K Number
K211604

Validate with FDA (Live)

Date Cleared
2021-10-27

(155 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs using ASTM D6978 and will be labeled with a statement of compliance and a summary of the testing results.

Device Description

Powder Free Nitrile Examination Gloves(Blue), Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes-Extra Small, Small, Medium, Large and Extra Large. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05(2019).

AI/ML Overview

The provided document describes the acceptance criteria and performance data for "Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs." This is a Class I medical device, and the testing focuses on its physical properties, biocompatibility, and resistance to chemotherapy drug permeation. It is crucial to note that this document pertains to a physical medical device (gloves) and not an AI/Software as a Medical Device (SaMD). Therefore, questions related to AI/SaMD specific aspects like expert data labeling, MRMC studies, training set data, etc., are not applicable to this submission. The information provided aligns with typical performance testing for physical devices.

Here's the breakdown of the acceptance criteria and reported performance, tailored to the available information:


Acceptance Criteria and Reported Device Performance

1. Table of Acceptance Criteria and the Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319-19LengthMinimum 230mmAll size ≥230mm
ASTM D6319-19Palm WidthXS: 70 ± 10mmS: 80 ± 10mmM: 95 ± 10mmL: 110 ± 10mmXL: 120 ± 10mmXS: 77-78mmS: 86-88 mmM: 96-98mmL: 108-110 mmXL: 116-117 mm
ASTM D6319-19ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)≥0.05mm
ASTM D6319-19, ASTM D412-16Tensile Strength, Before Aging14MPa, min≥14 MPa
ASTM D6319-19, ASTM D412-16Tensile Strength, After Accelerated Aging14MPa, min≥14 MPa
ASTM D6319-19, ASTM D412-16Ultimate Elongation, Before Aging500%, min≥500%
ASTM D6319-19, ASTM D412-16Ultimate Elongation, After Accelerated Aging400%, min≥400%
ASTM D5151-19Freedom from holesG-I, AQL 2.5Meet and above AQL2.5 requirements
ASTM D6124-06 (2017)Powder-Content≤2 mg per glove≤2 mg, meet requirements
ISO 10993-10:2010Skin Irritation StudyUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritant
ISO 10993-10:2010Maximization Sensitization StudyUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
ISO 10993-5:2009Cytotoxicity TestUnder the conditions of this study, the test article extract showed potential toxicityUnder the conditions of this study, the test article extract showed potential toxicity
ISO 10993-11:2017Systemic toxicityUnder the conditions of this study, there was no evidence of systemic toxicity.Under the conditions of this study, there was no evidence of systemic toxicity.
ASTM D6978-05 (2019)Tested for Use with Chemotherapy Drugs (Minimum BDT)No explicit numerical acceptance criteria for each drug are listed as "acceptance criteria" in the table. The table provides the reported values for the device. The "Acceptance Criteria" column for this test indicates "/", implying that the objective is to report the measured breakthrough times. Note: For Carmustine (BCNU) and Thio Tepa, the document explicitly states a warning due to "extremely low permeation times."Bleomycin (15.0mg/ml): >240Bortezomib (Velcade) 1mg/ml: >240Busulfan (6.0mg/ml): >240Carboplatin, (10.0mg/ml): >240Carmustine(BCNU) (3.3 mg/ml): 7.0 minutesChloroquine, (50.0mg/ml): >240Cisplatin (1mg/ml): >240Cyclophosphamide (Cytoxan) (20mg/ml): >240Cyclosporin, (100.0mg/ml): >240Cytarabine, (100.0mg/ml): >240Dacarbazine (10mg/ml): >240Daunorubicin, HCL (5.0mg/ml): >240Docetaxel, (10.0mg/ml): >240Doxorubicin HCL (2mg/ml): >240Epirubicin HCL (Ellence), (2.0mg/ml): >240Etoposide (20mg/ml): >240Fludarabine, (25.0mg/ml): >240Fluorouracil (50mg/ml): >240Gemcitabine (38.0mg/ml): >240Idarubicin HCL (1.0mg/ml): >240Ifosfamide, (50.0mg/ml): >240Irinotecan, (20.0mg/ml): >240Mechlorethamine HCI, (1.0mg/ml): >240Melphalan, (5.0mg/ml): >240Methotrexate (25mg/ml): >240Mitomycin C, (0.5mg/ml): >240Mitoxantrone, (2.0mg/ml): >240Oxaliplatin, (5.0mg/ml): >240Paclitaxel (Taxol) (6mg/ml): >240Paraplatin, (10.0mg/ml): >240Retrovir, (10.0mg/ml): >240Rituximab, (10.0mg/ml): >240Thio Tepa (10mg/ml): 23.0 minutesTopotecan HCL, (1.0mg/ml): >240Trisonex, (1.0mg/ml): >240Vincristine, (1.0mg/ml): >240

Regarding the AI/SaMD specific questions, the following points are not applicable as this document describes performance testing for a physical medical glove.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • The document does not specify the exact sample sizes for each physical test (e.g., how many gloves were tested for tensile strength, or how many samples for biocompatibility). Standard test methods (ASTM, ISO) typically define minimum sample sizes. The data provenance is implied to be from testing labs in relation to the manufacturer (Shanxi Hongjin Plastic Technology Co., Ltd. in China) and their US representative. This is a prospective testing of the manufactured product.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. Ground truth for physical device performance is established by standardized laboratory testing methods (ASTM, ISO) rather than expert human interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are typically relevant for human-in-the-loop assessments or expert consensus in image interpretation, not for physical material testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/SaMD.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an AI/SaMD. The "standalone" performance here refers to the physical properties of the glove itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The 'ground truth' for this device is based on objective measurements derived from standardized testing protocols (ASTM and ISO standards) measuring physical properties (like length, thickness, tensile strength, freedom from holes, powder content) and chemical resistance (breakthrough time for chemotherapy drugs), as well as biological responses (irritation, sensitization, cytotoxicity, systemic toxicity).

8. The sample size for the training set

  • Not applicable. This is not an AI/SaMD, so there is no "training set."

9. How the ground truth for the training set was established

  • Not applicable. As there is no "training set" for physical device testing, there is no ground truth establishment method for it.

Summary of the Study Proving Acceptance Criteria:

The study proving the device meets the acceptance criteria is a series of non-clinical performance tests conducted in accordance with recognized international (ISO) and national (ASTM) standards. These tests assess the physical, chemical, and biological properties of the nitrile examination gloves.

The "Non-Clinical Performance Data" section (Page 9 of the document) explicitly lists the test methods used, their purpose, the acceptance criteria, and the results, confirming that the proposed device complies with these standards. For example, ASTM D6319-19 was used for dimensions and tensile properties, ASTM D5151-19 for freedom from holes, ASTM D6124-06 for powder content, and ASTM D6978-05 (Reapproved 2019) for chemotherapy drug permeation. Biocompatibility was assessed using ISO 10993 series standards (Parts 5, 10, and 11) for cytotoxicity, irritation, sensitization, and systemic toxicity.

The conclusion states, "The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device." This indicates that the results met the established benchmarks (which are often derived from the requirements of the standards themselves or comparison to a predicate device).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 27, 2021

Shanxi Hongjin Plastic Technology Co., Ltd. Kathy Liu Project Manager Hongray USA Medical Products Inc. 3973 Schaefer Avenue Chino, California 91710

Re: K211604

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: July 23, 2021 Received: September 22, 2021

Dear Kathy Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical And Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211604

Device Name

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

Powder Free Nittile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs using ASTM D6978 and will be labeled with a statement of compliance and a summary of the testing results.

Chemotherapy Drug Permeation

The following chemicals have been tested with these gloves:

* Please note that the following drugs have extremely low permeation times:
Click Class of Children Carder Carder Carder1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Chemotherapy DrugMinimum BDT (Minutes
Bleomycin (15.0mg/ml)>240
Bortezomib (Velcade) 1mg/ml>240
Busulfan (6.0mg/ml)>240
Carboplatin, (10.0mg/ml)>240
Carmustine(BCNU) (3.3 mg/ml)7.0
Chloroquine, (50.0mg/ml)>240
Cisplatin (1mg/ml)>240
Cyclophosphamide (Cytoxan) (20mg/ml)>240
Cyclosporin, (100.0mg/ml)>240
Cytarabine, (100.0mg/ml)>240
Dacarbazine (10mg/ml)>240
Daunorubicin, HCL (5.0mg/ml)>240
Docetaxel, (10.0mg/ml)>240
Doxorubicin HCL (2mg/ml)>240
Epirubicin HCL (Ellence), (2.0mg/ml)>240
Etoposide (20mg/ml)>240
Fludarabine, (25.0mg/ml)>240
Fluorouracil (50mg/ml)>240
Gemcitabine (38.0mg/ml)>240
Idarubicin HCL (1.0mg/ml)>240
Ifosfamide, (50.0mg/ml)>240
Irinotecan, (20.0mg/ml)>240
Mechlorethamine HCI, (1.0mg/ml)>240
Melphalan, (5.0mg/ml)>240
Methotrexate (25mg/ml)>240
Mitomycin C, (0.5mg/ml)>240
Mitoxantrone, (2.0mg/ml)>240
Oxaliplatin, (5.0mg/ml)>240
Paclitaxel (Taxol) (6mg/ml)>240
Paraplatin, (10.0mg/ml)>240
Retrovir, (10.0mg/ml)>240
Rituximab, (10.0mg/ml)>240
Thio Tepa (10mg/ml)23.0
Topotecan HCL, (1.0mg/ml)>240
Trisonex, (1.0mg/ml)>240
Vincristine, (1.0mg/ml)>240

{3}------------------------------------------------

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 7.0 minutes, Thio Tepa: 23 minutes, Warning: Do not use with Carmustine (BCNU) and Thio Tepa.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

The assigned 510(K) numbers: K211604 Date Prepared: October 27, 2021

1. Owner's Identification:

Mr. Wu Zhigang Shanxi Hongjin Plastic Technology Co., Ltd. Coal Bed Gas Industrial Zone, Ou'e Town, Daning County, Linfen City, Shanxi Province

Tel: 86-311-66766067 Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Avenue, Chino, CA 91710, USA Tel: 909-590-1611 Email: kathyliu@hongrayusa.com or fdareg@126.com

2. Name of the Device:

Trade / Product Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs Common Name: Examination Gloves Classification Name: Patient Examination Glove Specialty Classification Regulation: 21 CFR 880.6250 Product Code: LZA LZC Classification Panel: General Hospital Device Class: Class I

3. Predicate Device Information:

Better Care Plastic Technology Co., Ltd Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs (Blue) (K182600)

4. Device Description:

Powder Free Nitrile Examination Gloves(Blue), Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes-Extra Small, Small, Medium, Large and Extra Large.

Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05(2019).

{5}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

5. Indications for Use:

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05(2019) and will be labeled with a statement of compliance and a summary of the testing results. Chemotherapy Drug Permeation

Chemotherapy DrugMinimum BDT (Minutes)
Bleomycin (15.0mg/ml)>240
Bortezomib (Velcade) 1mg/ml>240
Busulfan (6.0mg/ml)>240
Carboplatin, (10.0mg/ml)>240
Carmustine(BCNU) (3.3 mg/ml)7.0
Chloroquine, (50.0mg/ml)>240
Cisplatin (1mg/ml)>240
Cyclophosphamide (Cytoxan) (20mg/ml)>240
Cyclosporin, (100.0mg/ml)>240
Cytarabine, (100.0mg/ml)>240
Dacarbazine (10mg/ml)>240
Daunorubicin, HCL (5.0mg/ml)>240
Docetaxel, (10.0mg/ml)>240
Doxorubicin HCL (2mg/ml)>240
Epirubicin HCL (Ellence), (2.0mg/ml)>240
Etoposide (20mg/ml)>240
Fludarabine, (25.0mg/ml)>240
Fluorouracil (50mg/ml)>240
Gemcitabine (38.0mg/ml)>240
Idarubicin HCL (1.0mg/ml)>240
Ifosfamide, (50.0mg/ml)>240
Irinotecan, (20.0mg/ml)>240
Mechlorethamine HCI, (1.0mg/ml)>240
Melphalan, (5.0mg/ml)>240
Methotrexate (25mg/ml)>240
Mitomycin C, (0.5mg/ml)>240
Mitoxantrone, (2.0mg/ml)>240
Oxaliplatin, (5.0mg/ml)>240
Paclitaxel (Taxol) (6mg/ml)>240
Paraplatin, (10.0mg/ml)>240
Retrovir, (10.0mg/ml)>240
Rituximab, (10.0mg/ml)>240

The following chemicals have been tested with these gloves:

{6}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY

Thio Tepa (10mg/ml)23.0
Topotecan HCL, (1.0mg/ml)>240
Trisonex, (1.0mg/ml)>240
Vincristine, (1.0mg/ml)>240
  • Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 7.0 minutes, Thio Tepa: 23 minutes

6. Comparison of Subject Device and Predicate Device:

The proposed device will be known as Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs.

The following tables are summaries of the technological characteristics, biocompatibility and testing for use with chemotherapy drugs of the proposed and predicate devices.

Proposed Device K211604Predicate Device K182600Remark
Trade NamePowder Free NitrileExamination Gloves(Blue), Tested for Usewith ChemotherapyDrugsPowder Free NitrileExamination Gloves,Tested for Use withChemotherapy Drugs(Blue)Same
Product CodeLZA, LZCLZA, LZCSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame
Indications for UsePowder Free NitrileExamination Gloves (Blue),Tested for Use withChemotherapy Drugs is adisposable device intended formedical purposes that is wornon the examiner's hand toprevent contaminationbetween patient and examiner.Powder Free NitrileExamination Gloves, Tested forUse with Chemotherapy Drugs(Blue) is a disposable deviceintended for medical purposesthat is worn on the examiner'shand to prevent contaminationbetween patient and examiner.Same
Powder or PowderFreePowder FreePowder FreeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorBlueBlueSimilar
LabelingInformationSingle-use, powder freeglove size, quantity,Nitrile ExaminationGloves, Non SterileSingle-use, powder free glovesize, quantity, NitrileExamination Gloves, NonSterileSame
510(K) SUMMARYChemotherapy DrugSee below comparison
Permeation ClaimtableSee below comparison tableSame
Dimensions and Performance Comparison Table:
TechnologicalProposed DevicePredicate Device
CharacteristicsK211604K182600Remark
LengthMinimum 230mmMinimum 230mmSame
Palm Width (size) (mm)
XS70±1070±10Same
S80±1080±10Same
M95±1095±10Same
L110±10110±10Same
XL120±10120±10Same
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same
PalmMinimum 0.05Minimum 0.05Same
Tensile Strength,
Before Aging14MPa, min14MPa, minSame
Ultimate Elongation,
Before Aging500%, min500%, minSame
Tensile Strength, After
Accelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation,
After Accelerated400%, min400%, minSame
Aging
In accordance with ASTM DIn accordance with ASTM D
Freedom from holes5151-19, following ASTM5151-19, following ASTMSame
D6319- 19, G-I, AQL 2.5D6319- 19, G-I, AQL 2.5
Powder-Content< 2 mg per glove< 2 mg per gloveSimilar
10993-10:2010 SkinUnder the conditions of theUnder the conditions of the
Irritation Studystudy,study,Same
not an irritantnot an irritant
10993-10:2010Under the conditions of theUnder the conditions of the
Maximizationstudy,not a sensitizerstudy,not a sensitizerSame
Sensitization Study
10993-5:2009Under the conditions of thisUnder the conditions of
Cytotoxicity Teststudy, the test article extractthis study, no cytotoxicSame
showed potential toxicitypotential
ISO 10993 Part 11Under the conditions of this//
Systemic toxicitystudy, there was no evidence ofsystemic toxicity.
Tested Chemotherapy Drug andConcentrationMinimum BDT (Minutes)Remark
Proposed DeviceK211604Predicate DeviceK182600
Bleomycin (15.0mg/ml)>240>240Same
Bortezomib (Velcade) 1mg/ml>240>240Same
Busulfan (6.0mg/ml)>240>240Same
Carboplatin, (10.0mg/ml)>240>240Same
Carmustine(BCNU) (3.3 mg/ml)7.011.0similar
Chloroquine, (50.0mg/ml)>240>240Same
Cisplatin (1mg/ml)>240>240Same
Cyclophosphamide (Cytoxan) (20mg/ml)>240>240Same
Cyclosporin, (100.0mg/ml)>240>240Same
Cytarabine, (100.0mg/ml)>240>240Same
Dacarbazine (10mg/ml)>240>240Same
Daunorubicin, HCL (5.0mg/ml)>240>240Same
Docetaxel, (10.0mg/ml)>240>240Same
Doxorubicin HCL (2mg/ml)>240>240Same
Epirubicin HCL (Ellence), (2.0mg/ml)>240>240Same
Etoposide (20mg/ml)>240>240Same
Fludarabine, (25.0mg/ml)>240>240Same
Fluorouracil (50mg/ml)>240>240Same
Gemcitabine (38.0mg/ml)>240>240Same
Idarubicin HCL (1.0mg/ml)>240>240Same
Ifosfamide, (50.0mg/ml)>240>240Same
Irinotecan, (20.0mg/ml)>240>240Same
Mechlorethamine HCI, (1.0mg/ml)>240>240Same
Melphalan, (5.0mg/ml)>240>240Same
Methotrexate (25mg/ml)>240>240Same
Mitomycin C, (0.5mg/ml)>240>240Same
Mitoxantrone, (2.0mg/ml)>240>240Same
Oxaliplatin, (5.0mg/ml)>240>240Same
Paclitaxel (Taxol) (6mg/ml)>240>240Same
Paraplatin, (10.0mg/ml)>240>240Same
Retrovir, (10.0mg/ml)>240>240Same
Rituximab, (10.0mg/ml)>240>240Same
Thio Tepa (10mg/ml)23.028.8similar
Topotecan HCL, (1.0mg/ml)>240>240Same
Trisonex, (1.0mg/ml)>240>240Same
Vincristine, (1.0mg/ml)>240>240Same

General Comparison Table:

{7}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County,

Chemotherapy Permeation Comparison Claim:

{8}------------------------------------------------

Coal Bed Gas Ino Industrial Zone, Qu'e Town, Daning County,

510(K) SUMMARY

{9}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

510(K) SUMMARY 7. Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests For ● Irritation And Skin Sensitization.
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: Tests For In Vitro . Cytotoxicity
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for ● systemic toxicity
  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on ● Medical Gloves
  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ●
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ASTM D6978-05 (Reapproved 2019), Assessment of Reissuance of Medical Gloves ● to Permeation by Chemotherapy Drugs.
Test methodPurposeAcceptance CriteriaResults
ASTM D6319- 19LengthMinimum 230mmAll size ≥230
ASTM D6319- 19Palm WidthXS: 70 $\pm$ 10mm77-78mm
S: 80 $\pm$ 10mm86-88 mm
M:95 $\pm$ 10mm96 -98mm
L:110 $\pm$ 10mm108-110 mm
XL: 120 $\pm$ 10mm116-117 mm
ASTM D6319- 19ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)≥0.05mm
ASTM D6319-19ASTN D412-16Tensile Strength, BeforeAging14MPa, min≥14 MPa
ASTM D6319-19ASTN D412-16Tensile Strength, AfterAccelerated Aging14MPa, min≥14 MPa
ASTM D6319-19ASTN D412-16Ultimate Elongation,Before Aging500%, min≥500%
ASTM D6319-19ASTN D412-16Ultimate Elongation,After Accelerated Aging400%, min≥400%
ASTM D 5151-19Freedom from holesG-I, AQL 2.5Meet and above AQL2.5requirements
ASTM D 6124-06(2017)Powder-Content≤2 mg per glove≤2 mg, meet requirements

{10}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province

ISO 10993-10:2010Skin Irritation StudyUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritant
ISO 10993-10:2010MaximizationSensitization StudyUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
ISO 10993-5:2009Cytotoxicity TestUnder the conditions of this study, the test article extract showed potential toxicityUnder the conditions of this study, the test article extract showed potential toxicity
ISO 10993-11:2017Systemic toxicityUnder the conditions of this study, there was no evidence of systemic toxicity.Under the conditions of this study, there was no evidence of systemic toxicity.
ASTM D6978-05 (2019)Tested for Use withChemotherapy Drugs/Details refer to the above 5.0 result

510(K) SUMMARY

8. Clinical Performance Data

N/A

9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.