K Number
K191305
Date Cleared
2019-12-19

(219 days)

Product Code
Regulation Number
870.1210
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Poseidon Microcatheter is intended for use in peripheral and coronary vessels for diagnostic and interventional procedures including small or superselective anatomy.

Device Description

The Poseidon Microcatheter is a single lumen catheter constructed using an internal lubricious layer encapsulated by a metallic braid and lubricious outer layer. The distal tip includes a radiopaque marker band, and the distal 60 - 90 cm of the microcatheter has a hydrophilic coating.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the Heraeus Poseidon Microcatheter. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific performance acceptance criteria via a clinical study that would involve expert readers, complex adjudication, or MRMC studies.

Therefore, many of the questions asked cannot be answered directly from this type of FDA submission. The document details bench testing against general regulatory standards and comparisons to a predicate device's characteristics.

Here's a breakdown of what can and cannot be answered based on the provided text:

What can be answered:

  • Acceptance Criteria and Reported Device Performance (as listed in bench tests): The document provides a table of in vitro bench tests conducted and states that "Samples passed the acceptance criteria and the device was found to be [performance outcome]." The specific numerical acceptance criteria themselves are generally not detailed in this type of summary, but the qualitative outcome is stated as "passed."

    TestAcceptance Criteria (Qualitative)Reported Device Performance
    CytotoxicityNon-cytotoxicPassed acceptance criteria; non-cytotoxic
    SensitizationNo sensitization responsePassed acceptance criteria; did not elicit sensitization
    IrritationNo irritation responsePassed acceptance criteria; did not elicit irritation
    Systemic ToxicityNon-toxicPassed acceptance criteria; non-toxic
    PyrogenicityNon-pyrogenicPassed acceptance criteria; non-pyrogenic
    HemolysisNon-hemolyticPassed acceptance criteria; non-hemolytic
    HemocompatibilityComparable to predicate comparison articlePassed acceptance criteria; comparable to predicate
    ThrombogenicityNon-thrombogenicPassed acceptance criteria; non-thrombogenic
    Corrosion ResistanceNo corrosive responsePassed acceptance criteria; did not elicit corrosive response
    DimensionsComparable to predicatePassed acceptance criteria; comparable to predicate
    Package Integrity TestingComparable to predicatePassed acceptance criteria; comparable to predicate
    Tensile StrengthComparable to predicatePassed acceptance criteria; comparable to predicate
    Torque StrengthComparable to predicatePassed acceptance criteria; comparable to predicate
    Kink ResistanceComparable to predicatePassed acceptance criteria; comparable to predicate
    RadiopacityComparable to predicatePassed acceptance criteria; comparable to predicate
    Flow RateComparable to predicatePassed acceptance criteria; comparable to predicate
    Burst Pressure/Freedom from LeakageComparable to predicatePassed acceptance criteria; comparable to predicate
  • Sample size used for the test set: Not explicitly stated as "sample size" in the context of cases/patients, but rather in terms of number of device units or samples tested for each bench test. These are standard in vitro tests, not human studies.

  • Data provenance: For bench tests, this would be laboratory data.

  • Training Set (for AI-related questions): This device is a microcatheter, not an AI/ML device. Therefore, there is no training set in the context of an algorithm.

What cannot be answered from this document:

  • Number of experts used to establish ground truth: Not applicable for a physical medical device clearance based on substantial equivalence and bench testing.
  • Qualifications of those experts: Not applicable.
  • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, this is not an AI/imaging device.
  • Effect size of how much human readers improve with AI vs. without AI assistance: Not applicable.
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth in this context refers to the defined acceptance criteria for physical properties and biological responses established through international standards (ISO, ASTM) and FDA guidance for medical devices, not clinical outcomes or expert consensus on medical images.
  • Sample size for the training set: Not applicable (no AI/ML algorithm).
  • How the ground truth for the training set was established: Not applicable (no AI/ML algorithm).

In summary: The provided 510(k) summary demonstrates the safety and effectiveness of the Heraeus Poseidon Microcatheter through a comparison to a legally marketed predicate device and extensive in vitro bench testing against established international standards. It is not an AI/ML device, and thus the questions related to clinical study design, human readers, ground truth establishment by experts, and training/test sets for algorithms are not applicable to this submission.

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).