxide Sterilization") without deviations.

Summary of Substantial Equivalence

The subject device is substantially equivalent to the predicate devices with regard to device design, materials, intended use, and patient population. The conclusions drawn from the risk assessments, verification and validation testing conducted demonstrate that the subject device performs as designed and has an equivalent performance profile as the predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).