K Number
K180698
Date Cleared
2018-08-21

(158 days)

Product Code
Regulation Number
878.4780
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PICO 7 is indicated for patients who would benefit from a suction device (NPWT) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.

Appropriate wound types include:

  • Chronic
  • Acute
  • Traumatic
  • Subacute and dehisced wounds
  • Partial-thickness burns
  • Ulcers (such as diabetic or pressure)
  • Flaps and grafts
  • Closed Surgical incisions

PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

Device Description

PICO 7 Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable, electro-mechanical pump system connected through a flexible tube to a super-absorbent, gentle adhesive dressing. Secondary fixation strips are also provided to ensure an adequate seal is achieved. The pump, the dressing and fixation strips are supplied sterile and for single use. PICO 7 was designed to deliver negative pressure wound therapy to low and to moderate exudate levels of acute and chronic wound types. PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.

AI/ML Overview

The PICO 7 Single Use Negative Pressure Wound Therapy System is intended to promote wound healing by removing low-to-moderate levels of exudate and infectious materials. The device was evaluated against specific acceptance criteria for its performance indicators and safety. The study that proves the device meets the acceptance criteria is detailed in the "Non-Clinical Tests (Bench)" section of the 510(k) summary (pages 6-10).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Wound Model TestingDressings manage fluid at low (0.6 g/cm²/24 hours) and moderate (1.1 g/cm²/24 hours) exudate flow rates for a simulated wound of 25% of the dressing absorbent pad area. Dressings maintain negative pressure within -60 mmHg to -100 mmHg for a minimum of 95% of test time.All dressings managed fluid at low and moderate flow rates. Dressings maintained negative pressure within -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test.
Dressing Full IndicatorDevice detects blockages and activates the "dressing full" indicator no later than 2 hours after occlusion.All test results showed that when dressings reached maximum capacity, the device detected blockages and activated the "dressing full" indicator no later than 2 hours after the occlusion occurred.
Air Leak IndicatorAir leak indicators activate at specified time points when a high leak rate is introduced, in line with requirements for initial application (to achieve -60mmHg to -100mmHg) and maintenance of therapeutic pressure.Air leak indicator was activated at specified time points in line with test requirements, representing initial application to achieve the therapeutic negative pressure range of -60mmHg to -100mmHg (target -80mmHg), and maintenance of the therapeutic pressure range.
Low Battery Indicator"Battery low" indicator activates and flashes for a minimum of 12 hours prior to "critical low battery" state. Once in "critical low battery" state, pressing play/pause button enables "battery low" indicator to flash 3 times."Battery low" indicator activated and flashed for a minimum of 12 hours prior to the device entering "critical low battery" state. Once in "critical low battery" state, pressing the play/pause button enabled the "battery low" indicator to flash 3 times.
Acoustic TestingAverage noise level emitted during normal operation does not exceed recommended levels (e.g., in line with ANSI/AAMI HE75:2009/(R)/2013 for frequent peaks at night in a convalescent area). Maximum noise (LAFmax) does not exceed 35dB.Average noise level during normal operation was 27.6dB. Maximum noise (LAFmax) did not exceed 35dB, in accordance with ANSI/AAMI HE75:2009/(R)/2013.
Electromagnetic Compatibility & Electrical SafetyCompliance with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 62366 standards.The device was tested and found to be in compliance with all listed standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 62366.
Shelf LifeMaintain functionality and original performance criteria (fluid management, negative pressure maintenance) over the specified shelf life (initially 6, extended to 18 months).All dressings managed fluid at low and moderate flow rates. Dressings maintained negative pressure within -60 mmHg to -100 mmHg for a minimum of 95% of the test time. All specification test results complied with acceptance criteria. Functionality maintained over extended duration, allowing for an 18-month shelf life.

2. Sample Size Used for the Test Set and Data Provenance:

The document describes non-clinical (bench) tests. Specific numerical sample sizes for each test are not consistently provided in this summary. For example, "Wound model testing was conducted on PICO 7 Kits (10x20cm and 25x25cm dressings) at low and moderate flow rate..." implies testing was done on multiple kits/dressings but doesn't state the exact number. Similarly for "Dressing Full Indicator" and "Low Battery Indicator" tests, it states "All test results showed...", suggesting multiple tests were carried out, but without giving a precise count.

Data provenance: These are bench tests, meaning they are conducted in a controlled laboratory environment rather than on human subjects. Therefore, there is no country of origin for patient data, nor is it retrospective or prospective in the clinical sense.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. These were non-clinical, bench testing experiments conducted against engineering and performance specifications, not based on expert-established ground truth from clinical cases.

4. Adjudication Method for the Test Set:

Not applicable. This pertains to clinical studies often involving expert interpretation of medical images or patient outcomes, not bench testing against pre-defined engineering specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No. This 510(k) summary only describes non-clinical bench testing. MRMC studies are typically for evaluating diagnostic or predictive AI systems in a clinical context.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

The tests described are for the device's standalone performance, albeit within a simulated or controlled bench environment rather than a clinical human-in-the-loop setting. The device's indicators (dressing full, air leak, low battery) and its physical performance (fluid management, pressure maintenance, noise) are tested "algorithm only" in the sense that their automated function is evaluated.

7. The Type of Ground Truth Used:

The ground truth used for these non-clinical tests was based on pre-defined engineering specifications, performance metrics, and compliance standards. For example:

  • Wound model testing: Based on specified fluid flow rates (0.6 and 1.1 g/cm²/24 hours) and negative pressure ranges (-60 mmHg to -100 mmHg).
  • Indicator testing: Based on the device correctly activating indicators within specified timeframes or conditions (e.g., "dressing full" within 2 hours, "low battery" for 12 hours).
  • Acoustic testing: Based on noise level limits defined by ANSI/AAMI HE75:2009/(R)/2013 (e.g., maximum 35dB LAFmax).
  • EMC/Electrical Safety: Compliance with specific IEC and AAMI/ANSI standards.

8. The Sample Size for the Training Set:

Not applicable. This device is a mechanical/electronic negative pressure wound therapy system, not an AI/machine learning algorithm requiring a "training set" in the typical sense.

9. How the Ground Truth for the Training Set Was Established:

Not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 21, 2018

Smith & Nephew Medical Limited % Kulsum Master Director Regulatory Affairs, US Region Smith & Nephew 7000 West William Cannon Drive Austin, Texas 78735

Re: K180698

Trade/Device Name: PICO 7 Single Use Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: May 10, 2018 Received: May 14, 2018

Dear Kulsum Master:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Krause -S

  • for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180698

Device Name

PICO 7 Single Use Negative Pressure Wound Therapy System

Indications for Use (Describe)

PICO 7 is indicated for patients who would benefit from a suction device (NPWT) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.

Appropriate wound types include:

  • Chronic
  • Acute
  • Traumatic
  • Subacute and dehisced wounds
  • Partial-thickness burns
  • Ulcers (such as diabetic or pressure)
  • Flaps and grafts
  • Closed Surgical incisions

PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K180698

510(K) SUMMARY

PICO 7 Single Use Negative Pressure Wound Therapy System

General Information
Submitter Name/Address:Smith & Nephew Medical Limited
101 Hessle Road,
Hull
HU3 2BN
United Kingdom
Establishment Registration Number:8043484
Contact Person:Lavinia Tompkins, Senior Regulatory Affairs Specialist
Phone Number+44 1482 225181
Date Prepared:July 16, 2018
Application Correspondent:Smith & Nephew Inc.
7000 West William Cannon Drive,
Austin,
Texas, 78735,
USA
Contact Person:Kulsum Master, Director Regulatory Affairs, US Region
Phone Number:+1 512-358-5720

{4}------------------------------------------------

Device Description
510(k) Number:K180698
Trade Name:PICO 7 Single Use Negative PressureWound Therapy System
Common or Usual Name:Negative Pressure Wound Therapypowered suction pump
Classification Name:Powered suction pump (21 CFR878.4780)
Regulatory Class:Class II
Product Code:OMP

Predicate Device Information

Primary Predicate DeviceSecondary Predicate Device
510(k) Number:K172005K151436
Device:PICO 7 Single Use Negative Pressure Wound Therapy SystemPICO Single Use Negative Pressure Wound Therapy System
Clearance Date:February 02, 2018January 28, 2016

Device Description

PICO 7 Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable, electro-mechanical pump system connected through a flexible tube to a super-absorbent, gentle adhesive dressing. Secondary fixation strips are also provided to ensure an adequate seal is achieved. The pump, the dressing and fixation strips are supplied sterile and for single use. PICO 7 was designed to deliver negative pressure wound therapy to low and to moderate exudate levels of acute and chronic wound types. PICO 7 Single Use Negative

{5}------------------------------------------------

Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.

Indications for Use

PICO 7 is indicated for patients who would benefit from a suction device (NPWT) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.

Appropriate wound types include:

  • Chronic .
  • Acute .
  • Traumatic .
  • Subacute and dehisced wounds ●
  • Partial-thickness burns ●
  • Ulcers (such as diabetic or pressure) ●
  • Flaps and grafts ●
  • Closed surgical incisions .

PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.

Comparison between New and Primary Predicate Device

The Indications for Use statement of the PICO 7 Single Use Negative Pressure Wound Therapy System is identical to that of the primary and secondary predicate devices. The technological principal for delivering the negative pressure wound therapy for both the subject and predicate devices are identical. The main differences between the subject device and the primary predicate device are:

{6}------------------------------------------------

    1. Introduction of a check valve (1-way valve) into the PICO 7 device to isolate the internal air leak caused by the normal operation of the pump mechanism from the pressure sensor.
    1. To enable the addition of the check valve (1-way valve), the manifold within the device casing required resizing.
    1. Minor software amendments which have been submitted as part of this 510(k) application. The amendments do not affect the performance, function, or indicators of the device.
    1. Extension of shelf life of PICO 7 system from 6 to 18 months.

The following testing has been completed on the subject device:

Non-Clinical Tests (Bench)

Wound Model Testing

Wound model testing was conducted on PICO 7 Kits (10x20cm and 25x25cm dressings) at low and moderate flow rate in vertical and horizontal positions.

The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a low (0.6 g/cm²/24 hours) and a moderately exuding wound (1.1 g/cm²/24 hours) respectively for a simulated wound of 25% of the dressing absorbent pad area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the results of the tests it can be concluded that the PICO 7 systems can be used for their intended purposes.

Visual Indicator Performance Testing

  • . Dressing Full Indicator
    Testing was conducted on the PICO 7 system using water and simulated wound fluid to simulate a worst case scenario where the dressing reaches

{7}------------------------------------------------

maximum capacity and the filter in the dressing becomes blocked with fluid.

All test results showed when the dressings reached their maximum capacity the subject device was able to detect blockages and activate the "dressing full" indicator no later than 2 hours after the occlusion had occurred.

Based on the test results, it can be concluded that the PICO 7 devices can be used for their intended purposes.

Air Leak Indicator .

Air leak testing was conducted on PICO 7 by introducing deliberate air leaks into the system to show that when the system has a high leak rate, the leak indicators activate to alert the user to inspect the dressing application and address the point of air ingress/ air leak.

Test results showed that air leak indicator was activated at specified time points in line with the test requirements representing the initial application of the system to achieve the therapeutic negative pressure range of -60mmHg to -100mmHg (target -80mmHg); and maintenance of the therapeutic pressure range.

Based on the test results, it can be concluded that the PICO 7 devices can be used for their intended purposes.

Low Battery Indicator .

Low battery indicator testing of the PICO 7 system at pre-defined time periods which allow users to respond to the battery state.

The device was operated until the "battery low" indicator was activated, once activated, the device was continued to be operated for a minimum of a further 12 hours to ensure that the device reached "critical battery low" state.

{8}------------------------------------------------

All test results showed that the "battery low" indicator activated and flashed for a minimum of 12 hours prior to the device entering "critical low battery" state and that once the device had entered the "critical low battery" state, pressing the play/pause button of the device enabled the "battery low" indicator to flash 3 times.

Based on the test results, it can be concluded that the PICO 7 devices can be used for their intended purposes.

Acoustic Testing

Acoustic testing was conducted to demonstrate that the level of noise emitted from the subject device whilst addressing the internal air leak caused by the normal operation of the pump mechanism remained acceptable as defined within ANSI/AAMI HE75:2009/(R)/2013 in line with the recommendation for frequent peaks at night in a convalescent area.

Test results showed that during the normal operation of the pump mechanism, the subject device emitted an average noise level of 27.6dB and that the maximum noise emitted by the device (LAFmax) did not exceed 35dB in accordance with ANSI/AAMI HE75:2009/(R)/2013.

Based on the test results, it can be concluded that the PICO 7 devices can be used for their intended purposes.

Electromagnetic Compatibility and Electrical Safety Testing

PICO 7 Single Use Negative Pressure Wound Therapy System has been tested and assessed respectively and has been found to be in compliance with the following standards:

  • AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012 C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment -

{9}------------------------------------------------

Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)

  • IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment Part . 1-2: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
  • IEC 60601-1-6 Edition 3.1 2016-10 Medical Electrical Equipment Part . 1-6: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Usability
  • IEC 60601-1-11 Edition 2.0 2015-01 Medical Electrical Equipment Part . 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
  • IEC 62366 Medical Devices Application of Usability Engineering to . Medical Devices

Shelf Life

The PICO 7 system has been evaluated using real time and accelerating ageing challenge conditions and a combination of simulated wound model testing as well as specification testing to demonstrate system functionality over the duration of the shelf life.

Test results showed that all dressings managed fluid at a flow rate modelling a low (0.6 g/cm²/24 hours) and a moderately exuding wound (1.1 g/cm²/24 hours) respectively for a simulated wound of 25% of the dressing absorbent pad area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Furthermore, all specification test results complied with the acceptance criteria as defined within the test requirements. The results of all tests showed that the subject device was able to maintain its functionality over the duration of the ageing study and that a

{10}------------------------------------------------

shelf life of 18 months can be applied to the PICO 7 system. Therefore, it can be concluded that the PICO 7 systems can be used for their intended purposes.

Conclusions

In establishing substantial equivalence to the predicate device, Smith & Nephew Medical Ltd evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing, service life testing and electrical safety testing has been completed to demonstrate that the PICO 7 Single Use Negative Pressure Wound Therapy System is substantially equivalent to the primary and secondary predicate devices for the intended use.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.