(157 days)
PICO 14 is indicated for patients who would benefit from a suction device (NPWT) as it may promote would healing via removal of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcers (such as diabetic or pressure)
- Flaps and grafts
- Closed Surgical incisions
PICO 14 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.
PICO 14 Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable, electro-mechanical pump system connected through a flexible tube to a super-absorbent, gentle adhesive dressing. Secondary fixation strips are also provided to ensure an adequate seal is achieved. The pump, the dressing and fixation strips are supplied sterile and for single use.
The provided text describes the PICO 14 Single Use Negative Pressure Wound Therapy System and its substantial equivalence to the predicate device, PICO 7. However, the information presented primarily focuses on demonstrating equivalence through bench testing and a clinical study on patient outcomes rather than a traditional setup of acceptance criteria and a study dedicated to proving the device meets those specific criteria in a formal, AI-centric evaluation.
The device itself is a powered suction pump for wound therapy, not an AI/ML-driven diagnostic or treatment device, which explains the absence of some of the requested AI-specific information.
Here's an attempt to extract and present the information as requested, aligning it as closely as possible to the nature of the provided text:
Acceptance Criteria and Device Performance Study for PICO 14 Single Use Negative Pressure Wound Therapy System
The PICO 14 device is an electro-mechanical pump system and does not appear to involve AI/ML. Therefore, typical AI-specific acceptance criteria (like sensitivity, specificity, AUC) and an AI-specific study structure are not applicable to the information provided. Instead, "acceptance criteria" here refers to the demonstrated performance aspects that allow the device to be considered substantially equivalent and safe/effective for its intended use. The "study" refers to the non-clinical bench tests and the clinical evidence presented for the device.
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a medical device clearance and not an AI/ML product, the "acceptance criteria" are derived from the demonstrated functionalities and safety profiles required for its intended use and comparison to the predicate device.
| Acceptance Criteria Category | Specific Criterion (Implicit/Explicit) | Reported Device Performance/Evidence |
|---|---|---|
| Functional Performance | Ability to function effectively for 14 days without failure. | Bench tests successfully completed, representing worst-case scenarios. Wound model tests demonstrated the system can manage wound exudate and deliver Negative Pressure over 14 days. |
| Ability to manage wounds up to 4.5 cm deep. | Bench tests successfully completed, representing worst-case scenarios. Wound model tests demonstrated the system can manage wounds up to 4.5 cm deep. | |
| Maintain nominal negative pressure of -80mmHg and same pressure profile as predicate. | The PICO 14 system delivers the same nominal negative pressure of -80mmHg and the same pressure profile as the PICO 7 System. This was confirmed as the NPWT delivery profile is unchanged. | |
| Software changes do not impact the function of NPWT delivery. | Only minor convenience updates to software for maximum pump down times and 14-day use. No changes to the function of NPWT delivery. | |
| Safety and Biocompatibility | Biocompatibility of materials. | PICO 14 uses the same dressings and materials as the predicate PICO Single Use NPWT System; materials are identical, thus no new biocompatibility data required. Evaluated according to BS EN ISO 10993 and FDA guidance. |
| Electrical safety and Electromagnetic Compatibility (EMC). | PICO 14 has the same electrical characteristics as the predicate system (K180698). Software updates do not impact electrical safety. | |
| Acceptable adverse event rate. | In a clinical study, the adverse event rate in PICO groups (depth < 2cm: 1.2%, depth > 2cm: 2.5%) was lower than the non-PICO group (6.6%). No PICO device-related adverse events in the > 2cm group. | |
| Human Factors/Usability | No new user tasks or impact on critical tasks. | The minor changes are not considered to change the safety profile, raise new safety/efficacy questions, reflect new user tasks, or impact critical tasks. User interface remains similar with minor changes to appearance/packaging. |
| Duration of Use | Extended use time from 7 days to 14 days. | Bench tests demonstrated 14-day functionality. Clinical study data encompassed a longer observation period, and the device was able to manage exudate and pressure over 14 days. An additional set of batteries is provided to support the extended use. |
2. Sample Size and Data Provenance for the Test Set (Clinical Evidence)
- Sample Size for Clinical Evidence:
- Total patients followed: 3,159
- Patients using PICO device: 917
- PICO patients with wound depths > 2 cm: 409 (comprising 233 DFUs, 18 PUs, 147 dehisced surgical wounds, and 11 VLUs).
- Patients discontinuing PICO: 32 out of 917 (2.8%).
- Data Provenance: Retrospective, from two community care access centers.
- Country of Origin: Not explicitly stated but "community care access centers" suggests a real-world setting. Assuming within the country where the company operates or where the study was managed, which might be the UK (Smith & Nephew Medical Limited is in Hull, UK) or another location.
3. Number of Experts and Qualifications for Ground Truth for the Test Set
The clinical evidence provided appears to be an observational study of patient outcomes rather than a ground truth establishment by experts for a specific test set in the context of an AI device.
- Number of Experts: Not specified. The study involved patient care in community care access centers, suggesting healthcare professionals (e.g., nurses, doctors) were involved in patient admission, care, and data recording including wound depth and healing.
- Qualifications of Experts: Not specified, but implied to be healthcare professionals involved in wound care. Clinicians decided if exudate rate increased beyond indicated levels.
4. Adjudication Method for the Test Set
Not applicable. The clinical study was an observational study of patient outcomes, not an evaluation where expert adjudication of individual cases was performed to establish a definitive ground truth for device performance.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. The device is a physical therapy system, not an imaging or diagnostic AI tool that would typically involve human readers interpreting cases with and without AI assistance. The clinical evidence compared patient outcomes with PICO vs. non-PICO treatment.
6. If a Standalone (algorithm only without human-in-the-loop performance) was done
Yes, in a sense, the functionality of the device itself (the PICO 14 system delivering negative pressure) was evaluated in "standalone" bench tests, demonstrating its mechanical and software performance over 14 days and for deep wounds. This is not "algorithm only" in the AI sense, but rather the device's inherent function.
7. The Type of Ground Truth Used (Clinical Evidence)
The "ground truth" for the clinical evidence was based on patient outcomes data (time to healing, number and duration of dressing changes, and adverse events) and clinically observed wound parameters (wound depth, exudate levels, and clinicians' decisions).
8. Sample Size for the Training Set
Not applicable. The PICO 14 is an electro-mechanical device. There is no mention of an AI/ML algorithm that requires a "training set" in the context of machine learning. The "training" for the device's development would be engineering and design iterations.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no AI/ML component or "training set" described for this medical device in the provided text.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 5, 2019
Smith & Nephew Medical Limited Sam Greenhalgh Senior Regulatory Affairs Specialist 101 Hessle Road Hull, HU3 2BN Gb
Re: K191760
Trade/Device Name: PICO 14 Single Use Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP, Dated: November 1, 2019 Received: November 4, 2019
Dear Sam Greenhalgh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kimberly M. Ferlin, Ph.D. Assistant Director (acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
PICO 14 Single Use Negative Pressure Wound Therapy System
Indications for Use (Describe)
PICO 14 is indicated for patients who would benefit from a suction device (NPWT) as it may promote would healing via removal
of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcers (such as diabetic or pressure)
- Flaps and grafts
- Closed Surgical incisions
PICO 14 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY
PICO 14 Single Use Negative Pressure Wound Therapy System
| General Information | |
|---|---|
| Owner Name/Address: | Smith & Nephew Medical Limited101 Hessle Road,HullHU3 2BNUnited Kingdom |
| EstablishmentRegistration Number: | 8043484 |
| Contact Person: | Sam Greenhalgh, Senior Regulatory AffairsSpecialist |
| Phone Number | +44 1482 673436 |
| Date Prepared: | December 4th, 2019 |
| Device Description | |
| Trade Name: | PICO 14 Single Use Negative Pressure WoundTherapy System |
| Common or UsualName: | Negative Pressure Wound Therapy poweredsuction pump |
| Classification Name: | Powered suction pump (21 CFR 878.4780) |
| Medical Device Class: | Class II |
| Product Code: | OMP |
| Predicate Device Information | |
| 510(k) Number: | K180698 |
| Device name: | PICO 7 Single Use Negative Pressure WoundTherapy System |
| Clearance Date: | August 21, 2018 |
{4}------------------------------------------------
Device Description
PICO 14 Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable, electro-mechanical pump system connected through a flexible tube to a super-absorbent, gentle adhesive dressing. Secondary fixation strips are also provided to ensure an adequate seal is achieved. The pump, the dressing and fixation strips are supplied sterile and for single use.
Description of the Patient Population for which the device is intended:
The target patient population for which the device is intended is unchanged between the subject device (PICO 14) and the predicate device (PICO 7). The patient population is: Patients who would benefit from a suction device (negative pressure wound therapy).
Indications for Use
PICO 14 is indicated for patients who would benefit from a suction device (NPWT) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- . Chronic
- Acute .
- Traumatic ●
- Subacute and dehisced wounds ●
- Partial-thickness burns ●
- Ulcers (such as diabetic or pressure) ●
- Flaps and grafts ●
- Closed surgical incisions ●
PICO 14 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.
{5}------------------------------------------------
Comparison of Subject and Predicate Devices
| Comparison Between PICO 7 (Predicate Device) and PICO 14 (Subject Device) | |
|---|---|
| Software | The only changes to the software are minor convenience updates to the maximum pump down times and to allow for use over a 14 day period instead of 7 days (predicate device). There are no changes to the function of the system with respect to the delivery of NPWT, |
| Packaging format | The packaging format been modified to provide a more simple manufacturing process. The materials used in the packaging are unchanged however the IFU will now be in an A4 book format and the carton will be a single box with an insert rather than the book format. |
| Indications for Use statement | PICO 7 and PICO 14 are identical. |
| Instructions for use | The Instructions for Use are largely the same as the predicate system. The main changes are to provide clarity with respect to wound depth. The updated wording is clear that PICO 14 is suitable for use on wounds up to 4.5cm deep. Wounds greater than 0.5cm (1/4 in.) in depth are likely to require a foam or gauze negative pressure wound therapy filler to ensure adequate treatment of all the wound surface. Wounds treated with the PICO 14 system should generally be no more than 4.5cm (1 $1\frac{3}{4}$ in.) in depth and must not contain exposed arteries, veins, nerves or organs. |
| Technological principal for delivering the negative pressure wound therapy | PICO 7 and PICO 14 are identical |
| PICO Pump | Embossing and symbol changes to change from 7 day use to 14 day use and change to icon bar color (from gray to black). |
| Physical components of the pumps, dressings and tubing | PICO 7 and PICO 14 are identical |
| Batteries | While an additional set of batteries are now provided,The predicate system could require a battery change,which was already communicated to the user. Two setsof batteries will be provided in the case that the userneeds to change the batteries (the predicate deviceprovides one set of batteries). One of the battery setswill be provided sterile within the primary pump traypackaging (identical to the predicate device) and one setis provided non-sterile as an extra set. |
| System-use time | The system use time has changed from 7 days(predicate device) to 14 days (subject device). |
| User interface | PICO 14 remains similar, with only minor changes to theproduct appearance and packaging |
| Electrical Safety andElectromagneticCompatibility (EMC) | PICO 7 and PICO 14 are identical |
{6}------------------------------------------------
These minor changes discussed above are not considered to change the safety profile of the system, nor do they raise any new questions of safety or efficacy.
The changes do not reflect any new user tasks, nor do they impact any of the critical tasks associated with the use of PICO and the users are not expected to complete any new tasks.
Given the identical nature of the physical aspects of the predicate system and PICO 14, much of the data to support the product is the same as the predicate data.
{7}------------------------------------------------
Non-Clinical Tests (Bench)
The PICO 14 Single Use Negative Pressure Wound Therapy System delivers the same nominal negative pressure of -80mmHg and the same pressure profile as the PICO 7 System. The only functional changes are to increase the use time from 7 to 14 days, two repetitions of the dynamic duty cycle used in PICO 7 to maximize power use of 14 days instead of 7 days and changes to the maximum Initial Pump Down (IPD) and Maintenance Pump Down (MPD) times.
Since the system continues to act exactly as with the PICO 7 system, the bench testing provided is to demonstrate two aspects:
-
- The system can function for 14 days without failure
-
- The system continues to function when used on wounds up to 4.5cm deep
14 Day System Use
Bench tests have successfully been completed and represent worst case scenarios. The wound model tests demonstrate that, in a clinically representative scenario, the PICO 14 system can manage wound exudate and deliver Negative Pressure over 14 days.
Deep Wound Model Testing
Bench tests have successfully been completed and represent worst case scenarios. The wound model tests demonstrate that, in a clinically representative scenario, the PICO 14 system can manage wounds up to 4.5cm deep.
The testing utilized PICO pumps from the predicate submission; as discussed throughout this 510(k), the NPWT delivery profile is unchanged between the subject and predicate devices, hence the data from the predicate system as applicable to the subject system.
Clinical Evidence Summary
Clinical evidence was provided to support the use of PICO 14 in wounds with depths up to 4.5 cm. The study population included all patients at two community care access centers, who were admitted and received care for a chronic wound (pressure ulcer (PU), diabetic foot ulcer (DFU), venous leg ulcer (VLU), surgical ulcer (open incision)) from March 31, 2016 until March 31, 2018. 3,159 patients were followed from admission to healing. 917 patients used the PICO device and 409 of those patients had wound depths greater than 2 cm in which 233 were DFUs, 18 were PUs, 147 were dehisced surgical wounds, and 11 were VLUs. Time
{8}------------------------------------------------
K191760
to healing, number and duration of dressing changes, and adverse events were recorded.
The adverse event rate in the PICO groups (depth < 2cm 1.2% and depth >2cm 2.5%) was lower than the non-PICO group (6.6%). There were no PICO device related adverse events in the > 2cm group. Data were provided on the 32 of the 917 (2.8%) patients that discontinued PICO. 11 of those patients discontinued due to infection (1) or drainage (10) and for the drainage, the clinician decided the exudate rate had increased beyond the indicated low to moderate exudate rate. The data provided demonstrates safety of the device for wounds with depths up to 4.5 cm.
Biocompatibility, Sterilization and Electrical Testing
PICO 14 Single Use Negative Pressure Wound Therapy System has been evaluated according to the Biological Evaluation of Medical Devices Standard BS EN ISO 10993, with particular reference to Part 1 (2009): Evaluation and testing within a risk management process and FDA guidance 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". PICO 14 Single Use NPWT Use System uses the same dressings with soft port as the predicate PICO Single Use NPWT System; the materials are identical and therefore no new biocompatibility data are required. Additionally, the exterior pump components remain identical to the predicate device already cleared by the FDA.
The PICO 14 Single Use Negative Pressure Wound Therapy System has the same electrical characteristics as the predicate system (K180698), with the only changes being a software update. The software changes will not impact on the electrical safety of the system.
Conclusions
In establishing substantial equivalence to the predicate devices, Smith & Nephew Medical Ltd evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. It was confirmed that the PICO 14 Single Use Negative Pressure Wound Therapy System continues to use the same materials as the predicate system and provides the same user interface.
Performance testing has been completed to demonstrate that the PICO 14 Single Use Negative Pressure Wound Therapy System is substantially equivalent to the predicate for the intended use.
§ 878.4780 Powered suction pump.
(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.