K Number
K161064
Manufacturer
Date Cleared
2016-06-12

(58 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.

Device Description

The Penumbra System ACE 68 Reperfusion Catheter is a component to the currently available Penumbra System. The ACE 68 Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel in the neurovasculature. ACE 68 is provided sterile, non-pyrogenic, and intended for single use only.

AI/ML Overview

The Penumbra System ACE 68 Reperfusion Catheter received 510(k) clearance (K161064) based on substantial equivalence to the previously cleared Penumbra System ACE 68 Reperfusion Catheter (K152541). The study conducted in support of this clearance was a Design Verification (Bench-Top Testing), alongside Shelf Life Testing.

Here's a breakdown of the requested information based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

AttributeAcceptance CriteriaReported Device Performance
Packaging InspectionConfirm the packaging and dimensions of the units meet all product specifications.Pass
Dimensional/Visual(Implicitly covered by Packaging Inspection - meeting product specifications)Pass
Kink ResistanceNo kinking when formed in a defined radius.Pass
Markerband VisibilityThe markerband is fluoroscopically visible.Pass
Simulated Use (Intracranial Access, Vessel Access Entry Performance & Clot Removal)Simulated use testing of the Reperfusion Catheter and Separator was performed with accessory devices in an anatomical model which simulated the tortuosity of the neurovasculature. Devices were delivered through the tortuous anatomical model to evaluate the effectiveness of the devices to remove clots and that the Reperfusion Catheter does not collapse under vacuum.Pass (effectiveness of devices to remove clots and no collapse under vacuum confirmed)
Particulate testing: ≥ 10 um≤ 6000 particlesPass
Particulate testing: ≥ 25 um≤ 600 particlesPass
Particulate testing: ≥ 75 umMeasured for informational purposes only (FIPO)FIPO (results recorded, but no pass/fail criterion)
Particulate testing: ≥ 125 umMeasured for informational purposes only (FIPO)FIPO (results recorded, but no pass/fail criterion)
Coating IntegrityCoating has not delaminated, peeled, or flaked prior to or after simulated use particulate testing.Pass
Hub Air AspirationNo leaks detected when vacuum is pulled on the injection lumen.Pass
Static Burst Pressure Test45 psi for 30 sec minimum.Pass
ACE 68 / Sheath or 8F Guide Catheter Friction ForceMinimum value per specification.Pass
ACE 68 / 0.014 in. Guidewire Friction ForceMinimum value per specification.Pass
Joint sections bond strengthMinimum value per specification.Pass
Hub to shaft tensile strengthMinimum value per specification.Pass
Hub to hypotube tensile strengthMinimum value per specification.Pass
Elongation to failureElongation ≥ 5%.Pass
TorsionNumber of turns will be recorded for informational purposes only.FIPO (results recorded, but no pass/fail criterion)
CorrosionNo visible corrosion on Reperfusion Catheter immediately after corrosion testing procedure.Pass
Shelf Life TestingMet all established requirements (after transportation conditioning per ASTM D4169 and accelerated aging equal to 8 months).Successful testing verifies an 8-month shelf life.

2. Sample size used for the test set and data provenance

  • Sample Size for Test Set: The document does not specify the exact sample size (number of devices) used for each individual bench-top test. It generally states that "The physical and mechanical properties of the ACE 68 Reperfusion Catheter were assessed using standard test methods and pre-determined acceptance criteria." and "The subject ACE 68 Reperfusion Catheters met all established requirements" for shelf life.
  • Data Provenance: This was a non-clinical bench-top testing and shelf-life testing. The data is therefore from laboratory simulations and controlled environments, not human subjects. The country of origin for the data is not explicitly stated, but the sponsor is Penumbra, Inc. located in Alameda, CA, USA. This is retrospective in the sense that the device changes were made, and then these verification tests were performed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This type of bench-top engineering testing does not typically involve human experts establishing "ground truth" in the same way clinical studies do. The "truth" is established by adherence to predefined engineering specifications and standard test methods.
  • For the "Simulated Use" test, while it involves an "anatomical model which simulated the tortuosity of the neurovasculature," the assessment of "effectiveness of the devices to remove clots and that the Reperfusion Catheter does not collapse under vacuum" would be based on objective observation and measurement against predefined criteria rather than subjective expert consensus on clinical outcomes. The document does not mention the involvement or qualifications of experts for defining the ground truth for this non-clinical testing.

4. Adjudication method for the test set

  • Not applicable as this was bench-top testing against predefined specifications, not a clinical study requiring adjudication of outcomes by multiple readers or judges.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This document describes the non-clinical (bench-top) testing for a physical medical device (catheter), not an AI algorithm.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • No, this device is a physical medical instrument (catheter), not an algorithm or AI system. Therefore, standalone algorithm performance is not applicable.

7. The type of ground truth used

  • The "ground truth" for this non-clinical testing consisted of pre-determined, objective engineering specifications and acceptance criteria. For example, a specific burst pressure (45 psi for 30 sec minimum), specific particulate counts (≤ 6000 for ≥ 10 um), or visual confirmation (no kinking, fluoroscopically visible markerband).

8. The sample size for the training set

  • Not applicable. This is not an AI or machine learning study, so there is no "training set."

9. How the ground truth for the training set was established

  • Not applicable. This is not an AI or machine learning study.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2016

Penumbra, Inc. Mr. Charles DeNault Regulatory Affairs Specialist III One Penumbra Place Alameda, California 94502

Re: K161064

Trade/Device Name: Penumbra System ACE 68 Reperfusion Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: May 23, 2016 Received: May 24, 2016

Dear Mr. Charles DeNault:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S 同公

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161064

Device Name

Penumbra System ACE 68 Reperfusion Catheter

Indications for Use (Describe)

The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra. Inc. is providing the summary of Substantial Equivalence for the Penumbra System® ACE™ 68 Reperfusion Catheter.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Charles DeNault Regulatory Affairs Specialist III Phone: (510) 748-3302 FAX: (510) 217-6414 Email: cdenault@penumbrainc.com

Date of Preparation of 510(k) Summary 1.3

June 1, 2016

Device Trade or Proprietary Name 1.4

Penumbra System® ACE™ 68 Reperfusion Catheter

1.5 Device Classification

Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR §870.1250 Product Code: NRY (Catheter, Thrombus Removal)

1.6 Predicate Devices

510(k)NumberClearance DateName of Predicate DeviceName of Manufacturer
K152541January 13, 2016Penumbra System® ACE™ 68Reperfusion CatheterPenumbra, Inc.One Penumbra PlaceAlameda, CA 94502 USA

{4}------------------------------------------------

1.7 Predicate Comparison

Design modifications were made to the subject device. Changes include a reduction of the device markerband length as well as dimensional changes to the device PTFE liner, coil reinforcement filaments, and extrusions.

Predicate: ACE 68Subject: ACE 68
General
510(k) No.K152541K161064
ClassificationClass II, NRYSAME
IndicationThe Penumbra System is intended for usein the revascularization of patients withacute ischemic stroke secondary tointracranial large vessel occlusive disease(within the internal carotid, middle cerebral- M1 and M2 segments, basilar, andvertebral arteries) within 8 hours ofsymptom onset.SAME
Materials
Proximal hubGrilamid (TR55-LX)SAME
Strain Relief [HubSleeve]Grilamid (TR55)SAME
Strain Relief304 Stainless Steel (SS)SAME
ID BandPolyolefin, PET yellow [black ink]SAME
LinerPTFESAME
Coil Reinforcement304V SS, Nitinol (55% Ni, 45% Ti)SAME
Proximal ExtrusionsVestamid, Pebax 72D, Pebax 55D/72DblendSAME
Distal ExtrusionsPebax 63D, Pebax 55D, Pebax 40D/55DBlend, Pebax 40D, Pebax 35D/40D Blend,Pebax 35D, Tecoflex 80A/Pebax 35D,Tecoflex 80A, Pellethane 80ASAME
Extrusion ColorantsClear/Natural or PurpleSAME
MarkerbandPlatinum/Iridium (90% Pt, 10% Ir)SAME
CoatingSRDX Harmony (proprietary)SAME
Dimensions
Proximal OD0.084 in. maxSAME
Proximal ID0.068 in. minSAME
Distal OD0.084 in. maxSAME
Distal ID0.068 in. minSAME
Effective Lengths115, 120, 125, 127, 132 cmSAME
Distal Flex Length30 cmSAME
Coating Length30 cmSAME
Predicate: ACE 68Subject: ACE 68
Tip ShapeStraightSAME
Accessories
Peelable SheathPTFESAME
Rotating HemostasisValvePolycarbonate, silicone o-ringSAME
Shaping Mandrel0.038 in. OD stainless steelSAME
Packaging Materials
PouchPolyester/Polyethylene/TyvekSAME
Packaging HoopPolyethyleneSAME
Packaging Tray(Kit configuration)Polyethylene terephthalate, PolystyreneSAME
Packaging CardPolyethyleneSAME
Display CartonSBS PaperboardSAME
Other
SterilizationEOSAME
Shelf-Life36 Months8 Months
UseSingle use, disposableSAME

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Penumbra company logo. The word "Penumbra" is written in red font. To the right of the word is a red circle with a white "P" in the center.

Device Description 1.8

The Penumbra System ACE 68 Reperfusion Catheter is a component to the currently available Penumbra System. The ACE 68 Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel in the neurovasculature. ACE 68 is provided sterile, non-pyrogenic, and intended for single use only.

Indications for Use 1.9

The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.

1.10 Summary of Non-Clinical Data

As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the devices follows.

Included in this section is a summary description of the testing, which substantiates the performance of the subject ACE 68 Reperfusion Catheter as well as its substantial equivalence to the predicate devices:

  • Design Verification (Bench-Top Testing) .

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

  • . Shelf Life Testing
    The subject ACE 68 Reperfusion Catheters met all established requirements.

1.10.1 Design Verification (Bench-Top Testing)

The physical and mechanical properties of the ACE 68 Reperfusion Catheter were assessed using standard test methods and pre-determined acceptance criteria. The following tests were performed and all tests passed successfully:

AttributeSpecificationResults
Packaging InspectionConfirm the packaging and dimensions of the units meet all
Dimensional/Visualproduct specifications.Pass
Kink ResistanceNo kinking when formed in a defined radius
Markerband VisibilityThe markerband is fluoroscopically visible
Simulated Use (IntracranialAccess, Vessel Access EntryPerformance & ClotRemoval)Simulated use testing of the Reperfusion Catheter and Separatorwas performed with accessory devices in an anatomical modelwhich simulated the tortuosity of the neurovasculature. Deviceswere delivered through the tortuous anatomical model toevaluate the effectiveness of the devices to remove clots and thatthe Reperfusion Catheter does not collapse under vacuum.Pass
Particulate testing≥ 10 um will be ≤ 6000 particlesPass
≥ 25 um will be ≤ 600 particlesPass
≥ 75 um will be measured for informational purposes only(FIPO)FIPO
≥ 125 um will be measured for informational purposes only(FIPO)FIPO
Coating IntegrityCoating has not delaminated, peeled, or flaked prior to or aftersimulated use particulate testingPass
Hub Air AspirationNo leaks detected when vacuum is pulled on the injection lumenPass
Static Burst Pressure Test45 psi for 30 sec minimumPass
ACE 68 / Sheath or 8FGuide Catheter FrictionForceMinimum value per specificationPass
ACE 68 / 0.014 in.Guidewire Friction ForceMinimum value per specificationPass
Joint sections bond strengthMinimum value per specificationPass
Hub to shaft tensile strengthMinimum value per specificationPass
Hub to hypotube tensilestrengthMinimum value per specificationPass
Elongation to failureElongation ≥ 5%Pass
TorsionNumber of turns will be recorded for informational purposesonly.FIPO
CorrosionNo visible corrosion on Reperfusion Catheter immediately aftercorrosion testing procedurePass

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red sans-serif font, followed by a red circle with a white "P" inside. The logo is simple and modern, with a focus on the company name. The red color is bold and eye-catching, while the white "P" provides a contrast and adds visual interest. The overall effect is clean and professional.

1.10.2 Shelf Life Testing

The device stability and packaging integrity of the ACE 68 Reperfusion Catheter were assessed using standard test methods and pre-determined acceptance criteria. Devices underwent transportation conditioning per ASTM D4169 and accelerated aging equal to 8 months. Results of successful testing verify the ACE 68 Reperfusion Catheter may be labeled with an 8-month shelf life.

1.11 Summary of Substantial Equivalence

The subject Penumbra System ACE 68 Reperfusion Catheter is substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, shelf-life, packaging and sterilization processes.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).