(248 days)
Not Found
No
The description focuses on the physical characteristics and performance testing of a microcatheter, with no mention of AI or ML capabilities.
No.
The device is described as a microcatheter intended for diagnostic and interventional procedures, used to access small vessels. While it may be used in interventional procedures, its primary function as described is for access and delivery rather than directly providing a therapeutic effect itself (e.g., delivering a drug, ablating tissue, or performing a corrective action). It acts as a tool or conduit for a procedure, not a therapeutic device in its own right in this context.
Yes
The "Intended Use / Indications for Use" section explicitly states that the microcatheters are intended for "diagnostic and interventional procedures." Additionally, the "Device Description" reiterates their design for "diagnostic and interventional procedures." While they also have interventional uses, the diagnostic aspect makes them a diagnostic device.
No
The device description clearly details a physical medical device (braided microcatheters with hydrophilic coating, radiopaque marker band, specific dimensions) and performance studies related to physical properties and biological interactions, not software functionality.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for "diagnostic and interventional procedures including neuro, peripheral, or coronary use." This describes a device used within the body for visualization and treatment, not for testing samples outside the body.
- Device Description: The description details a physical catheter designed to be inserted into blood vessels. This is consistent with an interventional medical device, not an IVD.
- Input Imaging Modality: The device relies on "Fluoroscopic visualization," which is an imaging technique used to view structures inside the body in real-time. IVDs typically do not involve real-time imaging of the patient's internal anatomy.
- Anatomical Site: The device is used in "Small vessel or superselective anatomy including neuro, peripheral, or coronary use," which are locations within the human body. IVDs analyze samples taken from the body (like blood, urine, tissue), not the body itself.
IVD devices are designed to perform tests on samples taken from the human body to provide information about a person's health status. This device is a tool used during medical procedures within the body.
N/A
Intended Use / Indications for Use
The Cantata 2.9 Microcatheters are intended for use in small vessel or superselective anatomy for diagnostic and interventional procedures including neuro, peripheral, or coronary use.
Product codes (comma separated list FDA assigned to the subject device)
KRA, DQY
Device Description
The Cantata 2.9 Microcatheters are braided microcatheters with hydrophilic coating, designed for use in small vessel or superselective anatomy for diagnostic and interventional procedures including neuro, peripheral, or coronary use. The devices include one radiopaque marker band to assist in fluoroscopic visualization of the catheter during use. The catheters are available in 2.9 French shafts and are available in 100, 110, 135, and 150 centimeter lengths. The devices are compatible with a 0.038 inch (0.97 millimeters) diameter angiographic catheter and can accept 0.018 inch (0.46 millimeters) and 0.021 inch (0.53 millimeters) diameter wire guides.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Small vessel or superselective anatomy (neuro, peripheral, or coronary use)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The following tests were performed:
Air Leakage Testing: Test article is loaded with de-aerated processed water and subjected to a negative pressure after occluding the lumen. Visual inspection for air bubbles is subsequently conducted for fifteen seconds. Conclusion: Testing showed that the catheter does not exhibit air leakage during proper clinical use. Device met the predetermined acceptance criteria.
Liquid Leakage Testing: Test article is loaded with processed water and subjected to a pressure of 300 kPa to 320 kPa after occluding the lumen. Visual inspection for air bubbles is subsequently conducted for a minimum of 30 seconds. Conclusion: Testing shows that the catheter shall not leak liquid during proper clinical use. Testing demonstrated that the device met the predetermined acceptance criteria.
Tensile Testing: A uniaxial tensile load was applied to evaluate all four durometer transition zones of the test article shaft to ensure the resulting force required to break the shaft was greater than 5 Newtons. Conclusion: Testing showed that under proper clinical use of the device the peak load value shall be greater than 5 Newtons. In conformance with ISO 10555-1:1995, the predetermined acceptance criteria were met.
Kink Radius Testing: The distal segment of the test article shaft was passed through a kink radius fixture to create a loop in the device. Both end of the device were then slowly pulled to close the loop until a kink was formed in the shaft. This process was recorded by an image analysis system which was used to determine the radius of the loop immediately prior to kinking. Conclusion: Testing showed that under proper clinical use the kink radius of the device is
§ 870.1210 Continuous flush catheter.
(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows a logo for Cook Medical. The word "COOK" is in large, bold, sans-serif font at the top of the logo. Below "COOK" is the word "MEDICAL" in a smaller, bolder, sans-serif font. The logo is black and white.
510(K) Summary
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA 812.339.2235 touring 800,457,4500 WWW.COOKMEDICAL.COM
Cantata® 2.9 Microcatheter
Cook Incorporated
Bloomington, IN 47404 Jennifer A. Richardson, MA
(812) 339-2235 ext. 2370
jennifer.richardson@cookmedical.com
750 Daniels Way
Submitter Information:
Applicant: Address:
Contact: Email: Contact Phone Number: Contact Fax Number:
Date Prepared:
December 24, 2013
(812) 332-0281
Device Information:
Trade Name: Common Name: Classification Name: Cantata® 2.9 Microcatheter Continuous Flush Catheter Catheter, Continuous Flush KRA (21 CFR §870.1210)
Purpose of Submission:
The purpose of this submission is to expand the offering of COOK Inc. Cantata® Microcatheters to include a 2.9 French, 0.027 inch (0.69 millimeters) inner diameter microcatheter. The 2.9 French Cantata® Microcatheter will complement the existing 2.5 French and 2.8 French Cantata Microcatheters.
Indications for Use:
The Cantata 2.9 Microcatheters are intended for use in small vessel or superselective anatomy for diagnostic and interventional procedures including neuro, peripheral, or coronary use.
Predicate Device:
The Cantata 2.9 Microcatheters are identical in terms of intended use, principles of operation, and technological characteristics to the predicate devices. The device, subject of this submission, is substantially equivalent to the Cantata® Microcatheters cleared under 510(k) number K101450.
FEB 20 2014
1
K 131772 - Cantata® 2.9 Microcatheter 510(k) Summary 24 December 2013
Comparison to Predicate Device:
It has been demonstrated that the Cantata® 2.9 Microcatheters are comparable to the predicate device. The predicate devices and the device subject of this submission are intended for use in small vessel or superselective anatomy for diagnostic and interventional procedures including neuro, peripheral, or coronary use. The predicate and proposed devices are completely identical in terms of overall design and indications for use. .
Device Description:
The Cantata 2.9 Microcatheters are braided microcatheters with hydrophilic coating, designed for use in small vessel or superselective anatomy for diagnostic and interventional procedures including neuro, peripheral, or coronary use. The devices include one radiopaque marker band to assist in fluoroscopic visualization of the catheter during use. The catheters are available in 2.9 French shafts and are available in 100, 110, 135, and 150 centimeter lengths. The devices are compatible with a 0.038 inch (0.97 millimeters) diameter angiographic catheter and can accept 0.018 inch (0.46 millimeters) and 0.021 inch (0.53 millimeters) diameter wire guides.
Test Data:
The following tests were performed to demonstrate that the Cantata® 2.9 Microcatheters met applicable design and performance requirements and supports a determination of substantial equivalence.
Test | Methodology | Conclusions |
---|---|---|
Air Leakage Testing | Test article is loaded with de- | |
aerated processed water and | ||
subjected to a negative | ||
pressure after occluding the | ||
lumen. Visual inspection for | ||
air bubbles is subsequently | ||
conducted for fifteen seconds. | Testing showed that the | |
catheter does not exhibit air | ||
leakage during proper clinical | ||
use. Device met the | ||
predetermined acceptance | ||
criteria. | ||
Liquid Leakage Testing | Test article is loaded with | |
processed water and subjected | ||
to a pressure of 300 kPa to | ||
320 kPa after occluding the | ||
lumen. Visual inspection for | ||
air bubbles is subsequently | ||
conducted for a minimum of | ||
30 seconds. | Testing shows that the catheter | |
shall not leak liquid during | ||
proper clinical use. Testing | ||
demonstrated that the device | ||
met the predetermined | ||
acceptance criteria. | ||
Table I Cantata 2.9 Microcatheter Test Data (continued) | ||
Tensile Testing | A uniaxial tensile load was | |
applied to evaluate all four | ||
durometer transition zones of | ||
the test article shaft to ensure | ||
the resulting force required to | ||
break the shaft was greater | ||
than 5 Newtons. | Testing showed that under | |
proper clinical use of the | ||
device the peak load value | ||
shall be greater than 5 | ||
Newtons. In conformance | ||
with ISO 10555-1:1995, the | ||
predetermined acceptance | ||
criteria were met. | ||
Kink Radius Testing | The distal segment of the test | |
article shaft was passed | ||
through a kink radius fixture | ||
to create a loop in the device. | ||
Both end of the device were | ||
then slowly pulled to close the | ||
loop until a kink was formed | ||
in the shaft. This process was | ||
recorded by an image analysis | ||
system which was used to | ||
determine the radius of the | ||
loop immediately prior to | ||
kinking. | Testing showed that under | |
proper clinical use the kink | ||
radius of the device is ≤ 3 | ||
millimeters. Testing | ||
demonstrated that the device | ||
met the predetermined | ||
acceptance criteria. | ||
Dynamic Failure Pressure | ||
Testing | Test articles are subjected to a | |
dynamic flow of saline at | ||
1000 psi +50 psi / 0 psi and | ||
examined for leaking, rupture, | ||
or swelling of the shaft greater | ||
than twice the original | ||
diameter. | Testing demonstrated that | |
under proper clinical use the | ||
device can withstand a | ||
pressure of 1000 pounds per | ||
square inch without signs of | ||
failure. Testing demonstrated | ||
that the device met the | ||
predetermined acceptance | ||
criteria. | ||
Injection Pressures and Flow | ||
Rate Testing | Test article are subjected to | |
dynamic flow or various | ||
injection fluids at given | ||
pressure steps. The fluid | ||
through the lumen of the | ||
device during injection is | ||
measure and divided by the | ||
time that the pressure was | ||
applied to create a flow rate. | Injection pressures and flow | |
rates of the device were | ||
characterized. | ||
Embolic Particle Size Testing | Embolic particles are prepared | |
and delivered through the | ||
device per their corresponding | ||
Instructions for Use. Failure is | ||
considered to have occurred if | ||
the lumen of the device is | ||
completely blocked during | ||
delivery and the entire batch | ||
of particles cannot be | ||
delivered. | The device can deliver | |
embolic particles in the 710 – | ||
1000 micron range. | ||
Ancillary Device | ||
Compatibility Testing | The device was advanced over | |
a standard microwire guide | ||
and through a 0.038" inner | ||
diameter angiographic catheter | ||
placed in a simulated | ||
tortuosity fixture to ensure | ||
device compatibility. | Device is compatible with a | |
Standard micro wire guide and | ||
a 0.038 inch diameter | ||
angiographic catheter. | ||
Evaluation of Device in an | ||
Animal Model | The device was qualitatively | |
evaluated for performance | ||
criterion of preparation, | ||
introduction, pushability, | ||
trackability, and radiopacity in | ||
an animal model. | The device and the compatible | |
device/embolization medium | ||
must receive an acceptable | ||
rating in terms of preparation, | ||
introduction, pushability, | ||
trackability, and radiopacity. | ||
Trackability Testing | The device was inserted | |
through an angiographic | ||
catheter and tracked over a | ||
micro wire guide through a | ||
model of the internal carotid | ||
artery. Testing was considered | ||
successful if the device was | ||
able to be successfully tracked | ||
to the internal carotid artery | ||
siphon. | Testing showed that the device | |
is able to be consistently | ||
tracked to the location of the | ||
internal carotid artery siphon. | ||
Testing demonstrated that the | ||
device met the predetermined | ||
acceptance criteria. | ||
Cytotoxicity – ISO MEM | ||
Elution Assay | When scoring lysis, the test | |
article score a 0 (same as | ||
negative and cell control) at | ||
24, 48, and 72 ± 4 hours. | Non-cytotoxic | |
Table 1 Cantata 2.9 Microcatheter Test Data (continued) | ||
Sensitization | None of the negative animals | |
challenged with the control | ||
vehicle, or the animals | ||
challenged with the test article | ||
extracts were observed with a | ||
sensitization response great | ||
then '0'. The normal saline | ||
extract and cottonseed oil | ||
extract of the test material had | ||
a sensitization response of '0' | ||
under valid test conditions. | The test article did not elicit a | |
sensitization response | ||
Intracutaneous Reactivity | The differences in the mean | |
test and control scores of the | ||
extract dermal observations | ||
were less and 1.0 | The requirements of the ISO | |
Intracutaneous Reactivity Test | ||
have been met | ||
Systemic Toxicity | None of the test article treated | |
animals were observed with | ||
clinical signs consistent with | ||
toxicity at any of the | ||
observations periods. | The requirements of the ISO | |
Acute System Injection Test | ||
have been met | ||
Hemolysis | Test article exhibited an | |
Average Blank Corrected % | ||
Hemolytic Index of % (same | ||
as blank control) | Non-hemolytic | |
Complement Activation | Under conditions of the C3a | |
assay the test article and the | ||
comparison article activation | ||
at 2.1% and 3.5%, | ||
respectively, of the normalized | ||
C3a concentration produced | ||
by CVF | No ranges or levels | |
established as acceptable | ||
Partial Thromboplastin Time | The test article had an average | |
clotting time 297.0 seconds | ||
(99% of the negative control) | Minimal-activator of intrinsic | |
coagulation pathway |
Table 1 Cantata® 2.9 Microcatheter Test Data
2
Table 1 Cantata® 2.9 Microcatheter Test Data (continued)
3
Table 1 Cantata® 2.9 Microcatheter Test Data (continued)
4
Table 1 Cantata® 2.9 Microcatheter Test Data (continued)
In conclusion, the results of these tests provide reasonable assurance that the device is as safe and as effective as the predicate devices, and support a determination of substantial equivalence.
5
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a simple, graphic style.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002
February 20, 2014
Cook Inc. Ms. Jennifer A. Richardson Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47402
Re: K131772
Trade/Device Name: Cantata® 2.9 Microcatheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous flush catheter Regulatory Class: Class II Product Code: KRA, DQY Dated: June 14, 2013 Received: December 24, 2013
Dear Ms. Richardson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
6
Page 2 - Ms. Jennifer A. Richardson
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joyce M. Whang -S
for Carl.
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K131772
Device Name Cantata® 2.9 Microcatheters
Indications for Use (Describe)
The Cantata® 2.9 Microcatheters are in small vessel or superselective anatomy for diagnostic and interventional procedures including neuro, peripheral, or coronary use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED,
FOR FDA USE ONLY .. . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/11 description: The image shows the name "Joyce MFW hang -S" in a bold, sans-serif font. The letters are black and stand out against a white background. The letters "MFW" are stylized with a pattern that resembles a maze or a circuit board.
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
8
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."