K Number
K180340

Validate with FDA (Live)

Date Cleared
2018-09-26

(231 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of Glucose, Creatinine, and Blood Urea Nitrogen, in heparinized arterial and venous whole blood.

Glucose (Glu) Measurements are used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell tumor.
Creatinine (Creat) Measurements are used in the diagnosis and treatment of certain renal conditions and for monitoring adequacy of dialysis.
Blood Urea Nitrogen (BUN) Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

Device Description

The Stat Profile Prime Plus Analyzer System is designed to be a low cost, low maintenance analyzer for the hospital laboratory setting. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

The system architecture and user interface for this proposed device is based on the previously cleared Stat Profile Prime CCS Analyzer System (K131703). The primary predicate for this proposed device is the Stat Profile pHOx Ultra Analyzer System (K110648).

The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed. The reporting of CO-Oximeter parameters (or not reporting them) will also be determined by the selection of the Sensor Cards:

Primary Sensor Card Port:
There are two options for the primary sensor card:
. Primary Sensor Card 1 shall enable and report the following listed analytes:
o PO2, PCO2, pH, Hct, tHb, Na, Cl, K, iCa, iMg, Glu, SO2, O2Hb, COHb, MetHb, HHb
. Primary Sensor Card 2 shall enable and report the following listed analytes:
o PO2, PCO2, pH, Hct, tHb, Na, Cl, K, iCa, iMg, Glu, SO2

Auxiliary Sensor Card Port:
The reporting of Creatinine and BUN parameters (Or not reporting them) shall be determined by the selection of the Auxiliary Sensor Card
Auxiliary Sensor Card 1 enables Creatinine and BUN parameters
. Auxiliary Sensor Card 2 is a "dummy" sensor card, and will not report any parameters.

Similar to the primary predicate device, the Stat Profile Prime Plus Analyzer is a blood gas/cooximetry/electrolyte/chemistry and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control will be used, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly at all times.

The Stat Profile Prime Plus Analyzer accepts samples from syringes, open tubes, and small cups. The minimum sample size for analysis is 135 µL.

Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

Stat Profile Prime Plus Analyzer System Components:
The Stat Profile Prime Plus Analyzer System is comprised of the following components.
. Stat Profile Prime Plus Analyzer System
Primary Sensor Cartridge
. Auxiliary Sensor Cartridge
Stat Profile Prime Plus Auto-Cartridge Quality Control Pack
. Stat Profile Prime Plus Calibrator Cartridge
Stat Profile Prime Plus External Ampuled Control
IFU/Labeling

AI/ML Overview

The provided text describes the acceptance criteria and study proving the device meets those criteria. The device in question is the Stat Profile Prime Plus Analyzer System, and the studies focus on its performance for measuring Glucose, Creatinine, and Blood Urea Nitrogen (BUN) in whole blood.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria & Reported Device Performance

The document states that the "blood comparison data for Glucose, Creatinine, and BUN, for the Stat Profile Prime Plus analyzers meet the acceptance criteria." While explicit numerical acceptance criteria are not presented in a table form, the performance is described relative to a predicate device (Stat Profile pHOx Ultra Analyzer System) and internal specifications. The linearity and precision studies also explicitly state that the acceptance criteria were met.

For a clearer representation, we can infer the acceptance criteria and performance based on the descriptions:

Performance MetricAcceptance Criteria (Inferred from text)Reported Device Performance
Method ComparisonDemonstrated equivalence to the predicate device (Stat Profile pHOx Ultra Analyzer) for Glucose, Creatinine, and BUN measurements in heparinized whole blood."The blood comparison data for Glucose, Creatinine, and BUN, for the Stat Profile Prime Plus analyzers meet the acceptance criteria." (Implies equivalence was successfully demonstrated).
Precision (Within Run)Meet specified imprecision limits for within-run performance."The precision data for all parameters meet the within run imprecision specifications for the Stat Profile Prime Plus analyzers."
Precision (Run to Run)Meet specified imprecision limits for between-analyzer run-to-run performance."The precision data for all parameters meet the between analyzer run to run imprecision specifications for the Stat Profile Prime Plus analyzers."
LinearityGood correlation and linearity to reference analyzers across the claimed measurement range, meeting acceptance criteria."The linearity comparison data for all parameters for the Stat Profile Prime Plus analyzers shows good correlation and linearity to the reference analyzers across the claimed measurement range for all parameters and met the acceptance criteria."
Specificity/InterferenceIdentify significant interferents and determine concentrations at which interference occurs, with implied acceptable levels of bias for non-interfering substances.Specific interfering substances (Hydroxyurea, Oxalate, Thiocyanate) were identified for Glucose and Creatinine at certain concentrations, showing bias. Conversely, "No interference observed" below certain concentrations. (This implies that at clinically relevant concentrations, significant interference is either absent or managed). For example, Hydroxyurea for Glucose: no interference at 0.2 mg/dL, bias of 19.2% at 0.4 mg/dL.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:

    • Precision/Reproducibility - Within Run: 20 replicates for one run of each of the following: internal controls (Levels 1-5), ampuled controls (Levels 1-5), and two whole blood samples (from syringes).
    • Precision/Reproducibility - Run to Run: Triplicate analyses performed on a single whole blood sample in ten separate runs.
    • Linearity Testing: Not explicitly stated, but performed using "various medical decision limits" and comparing to reference analyzers.
    • Specificity/Interference Testing: Conducted using whole blood collected in lithium heparin vacutainers. Possible interfering substances tested at two analyte concentrations for each. The number of unique samples or overall sample size for this study is not explicitly stated beyond "many substances were screened".
    • Method Comparison Studies: Performed to compare the Stat Profile Prime Plus to the Nova Stat Profile pHOx Ultra analyzer in a clinical laboratory setting. The number of samples is not specified.
  • Data Provenance: The studies were performed in a "clinical laboratory setting" and involved "whole blood" samples. The document does not specify the country of origin for the data or explicitly state whether it was retrospective or prospective. Given the nature of performance testing for a medical device, it is typically prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not applicable (N/A) to this device and study type. This device is an in-vitro diagnostic (IVD) analyzer that quantitatively measures analytes (Glucose, Creatinine, BUN). The "ground truth" for its performance studies is established by:

  • Reference Methods/Analyzers: For method comparison and linearity studies.
  • Known Concentrations: For controls and linearity samples.
  • Established Analytical Specifications: For precision and interference studies.

There are no human experts classifying images or clinical conditions, so there is no panel of experts to establish a "ground truth" in the way described for AI/CADe systems.

4. Adjudication Method for the Test Set

This information is N/A. As explained above, the ground truth is based on quantitative measurements against reference methods or known concentrations, not on expert consensus requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This information is N/A. This device is an analytical instrument for quantitative measurements, not an AI/CADe system designed to assist human readers (e.g., radiologists) in interpreting medical images. Therefore, an MRMC study is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This information is N/A. This device is a standalone analyzer. Its "performance" is its standalone capability to accurately and precisely measure analytes. There is no human-in-the-loop component in its direct operation for measurement that would require a separate "human-in-the-loop performance" study. The studies described (method comparison, precision, linearity, interference) directly evaluate the standalone performance of the instrument.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The ground truth for this device's performance studies is based on:

  • Reference Measurement Methods/Instruments: For comparison studies (e.g., comparing the Prime Plus Analyzer's results to those from the predicate device or other established reference analyzers).
  • Known and Certified Control Materials: For quality control and precision studies, where the true concentration of analytes in the control samples is known.
  • Spiked Samples: For linearity and interference studies, where samples are spiked with known concentrations of analytes or interfering substances.

8. The Sample Size for the Training Set

This information is N/A. This document describes the validation of a hardware-based analytical instrument with established biochemical measurement principles, not an AI/machine learning algorithm that requires a "training set" in the computational sense. The device's measurement algorithms are based on scientific principles (e.g., enzymatic reactions, electrochemistry, Nernst equation) and are not "trained" on data in the way an AI model would be.

9. How the Ground Truth for the Training Set Was Established

This information is N/A for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 26, 2018

Nova Biomedical Corporation Cesidio Tempesta Senior Regulatory Affairs Specialist 200 Prospect St. Waltham, MA 02454

Re: K180340

Trade/Device Name: Stat Profile Prime Plus Analyzer System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: CGA, CDS, CGL Dated: August 24, 2018 Received: August 27, 2018

Dear Cesidio Tempesta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number K180340

Device Name

Stat Profile® Prime Plus Analyzer System

Indications for Use (Describe)

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of Glucose, Creatinine, and Blood Urea Nitrogen, in heparinized arterial and venous whole blood.

Glucose (Glu)Measurements are used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell tumor.
Creatinine (Creat)Measurements are used in the diagnosis and treatment of certain renal conditions and for monitoring adequacy of dialysis.
Blood Urea Nitrogen (BUN)Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.
Type of Use (Select one or both, as applicable)
-------------------------------------------------
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K180340

510(K) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Cesidio Tempesta, Regulatory Affairs Specialist
Date Prepared:September 25, 2018

Proprietary Name: Stat Profile Prime Plus Analyzer System

Common or Usual Name: Blood Gas Analyzer

Classification Name: Multiple

Classification NameRegulation #ClassProduct Code
Glucose test system862.1345IICGA
Creatinine test system862.1225IICGL
Urea Nitrogen Test System862.1770IICDS

Product Code: CGA

Predicate Device:

K110648 - Stat Profile pHOx Ultra Analyzer System

Device Description:

The Stat Profile Prime Plus Analyzer System is designed to be a low cost, low maintenance analyzer for the hospital laboratory setting. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

The system architecture and user interface for this proposed device is based on the previously cleared Stat Profile Prime CCS Analyzer System (K131703). The primary predicate for this proposed device is the Stat Profile pHOx Ultra Analyzer System (K110648).

The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed. The reporting of CO-Oximeter parameters (or not reporting them) will also be determined by the selection of the Sensor Cards:

Primary Sensor Card Port:

There are two options for the primary sensor card:

  • . Primary Sensor Card 1 shall enable and report the following listed analytes:
    • o PO2, PCO2, pH, Hct, tHb, Na, Cl, K, iCa, iMg, Glu, SO2, O2Hb, COHb, MetHb, HHb
  • . Primary Sensor Card 2 shall enable and report the following listed analytes:
    • o PO2, PCO2, pH, Hct, tHb, Na, Cl, K, iCa, iMg, Glu, SO2

{4}------------------------------------------------

Auxiliary Sensor Card Port:

The reporting of Creatinine and BUN parameters (Or not reporting them) shall be determined by the selection of the Auxiliary Sensor Card

  • Auxiliary Sensor Card 1 enables Creatinine and BUN parameters ●
  • . Auxiliary Sensor Card 2 is a "dummy" sensor card, and will not report any parameters.

Similar to the primary predicate device, the Stat Profile Prime Plus Analyzer is a blood gas/cooximetry/electrolyte/chemistry and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control will be used, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly at all times.

The Stat Profile Prime Plus Analyzer accepts samples from syringes, open tubes, and small cups. The minimum sample size for analysis is 135 µL.

Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

Stat Profile Prime Plus Analyzer System Components:

The Stat Profile Prime Plus Analyzer System is comprised of the following components.

  • . Stat Profile Prime Plus Analyzer System
  • Primary Sensor Cartridge ●
  • . Auxiliary Sensor Cartridge
  • Stat Profile Prime Plus Auto-Cartridge Quality Control Pack ●
  • . Stat Profile Prime Plus Calibrator Cartridge
  • Stat Profile Prime Plus External Ampuled Control ●
  • IFU/Labeling ●

Sample Types:

The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial and venous whole blood.

Measured Parameters:

The Stat Profile Prime Plus Analyzer measures:

  • . Glucose (Glu)
  • . Creatinine (Creat)
  • . Blood Urea Nitrogen (BUN)

Calculated Parameters:

The following parameters are calculated by the Prime Plus Analyzer based on results of the directly measured parameters.

  • BUN/Creat Ratio (BUN/Creat) ●
  • Blood Osmolality (OSM) ●
  • Estimated Glomerular Filtration Rate (eGFR) ●

Intended Use:

The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of Glucose, Creatinine, and Blood Urea Nitrogen, in heparinized arterial and venous whole blood.

Glucose (Glu)Measurements are used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell tumor.
Creatinine (Creat)Measurements are used in the diagnosis and treatment of certain renal conditions and for monitoring adequacy of dialysis.
Blood Urea Nitrogen (BUN)Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

{5}------------------------------------------------

Summary of the Technological Characteristics:

The Stat Profile Prime Plus Analyzer is substantially equivalent to the previously cleared for market Stat Profile pHOx Ultra Analyzer System (K110648) in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters. The External Control solutions for use with the Stat Profile Prime Plus Analyzer are substantially equivalent to those cleared for use with the predicate Stat Profile pHOx Ultra Analyzer System (K110648). See Table 1 for comparison of the predicate and proposed devices.

Principles of Measurement:

Glucose:

Glucose measurement is based on the level of H₂O₂ produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. At a constant potential of 0.70 volts, electroactive H3O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the platinum electrode is proportional to the glucose concentration of the sample.

Creatinine:

The Prime Plus Creatinine sensor uses 3 enzymes. These 3 enzymes catalyze the conversion of Creatinine, ultimately forming formaldehyde, glycine, and hydrogen peroxide.

At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the Creatinine concentration of the sample.

BUN:

The Prime Plus Analyzer uses urease, which has been chemically bonded to a membrane, to catalyze the conversion of urea present in the sample to ammonia and CO).

At the pH of the sample, ammonia converts predominantly to the ammonium ion. An ammonium ion selective electrode is used to detect the ammonium formed by the above reactions. This measurement is then related to the concentration of urea present in the original sample via the Nernst equation.

Summary of Performance Testing:

Performance testing was completed to demonstrate that the Stat Profile Prime Plus Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System. The performance testing included:

Method Comparison Studies:

A study was performed to compare the Stat Profile Prime Plus to the Nova Stat Profile pHOx Ultra analyzer to assess the equivalence of the analyzers in the measurement of Glucose, Creatinine, and BUN in heparinized whole blood in a clinical laboratory setting.

The blood comparison data for Glucose, Creatinine, and BUN, for the Stat Profile Prime Plus analyzers meet the acceptance criteria.

Precision/Reproducibility - Within Run and Run to Run Studies:

Within Run and Run to Run precision was evaluated by replication studies performed on three Stat Profile Prime Plus analyzers.

Within Run Precision testing consisted of one run of each of the following sample types and levels was performed, 20 replicates per run:

  • . Stat Profile Prime Plus Internal Controls: Levels 1-5
  • Stat Profile Prime Plus Ampuled Controls: Levels 1-5
  • Two whole bloods, sampled from syringes

To assess Run to Run Precision for whole blood, triplicate analyses were performed on a single whole blood sample in ten separate runs during a single day. The systems were recalibrated before each triplicate run.

{6}------------------------------------------------

The precision data for all parameters meet the within run imprecision specifications for the Stat Profile Prime Plus analyzers.

The precision data for all parameters meet the between analyzer run to run imprecision specifications for the Stat Profile Prime Plus analyzers.

Linearity Testing:

The study assessed the linearity of all parameters to establish and/or verify the Analytical Measurement Range (AMR) for the Stat Profile Prime Plus Analyzer on whole blood. The linearity was measured using method comparison. The evaluation of the linear range included lower and upper limits of the AMR and various medical decision limits.

All Stat Profile Prime Plus results were compared to the reference analyzer and/or the product specifications.

The linearity comparison data for all parameters for the Stat Profile Prime Plus analyzers shows good correlation and linearity to the reference analyzers across the claimed measurement range for all parameters and met the acceptance criteria.

Specificity / Interference Testing:

The purpose of this study was to identify substances that may interfere with the Stat Profile Prime Plus sensors. If interference was identified, a dose response study was performed to determine the concentration where the interfering substance may alter results.

This Interference Testing study was performed using whole blood collected in lithium heparin vacutainers. The possible interfering substances were tested at two analyte concentrations.

Many substances were screened as potential interferents through analysis on the Prime Plus analyzers. Substance screening was completed according to the CLSI EP7-A2 quideline. The interfering substances identified during the screening process are listed in the table below:

ParameterInterfering SubstanceConcentration of interferingsubstanceInterference
GlucoseHydroxyurea0.2 mg/dLNo interference observed
0.4 mg/dLBias of 19.2%
Oxalate125 mg/dLNo interference observed
250 mg/dLBias of -13.9%
Thiocyanate3.4 mmol/LNo interference observed
5.1 mmol/LBias of 13.5%
CreatinineHydroxyurea0.06 mg/dLNo interference observed
0.08 mg/dLBias of 15.5%
Thiocyanate1.7 mmol/LNo interference observed
3.4 mmol/LBias of 42.9%

Conclusion:

The results of software validation and performance verification testing confirmed that the Stat Profile Prime Plus Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime Plus Analyzer System is substantially equivalent to that of the predicate Stat Profile pHOx Ultra Analyzer System (K110648).

{7}------------------------------------------------

Table 1: Comparison of Predicate and Proposed devices

CharacteristicPredicate: K110648 - Stat Profile pHOx Ultra AnalyzerProposed: Stat Profile Prime Plus Analyzer
Indication For UseThe Stat Profile pHOx Ultra Analyzer without CO-Oximeter is intended for in vitrodiagnostic use by health care professionals and/or point-of-care usage in thequantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin(Hb) in heparinized whole blood; Na+, K'-, Cl-, Ca++, Mg++, Glucose (Glu), Lactate(Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, orplasma.The Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitroThe Stat Profile Prime Plus Analyzer System is indicated for use byhealthcare professionals in clinical laboratory settings for quantitativedetermination of Glucose, Creatinine, and Blood Urea Nitrogen, inheparinized arterial and venous whole blood.
diagnostic use by health care professionals and for point-of-care usage in thequantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), totalHemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb),Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) inheparinized whole blood; Na+, K'-, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac),BUN (Urea), and Creatinine (Creat) in Heparinized whole blood, serum, or plasma.Total Bilirubin (tBil) was not evaluated on neonatal samples.
Acceptable Samples
Sample TypesSodium or lithium heparinized whole blood, serum, or plasma samples from syringes,open tubes, small cups, and capillary tubes.Lithium heparin whole blood from syringes, open tubes, and small cups.
Sample Volumes60-200µL (dependent on panel selected)135µL
Measurement Range
Glu15 - 500 mg/dL15 - 500 mg/dL
Creat0.2 - 20.0 mg/dL0.2 - 12.0 mg/dL
BUN3 - 100 mg/dL3 - 100 mg/dL
Principles ofMeasurement
GluEnzymatic sensorSame
CreatImpedance sensorSame
BUNEnzymatic sensorSame
Touch Screen12.1" LCD, 1024x768 pixel, Resistive Touch10.1" WXGA 1280 x 800 color touch screen
MenuFully configurable test menu based on available sensorsSame
Bar Code ScannerInternal Integrated 1D/2DSame
Printer2" Roll, Thermal TransferSame
PumpPeristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT)Same
Analog BoardPrecision low level analog front end w/ amperometric and potentiometric amplifiers,air detector circuitry and temperature control circuitrySame

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.