(208 days)
Not Found
No
The device description and performance studies focus on the physical and mechanical properties of a balloon guide catheter, with no mention of software, algorithms, or data processing that would indicate AI/ML.
No.
This device is designed to facilitate procedures and temporarily occlude blood vessels, which are interventional actions, not therapeutic ones in themselves.
No
The BOBBY Balloon Guide Catheter is intended for facilitating the insertion and guidance of other intravascular catheters and for temporary vascular occlusion. It is a tool used during procedures, not a device that independently diagnoses a condition.
No
The device description clearly details a physical catheter with a balloon, lumens, radiopaque markers, and a hub. It is a hardware device used for physical intervention in blood vessels.
Based on the provided information, the BOBBY Balloon Guide Catheter is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use clearly states the device is for facilitating the insertion and guidance of catheters within the body (intravascular), providing temporary vascular occlusion, and acting as a conduit for retrieval devices. These are all in vivo procedures, meaning they are performed on a living organism.
- Device Description: The description details a physical catheter designed to be inserted into blood vessels. It describes features like lumens for guidewires and inflation, a balloon for occlusion, and a hydrophilic coating for insertion. These are all characteristics of a device used directly within the body.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue, etc.) outside of the body. IVDs are used to provide information about a patient's health status through the analysis of such specimens.
Therefore, the BOBBY Balloon Guide Catheter is an invasive medical device used for interventional procedures within the vascular system, not an IVD.
N/A
Intended Use / Indications for Use
The BOBBY Balloon Guide Catheter is intended: For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
Product codes
DQY, QJP
Device Description
The BOBBY Balloon Guide Catheter is a co-axial, braid-reinforced, variable stiffness catheter with an external hydrophilic coating. The BOBBY Balloon Guide Catheter incorporates a compliant balloon, radiopaque markers, and a bifurcated luer hub on the proximal end.
The BOBBY Balloon Guide Catheter has an inner lumen through which a guidewire and catheter can be inserted, and a co-axial outer lumen that is used to inflate and deflate the balloon with a syringe filled with contrast media. A bifurcated luer hub is attached to the proximal end of the balloon guide catheter to provide access to both the inner and outer lumens. In addition, a hydrophilic coating is applied to the distal end of the balloon guide catheter to provide a lubricious outer surface for catheter advancement in the vasculature. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. The balloon incorporates a distal air-purging system to purge air from the inflation lumen prior to use. The balloon catheter also incorporates radiopaque markers to facilitate fluoroscopic visualization and indication of the balloon position.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Peripheral and neuro vascular systems.
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance Testing Summary includes:
- Conditioning, Distribution, and Shelf Life Aging Verification: Pass. All samples met the pre-determined acceptance criteria.
- Packaging Visual Inspection: Pass. All samples met the pre-determined acceptance criteria.
- Visual Surface Requirements: Pass. All samples met the pre-determined acceptance criteria.
- Dimensional/Physical Attributes Inspection: Pass. All samples met the pre-determined acceptance criteria.
- Inflation Volume vs Balloon Diameter: Pass. All samples met the pre-determined acceptance criteria.
- Balloon Burst Volume: Pass. All samples met the pre-determined acceptance criteria.
- Tip Stiffness: Pass. All samples met the pre-determined acceptance criteria.
- Torque Testing: Pass. All samples met the pre-determined acceptance criteria.
- Force at Break: Pass. All samples met the pre-determined acceptance criteria.
- Small Bore Connector Compliance with Standard: Pass. All samples met the pre-determined acceptance criteria.
- Radiopacity: Pass. All samples met the pre-determined acceptance criteria.
- Particulate, Coating Integrity: Pass. All samples met the pre-determined acceptance criteria.
- Freedom from Liquid Leakage: Pass. All samples met the pre-determined acceptance criteria.
- Hub Aspiration Air Leakage: Pass. All samples met the pre-determined acceptance criteria.
- Balloon Fatigue Test: Pass. All samples met the pre-determined acceptance criteria.
- Simulated Use: Simulated use under in vitro conditions in a cerebral vascular model. Pass. All samples met the pre-determined acceptance criteria.
- Flexural Fatigue: Pass. All samples met the pre-determined acceptance criteria.
- Kink Resistance: Pass. All samples met the pre-determined acceptance criteria.
- Lubricity and durability of the hydrophilic coating: Pass. All samples met the pre-determined acceptance criteria.
- Guidewire lumen burst pressure (Static, dynamic): Pass. All samples met the pre-determined acceptance criteria.
- Balloon Deflation Time: Pass. All samples met the pre-determined acceptance criteria.
- Lumen Collapse: Pass. All samples met the pre-determined acceptance criteria.
Biocompatibility Testing Summary includes:
- Cytotoxicity: Pass. Noncytotoxic according to the predetermined acceptance criteria.
- Sensitization: Pass. Did not elicit a sensitization response according to the predetermined acceptance criteria.
- Intracutaneous Irritation: Pass. Test requirements for intracutaneous reactivity were met according to the predetermined acceptance criteria.
- Systemic Toxicity: Systemic Injection Test: Pass. Test requirements for systemic toxicity were met according to the predetermined acceptance criteria.
- Systemic Toxicity: Material Mediated Pyrogenicity: Pass. Non-pyrogenic, met the predetermined acceptance criteria.
- Hemocompatibility: Hemolysis: Pass. Non-hemolytic, met the predetermined acceptance criteria.
- Hemocompatibility: Complement Activation: Pass. Does not activate the complement system, met the predetermined acceptance criteria.
- Hemocompatibility: Thrombogenicity: Pass. Demonstrates similar thromboresistance characteristics as the control device, met the predetermined acceptance criteria.
- Hemocompatibility: In Vitro Hemocompatibility: Pass. Not expected to result in adverse effects in vivo, met the predetermined acceptance criteria.
- Hemocompatibility: Partial Thromboplastin Time: Pass. Does not have an effect on coagulation of human plasma, met the predetermined acceptance criteria.
No animal or clinical studies were required to demonstrate substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Concentric Medical, FlowGate2 (K153729)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 21, 2020
MicroVention, Inc. Stephanie Onstot Senior Regulatory Affairs Specialist 1311 Valencia Avenue Tustin, CA 92780
Re: K193607
Trade/Device Name: BOBBY Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: June 25, 2020 Received: June 26, 2020
Dear Stephanie Onstot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K193607
Device Name BOBBY Balloon Guide Catheter
Indications for Use (Describe) The BOBBY Balloon Guide Catheter is intended:
For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) |
---|
☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(K) SUMMARY
510(k) Owner Information:
510(k) Owner: | MicroVention, Inc. | |
---|---|---|
Address: | 1311 Valencia Ave | |
Tustin, California, USA | ||
Telephone: | 714-247-8000 | |
Fax: | 714-247-8014 | |
Contact Person | Stephanie Onstot | |
Telephone | 714-247-8000 x 8317 | |
Date Prepared: | June 25, 2020 | |
Device Trade Name: | BOBBY™ Balloon Guide Catheter | |
Classification: | Class II | |
Classification Name: | Percutaneous Catheter | |
Product Code(s): | DQY, QJP | |
Regulation Number(s) | 870.1250 | |
Predicate Devices: | Concentric Medical, FlowGate2™ (K153729) | |
Indication for Use:
The BOBBY Balloon Guide Catheter is intended:
For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.
Device Description
The BOBBY Balloon Guide Catheter is a co-axial, braid-reinforced, variable stiffness catheter with an external hydrophilic coating. The BOBBY Balloon Guide Catheter incorporates a compliant balloon, radiopaque markers, and a bifurcated luer hub on the proximal end.
The BOBBY Balloon Guide Catheter has an inner lumen through which a guidewire and catheter can be inserted, and a co-axial outer lumen that is used to inflate and deflate the balloon with a syringe filled with contrast media. A bifurcated luer hub is attached to the proximal end of the balloon guide catheter to provide access to both the inner and outer lumens. In addition, a hydrophilic coating is applied to the distal end of the balloon guide catheter to provide a lubricious outer surface for catheter advancement in the vasculature. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic
4
procedures. The balloon incorporates a distal air-purging system to purge air from the inflation lumen prior to use. The balloon catheter also incorporates radiopaque markers to facilitate fluoroscopic visualization and indication of the balloon position.
Performance Testing Summary:
The BOBBY Balloon Guide Catheter has successfully completed the following relevant performance testing demonstrating that the device is suitable for its intended use.
Performance Test Summary | |||
---|---|---|---|
Study Name | Description | Reference Standard | Results |
Conditioning, | |||
Distribution, and Shelf | |||
Life Aging Verification | To demonstrate that the product | ||
meets the packaging strength and | |||
packaging integrity following | |||
accelerated aging to a 1-year | |||
shelf life equivalent | ASTM F88 Seal | ||
Strength of Flexible | |||
Barrier materials | |||
ASTM F2096 Standard | |||
Test Method for | |||
Detecting Gross Leaks | |||
in Medical Packaging | Pass | ||
All samples met the pre- | |||
determined acceptance | |||
criteria | |||
Packaging Visual | |||
Inspection | To demonstrate that the product | ||
meets the packaging visual | |||
inspection requirements given | N/A | Pass | |
All samples met the pre- | |||
determined acceptance | |||
criteria | |||
Visual Surface | |||
Requirements | To demonstrate the product | ||
satisfies the visual surface | |||
requirements | ISO 10555-1:2013 | ||
Intravascular catheters |
- Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Dimensional/Physical
Attributes Inspection | To demonstrate that the product
meets the dimensional
specifications | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Inflation Volume vs
Balloon Diameter | To demonstrate that the product
meets the inflation volume vs
balloon diameter specifications | In consideration of
ISO 10555-4:2013
Intravascular Catheters - Sterile and Single-
Use Catheters - Part 4:
Balloon Dilation
Catheters | Pass
All samples met the pre-
determined acceptance
criteria |
| Balloon Burst Volume | To demonstrate that the Balloon
is capable of withstanding an
injection volume above the
maximum fill volume. | In consideration of
ISO 10555-4:2013
Intravascular Catheters - Sterile and Single-
Use Catheters - Part 4:
Balloon Dilation
Catheters | Pass
All samples met the pre-
determined acceptance
criteria |
| Tip Stiffness | To demonstrate that the stiffness
of the distal end of the product is
similar to other marketed devices. | N/A | Pass
All samples met the pre-
determined acceptance
criteria |
| Torque Testing | To demonstrate that the product
is capable of 720 degrees of
rotation about the central lumen
axis without failure. | N/A | Pass
All samples met the pre-
determined acceptance
criteria |
| Performance Test Summary | | | |
| Study Name | Description | Reference Standard | Results |
| Force at Break | To demonstrate the product
satisfies the force at break
requirements | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Small Bore Connector
Compliance with
Standard | To demonstrate that the product
meets the requirements | ISO 80369-7 2016,
Small-bore connectors
for liquids and gases in
healthcare applications - Part 7, Connectors
for intravascular or
hypodermic
applications. | Pass
All samples met the pre-
determined acceptance
criteria |
| Radiopacity | To determine the radiopaque
characteristics of the device. | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements
ASTM F640-12
Standard Test Methods
for Determining
Radiopacity for
Medical Use | Pass
All samples met the pre-
determined acceptance
criteria |
| Particulate, Coating
Integrity | This study was conducted to
determine the quantity and size of
particles generated during
simulated use | USP Particulate
Matter in Injections | Pass
All samples met the pre-
determined acceptance
criteria |
| Freedom from Liquid
Leakage | To demonstrate that the product
meets the liquid leakage
requirements given in ISO
10555- 1. | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Hub Aspiration Air
Leakage | To demonstrate that the product
meets the hub aspiration air
leakage requirements given in
ISO 10555-1. | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Balloon Fatigue Test | To demonstrate that there is no
degradation of the Balloon after
repetitive inflation cycles. | In consideration of
ISO 10555-4:2013
Intravascular Catheters - Sterile and Single-
Use Catheters - Part 4:
Balloon Dilation
Catheters | Pass
All samples met the pre-
determined acceptance
criteria |
| Simulated Use | Simulated use under in vitro
conditions in a cerebral vascular
model | In consideration of
ISO 10555-4:2013
Intravascular Catheters - Sterile and Single-
Use Catheters - Part 4:
Balloon Dilation
Catheters | Pass
All samples met the pre-
determined acceptance
criteria |
| Performance Test Summary | | | |
| Study Name | Description | Reference Standard | Results |
| Flexural Fatigue | To demonstrate that the product
does not lose structural integrity
when used in the tortuous path
model. | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Kink Resistance | To demonstrate that the device
has similar kink resistance
compared to the predicate device. | N/A | Pass
All samples met the pre-
determined acceptance
criteria |
| Lubricity and durability of
the hydrophilic coating | To demonstrate that the
hydrophilic coating is lubricious
and durable. | N/A | Pass
All samples met the pre-
determined acceptance
criteria |
| Guidewire lumen burst
pressure (Static, dynamic) | To demonstrate that the device
does not burst below rated burst
pressure. | ISO 10555-1:2013
Intravascular catheters - Sterile and single-use
catheters - Part 1:
General requirements | Pass
All samples met the pre-
determined acceptance
criteria |
| Balloon Deflation Time | To demonstrate that the device
has similar balloon deflation time
compared to the predicate device. | ISO10555-4:2013
Intravascular catheters - Sterile and single-use
catheters – Part 4:
Balloon dilation
catheters | Pass
All samples met the pre-
determined acceptance
criteria |
| Lumen Collapse | To demonstrate that the
guidewire lumen does not
collapse under aspiration. | N/A | Pass
All samples met the pre-
determined acceptance
criteria |
5
6
No animal or clinical studies were required to demonstrate substantial equivalence.
Biocompatibility Testing Summary
Categorized as Externally Communicating Device, Circulating Blood, Limited Contact (vitro cytotoxicity | Pass
Noncytotoxic according to the
predetermined acceptance criteria | |
| | Sensitization | Tested in accordance with ISO 10993-10, Biological
Evaluation of Medical Devices – Part 10 Tests for
Irritation and Skin Sensitization, Kligman
Maximization Test | Pass
Did not elicit a sensitization response
according to the predetermined
acceptance criteria |
| Intracutaneous
Irritation | | Tested in accordance with ISO 10993-10, Biological
Evaluation of Medical Devices – Part 10: Tests for
Irritation and Skin Sensitization | Pass
Test requirements for
intracutaneous reactivity were met
according to the predetermined
acceptance criteria |
7
Test Name | Test Method | Results |
---|---|---|
Systemic Toxicity: | ||
Systemic Injection | ||
Test | Tested in accordance with ISO 10993-11, Biological | |
Evaluation of Medical Devices - Part 11: Tests for | ||
Systemic Toxicity | Pass | |
Test requirements for systemic | ||
toxicity were met according to the | ||
predetermined acceptance criteria | ||
Systemic Toxicity: | ||
Material Mediated | ||
Pyrogenicity | Tested in accordance with ISO 10993-11, Biological | |
Evaluation of Medical Devices – Part 11: Tests for | ||
Systemic Toxicity and USP NF 36:2018 | ||
Pyrogen Test | Pass | |
Non-pyrogenic, met the | ||
predetermined acceptance criteria | ||
Hemocompatibility: | ||
Hemolysis | Tested in accordance with ASTM F756-17, | |
Standard Practice for Assessment of Hemolytic | ||
Properties of Materials and ISO 10993-4, Biological | ||
Evaluation of Medical Devices - Part 4: Selection of | ||
Tests for Interactions with Blood. Tests for | ||
Hemolytic Properties, Direct and Indirect Methods | Pass | |
Non-hemolytic, met the | ||
predetermined acceptance criteria | ||
Hemocompatibility: | ||
Complement | ||
Activation | Tested in accordance with ISO 10993-4, Biological | |
Evaluation of Medical Devices - Part 4: Selection of | ||
Tests for Interactions with Blood, SC5b-9 | ||
Complement Activation | Pass | |
Does not activate the complement | ||
system, met the predetermined | ||
acceptance criteria | ||
Hemocompatibility: | ||
Thrombogenicity | Tested in accordance with ISO 10994-4, Biological | |
Evaluation of Medical Devices - Part 4: Selection of | ||
Tests for Interactions with Blood, and ASTM | ||
F2888-19, Standard Practice for Platelet Leukocyte | ||
Count-An In Vitro Measure for Hemocompatibility | ||
Assessment of Cardiovascular Materials | Pass | |
Demonstrates similar | ||
thromboresistance characteristics as | ||
the control device, met the | ||
predetermined acceptance criteria | ||
Hemocompatibility: | ||
In Vitro | ||
Hemocompatibility | Tested in accordance with ISO 10993-4, Biological | |
Evaluation of Medical Devices - Part 4: Selection of | ||
Tests for Interactions with Blood, | ||
Hemocompatibility, Direct Contact Method | Pass | |
Not expected to result in adverse | ||
effects in vivo, met the | ||
predetermined acceptance criteria | ||
Hemocompatibility: | ||
Partial | ||
Thromboplastin Time | Tested in accordance with ISO 10994-4, Biological | |
Evaluation of Medical Devices - Part 4: Selection of | ||
Tests for Interactions with Blood and ASTM F2382- | ||
18, Standard Test Method for Assessment of | ||
Intravascular Medical Device Materials on Partial | ||
Thromboplastin Time (PTT) | Pass | |
Does not have an effect on | ||
coagulation of human plasma, met | ||
the predetermined acceptance | ||
criteria |
Predicate Device Comparison
The following table provides a comparison of the key characteristics of the BOBBY Balloon Guide Catheter to the predicate device.
Comparison Chart | ||||
---|---|---|---|---|
Feature | Subject Device | Predicate device | Comparison | |
BOBBY Balloon | ||||
Guide Catheter | ||||
K193607 | 8F FlowGate2™ | |||
Balloon Guide Catheter | ||||
K153729 | ||||
FDA Classification | Class II | Class II | Same | |
Product Code(s) | DQY, QJP | DQY | Same | |
Regulation Number | 870.1250 | 870.1250 | Same | |
Regulation Name | Percutaneous Catheter | Percutaneous Catheter | Same | |
Anatomical Locations | Peripheral and neuro vasculature | Peripheral and neuro vasculature | Same | |
Comparison Chart | ||||
Feature | Subject Device | |||
BOBBY Balloon | ||||
Guide Catheter | ||||
K193607 | Predicate device | |||
8F FlowGate™ | ||||
Balloon Guide Catheter | ||||
K153729 | Comparison | |||
Material | Commonly used | |||
medical grade plastics | ||||
(e.g. nylon, PTFE, | ||||
polyethylene, | ||||
polyolefin) and | ||||
stainless steel | Commonly used | |||
medical grade plastics | ||||
(e.g. nylon, PTFE, | ||||
polyethylene, | ||||
polyolefin) and | ||||
stainless steel | Similar, minor | |||
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy, confirmed | ||||
through | ||||
biocompatibility and | ||||
performance testing | ||||
Reinforced Catheter | ||||
Shaft | Stainless steel braid | |||
and coil reinforced | ||||
shaft | Stainless steel braid | |||
reinforced shaft | Similar, minor | |||
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy | ||||
Injection Port | Yes | Yes | Same | |
Radiopacity | Yes, shaft is visible | |||
due to distal tip Pt-Ir | ||||
marker bands | Yes, shaft material | |||
contains barium | ||||
sulfate, distal tip Pt-Ir | ||||
marker band | Similar, minor | |||
difference does not | ||||
raise new questions | ||||
regarding safety and | ||||
efficacy, both | ||||
devices are | ||||
radiopaque | ||||
Radiopaque Marker | ||||
Bands | 2 | 1 | Similar, minor | |
difference does not | ||||
raise new questions | ||||
regarding safety and | ||||
efficacy, both | ||||
devices utilize | ||||
radiopaque marker | ||||
bands | ||||
Compliant Balloon | Yes, polyurethane | Yes, silicone | Similar, material | |
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy, both | ||||
materials are used | ||||
for compliant | ||||
balloons for | ||||
intravascular use | ||||
Labeled Shaft Outer | ||||
Diameter | 0.110 in (2.8 mm) | |||
8Fr | 0.106 in (2.7 mm) | |||
8Fr | Similar, minor | |||
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy | ||||
Comparison Chart | ||||
Feature | Subject Device | |||
BOBBY Balloon | ||||
Guide Catheter | ||||
K193607 | Predicate device | |||
8F FlowGate2™ | ||||
Balloon Guide Catheter | ||||
K153729 | Comparison | |||
Labeled Shaft Inner | ||||
Diameter | 0.086 in (2.18 mm) | |||
6.5Fr | 0.084 in (2.1 mm) | |||
6.4Fr | Similar, minor | |||
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy | ||||
Effective length | 95 cm | |||
(37.4 in) | 85 cm, 95cm | |||
(33.5, 37.4 in) | Similar, minor | |||
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy | ||||
Tip Shape | Straight | Straight | Same | |
Coating | Hydrophilic Coating - | |||
Distal portion of shaft | None | Difference does not | ||
raise new questions | ||||
of safety and | ||||
efficacy. | ||||
Hydrophilic coatings | ||||
are commonly used | ||||
to lubricate vascular | ||||
catheters. | ||||
Internal Construction | Coaxial lumen | Coaxial lumen | Same | |
Accessories Supplied | Peel Away Sheath | Dilator, Rotating | ||
Hemostasis Valve, | ||||
Tuohy Borst Valve | ||||
with sideport, Peel | ||||
Away Sheaths, Luer- | ||||
Activated Valves | Similar, minor | |||
differences do not | ||||
raise new questions | ||||
of safety and | ||||
efficacy. | ||||
How Supplied | Sterile, single use | Sterile, single use | Same | |
Sterilization Method | EtO | EtO | Same | |
Sterility Assurance | ||||
Level | 10-6 | 10-6 | Same |
8
9
Conclusion:
MicroVention, Inc. concludes through a review of the benchtop assessments, the comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility that the BOBBY Balloon Guide Catheter is substantially equivalent to the predicate device, FlowGate2™ Balloon Guide Catheter. Any differences are minor and do not raise different questions of safety and effectiveness.