K Number
DEN180047
Manufacturer
Date Cleared
2019-01-23

(145 days)

Product Code
Regulation Number
866.3393
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aptima Mycoplasma genitalium assay is an in vitro nucleic acid amplification test (NAAT) for the qualitative detection of ribosomal RNA (rRNA) from Mycoplasma genitalium on the fully automated Panther system. It is intended for use as an aid in the diagnosis of M. genitalium urogenital infections in male and female patients suspected of M. genitalium infection.

The assay may be used to test the following specimens: clinician-collected and selfcollected vaginal swabs (in a clinical setting), clinician-collected endocervical swabs, female and male urine, clinician-collected male urethral swabs, and self-collected penile meatal swabs (in a clinical setting).

For females, a vaginal swab is the preferred specimen type due to higher clinical sensitivity for detecting M. genitalium than other specimen types: however, female urine or clinician-collected endocervical swabs may be used as alternative specimens when vaginal swab specimens are not available. If female urine or clinician-collected endocervical swab specimens test negative, testing with a vaginal swab may be indicated, if M. genitalium infection is suspected.

Device Description

The Aptima Mycoplasma genitalium Assay is a nucleic acid amplification test to qualitatively detect ribosomal RNA from Mycoplasma genitalium. The assay is performed using the Panther System and includes three main processing steps: target capture, transcription-mediated amplification (TMA), and detection by nucleic acid hybridization. The Aptima Mycoplasma genitalium Assay oligonucleotides are designed to specifically capture, amplify, and detect a highly conserved region of Mycoplasma genitalium and the internal control (IC).

The Aptima Mycoplasma genitalium Assay is provided as a 100-test kit. Following are the reagents and materials required to perform the assay on the Panther System:

Aptima Mycoplasma genitalium Assay Kit

  • Amplification Reagent
  • Enzyme Reagent
  • Probe Reagent
  • Internal Control Reagent
  • Selection Reagent
  • Target Capture Reagent
  • Amplification Reconstitution Solution
  • Enzyme Reconstitution Solution
  • Probe Reconstitution Solution

Aptima Mycoplasma genitalium Calibrators Kit

  • Negative Calibrator
  • Positive Calibrator

Materials required but sold separately:

  • Aptima Assay Fluids Kit
  • Aptima Auto Detect Kit
  • Aptima Multitest Swab Specimen Collection Kit
  • Aptima Unisex Swab Specimen Collection Kit for Endocervical and Male Urethral Swab Specimens
  • Aptima Urine Specimen Collection Kit
AI/ML Overview

Aptima Mycoplasma genitalium Assay Study Summary

This document describes the acceptance criteria and a prospective, multi-center clinical study conducted to establish the clinical performance characteristics of the Aptima Mycoplasma genitalium Assay.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are not explicitly stated as quantitative thresholds in the provided text (e.g., "sensitivity must be >X%"). Instead, the document presents detailed performance characteristics obtained from various studies, which are implicitly considered acceptable for the de novo classification. The key performance metrics are summarized below:

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Specificity from PIS Study)
Sensitivity (Female Specimens)High sensitivity desired across specimen types, with vaginal swabs as preferred.Clinician-Collected Vaginal Swab: 92.0% (86.9-95.1%)Patient-Collected Vaginal Swab: 98.9% (95.9-99.7%)Endocervical Swab: 81.5% (75.1-86.6%)Female Urine: 77.8% (71.1-83.3%)
Specificity (Female Specimens)High specificity desired across specimen types.Clinician-Collected Vaginal Swab: 98.0% (97.2-98.6%)Patient-Collected Vaginal Swab: 98.5% (97.7-99.0%)Endocervical Swab: 98.3% (97.5-98.8%)Female Urine: 99.0% (98.3-99.4%)
Sensitivity (Male Specimens)High sensitivity desired across specimen types.Urethral Swab: 98.2% (94.8-99.4%)Penile Meatal Swab: 88.4% (82.6-92.5%)Male Urine: 90.9% (85.5-94.4%)
Specificity (Male Specimens)High specificity desired across specimen types.Urethral Swab: 99.6% (99.1-99.8%)Penile Meatal Swab: 97.8% (96.9-98.5%)Male Urine: 99.4% (98.8-99.7%)
Positive Percent Agreement (PPA) (Specimen-Specific)High PPA desired across specimen types.Clinician-Collected Vaginal Swab: 98.3% (95.2-99.4%)Patient-Collected Vaginal Swab: 98.9% (95.9-99.7%)Endocervical Swab: 97.0% (93.2-98.7%)Female Urine: 94.2% (89.4-96.9%)Urethral Swab: 98.2% (94.8-99.4%)Penile Meatal Swab: 95.3% (91.0-97.6%)Male Urine: 97.5% (93.9-99.0%)
Negative Percent Agreement (NPA) (Specimen-Specific)High NPA desired across specimen types.Clinician-Collected Vaginal Swab: 98.9% (98.3-99.3%)Patient-Collected Vaginal Swab: 98.5% (97.7-99.0%)Endocervical Swab: 99.6% (99.1-99.8%)Female Urine: 99.4% (98.9-99.7%)Urethral Swab: 99.6% (99.1-99.8%)Penile Meatal Swab: 99.0% (98.3-99.4%)Male Urine: 99.9% (99.5-100%)
Limit of Detection (LoD)LoD should be low, indicating high analytical sensitivity.Vaginal Swab: 0.04 - 0.10 GE/mLFemale Urine: 0.04 - 0.12 GE/mLPenile Meatal Swab: 0.05 - 0.10 GE/mLMale Urine: 0.03 - 0.16 GE/mL
Precision/ReproducibilityLow variability across different conditions (instruments, operators, lots, days, runs).%CV for S/CO: 5.58% - 16.21% (total variability for positive samples)% Agreement: 100% for all negative panel members, 100% for positive panel members in reproducibility study.
Analytical Reactivity (Inclusivity)Ability to detect various Mycoplasma genitalium strains.9 Mycoplasma genitalium strains detected at ≥95% positivity, with most at low GE/mL concentrations.
Analytical Specificity (Cross-reactivity)No false positives with common genitourinary flora and related microorganisms.No false positive test results with any of 56 tested microorganisms. In silico analysis showed no significant interactions.
Microbial InterferenceNo, or minimal, interference from other microorganisms.No microbial interference observed, except with Mycoplasma pneumoniae (a limitation noted in the package insert).
Interfering SubstancesNo interference from common gynecological/feminine hygiene products, bodily fluids.No interference observed with listed interfering substances at specified concentrations, except with mucus at 0.3% w/v (a limitation noted in the package insert).
Carryover ContaminationNo carryover from high positive to negative samples.No carryover observed in checkerboard study.

2. Sample Size and Data Provenance (Clinical Study)

  • Sample Size (Test Set):
    • Subjects: 3300 evaluable subjects (1737 women, 1563 men).
    • Specimens with Valid Aptima Results: 11,774.
    • Specimens with Valid PIS Results (used for Sensitivity/Specificity): 11,557.
    • Specimens with Specimen-Specific Composite Comparator Results (used for PPA/NPA): 11,665.
  • Data Provenance: Prospective, multi-center clinical study conducted at 21 US sites. Testing was performed at three sites.

3. Number of Experts and Qualifications for Ground Truth (Clinical Study)

The ground truth ("patient infected status" - PIS) was established using a "composite comparator algorithm comprised of three validated transcription mediated amplification (TMA) based assays for Mycoplasma genitalium." The document specifies that these comparator TMA assays were developed by Hologic (the applicant), as shown in section "O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: Comparator TMA Assays validation."

Therefore, the ground truth was not established by human experts but by a composite of validated laboratory tests.

4. Adjudication Method (Clinical Study)

The PIS was determined using a composite comparator algorithm of three validated TMA assays.

  • Method: Two of the three comparator TMA assays had to be positive or negative to establish the infected or not-infected PIS, respectively.
  • Tie-breaker: The third TMA assay was run as a tie-breaker if the results of the first two TMA assays were discordant.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. This study evaluated the performance of a diagnostic device (the Aptima Mycoplasma genitalium Assay) against a composite reference standard (PIS), not the comparative effectiveness of human readers with or without AI assistance.

6. Standalone Performance Study

Yes, the study described is a standalone (algorithm only without human-in-the-loop performance) study. The Aptima Mycoplasma genitalium Assay is an in vitro diagnostic device (NAAT) whose performance characteristics (sensitivity, specificity, PPA, NPA) were directly evaluated against a composite reference standard. There is no indication of human interpretation or assistance being part of the device's output or the evaluation process.

7. Type of Ground Truth Used (Clinical Study)

The type of ground truth used was a composite comparator algorithm (laboratory-based), also referred to as "patient infected status (PIS)." This PIS was determined by the concordance of three validated transcription-mediated amplification (TMA) assays for Mycoplasma genitalium.

8. Sample Size for the Training Set

The document does not specify the sample size for a "training set." This is an in vitro diagnostic assay, and while it involves design and development, the primary clinical evaluation described is a prospective validation study. The analytical performance evaluations (e.g., LoD, inclusivity, cross-reactivity) involve testing various samples, but these are generally considered part of the verification and validation of the assay, not a distinct "training set" in the machine learning sense. The clinical study is a test set evaluation.

9. How the Ground Truth for the Training Set Was Established

As noted above, a distinct "training set" with independent ground truth establishment is not described in the context of this diagnostic assay's evaluation. The analytical studies establishing LoD, inclusivity, and specificity use characterized samples (e.g., spiked samples with known concentrations of M. genitalium or other microorganisms), which serve as their own form of "ground truth" for those specific analytical characteristics.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Aptima Mycoplasma genitalium Assay

DECISION SUMMARY

A. DEN Number:

DEN180047

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the Aptima Mycoplasma genitalium Assay.

C. Measurands:

Ribosomal RNA from Mycoplasma genitalium

D. Type of Test:

Nucleic acid amplification test (NAAT)

E. Applicant:

Hologic, Inc.

F. Proprietary and Established Names:

Aptima Mycoplasma genitalium Assay

G. Regulatory Information:

    1. Regulation section: 21 CFR 866.3393
    1. Classification: Class II (Special Controls)
    1. Product code: QEP
    1. Panel: 83 - Microbiology

H. Indications For Use:

{1}------------------------------------------------

1. Indications for use:

The Aptima Mycoplasma genitalium assay is an in vitro nucleic acid amplification test (NAAT) for the qualitative detection of ribosomal RNA (rRNA) from Mycoplasma genitalium on the fully automated Panther system. It is intended for use as an aid in the diagnosis of M. genitalium urogenital infections in male and female patients suspected of M. genitalium infection.

The assay may be used to test the following specimens: clinician-collected and selfcollected vaginal swabs (in a clinical setting), clinician-collected endocervical swabs, female and male urine, clinician-collected male urethral swabs, and self-collected penile meatal swabs (in a clinical setting).

For females, a vaginal swab is the preferred specimen type due to higher clinical sensitivity for detecting M. genitalium than other specimen types: however, female urine or clinician-collected endocervical swabs may be used as alternative specimens when vaginal swab specimens are not available. If female urine or clinician-collected endocervical swab specimens test negative, testing with a vaginal swab may be indicated, if M. genitalium infection is suspected.

    1. Special conditions for use statement(s):
      For in vitro diagnostic use

For prescription use only

3. Special instrument requirements:

The Aptima Mycoplasma genitalium Assay is performed on the Panther System.

I. Device Description:

The Aptima Mycoplasma genitalium Assay is a nucleic acid amplification test to qualitatively detect ribosomal RNA from Mycoplasma genitalium. The assay is performed using the Panther System and includes three main processing steps: target capture, transcription-mediated amplification (TMA), and detection by nucleic acid hybridization. The Aptima Mycoplasma genitalium Assay oligonucleotides are designed to specifically capture, amplify, and detect a highly conserved region of Mycoplasma genitalium and the internal control (IC).

The Aptima Mycoplasma genitalium Assay is provided as a 100-test kit. Following are the reagents and materials required to perform the assay on the Panther System:

Aptima Mycoplasma genitalium Assay Kit

  • Amplification Reagent ●
  • Enzyme Reagent
  • Probe Reagent ●
  • Internal Control Reagent ●

{2}------------------------------------------------

  • Selection Reagent
  • Target Capture Reagent
  • Amplification Reconstitution Solution
  • Enzyme Reconstitution Solution ●
  • Probe Reconstitution Solution

Aptima Mycoplasma genitalium Calibrators Kit

  • Negative Calibrator ●
  • . Positive Calibrator

Materials required but sold separately:

  • Aptima Assay Fluids Kit ●
  • Aptima Auto Detect Kit
  • Aptima Multitest Swab Specimen Collection Kit ●
  • Aptima Unisex Swab Specimen Collection Kit for Endocervical and Male Urethral ● Swab Specimens
  • Aptima Urine Specimen Collection Kit ●

J. Standard/Guidance Document Referenced (if applicable):

  • . EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures.
  • FDA Guidance Design Considerations for Pivotal Clinical Investigations for ● Medical Devices, 2013.

K. Test Principle:

The Aptima Mycoplasma genitalium Assay is comprised of three main processing steps that take place in a single tube on the Panther System: target capture, transcription-mediated amplification (TMA), and hybridization protection assay (HPA).

A collected specimen is transferred into an appropriate specimen transport tube containing transport solution that helps release the rRNA target and protects it from degradation during storage. The target rRNA, if present, is isolated by the use of a specific capture oligomer and magnetic microparticles in a method called target capture. During this step, the sequencespecific region of the capture oligomer binds to a specific region of the target molecule. The resulting capture oligomer:target complex is then captured out of solution by decreasing the temperature of the reaction to room temperature, which allows hybridization to occur between the deoxyadenosine region on the capture oligomer and the poly-deoxythymidine molecules that are covalently attached to the magnetic particles. The microparticles are then captured using magnets and are washed to remove residual specimen matrix that may contain amplification inhibitors. The target capture process isolates and purifies the rRNA from clinical specimens, which is then ready for amplification.

The TMA reaction amplifies a specific region of the small ribosomal subunit from Mycoplasma genitalium via DNA and RNA intermediates and generates multiple copies of

{3}------------------------------------------------

RNA amplicon.

Detection of the amplicon is achieved by HPA. During this step, a single-stranded chemiluminescent DNA probe labeled with an acridinium ester molecule combines with amplicon to form stable RNA:DNA hybrids. The light emitted from the labeled RNA:DNA hybrids is measured as photon signals called relative light units (RLU) in a luminometer. Assay results for each patient sample are determined based on the analyte signal-to-cutoff (S/CO) ratio.

Internal Control (IC) is added to each reaction via the Target Capture Reagent. The IC monitors nucleic acid capture, amplification, detection, and operator or instrument error.

L. Performance Characteristics:

1. Analytical performance:

a. Precision/Reproducibility:

Within Laboratory Precision:

The within laboratory precision study was conducted with two Panther Systems, two operators, and three lots of reagents over 12 days. The precision panel was prepared using two specimen matrices, urine and simulated vaginal swab samples, and included Mycoplasma genitalium negative, low positive, and moderate positive samples. The low and moderate positive panel members were prepared by spiking Mycoplasma genitalium into two specimen matrices at 1.5X Limit of Detection (LoD) and 3X LoD, respectively. Two operators tested five replicates of each panel member. The variability between instruments, operators, reagent lots, days, and between and within runs is shown in Table 1 below. Percent coefficient of variation (%CV) was calculated for all replicates based on S/CO values. Percent agreement was calculated based on the expected result (detected or not detected) for each panel member. Testing of all precision panel members generated acceptable results.

BetweenInstrumentsBetweenOperatorsBetweenLotsBetweenDaysBetweenRunsWithinRunsTotal
PanelN%Detected¹MeanS/COSD%CVSD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
NegativeUrine:UTM240100²0.00.00NC0.00NC0.00NC0.00NC0.00NC0.00NC0.00NC
1.5X LoDUrine:UTM24010024.70.000.000.000.000.000.000.200.810.471.882.088.432.148.68
3X LoDUrine:UTM24010025.20.080.330.160.620.000.000.271.050.190.761.365.381.415.58
NegativeSVM240100²0.00.00NC0.00NC0.00NC0.00NC0.00NC0.00NC0.00NC
1.5X LoDSVM24098.323.70.000.000.000.000.361.510.130.540.000.003.8316.133.8416.21

Table 1: Precision of the Aptima Mycoplasma genitalium Assay

{4}------------------------------------------------

3X LoD24010024.80.140.580.451.840.000.000.461.870.632.561.184.751.496.02
SVM

CV = coefficient of variation, LoD, limit of detection, S/CO = standard deviation, SVM = simulated vaginal matrix, UTM = urine transport medium.

1 Detected defined as S/CO > 1.0.

2 100% Mycoplasma genitalium negative.

Reproducibility:

The Aptima Mycoplasma genitalium Assay reproducibility was evaluated at three U.S. sites, by two operators at each site. Each operator performed one run per day over five days. Each run tested three replicates of each reproducibility panel member. Each testing site was provided with a 6-member reproducibility panel that included four panel members positive for Mycoplasma genitalium and two panel members negative for Mycoplasma genitalium. Positive panel members were created by spiking Mycoplasma genitalium into two different specimen matices: urine transport medium (UTM) and simulated vaginal swab matrix.

The variability between sites, between operators, between days, between runs, and within runs was calculated separately for each panel member and is presented in Table 2 below.

Between SitesBetween OperatorsBetween DaysBetween RunsWithin RunsTotal
PanelNMean S/COSD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
UTM Negative900.000.00NC0.00NC0.00NC0.00NC0.00NC0.00NC
UTM Low Pos9024.640.451.820.000.000.431.740.431.742.389.672.5910.51
UTM Mod Pos9025.540.000.000.000.000.000.000.000.001.204.711.415.51
SVM Negative900.00*0.00NC0.00NC0.00NC0.00NC0.00NC0.00NC
SVM Low Pos9024.050.000.000.481.980.000.000.000.001.857.672.128.83
SVM Mod Pos9025.140.000.000.481.910.562.250.562.251.144.531.656.58

Table 2: Reproducibility of the Aptima Mycoplasma genitalium Assay

CV = coefficient of variation, Mod = moderate, NC = not calculable, Pos = positive, S/CO = signal to cutoff, SD = standard deviation, SVM = simulated vaginal matrix, UTM = urine transport medium.

Note: In case variability from some factors may be numerically negative, SD and CV are shown as 0.00.

  • 1.1% (1 out of 90) results had S/CO value of 0.03 and 98.9% (89 out of 90) results had S/CO value of 0.

Agreement with the expected result, along with the corresponding 2-sided 95% score confidence interval, was also calculated for each panel member; agreement with the expected results was 100% (95% CI: 95.9-100) for all panel members.

  • b. Linearity/assay reportable range:

{5}------------------------------------------------

Not Applicable

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.

Internal Control:

An internal control consisting of a non-infectious RNA transcript is added to each reaction via the Target Capture Reagent. The IC monitors nucleic acid capture, amplification, detection, and operator or instrument error. During sample processing, IC acceptance criteria are automatically verified by the Panther System software. If an IC result is invalid, the sample result is invalidated. Every sample with an invalid IC result is retested.

External Controls:

External controls are not provided with the Aptima Mycoplasma genitalium Assay; however, testing of external positive and negative controls are recommended in the assay labeling. External controls should be tested in conformance with local, state, and/or federal regulations or accreditation requirements and each laboratory's standard quality control procedures.

Calibrators:

A positive and negative calibrator are run in duplicate each time a reagent kit is loaded on the Panther System. Assay calibration is required to generate valid results. The Aptima Mycoplasma genitalium Assay calibration is valid for up to 48 hours. Software on the Panther System alerts the operator when a new calibrator set should be run.

During sample processing, criteria for acceptance of the calibrator are automatically verified by the software on the Panther System. If two replicates are invalid for either the positive or negative calibrator, the software automatically invalidates the run. Samples in an invalidated run are retested using a freshly prepared set of calibrators.

Specimen Stability:

The specimen stability study indicated that it is acceptable to store swab and urine specimens at the following conditions prior to testing:

Swab specimens

  • . After collection, swab specimens in transport tubes can be stored at 2℃ to 30℃ for up to 60 days.
  • If longer storage is needed, swab specimens in transport tubes can be stored . at -20°C or -70°C for up to an additional 90 days.

Urine specimens

  • After collection, urine specimens in the primary collection container can be . stored at 2℃ to 30℃ for up to 24 hours before urine is transferred to the transport tube.
  • Processed urine in the transport tube can be stored at 2℃ to 30℃ for up ●

{6}------------------------------------------------

to 30 days (after transfer).

  • If longer storage is needed, processed urine in the transport tube can be . stored at -20°C or -70°C for up to an additional 90 days (after transfer).
    The specimen stability study evaluated vaginal swabs, penile meatal swabs, and male and female urine. For vaginal and penile meatal swabs. 20 pools of each of the swab types were spiked with Mycoplasma genitalium at 3X LoD and stored at the recommended storage conditions. For male and female urine. 20 pools of male urine and 20 pools of female urine, initially spiked with Mycoplasma genitalium at ~90 genome equivalents (GE)/mL, were diluted to < 2 GE/mL and tested at the recommended storage conditions. An aliquot from each pool was removed from the storage at specific time intervals and nine replicates of each pool (per temperature condition) were tested. An additional female urine stability study provided the support for longer storage, i.e., storage of female urine at -20°C or -70°C for up to an additional 90 days (following 31 days storage at 30°C). The concentration tested in this additional study was higher than the analytical LoD: however, the tested concentration is at the low range of the Mycoplasma genitalium concentration observed in the clinical samples.

  • d. Detection limit:
    The Limit of Detection (LoD), is the concentration at which the Aptima Mycoplasma genitalium Assay has positive results at least 95% of the time. The LoD of the Aptima Mycoplasma genitalium Assay was determined in pooled male urine, female urine, vaginal swab, and penile meatal swab specimens. Dilutions of two strains of Mycoplasma genitalium were tested in the study with at least (0) eplicates per concentration, per reagent lot, using two reagent lots for a tot 14 f at least (b) (4) replicates per strain. Testing was performed on two Panther Systems. The results of the LoD study are summarized below in Table 3:

Mycoplasma genitalium LoD (GE/mL)
Specimen TypeStrain 1Strain 2
Vaginal Swab0.040.10
Female Urine0.040.12
Penile meatal swab0.050.10
Male Urine0.030.16

Table 3: Limit of Detection of the Aptima Mycoplasma genitalium Assay

e. Analytical Reactivity (Inclusivity):

A study was performed to verify that the Aptima Mycoplasma genitalium Assay is able to detect various Mycoplasma genitalium strains. Nine Mycoplasma genitalium strains (b) (4)(b) (4)(b) (4)(b) (4)(b) (4)(b) (4)(b) (4)(b) (4)(b) (4) (b) (4)(b) (4)(b) (4) were evaluated in male and female urine, vaginal swabs, and penile meatal swabs in triplicates. Seven of the nine strains were detected at ≥95% positivity at <0.29-0.49 GE/mL in all four specimen types. One strain was

{7}------------------------------------------------

95% positive at 0.85-1.46 GE/mL in each of the four specimen types. The remaining strain was detected at 100% positivity at 1.16 and 1.46 GE/mL in vaginal and penile meatal swabs respectively, 100% positivity at 3.47 GE/mL in female urine, and 100% positivity at 8.50 GE/mL in male urine.

Analytical specificity: f.

Cross-reactivity

Cross-reactivity study was performed using a panel of 56 microorganisms, including common flora of the genitourinary tract and closely related microorganisms. The microorganisms were spiked into 3 clinical matrices: vaginal swabs, penile-meatal swabs, and female urine. Microorganisms were tested at 2.5x105 TCID50/mL for viruses, 1x10 CFU/mL for fungi and bacteria, 1x10 cells/mL for protozoa, or the highest concentration possible if these concentrations could not be achieved. All microorganisms were tested with three lots of reagents with three replicates per lot of reagent. The Aptima Mycoplasma genitalium Assay did not generate a false positive test result with any microorganism listed in Table 4.

In addition, an in silico analysis was performed to determine if the oligonucleotides of the Aptima Mycoplasma genitalium Assay could amplify and detect nucleic acid sequences from Human papillomavirus (HPV) type 31, HPV type 35, HPV type 54, Mycobacterium smegmatis. Chlamydia trachomatis serovars L1, L2, L3, and Treponema pallidum. No significant interactions were detected in the in silico analysis.

MicroorganismTest concentrationMicroorganismTest concentration
Acinetobacter lwoffii1x106CFU/mLHPV type 18 (HeLa cells)1x104 cells/mL
Actinomyces israelii1x106 CFU/mLHPV type 581x104 copies/mL
Alcaligenes faecalis1x106 CFU/mLHPV type 391x104 copies/mL
Atopobium vaginae1x109 rRNAcopies/mLHPV type 511x104 copies/mL
Bacteroides fragilis1x106 CFU/mLKlebsiella pneumoniae1x106 CFU/mL
Bifidobacteriumadolescentis1x106 CFU/mLLactobacillus acidophilus1x106 CFU/mL
Campylobacter jejuni1x106 CFU/mLLactobacillus crispatus1x106 CFU/mL
Candida albicans1x106 CFU/mLLeptotrichia buccalis1x105 CFU/mL
Chlamydia trachomatis1x104 IFU/mLListeria monocytogenes1x106 CFU/mL
Clostridium difficile1x106 CFU/mLMegasphaera type 11x109 copies/mL
Chromobacteriumviolaceum1x106 CFU/mLMobiluncus curtisii1x106 CFU/mL
Corynebacteriumgenitalium1x106 CFU/mLMobiluncus mulieris1x106 CFU/mL
Cryptococcusneoformans1x106 CFU/mLMycoplasma hominis1x109 copies/mL
Cytomegalovirus2.5 x104 TCID50/mLMycoplasma pneumoniae1x106 CFU/mL
Elizabethkingiameningosepticum1x106 CFU/mLNeisseria gonorrhoeae1x106 CFU/mL
Enterobacter cloacae1x106 CFU/mLPentatrichomonas hominis1x105 cells/mL
Enterococcus faecalis1x106 CFU/mLPrevotella bivia1x106 CFU/mL

Table 4: Microorganisms Tested in the Aptima Mycoplasma genitalium Assay

{8}------------------------------------------------

Escherichia coli1x106 CFU/mLPropionibacterium acnes1x106 cells/mL
Finegoldia magna1x106 copies/mLProteus vulgaris1x106 CFU/mL
Fusobacteriumnucleatum1x106 CFU/mLPseudomonas aeruginosa1x106 CFU/mL
Gardnerella vaginalis1x106 CFU/mLStaphylococcus aureus1x106 CFU/mL
Haemophilus ducreyi1x106 CFU/mLStaphylococcusepidermidis1x106 CFU/mL
Herpes simplex virustype 12.5 x103 TCID50/mLStaphylococcussaprophyticus1x106 CFU/mL
Herpes simplex virustype 22.5 x103 TCID50/mLStreptococcus agalactiae1x106 CFU/mL
HIV-11x106 copies/mLStreptococcus pyogenes1x106 CFU/mL
HPV type 61x106 copies/mLTrichomonas vaginalis1x105 cells/mL
HPV type 111x108 copies/mLUreaplasma parvum1x109 rRNA copies/mL
HPV type 16 (SiHa cells)1x104 cells/mLUreaplasma urealyticum1x109 rRNA copies/mL

Microbial Interference

The same organisms that were tested in the cross reactivity study were evaluated in the microbial interference study. Each sample tested contained Mycoplasma genitalium at a concentration of 3X LoD and microorganism concentration indicated above in Table 4. Once prepared, each sample was tested in triplicate. No microbial interference with the Aptima Mycoplasma genitalium Assay was observed, except in the presence of Mycoplasma pneumoniae, which is commonly found in the lower respiratory tract. A limitation has been added to the Aptima Mycoplasma genitalium Assay's package insert to mitigate this finding.

Interfering Substances

A study was performed to evaluate the performance of the Aptima Mycoplasma genitalium Assay in the presence of gynecological products, feminine hygiene products, whole blood, leukocytes, and other potentially interfering substances. Assay results were evaluated to determine if the presence of potentially interfering substances in Mycoplasma genitalium-negative or Mycoplasma genitalium-positive samples had an effect on the assay performance. Potentially interfering substances were diluted into relevant corresponding clinical matrices (penile meatal swab, vaginal swab, and female urine) and tested in the absence or presence of Mycoplasma genitalium at 3X LoD. A minimum of nine replicates of each panel member were tested with three reagent lots (three replicates per reagent lot). No interference was observed with the interfering substances at the concentration listed in Table 5 (i.e., all positive replicates tested produced positive Aptima Mycoplasma genitalium Assay results and all negative replicates tested produced negative Aptima Mycoplasma genitalium Assay results).

{9}------------------------------------------------

AssayPotential Interfering SubstanceTarget Concentration
(b) (4)

Table 5: Interfering Substances Tested in the Aptima Mycoplasma genitalium Assav

Interference in assay results was observed when mucus at a final concentration of 0.3% w/v was added to clinical specimen matrix. Interference was not observed when mucus at a final concentration of 0.03% w/v was added to clinical specimen matrix. A limitation has been added to the Aptima Mycoplasma genitalium Assay's package insert to reflect this finding.

  • g. Carryover Contamination:
    A series of alternating Mycoplasma genitalium negative and positive samples were processed using the Aptima Mycoplasma genitalium Assay. The positive samples consisted of simulated vaginal swab samples spiked with 6.1x106 GE/mL of Mycoplasma genitalium and the negative samples consisted of simulated vaginal swab samples without Mycoplasma genitalium. The checkerboard arrangement was tested on three Panther instruments, four runs/instrument, 40 negative, and 40 positive samples per run with one reagent lot. All samples tested produced expected results with all positive samples producing positive Aptima Mycoplasma genitalium Assay results, and all negative samples producing negative Aptima Mycoplasma genitalium Assay

{10}------------------------------------------------

results. No carryover was observed.

  • h. Assay cut-off:
    Assay result from each patient is determined by a cutoff based on the total RLU. A test may be negative, positive, or invalid, as determined by the IC RLU and Signal to Cutoff (S/CO) ratio for the detection of Mycoplasma genitalium rRNA. Table 6 below describes the criteria associated with each possible assay result.
Assay ResultCriteria
NegativeAnalyte S/CO < 1.0IC ≥ IC CutoffIC ≤ 1,200,000 RLU
PositiveAnalyte S/CO ≥ 1.0IC ≤ 1,200,000 RLUAnalyte ≤ 3,000,000 RLU
InvalidAnalyte S/CO < 1.0 and IC < IC CutoffOrIC > 1,200,000 RLUOrAnalyte > 3,000,000 RLU

Table 6: Result Internretation

    1. Comparison studies:
    • a. Method comparison with predicate device:

Not Applicable

  • b. Matrix comparison:
    Simulated Vaginal Swab Matrix Equivalency Study:

A study was performed to demonstrate matrix equivalency between simulated vaginal swab matrix and natural vaginal swab matrix via a head-to-head LoD study. Selfcollected natural vaginal swab samples were collected using the Aptima Multi-test Swab Collection Kit. The simulated vaginal swab matrix consists of Specimen Transport Media (STM) combined with a simulated vaginal fluid (SVF).

The LoD was determined by testing eight dilutions of Mycoplasma genitalium Ivsate prepared in both the natural vaginal swab and simulated vaginal swab matrices. A minimum of 20 replicates were tested for each panel member across three reagent lots. Probit analysis was used to determine the 95% LoD for Mycoplasma genitalium in each sample matrix. The LoDs of the Aptima Mycoplasma genitalium Assay for Mycoplasma genitalium in simulated vaginal swab matrix and natural vaginal swab matrix were found to be similar (~0.04 GE/mL).

{11}------------------------------------------------

3. Clinical studies:

Prospective Study:

A prospective, multi-center clinical study was conducted to establish the clinical performance characteristics of the Aptima Mycoplasma genitalium Assay on the Panther System. The clinical samples were collected at 21 US sites with testing performed at three sites. A total of 3393 subjects were enrolled, of which 93 were excluded (subjects withdrawn and unknown patient infected status). Of the 3300 remaining subjects, 1737 were women and 1563 were men. Four subjects were 15 to 17 years of age, 242 were 18 to 20 years of age, 483 were 21 to 24 years of age, 1954 were 25 to 44 years of age, 572 were 45 to 64 years of age, and 45 were ≥65 years of age. The study included collection of the following specimens in a clinical setting from symptomatic and asymptomatic subjects:

From males: self-collected penile meatal swab, clinician-collected urethral swab, and self-collected first catch urine.

From females: Self-collected first catch urine, self-collected vaginal swab, cliniciancollected vaginal swab, and clinician collected endocervical swab.

The clinical performance of the Aptima Mycoplasma genitalium Assay was evaluated against patient infected status (PIS) determined using a composite comparator algorithm comprised of three validated transcription mediated amplification (TMA) based assays for Mycoplasma genitalium (please see section O for validation of the comparator TMA assays). The results of the comparator TMA assays from self-collected vaginal swabs and male urethral swabs were used to define the PIS. Two of the three comparator TMA assays had to be positive or negative to establish the infected or not-infected PIS respectively. The third TMA assay was run, as a tie-breaker, if the results of the first two TMA assays were discordant.

Of the specimens collected, 11,827 were tested in valid Aptima Mycoplasma genitalium Assay runs. Of these, 11,774 (99.6%) had final valid Aptima Mycoplasma genitalium Assay results and 53 (0.4%) had final invalid results and were excluded from the analysis. Of 11,774 specimens, 11,557 had valid PIS results. Therefore, for the 3300 evaluable subjects, the following specimens were included in the analyses: 1709 clinician-collected vaginal swab, 1724 self-collected vaginal swab, 1715 endocervical swab, 1733 female urine, 1563 urethral swab, 1554 self-collected penile meatal swab, and 1559 male urine samples.

Performance in Female and Male Specimens based on PIS:

The sensitivity, specificity, PPV, and NPV of the Aptima Mycoplasma genitalium assay for Mycoplasma genitalium detection are shown for female and male specimens in Tables 7 and 8.

{12}------------------------------------------------

SpecimenTypeNTPFPTNFNPrev(%)Sensitivity %(95% CI) ¹Specificity %(95% CI) ¹PPV %(95% CI) ²NPV %(95% CI) ²
CVS17091603015051410.292.0(86.9-95.1)98.0(97.2-98.6)84.2(79.1-88.6)99.1(98.5-99.5)
PVS1724173241525210.298.9(95.9-99.7)98.5(97.7-99.0)87.8(83.1-91.7)99.9(99.5-100)
ES17151412615653210.181.5(75.1-86.6)98.3(97.5-98.8)84.4(78.9-89.1)97.9(97.2-98.5)
FU17331371615413910.277.8(71.1-83.3)99.0(98.3-99.4)89.5(84.3-93.6)97.5(96.8-98.2)
US156316261392310.698.2(94.8-99.4)99.6(99.1-99.8)96.4(92.7-98.6)99.8(99.4-100)
PM15541453013601910.688.4(82.6-92.5)97.8(96.9-98.5)82.9(77.4-87.6)98.6(97.9-99.1)
MU1559149913861510.590.9(85.5-94.4)99.4(98.8-99.7)94.3(90.0-97.2)98.9(98.3-99.4)

Table 7: Performance Characteristics Based on PIS

CI = confidence interval, CVS = clinician-collected vaginal swab, ES = endocervical swab, FN = false negative, FP = false positive, FU = female urine, MU = male urine, NPV = negative predictive value, PM = penile meatal swab, PPV = positive predictive value, Prev = prevalence, PVS = patient-collected vaginal swab, TN = true negative, TP = true positive, US = male urethral swab.

1 Score CI.

2 PPV 95% CI computed from the exact 95% CI for the positive likelihood ratio, NPV 95% CI computed from the exact 95% CI for the negative likelihood ratio.

SpecimenTypeSymptomStatusNTPFPTNFNPrev(%)Sensitivity %(95% CI) 1Specificity %(95% CI) 1PPV %(95% CI) 2NPV %(95% CI)2
CVSSym104011222898811.593.3(87.4-96.6)97.6(96.4-98.4)83.6(77.3-88.8)99.1(98.3-99.6)
Asym66948860768.188.9(77.8-94.8)98.7(97.5-99.3)85.7(75.8-92.9)99.0(98.0-99.6)
PVSSym104712118908011.6100(96.9-100)98.1(96.9-98.8)87.1(81.1-91.9)100(99.6-100)
Asym67752661728.096.3(87.5-99.0)99.0(97.9-99.6)89.7(80.4-95.7)99.7(98.9-100)
ESSym1046101179091911.584.2(76.6-89.6)98.2(97.1-98.9)85.6(79.1-90.8)98.0(97.0-98.7)
Asym669409607137.975.5(62.4-85.1)98.5(97.2-99.2)81.6(70.3-90.2)97.9(96.8-98.8)
FUSym105197159142511.679.5(71.5-85.7)98.4(97.4-99.0)86.6(80.0-91.8)97.3(96.3-98.2)
Asym682401627147.974.1(61.1-83.9)99.8(99.1-100)97.6(88.7-99.9)97.8(96.7-98.7)
USSym8661021761212.098.1(93.3-99.5)99.9(99.3-100)99.0(94.9-100)99.7(99.1-100)
Asym69760563118.898.4(91.3-99.7)99.2(98.2-99.7)92.3(84.0-97.3)99.8(99.2-100)

Table 8: Performance Characteristics Based on PIS by Symptom Status

{13}------------------------------------------------

PMSym86592177451111.989.3(81.9-93.9)97.8(96.5-98.6)84.4(77.5-90.0)98.5(97.6-99.2)
Asym689531361588.986.9(76.2-93.2)97.9(96.5-98.8)80.3(70.8-88.1)98.7(97.7-99.4)
MUSym8669377551112.089.4(82.0-94.0)99.1(98.1-99.6)93.0(86.9-96.9)98.6(97.6-99.3)
Asym69356263148.793.3(84.1-97.4)99.7(98.9-99.9)96.6(89.0-99.5)99.4(98.5-99.8)

Asym = asymptomatic, CI = confidence interval, CVS = clinician-collected vaginal swab, ES = endocervical swab, FN = false negative, FP = false positive, PM = penile meatal swab, PV=positive value, Prev=prevalence, PVS=Patient-collected vaginal swab, Sym=symptomatic, TN=true positive, US=male urethral swab 1 Score CI

2 PPV 95% CI computed from the exact 95% CI for the positive likelihood ratio, NPV 95% CI computed from the exact 95% CI for the negative likelihood ratio.

Specimen-specific Agreements:

Specimen-specific agreements were calculated by comparing the Aptima Mycoplasma genitalium Assay results and the composite comparator results, comprised of testing results from the earlier mentioned three validated TMA assays for the same specimen type. The sample was considered positive when two out of three TMA assay results were positive; the sample was considered negative when two out of three TMA assay results were negative. Of the 11.827 specimens from 3300 evaluable subjects. 11.665 specimens that had specimen-specific composite comparator results available were included in the specimen-specific agreement analysis. The positive (PPA) and negative (NPA) percent agreement of the Aptima Mycoplasma genitalium Assay for Mycoplasma genitalium detection are shown for female and male specimens in Tables 9 and 10.

SpecimenTypeNCompositecomparator+/Aptima+Compositecomparator-/Aptima+Compositecomparator-/Aptima-Compositecomparator+/Aptima-PPA(95% CI)¹NPA(95% CI)¹
CVS1729175171534398.3 (95.2-99.4)98.9 (98.3-99.3)
PVS1724173241525298.9 (95.9-99.7)98.5 (97.7-99.0)
ES173416371559597.0 (93.2-98.7)99.6 (99.1-99.8)
FU177414791609994.2 (89.4-96.9)99.4 (98.9-99.7)
US156316261392398.2 (94.8-99.4)99.6 (99.1-99.8)
PM1563162141379895.3 (91.0-97.6)99.0 (98.3-99.4)
MU157815921413497.5 (93.9-99.0)99.9 (99.5-100)

Table 9: Specimen-Specific Agreements

CI = confidence interval, CVS = clinician-collected vaginal swab, ES = endocervical swab, FU = female urine, MU = male urine, NPA = negative percent agreement, PM = penile meatal swab, PPA = positive percent agreement, PVS = patientcollected vaginal swab, US = male urethral swab

1 = Score 95% CI

{14}------------------------------------------------

SpecimenTypeSymptomStatusNCompositecomparator+/Aptima+Compositecomparator-/Aptima+Compositecomparator-/Aptima-Compositecomparator+/Aptima-PPA(95% CI)¹NPA(95% CI)¹
CVSSym105012312913298.4(94.4-99.6)98.7(97.7-99.3)
CVSAsym679525621198.1(90.1-99.7)99.2(98.1-99.7)
PVSSym1047121189080100(96.9-100)98.1(96.9-98.8)
PVSAsym677526617296.3(87.5-99.0)99.0(97.9-99.6)
ESSym10571154935397.5(92.8-99.1)99.6(98.9-99.8)
ESAsym677483624296.0(86.5-98.9)99.5(98.6-99.8)
FUSym10741067955694.6(88.8-97.5)99.3(98.5-99.6)
FUAsym700412654393.2(81.8-97.7)99.7(98.9-99.9)
USSym8661021761298.1(93.3-99.5)99.9(99.3-100)
USAsym697605631198.4(91.3-99.7)99.2(98.2-99.7)
PMSym8701018756595.3(89.4-98.0)99.0(97.9-99.5)
PMAsym693616623395.3(87.1-98.4)99.0(97.9-99.6)
MUSym874992770397.1(91.7-99.0)99.7(99.1-99.9)
MUAsym704600643198.4(91.3-99.7)100(99.4-100)

Table 10: Specimen-Specific Agreements by Symptom Status

Asym = asymptomatic, CI = confidence interval, CVS = clinician-collected vaginal swab, ES = endocervical swab, FU = female urine, MU = male urine, NPA = negative percent agreement, PM = penile meatal swab, PPA = positive percent agreement, PVS = patient-collected vaginal swab, Sym = symptomatic, US = male urethral swab. 1 Score 95% CI

4. Clinical cut-off:

Not Applicable

5. Expected values/Reference range:

The observed expected values (number of positive results detected by the Aptima Mycoplasma genitalium Assay) in this prospective clinical study were 11.1% in cliniciancollected vaginal swab samples, 11.4% in self-collected vaginal swab samples, 9.7% in endocervical swab samples, 8.8% in female urine, 10.7% in male urethral swab samples,

{15}------------------------------------------------

11.3% in self-collected penile meatal swab samples, and 10.1% in male urine samples.

M. Instrument Name:

The Panther System

N. System Descriptions:

    1. Modes of Operation:
      Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes

Level of Concern:

Moderate

Software Description:

There are two software components that comprise the Panther System which is required to perform the Aptima Mycoplasma genitalium Assay:

  • System Software: System software includes the Master Controller software and O instrument firmware. This software is assay independent and does not include assay specific processing parameters. The System software communicates with the Assay software to access assay specific parameters.
  • Assay Software: Assay software contains information that is specific to a given o assay and provides assay-specific configuration information to the instrument. Assay software includes assay specific parameters for reagent volumes, incubation times, incubation temperatures, and other assay parameters, such as the sequence of steps taken.

{16}------------------------------------------------

3. Specimen Identification:

Specimens are identified via barcode labels on the sample vial.

4. Specimen Sampling and Handling:

Clinician-collected and self-collected vaginal swab specimens, clinician-collected endocervical swab specimens, female and male urine specimens, clinician-collected male urethral swab specimens, and self-collected penile meatal swab specimens can be tested with the Aptima Mycoplasma genitalium assay. Assay performance has not been evaluated with specimens other than those collected with the following specimen collection kits:

  • · Aptima Unisex Swab Specimen Collection Kit for Endocervical and Male Urethral Swab Specimens
  • · Aptima Urine Collection Kit for Male and Female Urine Specimens
  • · Aptima Multitest Swab Specimen Collection Kit

For Specimen collection, refer to the appropriate specimen collection kit package insert for specific collection instructions.

Please refer to section L.1.c. for specimen transport and storage.

    1. Calibration:
      The Panther System is calibrated upon installation by Hologic Inc. Field service engineers as well as during preventive maintenance as scheduled.
    1. Quality Control:
      Please see section L.1.c.

O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Comparator TMA Assays validation:

The three comparator TMA assays used to determine the PIS status (for sensitivity and specificity calculation) and composite comparator result (for specimen-specific agreements calculation) detected regions of the rRNA gene different than the region detected by the Aptima Mycoplasma genitalium assay. The validation of these three comparator TMA assays include:

    1. LoD study: A complete LoD study was conducted with one strain of Mycoplasma genitalium using two Panther Systems with at leas (0) eplicates per dilution in four clinical matrices: vaginal swab, penile meatal swa (4) ale urine, and female urine. The LoD of each TMA assay was found comparable to the Aptima Mycoplasma genitalium assay.

{17}------------------------------------------------

    1. Inclusivity: Ten Mycoplasma genitalium strains were tested in this study in triplicates using three Panther Systems and one reagent lot for each comparator TMA assay. Each of the three TMA assays and the Aptima Mycoplasma genitalium assay detected the tested Myconlasma genitalium strains at a similar concentration.
    1. Cross-reactivit (b) icroorganisms tested in the cross-reactivity study for the Aptima Mycopl 41 genitalium assay were tested in triplicates for each comparator TMA assay. In addition, in silico analysis was performed similar to the Aptima Mycoplasma genitalium assay with eight microorganisms. No cross-reactivity was observed and the three comparator TMA assays were found specific by in silico analysis.
    1. Microbial Interference: microorganisms evaluated for cross-reactivity were also tested for microbial interestion in the presence of Mycoplasma genitalium at 3X LoD. Interference was observed with M. vneumoniae, commonly found in the lower respiratory tract, similar to what was observed with the Aptima Mycoplasma genitalium assay.
    1. Comparator TMA assays were further evaluated using (4) archived clinical specimens. The positive percent agreement between th ma Mycoplasma genitalium assay and each comparator TMA assay ranged from 94.4% to 97.7% and negative percent agreement ranged from 99.8% to 99.9%.

P. Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device.

O. Identified Risks to Health and Mitigation Measures:

Identified Risks to HealthMitigation Measures
Risk of false resultsGeneral controls and Special Controls (1), (2),(3), and (4)
Failure to correctly interpret test resultsGeneral Controls and Special Controls (1),(3)(iii), (3)(iv), and 3(v)
Failure to correctly operate the deviceGeneral Controls and Special Controls (1),(3)(i), and (4)

R. Benefit/Risk Analysis:

Summary of the Assessment of Benefit

There are no other FDA cleared or approved assays currently on the market for the detection of Mvcoplasma genitalium. The Aptima Mycoplasma genitalium Assay is the first qualitative nucleic acid amplification test (NAAT) to detect Mycoplasma genitalium nucleic acids directly from urogenital specimens. Use of the Aptima Mycoplasma genitalium assay will improve detection and treatment of Mycoplasma genitalium infections, and may help limit the spread of infection in the community and reduce the sequelae of infection. The performance demonstrated in the prospective clinical study suggests that the Aptima Mycoplasma genitalium assay will be highly sensitive and specific for detection of Mycoplasma genitalium in the preferred specimen types.

{18}------------------------------------------------

Summary of the Assessment of Risk

The risks associated with the device are those related to the risk of false test results, failure to correctly interpret test results, and failure to correctly operate the instrument. False-positive results and false-negative results are the primary risks associated with the use of the Aptima Mycoplasma genitalium assay. A false-positive result may lead to inappropriate or unnecessary antibiotic therapy with the potential for side effects and increasing antibiotic resistance, although the specificty of the assay suggests that false positives will be uncommon. A false-negative result may lead to a delayed or missed diagnosis with the potential for adverse reproductive health outcomes in women and urethritis in men. The lower performance in some claimed specimen types is mitigated by an additional statement included in the Indications for Use (IFU) and product labeling. The IFU statement emphasizes the higher risk of false-negative results in endocervical swab and female urine specimens and identifies vaginal swabs as the preferred specimen type. It also indicates that follow up testing could be necessary, if negative results are obtained with endocervical swabs or female urine and Mycoplasma genitalium infection is suspected. However, general controls alone are not sufficient to mitigate the risks described. The special controls include a statement that the labeling must include, for example, a detailed description of the device, detailed description of the test procedure and interpretation of test results, and appropriate limiting statements, and they include requirements for assay design verification and validation.

Summary of the Assessment of Benefit-Risk

The probable benefits of the Aptima Mycoplasma genitalium Assay outweigh the potential risks, given the proposed special controls and applicable general controls. The Aptima Mycoplasma genitalium Assay is the first qualitative NAAT to detect Mycoplasma genitalium in urogenital specimens and is likely to benefit patients by allowing a rapid diagnosis of an emerging sexually transmitted infection. The performance demonstrated during the clinical study suggest that errors will be uncommon and will be mitigated by the proposed special controls.

S. Patient Perspectives:

This submission did not include specific information on patient perspectives for this device.

T. Conclusion:

The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.3393. FDA believes that the special controls, in combination with the general controls, provide a reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code: QEP

Device to detect nucleic acids from non-viral microorganism(s) causing Device Type: sexually transmitted infections and associated resistance marker(s)

{19}------------------------------------------------

Class: II (special controls)

Regulation: 21 CFR 866.3393

  • (a) Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.
  • Classification: Class II (special controls). A device to detect nucleic acids from non-(b) viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) must comply with the following special controls:
  • (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
  • (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen tvpes claimed bv this device: alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
  • The 21 CFR 809.10(b) labeling must include: (3)
    • (i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
    • (ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including, but not limited to. Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate;
    • (iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing.
    • (iv) Limiting statements indicating that:

(A)a negative test result does not preclude the possibility of infection;

{20}------------------------------------------------

  • (B) the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
  • (C) reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe procedures in any one of these steps can lead to incorrect results; and
  • (D)if appropriate (e.g., recommended by CDC, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type.
  • (v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
    • (A)negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
    • (B) detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
    • (C) detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
    • (D) therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
  • (4) Design verification and validation must include:
    • (i) Detailed device description documentation, including, but not limited to, methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported result).
    • (ii) Detailed documentation of analytical studies including but not limited to, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate.
    • (iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical

{21}------------------------------------------------

samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.

  • A detailed description of the impact of any software, including, but not limited to, (iv) software applications and hardware-based devices that incorporate software, on the device's functions.

§ 866.3393 Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s).

(a)
Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.