K Number
K131703
Date Cleared
2014-03-25

(287 days)

Product Code
Regulation Number
862.1120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Glucose), in heparinized whole blood.

PCO2, PO2, pH: Whole blood measurement of certain gases in whole blood, or pH of whole blood, is used in the diagnosis and treatment of life-threatening acid-base disturbances.

Hct: Whole blood measurements of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).

Na +: Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.

K+: Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

Cl-: Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

ICa: Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Glu: Glucose measurement is used in the diagnosis and treatment of carbohydrate metabollsm disturbances Including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The Stat Profile Prime Auto QC Cartridge CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Calibrator Cartridge CCS is intended for the callbration of pH, PCO2, PO2, Hct, Na+, K+, C-, ICa, and Glucose using the Stat Profile Prime CCS Analyzer.

Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from callbrator cartridge or analytical instrument variation.

Device Description

The Stat Profile Prime CCS Analyzer is a small, low cost blood gas, metabolite and electrolyte analyzer for laboratory use. The sensors and flow path have been integrated into one replaceable microsensor card, which is replaced periodically according to usage. The product, consumables, installation instructions and packaging are designed for easy customer installation.

Whole blood specimens are aspirated into the analyzer's microsensor card from syringes, tubes, or capillary blood collection devices using a peristaltic pump and a sampling probe. The disposable microsensor card contains the analytical flowpath and all of the measurement sensors (pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Glucose). Once the analysis measurement is complete, the whole blood specimen is automatically flushed out of the microsensor card flowpath and into a self contained waste collection bag contained within the disposable calibrator carridge.

The Stat Profile Prime CCS Analyzer will have an enhanced test menu and multiple quality control options. Both traditional Internal and External liquid QC shall be offered, as well as an onboard Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly.

As with the predicate, the Stat Profile Prime CCS Analyzer is microprocessor-based and incorporates:

  • . traditional sensor technology to measure blood pO2
  • ion selective electrode technology to measure pH, pCO₂, blood sodium, potassium, . chloride, and ionized calcium
  • enzyme/Amperometric technology for glucose measurements .

Liquid quality control materials are available as internal auto-cartridge quality control packs and as external ampules. The sampling, calibration and quality control functions are fully automated.

Internal Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood gas electrodes using external compressed gas cylinders. The Calibration Cartridges contain aqueous solutions within individual flexible bags housed in a cardboard box and a flexible waste bag. Each bag includes a fitment with a septa that is pierced during the insertion of the cartridge into the analyzer. The Calibration Cartridge aqueous solutions allow for 2 point calibration of each parameter as follows.

  • Calibrator A pH, PCO2, Na, K, Cl, iCa, and Glu (Volume > 500 mL) .
  • Calibrator B pH, PO2, Na, K, Cl, iCa, and Glu (Volume > 250 mL) .
  • . Calibrator F - PCO2, PO2 (Volume > 720 mL)
  • Reference Solution KCI (Volume > 300 mL) .

The external glass ampule controls contain a buffered bicarbonate solution with a known pH and known levels of Na, K, CI, iCa, and Glu. The solutions are equilibrated with known levels of O2, CO2, and N2. Each ampule contains 1.7 ml volume.

The internal auto QC cartridge consists of 3 flexible bags within a cardboard carton. Each bag contains an aqueous quality control material for monitoring the measurement of pH, PCO2, PO2, hematocrit (Hct) Na, K, Cl, iCa, and Glucose (Glu). The aqueous quality control materials are composed of a buffered bicarbonate solution, each with a known pH and known level of Na, K, Cl, iCa, and Glu. Solutions are equilibrated with known levels of O2, CO2, and N2. Each bag contains a minimum volume of 100 mL. The aqueous quality control materials are formulated at three levels:

  • . Control 1: Acidosis, with High Electrolyte, Low Normal Glu
  • . Control 2: Normal pH, Low-Normal Hct, Normal Electrolyte, High Glu
  • Control 3: Alkalosis, High Hct, Low Electrolyte, High Abnormal Glu, .

Linearity Standard Set A consists of ampuled buffered solutions containing Ca++, Glu, K+, and Cl-. Each ampule contains 1.8 ml volume.

The Stat Profile Analyzer accepts Lithium heparin whole blood sample from syringes, open tubes, small cups, and capillary tubes. The minimum sample size for both syringe and capillary samples analysis is 100 µL.

Measured Parameters:

The Stat Profile Prime CCS Analyzer measures pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Note: Glucose is optional).

Calculated Parameters:

  • . pH, PCO2, PO2 (corrected to patient temperature)
  • . Bicarbonate level (HCO3-)
  • Total Carbon Dioxide (TCO2) .
  • Base Excess of the blood (BE-b) .
  • Base Excess of extracellular fluid (BE-ecf) .
  • Standard Bicarbonate Concentration (SBC) .
  • . Oxygen Content (O2Ct)
  • . Oxygen Capacity (O2Cap)
  • Alveolar Oxygen (A) .
  • Arterial Alveolar Oxygen Tension Gradient (AaDO2) .
  • Arterial Alveolar Oxygen Tension Ratio (a/A) .
  • Respiratory Index (RI) .
  • P50 .
  • PO2/FIO2 ratio .
  • Oxygen Saturation (SO2%) .
  • Hemoglobin .
  • Anion Gap .
  • Normalized Calcium, nCa ●
AI/ML Overview

The Nova Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use for the quantitative determination of blood gases (pH, PCO2, PO2), hematocrit (Hct), and electrolytes/metabolites (Na+, K+, Cl-, iCa, and Glucose) in heparinized whole blood. The acceptance criteria for the device's performance are established through bench testing, demonstrating substantial equivalence to its predicate device, the Nova Stat Profile pHOx Ultra Analyzer System (K110648).

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly defines acceptance criteria by comparing the proposed device to the predicate device and asserting substantial equivalence. While explicit numerical acceptance ranges are not provided for all parameters in a pass/fail format, the comparison table and the conclusion of the study indicate that the Stat Profile Prime CCS Analyzer met the performance levels demonstrated by the predicate.

CharacteristicPredicate: K110648 Stat Profile pHOx Ultra AnalyzerProposed: Stat Profile Prime CCS Analyzer (Reported Performance)
Measurement Range
pH6.500-8.000Same (6.500-8.000)
PCO23.0-200 mmHgSame (3.0-200 mmHg)
PO20-800 mmHg5-765 mmHg
Hct12%-70%Same (12%-70%)
Na+80-200 mmol/LSame (80-200 mmol/L)
K+1.0-20.0 mmol/LSame (1.0-20.0 mmol/L)
Cl-50-200 mmol/LSame (50-200 mmol/L)
Ca (Ca++) (iCa)0.10-2.70 mmol/L0.20-2.70 mmol/L
Glu15-500 mg/dLSame (15-500 mg/dL)
Other Performance
Method ComparisonPerformance demonstrated by predicateSubstantial equivalence demonstrated
Precision/ReproducibilityPerformance demonstrated by predicateSubstantial equivalence demonstrated
Run to Run PrecisionPerformance demonstrated by predicateSubstantial equivalence demonstrated
Within Run PrecisionPerformance demonstrated by predicateSubstantial equivalence demonstrated
Linearity TestingPerformance demonstrated by predicateSubstantial equivalence demonstrated
Specificity / Interference TestingPerformance demonstrated by predicateSubstantial equivalence demonstrated
Detection LimitPerformance demonstrated by predicateSubstantial equivalence demonstrated

2. Sample Sizes used for the Test Set and Data Provenance:

The document states that "Bench testing was completed to demonstrate that the Stat Profile Prime CCS Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System." However, it does not provide specific sample sizes for the test sets used in the method comparison, precision, linearity, specificity/interference, or detection limit studies.

The data provenance is not explicitly stated. Given that the testing was bench testing for regulatory submission, it is highly likely to be prospective data collected specifically for this submission, rather than retrospective. The country of origin for the data is also not specified, but the applicant, Nova Biomedical Corporation, is based in Waltham, MA, USA, suggesting the testing likely occurred in the USA.

3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:

This device is an in-vitro diagnostic (IVD) analyzer for quantitative measurements. The ground truth for such devices is typically established through reference methods or certified reference materials, not by expert consensus in the same way an image analysis AI might be. Therefore, the concept of "experts used to establish ground truth" with specific qualifications like "radiologist with 10 years of experience" does not apply in this context. The ground truth would be determined by the accuracy of the reference methods or the certified values of the reference materials used in the bench testing.

4. Adjudication Method for the test set:

Adjudication methods (e.g., 2+1, 3+1) are typically used in studies where human readers are making subjective interpretations, such as in medical imaging. For an IVD analyzer performing quantitative measurements, the ground truth is objective (e.g., a measured concentration or value). Therefore, the concept of an adjudication method does not apply to the bench testing conducted for this device. Discrepancies would be resolved through re-testing, calibration, or investigation of the reference method's accuracy.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

A multi-reader multi-case (MRMC) comparative effectiveness study is used to assess the impact of an AI algorithm on human readers' performance, typically in diagnostic imaging. This device is an automated IVD analyzer, not an AI-assisted diagnostic tool that aids human interpretation. Therefore, an MRMC study was not performed, and the concept of human readers improving with/without AI assistance does not apply.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The performance evaluation described ("Method Comparison Studies," "Precision/Reproducibility Studies," "Linearity Testing," "Specificity / Interference Testing," "Detection Limit") primarily assesses the standalone performance of the Stat Profile Prime CCS Analyzer. This type of testing evaluates the algorithm and hardware's ability to accurately and precisely measure the specified analytes without direct human intervention in the measurement process (beyond sample loading and initiation). The human role is in setting up the device and interpreting the final quantitative result.

7. The type of ground truth used:

The ground truth for the performance testing (method comparison, precision, linearity, etc.) would be established using:

  • Reference methods: Comparing the device's measurements to established, highly accurate laboratory methods for each analyte.
  • Certified reference materials: Using materials with known, precisely determined concentrations of the analytes.

The document does not explicitly state which specific reference methods or certified materials were used, but these are standard practices for IVD device validation.

8. The sample size for the training set:

This document describes a submission for an in-vitro diagnostic analyzer based on established sensor technologies and measurement algorithms, not a machine learning or AI model that requires a distinct "training set." Therefore, the concept of a "training set sample size" as understood in AI/ML development does not apply in this context. The underlying algorithms for measurement are based on known chemical and physical principles and extensive historical validation, not iterative training on a large dataset.

9. How the ground truth for the training set was established:

As mentioned above, this device does not utilize a training set in the AI/ML sense. The "ground truth" for the device's core measurement principles is based on validated scientific principles, sensor calibration procedures, and the established performance characteristics of the predicate device. Changes or improvements from the predicate would be validated against established reference methods.

{0}------------------------------------------------

510(k) Summary

510(K) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Paul W. MacDonald
Date Prepared:21 March 2014

Proprietary Name: Stat Profile® Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control ABG/CCS, Stat Profile Prime Calibrator Cartridge CCS/CCS Comp and Linearity Standard Set A

Common or Usual Name: Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer

Classification Name: Multiple

Classification Names:Class No.Reg. No.Class
Blood Gases and Blood pH system75CHL862.1120II
Sodium Test System75JGS862.1665II
Potassium Test System75CEM862.1600II
Calcium Test System75JFP862.1145II
Chloride Test System75CGZ862.1170II
Glucose Test System75CGA862.1345II
Instrument, Hematocrit, Automated81GKF864.5600II
Calibrators75JIX862.1150II
Quality Control Materials75JJS862.1660II

Product Codes: CHL, JGS, CEM, JFP, CGZ, CGA, GKF, JIX, JJS

Predicate Device: K110648 - Nova Stat Profile pHOx Ultra Analyzer System (including controls, calibrators and Linearity Standards)

Device Description:

The Stat Profile Prime CCS Analyzer is a small, low cost blood gas, metabolite and electrolyte analyzer for laboratory use. The sensors and flow path have been integrated into one replaceable microsensor card, which is replaced periodically according to usage. The product, consumables, installation instructions and packaging are designed for easy customer installation.

Whole blood specimens are aspirated into the analyzer's microsensor card from syringes, tubes, or capillary blood collection devices using a peristaltic pump and a sampling probe. The disposable microsensor card contains the analytical flowpath and all of the measurement sensors (pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Glucose). Once the analysis measurement is complete, the whole blood specimen is automatically flushed out of the microsensor card flowpath and into a self contained waste collection bag contained within the disposable calibrator carridge.

The Stat Profile Prime CCS Analyzer will have an enhanced test menu and multiple quality control options. Both traditional Internal and External liquid QC shall be offered, as well as an onboard Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly.

{1}------------------------------------------------

As with the predicate, the Stat Profile Prime CCS Analyzer is microprocessor-based and incorporates:

  • . traditional sensor technology to measure blood pO2
  • ion selective electrode technology to measure pH, pCO₂, blood sodium, potassium, . chloride, and ionized calcium
  • enzyme/Amperometric technology for glucose measurements .

Liquid quality control materials are available as internal auto-cartridge quality control packs and as external ampules. The sampling, calibration and quality control functions are fully automated.

Internal Calibration standards with dissolved gases are provided in sealed pouches eliminating the need for users to calibrate the blood gas electrodes using external compressed gas cylinders. The Calibration Cartridges contain aqueous solutions within individual flexible bags housed in a cardboard box and a flexible waste bag. Each bag includes a fitment with a septa that is pierced during the insertion of the cartridge into the analyzer. The Calibration Cartridge aqueous solutions allow for 2 point calibration of each parameter as follows.

  • Calibrator A pH, PCO2, Na, K, Cl, iCa, and Glu (Volume > 500 mL) .
  • Calibrator B pH, PO2, Na, K, Cl, iCa, and Glu (Volume > 250 mL) .
  • . Calibrator F - PCO2, PO2 (Volume > 720 mL)
  • Reference Solution KCI (Volume > 300 mL) .

The external glass ampule controls contain a buffered bicarbonate solution with a known pH and known levels of Na, K, CI, iCa, and Glu. The solutions are equilibrated with known levels of O2, CO2, and N2. Each ampule contains 1.7 ml volume.

The internal auto QC cartridge consists of 3 flexible bags within a cardboard carton. Each bag contains an aqueous quality control material for monitoring the measurement of pH, PCO2, PO2, hematocrit (Hct) Na, K, Cl, iCa, and Glucose (Glu). The aqueous quality control materials are composed of a buffered bicarbonate solution, each with a known pH and known level of Na, K, Cl, iCa, and Glu. Solutions are equilibrated with known levels of O2, CO2, and N2. Each bag contains a minimum volume of 100 mL. The aqueous quality control materials are formulated at three levels:

  • . Control 1: Acidosis, with High Electrolyte, Low Normal Glu
  • . Control 2: Normal pH, Low-Normal Hct, Normal Electrolyte, High Glu
  • Control 3: Alkalosis, High Hct, Low Electrolyte, High Abnormal Glu, .

Linearity Standard Set A consists of ampuled buffered solutions containing Ca++, Glu, K+, and Cl-. Each ampule contains 1.8 ml volume.

The Stat Profile Analyzer accepts Lithium heparin whole blood sample from syringes, open tubes, small cups, and capillary tubes. The minimum sample size for both syringe and capillary samples analysis is 100 µL.

Measured Parameters:

The Stat Profile Prime CCS Analyzer measures pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Note: Glucose is optional).

{2}------------------------------------------------

Calculated Parameters:

  • . pH, PCO2, PO2 (corrected to patient temperature)
  • . Bicarbonate level (HCO3-)
  • Total Carbon Dioxide (TCO2) .
  • Base Excess of the blood (BE-b) .
  • Base Excess of extracellular fluid (BE-ecf) .
  • Standard Bicarbonate Concentration (SBC) .
  • . Oxygen Content (O2Ct)
  • . Oxygen Capacity (O2Cap)
  • Alveolar Oxygen (A) .
  • Arterial Alveolar Oxygen Tension Gradient (AaDO2) .
  • Arterial Alveolar Oxygen Tension Ratio (a/A) .
  • Respiratory Index (RI) .
  • P50 .
  • PO2/FIO2 ratio .
  • Oxygen Saturation (SO2%) .
  • Hemoglobin .
  • Anion Gap .
  • Normalized Calcium, nCa ●

Intended Use:

The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Glucose), in heparinized whole blood.

PCO2,PO2,pHWhole blood measurement of certain gases in whole blood, or pH of whole blood, is used in thediagnosis and treatment of life-threatening acid-base disturbances.
HctWhole blood measurements of the packed red cell volume of a blood sample are used todistinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in thenumber of red cells).
Na+Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetesinsipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involvingelectrolyte imbalance.
K+Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment ofdisease conditions characterized by low or high potassium levels.
Cl-Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolicdisorders such as cystic fibrosis and diabetic acidosis.
iCaCalcium measurements are used in the diagnosis and treatment of parathyroid disease, avariety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions orspasms).
GluGlucose measurement is used in the diagnosis and treatment of carbohydrate metabolismdisturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia,and of pancreatic islet cell carcinoma.

Stat Profile Prime Auto QC Cartridge CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.

{3}------------------------------------------------

Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.

Stat Profile Prime Calibrator Cartridge CCS is intended for the calibration of pH, PCO2, PO2, Hct, Na*, K*, Cl , iCa, and Glucose, using the Stat Profile Prime CCS Analyzer.

Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.

Summary of the Technological Characteristics:

The Stat Profile Prime CCS Analyzer is substantially equivalent to the previously cleared for market Nova Stat Profile pHOx Ultra Analyzer System in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters. The Linearity Standard Set A for use with the Stat Profile Prime CCS Analyzer is identical to those cleared for use with the predicate Nova Stat Profile pHOx Ultra Analyzer System.

{4}------------------------------------------------

CharacteristicTable 1: Comparison of Predicate and Proposed devicesPredicate: K110648 Stat Profile pHOx Proposed: Stat Profile Prime CCS AnalyzerUltra Analyzer
Indication For UseThe STP pHOx Ultra Analyzer without CO-Oximeter is intended for in vitro diagnostic use by health care professionals and/or point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glu (Glucose), in heparinized whole blood.
PCO2, Whole blood measurement of blood gases is used in the diagnosis and treatment of life-threatening acid-base disturbances in critically ill patients with numerous metabolic and pulmonary diseases.
PO2,pH
Hct
Whole blood measurements of hematocrit are used to estimate that red blood cells are present in sufficient quantity to carry oxygen and carbon dioxide.
Na+
Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.
K+
Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.
Cl-
Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
iCa
Ionized Calcium measurement is used in the diagnosis and treatment of hypertension, renal disease, and vitamin D related disorders. Also useful in the diagnosis and treatment of patients with increased total protein and/or albumin levels, as in dehydration.
Glu
Glucose measurement is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Acceptable SamplesSodium or lithium heparinized wholeLithium heparinized whole blood from syringes, open tubes, small cups, and
CharacteristicPredicate: K110648 Stat Profile PHOxUltra AnalyzerProposed: Stat Profile Prime CCS Analyzer
blood, serum, or plasma samples fromsyringes, open tubes, small cups, andcapillary tubes.capillary tubes.
Sample Volumes100µL (syringe and capillary)100µL (syringe and capillary)
Measurement Range
pH6.500-8.000Same
PCO23.0 -200 mmHgSame
PO20-800 mmHg5-765 mmHg
Hct12%-70%Same
Na+80-200 mmol/LSame
K+1.0-20.0 mmol/LSame
Cl-50-200 mmol/LSame
Ca (Ca++)0.10-2.70 mmol/L0.20-2.70 mmol/L
Glu15-500 mg/dLSame
Principles ofMeasurement
pHHydrogen ion-selective sensorSame
PCO2Severinghaus-type sensorSame
PO2Polarographic Clark-type sensorSame
HctImpedance sensorSame
Na+Sodium ion-selective sensorSame
K+Potassium ion-selective sensorSame
Cl-Chloride ion-selective sensorSame
Ca (Ca++)Calcium ion-selective sensorSame
CharacteristicPredicate: K110648 Stat Profile pHOx Ultra AnalyzerProposed: Stat Profile Prime CCS Analyzer
GluGlucose Oxidase Enzymatic sensorSame
Touch Screen12.1" LCD, 1024x768 pixel, Resistive Touch5.7" VGA full color display with LED backlight and integrated touch panel
MenuFully configurable test menu based on above sensorsSame
Bar Code ScannerExternal (optional) 1DInternal Integrated 1D/2D
Printer2" Roll, Thermal TransferSame
PumpPeristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT)Same
Analog BoardPrecision low level analog front end w/ amperometric and potentiometric amplifiers, air detector circuitry and temperature control circuitrySame
CharacteristicPredicate: K110648 Stat Profile pHOx Ultra Calibrator CartridgeProposed: Stat Profile Prime Calibrator Cartridge CCS
Indication For UseThe intended use of the Nova STP pHOx Ultra Calibrator Cartridge is for the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin (Hb) in heparinized whole blood; Na+, K+, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.The Stat Profile Prime Calibrator Cartridge CCS is intended for the calibration of pH, PCO2, PO2, Hct, Na+, K+, Cl-, iCa, and Glucose using the Stat Profile Prime CCS Analyzer.
Configuration2 level calibration standards per analyte, and reference solutionSame
PackagingLiquid in Mylar bags inside cardboard container. Includes a waste collection bag. Self-contained, disposable packaging.Same
CharacteristicPredicate: K110648 Stat Profile pHOx Ultra/CCX ABG and CO-Oximeter ControlsProposed: Stat Profile Prime Auto QC Cartridge CCS
Indication For Use
Nova STP pHOx Ultra/CCX ABG and CO-Oximeter Controls andAutocartridge QC are intended for in vitro diagnostic use byhealthcare professionals for monitoring the performance of NovaSTP pHOx Ultra/CCX Analyzers.The Stat Profile Prime Auto QC Cartridge CCS is aquality control material intended for in vitro diagnosticuse by healthcare professionals for monitoring theperformance of the Stat Profile Prime CCS Analyzer.
Configuration3 level aqueous electrolyte, metabolite and gas solutions.Same
PackagingCartridge: Solution in Mylar bags inside cardboard container.Includes a waste collection bag. Self-contained, disposablepackaging.Same
CharacteristicPredicate: K110648 Stat Profile pHOx Ultra/CCX ABG and CO-Oximeter ControlsProposed: Stat Profile Prime Ampuled Control ABG/CCS
Indication For UseNova STP pHOx Ultra/CCX ABG and CO-Oximeter Controls and Autocartridge QC are intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Nova STP PHOx Ultra/CCX Analyzers.The Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.
Configuration3 level aqueous electrolyte, metabolite and gas solutions.Same
PackagingAmpules: Each glass ampule contains 1.7 ml volume.Same

Table 1: Comparison of Predicate and Proposed device

:

Page 5 of 9

:

{5}------------------------------------------------

:

. . . . . . .

.

Page 6 of 9

{6}------------------------------------------------

:

.

· ·

Page 7 of 9

{7}------------------------------------------------

:

CharacteristicPredicate: K110648 Stat Profile pHOx Ultra Linearity StandardsProposed: Linearity Standard Set A
Indication For UseFor in vitro diagnostic use with NOVA, Stat Profile/Ultra, Stat Profile pHOx Plus/L/C/M, Stat Profile pHOxUltra, Stat Profile Prime CCS, and Stat Profile CCX Analyzers to verify calibration, analytical linearity,estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.The Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzersto verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.
Configuration4 level aqueous solutions in glass ampules. Contain electrolyte, metabolite and gas solutions.Same
PackagingAmpules: Each glass ampule contains 1.8 ml volumeSame

Page 8 of 9

{8}------------------------------------------------

Summary of Performance Testing:

Bench testing was completed to demonstrate that the Stat Profile Prime CCS Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System.

The bench testing included:

  • . Method Comparison Studies
  • Precision/Reproducibility Studies .
  • Run to Run Precision .
  • Within Run Precision .
  • . Linearity Testing
  • Specificity / Interference Testing .
  • Detection Limit ●

The results of the testing confirmed that the performance of the Stat Profile Prime CCS Analyzer System is substantially equivalent to that of the Nova Stat Profile pHOx Ultra Analyzer System (predicate device).

Conclusion:

The results of software validation and performance verification testing confirmed that the Stat Profile Prime CCS Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime CCS Analyzer System is substantially equivalent to that of the Nova Stat Profile pHOx Ultra Analyzer System K110648 (predicate device).

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2014

NOVA BIOMEDICAL CORPORATION PAUL MACDONALD CHIEF OFFICER, QA & REG. AFFAIRS 200 PROSPECT ST WALTHAM MA 02454

Re: K131703

Trade/Device Name: Nova Stat Profile Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control ABG/CCS, Stat Profile Prime Ampuled Control ABG/CCS, Stat Profile Prime Calibrator Cartridge CCS/CCS Comp, Nova Linearity Standard Set A Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (PCO2, PO2) and blood pH test system Regulatory Class: II Product Code: CHL, JGS, CEM, JFP, CGZ, CGA, GKF, JIX, JJS Dated: February 19, 2014 Received: February 20, 2014

Dear Mr. MacDonald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (2) CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{10}------------------------------------------------

Page 2-Mr. MacDonald

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 80) and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known)

K131703

Device Name

Stat Profile® Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control ABG/CCS, Stat Profile Prime Calibrator Cartridge CCS/CCS Comp and Linearity Standard Set A

Indications for Use (Describe)

The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in cilnical faboratory settings for the quantitative determination of pH, PCO2, PO2, Hct, Na+, K+, Cl-, ICa, and Glu (Glucose), In heparinized whole blood.

PCO2, PO2, pH: Whole blood measurement of certain gases in whole blood, or pH of whole blood, is used in the diagnosis and treatment of life-threatening acid-base disturbances.

Hct: Whole blood measurements of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).

Na +: Sodium measurement is used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.

K+: Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

Cl-: Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

ICa: Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Glu: Glucose measurement is used in the diagnosis and treatment of carbohydrate metabollsm disturbances Including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The Stat Profile Prime Auto QC Cartridge CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.

The Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.

{12}------------------------------------------------

510(k) Number (if known)

K131703

Device Name

Stat Profile® Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control ABG/CCS, Stat Profile Prime Calibrator Cartridge CCS/CCS Comp and Linearity Standard Set A

The Stat Profile Prime Calibrator Cartridge CCS is intended for the callbration of pH, PCO2, PO2, Hct, Na+, K+, C-, ICa, and Glucose using the Stat Profile Prime CCS Analyzer.

Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from callbrator cartridge or analytical instrument variation.

Type of Use (Select one or both, as applicable)

8 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Yung W. Chan -S

{13}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.