K Number
K202977
Date Cleared
2022-04-29

(576 days)

Product Code
Regulation Number
866.3393
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Alinity m STI Assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System for the direct, qualitative detection and differentiation of ribosomal RNA from Chlamydia trachomatis (CT), DNA from Neisseria gonorrhoeae (NG), ribosomal RNA from Trichomonas vaginalis (TV), and ribosomal RNA from Mycoplasma genitalium (MG), to aid in the diagnosis of disease(s) caused by infection from these organisms. The assay may be used to test the following specimens from symptomatic and asymptomatic individuals for the following analytes:

CT: vaginal swabs (clinician-collected and self-collected in a clinical setting), endocervical swabs, male urine, oropharyngeal swabs, and rectal swabs

NG: vaginal swabs (clinician-collected and self-collected in a clinical setting). endocervical swabs, gynecological specimens in ThinPrep PreservCyt Solution, male urine, oropharyngeal swabs, and rectal swabs

TV: vaginal swabs (clinician-collected and self-collected in a clinical setting), endocervical swabs, gynecological specimens in ThinPrep PreservCyt Solution, female urine, and male urine

MG: vaginal swabs (clinician-collected and self-collected in a clinical setting), endocervical swabs, and male urine

A vaginal swab (self-collected or clinician-collected) is the preferred specimen type for MG testing in females due to higher clinical sensitivity compared to endocervical swabs. If endocervical swab specimens test negative, testing with a vaginal swab may be indicated if M. genitalium infection is suspected.

Device Description

The Alinity m STI Assay is a real time polymerase chain reaction (PCR) assay for the amplification and detection of Chlamydia trachomatis (CT) ribosomal RNA sequences, Neisseria gonorrhea (NG) genomic DNA sequences, Trichomonas vaginalis (TV) ribosomal RNA sequences, Mycoplasma genitalium (MG) ribosomal RNA sequences, and human genomic DNA sequences. The assay can be used with endocervical swab specimens, vaginal swab specimens, male and female urine specimens, gynecological specimens in ThinPrep® PreservCyt® Solution, oropharyngeal swab specimens, and rectal swab specimens. Endocervical swab, vaginal swab, oropharyngeal swab, rectal swab and urine specimens are collected with the Alinity m multi-Collect Specimen Collection Kit. PreservCyt Solution specimens are transferred to an Alinity m Transport Tube for processing on the Alinity m System.

The steps of the Alinity m STI Assay consist of sample preparation, RT-PCR assembly, amplification/detection, and result calculation and reporting. All stages of the Alinity m STI Assay procedure are executed automatically by the Alinity m System. No intermediate processing or transfer steps are performed by the user. The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m STI Assay in parallel with other Alinity m assays on the same instrument.

The Alinity m STI Assay requires two separate assay specific kits as follows:

  • . Alinity m STI AMP Kit, List No. 09N17-095 consisting of multi-well amplification plates containing lyophilized, unit-dose PCR amplification/detection reagents and multi-well activator plates containing liquid, unit-dose activation reagents (MgCl2, TMAC, and KCl). The intended storage condition for the Alinity m STI AMP Kit is 2℃ to 8℃.
  • . Alinity m STI CTRL Kit, List No. 09N17-085 consisting of negative controls and positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m STI Control Kit is -15°C to -25°C.

Nucleic acids from specimens are extracted automatically on-board the Alinity m System using the Alinity m Sample Prep Kit 1, Alinity m Lysis Solution, Alinity m Ethanol Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash and elution. The resulting purified nucleic acids are then combined with the liquid unit-dose activator reagent, lyophilized unit-dose Alinity m STI amplification reagents, and Alinity m Vapor Barrier Solution, and transferred by the instrument to an amplification/detection module for reverse transcription, PCR amplification, and real-time fluorescence detection.

Assay controls are tested at or above an established minimum frequency of every 24 hours to help ensure that instrument and reagent performance remain satisfactory. During each control event, a negative control and a positive control are processed through sample preparation and RT-PCR procedures that are identical to those used for specimens. Assay controls are used to demonstrate proper sample processing and assay validity. The controls do not indicate if bacterial cells have been adequately lysed.

The Alinity m STI amplification reagents include primers and a probe that amplify and detect the single copy human gene, ß-globin. Amplification and detection of the ß-globin gene demonstrates proper sample processing and adequate sample input. In addition, an exogenous internal control (containing an armored RNA sequence) is included in the lyophilized Alinity m STI amplification reagents to assess amplification efficiency and to confirm that no PCR inhibitors are present in the sample. The cellular control and internal control are both used to demonstrate assay validity.

The Alinity m STI Assay also utilizes the following accessories:

  • . Alinity m STI Assay Application Specification File, List No. 09N17-03A
  • . Alinity m System and System Software, List No. 08N53-002
  • Alinity m Sample Prep Kit 1, List No. 09N18-001 .
  • Alinity m multi-Collect Specimen Collection Kit, List No. 09N19-010 .
  • . Alinity m Tubes and Caps, List No. 09N49:
    • Alinity m Transport Tubes Pierceable Capped, List No. 09N49-010 .
    • . Alinity m Transport Tube, List No. 09N49-011
    • Alinity m Pierceable Cap, List No. 09N49-012 .
  • Alinity m System Solutions, List No. 09N20: .
    • Alinity m Lysis Solution, List No. 09N20-001 .
    • Alinity m Ethanol Solution, List No. 09N20-002 .
    • Alinity m Diluent Solution, List No. 09N20-003
    • Alinity m Vapor Barrier Solution, List No. 09N20-004 •
AI/ML Overview

The Abbott Alinity m STI Assay is an in vitro polymerase chain reaction (PCR) assay used with the automated Alinity m System for the direct, qualitative detection and differentiation of ribosomal RNA from Chlamydia trachomatis (CT), DNA from Neisseria gonorrhoeae (NG), ribosomal RNA from Trichomonas vaginalis (TV), and ribosomal RNA from Mycoplasma genitalium (MG), to aid in the diagnosis of sexually transmitted infections.

The acceptance criteria and the study results are detailed below:

1. Table of Acceptance Criteria and Reported Device Performance

The device performance is reported as Sensitivity (Positive Percent Agreement or PPA) and Specificity (Negative Percent Agreement or NPA). The document does not explicitly state pre-defined acceptance criteria (e.g., a specific threshold like "Sensitivity must be >= X%"). However, the reported performance values are the outcome of the clinical trials conducted to demonstrate the device's effectiveness.

Urogenital Specimens

TargetSpecimen TypeSymptom StatusSensitivity (%) Estimate (95% CI)Specificity (%) Estimate (95% CI)
CTClinician-collected Vaginal SwabAll98.0 (95.1, 99.2)99.2 (98.8, 99.4)
CTSelf-collected Vaginal SwabAll98.5 (95.8, 99.5)99.3 (98.9, 99.5)
CTEndocervical SwabAll94.5 (90.4, 96.9)99.4 (99.0, 99.6)
CTMale UrineAll97.2 (94.6, 98.6)99.5 (99.2, 99.7)
NGClinician-collected Vaginal SwabAll100.0 (91.4, 100.0)99.8 (99.5, 99.9)
NGSelf-collected Vaginal SwabAll100.0 (91.4, 100.0)99.7 (99.4, 99.8)
NGEndocervical SwabAll92.5 (80.1, 97.4)99.8 (99.5, 99.9)
NGPreservCytAll94.4 (81.9, 98.5)100.0 (99.8, 100.0)
NGMale UrineAll100.0 (96.4, 100.0)99.9 (99.7, 99.9)
TVClinician-collected Vaginal SwabAll99.7 (98.2, 99.9)97.2 (96.6, 97.8)
TVSelf-collected Vaginal SwabAll99.4 (97.7, 99.8)97.8 (97.2, 98.3)
TVEndocervical SwabAll97.7 (95.4, 98.9)96.9 (96.2, 97.5)
TVPreservCytAll95.6 (92.4, 97.4)99.4 (99.0, 99.7)
TVMale UrineAll98.7 (93.2, 99.8)99.2 (98.9, 99.5)
MGClinician-collected Vaginal SwabAll98.1 (95.5, 99.2)99.2 (98.8, 99.5)
MGSelf-collected Vaginal SwabAll95.4 (92.1, 97.3)98.6 (98.1, 99.0)
MGEndocervical SwabAll82.8 (77.6, 87.0)99.2 (98.7, 99.4)
MGMale UrineAll98.1 (95.3, 99.3)97.5 (96.9, 98.0)

Extragenital Specimens

TargetSpecimen TypeSymptom StatusSensitivity (%) Estimate (95% CI)Specificity (%) Estimate (95% CI)
CTOropharyngealAll93.3 (78.7, 98.2)99.9 (99.7, 100.0)
CTRectalAll94.5 (89.6, 97.2)99.6 (99.2, 99.8)
NGOropharyngealAll95.2 (89.3, 98.2)99.3 (98.9, 99.6)
NGRectalAll97.1 (91.9, 99.0)99.5 (99.1, 99.7)

2. Sample Size Used for the Test Set and Data Provenance

  • Urogenital Specimens: A total of 7,099 male and female subjects were enrolled for the urogenital clinical study. Specimens were collected across 33 geographically diverse sites in the United States, including STI clinics, primary care offices, and gynecology practices. This was a prospective study. Data provenance is United States, from prospective collection.
    • Number of results used in analysis for each analyte:
      • CT: 12,903
      • NG: 15,655
      • TV: 18,843
      • MG: 12,829
  • Extragenital Specimens: A total of 2,373 male and female subjects were enrolled. Specimens were previously collected and archived. Data provenance is United States, from archived specimens (retrospective).
    • Number of results used in analysis for each analyte:
      • CT (oropharyngeal): 2,316
      • CT (rectal): 2,053
      • NG (oropharyngeal): 2,312
      • NG (rectal): 2,049

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the "number of experts" or their "qualifications" involved in establishing the ground truth. Instead, the ground truth was established using comparator assays, which are commercially available nucleic acid amplification tests (NAATs) and, in some cases, culture. The results from these comparator assays were combined to derive a Patient Infected Status (PIS) or Composite Comparator (CC).

4. Adjudication Method for the Test Set

  • Urogenital Specimens (PIS):
    • CT or NG (Female): A minimum of 2 positive results (at least 1 from each comparator NAAT) for infection, or at least 1 comparator NAAT reported negative results for all sample types for not infected.
    • TV or MG (Female): First 2 swab comparator NAAT results both positive, or 2 of 3 swab comparator NAAT results positive (if 3rd NAAT was a tie-breaker) for infection. First 2 swab comparator NAAT results both negative, or 2 of 3 swab comparator NAAT results negative (if 3rd NAAT was a tie-breaker) for not infected.
    • CT, NG, TV, or MG (Male): A minimum of two comparator positive results for infection. If the comparator TV culture assay result was positive, the subject was categorized as infected for TV regardless of NAAT results. Two or more comparator NAAT results negative for not infected (for CT, NG, MG). For TV, negative culture AND one or more negative comparator NAATs for not infected.
  • Extragenital Specimens (CC): A specimen was categorized as infected (for CT or NG) if a minimum of 2 comparator positive results were reported. It was categorized as not infected if a minimum of 2 comparator negative results was reported.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The study compares the Alinity m STI Assay's performance against a composite ground truth derived from multiple established comparator assays, not directly evaluating human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, this was a standalone performance study. The Alinity m STI Assay is an automated PCR assay, and its results are directly compared to the established ground truth without involving human interpretation or modifications of its output in the primary performance analysis.

7. The Type of Ground Truth Used

The ground truth used was a Composite Comparator / Patient Infected Status (PIS/CC), derived from the combined results of multiple commercially available and clinically cleared comparator nucleic acid amplification tests (NAATs) and, for male TV, culture results.

8. The Sample Size for the Training Set

The document does not explicitly provide the sample size for the training set for the Alinity m STI Assay. The provided performance data (Sensitivity and Specificity tables) are from the validation (test) sets.

9. How the Ground Truth for the Training Set Was Established

Since the document does not provide information about a separate training set, it is assumed that the analytical studies and the design of the assay would have utilized reference materials and potentially early clinical samples for optimization and establishment of analytical performance characteristics (like LoD, inclusivity, cross-reactivity). However, specific details on how ground truth was established for a training set are not available in this document. The clinical studies described are for validation/testing, with ground truth established by comparator assays as detailed in point 4.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 29, 2022

Abbott Molecular, Inc. Stacy Ferguson Associate Director Regulatory Affairs 1300 E. Touhy Des Plains, Illinois 60018

Re: K202977

Trade/Device Name: Alinity m STI Assay Regulation Number: 21 CFR 866.3393 Regulation Name: Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) Regulatory Class: Class II Product Code: QEP, MKZ, LSL, OUY, LIO, OOI Dated: February 11, 2022 Received: February 14, 2022

Dear Stacy Ferguson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Himani Bisht, Ph.D. Assistant Director Viral Respiratory and HPV Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

Indications for Use (Describe)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary

Table of Contents

Page
5.0510(k) Summary
5.1Submitter
5.2Device Information
5.3Predicate Devices
5.4Device Description
5.5Intended Use
5.6Similarities and Differences to Predicate Devices
5.7Performance Data
5.7.1Specific Performance Characteristics
5.7.1.1 Analytical Sensitivity
5.7.1.2 Inclusivity
5.7.1.3 Evaluation of Potential Cross Reacting Microorganisms 14
5.7.1.4 Evaluation of Potential Interfering Substances 19
5.7.1.5 Competitive Interference Study
5.7.1.6 Within Laboratory Precision
5.7.1.7 Carryover
5.7.1.8 Reproducibility Study
5.7.2Clinical Performance
5.7.2.1 Clinical Study Results - Urogenital Specimens 35
5.7.2.2 Clinical Study Results - Extragenital Specimens 65
5.8Conclusions Drawn from the Studies

{4}------------------------------------------------

5.0 510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR Section 807.92(c).

Submitter 5.1

Applicants Name and Address:Abbott Molecular Inc.1300 E. Touhy AveAbbott Molecular Inc.
Contact Person:Stacy FergusonDirector Regulatory AffairsAbbott Molecular, Inc.1300 E. Touhy AvenueDes Plaines, IL 60018Phone: 224-361-7449Fax: 224-361-7269
Date Prepared:April 14, 2022

{5}------------------------------------------------

Trade NameRegulation NameProductCodeRegulationNumberClass
Alinity m STI AssayDevice to detect nucleic acids fromnon-viral microorganism(s) causingsexually transmitted infections andassociated resistance marker(s)QEP866.3393II

5.2 Device Information

5.3 Predicate Devices

Device NamePredicate Device510(k)Cleared
Alinity m STI AssayAbbott RealTime CT/NG(Primary Predicate for CT/NG)K1403545/9/14
Aptima Combo 2 AssayAptima Combo 2 Assay(Primary Predicate for CT/NG)K1905155/23/19
BD Max CT/GC/TVBD Max CT/GC/TV(Primary Predicate for TV)K1515899/6/16
Aptima Mycoplasma genitalium AssayAptima Mycoplasma genitalium Assay(Primary Predicate for MG)DEN1800471/23/19

{6}------------------------------------------------

5.4 Device Description

The Alinity m STI Assay is a real time polymerase chain reaction (PCR) assay for the amplification and detection of Chlamydia trachomatis (CT) ribosomal RNA sequences, Neisseria gonorrhea (NG) genomic DNA sequences, Trichomonas vaginalis (TV) ribosomal RNA sequences, Mycoplasma genitalium (MG) ribosomal RNA sequences, and human genomic DNA sequences. The assay can be used with endocervical swab specimens, vaginal swab specimens, male and female urine specimens, gynecological specimens in ThinPrep® PreservCyt® Solution, oropharyngeal swab specimens, and rectal swab specimens. Endocervical swab, vaginal swab, oropharyngeal swab, rectal swab and urine specimens are collected with the Alinity m multi-Collect Specimen Collection Kit. PreservCyt Solution specimens are transferred to an Alinity m Transport Tube for processing on the Alinity m System.

The steps of the Alinity m STI Assay consist of sample preparation, RT-PCR assembly, amplification/detection, and result calculation and reporting. All stages of the Alinity m STI Assay procedure are executed automatically by the Alinity m System. No intermediate processing or transfer steps are performed by the user. The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m STI Assay in parallel with other Alinity m assays on the same instrument.

The Alinity m STI Assay requires two separate assay specific kits as follows:

  • . Alinity m STI AMP Kit, List No. 09N17-095 consisting of multi-well amplification plates containing lyophilized, unit-dose PCR amplification/detection reagents and multi-well activator plates containing liquid, unit-dose activation reagents (MgCl2, TMAC, and KCl). The intended storage condition for the Alinity m STI AMP Kit is 2℃ to 8℃.
  • . Alinity m STI CTRL Kit, List No. 09N17-085 consisting of negative controls and positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m STI Control Kit is -15°C to -25°C.

{7}------------------------------------------------

Nucleic acids from specimens are extracted automatically on-board the Alinity m System using the Alinity m Sample Prep Kit 1, Alinity m Lysis Solution, Alinity m Ethanol Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash and elution. The resulting purified nucleic acids are then combined with the liquid unit-dose activator reagent, lyophilized unit-dose Alinity m STI amplification reagents, and Alinity m Vapor Barrier Solution, and transferred by the instrument to an amplification/detection module for reverse transcription, PCR amplification, and real-time fluorescence detection.

Assay controls are tested at or above an established minimum frequency of every 24 hours to help ensure that instrument and reagent performance remain satisfactory. During each control event, a negative control and a positive control are processed through sample preparation and RT-PCR procedures that are identical to those used for specimens. Assay controls are used to demonstrate proper sample processing and assay validity. The controls do not indicate if bacterial cells have been adequately lysed.

The Alinity m STI amplification reagents include primers and a probe that amplify and detect the single copy human gene, ß-globin. Amplification and detection of the ß-globin gene demonstrates proper sample processing and adequate sample input. In addition, an exogenous internal control (containing an armored RNA sequence) is included in the lyophilized Alinity m STI amplification reagents to assess amplification efficiency and to confirm that no PCR inhibitors are present in the sample. The cellular control and internal control are both used to demonstrate assay validity.

{8}------------------------------------------------

The Alinity m STI Assay also utilizes the following accessories:

  • . Alinity m STI Assay Application Specification File, List No. 09N17-03A
  • . Alinity m System and System Software, List No. 08N53-002
  • Alinity m Sample Prep Kit 1, List No. 09N18-001 .
  • Alinity m multi-Collect Specimen Collection Kit, List No. 09N19-010 .
  • . Alinity m Tubes and Caps, List No. 09N49:
    • Alinity m Transport Tubes Pierceable Capped, List No. 09N49-010 .
    • . Alinity m Transport Tube, List No. 09N49-011
    • Alinity m Pierceable Cap, List No. 09N49-012 .
  • Alinity m System Solutions, List No. 09N20: .
    • Alinity m Lysis Solution, List No. 09N20-001 .
    • Alinity m Ethanol Solution, List No. 09N20-002 .
    • Alinity m Diluent Solution, List No. 09N20-003
    • Alinity m Vapor Barrier Solution, List No. 09N20-004 •

{9}------------------------------------------------

5.5 Intended Use

Alinity m STI AMP Kit:

The Alinity m STI Assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System for the direct, qualitative detection and differentiation of ribosomal RNA from Chlamydia trachomatis (CT), DNA from Neisseria gonorrhoeae (NG), ribosomal RNA from Trichomonas vaginalis (TV), and ribosomal RNA from Mycoplasma genitalium (MG), to aid in the diagnosis of disease(s) caused by infection from these organisms. The assay may be used to test the following specimens from symptomatic and asymptomatic individuals for the following analytes:

CT: vaginal swabs (clinician-collected and self-collected in a clinical setting), endocervical swabs, male urine, oropharyngeal swabs, and rectal swabs

NG: vaginal swabs (clinician-collected and self-collected in a clinical setting). endocervical swabs, gynecological specimens in ThinPrep PreservCyt Solution, male urine, oropharyngeal swabs, and rectal swabs

TV: vaginal swabs (clinician-collected and self-collected in a clinical setting), endocervical swabs, gynecological specimens in ThinPrep PreservCyt Solution, female urine, and male urine

MG: vaginal swabs (clinician-collected and self-collected in a clinical setting), endocervical swabs, and male urine

A vaginal swab (self-collected or clinician-collected) is the preferred specimen type for MG testing in females due to higher clinical sensitivity compared to endocervical swabs. If endocervical swab specimens test negative, testing with a vaginal swab may be indicated if M. genitalium infection is suspected.

{10}------------------------------------------------

Alinity m multi-Collect Specimen Collection Kit:

The Alinity m multi-Collect Specimen Collection Kit is intended for the collection and transportation of male and female urine specimens, endocervical swab specimens, vaginal swab specimens, oropharyngeal swab specimens, and rectal swab specimens to stabilize nucleic acid for testing with the Alinity m STI Assay. Refer to the Alinity m STI Assay package insert for additional information.

The Alinity m multi-Collect Specimen Collection Kit is not intended for home use.

5.6 Similarities and Differences to Predicate Devices

The primary functional components of the Alinity m STI Assay are substantially equivalent to other legally marketed nucleic acid amplification tests (NAAT) intended for the qualitative detection of Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), Trichomonas vaginalis (TV), and Mycoplasma genitalium (MG).

The Alinity m STI Assay has the same general intended use as the predicate devices. Although there are some technological differences between the Alinity m STI Assay and the predicate devices, these differences do not raise new types of safety or effectiveness questions.

These devices are similar in that they are designed to prepare nucleic acids for amplification, amplify specific CT, NG, TV, and MG sequences, detect the amplified products, and report qualitative results.

The primary similarities and differences between the Alinity m STI Assay and the NAAT predicate devices are shown in Table 1. The primary similarities and differences between the Alinity m multi-Collect Specimen Collection Kit and the predicate device are shown in .

{11}------------------------------------------------

Table 1. Similarities and Differences Between Alinity m STI Assay and Nucleic Acid Amplification Tests-Predicate Devices
FeatureCurrent ApplicationPredicate Devices
Alinity m STI AssayAbbott RealTime CT/NG(Predicate for CT/NG)Aptima Combo 2 Assay(Predicate for CT/NG)Becton DickinsonMAX CT/GC/TV(Predicate for TV)Aptima Mycoplasma genitaliumAssay(Predicate for MG)
Intended UseThe Alinity m STI Assay is an invitro polymerase chain reaction(PCR) assay for use with theautomated Alinity m System forthe direct, qualitative detectionand differentiation of ribosomalRNA from Chlamydiatrachomatis (CT), DNA fromNeisseria gonorrhoeae (NG),ribosomal RNA fromTrichomonas vaginalis (TV), andribosomal RNA fromMycoplasma genitalium (MG) toaid in the diagnosis of urogenitaldisease(s) caused by infectionfrom these organisms. The assaymay be used to test the followingspecimens from symptomatic andasymptomatic individuals for thefollowing analytes:CT: vaginal swabs (clinician-collected and self-collected in aclinical settings), endocervicalswabs, male urine, oropharyngealswabs, and rectal swabsNG: vaginal swabs (clinician-collected and self-collected in aclinical settings), endocervicalswabs, gynecological specimensin ThinPrep PreservCyt Solution,male urine, oropharyngeal swabs,and rectal swabsTV: vaginal swabs (clinician-collected and self-collected in aThe Abbott RealTime CT/NG assayis an in vitro polymerase chainreaction (PCR) assay for the direct,qualitative detection of the plasmidDNA of Chlamydia trachomatisand the genomic DNA of Neisseriagonorrhoeae. The assay may beused to test the followingspecimens from symptomaticindividuals: female endocervicalswab, clinician-collected vaginalswab, patient-collected vaginalswab specimens; male urethralswab specimens; and female andmale urine specimens. The assaymay be used to test the followingspecimens from asymptomaticindividuals: clinician-collectedvaginal swab and patient-collectedvaginal swab specimens; femaleand male urine specimens.The Aptima Combo 2 Assay is atarget amplification nucleic acidprobe test that utilizes target capturefor the in vitro qualitative detectionand differentiation of ribosomalRNA (rRNA) from Chlamydiatrachomatis (CT) and/or Neisseriagonorrhoeae (GC) to aid in thediagnosis of chlamydial and/orgonococcal disease using the PantherSystem as specified.On the Panther System, the assaymay be used to test the followingspecimens from symptomatic andasymptomatic individuals: clinician-collected endocervical, vaginal,throat, rectal, and male urethral swabspecimens, clinician-collectedgynecological specimens collected inthe PreservCyt Solution, patient-collected vaginal swab specimens,1and female and male urinespecimens.Patient-collected vaginal swabspecimens are an option forscreening women when a pelvicexam is not otherwise indicated. TheAptima Multitest Swab SpecimenCollection kit has not been evaluatedfor home use.The BD MAX CT/GC/TV assay, asperformed using the BD MAXSystem incorporates automatedDNA extraction and real-timepolymerase chain reaction (PCR)for the direct, qualitative detectionof DNA from Chlamydiatrachomatis (CT), Neisseriagonorrhoeae (GC) and/orTrichomonas vaginalis (TV). Theassay may be used for detection ofCT and/or GC DNA in male urinespecimens, and the detection of CT,GC and/or TV DNA in femaleurine specimens, clinician-collectedfemale endocervical swabspecimens and patient-collectedvaginal swab specimens (in aclinical setting). The assay isindicated for use to aid in thediagnosis of chlamydial urogenitaldisease, gonococcal urogenitaldisease and/or trichomoniasis inasymptomatic and symptomaticindividuals.The Aptima Mycoplasma genitaliumassay is an in vitro nucleic acidamplification test (NAAT) for thequalitative detection of ribosomalRNA (rRNA) from Mycoplasmagenitalium on the fully automatedPanther system. It is intended for useas an aid in the diagnosis of M.genitalium urogenital infections inmale and female patients suspected ofM. genitalium infection.The assay may be used to test thefollowing specimens: clinician-collected and self-collected vaginalswabs (in a clinical setting), clinician-collected endocervical swabs, femaleand male urine, clinician-collectedmale urethral swabs, and self-collected penile meatal swabs (in aclinical setting).For females, a vaginal swab is thepreferred specimen type due to higherclinical sensitivity for detecting M.genitalium than other specimen types;however, female urine or clinician-collected endocervical swabs may beused as alternative specimens whenvaginal swab specimens are notavailable. If female urine or clinician-collected endocervical swabspecimens test negative, testing with avaginal swab may be indicated, if M.genitalium infection is suspected.
FeatureCurrent ApplicationPredicate Devices
Alinity m STI AssayAbbott RealTime CT/NG(Predicate for CT/NG)Aptima Combo 2 Assay(Predicate for CT/NG)Becton DickinsonMAX CT/GC/TV(Predicate for TV)Aptima Mycoplasma genitaliumAssay(Predicate for MG)
clinical settings), endocervicalswabs, gynecological specimensin ThinPrep PreservCyt Solution,female urine, and male urine
MG : vaginal swabs (cliniciancollected and self-collected in aclinical settings), endocervicalswabs, and male urine
A vaginal swab (self-collected orclinician-collected) is thepreferred specimen type for MGtesting in females due to higherclinical sensitivity compared toendocervical swabs. Ifendocervical swab specimens testnegative, testing with a vaginalswab may be indicated if M.genitalium infection is suspected.
Table 1. Similarities and Differences Between Alinity m STI Assay and Nucleic Acid Amplification Tests-Predicate Devices
FeatureCurrent ApplicationPredicate Devices
Alinity m STI AssayAbbott RealTime CT/NG(Predicate for CT/NG)Aptima Combo 2 Assay(Predicate for CT/NG)Becton DickinsonMAX CT/GC/TV(Predicate for TV)Aptima Mycoplasma genitaliumAssay(Predicate for MG)
Assay TypeQualitativeQualitativeQualitativeQualitativeQualitative
Assay TargetsCT ribosomal RNACT cryptic plasmid DNACT ribosomal RNACT genomic DNANA
NG genomic DNANG genomic DNANG ribosomal RNANG genomic DNANA
TV ribosomal RNANANATV genomic DNANA
MG ribosomal RNANANANAMG ribosomal RNA
Specimen TypesEndocervical swabEndocervical swabEndocervical swabEndocervical swabEndocervical swab
Self-collected vaginal swabSelf-collected vaginal swabSelf-collected vaginal swabSelf-collected vaginal swabSelf-collected vaginal swab
Clinician-collected vaginalswabClinician-collected vaginalswabClinician-collected vaginal swabNAClinician-collected vaginal swab
Male and female urineMale and female urineMale and female urineMale urine (for CT and NGonly ) and female urine (CT,NG and TV)Male and female urine
NAMale urethral swabsMale urethral swabNAMale urethral swab
Gynecological specimens inPreservCyt SolutionNAGynecological specimens inPreservCyt SolutionNANA
NANANANASelf-collected penile meatalswabs (in a clinical setting)
Oropharyngeal swabNAOropharyngeal swabNANA
Rectal swabNARectal swabNANA
Table 1. Similarities and Differences Between Alinity m STI Assay and Nucleic Acid Amplification Tests-Predicate Devices
FeatureCurrent ApplicationPredicate Devices
Alinity m STI AssayAbbott RealTime CT/NG(Predicate for CT/NG)Aptima Combo 2 Assay(Predicate for CT/NG)Becton DickinsonMAX CT/GC/TV(Predicate for TV)Aptima Mycoplasma genitaliumAssay(Predicate for MG)
SamplePreparationProcedureAutomatedAutomatedAutomatedAutomatedAutomated
AmplificationTechnologyReal-Time Polymerase ChainReaction (PCR)Real-Time PCRTarget Capture (TC),Transcription-mediatedAmplification (TMA),Dual Kinetic Assay (DKA)Real-Time PCRTarget Capture (TC),Transcription-mediatedAmplification (TMA),Hybridization Protection Assay(HPA)
Assay ControlsNegative Control Positive Control Internal Control (IC) Cellular Control (CC)Negative Control Cutoff Control Internal Control (IC)CT Positive/NG Negative Control CT Negative/NG Positive ControlNo Negative Control No Positive Control Sample Processing ControlNo Negative Control No Positive Control Internal Control (IC)

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

5.7 Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Specific Performance Characteristics 5.7.1

5.7.1.1 Analytical Sensitivity

Urogenital Specimens

The limit of detection (LoD) for urogenital specimens was determined by testing dilutions of CT, NG, TV, and MG organisms in pooled negative vaginal swab matrix, pooled negative gynecological PreservCyt matrix, and pooled negative urine matrix. Two different strains were evaluated for each organism in each matrix: serovars D and E for CT (ATCC), strains Z437 and Z433 for NG (Zeptometrix), strains 30001 (ATCC) and MTZ (metronidazole-resistant, Zeptometrix) for TV, and strains SEA-1 and MEGA 216 (azithromycin-resistant) for MG (non-commercial source). For each strain and matrix, a minimum of 6 target levels were evaluated in at least 20 replicates using each of 2 lots. Probit analysis was performed to estimate the LoD for each strain and matrix for each lot. In cases where the probit model did not fit the data, the LoD was determined for each lot to be the target concentration with a detection rate 95% or greater.

The LoD claim for the Alinity m STI Assay is as follows for each of the organisms in urogenital specimen types:

  • CT: 17.0 Elementary Bodies (EB)/mL .
  • NG: 7.5 Colony Forming Units (CFU)/mL .
  • . TV: 0.1 TV/mL
  • . MG: 165 Genome Equivalents (GE)/mL

Extragenital Specimens

The LoD for extragenital specimens was determined by testing dilutions of CT and NG organisms in pooled negative oropharyngeal swab matrix and pooled negative rectal

{16}------------------------------------------------

swab matrix. Two different strains were evaluated for each organism in each matrix: serovars D and E for CT (ATCC) and strains Z437 and Z433 (Zeptometrix) for NG. For each strain and matrix, a minimum of 3 target levels were evaluated in 20 replicates using one reagent lot. The LoD level was determined to be the concentration with a detection rate of 95% or greater where the target level 2-fold below was less than 95%.

The LoD claim for the Alinity m STI Assay is as follows for each of the organisms in extragenital specimen types:

  • . CT: 17.0 EB/mL
  • . NG: 7.5 CFU/mL

5.7.1.2 Inclusivity

The analytical sensitivity claim of the Alinity m STI Assay was confirmed by testing at least 20 replicates of the following additional strains:

  • . CT: Serovars A, B, Ba, C, F, G, H, I, J, K, L1, L2, L3, and E nvCT at 17.0 EB/mL or less. All serovars were detected at ≥95.0% across swab and urine matrices.
  • . NG: 28 additional isolates at 7.5 CFU/mL. All isolates were detected at ≥95.0% across swab, urine, and PreservCyt matrices.
  • . TV: Strains Z070, Z158, and Z159 at 0.1 TV/mL. All strains were detected at ≥ 95.0% across swab, urine, and PreservCyt matrices.
  • MG: Strains MEGA 601, 10008, MEGA 1256, MEGA 1272, MEGA 1404 and . SEA-2 at 165 GE/mL. All strains were detected at ≥ 95.0% across swab and urine matrices.

5.7.1.3 Evaluation of Potential Cross Reacting Microorganisms

Urogenital Specimens

A total of 71 potential cross reacting microorganisms that are phylogenetically related to CT, NG, TV, or MG or that may be found in the urogenital tract were evaluated with the Alinity m STI Assay (Table 2). The microorganisms were tested at 105 units/mL for

{17}------------------------------------------------

viruses and eukaryotes and 106 units/mL for bacteria. The unit of measure was specific to each microorganism. No cross reactivity was observed for CT, NG, TV, or MG in the presence of these microorganisms.

A subset of the microorganisms closely related to the STI analytes (asterisk in Table 2) was also assessed in CT, NG, TV, and MG positive samples. The positive samples contained CT, NG, TV, and MG organisms at 2 times the claimed LoD. The potential cross reacting microorganisms were tested at 105 units/mL for viruses and eukaryotes, and 106 units/mL for bacteria. All positive samples reported positive results for CT, NG, TV, and MG in the presence of these microorganisms.

{18}------------------------------------------------

Table 2. Urogenital Microorganisms

Acinetobacter lwoffi
Actinomyces israelii
Atopobium vaginae
Bacteroides fragilis
Bacteroides ureolyticus (Campylobacterureolyticus)
Bifidobacterium longum
Candida albicans
Candida glabrata
Candida parapsilosis
Candida tropicalis
Chlamydia pneumoniae*
Chlamydia psittaci*
Clostridium difficile
Clostridium perfringens
Corynebacterium genitalium
Corynebacterium xerosis
Cryptococcus neoformans
Dientamoeba fragilisa
Enterococcus faecalis
Enterobacter aerogenes
Enterobacter cloacae
Escherichia coli
Fusobacterium nucleatum
Gardnerella vaginalis
Haemophilus ducreyia
Herpes simplex virus I
Herpes simplex virus II
Human Immunodeficiency virus I
Human papilloma virus 16
Kingella denitrificans
Klebsiella oxytoca
Klebsiella pneumoniae
Lactobacillus acidophilus
Lactobacillus brevis
Lactobacillus jensenii
Lactobacillus lactis

Lactobacillus vaginalis Listeria moncytogenes Mobiluncus curtisii Mobiluncus mulieris Mycoplasma hominis* Mycoplasma pneumoniae Neisseria cinerea* Neisseria elongata* Neisseria flava* Neisseria flavescens* Neisseria lactamica* Neisseria mucosa* Neisseria meningitidis Serogroup A * Neisseria meningitidis Serogroup B* Neisseria meningitidis Serogroup C* Neisseria meningitidis Serogroup D* Neisseria meningitidis Serogroup W135 * Neisseria meningitidis Serogroup Y* Neisseria perflava* Neisseria polysaccharea* Neisseria sicca* Neisseria subflava* Pentatrichomonas hominis Peptostreptococcus anaerobius Prevotella bivia Propionibacterium acnes Proteus mirabilis Pseudomonas aeruginosa Staphylococcus aureus Staphylococcus epidermidis Streptococcus agalactiae Streptococcus pyogenes Tricomonas tenax Ureaplasma parvum Ureaplasma urealyticum

ªEvaluated using purified genomic DNA

Note: Microorganisms with asterisks were tested with both CT/NG/TV/MG negative and positive samples. Microorganisms without asterisks were tested with only CT/NG/TV/MG negative samples.

{19}------------------------------------------------

Extragenital Specimens

A total of 56 additional potential cross reacting microorganisms that may be found in oropharyngeal or rectal specimens were evaluated with the Alinity m STI Assay (Table 3). The microorganisms were tested at 105 units/mL for viruses and eukaryotes and 106 units/mL for bacteria. No cross reactivity was observed for CT or NG in the presence of these microorganisms.

A subset of the microorganisms (asterisk in Table 3) was also assessed in CT and NG positive samples. The positive samples contained CT and NG organisms at 2 times the claimed LoD. The potential cross reacting microorganisms were tested at 105 units/mL for viruses and eukaryotes, and 106 units/mL for bacteria. All positive samples reported positive results for CT and NG in the presence of these microorganisms.

{20}------------------------------------------------

Table 3. Extragenital Microorganisms
Adenovirus*Moraxella catarrhalis
Acinetobacter baumanniiMorganella morganii
Aggregatibacter actinomycetemcomitansMycoplasma pneumoniae
Anaerococcus prevotiiNorovirus
Arcanobacterium haemolyticumParvinomas micra
Bifidobacterium adolescentisPlesiomonas shigelloides
Bordetella pertussisPorphyromonas gingivalis
Campylobacter jejuni*Prevotella oralis
Campylobacter rectusProvidencia stuartii
Citrobacter freundiiRespiratory syncytial virus
Clostridium difficileSaccharomyces cerevisiae
Corynebacterium diphtheriae*Salmonella enterica
Entamoeba histolytica*Shigella flexneri
Enterobacter cloacaeShigella sonnei
Enterococcus faeciumStreptococcus anginosus
EnterovirusStreptococcus dysgalactiae*
Fusobacterium necrophorumStreptococcus mitis
Fusobacterium nucleatumStreptococcus mutans
Giardia lambliaStreptococcus pneumoniae
Haemophilus influenzaeStreptococcus pyogenes*
Helicobacter pyloriStreptococcus salivarius
Human Coronavirus*Streptococcus sanguinis
Human influenza virus ATannerella forsythia
Human influenza virus BTreponema denticola
Human metapneumovirusVeillonella parvula
Human rhinovirus 57*Vibrio cholerae
Klebsiella oxytocaVibrio parahaemolyticus
Listeria monocytogenesYersinia enterocolitica

Note: Microorganisms with asterisks were tested with both CT/NG negative and positive samples. Microorganisms without asterisks were tested with only CT/NG negative samples.

{21}------------------------------------------------

5.7.1.4 Evaluation of Potential Interfering Substances

Urogenital Specimens

The potential for interference in the Alinity m STI Assay was assessed with 35 substances that may be found in urine, vaginal swab, endocervical swab, and/or PreservCyt samples (Table 4). Substances were diluted into pooled vaginal swab, pooled gynecological PreservCyt, and/or pooled urine matrices. For each substance and matrix, both CT, NG, TV, and MG positive and negative samples were tested. The positive matrices contained CT, NG, TV, and MG organisms at 3 times the claimed assay LoD. Two different strains of each organism were used in this study (CT serovars D and E, NG strains Z433 and Z437, TV strains 30001 and MTZ, and MG strains SEA-1 and MEGA 216). No interference was observed in the presence of any of the substances at the concentrations shown in Table 4 for all positive and negative samples. Assay interference was observed in the presence of seminal fluid at concentrations greater than 3.0% in PreservCyt samples. Cycle number delays were observed for seminal fluid. yglobulin, and glucose in urine and seminal fluid. mucus, and Preparation H Hemorrhoidal Cream in PreservCyt, which could result in interference at lower target levels.

Table 4. Potentially Interfering Substances in Urogenital Specimens
SubstanceMatrixaTest Level
BloodU, S, P5.0% v/v
Norforms Deodorant SuppositoriesU, S, P0.25% w/v
ProgesteroneU, S, P20 ng/mL
Beta EstradiolU, S, P1.2 ng/mL
LeukocytesU, S, P106 cells/mL
MucusU0.2% v/v
S, Pb0.8% v/v
Seminal FluidUb, S5.0% v/v
Pb3.0% v/v
AzithromycinU12.0 µg/mL
DoxycyclineU31.2 µg/mL
AcetaminophenU196.5 µg/mL
AspirinU652.2 µg/mL
Vagisil Feminine PowderU0.25% w/v

Table 4 Potentially Interfering Substances in Urogenital Speciments

{22}------------------------------------------------

SubstanceMatrixaTest Level
AlbuminU60 mg/mL
γ-globulinUb60 mg/mL
GlucoseUb1.2 mg/mL
Acidic urineUpH 4.0
Alkaline urineUpH 9.0
BilirubinU72.5 µg/mL
Candida albicans(Urinary tract infection organism)U3x104 CFU/mL
Staphylococcus saprophyticus(Urinary tract infection organism)U3x104 CFU/mL
Escherichia coli(Urinary tract infection organism)U3x104 CFU/mL
IbuprofenU495.1 µg/mL
Phenazopyridine HydrochlorideU80 µg/mL
Clotrimazole Vaginal CreamS, P0.25% w/v
KY Jelly Personal LubricantS, P0.25% w/v
MetronidazoleS, P40.1 µg/mL
Miconazole-3S, P0.25% w/v
Monistat-1S, P0.25% w/v
Terconazole Vaginal CreamS, P0.25% w/v
Preparation H Hemorroidal CreamS, Pb0.25% w/v
Vagisil Anti-Itch CreamS, P0.25% w/v
Vagisil Moisturizing GelS, P0.25% w/v
Povidone-Iodine Medicated DoucheS, P0.25% w/v
Yeast Gard DoucheS, P0.25% w/v
Vaginal Contraceptive FoamS, P0.25% w/v

Table 4. Potentially Interfering Substances in Urogenital Specimens

a U=Urine, S = Swab, P = PreservCyt

bCycle number delays were observed for seminal fluid, y-globulin, and glucose in urine and seminal fluid, mucus, and Preparation H Hemorrhoidal Cream in PreservCyt, which could result in interference at lower target levels.

{23}------------------------------------------------

Extragenital Specimens

The potential for interference in the Alinity m STI Assay was assessed with 18 substances that may be found in oropharyngeal or rectal swab samples (Table 5). Substances were diluted into pooled oropharyngeal swab or pooled rectal swab matrices. For each substance and matrix, both CT and NG positive and negative samples were tested. The positive matrices contained CT and NG at 3 times the claimed LoD. Two different strains of CT and NG were used in this study (CT serovars D and E and NG strains Z433 and Z437). No interference was observed in the presence of the substances at the concentrations shown in Table 5 for all positive and negative samples. Cycle number delays were observed for Sensodyne Repair & Protect Sensitive Toothpaste in oropharyngeal specimens and feces in rectal specimens, which could result in interference at lower target levels.

SubstanceMatrixᵃTest Level
BloodO5.0% v/v
MucusO25 mg/mL
Listerine Antiseptic MouthwashO5.0% v/v
Mucinex Dextromethorphan Cough SuppressantO100 ug/mL
Chloraseptic Sore Throat Spray, MentholO5.0% v/v
Abreva Cold Sore MedicationO5.0% w/v
Colgate Total Whitening ToothpasteO0.25% w/v
Arm & Hammer PeroxiCare Deep Clean ToothpasteO0.25% w/v
Sensodyne Repair & Protect Sensitive ToothpasteOᵇ0.25% w/v
Preparation H Hemorrhoidal CreamR0.25% w/v
Barium SulfateR0.25% w/v
Dulcolax Laxative SuppositoryR0.25% w/v
K-Y Jelly Personal LubricantR0.25% w/v
Phillips' Milk of Magnesia Liquid LaxativeR0.25% w/v
Pepto-BismolR0.25% w/v
ImodiumR0.25% w/v
FecesRᵇ1.0% w/v
LotriminR0.25% w/v

Table 5. Potentially Interfering Substances in Oropharyngeal or Rectal Specimens

ª O = Oropharyngeal, R = Rectal

b Cycle number delays were observed for Sensodyne Repair & Protect Sensitive Toothpaste in

{24}------------------------------------------------

oropharyngeal specimens and feces in rectal specimens, which could result in interference at lower target levels.

Competitive Interference Study 5.7.1.5

A competitive interference study was conducted to challenge the performance of the Alinity m STI Assay with swab, PreservCyt, and urine samples containing CT, NG, TV, or MG at 3 times the claimed LoD in the presence of high concentrations of the other three organisms. The high positive targets were prepared with titers of 3.0x10 EB/mL for CT, 1.6x104 CFU/mL for NG, 6.0x104 TV/mL for TV, and 1.1x106 GE/mL for MG. Four conditions were evaluated:

  • CT at 3 times the claimed LoD and NG, TV, and MG at high concentrations .
  • NG at 3 times the claimed LoD and CT, TV, and MG at high concentrations .
  • TV at 3 times the claimed LoD and CT, NG, and MG at high concentrations .
  • . MG at 3 times the claimed LoD and CT, NG, and TV at high concentrations

For each analyte at the low concentration, 100% (20/20) of replicates were detected in each matrix.

Within Laboratory Precision 5.7.1.6

Alinity m STI Assay within laboratory precision was evaluated by testing panel members in 3 different sample matrices that represent urogenital specimen types used in the Alinity m STI Assay: urine, swab, and PreservCyt matrix. For each applicable specimen matrix, 13 panel members were prepared with combinations of CT, NG, TV, and MG at sub-LoD (high negative), claimed LoD, low positive (2x claimed LoD), high positive, and negative target levels (Table 6). Each panel member was tested in 2 replicates, twice each day for 12 days, on 3 Alinity m Systems with 3 reagent lots by 3 operators, for a total of 144 replicates. The results for CT, NG, TV, and MG are summarized in Table 7, Table 8, Table 9, and Table 10 respectively.

{25}------------------------------------------------

PanelMemberCTNGTVMG
1NegativeNegativeNegativeNegative
22X LoD ClaimNegativeNegativeNegative
3Negative2X LoD ClaimNegativeNegative
4NegativeNegative2X LoD ClaimNegative
5NegativeNegativeNegative2X LoD Claim
62X LoD Claim2X LoD Claim2X LoD Claim2X LoD Claim
7High Positivea2X LoD Claim2X LoD Claim2X LoD Claim
82X LoD ClaimHigh Positiveb2X LoD Claim2X LoD Claim
92X LoD Claim2X LoD ClaimHigh Positivec2X LoD Claim
102X LoD Claim2X LoD Claim2X LoD ClaimHigh Positived
11High PositiveaHigh PositivebHigh PositivecHigh Positived
12LoD ClaimLoD ClaimLoD ClaimLoD Claim
13Sub-LoD(High Negative)Sub-LoD(High Negative)Sub-LoD(High Negative)Sub-LoD(High Negative)

Table 6. Precision Panel Composition

ªConcentration for High CT is 4.4 X 104 IFU/mL or 8.8 X 103 IFU/assay (3.0x105 EB/mL or 6.0x10* EB/assay) bConcentration for High NG is 1.6 X 104 CFU/mL or 3.2 X 103 CFU/assay

℃oncentration for High TV is 6.0 X 104 TV/mL or 1.2 X 104 TV/assay

4Concentration for High MG is 1.1 X 106 genome copies/mL or 2.2 X 105 genome copies/assay

{26}------------------------------------------------

Table 7. Within Lab Precision: CT Results

Within-RunComponentBetween-RunComponentBetween-DayComponentWithin-LaboratoryBetween-Instrument/LotComponentTotald
MatrixPanel DescriptionNanbAgreement(n/N)MeanCNSD% CVSD% CVSD% CVSD% CVSD% CVSD% CV
UrineCT High Pos(NG, TV & MG High Pos)144144100.0%16.620.1911.20.1330.80.1330.80.6553.90.0940.60.6624.0
CT High Pos(NG, TV & MG at 2X LOD Claim)144144100.0%17.030.1621.00.1380.80.1380.80.4572.70.2291.30.5113.0
CT at 2X LOD Claim(TV High Pos, NG & MG at 2X LOD Claim)144144100.0%30.180.2460.80.1210.40.1210.40.6132.00.1600.50.6342.1
CT at 2X LOD Claim(NG High Pos, TV & MG at 2X LOD Claim)144144100.0%30.650.1710.60.1250.40.1250.40.3521.10.2320.80.4221.4
CT at 2 X LOD Claim(MG High Pos, NG & TV at 2X LOD Claim)144144100.0%30.370.2130.70.0690.20.0690.20.3111.00.2310.80.3881.3CT at 2X LOD Claim(NG, TV & MG at 2X LOD Claim)144144100.0%30.800.3801.20.0790.30.0790.30.4891.60.3211.00.5851.9CT at 2X LOD Claim(CT only)144144100.0%30.230.1800.60.1650.50.1650.50.2870.90.1770.60.3371.1CT at LOD Claim144144100.0%31.730.1640.50.1120.40.1120.40.4041.30.2040.60.4531.4CT at Sub-LOD(High Negative)1447652.8%37.080.5221.40.1010.30.1010.30.6301.70.2400.60.6751.8CT Negativee57657599.8%SwabCT High Pos(NG, TV & MG High Pos)144144100.0%16.940.2191.31.6459.71.6459.71.6599.80.2531.51.679f9.9CT High Pos(NG, TV & MG at 2X LOD Claim)144144100.0%16.970.0830.50.0460.30.0460.30.1030.60.1090.60.1500.9CT at 2X LOD Claim(TV High Pos, NG & MG at 2X LOD Claim)144144100.0%29.580.1020.30.0000.00.0000.00.1160.40.1130.40.1620.5CT at 2X LOD Claim(NG High Pos, TV & MG at 2X LOD Claim)144144100.0%29.530.1020.30.0370.10.0370.10.1250.40.1400.50.1880.6CT at 2 X LOD Claim(MG High Pos, NG & TV at 2X LOD Claim)144144100.0%29.620.1130.40.0270.10.0270.10.1190.40.1370.50.1810.6CT at 2X LOD Claim(NG, TV & MG at 2X LOD Claim)144144100.0%29.600.1630.50.0710.20.0710.20.1830.60.1700.60.2490.8CT at 2X LOD Claim(CT only)144144100.0%29.710.0840.30.0220.10.0220.10.0900.30.1050.40.1380.5CT at LOD Claim144144100.0%30.510.2300.80.0360.10.0360.10.2360.80.1170.40.2640.9CT at Sub-LOD(High Negative)1448861.1%36.600.5431.50.1930.50.1930.50.5761.60.2980.80.6491.8CT Negativee57657599.8%
CT at 2 X LOD Claim(MG High Pos, NG & TV at 2X LOD Claim)144144100.0%30.370.2130.70.0690.20.0690.20.3111.00.2310.80.3881.3
CT at 2X LOD Claim(NG, TV & MG at 2X LOD Claim)144144100.0%30.800.3801.20.0790.30.0790.30.4891.60.3211.00.5851.9
CT at 2X LOD Claim(CT only)144144100.0%30.230.1800.60.1650.50.1650.50.2870.90.1770.60.3371.1
CT at LOD Claim144144100.0%31.730.1640.50.1120.40.1120.40.4041.30.2040.60.4531.4
CT at Sub-LOD(High Negative)1447652.8%37.080.5221.40.1010.30.1010.30.6301.70.2400.60.6751.8
CT Negativee57657599.8%
SwabCT High Pos(NG, TV & MG High Pos)144144100.0%16.940.2191.31.6459.71.6459.71.6599.80.2531.51.679f9.9
CT High Pos(NG, TV & MG at 2X LOD Claim)144144100.0%16.970.0830.50.0460.30.0460.30.1030.60.1090.60.1500.9
CT at 2X LOD Claim(TV High Pos, NG & MG at 2X LOD Claim)144144100.0%29.580.1020.30.0000.00.0000.00.1160.40.1130.40.1620.5
CT at 2X LOD Claim(NG High Pos, TV & MG at 2X LOD Claim)144144100.0%29.530.1020.30.0370.10.0370.10.1250.40.1400.50.1880.6
CT at 2 X LOD Claim(MG High Pos, NG & TV at 2X LOD Claim)144144100.0%29.620.1130.40.0270.10.0270.10.1190.40.1370.50.1810.6
CT at 2X LOD Claim(NG, TV & MG at 2X LOD Claim)144144100.0%29.600.1630.50.0710.20.0710.20.1830.60.1700.60.2490.8
CT at 2X LOD Claim(CT only)144144100.0%29.710.0840.30.0220.10.0220.10.0900.30.1050.40.1380.5
CT at LOD Claim144144100.0%30.510.2300.80.0360.10.0360.10.2360.80.1170.40.2640.9
CT at Sub-LOD(High Negative)1448861.1%36.600.5431.50.1930.50.1930.50.5761.60.2980.80.6491.8
CT Negativee57657599.8%

a N:Total number of replicates

bn: Number of replicates with deectable parel and non-dected for negative parel; the number of replicates vere used for the Mean and SD calculation.

C Within-Laboratory includes Within-Run, Between-Run and Between-Day Components.

d Total includes Within-Run, Between-Run, Between-Day and Between-Instrument/Lot Components.

€ The negative panel included 4 panel members negative for CT

Tw samles ladedlular contr (CC) filines and very has a minim in ST assy reports prairie cestle centrif the Child the tod SD and W. Whot those samples, the total SD was 0.143 and the total % CV was 0.9.

{27}------------------------------------------------

Table 8. Within Lab Precision: NG Results

Within-RunComponentBetween-RunComponentBetween-DayComponentWithin-LaboratoryBetween-Instrument/LotComponentTotalª
MatrixPanel DescriptionNbAgreement(n/N)MeanCNSD% CVSD% CVSD% CVSD% CVSD% CVSD% CV
NG High Pos (CT, TV & MG High Pos)144144100.0%21.890.1830.80.1380.60.1380.60.3221.50.2091.00.3841.8
NG High Pos(CT, TV & MG at 2X LOD Claim)144144100.0%22.410.2301.00.2060.90.2060.90.4161.90.1930.90.4582.0
NG at 2X LOD Claim(MG High Pos, CT & TV at 2X LOD Claim)144144100.0%30.950.2870.90.0440.10.0440.10.3911.30.1400.50.4151.3
NG at 2X LOD Claim(TV High Pos, CT & MG at 2X LOD Claim)144144100.0%30.720.2450.80.0000.00.0000.00.3181.00.1970.60.3741.2
UrineNG at 2X LOD Claim(CT, TV & MG at 2X LOD Claim)144144100.0%31.370.2910.90.2070.70.2070.70.4331.40.0220.10.4341.4
NG at 2X LOD Claim(CT High Pos, TV & MG at 2X LOD Claim)144144100.0%31.240.2120.70.1440.50.1440.50.3461.10.0930.30.3591.1
NG at 2X LOD Claim (NG only)144144100.0%31.230.1890.60.0000.00.0000.00.2410.80.0830.30.2550.8
NG at LOD Claim144144100.0%32.320.2170.70.1490.50.1490.50.3841.20.1260.40.4041.3
NG at Sub-LOD(High Negative)14411982.6%37.800.8492.20.1910.50.1910.50.8702.30.5271.41.0172.7
NG Negative576576100.0%
NG High Pos (CT, TV & MG High Pos)144144100.0%21.310.1350.60.1170.60.1170.60.1940.90.2321.10.3031.4
NG High Pos(CT, TV & MG at 2X LOD Claim)144144100.0%22.070.1650.70.1840.80.1840.80.3421.50.1340.60.3671.7
NG at 2X LOD Claim(MG High Pos, CT & TV at 2X LOD Claim)144144100.0%31.450.1790.60.0920.30.0920.30.2380.80.1260.40.2700.9
NG at 2X LOD Claim(TV High Pos, CT & MG at 2X LOD Claim)144144100.0%31.320.1590.50.1660.50.1660.50.2890.90.1360.40.3191.0
SwabNG at 2X LOD Claim(CT, TV & MG at 2X LOD Claim)NG at 2X LOD Claim144144100.0%31.400.1730.50.2580.80.2580.80.3231.00.0000.00.3231.0
(CT High Pos, TV & MG at 2X LOD Claim)144144100.0%31.330.1920.60.1190.40.1190.40.2760.90.1710.50.3241.0
NG at 2X LOD Claim (NG only)144144100.0%31.490.1730.60.0960.30.0960.30.2050.60.1960.60.2830.9
NG at LOD Claim144144100.0%32.160.2010.60.1030.30.1030.30.2430.80.1270.40.2740.9
NG at Sub-LOD(High Negative)14412284.7%36.760.7302.00.1980.50.1980.50.7572.10.2660.70.8022.2
NG Negative®576576100.0%
NG High Pos (CT, TV & MG High Pos)144144100.0%21.280.1470.70.1770.80.1770.80.2301.10.1620.80.2811.3
NG High Pos(CT, TV & MG at 2X LOD Claim)144144100.0%24.040.1910.80.0630.30.0630.30.2050.90.2321.00.3101.3
NG at 2X LOD Claim(MG High Pos, CT & TV at 2X LOD Claim)144144100.0%30.660.1810.60.0820.30.0820.30.1990.60.1290.40.2370.8
NG at 2X LOD Claim(TV High Pos, CT & MG at 2X LOD Claim)144144100.0%30.910.1640.50.0860.30.0860.30.1980.60.2200.70.2961.0
PreservCytNG at 2X LOD Claim(CT, TV & MG at 2X LOD Claim)144144100.0%31.800.2530.80.0810.30.0810.30.2780.90.1340.40.3091.0
NG at 2X LOD Claim(CT High Pos, TV & MG at 2X LOD Claim)144144100.0%31.350.2190.70.1030.30.1030.30.2420.80.1950.60.3111.0
NG at 2X LOD Claim (NG only)144144100.0%31.180.2240.70.1090.40.1090.40.2660.90.2500.80.3651.2
NG at LOD Claim144144100.0%32.870.2720.80.1850.60.1850.60.3291.00.1530.50.3631.1
NG at Sub-LOD(High Negative)144ર્ ર45.1%37.310.6131.60.3320.90.3320.90.6981.90.2320.60.7352.0
NG Negative®576576100.0%……:……
Panel DescriptionNbWithin-RunComponentBetween-RunComponentBetween-DayComponentWithin-LaboratoryBetween-Instrument/LotComponentTotalª
MatrixAgreement(n/N)MeanCNSD% CVSD% CVSD% CVSD% CVSD% CVSD% CV
TV High Pos(CT, NG & MG High Pos)144144100.0%9.850.2742.80.1781.80.1781.80.7067.20.2562.60.7517.6
TV High Pos(CT, NG & MG at 2X LOD Claim)144144100.0%9.850.3533.60.2042.10.2042.10.7587.70.3573.60.8378.5
TV at 2X LOD Claim(MG High Pos, CT & NG at 2X LOD Claim)144144100.0%27.300.2450.90.1010.40.1010.40.3601.30.2951.10.4651.7
TV at 2X LOD Claim(NG High Pos, CT & MG at 2X LOD Claim)144144100.0%27.600.1860.70.1750.60.1750.60.3971.40.3531.30.5311.9
UrineTV at 2X LOD Claim(CT High Pos, NG & MG at 2X LOD Claim)144144100.0%27.580.2240.80.0610.20.0610.20.4611.70.2971.10.5482.0
TV at 2X LOD Claim(CT, NG & MG at 2X LOD Claim)14414399.3%27.840.2100.80.0000.00.0000.00.3681.30.3411.20.5021.8
TV at 2X LOD Claim(TV only)144144100.0%27.220.2661.00.0000.00.0000.00.3931.40.3751.40.5432.0
TV at LOD Claim144144100.0%28.730.3551.20.1900.70.1900.70.5201.80.3211.10.6112.1
TV at Sub-LOD(High Negative)1443423.6%33.890.9562.80.0000.00.0000.00.9782.90.3551.01.0413.1
TV Negative®57657499.7%
TV High Pos(CT, NG & MG High Pos)144144100.0%10.440.2672.61.67916.11.67916.11.70416.30.5835.61.80117.3
TV High Pos(CT, NG & MG at 2X LOD Claim)144144100.0%10.130.1701.70.0250.20.0250.20.2212.20.3713.70.4324.3
TV at 2X LOD Claim(MG High Pos, CT & NG at 2X LOD Claim)144144100.0%29.310.3761.30.1590.50.1590.50.4091.40.2951.00.5041.7
TV at 2X LOD Claim(NG High Pos, CT & MG at 2X LOD Claim)144144100.0%29.110.1920.70.0930.30.0930.30.2650.90.3301.10.4231.5
SwabTV at 2X LOD Claim(CT High Pos, NG & MG at 2X LOD Claim)144144100.0%29.240.2000.70.1610.60.1610.60.2801.00.2981.00.4091.4
TV at 2X LOD Claim(CT, NG & MG at 2X LOD Claim)144144100.0%28.940.3381.20.1650.60.1650.60.4021.40.2750.90.4871.7
TV at 2X LOD Claim(TV only)144144100.0%26.990.1900.70.0540.20.0540.20.2070.80.2641.00.3361.2
TV at LOD Claim144144100.0%29.750.3341.10.0550.20.0550.20.3521.20.2590.90.4371.5
TV at Sub-LOD(High Negative)14413593.8%33.550.3781.10.0000.00.0000.00.4111.20.1510.50.4381.3
TV Negative®57657599.8%
TV High Pos(CT, NG & MG High Pos)144144100.0%9.110.1822.00.1521.70.1521.70.2382.60.3944.30.4605.1
TV High Pos(CT, NG & MG at 2X LOD Claim)144144100.0%8.540.1661.90.0991.20.0991.20.1932.30.3564.20.4054.7
TV at 2X LOD Claim(MG High Pos, CT & NG at 2X LOD Claim)144144100.0%26.980.2911.10.1260.50.1260.50.3171.20.3681.40.4861.8
TV at 2X LOD Claim(NG High Pos, CT & MG at 2X LOD Claim)144144100.0%27.410.2160.80.2060.80.2060.80.3041.10.3761.40.4831.8
PreservCytTV at 2X LOD Claim(CT High Pos, NG & MG at 2X LOD Claim)144144100.0%26.590.166().60.0630.20.0630.20.1800.70.3151.20.3631.4
TV at 2X LOD Claim(CT, NG & MG at 2X LOD Claim)144144100.0%28.110.2891.00.1240.40.1240.40.3141.10.2480.90.4011.4
TV at 2X LOD Claim(TV only)144144100.0%27.880.1280.50.1010.40.1010.40.2010.70.3311.20.3881.4
TV at LOD Claim144144100.0%29.010.4521.60.0000.00.0000.00.4661.60.2470.90.5271.8
TV at Sub-LOD(High Negative)1448559.0%33.950.3521.00.0000.00.0000.00.3731.10.0960.30.3851.1
TV Negative®57657599.8%
Panel DescriptionNᵇAgreement(n/N)Within-RunComponentBetween-RunComponentBetween-DayComponentWithin-LaboratoryᶜBetween-Instrument/LotComponentTotalᵈ
MatrixNᵃMeanCNSD% CVSD% CVSD% CVSD% CVSD% CVSD% CV
MG High Pos(CT, NG & TV High Pos)144144100.0%20.570.1991.00.1490.70.1490.70.6603.20.4232.10.7843.8
UrineMG High Pos (CT, NG & TV at 2X LOD Claim)144144100.0%21.630.2191.00.0980.50.0980.50.3521.60.4722.20.5892.7
MG at 2X LOD Claim(TV High Pos, CT & NG at 2X LOD Claim)144144100.0%31.390.3151.00.1480.50.1480.50.6922.20.4431.40.8222.6
MG at 2X LOD Claim(NG High Pos, CT & TV at 2X LOD Claim)144144100.0%31.760.2420.80.1820.60.1820.60.4381.40.4981.60.6632.1
MG at 2X LOD Claim(CT High Pos, NG & TV at 2X LOD Claim)144144100.0%31.740.2420.80.1120.40.1120.40.5171.60.4401.40.6792.1
MG at 2X LOD Claim(CT, NG & TV at 2X LOD Claim)144144100.0%31.770.6722.10.0000.00.0000.00.7582.40.5591.80.9423.0
MG at 2X LOD Claim (MG only)144144100.0%31.730.1390.40.0870.30.0870.30.3051.00.3761.20.4841.5
MG at LOD Claim144144100.0%32.900.3171.00.1260.40.1260.40.5891.80.5301.60.7922.4
MG at Sub-LOD(High Negative)14411177.1%38.851.0842.80.1250.30.1250.31.3643.50.0000.01.3643.5
MG Negativeᵉ576576100.0%.......................................
MG High Pos(CT, NG & TV High Pos)144144100.0%19.850.4042.01.8269.21.8269.21.8709.40.5742.91.9569.9ᶠ
MG High Pos (CT, NG & TV at 2X LOD Claim)144144100.0%20.580.1410.70.0810.40.0810.40.1630.80.3821.90.4152.0
MG at 2X LOD Claim(TV High Pos, CT & NG at 2X LOD Claim)144144100.0%30.900.1410.50.0170.10.0170.10.1530.50.3371.10.3701.2
MG at 2X LOD Claim(NG High Pos, CT & TV at 2X LOD Claim)144144100.0%30.760.1580.50.0380.10.0380.10.1820.60.3241.10.3711.2
SwabMG at 2X LOD Claim(CT High Pos, NG & TV at 2X LOD Claim)144144100.0%30.860.1230.40.0720.20.0720.20.1470.50.3071.00.3401.1
MG at 2X LOD Claim(CT, NG & TV at 2X LOD Claim)144144100.0%30.720.1990.60.1300.40.1300.40.2430.80.3271.10.4071.3
MG at 2X LOD Claim (MG only)144144100.0%31.310.1370.40.0790.30.0790.30.1600.50.2520.80.2981.0
MG at LOD Claim144144100.0%31.550.2480.80.0000.00.0000.00.2480.80.2700.90.3671.2
MG at Sub-LOD(High Negative)1449465.3%35.950.4091.10.2630.70.2630.70.4871.40.1710.50.5161.4
MG Negativeᵉ576576100.0%.......................................

a N: total number of replicates

b n: Number of replicates with detectable parel and non-detected for negative panel; the number of replicates were used for the Mean and SD calculation.

C Within-Laboratory includes Within-Run, Between-Run and Between-Day Components.

d Total includes Within-Run, Between-Run, Between-Day and Between-Instrument/Lot Components.

€ The negative panel included 4 panel members negative for NG.

{28}------------------------------------------------

Table 9. Within Lab Precision: TV Results

a N: Total number of replicates,

b n: Number of replicates with detectable parel and non-etected for negative parel; the nurber of replicates were used for the Mean and SD calculation.

C Within-Laboratory includes Within-Run, Between-Run and Between-Day Components.

d Total includes Within-Run, Between-Run, Between-Day and Between-Instrument/Lot Components.

e The negative panel included 4 panel members negative for TV.

"Two same hadellate control (C) filimes and very Let any envire positives and the CCails, the Child Me call Del MCV. West Dan MCV. Web the sappes, per and sole in the sample, was 0.402 and the total %CV was 3.9.

{29}------------------------------------------------

Table 10. Within Lab Precision: MG Results

a N: Total number of replicates

b n . Number of replicates with detectable parel and nor-ecected for negative panel, the number of replicates were used for the Mean and SD calculation.

C Within-Laboratory includes Within-Run, Between-Run and Between-Day Components.

d Total includes Within-Run, Between-Run, Between-Day and Between-Instrument/Lot Components

e The negative panel included 4 panel members negative for MG.

" Try samles hadeellar contribute in the argent and the M. Assy contre ends event to C. C. C. in the collection versions were included in the cold S. What unes samps, the onl SD was 0.361 and the total %CV was 1.8.

{30}------------------------------------------------

5.7.1.7 Carryover

The carryover rate for Alinity m STI Assay was determined by testing alternating replicates of STI high positive samples and STI negative samples across multiple runs. The high positive samples were prepared with titers greater than or equal to 3.0x105 EB/mL for CT, 1.6x104 CFU/mL for NG, 2.3x104 TV/mL for TV, and 1.1x106 GE/mL for MG. Of the 548 negative samples tested, 1 sample was reported positive for TV and 1 sample was reported as positive for both CT and TV. All negative samples were reported negative for NG and MG. The overall carryover rate for CT was 0.2% (1/548, 95% CI: 0.0% to 1.0%), the overall carryover rate for TV was 0.4% (2/548, 95% CI: 0.1% to 1.3% ), and the overall carryover rate for NG and MG was 0.0% (0/548, 95% CI: 0.0% to 0.7%). The overall sample carryover rate was 0.4% (2/548, 95% CI: 0.1% to 1.3%).

Reproducibility Study 5.7.1.8

Reproducibility performance of the Alinity m STI Assay was evaluated by testing panel members in 3 different sample matrices: urine matrix, swab matrix, and PreservCyt matrix. For each applicable specimen matrix, a 13-member panel was prepared with combinations of CT, NG, TV, and MG at sub-LoD (High Negative), claimed LoD, low positive (2x claimed LoD), high positive, and negative target levels. A total of 3 Alinity m STI AMP Kit lots were used. Each of the 3 clinical sites tested 2 Alinity m STI AMP Kit lots, on 5 non-consecutive days for each lot. A total of 5 replicates of each panel member were tested on each of 5 days. Each of the 3 clinical sites used different lots of Alinity m STI CTRL Kits and Alinity m Sample Prep Kit 1. The reproducibility results for CT, NG, TV, and MG are summarized in Table 12, Table 13, and Table 14, respectively.

{31}------------------------------------------------

Table 11. Reproducibility Analysis: CT Results
----------------------------------------------------
MatrixPanel DescriptionNanbAgreement(n/N)MeanCNWithin-Run/DayComponentBetween-Run/DayComponentBetween-LotComponentBetween-SiteComponentTotalc
UrineCT High Pos (NG, TV & MG High Pos)150150100.0%19.310.4052.10.2781.40.1440.70.9565.01.0845.6
CT High Pos (NG, TV & MG at 2X LOD Claim)150150100.0%19.180.2241.20.2721.40.1430.70.6973.60.7944.1
CT at 2X LOD Claim (MG High Pos, NG & TV at 2X LOD Claim)150150100.0%31.500.2430.80.2090.70.1140.40.5921.90.6832.2
CT at 2X LOD Claim (TV High Pos, NG & MG at 2X LOD Claim)150150100.0%31.810.2850.90.2720.90.1280.40.7742.40.8782.8
CT at 2X LOD Claim (NG High Pos, TV & MG at 2X LOD Claim)150150100.0%31.470.1890.60.2700.90.0000.00.6091.90.6922.2
CT at 2X LOD Claim (NG, TV & MG at 2X LOD Claim)150150100.0%31.340.2000.60.2620.80.0630.20.5611.80.6542.1
CT at 2X LOD Claim (CT only)150150100.0%31.290.1820.60.2440.80.1480.50.4021.30.5261.7
CT at LOD Claim15014999.3%32.050.3441.10.1570.50.2450.80.4641.40.6472.0
CT at Sub-LOD (High Negative)1505436.0%36.850.5431.50.0160.00.5721.60.5501.50.9622.6
CT Negatived600600100.0%
SwabCT High Pos (NG, TV & MG High Pos)150150100.0%16.800.1851.10.0750.40.1510.90.0000.00.2501.5
CT High Pos (NG, TV & MG at 2X LOD Claim)150150100.0%16.950.1200.70.0490.30.1230.70.0000.00.1791.1
CT at 2X LOD Claim (MG High Pos, NG & TV at 2X LOD Claim)150150100.0%29.590.1470.50.0970.30.1330.40.0000.00.2210.7
CT at 2X LOD Claim (TV High Pos, NG & MG at 2X LOD Claim)150150100.0%29.710.4281.40.0000.00.1150.40.0000.00.4431.5
CT at 2X LOD Claim (NG High Pos, TV & MG at 2X LOD Claim)150150100.0%29.530.1590.50.1040.40.1300.40.0000.00.2300.8
CT at 2X LOD Claim (NG, TV & MG at 2X LOD Claim)150150100.0%29.570.1300.40.0910.30.1110.40.0000.00.1930.7
CT at 2X LOD Claim (CT only)150150100.0%29.450.1290.40.0880.30.1330.50.0000.00.2060.7
CT at LOD Claim150150100.0%30.470.2220.70.0690.20.1690.60.0000.00.2880.9
CT at Sub-LOD (High Negative)1505436.0%36.940.9002.40.0000.00.9402.50.0000.01.3013.5
CT Negatived60059599.2%
Within-Run/DayComponentBetween-Run/DayComponentBetween-LotComponentBetween-SiteComponentTotalc
MatrixPanel DescriptionNanbAgreement(n/N)MeanCNSD% CVSD% CVSD% CVSD% CVSD% CV
UrineNG High Pos (CT, TV & MG High Pos)150150100.0%24.850.5132.11.1624.70.5472.20.4481.81.4535.8
NG High Pos (CT, TV & MG at 2X LOD Claim)150150100.0%26.050.4361.71.1364.40.5942.30.9743.71.6686.4
NG at 2X LOD Claim (MG High Pos, CT & TV at 2X LOD Claim)15014898.7%35.290.7642.21.0733.00.5401.51.0583.01.7745.0
NG at 2X LOD Claim (TV High Pos, CT & MG at 2X LOD Claim)150150100.0%34.700.7352.11.0893.10.1600.50.7602.21.5264.4
NG at 2X LOD Claim (CT High Pos, TV & MG at 2X LOD Claim)15014999.3%34.680.6621.91.0182.90.4011.20.9112.61.5704.5
NG at 2X LOD Claim (CT, TV & MG at 2X LOD Claim)15014798.0%35.371.0042.81.0933.10.6791.91.0372.91.9345.5
NG at 2X LOD Claim (NG only)15014596.7%35.840.7672.10.6771.90.7962.21.0252.91.6524.6
NG at LOD Claim15014194.0%36.200.9312.61.1103.10.5261.50.8702.41.7704.9
NG at Sub-LOD (High Negative)150149.3%37.820.0000.00.7351.90.6171.60.4431.21.0562.8
NG Negatived600600100.0%-----------
SwabNG High Pos (CT, TV & MG High Pos)150150100.0%21.770.2591.20.1720.80.1120.50.0000.00.3311.5
NG High Pos (CT, TV & MG at 2X LOD Claim)150150100.0%22.510.2891.30.2681.20.0000.00.1970.90.4402.0
NG at 2X LOD Claim (MG High Pos, CT & TV at 2X LOD Claim)150150100.0%31.630.3161.00.1590.50.1550.50.1390.40.4111.3
NG at 2X LOD Claim (TV High Pos, CT & MG at 2X LOD Claim)150150100.0%31.440.2750.90.1590.50.1470.50.0000.00.3501.1
NG at 2X LOD Claim (CT High Pos, TV & MG at 2X LOD Claim)150150100.0%31.680.2940.90.2620.80.0000.00.1360.40.4171.3
NG at 2X LOD Claim (CT, TV & MG at 2X LOD Claim)150150100.0%31.540.3061.00.2460.80.0000.00.1110.40.4081.3
NG at 2X LOD Claim (NG only)150150100.0%31.780.2850.90.1800.60.1170.40.1080.30.3731.2
NG at LOD Claim150150100.0%32.490.3361.00.1850.60.0530.20.1160.40.4051.2
NG at Sub-LOD (High Negative)1508053.3%36.950.9342.50.0000.00.1260.30.1630.40.9562.6
NG Negatived600600100.0%-----------
PreservCytNG High Pos (CT, TV & MG High Pos)150150100.0%21.640.3221.50.1820.80.0860.40.1690.80.4151.9
NG High Pos (CT, TV & MG at 2X LOD Claim)150150100.0%25.300.2390.90.1980.80.0550.20.0970.40.3291.3
NG at 2X LOD Claim (MG High Pos, CT & TV at 2X LOD Claim)150150100.0%31.080.2850.90.2310.70.1420.50.0000.00.3941.3
NG at 2X LOD Claim (TV High Pos, CT & MG at 2X LOD Claim)150150100.0%31.160.2710.90.1760.60.0000.00.0850.30.3341.1
NG at 2X LOD Claim (CT High Pos, TV & MG at 2X LOD Claim)150150100.0%30.390.3281.10.1630.50.1370.50.0000.00.3921.3
MatrixPanel DescriptionNanbAgreement(n/N)MeanCNWithin-Run/DayComponentBetween-Run/DayComponentBetween-LotComponentBetween-SiteComponentTotalc
NG at 2X LOD Claim (CT, TV & MG at 2X LOD Claim)150150100.0%32.060.3761.20.3091.00.2250.70.0000.00.5361.7
NG at 2X LOD Claim (NG only)150150100.0%30.480.3181.00.2450.80.2400.80.1280.40.4851.6
NG at LOD Claim150150100.0%33.100.4131.20.1660.50.1930.60.0000.00.4861.5
NG at Sub-LOD (High Negative)1506342.0%37.460.7422.00.0000.00.3230.90.0000.00.8102.2
NG Negatived60059899.7%.................................
Panel DescriptionNanbAgreement(n/N)MeanCNWithin-Run/DayComponentBetween-Run/DayComponentBetween-LotComponentBetween-SiteComponentTotalC
MatrixSD% CVSD% CVSD% CVSD% CVSD% CV
UrineTV High Pos (CT, NG & MG High Pos)150150100.0%11.580.6065.20.3873.30.0000.00.9798.51.21510.5
TV High Pos (CT, NG & MG at 2X LOD Claim)150150100.0%11.450.3282.90.4093.60.0000.00.7806.80.9408.2
TV at 2X LOD Claim (MG High Pos, CT & NG at 2X LOD Claim)150150100.0%31.610.3641.20.2510.80.2530.80.6872.20.8552.7
TV at 2X LOD Claim (NG High Pos, CT & MG at 2X LOD Claim)15014999.3%31.380.4451.40.3691.20.0760.20.5681.80.8142.6
TV at 2X LOD Claim (CT High Pos, NG & MG at 2X LOD Claim)150150100.0%31.560.3971.30.2060.70.3211.00.6922.20.8842.8
TV at 2X LOD Claim (CT, NG & MG at 2X LOD Claim)150150100.0%31.090.3151.00.2780.90.1030.30.6422.10.7752.5
TV at 2X LOD Claim (TV only)150150100.0%31.340.4651.50.0870.30.3071.00.2050.70.6001.9
TV at LOD Claim15014999.3%31.750.3571.10.1560.50.2290.70.5211.60.6902.2
TV at Sub-LOD (High Negative)1505637.3%34.880.9372.70.2270.70.4851.40.3140.91.1243.2
TV Negatived60059699.3%
SwabTV High Pos (CT, NG & MG High Pos)150150100.0%10.430.3123.00.1451.40.3223.10.0000.00.4724.5
TV High Pos (CT, NG & MG at 2X LOD Claim)150150100.0%10.470.7156.80.1311.20.2752.60.0000.00.7787.4
TV at 2X LOD Claim (MG High Pos, CT & NG at 2X LOD Claim)150150100.0%29.290.4941.70.2831.00.1160.40.0000.00.5812.0
TV at 2X LOD Claim (NG High Pos, CT & NG at 2X LOD Claim)150150100.0%29.220.3361.10.2340.80.3641.20.0000.00.5481.9
TV at 2X LOD Claim (CT High Pos, NG & MG at 2X LOD Claim)150150100.0%29.330.3491.20.1340.50.3641.20.0000.00.5221.8
TV at 2X LOD Claim (CT, NG & MG at 2X LOD Claim)150150100.0%29.050.2000.70.1420.50.2730.90.0000.00.3671.3
TV at 2X LOD Claim (TV only)150150100.0%29.690.2690.90.1730.60.2220.70.0640.20.3941.3
TV at LOD Claim150150100.0%29.770.3951.30.6582.20.0000.00.0000.00.7672.6
TV at Sub-LOD (High Negative)1507449.3%34.221.2723.70.9182.70.5221.50.0000.01.6534.8
TV Negatived60059699.3%
PreservCytTV High Pos (CT, NG & MG High Pos)150150100.0%9.530.2352.50.1121.20.3713.90.0000.00.4534.8
TV High Pos (CT, NG & MG at 2X LOD Claim)150150100.0%8.960.2332.60.1371.50.3003.30.0000.00.4044.5
TV at 2X LOD Claim (MG High Pos, CT & NG at 2X LOD Claim)150150100.0%27.320.4191.50.1660.60.2260.80.0000.00.5051.8
TV at 2X LOD Claim (NG High Pos, CT & NG at 2X LOD Claim)150150100.0%27.610.2070.70.1700.60.1140.40.0660.20.2991.1
TV at 2X LOD Claim (CT High Pos, NG & MG at 2X LOD Claim)150150100.0%26.770.5592.10.2621.00.2290.90.0000.00.6582.5
MatrixPanel DescriptionNanbAgreement(n/N)MeanCNWithin-Run/DayComponentBetween-Run/DayComponentBetween-LotComponentBetween-SiteComponentTotalc
TV at 2X LOD Claim (CT, NG & MG at 2X LOD Claim)150150100.0%28.390.3801.30.2130.70.1520.50.0000.00.4611.6
TV at 2X LOD Claim (TV only)150150100.0%28.370.3271.20.0700.20.1670.60.0000.00.3741.3
TV at LOD Claim150150100.0%29.320.5992.00.0000.00.1840.60.0810.30.6322.2
TV at Sub-LOD (High Negative)1504429.3%34.771.0453.00.1780.50.2220.60.0000.01.0833.1
TV Negatived60058196.8%.................................

a N: Total number of replicates

º 1: Number of replicats with deectable and non-detected for negative parel; the number of replicates were used for the Mean and SD calculation.

& Total includes Within-Run/Day, Between-Run/Day, Between-Lot and Between-Site Components.

d The negative panel included 4 panel members negative for CT

{32}------------------------------------------------

Table 12 Reproducibility Analysis: NG Results

{33}------------------------------------------------

Table 12 Reproducibility Analysis: NG Results

a N: total numberof replicates

b n: Number of replicates with detectable and non-letected for negative parel; the minter of replicates were used for the Mean and SD calculation.

C Total includes Within-Run/Day, Between-Run/Day, Between-Lot and Between-Site Components.

d The negative panel included 4 panel members negative for NG.

{34}------------------------------------------------

Table 13. Reproducibility Analysis: TV Results

{35}------------------------------------------------

Table 13. Reproducibility Analysis: TV Results

a N: Total number of replicates

b n: Number of replicates with deectable and non-letected for negative parel; the minter of replicates were used for the Mean and SD calculation.

C Total includes Within-Run/Day, Between-Run/Day, Between-Lot and Between-Site Components.

d The negative panel included 4 panel members negative for TV.

{36}------------------------------------------------

Within-Run/DayComponentBetween-Run/DayComponentBetween-LotComponentBetween-SiteComponentTotalc
MatrixPanel DescriptionNanbAgreement(n/N)MeanCNSD% CVSD% CVSD% CVSD% CVSD% CV
UrineMG High Pos (CT, NG & TV High Pos)150150100.0%22.660.4281.90.2281.00.3291.50.9064.01.0794.8
MG High Pos (CT, NG & TV at 2X LOD Claim)150150100.0%22.670.2921.30.2661.20.2311.00.7083.10.8433.7
MG at 2X LOD Claim (TV High Pos, CT & NG at 2X LOD Claim)150150100.0%33.730.4971.50.3090.90.3411.00.9452.81.1633.4
MG at 2X LOD Claim (NG High Pos, CT & TV at 2X LOD Claim)150150100.0%33.180.3821.20.3040.90.0330.10.8252.50.9592.9
MG at 2X LOD Claim (CT High Pos, NG & TV at 2X LOD Claim)150150100.0%33.180.3681.10.2570.80.3531.10.7042.10.9072.7
MG at 2X LOD Claim (CT, NG & TV at 2X LOD Claim)150150100.0%32.850.3120.90.2920.90.2670.80.7342.20.8902.7
MG at 2X LOD Claim (MG only)150150100.0%33.010.2740.80.1990.60.2950.90.3401.00.5641.7
MG at LOD Claim150150100.0%33.710.6451.90.1180.40.3621.10.7232.11.0413.1
MG at Sub-LOD (High Negative)1505536.7%40.291.0212.50.0760.20.8542.10.0000.01.3333.3
MG Negatived600600100.0%
SwabMG High Pos (CT, NG & TV High Pos)150150100.0%19.730.2251.10.1010.50.2921.50.0000.00.3821.9
MG High Pos (CT, NG & TV at 2X LOD Claim)150150100.0%20.640.1860.90.1270.60.3201.60.0000.00.3911.9
MG at 2X LOD Claim (TV High Pos, CT & NG at 2X LOD Claim)150150100.0%31.030.4711.50.0000.00.1960.60.0000.00.5111.6
MG at 2X LOD Claim (NG High Pos, CT & TV at 2X LOD Claim)150150100.0%30.800.1890.60.1210.40.2500.80.0000.00.3361.1
MG at 2X LOD Claim (CT High Pos, NG & TV at 2X LOD Claim)150150100.0%30.870.1590.50.0460.10.2360.80.0000.00.2880.9
MG at 2X LOD Claim (CT, NG & TV at 2X LOD Claim)150150100.0%30.690.1650.50.1240.40.2250.70.0000.00.3051.0
MG at 2X LOD Claim (MG only)150150100.0%31.400.1800.60.0830.30.2380.80.0000.00.3101.0
MG at LOD Claim150150100.0%31.580.2440.80.0340.10.3041.00.0000.00.3911.2
MG at Sub-LOD (High Negative)15010268.0%36.540.5711.60.1510.40.0480.10.1640.40.6141.7
MG Negatived60059999.8%

Table 14. Reproducibility Analysis: MG Results

a N: total numberof replicates

b n: Number of replicates with deectable and non-detected for negative parel; the minter of replicates were used for the Mean and SD calculation

C Total includes Within-Run/Day, Between-Run/Day, Between-Lot and Between-Site Components

d The negative panel included 4 panel members negative for MG.

{37}------------------------------------------------

5.7.2 Clinical Performance

Clinical Study Results - Urogenital Specimens 5.7.2.1

Performance characteristics of the Alinity m STI Assay with urogenital specimens were established in a multicenter clinical study conducted in the United States. Specimens were collected from subjects 14 years of age or older at 33 geographically diverse sites that included but were not limited to STI clinics, primary care offices, and gynecology practices. A total of 7,099 male and female, asymptomatic and symptomatic subjects were enrolled. Study subjects were classified as symptomatic if the subject reported STI related symptoms.

Each female subject provided up to 8 specimens, including 1 urine specimen, 1 selfcollected vaginal swab, 3 clinician-collected vaginal swabs, 2 endocervical swabs, and 1 ThinPrep (PreservCyt) cytology specimen. Each male provided 1 urine specimen.

Specimen testing methods included the Alinity m STI Assay and comparator assays for CT, NG, TV, and MG. Alinity m STI testing was performed at 3 external clinical testing sites. Comparator assays for CT and NG included 2 commercially available nucleic acid amplification tests (NAAT) for females (each NAAT tested with 1 swab and 1 urine specimen) and 3 commercially available NAATs for males (each NAAT tested with urine). Comparator assays for TV included 3 NAATs tested with swab specimens and 2 NAATs tested with urine specimens for females and 2 NAATs and culture tested with urine for males. Comparator assays for MG included 3 NAATs for females (each tested with a vaginal swab) and males (each tested with urine).

For each subject, a patient infected status (PIS) was determined based on the combined results from the comparator assays. A female subject was categorized as infected for CT or NG if a minimum of 2 positive results (at least 1 from each comparator NAAT) was reported. For CT, female subjects with positive results on both comparator urine specimens and negative results on both comparator swab specimens (endocervical swab from NAAT 1 and clinician-collected vaginal swab specimens from NAAT 2) were categorized as infected for urine and not infected for swab specimens (there were 2 female subjects that were negative for CT in the swab samples and positive in urine by

{38}------------------------------------------------

both comparator NAATs). A female subject was categorized as not infected for CT or NG if at least 1 of the comparator NAATs reported negative results for all sample types. Refer to Table 20, Table 21, Table 24 and Table 25 for the female PIS algorithm. A female subject was categorized as infected for TV or MG if the first 2 swab comparator NAAT results were both positive or if 2 of the 3 swab comparator NAAT results were positive in cases where the 3ª NAAT was used as a tie-breaker. A female subject was categorized as not infected for TV or MG if the first 2 swab comparator NAAT results were both negative or if 2 of the 2 swab comparator NAAT results were negative in cases where 3rd NAAT was used as a tie-breaker. Refer to Table 28, Table 32 and Table 33 for the female PIS algorithm.

A male subject was categorized as infected for CT, NG, TV, or MG if a minimum of two comparator positive results was reported. If the comparator TV culture assay result was positive, the subject was categorized as infected for TV regardless of NAAT results. A male subject was categorized as not infected for CT, NG, or MG if two or more comparator NAAT results were negative. For TV, a male subject was categorized as not infected if the TV culture assay result was negative, and if one or more comparator NAATs were negative. Refer to Table 22, Table 26, Table 26, Table 27, Table 30, Table 31, Table 34 and Table 35 for the male PIS algorithm.

If a PIS could not be determined due to missing and/or indeterminate results from the comparator assays, the subject was excluded from the analysis for that analyte. Out of 7099 subjects, PIS could not be determined for 1 or more analytes for 171 subjects (42 subjects for CT, 32 subjects for NG, 73 subjects for TV, 93 subjects for MG).

Alinity m STI test results were compared to the PIS for calculation of assay sensitivity and specificity. A total of 12,903 CT, 15,655 NG, 18,843 TV, and 12,829 MG results were used in the analysis. The results were analyzed by gender, specimen type, and the presence of symptoms. CT sensitivity and specificity for female and male subjects by specimen type and by symptom status are presented in Table 15. NG sensitivity and specificity for female and male subjects by specimen type and by symptom status are presented in Table 16. TV sensitivity and specificity for female and male subjects by

{39}------------------------------------------------

specimen type and by symptom status are presented in Table 17. MG sensitivity and specificity for female and male subjects by specimen type and by symptom status are presented in Table 19.

For the urogenital study, approximately 4.7% of sample results were invalid on the initial test. Specimens with initial invalid results were retested. Out of a total of 20,145 specimens, the number of specimens with final invalid results for CT, NG, TV and MG were 37, 49, 54 and 38, respectively, or approximately 0.3%.

{40}------------------------------------------------

GenderSpecimen TypeSymptom StatusNTPFPTNFNSensitivity (%)Estimate (95% CI)n / NSpecificityEstimate (95% CI)n / N
FemaleClinician-collectVaginal SwabSymptomatic1517119181378298.3 (94.2,99.5)119/12198.7 (98.0,99.2)1378/1396
Asymptomatic16498271558297.6 (91.7,99.3)82/8499.6 (99.1,99.8)1558/1565
All3166201252936498.0 (95.1,99.2)201/20599.2 (98.8,99.4)2936/2961
Self-collectVaginal SwabSymptomatic1521121141383397.6 (93.1,99.2)121/12499.0 (98.3,99.4)1383/1397
Asymptomatic164282815520100.0 (95.5,100.0)82/8299.5 (99.0,99.7)1552/1560
All3163203222935398.5 (95.8,99.5)203/20699.3 (98.9,99.5)2935/2957
Endocervical SwabSymptomatic149511381369595.8 (90.5,98.2)113/11899.4 (98.9,99.7)1369/1377
Asymptomatic159275101501692.6 (84.8,96.6)75/8199.3 (98.8,99.6)1501/1511
All30871881828701194.5 (90.4,96.9)188/19999.4 (99.0,99.6)2870/2888
MaleMale UrineSymptomatic11071232979397.6 (93.2,99.2)123/12699.8 (99.3,99.9)979/981
Asymptomatic2380155142206596.9 (92.9,98.7)155/16099.4 (98.9,99.6)2206/2220
All3487278163185897.2 (94.6,98.6)278/28699.5 (99.2,99.7)3185/3201

Table 15. CT Clinical Sensitivity and Specificity by Gender, Specimen Type and Symptom Status for Urogenital Specimens

Two subjects tested negative for CT by both comparators in the swab specimens in urine specimens. For calculations of performance, these samples were consider for swab samples and PIS CT Positive for urine samples.

{41}------------------------------------------------

GenderSpecimen TypeSymptom StatusNTPFPTNFNSensitivity (%)Estimate(95% CI)n/NSpecificity (%)Estimate(95% CI)n/N
FemaleClinician-collectVaginal SwabSymptomatic151923314930100.0 (85.7,100.0)23/2399.8 (99.4,99.9)1493/1496
Asymptomatic165118416290100.0 (82.4,100.0)18/1899.8 (99.4,99.9)1629/1633
All317041731220100.0 (91.4,100.0)41/4199.8 (99.5,99.9)3122/3129
Self-collectVaginal SwabSymptomatic152323514950100.0 (85.7,100.0)23/2399.7 (99.2,99.9)1495/1500
Asymptomatic164318516200100.0 (82.4,100.0)18/1899.7 (99.3,99.9)1620/1625
All3166411031150100.0 (91.4,100.0)41/4199.7 (99.4,99.8)3115/3125
Endocervical SwabSymptomatic149719514703a86.4 (66.7,95.3)19/2299.7 (99.2,99.9)1470/1475
Asymptomatic159318215730100.0 (82.4,100.0)18/1899.9 (99.5,100.0)1573/1575
All309037730433a92.5 (80.1,97.4)37/4099.8 (99.5,99.9)3043/3050
PreservCytSymptomatic13491911327290.5 (71.1,97.3)19/2199.9 (99.6,100.0)1327/1328
Asymptomatic138715013720100.0 (79.6,100.0)15/15100.0 (99.7,100.0)1372/1372
All27363412699294.4 (81.9,98.5)34/36100.0 (99.8,100.0)2699/2700
MaleMale UrineSymptomatic110974210330100.0 (95.1,100.0)74/7499.8 (99.3,99.9)1033/1035
Asymptomatic238428323530100.0 (87.9,100.0)28/2899.9 (99.6,100.0)2353/2356
All3493102533860100.0 (96.4,100.0)102/10299.9 (99.7,99.9)3386/3391

Table 16. NG Clinical Sensitivity and Specificity by Gender, Specimen Type and Symptom Status for Urogenital Specimens

ª A false negative test result was obtainen which was collected after 2 other swabs were collected from the cervix (ic, last of the 3 swabs). Based on the CN values from the Cellular Control, the false negative result was likely due to insufficient cellular material collected. Without this sample, the overall sensitivity of NG is 94.9% (37/39) with a 95% confidence interval of (83.1%, Refer to the limitations section for caution when collecting multiple cervical specimens.

{42}------------------------------------------------

Sensitivity (%)Specificity (%)
GenderSpecimen TypeSymptom StatusNTPFPTNFNEstimate(95% CI)n / NEstimate(95% CI)n / N
FemaleClinician-collectVaginal SwabSymptomatic1518157351325199.4 (96.5,99.9)157/15897.4 (96.4,98.1)1325/1360
Vaginal SwabAsymptomatic16541594414510100.0 (97.6,100.0)159/15997.1 (96.1,97.8)1451/1495
All3172316792776199.7 (98.2,99.9)316/31797.2 (96.6,97.8)2776/2855
Self-collectVaginal SwabSymptomatic1522157271337199.4 (96.5,99.9)157/15898.0 (97.1,98.6)1337/1364
Vaginal SwabAsymptomatic1644155351453199.4 (96.5,99.9)155/15697.6 (96.7,98.3)1453/1488
All3166312622790299.4 (97.7,99.8)312/31497.8 (97.2,98.3)2790/2852
Endocervical SwabSymptomatic1496152461295398.1 (94.5,99.3)152/15596.6 (95.5,97.4)1295/1341
Asymptomatic1597152391402497.4 (93.6,99.0)152/15697.3 (96.3,98.0)1402/1441
All3093304852697797.7 (95.4,98.9)304/31196.9 (96.2,97.5)2697/2782
PreservCytSymptomatic133713151197497.0 (92.6,98.8)131/13599.6 (99.0,99.8)1197/1202
Asymptomatic138612791242894.1 (88.7,97.0)127/13599.3 (98.6,99.6)1242/1251
All27232581424391295.6 (92.4,97.4)258/27099.4 (99.0,99.7)2439/2453
MaleMale UrineSymptomatic110924910760100.0 (86.2,100.0)24/2499.2 (98.4,99.6)1076/1085
Asymptomatic238554172313198.2 (90.4,99.7)54/5599.3 (98.8,99.5)2313/2330
All349478263389198.7 (93.2,99.8)78/7999.2 (98.9,99.5)3389/3415

Table 17. TV Clinical Sensitivity and Specificity by Gender, Specimen Type and Symptom Status for Urogenital Specimens

For female urine, the TV result of the Alinity m STI Assay was compared against a specimen-specific agreement.. A female urine specimen was considered positive for TV if at least 1 comparator NAAT was positive for the urine specimen. A female urine specimen was considered negative if both comparator NAATs were negative for the urine specimen. If a TV specimen-specific status for female urine could not be determined due to missing and/or indeterminate results from the urine comparator assays, the subject was excluded from the analysis for that analyte. The specimen-specific TV status could not be determined for 37 subjects. TV specimen-specific positive and negative agreement for female urine by symptom status are presented in Table 18. The TV clinical sensitivity based on the PIS was up to 6.6% lower in female urine than in vaginal swab specimens.

{43}------------------------------------------------

PPANPA
AnalyteSpecimenTypeSymptomStatusNAlinity +CCA+Alinity+CCA-Alinity-CCA-Alinity-CCA+Estimate(95% CI)n / NEstimate(95% CI)n / N
TVFemaleUrineSymptomatic1507141161346497.2(93.1,98.9)141/14598.8(98.1,99.3)1346/1362
Asymptomatic1651154101484398.1(94.5,99.3)154/15799.3(98.8,99.6)1484/1494
All3158295262830797.7(95.3,98.9)295/30299.1(98.7,99.4)2830/2856

Table 19. MG Clinical Sensitivity and Specificity by Gender, Specimen Type and Symptom Status for Urogenital Specimens

Sensitivity (%)Specificity (%)
GenderSpecimen TypeSymptom StatusNTPFPTNFNEstimate(95% CI)n / NEstimate(95% CI)n / N
FemaleClinician-collectVaginal SwabSymptomatic1514148151350199.3 (96.3,99.9)148/14998.9 (98.2,99.3)1350/1365
Vaginal SwabAsymptomatic164310581526496.3 (90.9,98.6)105/10999.5 (99.0,99.7)1526/1534
All3157253232876598.1 (95.5,99.2)253/25899.2 (98.8,99.5)2876/2899
Self-collectVaginal SwabSymptomatic1517144211346696.0 (91.5,98.2)144/15098.5 (97.7,99.0)1346/1367
Vaginal SwabAsymptomatic1632104201502694.5 (88.6,97.5)104/11098.7 (98.0,99.1)1502/1522
All31492484128481295.4 (92.1,97.3)248/26098.6 (98.1,99.0)2848/2889
Endocervical SwabSymptomatic14921251013362185.6 (79.0,90.4)125/14699.3 (98.6,99.6)1336/1346
Asymptomatic1584821414662278.8 (70.0,85.6)82/10499.1 (98.4,99.4)1466/1480
All30762072428024382.8 (77.6,87.0)207/25099.2 (98.7,99.4)2802/2826
MaleMale UrineSymptomatic10999940958298.0 (93.1,99.5)99/10196.0 (94.6,97.0)958/998
Asymptomatic2348110412195298.2 (93.7,99.5)110/11298.2 (97.5,98.6)2195/2236
All3447209813153498.1 (95.3,99.3)209/21397.5 (96.9,98.0)3153/3234

The numbers of specimens in all combinations of PIS, individual comparator results and Alinity m STI result are summarized. CT results for infected and non-infected female subjects are presented in Table 20 and Table 21, and for infected and non-infected male subjects in Table 22 and Table 23. NG results for infected and non-infected female subjects are presented in Table 24 and Table 25, and for infected and non-infected male subjects in Table 26 and Table 27. TV results for infected and non-infected female subjects are presented in Table 28 and Table 29, and for infected and non- infected male subjects in Table 30 and Table 31. MG results for infected and non-infected female subjects are presented in Table 32 and Table 33, and for infected and non-infected male subjects in Table 34 and Table 35.

{44}------------------------------------------------

NAAT 1NAAT 2Alinity m STINo. of Subjects
EFUVSFUESCVCCVSymptomaticAsymptomaticTotal
+++++++9565160
++++N/A++224
+++++N/A+011
++++N/AN/A+011
++++N/A+N/A101
++++-++112
++++++-011
N/A++++++213
N/A+++N/A++101
++N/A++++101
+++N/A+++224
+N/A+N/A+++101
-++++++314
-+++N/A+N/A101
-+++-++033
-+++-+-011
+-+++++213
+-+++-+101
+++-+++213
+++-N/A+N/A101
+++--++112
-+N/A+---101
+N/A+-+++011
-++-+++202
-++--+-101
+-+-+++112
+-+--++101
+-+-+-+101
-+-+-+-202

Table 20. CT Analysis Per Patient Infected Status- INFECTED FEMALE Subjects (Urogenital)

E = Endocervical Swab, VS = Vaginal Swab, FU = Fenale Urine, CCV = Clinician-Collected Vaginal Swab, SCV = Self Collected Vaginal Swab, N/A = Not Available

Two subjects tested negative for CT by both comparators in the swab specimens and positive by both comparators in urine specimens. For calculations of performance, these samples were considere for swab samples and PIS CT Positive for urine samples.

{45}------------------------------------------------

NAAT 1NAAT 2Alinity m STINo. of Subjects
EFUVSFUESCVCCVSymptomaticAsymptomaticTotal
-------126714052672
----N/A--164460
-----N/A-92231
------N/A6915
----N/A-N/A369
-----N/AN/A202
----+--7815
-----+-4610
------+9413
-----++101
N/A------151732
N/A---N/A--111728
N/A----N/A-101
N/A---N/A-N/A235
-N/A-----6713
--N/A----81119
--N/A---+101
--N/A--++101
---N/A---10818
N/AN/A-----9312
N/AN/A--N/A--011
N/AN/A---N/A-011
N/AN/A--+--011
+----628
+-N/A--011
+---+-101
+----++101
-+-----325
-+---++101
--+----022

Table 21. CT Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

{46}------------------------------------------------

NAAT 1NAAT 2Alinity m STINo. of Subjects
EFUVSFUESCVCCVSymptomaticAsymptomaticTotal
--+--+-101
--+---+213
--+--++112
--+-+++011
---+N/A-N/A101
++--+--101
--++-++101

Table 21. CT Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

E = Endocervical Swab, VS = Vaginal Swab, FU = Female Urine, CCV = Clinician-Collected Vaginal Swab, SCV = Self-Collected Vaginal Swab, N/A = Not Available

{47}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
++++105131236
N/A+++011
+N/A++325
++N/A+81523
-+++224
+-++235
++-+314
+++-235
++N/A-022
+-+-101

Table 22. CT Analysis Per Patient Infected Status- INFECTED MALE Subjects (Urogenital)

MU = Male Urine, N/A = Not Available

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
----87416652539
N/A---31215
-N/A--41115
--N/A-96510606
+---145
-+--112
--+-033
---+21214
-+-+022

Table 23 CT Ar 187 cia Da Patie nt Infected Stati NOT INFECTED MALE Subje

MU = Male Urine, N/A = Not Available

{48}------------------------------------------------

NAAT 1NAAT 2Alinity m STINo. of Subjects
EFUVSFUESCVCCVPreservCytSymptomaticAsymptomaticTotal
++++++++121224
+++++++N/A022
++++N/A++N/A101
N/A+++++++101
+++N/A++++022
+++N/A+++N/A101
+N/A+N/A++++101
+-++++++314
+-+++++-101
+++--++-101
+-+N/A++++101
+-+-+++N/A011
+-+-++++101
NAAT 1NAAT 2Alinity m STINo. of Subjects
EFUVSFUESCVCCVPreservCytSymptomaticAsymptomaticTotal
--------121912612480
----N/A---7916
-----N/A--102131
------N/A-6814
-------N/A169223392
----N/AN/A--011
----N/A-N/A-011
----N/A--N/A103848
-----N/AN/A-202
-----N/A-N/A022
----N/AN/AN/A-033
----N/A-N/AN/A7512
----+---415
-----+--415
------+-314
-------+101
----+--N/A101
-----+N/A-011
-----+-N/A101
-----++-011
-----++N/A011
N/A-------211940
N/A---N/A---6612
N/A----N/A--101
N/A------N/A112
N/A---N/A-N/A-235
N/A---N/A--N/A61117
N/A---N/AN/AN/A-11213
N/A---N/AN/A-N/A011
-N/A------5611
-N/A-----N/A123
NAAT 1NAAT 2Alinity m STINo. of Subjects
EFUVSFUESCVCCVPreservCytSymptomaticAsymptomaticTotal
--N/A-----4913
--N/A----N/A011
--N/A-N/AN/AN/A-011
---N/A----11819
N/AN/A------10414
N/AN/A--N/A---011
N/AN/A---N/A--011
-+------011
--+-----011
--+-+++-011
++------101

Table 24. NG Analysis Per Patient Infected Status- INFECTED FEMALE Subjects (Urogenital)

E = Endocervical Swab, VS = Vaginal Swab, FU = Female Urine, CCV = Clinician-Collected Vaginal Swab, SCV = Self Collected

Vaginal Swab, N/A = Not Available

{49}------------------------------------------------

Table 25. NG Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

{50}------------------------------------------------

Table 25. NG Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

E = Endocervical Swab, VS = Vaginal Swab, FU = Female Urine, CCV = Clinician-Collected Vaginal Swab, N/A = Not Available

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
++++692190
N/A+++011
+N/A++101
++N/A+369
++-+101

Table 26. NG Analysis Per Patient Infected Status- INFECTED MALE Subjects (Urogenital)

MU = Male Urine, N/A = Not Available

{51}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
----92118002721
N/A---31215
-N/A--41216
--N/A-102525627
+---112
-+--112
--+-123
---+235

Table 27. NG Analysis Per Patient Infected Status- NOT INFECTED MALE Subjects (Urogenital)

MU = Male Urine, N/A = Not Available

{52}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
EVSVSESCVCCVFUPreservCytSymptomaticAsymptomaticTotal
++N/A+++++113109222
++N/A+N/A+++235
++N/A+++N/A+213
++N/A++++N/A191635
++N/AN/A+++N/A123
++N/AN/A+N/A+N/A101
++N/AN/AN/AN/AN/AN/A033
++N/A+++++538
++N/A+++++022
++N/A++++N/A022
++N/A+++--011
N/A+++++++235
N/A++N/A+N/A++101
N/A++N/A+++N/A112
+N/A++++++112
-+++++++257
-++++++N/A011
-++N/A+++N/A101
-+++++-+213
-++++++-123
-++-+++N/A202
-+++++-N/A022
-++-++-+011
-++-+++-022
-+++++--202
-++--+--112
+-++++-+101
NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
EVSVSESCVCCVFUPreservCytSymptomaticAsymptomaticTotal
--N/A-----104110342075
--N/AN/A----7815
--N/A-N/A---71522
--N/A--N/A--5813
--N/A---N/A-142943
--N/A----N/A120175295
--N/AN/AN/AN/A--011
--N/AN/A-N/A--011
--N/AN/A---N/A63642
--N/A-N/AN/A--202
--N/A-N/A-N/A-022
--N/A-N/A--N/A022
--N/A--N/AN/A-112
--N/A---N/AN/A145
--N/AN/AN/AN/AN/A-011
--N/AN/A-N/A-N/A5510
--N/AN/A--N/AN/A101
--N/AN/AN/AN/AN/A-011
--N/AN/AN/AN/A-N/A133144
--N/A-+---191635
--N/A-+---151732
--N/A--+--152237
--N/A---+-8917
--N/A----+4812
--N/A+--N/A-101
--N/A+---N/A14721
--N/A-+-N/A-101
--N/A-+--N/A3710
--N/AN/A-+--011

Table 28. TV Analysis Per Patient Infected Status- INFECTED FEMALE Subjects (Urogenital)

E = Endocervical Swab, VS = Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, SCV = Self Collected Vaginal Swab, FU =
Female Urine, N/A = Not Available

{53}------------------------------------------------

Table 29. TV Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

{54}------------------------------------------------

-N/A-N/A+--101
-N/A--+-N/A336
-N/A-N/A+--011
-N/A--+N/AN/A101
-N/AN/A+--N/A101
-N/AN/A-+-N/A101
-N/A-++--101
-N/A-+-+-4610
-N/A-++--033
-N/A--+-+101
-N/A-+-+N/A112
-N/A--+N/A-011
-N/A-++N/A-101
-N/A-+++-224
-N/A-+-++011

E = Endocervical Swab, VS = Vaginal Swab, CCV = Clinician-Collected Vaginal Swab,
SCV = Self Collected Vaginal Swab, FU = Female Urine, N/A = Not Available

{55}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
EVSVSESCVCCVFUPreservCytSymptomaticAsymptomaticTotal
--N/A-+++-011
--N/A+++-N/A123
--N/A++++-101
N/A------N/A112
N/A--N/A---N/A5914
N/A--N/A----6713
N/A---N/A---112
N/A--N/AN/A--N/A011
N/A--N/A-N/A--123
-N/A------62127
-N/A-N/A-N/A-N/A101
-N/A-----N/A5914
N/A-------271946
-N/A-+----011
-N/A---+--202
-N/A----+-202
+-------213
+--+----011
+----+--101
-+------257
-+-----N/A101
-+--+---101
-+---+--011
-+-+---N/A112
-+-+-+--101
-+--+-+-022

Table 29 Continued. TV Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

E = Endocervical Swab, VS = Vaginal Swab, CCV = Clinician-Collected Vaginal Swab,
SCV = Self Collected Vaginal Swab, FU = Female Urine

SCV = Self Collected Vaginal Swab, FU = Female Urine

{56}------------------------------------------------

CultureNAAT 1NAAT 2Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
++++153247
N/A+++112
++N/A+011
-+++61824
+-++101
++-+011
+--+112
-++-011

V Analysis Per Patient Intected Status- INFECTED MALE Subjects

MU = Male Urine, N/A = Not Available

CultureNAAT 1NAAT 2Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
----104722113258
N/A---175976
-N/A--51924
--N/A-62228
-+--101
--+-022
---+61521
-N/A-+101
--++224

MU = Male Urine, N/A = Not Available

{57}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINo. of Subjects
VSVSVSESCVCCVSymptomaticAsymptomaticTotal
++N/A+++7558133
++N/AN/A++224
++N/A+N/A+011
++N/A++N/A101
++N/A+++61117
++N/A+-+325
++N/A++N/A101
N/A+++++8311
N/A++-++011
+N/A++++101
+N/A+N/A+N/A011
+-++++341448
+-+N/A++325
+-++N/A+101
+-+N/A+N/A011
+-+-++12921
+-++-+123
+-+++-112
+-+--+202
+-+-+-011
+-++--011
+-+N/A--011

Table 32. MG Analysis Per Patient Infected Status- INFECTED FEMALE Subjects (Urogenital)

VS = Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, SCV = Self-Collected Vaginal Swab, NA = Not Available

{58}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINo. of Subjects
VSVSVSESCVCCVSymptomaticAsymptomaticTotal
--N/A---113512812416
--N/AN/A--204666
--N/A-N/A-82230
--N/A--N/A279
--N/AN/AN/A-011
--N/AN/A-N/A549
--N/A-N/AN/A202
--N/A+--224
--N/A-+-718
--N/A--+112
--N/A++-022
--N/A-++022
N/A-----142111253
N/A--N/A--224
N/A---N/A-213
N/A----N/A101
N/A----+101
N/A--+-+101
N/A---++101
-N/A----131629
-N/A-N/A--31013
-N/A--N/A-011
-N/A-N/A-N/A325
+-----51116
+--N/A--011
+----N/A101
+--+--235
+---+-4610
+----+213
+--N/A-+011
+--++-268
NAAT 1NAAT 2NAAT 3Alinity m STINo. of Subjects
VSVSVSESCVCCVSymptomaticAsymptomaticTotal
+--+-+202
+---++628
+--+++112
-+----033
------303

Table 33. MG Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

{59}------------------------------------------------

Table 33. MG Analysis Per Patient Infected Status- NOT INFECTED FEMALE Subjects (Urogenital)

VS = Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, SCV = Self-Collected Vaginal Swab, FU = Female Uine, N/A = Not Available

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
++++5669125
N/A+++15823
+N/A++279
++N/A+325
-+++202
+-++6915
++-+151530
+++-112
++--112

Table 34. MG Analysis Per Patient Infected Status-INFECTEDMALE Subjects (Urogenital)

MU = Male Urine, N/A = Not Available

{60}------------------------------------------------

NAAT 1NAAT 2NAAT 3Alinity m STINumber of Subjects
MUMUMUMUSymptomaticAsymptomaticTotal
----79017922582
N/A---123187310
-N/A--8141149
--N/A-244872
+---132134
-+--011
--+-055
---+123
N/A--+347
+--+363571

Table 35. MG Analysis Per Patient Infected Status- NOTINFECTED MALE Subjects (Urogenital)

MU = Male Urine, N/A = Not Available

Expected Values

The prevalence of CT,NG, TV, and MG in this study was dependent on several factors including age, gender, clinic type, presence of symptoms, and the method of testing. A summary of the positivity for CT,NG, TV, and MG, as determined by the Alinity m STI Assay for each specimen type, is presented by collection site and overall in Table 36 through Table 39.

{61}------------------------------------------------

% Positivity (Number Positive / Number Tested with Valid Results)% Positivity (Number Positive / Number Tested with Valid Results)Collection SiteESCVCCVFUPreservCytMU
Collection SiteESCVCCVMUCollection SiteESCVCCVPreservCytMU0112.2 (5/41)12.2 (5/41)15.4 (6/39)12.2 (5/41)11.1 (4/36)7.0 (4/57)
012.4 (1/41)2.4 (1/41)5.1 (2/39)15.3 (9/59)010.0 (0/41)2.4 (1/41)2.6 (1/39)0.0 (0/36)20.3 (12/59)0221.8 (121/554)16.8 (96/570)17.8 (101/569)12.9 (75/581)14.9 (40/269)3.1 (28/890)
024.0 (22/552)4.6 (26/568)4.2 (24/567)4.7 (42/889)020.9 (5/553)1.1 (6/569)0.9 (5/568)0.4 (1/268)0.4 (4/889)0314.2 (32/225)14.4 (32/222)14.2 (33/233)13.7 (32/233)14.3 (32/224)2.3 (6/260)
032.7 (6/225)4.1 (9/222)4.3 (10/232)3.5 (9/260)030.9 (2/225)0.9 (2/222)0.9 (2/233)0.9 (2/224)0.4 (1/260)0415.0 (3/20)15.0 (3/20)15.0 (3/20)15.0 (3/20)15.0 (3/20)2.6 (3/116)
0421.1 (4/19)15.8 (3/19)21.1 (4/19)17.2 (20/116)040.0 (0/20)5.0 (1/20)0.0 (0/20)0.0 (0/20)8.6 (10/116)058.4 (20/237)7.5 (19/253)9.6 (24/251)7.9 (20/254)6.0 (15/249)1.3 (3/230)
057.2 (17/237)7.1 (18/254)7.2 (18/251)3.9 (9/233)050.4 (1/237)0.0 (0/254)0.0 (0/251)0.0 (0/263)0.9 (2/233)0625.0 (1/4)0.0 (0/3)0.0 (0/4)25.0 (1/4)0.0 (0/4)0.0 (0/24)
060.0 (0/4)0.0 (0/3)0.0 (0/4)0.0 (0/24)060.0 (0/4)0.0 (0/3)0.0 (0/4)0.0 (0/4)0.0 (0/24)0715.4 (61/395)16.0 (70/438)16.6 (72/433)11.8 (55/466)15.2 (47/310)4.1 (25/609)
073.5 (14/395)3.4 (15/438)3.2 (14/433)3.6 (22/606)070.3 (1/395)0.5 (2/438)0.7 (3/433)0.0 (0/310)1.0 (6/608)0816.5 (17/103)19.2 (20/104)22.3 (23/103)16.7 (17/102)16.2 (17/105)8.9 (7/79)
082.9 (3/103)2.9 (3/105)2.9 (3/104)15.2 (12/79)083.9 (4/103)2.9 (3/105)2.9 (3/104)2.9 (3/105)13.9 (11/79)098.4 (9/107)10.3 (11/107)10.3 (11/107)7.7 (8/104)7.3 (8/109)0.0 (0/47)
096.6 (7/106)7.5 (8/107)6.6 (7/106)4.3 (2/47)090.0 (0/106)0.0 (0/107)0.0 (0/106)0.9 (1/108)0.0 (0/47)106.7 (7/105)7.6 (8/105)10.6 (11/104)6.6 (7/106)6.6 (7/106)0.0 (0/112)
104.8 (5/105)3.8 (4/105)4.8 (5/104)12.7 (14/110)103.8 (4/104)6.7 (7/104)5.8 (6/103)3.8 (4/105)3.6 (4/111)110.0 (0/18)0.0 (0/18)0.0 (0/18)0.0 (0/18)0.0 (0/18)6.3 (1/16)
110.0 (0/18)0.0 (0/18)0.0 (0/18)6.3 (1/16)110.0 (0/18)0.0 (0/18)0.0 (0/18)0.0 (0/18)0.0 (0/16)1210.4 (5/48)10.4 (5/48)10.4 (5/48)8.3 (4/48)10.4 (5/48)5.4 (2/37)
124.2 (2/48)4.2 (2/48)4.2 (2/48)5.6 (2/36)122.1 (1/48)0.0 (0/48)0.0 (0/48)0.0 (0/48)8.3 (3/36)1313.5 (7/52)13.7 (7/51)11.5 (6/52)13.2 (5/38)13.6 (6/44)0.0 (0/33)
137.7 (4/52)9.8 (5/51)7.7 (4/52)18.8 (6/32)130.0 (0/52)0.0 (0/51)0.0 (0/52)0.0 (0/44)6.3 (2/32)145.0 (2/40)5.1 (2/39)7.5 (3/40)5.1 (2/39)5.0 (2/40)0.0 (0/25)
140.0 (0/39)2.6 (1/38)0.0 (0/39)8.0 (2/25)142.5 (1/40)2.6 (1/39)2.5 (1/40)2.5 (1/40)0.0 (0/25)1516.2 (6/37)10.8 (4/37)10.8 (4/37)5.4 (2/37)5.4 (2/37)4.4 (3/68)
158.1 (3/37)10.8 (4/37)8.1 (3/37)14.7 (10/68)152.7 (1/37)5.4 (2/37)5.4 (2/37)5.4 (2/37)4.4 (3/68)160.0 (0/16)0.0 (0/15)0.0 (0/15)0.0 (0/16)0.0 (0/16)0.0 (0/7)
1611.8 (2/17)12.5 (2/16)12.5 (2/16)16.7 (1/6)160.0 (0/17)0.0 (0/16)0.0 (0/16)0.0 (0/17)0.0 (0/6)179.3 (4/43)11.6 (5/43)11.6 (5/43)9.3 (4/43)9.8 (4/41)4.2 (1/24)
170.0 (0/43)0.0 (0/43)0.0 (0/43)8.3 (2/24)170.0 (0/43)0.0 (0/43)0.0 (0/43)0.0 (0/41)8.3 (2/24)180.0 (0/6)0.0 (0/6)0.0 (0/6)0.0 (0/5)0.0 (0/6)0.0 (0/1)
180.0 (0/5)0.0 (0/5)0.0 (0/5)0.0 (0/1)180.0 (0/5)0.0 (0/5)0.0 (0/5)0.0 (0/5)0.0 (0/1)1910.0 (1/10)10.0 (1/10)11.1 (1/9)11.1 (1/9)10.0 (1/10)7.7 (1/13)
1911.1 (1/9)11.1 (1/9)12.5 (1/8)7.7 (1/13)190.0 (0/9)0.0 (0/9)0.0 (0/8)0.0 (0/9)15.4 (2/13)20-----0.0 (0/1)
20---0.0 (0/1)200.0 (0/1)211.5 (2/134)3.8 (5/132)2.2 (3/134)3.0 (4/134)1.5 (2/133)0.0 (0/77)
217.5 (10/133)9.9 (13/131)9.8 (13/133)14.3 (11/77)212.2 (3/134)1.5 (2/132)1.5 (2/134)1.5 (2/133)3.9 (3/77)222.9 (1/34)2.9 (1/35)0.0 (0/35)0.0 (0/34)0.0 (0/34)0.0 (0/26)
225.9 (2/34)5.7 (2/35)8.6 (3/35)19.2 (5/26)220.0 (0/34)0.0 (0/35)0.0 (0/35)0.0 (0/34)0.0 (0/26)234.2 (5/118)4.2 (5/120)4.1 (5/121)5.8 (7/121)5.8 (7/120)0.0 (0/45)
237.6 (9/118)9.2 (11/120)10.7 (13/121)4.4 (2/45)233.4 (4/118)3.3 (4/120)3.3 (4/121)3.3 (4/122)6.7 (3/45)245.4 (2/37)5.6 (2/36)8.3 (3/36)5.6 (2/36)5.4 (2/37)8.2 (5/61)
245.3 (2/38)5.4 (2/37)8.1 (3/37)8.2 (5/61)240.0 (0/38)2.7 (1/37)2.7 (1/37)0.0 (0/38)0.0 (0/61)250.0 (0/50)2.0 (1/50)0.0 (0/51)2.2 (1/45)0.0 (0/47)0.0 (0/49)
2512.0 (6/50)12.0 (6/50)15.7 (8/51)4.1 (2/49)250.0 (0/50)0.0 (0/50)0.0 (0/51)0.0 (0/47)0.0 (0/49)266.9 (2/29)6.9 (2/29)6.9 (2/29)7.1 (2/28)7.4 (2/27)0.0 (0/33)
2610.3 (3/29)6.9 (2/29)10.3 (3/29)18.2 (6/33)266.9 (2/29)6.9 (2/29)6.9 (2/29)7.4 (2/27)15.2 (5/33)277.1 (6/84)4.7 (4/85)4.7 (4/85)4.8 (4/84)4.8 (4/83)1.8 (1/56)
276.0 (5/84)4.7 (4/85)4.7 (4/85)7.3 (4/55)271.2 (1/84)2.4 (2/85)2.4 (2/85)1.2 (1/83)1.8 (1/56)2810.2 (23/225)7.4 (17/231)9.5 (22/231)7.8 (18/230)8.3 (19/230)1.4 (2/138)
2812.9 (29/225)13.4 (31/231)13.4 (31/231)14.7 (20/136)282.7 (6/225)2.6 (6/231)2.6 (6/231)2.6 (6/230)7.2 (10/138)293.4 (3/87)3.4 (3/87)4.7 (4/86)3.5 (3/85)2.3 (2/86)0.8 (1/119)
299.2 (8/87)9.2 (8/87)8.1 (7/86)16.8 (20/119)291.1 (1/87)1.1 (1/87)1.2 (1/86)1.2 (1/86)8.4 (10/119)3021.5 (14/65)23.4 (15/64)20.3 (13/64)20.0 (13/65)20.0 (13/65)3.4 (2/59)
3020.0 (13/65)25.0 (16/64)21.9 (14/64)25.4 (15/59)303.1 (2/65)3.1 (2/64)4.7 (3/64)3.1 (2/65)6.8 (4/59)3126.8 (19/71)25.7 (18/70)27.8 (20/72)25.0 (18/72)23.9 (17/71)8.5 (5/59)
3114.1 (10/71)12.9 (9/70)12.5 (9/72)18.6 (11/59)312.8 (2/71)1.4 (1/70)1.4 (1/72)0.0 (0/71)5.1 (3/59)3211.8 (6/51)14.0 (7/50)12.0 (6/50)10.0 (5/50)9.8 (5/51)5.9 (3/51)
329.8 (5/51)10.0 (5/50)10.0 (5/50)19.6 (10/51)320.0 (0/51)2.0 (1/50)0.0 (0/50)0.0 (0/51)2.0 (1/51)3310.6 (5/47)12.8 (6/47)10.6 (5/47)10.6 (5/47)12.8 (6/47)1.4 (1/72)
3327.7 (13/47)29.8 (14/47)29.8 (14/47)26.4 (19/72)336.4 (3/47)8.5 (4/47)6.4 (3/47)6.4 (3/47)6.9 (5/72)All12.6 (380/3003)11.8 (374/3166)12.5 (395/3172)10.1 (323/3195)10.0 (272/2723)3.0 (104/3494)
All6.7 (206/3087)7.1 (225/3163)7.1 (226/3166)8.4 (294/3487)All1.4 (44/3090)1.6 (51/3166)1.5 (48/3170)1.3 (35/2736)3.1 (107/3493)

Table 36 Positivity of CT as Determined by the Alinity m STI Assay by Specimen Type and Clinical Site for Urogenital Specimens

E = Endocervical Swab, SCV = Self-Collected Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, MU = Male Urine

{62}------------------------------------------------

Table 37 Positivity of NG as Determined by the Alinity m STI Assay by Specimen Type and Clinical Site for Urogenital Specimens

E = Endocervical Swab, SCV = Self-Collected Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, MU = Male Urine

{63}------------------------------------------------

Table 38. Positivity of TV as Determined by the Alinity m STI Assay by Specimen Type and Clinical Site for Urogenital Specimens

E = Endocervical Swab, SCV = Self-Collected Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, FU = Female Urine, MU = Male Urine

{64}------------------------------------------------

Collection Site% Positivity (Number Positive / Number Tested with Valid Results)ESCVCCVMU
0117.5 (7/40)22.5 (9/40)23.7 (9/38)29.3 (17/58)
026.5 (36/551)7.1 (40/567)7.8 (44/566)6.7 (60/890)
035.8 (13/224)5.4 (12/221)5.6 (13/231)4.6 (12/259)
0415.0 (3/20)15.0 (3/20)15.0 (3/20)19.1 (22/115)
057.6 (18/236)9.1 (23/252)9.2 (23/251)4.3 (10/233)
060.0 (0/4)0.0 (0/3)0.0 (0/4)12.5 (3/24)
077.1 (28/392)7.4 (32/434)6.0 (26/430)5.1 (29/574)
085.8 (6/103)5.7 (6/105)6.7 (7/104)19.0 (15/79)
0910.5 (11/105)14.3 (15/105)13.2 (14/106)10.4 (5/48)
108.7 (9/104)13.5 (14/104)11.7 (12/103)6.4 (7/110)
115.6 (1/18)11.1 (2/18)5.6 (1/18)25.0 (4/16)
124.2 (2/48)4.2 (2/48)4.2 (2/48)2.8 (1/36)
130.0 (0/52)3.9 (2/51)3.8 (2/52)12.1 (4/33)
147.9 (3/38)10.8 (4/37)10.5 (4/38)0.0 (0/22)
1513.9 (5/36)16.7 (6/36)11.1 (4/36)20.6 (14/68)
165.9 (1/17)0.0 (0/16)6.3 (1/16)0.0 (0/5)
1711.6 (5/43)16.3 (7/43)18.6 (8/43)25.0 (6/24)
180.0 (0/6)0.0 (0/6)0.0 (0/6)0.0 (0/1)
1920.0 (2/10)30.0 (3/10)33.3 (3/9)30.8 (4/13)
20---0.0 (0/1)
217.5 (10/134)9.1 (12/132)7.5 (10/134)5.2 (4/77)
225.7 (2/35)11.1 (4/36)8.3 (3/36)4.0 (1/25)
236.8 (8/117)8.4 (10/119)8.3 (10/120)8.9 (4/45)
2410.5 (4/38)13.5 (5/37)10.8 (4/37)9.8 (6/61)
250.0 (0/49)0.0 (0/49)0.0 (0/50)6.1 (3/49)
266.9 (2/29)6.9 (2/29)6.9 (2/29)6.1 (2/33)
2716.7 (14/84)20.0 (17/85)15.3 (13/85)9.1 (5/55)
286.7 (15/224)9.1 (21/230)9.1 (21/230)7.4 (10/135)
294.6 (4/87)6.9 (6/87)7.0 (6/86)7.6 (9/119)
309.2 (6/65)10.9 (7/64)12.5 (8/64)11.9 (7/59)
3110.0 (7/70)15.9 (11/69)12.7 (9/71)8.5 (5/59)
323.9 (2/51)10.0 (5/50)12.0 (6/50)14.6 (7/48)
3315.2 (7/46)19.6 (9/46)17.4 (8/46)19.2 (14/73)
All7.5 (231/3076)9.2 (289/3149)8.7 (276/3157)8.4 (290/3447)

Table 39. Positivity of MG as Determined by the Alinity m STI Assay by Specimen Type and Clinical Site for Urogenital Specimens

E = Endocervical Swab, SCV = Self-Collected Vaginal Swab, CCV = Clinician-Collected Vaginal Swab, MU = Male Urine

{65}------------------------------------------------

Positive and Negative Predictive Values for Hypothetical Prevalence Rates

The Positive and Negative Predictive Values (PPV and NPV) were calculated using hypothetical prevalence rates and the Alinity m STI Assay sensitivity and specificity determined from the clinical study. Estimates of the PPV and NPV for the Alinity m STI Assay for urogenital specimens are presented in Table 40 through Table 43 for CT,NG, TV, and MG, respectively.

Table 40. CT Positive and Negative Predictive Value Using Hypothetical Prevalence for Urogenital Specimens
SpecimenTypeCategory0.5%1.0%2.0%5.0%10.0%15.0%20.0%25.0%30.0%
CCVPPV (%)36.954.070.385.992.895.396.797.598.0
NPV (%)100.0100.0100.099.999.899.799.599.399.2
SCVPPV (%)40.057.273.087.593.695.997.197.898.3
NPV (%)100.0100.0100.099.999.899.799.699.599.4
EPPV (%)43.260.575.688.994.496.497.498.198.5
NPV (%)100.099.999.999.799.499.098.698.297.7
MUPPV (%)49.466.379.991.195.697.298.098.598.8
NPV (%)100.0100.099.999.999.799.599.399.198.8
NPV (%)100.0100.099.999.999.799.599.399.198.8

CCV = Clinician-Collected Vaginal Swab, SCV = Self-Collected Vaginal Swab, E = Endocervical Swab, MU = Male Urine

Table 41. NG Positive and Negative Value Using Hypothetical Prevalence for Urogenital Specimens
SpecimenTypeCategory0.5%1.0%2.0%5.0%10.0%15.0%20.0%25.0%30.0%
CCVPPV (%)69.281.990.195.998.098.799.199.399.5
NPV (%)100.0100.0100.0100.0100.0100.0100.0100.0100.0
SCVPPV (%)61.175.986.494.397.298.298.799.099.3
NPV (%)100.0100.0100.0100.0100.0100.0100.0100.0100.0
EPPV (%)66.980.389.295.597.898.699.099.399.4
NPV (%)100.099.999.899.699.298.798.297.696.9
PreservCytPPV (%)92.896.398.199.399.699.899.899.999.9
NPV (%)100.099.999.999.799.499.098.698.297.7
MUPPV (%)77.387.393.397.398.799.299.499.699.7
NPV (%)100.0100.0100.0100.0100.0100.0100.0100.0100.0

CCV = Clinician-Collected Vaginal Swab, SCV = Self-Collected Vaginal Swab, E = Endocervical Swab, MU = Male Urine

{66}------------------------------------------------

SpecimenTypeCategory0.5%1.0%2.0%5.0%10.0%15.0%20.0%25.0%30.0%
CCVPPV (%)15.326.742.465.580.086.490.092.393.9
NPV (%)100.0100.0100.0100.0100.099.999.999.999.9
SCVPPV (%)18.731.648.370.683.589.092.093.895.1
NPV (%)100.0100.0100.0100.099.999.999.899.899.7
EPPV (%)13.924.439.562.778.085.088.991.493.2
NPV (%)100.0100.0100.099.999.799.699.499.299.0
PreservCytPPV (%)45.762.877.489.894.996.797.798.298.6
NPV (%)100.0100.099.999.899.599.298.998.598.1
FUPPV (%)34.151.067.884.492.094.896.397.297.8
NPV (%)100.099.999.999.699.298.898.397.797.1
MUPPV (%)39.556.772.687.293.595.897.097.798.2
NPV (%)100.0100.0100.099.999.999.899.799.699.5

Table 42. TV Positive and Negative Value Using Hypothetical Prevalence for Urogenital Specimens


SpecimenTypeCategory0.5%1.0%2.0%5.0%10.0%15.0%20.0%25.0%30.0%
CCVPPV (%)38.355.571.686.793.295.696.997.698.1
NPV (%)100.0100.0100.099.999.899.799.599.499.2
SCVPPV (%)25.240.457.878.088.292.294.495.796.6
NPV (%)100.0100.099.999.899.599.298.898.598.0
EPPV (%)32.949.666.683.791.594.596.197.097.7
NPV (%)99.999.899.699.198.197.095.894.593.1
MUPPV (%)16.428.444.467.381.387.490.792.994.4
NPV (%)100.0100.0100.099.999.899.799.599.499.2

CCV = Clinician-Collected Vaginal Swab, SCV = Self-Collected Vaginal Swab, E = Endocervical Swab, MU = Male Urine

{67}------------------------------------------------

Clinical Study Results - Extragenital Specimens 5.7.2.2

Performance characteristics of the Alinity m STI Assay with extragenital specimens were established in a multicenter clinical study conducted in the United States. Archived specimens were previously collected with an IRB/IEC approved consent. A total of 2,373 male and female, asymptomatic and symptomatic subjects were enrolled. Three oropharyngeal and 3 rectal specimens were collected from each subject using a common collection device created specifically for the study.

Specimen testing methods included the Alinity m STI Assay and 3 commercially available CT/NG NAAT comparator assays for each oropharyngeal and rectal specimen.

Comparator NAAT results were used to establish an anatomic site-specific composite comparator (CC). A specimen was categorized as infected for CT or NG if a minimum of 2 comparator positive results were reported and as not infected for CT or NG if a minimum of 2 comparator negative results was reported. Refer to Table 46 through Table 49 for the CC algorithm. If the specimen CC could not be determined for a given analyte (CT or NG) due to missing and/or indeterminate results from the comparator assays, the specimen was excluded from the analysis for that analyte. CC could not be determined for 6 oropharyngeal and 8 rectal specimens for CT, and 10 oropharyngeal and 6 rectal specimens for NG.

Alinity m STI CT and NG test results were compared to the CC for calculation of assay sensitivity and specificity. A total of 2,316 CT and 2,312 NG results from oropharyngeal specimens, and a total of 2,053 CT and 2,049 NG results from rectal specimens were used in the analysis. The results were analyzed by specimen type and the presence of symptoms. Sensitivity and specificity for CT for oropharyngeal and rectal specimens are presented in Table 44. Sensitivity and specificity for NG for oropharyngeal and rectal specimens are presented in Table 45.

{68}------------------------------------------------

Clinical Sensitivity and Specificity by Specimen Type and Symptom Status for pecimens

Specimen TypeSymptom StatusNTPFPTNFNSensitivity (%)Specificity (%)
Estimate(95% CI)n / NEstimate(95% CI)n / N
OropharyngealSymptomatic741807330100.0 (67.6,100.0)8/8100.0 (99.5,100.0)733/733
OropharyngealAsymptomatic15752021551290.9 (72.2,97.5)20/2299.9 (99.5,100.0)1551/1553
OropharyngealAll23162822284293.3 (78.7, 98.2)28/3099.9 (99.7, 100.0)2284/2286
RectalSymptomatic668552610198.2 (90.6,99.7)55/5699.7 (98.8,99.9)610/612
RectalAsymptomatic13858361289792.2 (84.8,96.2)83/9099.5 (99.0,99.8)1289/1295
RectalAll205313881899894.5 (89.6, 97.2)138/14699.6 (99.2, 99.8)1899/1907

Table 45 NG Clinical Sensitivity and Specificity by Specimen Type and Symptom Status for Extragenital Specimens

Sensitivity (%)Specificity (%)
Estimate(95% CI)n / NEstimate(95% CI)n / N
Specimen TypeSymptom StatusNTPFPTNFN
OropharyngealSymptomatic738536676394.6 (85.4,98.2)53/5699.1 (98.1,99.6)676/682
Asymptomatic15744791516295.9 (86.3,98.9)47/4999.4 (98.9,99.7)1516/1525
All2312100152192595.2 (89.3,98.2)100/10599.3 (98.9,99.6)2192/2207
RectalSymptomatic670467615295.8 (86.0,98.8)46/4898.9 (97.7,99.5)615/622
Asymptomatic13795631319198.2 (90.7,99.7)56/5799.8 (99.3,99.9)1319/1322
All2049102101934397.1 (91.9,99.0)102/10599.5 (99.1, 99.7)1934/1944

{69}------------------------------------------------

A comparison of CC, individual test results from the comparator assays and Alinity m STI Assay was performed. CT results for infected and non-infected oropharyngeal specimens are presented in Table 46, and for infected and non-infected rectal specimens in Table 47. NG results for infected and non-infected oropharyngeal specimens are presented in

CCNAAT1NAAT2NAAT 3Alinity mSTISymptomaticAsymptomaticTotal
Infected++++81725
N/A+++011
-+++022
+++-011
-++-011
Not Infected----69715102207
N/A---101
-N/A--121628
--N/A-7613
+---131831
-+--112
--+-202
---+011
-+-+011

Table 48, and for infected and non-infected rectal specimens in Table 49.

{70}------------------------------------------------

Number of Subjects
CCNAAT1NAAT2NAAT 3Alinity mSTISymptomaticAsymptomaticTotal
Infected++++5079129
N/A+++101
+N/A++101
-+++112
+-++011
++-+224
+++-022
-++-101
+-+-044
++--011
Not Infected----58012611841
N/A---011
-N/A--13821
--N/A-8614
+---4913
-+--145
--+-404
---+213
+--+022
-+-+011
--++022

{71}------------------------------------------------

Number of Subjects
CCNAAT 1NAAT 2NAAT 3Alinity m STISymptomaticAsymptomaticTotal
Infected++++493786
N/A+++101
+N/A++033
-+++022
+-++358
+++-112
+-+-213
Not Infected----64514762121
N/A---202
-N/A--111223
--N/A-6612
+---224
-+--51217
--+-5813
---+347
-N/A-+011
+--+213
-+-+134

Table 48. NG Analysis per Composite Comparator – Oropharyngeal Specimens

{72}------------------------------------------------

Number of Subjects
CCNAAT 1NAAT 2NAAT 3Alinity m STISymptomaticAsymptomaticTotal
Infected++++415394
+N/A++022
+-++415
++-+101
+-+-213
Not Infected----58713031890
-N/A--14721
--N/A-8614
+---303
-+--022
--+-314
---+325
+--+202
--++213

Table 49. NG Analysis per Composite Comparator – Rectal Specimens

{73}------------------------------------------------

Expected Values

The prevalence of CT and NG in this study was dependent on several factors including age, gender, clinic type, presence of symptoms, and the method of testing. A summary of the positivity of CT and NG detection, as determined by the Alinity m STI Assay, is presented by collection site and overall in Table 50.

% Positivity(Number Positive / Number Tested with Valid Results) for CT% Positivity(Number Positive / Number Tested with Valid Results) for NG
CollectionSiteOropharyngealRectalOropharyngealRectal
011.2 (2/170)4.0 (6/149)2.9 (5/170)2.0 (3/149)
021.1 (1/91)3.4 (3/87)5.6 (5/90)3.4 (3/87)
031.0 (4/385)14.1 (46/326)12.6 (48/382)15.4 (50/324)
040.6 (1/168)2.9 (4/140)0.6 (1/168)2.9 (4/138)
050.9 (2/227)6.6 (14/212)7.9 (18/227)7.0 (15/213)
062.4 (10/413)9.6 (34/355)3.1 (13/413)4.2 (15/354)
071.8 (7/396)4.6 (18/393)2.5 (10/396)2.0 (8/394)
080.6 (3/466)5.4 (21/391)3.2 (15/466)3.6 (14/390)
All1.3 (30/2332)7.1 (146/2053)5.0 (115/2312)5.5(112/2049)

{74}------------------------------------------------

Positive and Negative Predictive Values for Hypothetical Prevalence Rates

The Positive and Negative Predictive Values (PPV and NPV) were calculated using hypothetical prevalence rates and the Alinity m STI Assay sensitivity and specificity determined from the clinical study. Estimates of the PPV and NPV for the Alinity m STI Assay for extragenital specimens are presented in Table 51 and Table 52 for CT and NG, respectively.

Table 51. CT Positive and Negative Predictive Value Using Hypothetical Prevalence for Extragenital Specimens
Specimen TypeCategory0.5%1.0%2.0%5.0%10.0%15.0%20.0%25.0%30.0%
OropharyngealPPV (%)84.391.595.698.399.299.599.699.799.8
NPV (%)100.099.999.999.799.398.898.497.897.2
RectalPPV (%)53.169.582.192.296.297.598.398.799.0
NPV (%)100.099.999.999.799.499.098.698.297.7
Table 52. NG Positive and Negative Predictive Value Using Hypothetical Prevalence for Extragenital Specimens
Specimen TypeCategory0.5%1.0%2.0%5.0%10.0%15.0%20.0%25.0%30.0%
OropharyngealPPV (%)41.358.674.188.194.096.197.297.998.4
OropharyngealNPV (%)100.0100.099.999.799.599.298.898.498.0
RectalPPV (%)48.765.679.490.995.597.197.998.498.8
RectalNPV (%)100.0100.099.999.899.799.599.399.198.8

{75}------------------------------------------------

5.8 Conclusions Drawn from the Studies

The analytical and clinical study results demonstrate that the Alinity m STI Assay on the Alinity m System performs comparably to the predicate devices in detecting Chlamydia trachomatis, Neisseria gonorrhea, Trichomonas vaginalis, and Mycoplasma genitalium microorganisms that cause sexually transmitted infections and support a substantial equivalence decision.

§ 866.3393 Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s).

(a)
Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.